§ 89.321-96 - Oxides of nitrogen analyzer calibration.  


Latest version.
  • (a) The chemiluminescent oxides of nitrogen analyzer shall receive the initial and periodic calibration described in this section.

    (b) Prior to its introduction into service, and monthly thereafter, the chemiluminescent oxides of nitrogen analyzer is checked for NO2 to NO converter efficiency according to § 89.317-96.

    (c) Initial and periodic calibration. Prior to its introduction into service, and monthly thereafter, the chemiluminescent oxides of nitrogen analyzer shall be calibrated on all normally used instrument ranges. Use the same flow rate as when analyzing samples. Proceed as follows:

    (1) Adjust analyzer to optimize performance.

    (2) Zero the oxides of nitrogen analyzer with zero-grade air or zero-grade nitrogen.

    (3) Calibrate on each normally used operating range with NO-in-N2 calibration gases with nominal concentrations starting at between 10 and 15 percent and increasing in at least six incremental steps to 90 percent of that range. The incremental steps are to be spaced to represent good engineering practice. For each range calibrated, if the deviation from a least-squares best-fit straight line is 2 percent or less of the value at each data point, concentration values may be calculated by use of a single calibration factor for that range. If the deviation exceeds 2 percent at any point, the best-fit non-linear equation which represents the data to within 2 percent of each test point shall be used to determine concentration.

    (d) The initial and periodic interference, system check, and calibration test procedures specified in part 86, subpart D of this chapter may be used in lieu of the procedures specified in this section.