§ 84.177 - Total noise level test - PAPR classes HE and PAPR100.
Latest version.
-
§ 84.177 Inhalation and exhalation valves; minimum requirements.
(a) Inhalation and exhalation valves shall be protected against distortion.
(b) Inhalation valves shall be designed and constructed and provided where necessary to prevent excessive exhaled air from adversely affecting filters, except where filters are specifically designed to resist moisture.
(c) Exhalation valves shall be:
(1) Provided where necessary;
(3) Designed and constructed to prevent inward leakage of contaminated air(2) Protected against damage and external influence; and
Total noise level test - PAPR classes HE and PAPR100.
Noise levels generated by any powered air-purifying respirators that cover the ears (i.e., hood or helmet) will be measured at the entrance to each ear at maximum airflow obtainable and must not exceed 80 dBA.