E7-12789. Medical Devices; General Hospital and Personal Use Devices; Classification of the Filtering Facepiece Respirator for Use by the General Public in Public Health Medical Emergencies  

  • Start Preamble

    AGENCY:

    Food and Drug Administration, HHS.

    ACTION:

    Final rule.

    SUMMARY:

    The Food and Drug Administration (FDA) is classifying the filtering facepiece respirator for use by the general public in public health medical emergencies into class II (special controls). The agency is classifying these devices into class II (special controls) in order to provide a reasonable assurance of the safety and effectiveness of these devices and is specifying what those special controls are.

    Elsewhere in this issue of the Federal Register, FDA is announcing the availability of a guidance document entitled, “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Filtering Facepiece Respirator for Use by the General Public in Public Health Medical Emergencies.” This guidance document will serve as one of the special controls, along with certification of the respirator by the National Institute for Occupational Safety and Health (NIOSH) in accordance with its regulations for non-powered air-purifying particulate respirators, found in 42 CFR part 84, as specified in the classification regulation.

    DATES:

    This rule is effective August 2, 2007. The classification was effective May 8, 2007.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    Sheila A. Murphey, Center for Devices and Radiological Health (HFZ-480), Food and Drug Administration, 9200 Corporate Blvd., Rockville, MD 20850, 240-276-3700.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    I. What is the Background of this Rulemaking?

    In accordance with section 513(f)(1) of the Federal Food, Drug, and Cosmetic Act (the act) (21 U.S.C. 360c(f)(1)), devices that were not in commercial distribution before May 28, 1976, the date of enactment of the Medical Device Amendments of 1976 (the amendments), generally referred to as postamendments devices, are classified automatically by statute into class III without any FDA rulemaking process. These devices remain in class III and require premarket approval, unless the device is classified or reclassified into class I or class II, or FDA issues an order finding the device to be substantially equivalent, in accordance with section 513(i) of the act, to a predicate device that does not require premarket approval. The agency determines whether new devices are substantially equivalent to predicate devices by means of premarket notification procedures in section 510(k) of the act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807) of FDA's regulations.

    Section 513(f)(2) of the act provides that any person who submits a premarket notification under section 510(k) of the act for a device that has not previously been classified may, within 30 days after receiving an order classifying the device in class III under section 513(f)(1) of the act, request FDA to classify the device under the criteria set forth in section 513(a)(1) of the act. FDA shall, within 60 days of receiving such a request, classify the device by written order. This classification shall be the initial classification of the device type. Within 30 days after the issuance of an order classifying the device, FDA must publish a notice in the Federal Register announcing such classification (section 513(f)(2) of the act).

    In accordance with section 513(f)(1) of the act, FDA issued an order on August 30, 2006, classifying the two 3M filtering facepiece respirators intended for use by the general public in public health medical emergencies (designated at that time as the 3MTM N95 Home Respirator with Fluid Resistance and 3MTM N95 Home Respirator) in class III, because each device was not substantially equivalent to a device that was introduced or delivered for introduction into interstate commerce for commercial distribution before May 28, 1976, or a device that was subsequently reclassified into class I or class II. On October 3, 2006, 3M Inc. submitted a petition requesting initial classification of these devices under section 513(f) (2) of the act. The manufacturer recommended that the Start Printed Page 36361devices be classified into class II (Ref. 1). In response to FDA requests for additional information, 3M supplemented its petition on March 22, 2007.

    In accordance with section 513(f) (2) of the act, FDA reviewed the petition in order to classify the devices under the criteria for classification set forth in 513(a)(1) of the act. Devices are to be classified into class II if general controls, by themselves, are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls to provide reasonable assurance of the safety and effectiveness of the device for its intended use. After review of the information submitted in the petition and its supplements, FDA determined that the 3MTM filtering facepiece respirator devices, now known as the 3MTM 8612F Respirator for Use by the General Public in Public Health Medical Emergencies and 3MTM 8670F Respirator for Use by the General Public in Public Health Medical Emergencies can be classified into class II with the establishment of special controls. FDA believes that special controls, in addition to general controls, are adequate to provide reasonable assurance of the safety and effectiveness of the device and that there is sufficient information to establish special controls to provide such assurance.

    The device is assigned the generic name “Filtering Facepiece Respirator for use by the General Public in Public Health Medical Emergencies” and is identified as a device that is a disposable half-facepiece non-powered air-purifying particulate respirator intended for use to cover the nose and mouth of the wearer to help reduce wearer exposure to pathogenic biological airborne particulates during a public health medical emergency. The device is made of polymeric materials and is intended to fit closely to the face and to function by filtering particulate material.

    FDA believes that special controls are needed to help address the following issues affecting the safety and effectiveness of the filtering facepiece respirator for use by the general public in public health medical emergencies.

    A. Assuring Filtration and Breathability

    For this type of respirator to reduce wearer exposure to pathogenic biological airborne particulates, it must be made of filter material that is highly efficient in filtering such particles. At the same time, because this type of device depends on the wearer's normal respiration to draw ambient air through the respirator materials and into the lungs, the respirator material must also permit adequate respiration.

