96-25720. Medical Devices; Current Good Manufacturing Practice (CGMP) Final Rule; Quality System Regulation  

  • [Federal Register Volume 61, Number 195 (Monday, October 7, 1996)]
    [Rules and Regulations]
    [Pages 52602-52662]
    From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
    [FR Doc No: 96-25720]
    
    
          
    
    [[Page 52601]]
    
    
    _______________________________________________________________________
    
    Part VII
    
    
    
    
    
    Department of Health and Human Services
    
    
    
    
    
    _______________________________________________________________________
    
    
    
    Food and Drug Administration
    
    
    
    _______________________________________________________________________
    
    
    
    21 CFR Parts 808, 812, and 820
    
    
    
    Medical Devices; Current Good
    
    
    
    Manufacturing Practice (CGMP); Final Rule
    
    Federal Register / Vol. 61, No. 195 / Monday, October 7, 1996 / Rules 
    and Regulations
    
    [[Page 52602]]
    
    
    
    DEPARTMENT OF HEALTH AND HUMAN SERVICES
    
    Food and Drug Administration
    
    21 CFR Parts 808, 812, and 820
    
    [Docket No. 90N-0172]
    RIN 0910-AA09
    
    
    Medical Devices; Current Good Manufacturing Practice (CGMP) Final 
    Rule; Quality System Regulation
    
    AGENCY: Food and Drug Administration, HHS.
    
    ACTION: Final rule.
    
    -----------------------------------------------------------------------
    
    SUMMARY: The Food and Drug Administration (FDA) is revising the current 
    good manufacturing practice (CGMP) requirements for medical devices and 
    incorporating them into a quality system regulation. The quality system 
    regulation includes requirements related to the methods used in, and 
    the facilities and controls used for, designing, manufacturing, 
    packaging, labeling, storing, installing, and servicing of medical 
    devices intended for human use. This action is necessary to add 
    preproduction design controls and to achieve consistency with quality 
    system requirements worldwide. This regulation sets forth the framework 
    for device manufacturers to follow and gives them greater flexibility 
    in achieving quality requirements.
    
    DATES: The regulation is effective June 1, 1997. For more information 
    on compliance with 21 CFR 820.30 see section IV. of this document.
        Written comments on the information collection requirements should 
    be submitted by December 6, 1996.
    
    ADDRESSES: Submit written comments on the information collection 
    requirements to the Dockets Management Branch (HFA-305), Food and Drug 
    Administration, 12420 Parklawn Dr., rm. 1-23, Rockville, MD 20857. All 
    comments should be identified with the docket number found in brackets 
    in the heading of this document.
    
    FOR FURTHER INFORMATION CONTACT: Kimberly A. Trautman, Center for 
    Devices and Radiological Health (HFZ-341), Food and Drug 
    Administration, 2098 Gaither Rd., Rockville, MD 20850, 301-594-4648.
    
    SUPPLEMENTARY INFORMATION:
    
    I. Background
    
        Manufacturers establish and follow quality systems to help ensure 
    that their products consistently meet applicable requirements and 
    specifications. The quality systems for FDA-regulated products (food, 
    drugs, biologics, and devices) are known as CGMP's. CGMP requirements 
    for devices in part 820 (21 CFR part 820) were first authorized by 
    section 520(f) of the Federal Food, Drug, and Cosmetic Act (the act) 
    (21 U.S.C. 360j(f)), which was among the authorities added to the act 
    by the Medical Device Amendments of 1976 (Pub. L. 94-295).
        Under section 520(f) of the act, FDA issued a final rule in the 
    Federal Register of July 21, 1978 (43 FR 31 508), prescribing CGMP 
    requirements for the methods used in, and the facilities and controls 
    used for the manufacture, packing, storage, and installation of medical 
    devices. This regulation became effective on December 18, 1978, and is 
    codified under part 820. Except for editorial changes to update 
    organizational references in the regulation and revisions to the list 
    of critical devices that was included in the preamble to the final 
    regulation, the device CGMP requirements have not been revised since 
    1978. This final rule is the result of an extensive effort begun in 
    1990 to revise this regulation.
        The Safe Medical Devices Act of 1990 (the SMDA) (Pub. L. 101-629), 
    enacted on November 28, 1990, amended section 520(f) of the act, 
    providing FDA with the authority to add preproduction design controls 
    to the CGMP regulation. This change in law was based on findings that a 
    significant proportion of device recalls were attributed to faulty 
    design of product. Specifically, in January 1990, FDA published the 
    results of an evaluation of device recalls that occurred from October 
    1983 through September 1989, in a report entitled ``Device Recalls: A 
    Study of Quality Problems'' (Ref. 1). (See 55 FR 21108, May 22, 1990, 
    where FDA announced the availability of the report.) FDA found that 
    approximately 44 percent of the quality problems that led to voluntary 
    recall actions during this 6-year period were attributed to errors or 
    deficiencies that were designed into particular devices and may have 
    been prevented by adequate design controls. These design-related 
    defects involved both noncritical devices (e.g., patient chair lifts, 
    in vitro diagnostics, and administration sets) and critical devices 
    (e.g., pacemakers and ventilators). Also in 1990, the Department of 
    Health and Human Services' Inspector General conducted a study entitled 
    ``FDA Medical Device Regulation From Premarket Review to Recall'' (Ref. 
    2), which reached similar conclusions. With respect to software used to 
    operate medical devices, the data were even more striking. A subsequent 
    study of software-related recalls for the period of fiscal year (FY) 
    1983 through FY 1991 indicated that over 90 percent of all software-
    related device failures were due to design-related errors, generally, 
    the failure to validate software prior to routine production (Ref. 3).
        The SMDA also added new section 803 to the act (21 U.S.C. 383) 
    which, among other things, encourages FDA to work with foreign 
    countries toward mutual recognition of CGMP requirements. FDA undertook 
    the revision of the CGMP regulation to add the design controls 
    authorized by the SMDA to the CGMP regulation, as well as because the 
    agency believed that it would be beneficial to the public and the 
    medical device industry for the CGMP regulation to be consistent, to 
    the extent possible, with the requirements for quality systems 
    contained in applicable international standards, primarily, the 
    International Organization for Standards (ISO) 9001:1994 ``Quality 
    Systems--Model for Quality Assurance in Design, Development, 
    Production, Installation, and Servicing'' (Ref. 4), and the ISO 
    committee draft (CD) revision of ISO/CD 13485 ``Quality Systems--
    Medical Devices--Supplementary Requirements to ISO 9001'' (Ref. 5).
        This action is being taken under those provisions of the SMDA and 
    in response to the following: (1) Notices that appeared in the Federal 
    Register of April 25, 1990 (55 FR 17502), and in the Federal Register 
    of April 17, 1991 (56 FR 15626), that announced meetings of the 
    agency's Device Good Manufacturing Practice Advisory Committee (GMP 
    Advisory Committee), at which the need for revisions to the CGMP 
    regulation was explored; (2) an advance notice of proposed rulemaking 
    (ANPRM) that appeared in the Federal Register of June 15, 1990 (55 FR 
    24544), that announced the agency's intent to revise the CGMP 
    regulation; (3) a notice of availability of a document that appeared in 
    the Federal Register of November 30, 1990 (55 FR 49644), entitled 
    ``Medical Devices; Current Good Manufacturing Practices (CGMP) 
    Regulations Document; Suggested Changes; Availability'' (Ref. 6) and 
    comments solicited from the public about the document; (4) a proposed 
    rule in the Federal Register of November 23, 1993 (58 FR 61952), (Ref. 
    7) and comments solicited from the public about the proposal; (5) a 
    notice of availability that appeared in the Federal Register of July 
    24, 1995 (60 FR 37856), announcing the availability of the ``Working 
    Draft of the Current Good Manufacturing Practice (CGMP) Final Rule'' 
    (hereinafter referred to as the Working Draft) (Ref. 8) and comments
    
    [[Page 52603]]
    
    solicited from the public about the Working Draft; (6) testimony at an 
    August 23, 1995, open public meeting announced in the Federal Register 
    (60 FR 37856); (7) and testimony and advisory committee recommendations 
    from the September 13 and 14, 1995, meeting of the GMP Advisory 
    Committee announced in the Federal Register of August 24, 1995 (60 FR 
    44036). Thus, FDA's decision to revise the CGMP regulation is based on 
    changes in the law made by the SMDA, the agency's discussions with 
    others including its GMP Advisory Committee, responses to the Federal 
    Register notices on this matter, FDA's analysis of recall data, its 
    experience with the regulatory application of the original CGMP 
    regulation, and its assessment of international quality standards.
        The agency's final rule embraces the same ``umbrella'' approach to 
    the CGMP regulation that is the underpinning of the original CGMP 
    regulation. Because this regulation must apply to so many different 
    types of devices, the regulation does not prescribe in detail how a 
    manufacturer must produce a specific device. Rather, the regulation 
    provides the framework that all manufacturers must follow by requiring 
    that manufacturers develop and follow procedures and fill in the 
    details that are appropriate to a given device according to the current 
    state-of-the-art manufacturing for that specific device. FDA has made 
    changes to the proposed regulation and the Working Draft, as the final 
    rule evidences, to provide manufacturers with even greater flexibility 
    in achieving the quality requirements.
        The Supreme Court recently addressed the preemptive effect, under 
    section 521 of the act (21 U.S.C. 360k), of the original CGMP 
    regulation and other FDA requirements for medical devices on State tort 
    actions. In Medtronic, Inc. v. Lohr, 116 S. Ct. 2240 (1996), the 
    Supreme Court gave substantial deference to the agency's interpretation 
    of section 521 of the act found at Sec. 808.1 (21 CFR 808.1). The Court 
    noted that CGMP requirements are general rather than ``specific 
    requirements applicable to a particular device,'' and that State common 
    law remedies are similarly general, and do not establish a 
    ``substantive requirement for a specific device.'' (Lohr at 2257; see 
    also Sec. 808.1(d) and (d)(6)(ii).) Moreover, the Court drew a 
    distinction between remedies and requirements, noting that while common 
    law tort actions may provide remedies different from those available 
    under the act, no preemption occurs unless the substantive requirements 
    of the State law are ``different from, or in addition to,'' those 
    imposed by the act. (See Lohr at 2255.) Under the Supreme Court's 
    analysis in Lohr, the requirements imposed by the original CGMP 
    regulation would rarely have preemptive effect.
        FDA believes that the reasoning of Medtronic v. Lohr applies 
    equally to the new quality system regulation, which, as does the 
    original CGMP regulation, prescribes requirements that apply to medical 
    devices in general, rather than to any particular medical device. 
    Therefore, FDA has concurrently amended part 808 (21 CFR part 808) to 
    make clear the new quality system regulation does not preempt State 
    tort and common law remedies.
    
