§ 1065.945 - Specifications for CO analyzers.  


Latest version.
  • (a) Use a non-dispersive infrared (NDIR) detector that is compensated for CO2 and water interference.

    (b) The analyzer must have an accuracy and precision of ±2 percent of point or better under steady-state laboratory conditions.

    (c) The analyzer must reach at least 90 percent of its final response within 5.0 second after any step change to the input concentration greater than or equal 80 percent of full scale.

    (d) The analyzer must be zeroed and spanned daily during testing, and must be calibrated according to the analyzer manufacturer's specifications.