§ 89.322-96 - Carbon dioxide analyzer calibration.  


Latest version.
  • (a) Prior to its introduction into service, and monthly thereafter, the NDIR carbon dioxide analyzer shall be calibrated as follows:

    (1) Follow good engineering practices for instrument start-up and operation. Adjust the analyzer to optimize performance.

    (2) Zero the carbon dioxide analyzer with either zero-grade air or zero-grade nitrogen.

    (3) Calibrate on each normally used operating range with carbon dioxide-in-N2 calibration or span gases having nominal concentrations starting between 10 and 15 percent and increasing in at least six incremental steps to 90 percent of that range. The incremental steps are to be spaced to represent good engineering practice. For each range calibrated, if the deviation from a least-squares best-fit straight line is 2 percent or less of the value at each data point, concentration values may be calculated by use of a single calibration factor for that range. If the deviation exceeds 2 percent at any point, the best-fit non-linear equation which represents the data to within 2 percent of each test point shall be used to determine concentration.

    (b) The initial and periodic interference, system check, and calibration test procedures specified in part 86, subpart D of this chapter may be used in lieu of the procedures in this section.