    II. Decision to Make a Working Draft Available for Comment
    
        In the Federal Register of November 23, 1993, the agency issued the 
    proposed revisions to the CGMP regulation, entitled ``Medical Devices; 
    Current Good Manufacturing Practice (CGMP) Regulations; Proposed 
    Revisions; Request for Comments,'' and public comment was solicited. 
    After the proposal issued, FDA met with the Global Harmonization Task 
    Force (the GHTF) Study Group in early March 1994, in Brussels, to 
    compare the provisions of the proposal with the provisions of ISO 
    9001:1994 and European National Standard (EN) 46001 ``Quality Systems--
    Medical Devices--Particular Requirements for the Application of EN 
    29001'' (Ref. 9). ISO 9001:1994 and EN 46001:1994 are written as 
    voluntary standards, but when used to fulfill the requirements of the 
    European Medical Device Directives, or other national regulations, 
    these standards are mandatory requirements similar to the CGMP 
    requirements. The GHTF includes: Representatives of the Canadian 
    Ministry of Health and Welfare, the Japanese Ministry of Health and 
    Welfare, FDA, and industry members from the European Union (EU), 
    Australia, Canada, Japan, and the United States. The participants at 
    the GHTF meeting favorably regarded FDA's effort toward harmonization 
    with international standards. The GHTF submitted comments, however, 
    noting where FDA could more closely harmonize to achieve consistency 
    with quality system requirements worldwide. Since the proposal 
    published, FDA has also attended numerous industry and professional 
    association seminars and workshops, including ISO Technical Committee 
    (TC) 210 ``Quality Management and Corresponding General Aspects for 
    Medical Devices'' meetings, where the proposed revisions were 
    discussed.
        The original period for comment on the proposal closed on February 
    22, 1994, and was extended until April 4, 1994. Because of the heavy 
    volume of comments and the desire to increase public participation in 
    the development of the quality system regulation, FDA decided to 
    publish the notice of availability in the Federal Register to allow 
    comment on the Working Draft before issuing a final regulation.
        The Working Draft represented the agency's views at the time on how 
    it would respond to the many comments received, and on how the agency 
    believed a final rule should be framed. FDA solicited public comment on 
    the Working Draft until October 23, 1995, to determine if the agency 
    had adequately addressed the many comments received and whether the 
    agency had framed a final rule that achieved the public health goals to 
    be gained from implementation of quality systems in the most efficient 
    manner.
    
    III. Open Public Meeting and GMP Advisory Committee Meeting
    
        FDA held an open public meeting on the quality system regulation on 
    August 23, 1995. The public meeting consisted of prepared presentations 
    followed by an open discussion period. Both the agency and the 
    participants found the meeting to be very productive in focusing 
    attention on the few main areas of concern in the Working Draft. The 
    main issues were: The application of the regulation to component 
    manufacturers; the application of the regulation to third party 
    servicers and refurbishers; and the implementation timeframe of the 
    final rule. A transcript of the proceedings of the public meeting, as 
    well as data and information submitted to FDA during the public 
    meeting, are available from the Dockets Management Branch (HFA-305), 
    Food and Drug Administration, 12420 Parklawn Dr., rm. 1-23, Rockville, 
    MD 20857, between 9 a.m. and 4 p.m., Monday through Friday.
        There also was a meeting of the GMP Advisory Committee on the 
    Working Draft on September 13 and 14, 1995. A notice of the meeting was 
    published in the Federal Register of August 24, 1995. FDA made a brief 
    presentation to the committee on the changes from the 1993 proposal to 
    the 1995 Working Draft and discussed some changes that FDA was 
    recommending as a result of the August 1995 meeting. Two consultants 
    also made presentations to the committee, one a representative from ISO 
    TC 176 (the TC that authored the ISO 9000 series) and the other a 
    representative from the European Committee for Standardization (CEN). 
    The remainder of the meeting consisted of prepared
    
    [[Page 52604]]
    
    presentations from the public and the committee's discussion on the 
    main issues.
        The overwhelming majority of the committee members believed that 
    the Working Draft met the public health needs, gave manufacturers 
    sufficient flexibility to comply with the regulation, and met the 
    agency's goal of harmonizing the quality system requirements with those 
    of other countries. The GMP Advisory Committee strongly supported FDA's 
    recommendation, in response to the August 1995 public meeting, to not 
    include component manufacturers under this final rule. However, the GMP 
    Advisory Committee was clearly divided on several issues related to the 
    proposed regulation of third party servicers and refurbishers. A 
    transcript of the proceedings of the GMP Advisory Committee meeting, as 
    well as data and information submitted to FDA during the meeting, are 
    available from the Dockets Management Branch (address above).
        After considering the written comments and the views expressed at 
    meetings with the GHTF, at the August 1995 public meeting, and at the 
    September 1995 GMP Advisory Committee meeting, FDA is publishing this 
    final rule. A summary of changes from the July 1995 Working Draft to 
    the final rule is contained at the end of this preamble.
    
    IV. Implementation of the Final Rule
    
        FDA has decided, in response to the many comments and concerns 
    expressed about the need for more time to implement design controls, to 
    implement the final rule in two stages. Under stage one, on June 1, 
    1997, approximately 1 year after this rule is published in the Federal 
    Register, all elements of the final rule become effective. However, 
    with respect to the design control requirements in Sec. 820.30, as long 
    as manufacturers are taking reasonable steps to come into compliance, 
    FDA will implement a special 1-year transition program, with a 
    midcourse review, during which official agency action will not be 
    initiated, including FDA Form 483 observations, warning letters, or 
    enforcement cases, based on failure to comply with Sec. 820.30. Under 
    stage two, beginning June 1, 1998, FDA will treat noncompliance with 
    design control requirements in Sec. 820.30 the same as noncompliance 
    with other provisions of the CGMP regulation.
        To prepare for stage one of this implementation plan, FDA intends 
    to develop, by April of 1997, a strategy for inspecting the design 
    control requirements. Both industry and FDA field investigators will 
    then be trained on this inspectional strategy for design controls 
    during April and May 1997. Starting June 1, 1997, manufacturers will be 
    inspected for compliance with all the new quality system requirements, 
    including design controls, in the manner described in the inspectional 
    strategy. However, as part of the transition program, from June 1, 
    1997, for a period of 1 year, although FDA will inspect firms for 
    compliance with the design control requirements, the field will issue 
    any observations to the manufacturer on a separate design control 
    inspectional strategy report, not on FDA Form 483. The design control 
    inspectional strategy report will be made a part of the manufacturer's 
    establishment inspection report (EIR), but the observations relating to 
    Sec. 820.30 will not be included in any warning letters or regulatory 
    actions during this initial 1-year period. FDA notes that it can, at 
    any time, take action against unsafe or adulterated medical devices 
    under different regulatory or statutory authorities. FDA wants to 
    emphasize that manufacturers are required to take reasonable steps to 
    come into compliance with the design control requirements during the 
    June 1, 1997, to June 1, 1998, period.
        FDA also emphasizes that this transition period relates only to the 
    design control requirements of Sec. 820.30, and that beginning June 1, 
    1997, the agency will issue observations on FDA Form 483's, issue 
    warning letters, and take any necessary regulatory action for 
    violations of all other provisions of the CGMP final rule. The time 
    period from June 1, 1997, to June 1, 1998, is intended to allow both 
    the industry and FDA field investigators time to become familiar with 
    the design control requirements and the enforcement aspects of this new 
    area.
        Finally, as described elsewhere in this preamble, FDA intends to 
    conduct a midcourse review of the new design control requirements 
    during the transition year (June 1997 to June 1998). Specifically, the 
    results of the first several months of design control inspections will 
    be reviewed by early 1998. FDA will review all of the completed design 
    control inspectional strategy reports that were given to manufacturers 
    from between June 1, 1997, through December 1, 1997. The completed 
    strategy reports will be reviewed with particular attention paid to 
    clarity of information obtained, the appropriateness of the information 
    collected with respect to the design control requirements, the 
    appropriateness of the questions on the inspectional strategy, the 
    manner in which the investigators are writing out their observations, 
    and any requirements that seem to be giving manufacturers a problem or 
    where there might be misunderstandings as to what the regulation 
    requires. FDA will then hold an open public meeting in early 1998 to 
    discuss with industry these findings and to further explore any 
    concerns industry might be having in implementing the new design 
    control requirements. As a result of the midcourse review and open 
    public meeting, FDA might hold additional workshops, meetings, and/or 
    training sessions.
        Any midcourse adjustments to the inspectional strategy will be 
    instituted and made public by the spring of 1998. Also during this 
    midcourse review, FDA will evaluate the information gathered at that 
    point and determine if the design control requirements as written in 
    this final rule are appropriate to obtain the goals expressed in this 
    preamble. FDA will consider minor or even major changes, based on 
    experience to date. Any necessary adjustments or proposed revisions 
    will be published in the Federal Register and comments will be 
    solicited as necessary during the spring of 1998. This implementation 
    strategy is responsive to requests by industry for FDA to harmonize the 
    quality system regulation's implementation with the mandatory date for 
    implementation of the EU's Medical Device Directive, which is June 
    1998. However, if during the midcourse review of stage one it is 
    determined that the industry and/or FDA needs more time to fully 
    implement the design control requirements, FDA will publish an 
    extension of the regulatory implementation date for design control 
    requirements prior to June 1, 1998.
    
    V. Response to Comments and Rationale for Changes
    
        Approximately 280 separate individuals or groups commented on the 
    proposal published in the Federal Register of November 23, 1993, and 
    approximately 175 separate individuals or groups commented on the 
    Working Draft that was announced in a notice of availability published 
    in the Federal Register on July 24, 1995. FDA made many changes in 
    response to the comments. Most of the changes were made in response to 
    specific comments, in response to comments for clarity, understanding, 
    and readability, or to further harmonize FDA requirements with 
    international standards, as many comments requested.
        Numerous comments stated that industry was very pleased with FDA's
    
    [[Page 52605]]
    
    Working Draft and the effort that was made to harmonize with ISO, as 
    well as to engage industry in commenting on the Working Draft through 
    the open public meeting and the GMP Advisory Committee meeting that 
    were held in August and September 1995, respectively.
        FDA's responses to the comments received on the proposal and the 
    Working Draft, as well as explanations for the changes made, follow.
    
