2018-01638. Medical Devices; Cardiovascular Devices; Classification of the Temporary Catheter for Embolic Protection During Transcatheter Intracardiac Procedures  

  • Start Preamble

    AGENCY:

    Food and Drug Administration, HHS.

    ACTION:

    Final order.

    SUMMARY:

    The Food and Drug Administration (FDA or we) is classifying the temporary catheter for embolic protection during transcatheter intracardiac procedures into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the temporary catheter for embolic protection during transcatheter intracardiac procedures' classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.

    DATES:

    This order is effective January 30, 2018. The classification was applicable on June 1, 2017.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    Sadaf Toor, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 1202, Silver Spring, MD 20993-0002, 301-796-6381, Sadaf.Toor@fda.hhs.gov.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    I. Background

    Upon request, FDA has classified the temporary catheter for embolic protection during transcatheter intracardiac procedures as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness. In addition, we believe this action will enhance patients' access to beneficial innovation, in part by reducing regulatory burdens by placing the device into a lower device class than the automatic class III assignment.

    The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).

    FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate by means of the procedures for premarket notification under section 510(k) of the FD&C Act and part 807 (21 U.S.C. 360(k) and 21 CFR part 807, respectively).

    FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act. Section 207 of the Food and Drug Administration Modernization Act of 1997 established the first procedure for De Novo classification (Pub. L. 105-115). Section 607 of the Food and Drug Administration Safety and Innovation Act modified the De Novo application process by adding a second procedure (Pub. L. 112-144). A device sponsor may utilize either procedure for De Novo classification.

    Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).

    Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.Start Printed Page 4140

    Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically within class III, the De Novo classification is considered to be the initial classification of the device.

    We believe this De Novo classification will enhance patients' access to beneficial innovation, in part by reducing regulatory burdens. When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application in order to market a substantially equivalent device (see 21 U.S.C. 360c(i), defining “substantial equivalence”). Instead, sponsors can use the less-burdensome 510(k) process, when necessary, to market their device.

    II. De Novo Classification

    On September 20, 2016, Claret Medical, Inc., submitted a request for De Novo classification of the Sentinel® Cerebral Protection System. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.

    We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.

    Therefore, on June 1, 2017, FDA issued an order to the requester classifying the device into class II. FDA is codifying the classification of the device by adding 21 CFR 870.1251. We have named the generic type of device temporary catheter for embolic protection during transcatheter intracardiac procedures, and it is identified as a single use percutaneous catheter system that has (a) blood filter(s) at the distal end. This device is indicated for use while performing transcatheter intracardiac procedures. The device is used to filter blood in a manner that may prevent embolic material (thrombus/debris) from the transcatheter intracardiac procedure from traveling towards the cerebral circulation.

    FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.

    Table 1—Temporary Catheter for Embolic Protection During Transcatheter Intracardiac Procedures Risks and Mitigation Measures

    Identified risksMitigation measures
    Device failure leading to debris embolization and stroke or deathNon-clinical performance testing, Animal testing, and Clinical performance testing.
    Impeded or disrupted blood flow leading to peripheral ischemiaNon-clinical performance testing, Animal testing, Clinical performance testing, and Labeling.
    Device incompatibility with transcatheter intracardiac procedure device leading to prolonged treatment time or device failureNon-clinical performance testing, Animal testing, Clinical performance testing, and Labeling.
    Adverse tissue reactionBiocompatibility evaluation.
    InfectionSterilization validation, Shelf life testing, and Labeling.
    Vascular injury due to device delivery, deployment, placement, or retrievalNon-clinical performance testing, Animal testing, Clinical performance testing, and Labeling.

    FDA has determined that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of safety and effectiveness. For a device to fall within this classification, and thus avoid automatic classification in class III, it would have to comply with the special controls named in this final order. The necessary special controls appear in the regulation codified by this order. This device is subject to premarket notification requirements under section 510(k) of the FD&C Act.

