2018-22837. Medical Devices; General and Plastic Surgery Devices; Classification of the Wound Autofluorescence Imaging Device
-
Start Preamble
AGENCY:
Food and Drug Administration, HHS.
ACTION:
Final order.
SUMMARY:
The Food and Drug Administration (FDA or we) is classifying the wound autofluorescence imaging device into class I. We are taking this action because we have determined that classifying the device into class I will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.
DATES:
This order is effective October 19, 2018. The classification was applicable on July 31, 2018.
Start Further InfoFOR FURTHER INFORMATION CONTACT:
Yasaman Ardeshirpour, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. G455, Silver Spring, MD, 20993-0002, 240-402-3706, Yasaman.Ardeshirpour@fda.hhs.gov.
End Further Info End Preamble Start Supplemental InformationSUPPLEMENTARY INFORMATION:
Start Printed Page 52967
I. Background
Upon request, FDA has classified the wound autofluorescence imaging device as class I (general controls), which we have determined will provide a reasonable assurance of safety and effectiveness. In addition, we believe this action will enhance patients' access to beneficial innovation, in part by reducing regulatory burdens by placing the device into a lower device class than the automatic class III assignment.
The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act to a predicate device that does not require premarket approval (see 21 U.S.C. 360c(i)). We determine whether a new device is substantially equivalent to a predicate by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act (21 U.S.C. 360c(f)(2)). Section 207 of the Food and Drug Administration Modernization Act of 1997 established the first procedure for De Novo classification (Pub. L. 105-115). Section 607 of the Food and Drug Administration Safety and Innovation Act modified the De Novo application process by adding a second procedure (Pub. L. 112-144). A device sponsor may utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).
Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA shall classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act (21 U.S.C. 360c(a)(1)). Although the device was automatically within class III, the De Novo classification is considered to be the initial classification of the device.
We believe this De Novo classification will enhance patients' access to beneficial innovation, in part by reducing regulatory burdens. When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a result, other device sponsors do not have to submit a De Novo request or PMA in order to market a substantially equivalent device (see 21 U.S.C. 360c(i), defining “substantial equivalence”). Instead, sponsors can use the less-burdensome 510(k) process, when necessary, to market their device.
II. De Novo Classification
On February 16, 2018, MolecuLight, Inc. submitted a request for De Novo classification of the MolecuLight i:X. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class I if general controls are sufficient to provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(A)). After review of the information submitted in the request, we determined that the device can be classified into class I. FDA has determined that general controls will provide reasonable assurance of the safety and effectiveness of the device.
Therefore, on July 31, 2018, FDA issued an order to the requester classifying the device into class I. FDA is codifying the classification of the device by adding 21 CFR 878.4165. We have named the generic type of device wound autofluorescence imaging device, and it is identified as a tool to view autofluorescence images from skin wounds that are exposed to an excitation light. The device is not intended to provide quantitative or diagnostic information.
FDA has identified the following risks to health associated specifically with this type of device: electrical/mechanical/thermal, electromagnetic compatibility (EMC) and optical safety of the device, and the error in fluorescence detection from the wound.
Section 510(l)(1) of the FD&C Act provides that a device within a type that has been classified into class I under section 513 of the FD&C Act is exempt from premarket notification under section 510(k), unless the device is of substantial importance in preventing impairment of human health or presents a potentially unreasonable risk of illness or injury (21 U.S.C. 360(l)(1)). Devices within this type are exempt from the premarket notification requirements under section 510(k), subject to the limitations of exemptions in 21 CFR 878.9.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order refers to previously approved collections of information found in other FDA regulations and guidance. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections of information in the guidance document “De Novo Classification Process (Evaluation of Automatic Class III Designation)” have been approved under OMB control number 0910-0844; the collections of information in 21 CFR part 814, subparts A through E, regarding premarket approval, have been approved under OMB control number 0910-0231; the collections of information in 21 CFR part 820, regarding current good manufacturing practices, have been approved under OMB control number 0910-0073; and the collections of information in part 807, subpart E, regarding premarket notification submissions, have been approved under OMB control number 0910-0120; and the collections of information in 21 CFR part 801, regarding labeling, have been approved under OMB control number 0910-0485.
Start List of Subjects Start Printed Page 52968List of Subjects in 21 CFR Part 878
- Medical devices
Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, part 878 is amended as follows:
Start PartPART 878—GENERAL AND PLASTIC SURGERY DEVICES
End Part Start Amendment Part1. The authority citation for part 878 continues to read as follows:
End Amendment Part Start Amendment Part2. Add § 878.4165 to subpart E to read as follows:
End Amendment PartWound autofluorescence imaging device.(a) Identification. A wound autofluorescence imaging device is a tool to view autofluorescence images from skin wounds that are exposed to an excitation light. The device is not intended to provide quantitative or diagnostic information.
(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.
Dated: October 16, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018-22837 Filed 10-18-18; 8:45 am]
BILLING CODE 4164-01-P
Document Information
- Effective Date:
- 10/19/2018
- Published:
- 10/19/2018
- Department:
- Food and Drug Administration
- Entry Type:
- Rule
- Action:
- Final order.
- Document Number:
- 2018-22837
- Dates:
- This order is effective October 19, 2018. The classification was applicable on July 31, 2018.
- Pages:
- 52966-52968 (3 pages)
- Docket Numbers:
- Docket No. FDA-2018-N-3696
- Topics:
- Medical devices
- PDF File:
- 2018-22837.pdf
- CFR: (1)
- 21 CFR 878.4165