    A. General Provisions (Subpart A)
    
    i. Scope (Sec. 820.1)
        1. The title of the regulation, as reflected in this section, has 
    been changed from the ``Current Good Manufacturing Practices (CGMP)'' 
    regulation to the ``Quality System'' regulation. This revision follows 
    the suggestion underlying many comments on specific provisions that FDA 
    generally harmonize the CGMP requirements and terminology with 
    international standards. ISO 9001:1994, ISO/CD 13485, and EN 46001 
    employ this terminology to describe the CGMP requirements. In addition, 
    this title accurately describes the sum of the requirements, which now 
    include the CGMP requirements for design, purchasing, and servicing 
    controls. CGMP requirements now cover a full quality system.
        FDA notes that the principles embodied in this quality system 
    regulation have been accepted worldwide as a means of ensuring that 
    acceptable products are produced. While the regulation has been 
    harmonized with the medical device requirements in Europe, Australia, 
    and Japan, as well as the requirements proposed by Canada, it is 
    anticipated that other countries will adopt similar requirements in the 
    near future.
        FDA, however, did not adopt ISO 9001:1994 verbatim for two reasons. 
    First, there were complications in dealing with the issue of copyrights 
    and, second, FDA along with health agencies of other governments does 
    not believe that for medical devices ISO 9001:1994 alone is sufficient 
    to adequately protect the public health. Therefore, FDA has worked 
    closely with the GHTF and TC 210 to develop a regulation which is 
    consistent with both ISO 9001:1994 and ISO/CD 13485. FDA made several 
    suggestions to TC 210 on the drafts of the ISO/CD 13485 document in 
    order to minimize differences and move closer to harmonization. In some 
    cases, FDA has explicitly stated requirements that many experts believe 
    are inherent in ISO 9001:1994. Through the many years of experience 
    enforcing and evaluating compliance with the original CGMP regulation, 
    FDA has found that it is necessary to clearly spell out its 
    expectations. This difference in approach does not represent any 
    fundamentally different requirements that would hinder global 
    harmonization. In fact, numerous comments expressed their approval and 
    satisfaction with FDA's effort to harmonize the quality system 
    requirements with those of ISO 9001:1994 and ISO/CD 13485.
        2. One comment suggested that the term ``purchasing'' in the scope 
    be deleted because it could be interpreted to mean the purchase of 
    finished medical devices by health care institutions and medical 
    professionals, instead of the purchase of components and manufacturing 
    materials as intended.
        FDA agrees and has deleted the term ``purchasing'' throughout the 
    regulation when used in this context.
        3. Several comments suggested that Sec. 820.1(a)(1) should not 
    state that the regulation establishes the ``minimum'' requirements 
    because it implies that compliance with the stated requirements may be 
    insufficient. They asked that FDA delete the word ``minimum,'' to avoid 
    having auditors search for additional requirements.
        FDA does not believe that the provision would have required that 
    manufacturers meet additional requirements not mandated by the 
    regulation but has modified the section to clarify its intent by 
    stating that the regulation establishes the ``basic'' requirements for 
    manufacturing devices. The quality system regulation provides a 
    framework of basic requirements for each manufacturer to use in 
    establishing a quality system appropriate to the devices designed and 
    manufactured and the manufacturing processes employed. Manufacturers 
    must adopt current and effective methods and procedures for each device 
    they design and manufacture to comply with and implement the basic 
    requirements. The regulation provides the flexibility necessary to 
    allow manufacturers to adopt advances in technology, as well as new 
    manufacturing and quality system procedures, as they become available.
        During inspections, FDA will assess whether a manufacturer has 
    established procedures and followed requirements that are appropriate 
    to a given device under the current state-of-the-art manufacturing for 
    that specific device. FDA investigators receive extensive training to 
    ensure uniform interpretation and application of the regulation to the 
    medical device industry. Thus, the agency does not believe that FDA 
    investigators will cite deviations from requirements not contained in 
    this part. However, as noted above, FDA has altered the language of the 
    scope to make clear that additional, unstated requirements do not 
    exist.
        4. A few comments suggested eliminating the distinction between 
    critical and noncritical devices, thus eliminating the need for 
    distinct requirements for critical devices. Other comments disagreed, 
    asserting that eliminating the distinction would increase the cost of 
    production of low-risk devices without improving their safety and 
    effectiveness.
        FDA agrees in part with the comments that suggest eliminating the 
    distinction between critical and noncritical devices and has eliminated 
    the term ``critical device'' from the scope, definitions, and 
    regulation in Secs. 820.65 Critical devices, traceability and 820.165 
    Critical devices, labeling. However, FDA has retained the concept of 
    distinguishing between devices for the traceability requirements in 
    Sec. 820.65. As addressed in the discussion under that section, FDA 
    believes that it is imperative that manufacturers be able to trace, by 
    control number, any device, or where appropriate component of a device, 
    that is intended for surgical implant into the body or to support or 
    sustain life whose failure to perform when properly used in accordance 
    with instructions for use provided in the labeling can be reasonably 
    expected to result in a significant injury to the user.
        The deletion of the terminology will bring the regulation in closer 
    harmony with ISO 9001:1994 and the quality system standards or 
    requirements of other countries.
        Finally, FDA notes that eliminating the term ``critical device'' 
    and the list of critical devices does not result in the imposition of 
    new requirements. In fact the new regulation is less prescriptive and 
    gives the manufacturer the flexibility to determine the controls that 
    are necessary commensurate with risk. The burden is on the 
    manufacturer, however, to describe the types and degree of controls and 
    how those controls were decided upon. Such determinations are made in 
    accordance with standard operating procedures (SOP's) established by 
    the manufacturer.
        5. In response to numerous comments, FDA has added the sentence 
    ``If a person engages in only some operations subject to the 
    requirements in this part, and not in others, that person need only 
    comply with those requirements applicable to the operations in which he 
    or she is engaged.'' This sentence was added to clarify the scope of 
    the regulation and
    
    [[Page 52606]]
    
    the responsibility of those who fall under this regulation. The wording 
    is the same as that used in the drug CGMP.
        6. Several comments recommended that the short list of class I 
    devices subject to design control requirements be deleted from the 
    regulation and be placed in the preamble, to allow additions or 
    deletions without requiring a change to the entire regulation. Others 
    commented that the list of class I devices should be entirely 
    eliminated to harmonize with Europe and Japan.
        FDA disagrees that the list of devices subject to design control 
    requirements should be deleted from the regulation. FDA has experienced 
    problems or has concerns with the class I devices listed and has 
    determined that design controls are needed for the listed devices. 
    Further, placing the list in the regulation establishes the 
    requirements related to those devices, and is convenient for use by 
    persons who are not familiar with, or who do not have access to, the 
    preamble. Further, FDA notes that individual sections of a regulation 
    may be revised independent of the remainder of the regulation.
        7. Numerous written comments and persons who testified at the 
    August and September 1995 meetings stated that application of the 
    regulation to component manufacturers would increase product cost, with 
    questionable value added to device safety and effectiveness, and that 
    many component suppliers would refuse to supply components or services 
    to the medical device industry. This would be especially likely to 
    occur, it was suggested, where medical device manufacturers account for 
    a small fraction of the supplier's sales.
        FDA believes that because of the complexity of many components used 
    in medical devices, their adequacy cannot always be assured through 
    inspection and testing at the finished device manufacturer. This is 
    especially true of software and software-related components, such as 
    microprocessors and microcircuits. Quality must be designed and built 
    into components through the application of proper quality systems.
        However, FDA notes that the quality system regulation now 
    explicitly requires that the finished device manufacturer assess the 
    capability of suppliers, contractors, and consultants to provide 
    quality products pursuant to Sec. 820.50 Purchasing controls. These 
    requirements supplement the acceptance requirements under Sec. 820.80. 
    Manufacturers must comply with both sections for any incoming component 
    or subassembly or service, regardless of the finished device 
    manufacturer's financial or business affiliation with the person 
    providing such products or services. FDA believes that these purchasing 
    controls are sufficient to provide the needed assurance that suppliers, 
    contractors, and consultants have adequate controls to produce 
    acceptable components.
        Therefore, balancing the many concerns of the medical device 
    industry and the agency's public health and safety concerns, FDA has 
    decided to remove the provision making the CGMP regulation applicable 
    to component manufacturers and return to the language in the original 
    CGMP regulation. This approach was unanimously endorsed by the members 
    of the GMP Advisory Committee at the September 1995 meeting. FDA will 
    continue to focus its inspections on finished device manufacturers and 
    expects that such manufacturers will properly ensure that the 
    components they purchase are safe and effective. Finished device 
    manufacturers who fail to comply with Secs. 820.50 and 820.80 will be 
    subject to enforcement action. FDA notes that the legal authority 
    exists to cover component manufacturers under the CGMP regulation 
    should the need arise.
        8. One comment stated that proposed Sec. 820.1(a)(2) should be 
    revised to include the District of Columbia and the Commonwealth of 
    Puerto Rico, as in the original CGMP regulation.
        FDA agrees with the comment. These localities were inadvertently 
    omitted and have been added to the regulation.
        9. FDA added Sec. 820.1(a)(3) on how to interpret the phrase 
    ``where appropriate'' in the regulation, as recommended by the GMP 
    Advisory Committee. This section is consistent with the statement in 
    ISO/CD 13485.
        10. Some comments on proposed Sec. 820.1(c) recommended that the 
    section be deleted as it already appears in the act. Others stated that 
    the provision implies that FDA will subject devices or persons to legal 
    action, regardless of the level of noncompliance. Still others 
    suggested that only intentional violations of the regulation should 
    give rise to regulatory action.
        FDA disagrees with these comments. The consequences of the failure 
    to comply, and the legal authority under which regulatory action may be 
    taken, are included in the regulation so that the public may be fully 
    apprised of the possible consequences of noncompliance and understand 
    the importance of compliance. FDA notes that the agency exercises 
    discretion when deciding whether to pursue a regulatory action and does 
    not take enforcement action for every violation it encounters. Further, 
    FDA generally provides manufacturers with warning prior to initiating 
    regulatory action and encourages voluntary compliance. The agency also 
    notes, however, that violations of this regulation need not be 
    intentional to place the public at serious risk or for FDA to take 
    regulatory action for such violations.
        In response to the concerns regarding the tone of the section, 
    however, the title has been changed. FDA has also deleted the specific 
    provisions referenced in the proposed section with which the failure to 
    comply would render the devices adulterated. The term ``part'' includes 
    all of the regulation's requirements.
        11. A few comments on proposed Sec. 820.1(c)(2), now Sec. 820.1(d), 
    requested that the agency clarify what is meant by requiring that 
    foreign manufacturers ``schedule'' an inspection. A few comments stated 
    that FDA was adding new requirements for foreign manufacturers in this 
    section. Others stated that the proposed language would prohibit global 
    harmonization because it would limit third party audits in place of FDA 
    inspections.
        FDA has moved the provision related to foreign manufacturers into a 
    separate section and has modified the language. The language in the 
    regulation reflects the language in section 801(a) of the act (21 
    U.S.C. 381(a)). FDA disagrees that it is adding new requirements for 
    foreign manufacturers in Sec. 820.1(d) because the section recites the 
    current requirement and standard used, and is consistent with current 
    agency policy. The agency believes that it is imperative that foreign 
    facilities be inspected for compliance with this regulation and that 
    they be held to the same high standards to which U.S. manufacturers are 
    held. Otherwise, the U.S. public will not be sufficiently protected 
    from potentially dangerous devices, and the U.S. medical device 
    industry will be at a competitive disadvantage.
        FDA intends to continue scheduling inspections of foreign 
    manufacturers in advance to assure their availability and avoid 
    conflicts with holidays and shut down periods. However, the language 
    pertaining to the ``scheduling'' of such inspections has been deleted 
    to allow flexibility in scheduling methods.
        FDA disagrees that, as written, the language would prohibit 
    inspections by third parties. FDA may use third party inspections, as 
    it uses other compliance information, in setting its priorities and 
    utilizing its resources related to foreign inspections. In this regard, 
    FDA looks forward to entering into agreements with foreign countries 
    related to CGMP
    