    III. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

    IV. Paperwork Reduction Act of 1995

    This final order establishes special controls that refer to previously approved collections of information found in other FDA regulations. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections of information in the guidance document “De Novo Classification Process (Evaluation of Automatic Class III Designation)” have been approved under OMB control number 0910-0844; the collections of information in part 814, subparts A through E, regarding premarket approval, have been approved under OMB control number 0910-0231; the collections of information in part 807, subpart E, regarding premarket notification submissions, have been approved under OMB control number 0910-0120; and the collections of information in 21 CFR part 801, regarding labeling, have been approved under OMB control number 0910-0485.

    Start List of Subjects

    List of Subjects in 21 CFR Part 870

    • Medical devices
    End List of Subjects

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 870 is amended as follows:

    Start Part

    PART 870—CARDIOVASCULAR DEVICES

    End Part Start Amendment Part

    1. The authority citation for part 870 continues to read as follows:

    End Amendment Part Start Authority

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360 l, 371.

    End Authority Start Amendment Part

    2. Add § 870.1251 to subpart B to read as follows:

    End Amendment Part
    Start Printed Page 4141
    Temporary catheter for embolic protection during transcatheter intracardiac procedures.

    (a) Identification. This device is a single use percutaneous catheter system that has (a) blood filter(s) at the distal end. This device is indicated for use while performing transcatheter intracardiac procedures. The device is used to filter blood in a manner that may prevent embolic material (thrombus/debris) from the transcatheter intracardiac procedure from traveling towards the cerebral circulation.

    (b) Classification. Class II (special controls). The special controls for this device are:

    (1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:

    (i) Simulated-use testing in a clinically relevant bench anatomic model to assess the following:

    (A) Delivery, deployment, and retrieval, including quantifying deployment and retrieval forces, and procedural time; and

    (B) Device compatibility and lack of interference with the transcatheter intracardiac procedure and device.

    (ii) Tensile strengths of joints and components, tip flexibility, torque strength, torque response, and kink resistance.

    (iii) Flow characteristics.

    (A) The ability of the filter to not impede blood flow.

    (B) The amount of time the filter can be deployed in position and/or retrieved from its location without disrupting blood flow.

    (iv) Characterization and verification of all dimensions.

    (2) Animal testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be assessed:

    (i) Delivery, deployment, and retrieval, including quantifying procedural time.

    (ii) Device compatibility and lack of interference with the transcatheter intracardiac procedure and device.

    (iii) Flow characteristics.

    (A) The ability of the filter to not impede blood flow.

    (B) The amount of time the filter can be deployed in position and/or retrieved from its location without disrupting blood flow.

    (iv) Gross pathology and histopathology assessing vascular injury and downstream embolization.

    (3) All patient contacting components of the device must be demonstrated to be biocompatible.

    (4) Performance data must demonstrate the sterility of the device components intended to be provided sterile.

    (5) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.

    (6) Labeling for the device must include:

    (i) Instructions for use;

    (ii) Compatible transcatheter intracardiac procedure devices;

    (iii) A detailed summary of the clinical testing conducted; and

    (iv) A shelf life and storage conditions.

    (7) Clinical performance testing must demonstrate:

    (i) The ability to safely deliver, deploy, and remove the device;

    (ii) The ability of the device to filter embolic material while not impeding blood flow;

    (iii) Secure positioning and stability of the position throughout the transcatheter intracardiac procedure; and

    (iv) Evaluation of all adverse events including death, stroke, and vascular injury.

    Start Signature

    Dated: January 24, 2018.

    Leslie Kux,

    Associate Commissioner for Policy.

    End Signature End Supplemental Information

    [FR Doc. 2018-01638 Filed 1-29-18; 8:45 am]

    BILLING CODE 4164-01-P

Document Information

Effective Date:
1/30/2018
Published:
01/30/2018
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final order.
Document Number:
2018-01638
Dates:
This order is effective January 30, 2018. The classification was applicable on June 1, 2017.
Pages:
4139-4141 (3 pages)
Docket Numbers:
Docket No. FDA-2017-N-6285
Topics:
Medical devices
PDF File:
2018-01638.pdf
CFR: (1)
21 CFR 870.1251