2017-15901. Medical Devices; Cardiovascular Devices; Classification of the Adjunctive Cardiovascular Status Indicator  

  • Start Preamble

    AGENCY:

    Food and Drug Administration, HHS.

    ACTION:

    Final order.

    SUMMARY:

    The Food and Drug Administration (FDA) is classifying the adjunctive cardiovascular status indicator into class II (special controls). The special controls that will apply to the device are identified in this order and will be part of the codified language for the adjunctive cardiovascular status indicator's classification. The Agency is classifying the device into class II (special controls) in order to provide a reasonable assurance of safety and effectiveness of the device.

    DATES:

    This order is effective July 28, 2017. The classification was applicable on December 21, 2016.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    Nathalie Yarkony, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 1254, Silver Spring, MD 20993-0002, 301-796-1235, nathalie.yarkony@fda.hhs.gov.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    I. Background

    In accordance with section 513(f)(1) of the Federal Food, Drug, and Cosmetic Act (the FD&C Act) (21 U.S.C. 360c(f)(1)), devices that were not in commercial distribution before May 28, 1976 (the date of enactment of the Medical Device Amendments of 1976), generally referred to as postamendments devices, are classified automatically by statute into class III without any FDA rulemaking process. These devices remain in class III and require premarket approval unless and until the device is classified or reclassified into class I or II, or FDA issues an order finding the device to be substantially equivalent, in accordance with section 513(i) of the FD&C Act, to a predicate device that does not require premarket approval. The Agency determines whether new devices are substantially equivalent to predicate devices by means of premarket notification procedures in section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807) of the regulations.Start Printed Page 35066

    Section 513(f)(2) of the FD&C Act, also known as De Novo classification, as amended by section 607 of the Food and Drug Administration Safety and Innovation Act (Pub. L. 112-144), provides two procedures by which a person may request FDA to classify a device under the criteria set forth in section 513(a)(1). Under the first procedure, the person submits a premarket notification under section 510(k) of the FD&C Act for a device that has not previously been classified and, within 30 days of receiving an order classifying the device into class III under section 513(f)(1) of the FD&C Act, the person requests a classification under section 513(f)(2). Under the second procedure, rather than first submitting a premarket notification under section 510(k) of the FD&C Act and then a request for classification under the first procedure, the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence and requests a classification under section 513(f)(2) of the FD&C Act. If the person submits a request to classify the device under this second procedure, FDA may decline to undertake the classification request if FDA identifies a legally marketed device that could provide a reasonable basis for review of substantial equivalence with the device or if FDA determines that the device submitted is not of “low-moderate risk” or that general controls would be inadequate to control the risks and special controls to mitigate the risks cannot be developed.

    In response to a request to classify a device under either procedure provided by section 513(f)(2) of the FD&C Act, FDA shall classify the device by written order within 120 days. This classification will be the initial classification of the device.

    On May 24, 2016, Flashback Technologies submitted a request for classification of the CipherOx CRI Tablet under section 513(f)(2) of the FD&C Act.

    In accordance with section 513(f)(2) of the FD&C Act, FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1). FDA classifies devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls to provide reasonable assurance of the safety and effectiveness of the device for its intended use. After review of the information submitted in the request, FDA determined that the device can be classified into class II with the establishment of special controls. FDA believes these special controls, in addition to general controls, will provide reasonable assurance of the safety and effectiveness of the device.

    Therefore, on December 21, 2016, FDA issued an order to the requestor classifying the device into class II. FDA is codifying the classification of the device by adding § 870.2200.

    Following the effective date of this final classification order, any firm submitting a premarket notification (510(k)) for an adjunctive cardiovascular status indicator will need to comply with the special controls named in this final order. A De Novo classification decreases regulatory burdens. When FDA classifies a device type as class I or II via the De Novo pathway, other manufacturers do not have to submit a De Novo request or premarket approval application in order to market the same type of device, unless the device has a new intended use or technological characteristics that raise different questions of safety or effectiveness. Instead, manufacturers can use the less burdensome 510(k) pathway, when necessary, to market their device, and the device that was the subject of the original De Novo classification can serve as a predicate device for additional 510(k)s from other manufacturers.

    The device is assigned the generic name adjunctive cardiovascular status indicator, and it is identified as a prescription device based on sensor technology for the measurement of a physical parameter(s). This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.

    FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1:

    Table 1—Adjunctive Cardiovascular Status Indicator Risks and Mitigation Measures

    Identified riskMitigation measures
    Delayed or incorrect treatment due to erroneous output as a result of software malfunction or algorithm errorSoftware verification, validation, and hazard analysis. Non-clinical performance testing. Clinical performance testing. Labeling.
    Delayed or incorrect treatment due to user misinterpretationUsability assessment. Labeling.

