Table B-1 to Subpart B of Part 53 - —Performance Limit Specifications for Automated Methods
Latest version.
-
Table B-1 to Subpart B of Part 53 - Performance Limit Specifications for Automated Methods
Performance
parameterUnits1 SO2 O3 CO NO2
(Std.
range)Definitions and test
proceduresStd. range3 Lower range 2 3 Std. range3 Lower range 2 3 Std. range3 Lower range 2 3 1. Range ppm 0-0.5 <0.5 0-0.5 <0.5 0-50 <50 0-0.5 Sec. 53.23(a) 2. Noise ppm 0.001 0.0005 0.0025 0.001 0.2 0.1 0.005 Sec. 53.23(b) 3. Lower detectable limit ppm 0.002 0.001 0.005 0.002 0.4 0.2 0.010 Sec. 53.23(c) 4. Interference equivalent Each interferent ppm ±0.005 4 ±0.005 ±0.005 ±0.005 ±1.0 ±0.5 ±0.02 Sec. 53.23(d) Total, all interferents ppm - - - - - - 0.04 Sec. 53.23(d) 5. Zero drift, 12 and 24 hour ppm ±0.004 ±0.002 ±0.004 ±0.002 ±0.5 ±0.3 ±0.02 Sec. 53.23(e) 6. Span drift, 24 hour 20% of upper range limit Percent - - - - - - ±20.0 Sec. 53.23(e) 80% of upper range limit Percent ±3.0 ±3.0 ±3.0 ±3.0 ±2.0 ±2.0 ±5.0 Sec. 53.23(e) 7. Lag time Minutes 2 2 2 2 2.0 2.0 20 Sec. 53.23(e) 8. Rise time Minutes 2 2 2 2 2.0 2.0 15 Sec. 53.23(e) 9. Fall time Minutes 2 2 2 2 2.0 2.0 15 Sec. 53.23(e) 10. Precision 20% of upper range limit - - - - - - Sec. 53.23(e) Percent5 2 2 2 2 1.0 1.0 4 Sec. 53.23(e) 80% of upper range limit - - - - - - Sec. 53.23(e) Percent5 2 2 2 2 1.0 1.0 6 Sec. 53.23(e)
Sec. 53.23(e)[80 FR 65460, Oct. 26, 2015]