    B. Assuring Proper Fit

    The device must fit closely to the wearer's face without any gaps that would allow air to reach the wearer's respiratory tract without passing through the filter material. Otherwise, improper fit of the respirator could result in inhalation of pathogenic biological airborne particulates carried in air that passes around the sides of the device.

    C. Avoiding Adverse Skin Reaction

    Reducing wearer exposure to pathogenic biological airborne particulates requires that the device be properly fitted to the face. If the respirator material in contact with the skin is not biocompatible, it may cause adverse reactions such as redness, pruritus, and skin irritation.

    D. Assuring Proper Use

    While a filtering facepiece respirator for use by the general public in public health medical emergencies can help to reduce wearer exposure to pathogenic biological airborne particulates in a public health medical emergency where there is a serious risk from such exposure, these devices do not provide complete protection against infection. Even when used correctly and consistently, a filtering facepiece respirator does not eliminate all respiratory exposure, and for many pathogens that may be transmitted through airborne particulates, transmission via other routes is also possible. (Because filtering facepiece respirators for use by the general public in public health medical emergencies have not been tested against specific microorganisms, the extent of protection to be expected against specific pathogens is not known and would vary with particular conditions in any event.)

    The respirator should always be used in conjunction with other infection control and respiratory protection measures. In addition, because the outside of the respirator may be contaminated with infectious materials during normal use, proper handling and disposal is important to avoid the respirator itself becoming a vector of transmission of infectious agents.

    Further, failure of the user to assure proper fit of the respirator could result in exposure to pathogenic biological airborne particles. Certain populations such as children will be unlikely to achieve a proper fit because respirators are designed and sized for adults.

    For users with certain underlying cardiac, pulmonary or related medical conditions, achieving the fit necessary to help reduce their exposure to pathogenic biological airborne particulates may exacerbate their underlying medical conditions raising a concern about their safe use for these populations.

    Finally, these respirators have not been established to be safe or effective if reused, and use of a single respirator by multiple users may result in the respirator itself becoming a vector of transmission.

    To address these issues, the class II special controls guidance document provides recommendations for labeling and for information to be provided to meet premarket notification (510(k)) submission requirements for the device, including recommendations for fit testing and biocompatibility testing. In addition, this classification regulation specifies another special control, certification of the respirator by NIOSH as a non-powered air-purifying particulate respirator with a minimum filtration efficiency classification of N95, in accordance with 42 CFR part 84. The respirators that were the subject of the initial classification order described in this document and that are the initial legal predicate devices for this new device classification were certified by NIOSH under these requirements, as revised as of October 1, 2006. FDA's determination that NIOSH certification is an appropriate special control to help assure the safety and effectiveness of the respirator for its intended use under this classification rests on the assurance of filtration efficiency and breathability provided by NIOSH certification under these requirements, as effective on May 8, 2007, the date of FDA's classification order. Should NIOSH revise the requirements for certification in the future, FDA will evaluate whether certification under such revised NIOSH regulations is an appropriate special control for devices within this classification and may revise FDA's regulation using appropriate procedures.

    FDA believes that these special controls, designated in this rule, in addition to general controls, address the issues identified previously and provide reasonable assurance of the safety and effectiveness of the device type. Thus, on May 8, 2007, FDA issued an order to the petitioner classifying the device type into class II. FDA is codifying this classification at 21 CFR 880.6260.

    Following the effective date of the final classification rule, manufacturers will need to demonstrate NIOSH certification of any filtering facepiece respirator for use by the general public in public health medical emergencies, Start Printed Page 36362as set forth in 21 CFR 880.6260(b)(1), and address the issues covered in the special controls guidance. With respect to the issues addressed only in the special control guidance, however, the manufacturer need only show that its device meets the recommendations of the guidance or in some other way provides equivalent assurance of safety and effectiveness.

    Section 510(m) of the act provides that FDA may exempt a class II device from the premarket notification requirements under section 510(k) of the act, if FDA determines that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device. For filtering facepiece respirators for use by the general public in public health medical emergencies, however, FDA has determined that premarket review of fit performance data, labeling, and other requirements as outlined in 21 CFR 807.87, is necessary to provide reasonable assurance that acceptable levels of performance for both safety and effectiveness will be addressed before marketing clearance. Thus, persons who intend to market this type of device must submit to FDA a premarket notification, prior to marketing the device type.

    II. What is the Environmental Impact of This Rule?

    The agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Thus, neither an environmental assessment nor an environmental impact statement is required.

    III. What is the Economic Impact of This Rule?

    FDA has examined the impacts of the final rule under Executive Order 12866 and the Regulatory Flexibility Act (5 U.S.C. 601-612), and the Unfunded Mandates Reform Act of 1995 (Public Law 104-4). Executive Order 12866 directs agencies to assess all costs and benefits of available regulatory alternatives and, when regulation is necessary, to select regulatory approaches that maximize net benefits (including potential economic, environmental, public health and safety, and other advantages; distributive impacts; and equity). The agency believes that this final rule is not a significant regulatory action as defined by the Executive order.