    FDA believes that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of the safety and effectiveness.

    Adjunctive cardiovascular status indicators are not safe for use except under the supervision of a practitioner licensed by law to direct the use of the device. As such, the device is a prescription device and must satisfy prescription labeling requirements (see 21 CFR 801.109 Prescription devices).

    Section 510(m) of the FD&C Act provides that FDA may exempt a class II device from the premarket notification requirements under section 510(k), if FDA determines that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device. For this type of device, FDA believes premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device type and, therefore, is planning to exempt the device from the premarket notification requirements under section 510(m) of the FD&C Act. Once finalized, persons who intend to market this device type need not submit a 510(k) premarket notification containing information on the adjunctive cardiovascular status indicator prior to marketing the device.

    II. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

    III. Paperwork Reduction Act of 1995

    This final administrative order establishes special controls that refer to previously approved collections of Start Printed Page 35067information found in other FDA regulations. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections of information in part 807, subpart E, regarding premarket notification submissions have been approved under OMB control number 0910-0120, and the collections of information in 21 CFR part 801, regarding labeling have been approved under OMB control number 0910-0485.

    Start List of Subjects

    List of Subjects in 21 CFR Part 870

    • Medical devices
    End List of Subjects

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 870 is amended as follows:

    Start Part

    PART 870—CARDIOVASCULAR DEVICES

    End Part Start Amendment Part

    1. The authority citation for part 870 continues to read as follows:

    End Amendment Part Start Authority

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360 l, 371.

    End Authority Start Amendment Part

    2. Add § 870.2200 to subpart C to read as follows:

    End Amendment Part
    Adjunctive cardiovascular status indicator.

    (a) Identification. The adjunctive cardiovascular status indicator is a prescription device based on sensor technology for the measurement of a physical parameter(s). This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.

    (b) Classification. Class II (special controls). The special controls for this device are:

    (1) Software description, verification, and validation based on comprehensive hazard analysis must be provided, including:

    (i) Full characterization of technical parameters of the software, including any proprietary algorithm(s);

    (ii) Description of the expected impact of all applicable sensor acquisition hardware characteristics on performance and any associated hardware specifications;

    (iii) Specification of acceptable incoming sensor data quality control measures; and

    (iv) Mitigation of impact of user error or failure of any subsystem components (signal detection and analysis, data display, and storage) on accuracy of patient reports.

    (2) Scientific justification for the validity of the status indicator algorithm(s) must be provided. Verification of algorithm calculations and validation testing of the algorithm using a data set separate from the training data must demonstrate the validity of modeling.

    (3) Usability assessment must be provided to demonstrate that risk of misinterpretation of the status indicator is appropriately mitigated.

    (4) Clinical data must be provided in support of the intended use and include the following:

    (i) Output measure(s) must be compared to an acceptable reference method to demonstrate that the output measure(s) represent(s) the predictive measure(s) that the device provides in an accurate and reproducible manner;

    (ii) The data set must be representative of the intended use population for the device. Any selection criteria or limitations of the samples must be fully described and justified;

    (iii) Agreement of the measure(s) with the reference measure(s) must be assessed across the full measurement range; and

    (iv) Data must be provided within the clinical validation study or using equivalent datasets to demonstrate the consistency of the output and be representative of the range of data sources and data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment.

    (5) Labeling must include the following:

    (i) The type of sensor data used, including specification of compatible sensors for data acquisition;

    (ii) A description of what the device measures and outputs to the user;

    (iii) Warnings identifying sensor reading acquisition factors that may impact measurement results;

    (iv) Guidance for interpretation of the measurements, including warning(s) specifying adjunctive use of the measurements;

    (v) Key assumptions made in the calculation and determination of measurements;

    (vi) The measurement performance of the device for all presented parameters, with appropriate confidence intervals, and the supporting evidence for this performance; and

    (vii) A detailed description of the patients studied in the clinical validation (e.g., age, gender, race/ethnicity, clinical stability) as well as procedural details of the clinical study.

    Start Signature

    Dated: July 24, 2017.

    Leslie Kux,

    Associate Commissioner for Policy.

    End Signature End Supplemental Information

    [FR Doc. 2017-15901 Filed 7-27-17; 8:45 am]

    BILLING CODE 4164-01-P

Document Information

Effective Date:
7/28/2017
Published:
07/28/2017
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final order.
Document Number:
2017-15901
Dates:
This order is effective July 28, 2017. The classification was applicable on December 21, 2016.
Pages:
35065-35067 (3 pages)
Docket Numbers:
Docket No. FDA-2017-N-1620
Topics:
Medical devices
PDF File:
2017-15901.pdf
CFR: (1)
21 CFR 870.2200