§ 35.2643 - Records of periodic spot-checks for remote afterloader units.  


Latest version.
  • § 35.2643 Records of periodic spot-checks for remote afterloader units.

    (a) A licensee shall retain a record of each spot-check for remote afterloader units required by § 35.643 for 3 years.

    (b) The record must include, as applicable -

    (1) The date of the spot-check;

    (2) The manufacturer's name, model number, and serial number for the remote afterloader unit and source;

    (3) An assessment of timer accuracy;

    (4) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems, and clock and decayed source activity in the unit's computer; and

    (5) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.

    (c) A licensee shall retain a copy of the procedures required by § 35.643(b) until the licensee no longer possesses the remote afterloader unit.