    [[Page 52607]]
    
    inspections that would provide FDA with reliable inspectional 
    information.
        12. Two comments stated that the section on ``Exemptions or 
    variances,'' now Sec. 820.1(e), should require that FDA provide a 
    decision on petitions within 60 days of receipt and state that the 
    agency will take no enforcement action with respect to the subject of 
    the petition until a decision is rendered. The comments said that the 
    petition process is long, arduous, and not practical.
        FDA disagrees with the comments. Currently, FDA is required by 
    section 520(f)(2)(B) of the act to respond within 60 days of receipt of 
    the petition, unless the petition is referred to an advisory committee. 
    When the 1978 CGMP regulation was published, there was a prediction 
    that FDA would be overwhelmed with petitions for exemption and variance 
    from the regulation. Over the past 18 years, since the CGMP regulation 
    first became effective, FDA has only received approximately 75 
    petitions. It is FDA's opinion that few petitions have been received 
    because of the flexible nature of the CGMP regulation. FDA has 
    attempted to write the current regulation with at least the same degree 
    of flexibility, if not more, to allow manufacturers to design a quality 
    system that is appropriate for their devices and operations and that is 
    not overly burdensome.
        Guidelines for the submission of petitions for exemption or 
    variance are available from the Division of Small Manufacturers 
    Assistance (the DSMA). The petition guidelines state that FDA will not 
    process a petition for exemption or variance while an FDA inspection of 
    a manufacturer is ongoing. Until FDA has approved a petition for an 
    exemption or variance, a manufacturer should not deviate from the 
    requirements of this regulation. FDA must first have the opportunity to 
    ensure that the manufacturer has established that an exemption or 
    variance is warranted, to carry out its obligation of ensuring that 
    devices are safe and effective.
        13. Several comments stated that the proposed requirements are not 
    necessary for all manufacturers, particularly small manufacturers with 
    few employees and low-risk devices. Other comments stated that the 
    documentation requirements are excessive.
        FDA generally disagrees with these comments. The regulation 
    provides the ``basic'' requirements for the design and manufacture of 
    medical devices. And, as noted in the previous response, the 
    requirements are written in general terms to allow manufacturers to 
    establish procedures appropriate for their devices and operations. 
    Also, as discussed above, a manufacturer need only comply with those 
    requirements applicable to the operations in which he or she is 
    engaged. However, because the regulation requirements are basic, they 
    will apply in total to most manufacturers subject to the regulation. 
    The extent of the documentation necessary to meet the regulation 
    requirements may vary with the complexity of the design and 
    manufacturing operations, the size of the firm, the importance of a 
    process, and the risk associated with the failure of the device, among 
    other factors. Small manufacturers may design acceptable quality 
    systems that require a minimum of documentation and, where possible, 
    may automate documentation. In many situations, documentation may be 
    kept at a minimum by combining many of the recordkeeping requirements 
    of the regulation, for example, the production SOP's, handling, and 
    storage procedures. When manufacturers believe that the requirements 
    are not necessary for their operations, they may petition for an 
    exemption or variance from all or part of the regulation pursuant to 
    section 520(f)(2) of the act.
        In addition, FDA has added a variance provision in Sec. 820.1(e)(2) 
    under which the agency can initiate a variance when it is in the best 
    interest of the public health. Under this provision, for instance, the 
    agency may initiate and grant a variance to manufacturers of devices 
    during times of product shortages, where the devices are needed by the 
    public and may not otherwise be made available, if such manufacturers 
    can adequately assure that the resulting devices are safe and 
    effective. The agency envisions this provision as a bridge, providing a 
    manufacturer with the time necessary to fulfill the requirements in the 
    regulation while providing important and needed devices to the public. 
    Thus, the variance would only be granted for a short period of time, 
    and only while the devices remained necessary and in short supply. 
    Under this provision, FDA will require a manufacturer to submit a plan 
    detailing the action it is taking to assure the safety and 
    effectiveness of the devices it manufactures and to meet the 
    requirements of the regulation.
        This agency initiated variance provision is in accordance with 
    section 520(f) of the act which permits, but does not require, FDA to 
    promulgate regulations governing the good manufacturing practices for 
    devices and section 701(a) of the act (21 U.S.C. 371(a)), which permits 
    FDA to promulgate regulations for the efficient enforcement of the act. 
    Because the statute does not mandate that the agency establish any 
    requirements for device CGMP's, the agency has the authority to 
    determine that the manufacturers of certain devices need not follow 
    every requirement of the regulation.
        Further, the agency initiated variance provision is in keeping with 
    the intent of Congress that FDA prevent hazardous devices from reaching 
    the marketplace, H. Rept. 853, 94th Cong., 2d sess. 25-26 (1976), and 
    the general intent of the act that the agency undertake to protect the 
    public health. The agency will only initiate such a variance where the 
    devices are needed and may not otherwise be made available, and the 
    manufacturer can assure the agency that its procedures are likely to be 
    adequate and that it is actively pursuing full compliance. The 
    variances will only be in effect for a limited time.
        Section 820.1(e) has been modified to include the above addition, 
    to reflect the title change of the regulation, and to provide the most 
    current address for the DSMA.
    ii. Definitions (Sec. 820.3)
        14. Several comments were received regarding the definition of 
    ``complaint.'' Comments generally believed that the definition was 
    unclear and could be interpreted to include routine service requests, 
    communications from customers unrelated to the quality, safety, or 
    effectiveness of the device, and internal communications.
        FDA agrees with the comments in part and has modified the 
    definition to make clear that a communication would be considered a 
    ``complaint'' only if the communication alleged some deficiency related 
    to the identity, quality, durability, reliability, safety, 
    effectiveness, or performance of the device after it is released for 
    distribution. The definition is now very similar to the definition used 
    in ISO/CD 13485.
        The regulation addresses service requests and in-house indications 
    of dissatisfaction under Sec. 820.100 Corrective and preventive action. 
    This section requires manufacturers to establish procedures to identify 
    quality problems and process the information received to detect and 
    correct quality problems. Information generated in-house relating to 
    quality problems should be documented and processed as part of this 
    corrective and preventive action program.
        With respect to service requests, Sec. 820.200 Servicing states 
    that a service report that represents an event which
    
    [[Page 52608]]
    
    must be reported to the FDA under part 803 or 804 (21 CFR part 803 or 
    804) shall automatically be considered a complaint. All other service 
    reports must be analyzed for trends or systemic problems and when 
    found, these trends or systemic problems must be investigated according 
    to the provisions of Sec. 820.100 Corrective and preventive action.
        15. One comment suggested that the agency delete the phrase ``used 
    during device manufacturing'' in the definition of ``component'' 
    because it was confusing and may cause problems with certain aspects of 
    distributor operations.
        FDA agrees and has deleted the words ``used during device 
    manufacturing'' from the definition because it was not intended to 
    differentiate between distributors and manufacturers. Further, FDA 
    deleted the term ``packaging'' to clarify that every piece of packaging 
    is not necessarily a component. Only the materials that are part of the 
    ``finished, packaged, and labeled device'' are considered to be 
    components.
        16. Several comments stated that the term ``complete history'' in 
    the definition of ``control number'' should be clarified or deleted 
    because it is unclear what a complete production history is, and the 
    term could be construed to require full traceability for all component 
    lots of any product containing a control number.
        FDA agrees in part with the comments. The control number is the 
    means by which the history of the device, from purchase of components 
    and materials through distribution, may be traced, where traceability 
    is required. The definition does not require that a manufacturer be 
    able to trace the device whenever control numbers are used. In fact, 
    the definition itself does not establish any requirements. The agency 
    notes, however, that the manufacturer's traceability procedures should 
    ensure that a complete history of the device, including environmental 
    conditions which could cause the device to fail to conform to its 
    specified requirements, can be traced and should facilitate 
    investigation of quality problems and corrective action. FDA notes, 
    however, that the level of detail required for this history is 
    dependent on the nature of the device, its intended use, and its 
    complexity. Therefore, FDA has removed the term ``complete'' in the 
    definition for clarity and flexibility.
        FDA has also amended the definition for added flexibility, to state 
    that symbols may be used and has included the term ``unit'' for any 
    device that is not manufactured as a lot or batch.
        17. The definition of ``critical device'' has been deleted for the 
    reasons discussed above.
        18. Several comments stated that the term ``design history record'' 
    should be changed because the acronym for the term is the same as that 
    for device history record (the DHR). Other comments said the ``design 
    history record'' should not need to contain documentation of a 
    ``complete'' design history. One comment stated that the definition 
    should allow reference to records containing the design history of the 
    device. A few comments stated that the term should be deleted 
    altogether because it is redundant with the definition of device master 
    record (the DMR).
        FDA agrees in part with these comments and has changed the term 
    ``design history record'' to ``design history file.'' In addition, FDA 
    has amended the provisions to require that the file describe the design 
    history, as it may not be necessary to maintain a record of every step 
    in the design phase, although the ``entire history'' should be apparent 
    from the document. Section 820.30(j) further delineates what should be 
    in the design history file (the DHF), specifically records sufficient 
    to verify that the design was developed in accordance with the design 
    and development plan and other applicable design requirements of the 
    regulation.
        FDA does not agree that the definitions of the DHF and the DMR are 
    redundant. The DHF for each type of device should include, for example, 
    the design and development plan, design review results, design 
    verification results, and design validation results, as well as any 
    other data necessary to establish compliance with the design 
    requirements. The DMR should contain all of the procedures related to 
    each type of device as required by this part and the most current 
    manufacturing specifications of the device, once the design 
    specifications have been transferred into production.
        19. One comment on ``design input'' stated it was confused by the 
    term ``requirements'' and wanted to know whose requirements are 
    encompassed in this definition.
        The term ``requirement'' is meant in the broadest sense, to 
    encompass any internally or externally imposed requirements such as 
    safety, customer-related, and regulatory requirements. All of these 
    requirements must be considered as design inputs. How these 
    requirements are handled and dealt with is up to the manufacturer.
        20. Two comments stated that the definition of ``design output'' 
    should be revised because it is not necessary, and would be burdensome, 
    to keep records of and review the ``results of a design effort at each 
    design phase and at the end.'' Other comments suggested that the design 
    output definition should be restricted to physical characteristics of 
    the device.
        FDA agrees in part, but has not deleted the phrase ``results of a 
    design effort at each design phase and at the end'' from the 
    definition. The intent was not to dictate when design phases would 
    occur. Such phases will be defined in the design and development plan. 
    For example, a manufacturer may only have a few design phases for a new 
    type of syringe. Thus, design output would be the results of those few 
    efforts. The results of each design phase constitute the total design 
    output. The definition has been amended, however, to clarify that the 
    finished design output is the basis for the DMR.
        FDA disagrees with the comments that suggest that the design output 
    should be restricted to physical characteristics of the device. Design 
    output is more than just the device specifications. Design output 
    includes, among other things, the specifications for the manufacturing 
    process, the quality assurance testing, and the device labeling and 
    packaging. It is important to note that the design effort should not 
    only control the design aspects of the device during the original 
    development phase, but also all subsequent design and development 
    activities including any redesign or design changes after the original 
    design is transferred to production.
        21. A few comments on the definition of ``design review'' stated 
    that proposing solutions to problems is not part of the design review 
    activity. Two other comments expressed concern that the definition 
    would require that each design review be ``comprehensive.'' 
        In response to the comments on the proper role of design review, 
    FDA agrees that the design review participants are typically not 
    responsible for establishing solutions, although they may do so in many 
    small operations. The definition has been amended, but FDA wants to 
    make clear that although the design review participants need not 
    propose solutions, they should ensure that solutions to any identified 
    problems are adequate and implemented appropriately.
        Regarding the scope of design review, each design review need not 
    be ``comprehensive'' for the entire design process but must be 
    ``comprehensive'' for the design phase being reviewed. However, at the 
    end of the design process when the design is transferred
    
    [[Page 52609]]
    
    to production, all aspects of the design process should have been 
    reviewed.
        A few other changes were made to harmonize with the definition in 
    ISO 8402:1994 ``Quality--Vocabulary.''
        22. Comments on the definition of ``device master record'' pointed 
    out that the definition is not consistent with the requirements of 
    Sec. 820.181 Device master record. Other comments stated that the 
    definition should allow reference to records. One comment stated that 
    ``all'' procedures related to a specific finished device need not be 
    included in the DMR, such as the procedures for the design and 
    development, since they may be in the DHF.
        FDA agrees in part with the comments that found the DMR definition 
    and requirements to be inconsistent and has amended the definition to 
    be consistent with the requirements set forth in Sec. 820.181. FDA does 
    not believe, however, that it is necessary to modify the definition to 
    include the referencing of records because the DMR requirements in 
    Sec. 820.181 state that the DMR ``shall include or refer to the 
    location of'' the required information. FDA agrees that the term 
    ``all'' is not necessary and has deleted it in order to give 
    manufacturers the necessary flexibility.
        23. The definition for the term ``end-of-life'' was added to the 
    Working Draft because this term was used in the definitions for 
    ``refurbisher'' and ``servicing'' to help distinguish the activities of 
    refurbishing from those of servicing. FDA determined that such a 
    distinction was necessary, due to comments and ongoing confusion 
    regarding the difference between the two functions, and the different 
    requirements applicable to the functions.
        Many written comments and persons who testified at the August and 
    September 1995 meetings stated that the term was confusing, 
    unnecessary, and introduced many new legal and liability issues. FDA 
    agrees with these comments and has deleted the term throughout the 
    regulation. FDA has also deleted definitions for ``refurbisher'' and 
    ``servicing'' for the reasons discussed below.
        24. The few comments received on the definition of ``establish'' 
    indicated a concern that the regulation requires too much documentation 
    and is more onerous than ISO 9001 requirements.
        FDA disagrees with the comments. The term ``establish'' is only 
    used where documentation is necessary. FDA also notes that the quality 
    system regulation is premised on the theory that adequate written 
    procedures, which are implemented appropriately, will likely ensure the 
    safety and effectiveness of the device. ISO 9001:1994 relies on the 
    same premise. The 1994 version of ISO 9001 broadly requires the 
    manufacturer to ``establish, document, and maintain a quality system,'' 
    which includes documenting procedures to meet the requirements.
        The definition has been amended, however, in response to general 
    comments received, to clarify that a ``document'' may be in writing or 
    on electronic media, to allow flexibility for any type of recorded 
    media.
        25. FDA received comments questioning the inclusion of a device 
    that is intended to be sterile, but that is not yet sterile, in the 
    definition of ``finished device.'' A few comments stated that ``capable 
    of functioning'' is ambiguous, and ``suitable for use'' is not 
    necessary. Another comment requested that the term ``accessory'' be 
    defined.
        FDA disagrees with the comments, but has amended the definition for 
    clarification. Since the 1978 CGMP regulation was promulgated, FDA has 
    been repeatedly asked whether devices intended to be sold as sterile 
    are considered subject to the CGMP requirements, even though they have 
    not yet been sterilized. The agency had intended the new definition to 
    make explicit the application of the regulation to the manufacture of 
    sterile devices that have yet to be sterilized. Although FDA believes 
    it should be obvious that such devices are subject to CGMP 
    requirements, some manufacturers have taken the position that the 
    regulation does not apply because the device is not ``finished'' or 
    ``suitable for use'' until it has been sterilized.
        To better clarify its intent, FDA has amended the definition to add 
    that all devices that are capable of functioning, including those 
    devices that could be used even though they are not yet in their final 
    form, are ``finished devices.'' For example, devices that have been 
    manufactured or assembled, and need only to be sterilized, polished, 
    inspected and tested, or packaged or labeled by a purchaser/
    manufacturer are clearly not components, but are now in a condition in 
    which they could be used, therefore meeting the definition of 
    ``finished device.''
        The distinction between ``components'' and ``finished devices'' was 
    not intended to permit manufacturers to manufacture devices without 
    complying with CGMP requirements by claiming that other functions, such 
    as sterilization, incoming inspection (where sold for subsequent minor 
    polishing, sterilization, or packaging), or insertion of software, will 
    take place. The public would not be adequately protected in such cases 
    if a manufacturer could claim that a device was not a ``finished'' 
    device subject to the CGMP regulation because it was not in its 
    ``final'' form.
        The phrase ``for commercial distribution'' was deleted from the 
    proposed definition of ``finished device'' because it is not necessary 
    for a device to be in commercial distribution to be considered a 
    finished device. Further, FDA notes that the term ``accessory'' is 
    described in Sec. 807.20(a)(5) (21 CFR 807.20(a)(5)).
        26. Two comments on the definition of ``lot or batch'' requested 
    that the definition be clarified: One to reflect that single units may 
    be produced for distribution, the other to indicate that what 
    constitutes a lot or a batch may vary depending on the context.
        In response to the comments, FDA has modified the definition to 
    make clear that a lot or batch may, depending on circumstances, be 
    comprised of one finished device. Whether for inspection or for 
    distribution, a lot or batch is determined by the factors set forth in 
    the definition; of course, a manufacturer may determine the size of the 
    lot or batch, as appropriate.
        27. Several comments received on the definition of ``executive 
    management'' objected that the definition is inconsistent with ISO 
    9001. Others thought that FDA should better define the level of 
    management the term was intended to describe.
        FDA agrees with both concerns and has modified the definition by 
    deleting the second half, which appeared to bring executive authority 
    and responsibility too far down the organization chart. The term was 
    intended to apply only to management that has the authority to bring 
    about change in the quality system and the management of the quality 
    system. Although such management would clearly have authority over, for 
    example, distribution, those who may have delegated management 
    authority over distribution would not necessarily have authority over 
    the quality system and quality policy. Accordingly, the definition has 
    been modified to include only those who have the authority and 
    responsibility to establish and make changes to the quality policy and 
    quality system. It is the responsibility of top management to establish 
    and communicate the quality policy. In addition, the term ``executive 
    management'' has been changed to ``management with executive 
    responsibility,'' to harmonize with ISO 9001:1994.
        28. Several comments in response to the proposed definition of
    
    [[Page 52610]]
    
    ``manufacturer'' stated that refurbishers and servicers should be added 
    to the definition of a ``manufacturer.'' Other comments recommended 
    adding the term ``remanufacturer.'' Other comments requested deletion 
    of contract sterilizers, installers, specification developers, 
    repackagers, relabelers, and initial distributors from the definition. 
    One comment stated that the phrase ``processes a finished device'' 
    should be explained in the definition of manufacturer.
        FDA's Compliance Policy Guide (CPG) 7124.28 contains the agency's 
    policy regarding the provisions of the act and regulations with which 
    persons who recondition or rebuild used devices are expected to comply. 
    This CPG is in the process of being revised in light of FDA's 
    experience in this area. FDA is not including the terms ``servicer'' or 
    ``refurbisher,'' as they relate to entities outside the control of the 
    original equipment manufacturer, in this final regulation, even though 
    it believes that persons who perform such functions meet the definition 
    of manufacturer. Because of a number of competitive and other issues, 
    including sharply divided views by members of the GMP Advisory 
    Committee at the September 1995 meeting, FDA has elected to address 
    application of the CGMP requirements to persons who perform servicing 
    and refurbishing functions outside the control of the original 
    manufacturer in a separate rulemaking later this year, with another 
    opportunity for public comment.
        FDA agrees that the term ``remanufacturing'' should be added to the 
    definition of ``manufacturer'' and has separately defined the term. A 
    remanufacturer is defined as ``any person who processes, conditions, 
    renovates, repackages, restores, or does any other act to a finished 
    device that significantly changes the finished device's performance or 
    safety specifications, or intended use.''
        However, FDA disagrees that contract sterilizers, installers, 
    specification developers, repackagers, relabelers, and initial 
    distributors should be deleted from the definition, primarily because 
    all such persons may have a significant effect on the safety and 
    effectiveness of a device and on the public health. All persons who 
    perform these functions meet the definition of manufacturer, and 
    therefore should be inspected to ensure that they are complying with 
    the applicable provisions. For example, a specification developer 
    initiates the design requirements for a device that is manufactured by 
    a second party for subsequent commercial distribution. Such a developer 
    is subject to design controls. Further, those that perform the 
    functions of contract sterilization, installation, relabeling, 
    remanufacturing, and repacking have routinely been considered to be 
    manufacturers under the original CGMP definition, and the agency has 
    treated them as such by inspecting them to ensure that they comply with 
    the appropriate portions of the original CGMP. By explicitly including 
    them in the definition of ``manufacturer'' the agency has simply 
    codified its longstanding policy and interpretation of the original 
    regulation.
        The phrase ``processes a finished device'' applies to a finished 
    device after distribution. Again, this phrase has been part of the CGMP 
    regulation definition of ``manufacturer'' for 18 years.
        29. A number of comments on the definition of ``manufacturing 
    material,'' and on other parts of the proposal containing requirements 
    for ``manufacturing material,'' stated that while the control of 
    manufacturing material is important, it need not be as extensive as 
    required throughout the regulation. Other comments stated that the 
    meaning of the phrase ``or other byproducts of the manufacturing 
    process'' is unclear, and should be deleted. One comment suggested that 
    the definition be modified to separate the definition from the 
    examples.
        FDA agrees that, depending on the manufacturing material and the 
    device, the degree of control that is needed will vary. FDA believes 
    that manufacturing materials must be assessed, found acceptable for 
    use, and controlled. Therefore, the regulation requires manufacturers 
    to assess, assure acceptability of, and control manufacturing materials 
    to the degree necessary to meet the specified requirements. The agency 
    notes that international standards such as ISO 8402:1994 include 
    manufacturing material in their definition of ``product,'' to which all 
    requirements apply, and notes that FDA has added the same definition in 
    Sec. 820.3(r) in its effort toward harmonization.
        FDA amended the definition of manufacturing material to read ``a 
    concomitant constituent, or a byproduct constituent produced during the 
    manufacturing process'' to help clarify this definition. These terms 
    refer to those materials or substances that naturally occur as a part 
    of the material or during the manufacturing process which are intended 
    to be removed or reduced in the finished device. For example, some 
    components, such as natural rubber latex, contain allergenic proteins 
    that must be reduced or removed from the finished devices. The 
    definition has been modified to include ``concomitant constituents'' to 
    clarify the meaning.
        In addition to clarifying the definition, FDA has deleted the 
    specific examples. Therefore, FDA notes that cleaning agents, mold 
    release agents, lubricating oils, latex proteins, and sterilant 
    residues are just some examples of manufacturing materials.
        30. The comments received on the definition for ``nonconforming'' 
    conveyed a general sense that the definition was confusing, with 
    various comments suggesting that different parts of the definition 
    should be deleted and one suggesting that the definition be deleted 
    altogether.
        In response to these comments, the definition of ``nonconforming'' 
    has been deleted. However, the definition from ISO 8402:1994 for 
    ``nonconformity'' was added to ensure that the requirements in the 
    regulation, especially those in Secs. 820.90 Nonconforming product and 
    820.100 Corrective and preventive action are understood. FDA emphasizes 
    that a ``nonconformity'' may not always rise to the level of a product 
    defect or failure, but a product defect or failure will typically 
    constitute a nonconformity.
        31. Several comments requested various revisions to the definition 
    of ``production'' to make it more clear, and one thought that it was a 
    common term and should be deleted.
        In response, FDA has deleted the definition for ``production'' 
    because it should be commonly understood.
        As noted in response to comments on the definition of manufacturing 
    material, FDA has added a definition of ``product'' to conform to the 
    definition in ISO 8402:1994 and to avoid the necessity of repeating the 
    individual terms throughout the regulation. Whenever a requirement is 
    not applicable to all types of product, the regulation specifically 
    states the product(s) to which the requirement is applicable.
        It should be noted that the regulation has acceptance requirements 
    for incoming ``product'' and other requirements for ``product,'' which 
    by definition includes manufacturing materials. Manufacturing materials 
    should be controlled in a manner that is commensurate with their risk 
    as discussed above. However, for manufacturing materials that are 
    ``concomitant constituents,'' FDA realizes that incoming acceptance, 
    identification, etc., may not be feasible. The important control 
    measure for ``concomitant constituents'' is the
    
    [[Page 52611]]
    
    reduction or removal requirement found in Sec. 820.70(h).
        32. A few comments stated that the definition of ``quality'' should 
    be changed to be identical to ISO 8402. Others stated that the 
    terminology adopted from ISO 8402, ``that bear on,'' is too broad and 
    could cover every potential and imaginable factor. Still others wanted 
    to add the phrase, ``as defined by the manufacturer'' to the end of the 
    sentence.
        FDA disagrees with the comments and believes that the definition is 
    closely harmonized to that in ISO 8402:1994. FDA believes that the 
    definition appropriately defines quality in the context of a medical 
    device and believes that the phrase from ISO 8402:1994, ``stated and 
    implied needs,'' has the same meaning as the phrase ``fitness-for-use, 
    including safety and performance'' in the context of the Quality System 
    regulation. Further, ``quality'' is not just ``defined by the 
    manufacturer'' but is also defined by customer need and expectation.
        33. Many comments received on the ``quality audit'' definition 
    suggested that the definition should not state that it is an 
    examination of the ``entire'' quality system because that would require 
    that every audit include the ``entire'' quality system. Other comments 
    on ``quality audit'' stated that it is unclear what is meant by the 
    last sentence of the proposed definition, namely, that `` `[q]uality 
    audit' is different from * * * other quality system activities required 
    by or under this part.''
        FDA agrees that while the quality audit is an audit of the 
    ``entire'' quality system, audits may be conducted in phases, with some 
    areas requiring more frequent audits than other areas, and that each 
    audit need not review the whole system. The frequency of internal 
    quality audits should be commensurate with, among other things, the 
    importance of the activity, the difficulty of the activity to perform, 
    and the problems found. To avoid any misunderstanding, the word 
    ``entire'' before quality system has been deleted.
        FDA emphasizes that if conducted properly, internal quality audits 
    can prevent major problems from developing and provide a foundation for 
    the management review required by Sec. 820.20(c), ``Management 
    review.''
        In response to the confusion about the last sentence of the 
    proposed definition, FDA has deleted the last sentence. The purpose of 
    the sentence was to clarify that the internal audit requirement is 
    different from, and in addition to, the requirements for establishing 
    quality assurance procedures and recording results. On occasion, 
    manufacturers have attempted to prevent FDA investigators from 
    reviewing such quality assurance procedures and results (for example, 
    trend analysis results) by stating that they are part of the internal 
    quality audit report and not subject to review during a CGMP 
    inspection. FDA disagrees with this position. To clarify which records 
    are exempt from routine FDA inspection, FDA has added Sec. 820.180(c).
        34. One comment said that the word ``executive'' should be deleted 
    from the definition of ``quality policy'' because quality policy should 
    be supported by all personnel, not just those in executive management. 
    A few comments stated that ``formally expressed'' should be deleted 
    because it is incompatible with the requirements in Sec. 820.20(a) and 
    (c) which require that the quality policy be ``established.'' Other 
    comments stated that the ``quality'' before ``intentions'' was 
    tautological.
        FDA agrees that all company personnel must follow the quality 
    policy. However, the definition is intended to make clear that the 
    quality policy must be established by top management. Therefore it has 
    been retained. The term ``executive management'' has been modified to 
    ``management with executive responsibility'' to be consistent with the 
    revised ISO 9001:1994. FDA agrees with the remaining comments and has 
    changed ``formally expressed'' to ``established'' for consistency and 
    has deleted the ``quality'' before ``intentions.''
        35. A few comments suggested using the definition of ``quality 
    systems'' from ISO 8402 and 9001. Other comments on the definition of 
    ``quality system'' said that the term ``quality management'' should be 
    defined.
        FDA agrees in part with the comments. The term ``specifications'' 
    has been deleted to harmonize the definition with ISO 8402:1994. FDA 
    does not agree that the term ``quality management'' must be defined. A 
    definition can be found in ISO 8402:1994 that is consistent with FDA's 
    use of the term.
        36. Many comments on the definition of ``record'' were received. 
    Some thought the term was too broad, giving FDA access to all documents 
    and exceeding FDA's inspection authority. Others thought that the 
    definition of ``record'' would tremendously increase the recordkeeping 
    burden. Several comments recommended that FDA adopt the ISO definition.
        The definition of ``record'' was deleted because it seemed to add 
    more confusion than clarity. The definition was intended to clarify 
    that ``records'' may include more than the traditional hardcopy 
    procedures and SOP's, for example, plans, notes, forms, data, etc. FDA 
    was trying to clarify that ``records'' could be written, electronic, 
    optical, etc., as long as they could be stored and controlled. FDA 
    could not adopt the ISO 8402:1994 definition because of how the term 
    ``record'' is used in the act, which is broader than the ISO 
    definition. Therefore, FDA will allow the act and case law to continue 
    to define the term.
        37. The definition in the Working Draft of ``refurbisher'' was 
    deleted and will be addressed in the separate rulemaking described 
    above.
        38. FDA added the definition of ``remanufacturer'' to codify FDA's 
    longstanding policy and interpretation of the original CGMP. The 
    language is consistent with the 510(k) provisions and the premarket 
    approval amendment/supplement requirements, because FDA has always 
    considered remanufacturers in fact to be manufacturers of a new device.
        39. Several comments on the definition of ``reprocessing'' 
    requested clarification of the difference between that term and 
    ``refurbishing.'' Several other comments on the definition of 
    ``reprocessing'' stated that FDA should clarify that ``reprocessing'' 
    is an activity performed before a device is distributed. Others 
    commented that the term ``rework'' should be used instead of the term 
    ``reprocessing,'' to be consistent with ISO terminology.
        FDA agrees with the comments and has changed the term to 
    ``rework,'' adopted the ISO 8402:1994 definition, and added that 
    ``rework'' is performed according to specified DMR requirements before 
    the device is released for distribution.
        40. A few comments stated that including the term ``maintenance'' 
    in the proposed definition of ``servicing'' implies that preventative 
    maintenance would be subject to the regulation. Other comments said 
    that it may not be desirable to return old devices or devices that have 
    received field modifications to the original specifications. Therefore, 
    the comments suggested deleting the last part of the definition that 
    states that ``servicing'' is returning a device to its specifications.
        FDA has deleted the definition of ``servicing'' and has not added a 
    definition of ``servicer'' because this will be covered in the separate 
    rulemaking discussed above. FDA notes, however, that servicing 
    performed by manufacturers and remanufacturers is subject to the 
    requirements in Sec. 820.200 Servicing. These requirements are a 
    codification of longstanding interpretations of the original CGMP,
    
    [[Page 52612]]
    
    Sec. 820.20(a)(3), and current agency policy.
        41. Several comments were received on the proposed definition of 
    ``special process.'' Many asked for clarification or adoption of the 
    ISO definition. Some stated that it is impossible to completely verify 
    processes in every instance.
        FDA has deleted the definition because the term ``special process'' 
    is no longer used in ISO 9001:1994, except in a note. FDA has, however, 
    modified the requirements of the regulation to reflect that, in many 
    cases, testing and inspecting alone may be insufficient to prove the 
    adequacy of a process. One of the principles on which the quality 
    systems regulation is based is that all processes require some degree 
    of qualification, verification, or validation, and manufacturers should 
    not rely solely on inspection and testing to ensure processes are 
    adequate for their intended uses.
        42. Several comments on the definition of ``specification'' 
    suggested that the term should not apply to quality system 
    requirements. One comment suggested that the phrase ``other activity'' 
    be deleted because it is too broad. Another comment noted that the 
    definition in ISO 9001 pertains to requirements, not only documents.
        In response, FDA has amended the definition to make clear that it 
    applies to the requirements for a product, process, service, or other 
    activity. The reference to the quality system has been deleted. FDA 
    disagrees that the definition is too broad and has not deleted the term 
    ``other activity'' because a specification can be developed for 
    anything the manufacturer chooses. FDA notes, however, that ISO 
    9001:1994 does not contain a definition for ``specification'' but uses 
    the definition found in ISO 8402:1994.
        43. Numerous comments were received on the definitions of 
    ``validation'' and ``verification.'' Almost all stated that the two 
    definitions overlapped and that there was a need to rewrite the 
    definitions to prevent confusion. Many suggested that the ISO 
    definitions be adopted. Others stated that there was a need to 
    distinguish between design validation and process validation.
        FDA agrees with the comments and has rewritten the two definitions 
    to better reflect the agency's intent. FDA has adopted the ISO 
    8402:1994 definition of validation. ``Validation'' is a step beyond 
    verification to ensure the user needs and intended uses can be 
    fulfilled on a consistent basis. FDA has further distinguished 
    ``process validation'' from ``design validation'' to help clarify these 
    two types of ``validation.'' The ``process validation'' definition 
    follows from FDA's ``Guidelines on General Principles of Process 
    Validation'' (Ref. 10). The definition for ``design validation'' is 
    consistent with the requirements contained in Sec. 820.30 Design 
    controls.
        The ISO 8402:1994 definition of ``verification'' has been adopted. 
    ``Verification'' is confirmation by examination and provision of 
    objective evidence that specified requirements for a particular device 
    or activity at hand have been met.
    iii. Quality System (Sec. 820.5)
        44. Several comments suggested that the requirement should be more 
    general, in that the requirement that devices be safe and effective is 
    covered elsewhere in the regulation. The comments recommended that the 
    quality system requirements be harmonized with international standards 
    and focus on requiring that a system be established that is appropriate 
    to the specific device and that meets the requirements of the 
    regulation.
        FDA agrees in part with the comments and has modified the language 
    as generally suggested by several comments to require that the quality 
    system be ``appropriate for the specific medical device(s) designed or 
    manufactured, and [] meet[] the requirements of this part.'' This is 
    essentially the requirement of the original CGMP regulation with the 
    added reference to design control.
        The requirements that effective quality system instructions and 
    procedures be established and effectively maintained are retained; 
    however, they were moved to Sec. 820.20(b)(3)(i). As previously noted, 
    the quality system regulation is premised on the theory that the 
    development, implementation, and maintenance of procedures designed to 
    carry out the requirements will assure the safety and effectiveness of 
    devices. Thus, the broad requirements in Sec. 820.5 are in a sense the 
    foundation on which the remaining quality system requirements are 
    built.
    
    B. Quality System Requirements (Subpart B)
    
    i. Management Responsibility (Sec. 820.20)
        45. Several comments on Sec. 820.20(a), ``Quality policy,'' related 
    to the use of the term ``executive management.'' A few comments stated 
    that quality system development and implementation are the 
    responsibility of the chief executive officer, but how he or she 
    chooses to discharge the responsibility should be left to the 
    discretion of the manufacturer. Other comments stated that the 
    requirement that executive management ensure that the quality policy is 
    understood is impossible and should be deleted or rewritten.
        FDA agrees in part with the comments. In response to the comments, 
    FDA has deleted the term ``executive management'' and replaced it with 
    ``management with executive responsibility,'' which is consistent with 
    ISO 9001:1994. Management with executive responsibility is that level 
    of management that has the authority to establish and make changes to 
    the company quality policy. The establishment of quality objectives, 
    the translation of such objectives into actual methods and procedures, 
    and the implementation of the quality system may be delegated. The 
    regulation does not prohibit the delegation. However, it is the 
    responsibility of the highest level of management to establish the 
    quality policy and to ensure that it is followed. (See United States v. 
    Dotterweich, 320 U.S. 277 (1943), and United States v. Park, 421 U.S. 
    658 (1975).)
        For this reason, FDA disagrees that the requirement that management 
    ensure that the quality policy is understood should be deleted. It is 
    without question management's responsibility to undertake appropriate 
    actions to ensure that employees understand management's policies and 
    objectives. Understanding is a learning process achieved through 
    training and reinforcement. Management reinforces understanding of 
    policies and objectives by demonstrating a commitment to the quality 
    system visibly and actively on a continuous basis. Such commitment can 
    be demonstrated by providing adequate resources and training to support 
    quality system development and implementation. In the interest of 
    harmonization, the regulation has been amended to be very similar to 
    ISO 9001:1994.
        46. A few comments stated that the words ``adequate'' and 
    ``sufficient'' should be deleted from Sec. 820.20(b) ``Organization,'' 
    as they are subjective and too difficult to define. One comment thought 
    that the general requirements in the paragraphs are addressed by 
    Sec. 820.25 Personnel. Another comment stated that ``designed'' should 
    be added prior to ``produced'' for consistency with the scope.
        FDA agrees that the requirement for ``sufficient personnel'' is 
    covered in Secs. 820.20(b)(2), ``Resources,'' and 820.25 Personnel, 
    both of which require manufacturers to employ sufficient personnel with 
    the training and
    
    [[Page 52613]]
    
    experience necessary to carry out their assigned activities properly. 
    The phrase is, therefore, deleted. However, FDA has retained the 
    requirement for establishing an ``adequate organizational structure'' 
    to ensure compliance with the regulation, because such an 
    organizational structure is fundamental to a manufacturer's ability to 
    produce safe and effective devices. The organizational structure should 
    ensure that the technical, administrative, and human factors functions 
    affecting the quality of the device will be controlled, whether these 
    functions involve hardware, software, processed materials, or services. 
    All such control should be oriented towards the reduction, elimination, 
    or ideally, prevention of quality nonconformities. Further, the agency 
    does not believe that the term is ambiguous. The organizational 
    structure established will be determined in part by the type of device 
    produced, the manufacturer's organizational goals, and the expectations 
    and needs of customers. What may be an ``adequate'' organizational 
    structure for manufacturing a relatively simple device may not be 
    ``adequate'' for the production of a more complex device, such as a 
    defibrillator. FDA has also added ``designed'' prior to ``produced'' to 
    be consistent with the scope of the regulation.
        47. A number of comments on proposed Sec. 820.20 (b)(1)(i) through 
    (b)(1)(v), ``Responsibility and authority,'' objected to the section, 
    stating that it was too detailed and confusing and that the wording was 
    redundant with other sections of the proposal.
        FDA agrees generally with the comments in that the proposed 
    paragraphs set forth examples of situations in which independence and 
    authority are important. Therefore, the examples provided in 
    Sec. 820.20 (b)(1)(i) through (b)(1)(v) are deleted. However, FDA has 
    retained the broad requirement that the necessary independence and 
    authority be provided as appropriate to every function affecting 
    quality. FDA emphasizes that it is crucial to the success of the 
    quality system for the manufacturer to ensure that responsibility, 
    authority, and organizational freedom (or independence) is provided to 
    those who initiate action to prevent nonconformities, identify and 
    document quality problems, initiate, recommend, provide, and verify 
    solutions to quality problems, and direct or control further 
    processing, delivery, or installation of nonconforming product. 
    Organizational freedom or independence does not necessarily require a 
    stand-alone group, but responsibility, authority, and independence 
    should be sufficient to attain the assigned quality objectives with the 
    desired efficiency.
        48. Several comments on proposed Sec. 820.20(b)(2), ``Verification 
    resources and personnel,'' stated that requiring ``adequately'' trained 
    personnel was subjective and that the section was not consistent with 
    ISO 9001.
        FDA agrees that the section is not consistent with ISO 9001, and 
    has adopted the language used in ISO 9001:1994, section 4.1.2.2, 
    ``Resources,'' and has renamed the section ``Resources.'' The provision 
    is now a broad requirement that the manufacturer provide adequate 
    resources for the quality system and is not restricted to the 
    verification function. FDA acknowledges that Sec. 820.25(a), 
    ``General,'' requires that sufficiently trained personnel be employed. 
    However, Sec. 820.20(b)(2), ``Resources,'' emphasizes that all resource 
    needs must be provided for, including monetary, supplies, etc., as well 
    as personnel resources. In contrast, Sec. 820.25(a) specifically 
    addresses education, background, training, and experience requirements 
    for personnel.
        49. Comments on Sec. 820.20(b)(3), ``Management representative,'' 
    stated that the management representative should not be limited to 
    ``executive'' management. A few comments stated that the appointment 
    should be documented. In addition, a few comments from proposed 
    Sec. 820.5 stated that the terms ``effective'' and ``effectively'' 
    should be defined.
        The agency agrees that the responsibility need not be assigned to 
    ``executive'' management and has modified the requirement to allow 
    management with executive responsibility to appoint a member of 
    management. When a member of management is appointed to this function, 
    potential conflicts of interest should be examined to ensure that the 
    effectiveness of the quality system is not compromised. In addition, in 
    response to many comments, the requirement was amended to make clear 
    that the appointment of this person must be documented, moving the 
    requirement up from Sec. 820.20(b)(3)(ii). The amended language is 
    consistent with ISO 9001:1994. Further, FDA has amended this section to 
    change ``executive management'' to ``management with executive 
    responsibility'' for consistency with the definition.
        The terms ``effective'' and ``effectively'' are no longer used in 
    Sec. 820.5 but ``effectively'' is found in Sec. 820.20(b)(3)(i). FDA 
    does not believe that these terms require a definition. Instructions 
    and procedures must be defined, documented, implemented, and maintained 
    in such a way that the requirements of this part are met. If they are, 
    they will be ``effective.''
        50. A few comments stated that the improvement of the quality 
    system is not a requirement under the act and the reference to such 
    improvement in Sec. 820.20(b)(3)(ii) should, therefore, be deleted.
        FDA agrees in part with the comments and has deleted the 
    requirement that the person appointed under this section provide 
    information for improving the quality system. The provision implied 
    that the manufacturer must go beyond the requirements of the 
    regulation. FDA notes, however, that information collected in complying 
    with Secs. 820.20(b)(3)(ii) and 820.100 Corrective and preventive 
    action, should be used not only for detecting deficiencies and for 
    subsequent correction of the deficiencies but also to improve the 
    device and quality system.
        51. Many comments stated that the report required by 
    Sec. 820.20(c), ``Management review,'' should not be subject to FDA 
    review, due to the same liability and self-incrimination concerns 
    related to the internal audit.
        FDA agrees in part with the comments. The proposed regulation did 
    not state FDA's intentions with respect to inspectional review of the 
    results of the required management review. After careful consideration 
    of the comments, FDA agrees that it will not request to inspect and 
    copy the reports of reviews required by Sec. 820.20(c) when conducting 
    routine inspections to determine compliance with this part. FDA 
    believes that refraining from routinely reviewing these reports may 
    help ensure that the audits are complete and candid and of maximum use 
    to the manufacturer. However, FDA believes that it is important that 
    the dates and results of quality system reviews be documented, and FDA 
    may require that management with executive responsibility certify in 
    writing that the manufacturer has complied with the requirements of 
    Sec. 820.20(c). FDA will also review the written procedures required by 
    Sec. 820.20(c), as well as all other records required under 
    Sec. 820.20.
        52. A few comments stated that the management review should not be 
    dictated by established review procedures because management level 
    employees should be fully capable of reviewing documents without a 
    written procedure.
        As noted above, FDA has retained the requirement for establishing 
    procedures to conduct the required management review in Sec. 820.20(c). 
    FDA believes that
    
    [[Page 52614]]
    
    a manufacturer can establish procedures flexible enough for management 
    to vary the way in which a review is conducted, as appropriate. 
    Procedures should require that the review be conducted at appropriate 
    intervals and should be designed to ensure that all parts of the 
    quality system are adequately reviewed. A manufacturer may, of course, 
    develop procedures that permit review of different areas at different 
    times, so long as such reviews are sufficient to carry out the 
    objectives of this section. If there are known problems, for example, a 
    ``sufficient frequency'' may be fairly frequent. Further, because FDA 
    will not be reviewing the results of such reviews, FDA must be assured 
    that this function will occur in a consistent manner.
        53. A few comments stated that Sec. 820.20(c) should be deleted 
    because it duplicates the quality audit required by Sec. 820.22.
        FDA disagrees that Sec. 820.20(c) duplicates the requirements in 
    Sec. 820.22. The purpose of the management reviews required by 
    Sec. 820.20(c) is to determine if the manufacturer's quality policy and 
    quality objectives are being met, and to ensure the continued 
    suitability and effectiveness of the quality system. An evaluation of 
    the findings of internal and supplier audits should be included in the 
    Sec. 820.20(c) evaluation. The management review may include a review 
    of the following: (1) The organizational structure, including the 
    adequacy of staffing and resources; (2) the quality of the finished 
    device in relation to the quality objectives; (3) combined information 
    based on purchaser feedback, internal feedback (such as results of 
    internal audits), process performance, product (including servicing) 
    performance, among other things; and (4) internal audit results and 
    corrective and preventive actions taken. Management reviews should 
    include considerations for updating the quality system in relation to 
    changes brought about by new technologies, quality concepts, market 
    strategies, and other social or environmental conditions. Management 
    should also review periodically the appropriateness of the review 
    frequency, based on the findings of previous reviews. The quality 
    system review process in Sec. 820.20(c), and the reasons for the 
    review, should be understood by the organization.
        The requirements under Sec. 820.22 Quality audit are for an 
    internal audit and review of the quality system to verify compliance 
    with the quality system regulation. The review and evaluations under 
    Sec. 820.22 are very focused. During the internal quality audit, the 
    manufacturer should review all procedures to ensure adequacy and 
    compliance with the regulation, and determine whether the procedures 
    are being effectively implemented at all times. In contrast, as noted 
    above, the management review under Sec. 820.20(c) is a broader review 
    of the organization as a whole to ensure that the quality policy is 
    implemented and the quality objectives are met. The reviews of the 
    quality policy and objectives (Sec. 820.20(c)) should be carried out by 
    top management, and the review of supporting activities (Sec. 820.22) 
    should be carried out by management with executive responsibility for 
    quality and other appropriate members of management, utilizing 
    competent personnel as decided on by the management.
        54. Some comments suggested that the requirements in 
    Sec. 820.186(a) and (d) be moved to Sec. 820.20 for clarity and to 
    better align with the structure of ISO 9001:1994 and ISO/CD 13485.
        FDA agrees and has moved the specific requirements from 
    Sec. 820.186 and rewritten them into new Sec. 820.20 (d) and (e) for 
    clarity, better organization, and closer harmonization. Therefore, 
    Sec. 820.20(d) is consistent with ISO 9001:1994, section 4.2.3, 
    ``Quality planning,'' and Sec. 820.20(e) is consistent with ISO 
    9001:1994, sections 4.2.1, ``General,'' and 4.2.2, ``Quality-system 
    procedures.'' Section 820.20(e) discusses ``[a]n outline of the 
    structure of the documentation used in the quality system.'' FDA 
    believes that outlining the structure of the documentation is 
    beneficial and, at times, may be critical to the effective operation of 
    the quality system. FDA recognizes, however, that it may not be 
    necessary to create an outline in all cases. For example, it may not be 
    necessary for smaller manufacturers and manufacturers of less 
    complicated devices. Thus, the outline is only required where 
    appropriate.
    ii. Quality Audit (Sec. 820.22)
        55. A few comments suggested that FDA delete the requirement that 
    persons conducting the audit be ``appropriately trained'' from the 
    second sentence of proposed Sec. 820.22(a), because it is subjective 
    and not consistent with ISO 9001.
        FDA has deleted the requirement from Sec. 820.22(a) because 
    Sec. 820.25 Personnel requires that such individuals be appropriately 
    trained. Further, FDA has attempted to better harmonize with ISO 
    9001:1994, which does not explicitly state personnel qualifications in 
    each provision. Similarly, in response to general comments suggesting 
    better harmonization, FDA has added the requirement that the audit 
    ``determine the effectiveness of the quality system'' as required by 
    ISO 9001:1994. This requirement underscores that the quality audit must 
    not only determine whether the manufacturer's requirements are being 
    carried out, but whether the requirements themselves are adequate.
        56. Some comments stated that requiring ``individuals who do not 
    have direct responsibility for the matters being audited'' to conduct 
    the audits is impractical and burdensome, particularly for small 
    manufacturers.
        FDA disagrees with the comments. Both small and large manufacturers 
    have been subject to the identical requirement since 1978 and FDA knows 
    of no hardship, on small or large manufacturers, as a result. Small 
    manufacturers must generally establish independence, even if it means 
    hiring outside auditors, because the failure to have an independent 
    auditor could result in an ineffective audit.
        Manufacturers must realize that conducting effective quality audits 
    is crucial. Without the feedback provided by the quality audit and 
    other information sources, such as complaints and service records, 
    manufacturers operate in an open loop system with no assurance that the 
    process used to design and produce devices is operating in a state of 
    control. ISO 9001:1994 has the same requirement for independence from 
    the activity being audited.
        57. Several comments claimed that the last sentence in proposed 
    Sec. 820.22(a), which required that followup corrective action be 
    documented in the audit report, made no sense. The comments said that 
    corrective action would be the subject of a followup report.
        It was the agency's intent that the provision require that where 
    corrective action was necessary, it would be taken and documented in a 
    reaudit report. The provision has been rewritten to make that clear. 
    New Sec. 824.22 also clarifies that a reaudit is not always required, 
    but where it is indicated, it must be conducted. The report should 
    verify that corrective action was implemented and effective. Because 
    FDA does not review these reports, the date on which the audit and 
    reaudit were performed must be documented and will be subject to FDA 
    review. The revised reaudit provision is consistent with ISO 9001:1994.
        58. