2011-18391. Cardiovascular Devices; Classification of Electrocardiograph Electrodes  

  • Start Preamble

    AGENCY:

    Food and Drug Administration, HHS.

    ACTION:

    Final rule.

    SUMMARY:

    The Food and Drug Administration (FDA) is classifying the electrocardiograph electrode, intended to acquire and transmit the electrical signal at the body surface to a processor that produces an electrocardiogram (ECG) or vectorcardiogram, into class II (special controls). FDA is also exempting this device from the premarket notification requirement.

    DATES:

    This rule is effective August 22, 2011.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    Sharon Lappalainen, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, rm. 1238, Silver Spring, MD 20993, 301-796-6322.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    I. Regulatory Authorities

    The Federal Food, Drug, and Cosmetic Act (the FD&C Act) (21 U.S.C. 301 et seq.), as amended by the Medical Device Amendments of 1976 (the 1976 amendments) (Pub. L. 94-295), the Safe Medical Devices Act of 1990 (SMDA) (Pub. L. 101-629), the Food and Drug Administration Modernization Act of 1997 (FDAMA) (Pub. L. 105-115), and the Medical Device User Fee and Modernization Act of 2002 (MDUFMA) (Pub. L. 107-250), established a comprehensive system for the regulation Start Printed Page 43583of medical devices intended for human use. Section 513 of the FD&C Act (21 U.S.C. 360c) established three categories (classes) of devices, defined by the regulatory controls needed to provide reasonable assurance of their safety and effectiveness. The three categories of devices are class I (general controls), class II (special controls), and class III (premarket approval).

    Under section 513 of the FD&C Act, FDA refers to devices that were in commercial distribution before May 28, 1976 (the date of enactment of the 1976 amendments), as “preamendments devices.” FDA classifies these devices after it takes the following steps: (1) Receives a recommendation from a device classification panel (an FDA advisory committee); (2) publishes the panel's recommendation for comment, along with a proposed regulation classifying the device; and (3) publishes a final regulation classifying the device. FDA has classified most preamendments devices under these procedures.

    Devices that were not in commercial distribution before May 28, 1976, generally referred to as postamendments devices, are classified automatically under section 513(f) of the FD&C Act into class III without any FDA rulemaking process. Those devices remain in class III until FDA does the following: (1) Reclassifies the device into class I or II; (2) issues an order classifying the device into class I or II in accordance with section 513(f)(2) of the FD&C Act; or (3) issues an order finding the device to be substantially equivalent, in accordance with section 513(i) of the FD&C Act, to a legally marketed device that has been classified into class I or class II.

    The Agency determines whether new devices are substantially equivalent to previously marketed devices by means of premarket notification procedures in section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and 21 CFR part 807 of the regulations. FDAMA added a new section 510(m) to the FD&C Act. New section 510(m) provides that a class II device may be exempted from the premarket notification requirements under section 510(k) of the FD&C Act, if the Agency determines that premarket notification is not necessary to assure the safety and effectiveness of the device. FDA has determined that premarket notification is not necessary to assure the safety and effectiveness of electrocardiograph electrodes.

    Under the 1976 amendments, class II devices were defined as devices for which there was insufficient information to show that general controls themselves would provide reasonable assurance of safety and effectiveness, but for which there was sufficient information to establish performance standards to provide such assurance. SMDA broadened the definition of class II devices to mean those devices for which the general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but for which there is sufficient information to establish special controls to provide such assurance, including performance standards, postmarket surveillance, patient registries, development and dissemination of guidelines, recommendations, and any other appropriate actions FDA deems necessary (section 513(a)(1)(B) of the FD&C Act). Elsewhere in this issue of the Federal Register, FDA is announcing the availability of the guidance document that will serve as the special control for this device. This action is being taken under the FD&C Act, as amended by the 1976 amendments, SMDA, FDAMA, and MDUFMA.

    II. Regulatory History of the Device

    In the Federal Register of October 4, 2007 (72 FR 56702; Docket No. 2007N-0308), FDA proposed to classify electrocardiograph electrodes, intended to acquire and transmit the electrical signal at the body surface to a processor that produces an electrocardiogram (ECG) or vectorcardiogram, into class II. FDA also proposed to exempt this device from premarket notification requirements. FDA invited interested persons to comment on the proposed regulation by January 2, 2008. FDA received seven comments on the proposed rule.

    III. Summary of Final Rule

    FDA is amending the classification regulation for electrocardiograph electrodes into class II, by making this device exempt from 510(k) premarket notification requirements and subject to the new special controls described in the special controls guidance document entitled “Guidance for Industry and Food and Drug Administration Staff: Class II Special Controls Guidance Document: Electrocardiograph Electrodes.”

    As described in that special controls guidance document, the special controls include the following:

    • Documentation of device description, which includes compliance with 21 CFR 820.181(a) to maintain a device master record;
    • Documentation of performance characteristics, which includes documentation on biocompatibility, electrical performance, adhesive performance, shelf life, reuse, electrodes intended for use in specified procedures, sterility, and, with respect to electrode lead wires and patient cables, compliance with 21 CFR part 898; and
    • Specific labeling, including indications for use, cautions, precautions, and adverse reactions.

    Manufacturers must comply with the special controls as identified in that special controls guidance document, either by meeting the recommendations in the guidance document or by some other means that provides equivalent assurance of safety and effectiveness.

    IV. Discussion of Comments

    The public comments received in response to the proposed rule addressed issues pertaining to labeling, the scope of the devices subject to the classification rule, and testing as follows:

    Regarding labeling, the comments requested the mandatory product labeling of all adhesive coated disposable electrocardiograph electrodes and the establishment of template labeling with which electrocardiograph electrodes should comply. In response, FDA has revised the labeling section of the special controls guidance document; however, FDA has not established template labeling.

    Regarding the scope of the devices subject to the notice, the comments requested a products-based definition or listing of examples to flesh out the “type” of devices that are consistent with Agency intent, requested a clarification of the type of sensor that is included in the scope of the classification, and requested a clarification if the list of environmental conditions is intended to be an exclusive list. In response, FDA has revised the special controls guidance document to clarify what the scope of this classification rule includes and excludes and to clarify what labeling should be reported regarding conditions of use.

    Regarding testing, the comments requested clarification of the shelf life, storage condition testing, and human clinical testing required to establish sensitivity and irritation for all adhesives. In response, FDA has revised the special controls guidance document to clarify testing for shelf life and has clarified the testing for biocompatibility.

    FDA is adopting in final form the assessment of the risks to public health stated in the proposed rule published on October 4, 2007, and agrees that the risk of electrical shock should also be taken Start Printed Page 43584into account for purposes of this assessment. The guidance document has been revised accordingly. FDA is issuing this final rule which classifies the generic type of device, electrocardiograph electrode, into class II subject to special controls.

    V. Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

    VI. Analysis of Impacts

    FDA has examined the impacts of the final rule under Executive Order 12866, Executive Order 13563, the Regulatory Flexibility Act (5 U.S.C. 601-612), and the Unfunded Mandates Reform Act of 1995 (Pub. L. 104-4). Executive Orders 12866 and 13563 direct Agencies to assess all costs and benefits of available regulatory alternatives and, when regulation is necessary, to select regulatory approaches that maximize net benefits (including potential economic, environmental, public health and safety, and other advantages; distributive impacts; and equity). The Agency believes that this final rule is not a significant regulatory action under the Executive Order 12866.

    The Regulatory Flexibility Act requires Agencies to analyze regulatory options that would minimize any significant impact of a rule on small entities. Because manufacturers are already substantially in compliance with the recommendations in the guidance document and exemption from the premarket notification requirements for the devices following the specific measures recommended in the special control will simplify the entry to market for other manufacturers, the Agency certifies that the final rule will not have a significant economic impact on a substantial number of small entities. In addition, this final rule will not impose costs of $100 million or more on either the private sector or state, local, and tribal governments in the aggregate, and therefore a written statement under section 202(a) of the Unfunded Mandates Reform Act of 1995 is not required.

    Section 202(a) of the Unfunded Mandates Reform Act of 1995 requires that Agencies prepare a written statement, which includes an assessment of anticipated costs and benefits, before proposing “any rule that includes any Federal mandate that may result in the expenditure by State, local, and tribal governments, in the aggregate, or by the private sector, of $100,000,000 or more (adjusted annually for inflation) in any one year.” The current threshold after adjustment for inflation is $136 million, using the most current (2010) Implicit Price Deflator for the Gross Domestic Product. FDA does not expect this final rule to result in any 1-year expenditure that would meet or exceed this amount.

    The specific measures recommended largely reflect current practices. With most manufacturers complying with most of the recommendations in the guidance document, any additional burden brought about by the final rule and guidance will likely be small.

    VII. Federalism

    FDA has analyzed this final rule in accordance with the principles set forth in Executive Order 13132. Section 4(a) of the Executive order requires Agencies to “construe* * * a Federal statute to preempt State law only where the statute contains an express preemption provision or there is some other clear evidence that the Congress intended preemption of State law, or where the exercise of State authority conflicts with the exercise of Federal authority under the Federal statute.” Federal law includes an express preemption provision that preempts certain state requirements “different from or in addition to” certain Federal requirements applicable to devices. 21 U.S.C. 360k. See Medtronic v. Lohr, 518 U.S. 470 (1996); and Riegel v. Medtronic, 552 U.S. 312 (2008). The special controls established by this final rule create “requirements” to address each identified risk to health presented by these specific medical devices under 21 U.S.C. 360k, even though product sponsors may have flexibility in how they meet those requirements Cf. Papike v. Tambrands, Inc., 107 F.3d 737, 740-42 (9th Cir. 1997).

    VIII. Paperwork Reduction Act of 1995

    FDA concludes that this final rule contains no new collections of information. Therefore, clearance by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (the PRA) (44 U.S.C. 3501-3520) is not required.

    This final rule designates a guidance document as a special control. FDA also concludes that the special control guidance document does not contain new information collection provisions that are subject to review and clearance by OMB under the PRA.

    IX. Electronic Access

    For access to the docket to read references or the public comments received, go to http://www.regulations.gov or go to the Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852. A hard copy of any document can be obtained by submitting a Freedom of Information Act request to the Division of Freedom of Information (ELEM-1029), Food and Drug Administration, 12420 Parklawn Dr., Element Bldg., Rockville, MD 20857.

    X. References

    The following references have been placed on display in the Division of Dockets Management (see section IX of this document) and may be seen by interested persons between 9 a.m. and 4 p.m., Monday through Friday, or can be obtained in hardcopy by submitting a Freedom of Information Act request to the Division of Freedom of Information (see section IX of this document). (FDA has verified the Web site address, but we are not responsible for any subsequent changes to the Web site after this document publishes in the Federal Register.)

    1. “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.”

    2. ANSI/AAMI/ISO 10993-1, “Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing.”

    3. ISO 10993-5, “Biological Evaluation of Medical Devices—Part 5: Tests for In Vitro Cytotoxicity.”

    4. ISO 10993-10, “Biological Evaluation of Medical Devices—Part 10: Tests for Irritation and Delayed Type Hypersensitivity.”

    5. ANSI/AAMI EC12, “Disposable ECG Electrodes.”

    6. AAMI/ANSI/ISO ST79, “Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities.”

    7. “Updated 520(k) Sterility Review Guidance K90-1.”

    8. IEC 601-1.56.3(c), “Medical Electrical Equipment—Part 1, General Requirements for Safety.”

    9. ANSI/AAMI EC53, “ECG Cables and Leadwires.”

    10. FDA Guidance 89-4203, “Labeling—Regulatory Requirements for Medical Devices,” available at http://www.fda.gov/​cdrh/​dsma/​470.pdf.

    Start List of Subjects

    List of Subjects in 21 CFR Part 870

    • Medical devices
    End List of Subjects

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 870 is amended as follows:

    Start Part Start Printed Page 43585

    PART 870—CARDIOVASCULAR DEVICES

    End Part Start Amendment Part

    1. The authority citation for 21 CFR part 870 continues to read as follows:

    End Amendment Part Start Authority

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

    End Authority Start Amendment Part

    2. Revise paragraph (b) in § 870.2360 to read as follows:

    End Amendment Part
    Electrocardiograph electrode.
    * * * * *

    (b) Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.

    Start Signature

    Dated: July 18, 2011.

    Nancy K. Stade,

    Deputy Director for Policy, Center for Devices and Radiological Health.

    End Signature End Supplemental Information

    [FR Doc. 2011-18391 Filed 7-20-11; 8:45 am]

    BILLING CODE 4160-01-P

Document Information

Comments Received:
0 Comments
Effective Date:
8/22/2011
Published:
07/21/2011
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final rule.
Document Number:
2011-18391
Dates:
This rule is effective August 22, 2011.
Pages:
43582-43585 (4 pages)
Docket Numbers:
Docket No. FDA-2007-N-0092
Topics:
Medical devices
PDF File:
2011-18391.pdf
CFR: (1)
21 CFR 870.2360