    The Regulatory Flexibility Act requires agencies to analyze regulatory options that would minimize any significant impact of a rule on small entities. Because classification of this device into class II will relieve manufacturers of the cost of complying with the premarket approval requirements of section 515 of the act (21 U.S.C. 360e), and may permit small potential competitors to enter the marketplace by lowering their costs, the agency certifies that the final rule will not have a significant economic impact on a substantial number of small entities.

    Section 202(a) of the Unfunded Mandates Reform Act of 1995 requires that agencies prepare a written statement, which includes an assessment of anticipated costs and benefits, before proposing “any rule that includes any Federal mandate that may result in the expenditure by State, local, and tribal governments, in the aggregate, or by the private sector, of $100,000,000 or more (adjusted annually for inflation) in any one year.” The current threshold after adjustment for inflation is $122 million, using the most current (2005) Implicit Price Deflator for the Gross Domestic Product. FDA does not expect this final rule to result in any 1-year expenditure that would meet or exceed this amount.

    IV. Does This Final Rule Have Federalism Implications?

    FDA has analyzed this final rule in accordance with the principles set forth in Executive Order 13132. FDA has determined that the rule does not contain policies that have substantial direct effects on the States, on the relationship between the National Government and the States, or on the distribution of power and responsibilities among the various levels of government. Accordingly, the agency has concluded that the rule does not contain policies that have federalism implications as defined in the Executive order and, consequently, a federalism summary impact statement is not required.

    V. How Does This Rule Comply with the Paperwork Reduction Act of 1995?

    This final rule contains no new information collection provisions but refers to NIOSH regulations in 42 CFR part 84 that contain information collection provisions that have been reviewed and approved by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520), under OMB Control No. 0920-0109. Based on information from NIOSH regarding submissions for respirator certification received in the past 3 years, FDA concludes that specification of NIOSH certification as a special control will not result in the collection of any additional information by NIOSH not already covered by NIOSH's burden estimates. This final rule also designates a guidance document as a special control. Elsewhere in this issue of the Federal Register, FDA is publishing a notice announcing the availability of that guidance document, “Class II Special Controls Guidance Document: Filtering Facepiece Respirators for Use by the General Public in Public Health Medical Emergencies,” which contains a Paperwork Reduction Act analysis for that guidance.

    VI. What References are on Display?

    The following references have been placed on display in the Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852, and may be seen by interested persons between 9 a.m. and 4 p.m., Monday through Friday.

    1. Petition from 3M.

    2. 42 CFR part 84, as revised as of October 1, 2006.

    Start List of Subjects

    List of Subjects in 21 CFR part 880

    • Medical devices
    End List of Subjects Start Amendment Part

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 880 is amended as follows:

    End Amendment Part Start Part

    PART 880—GENERAL HOSPITAL AND PERSONAL USE DEVICES

    End Part Start Amendment Part

    1. The authority citation for 21 CFR part 880 continues to read as follows:

    End Amendment Part Start Authority

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

    End Authority Start Amendment Part

    2. Section 880.6260 is added to subpart G to read as follows:

    End Amendment Part
    Filtering facepiece respirator for use by the general public in public health medical emergencies.

    (a) Identification. A filtering facepiece respirator for use by the general public in public health medical emergencies is a device that is a disposable half-facepiece non-powered air-purifying particulate respirator intended for use to cover the nose and mouth of the wearer to help reduce wearer exposure to pathogenic biological airborne particulates during a public health medical emergency. The device is made of polymeric materials and is intended to fit closely to the face and to function by filtering particulate material.

    (b) Classification. Class II (special controls). The special controls are:

    (1) Certification by the National Institute for Occupational Safety and Health (NIOSH) as a non-powered air-Start Printed Page 36363purifying particulate respirator with a minimum filtration efficiency classification of N95, in accordance with 42 CFR part 84.

    (2) The FDA guidance document entitled: “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Filtering Facepiece Respirator for use by the General Public in Public Health Medical Emergencies.” See § 880.1(e) for information on obtaining a copy of this guidance document.

    Start Signature

    Dated: June 22, 2007.

    Linda S. Kahan,

    Deputy Director, Center for Devices and Radiological Health.

    End Signature End Supplemental Information

    [FR Doc. E7-12789 Filed 7-2-07; 8:45 am]

    BILLING CODE 4160-01-S

Document Information

Effective Date:
8/2/2007
Published:
07/03/2007
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final rule.
Document Number:
E7-12789
Dates:
This rule is effective August 2, 2007. The classification was effective May 8, 2007.
Pages:
36360-36363 (4 pages)
Docket Numbers:
Docket No. 2007N-0198
PDF File:
e7-12789.pdf
CFR: (1)
21 CFR 880.6260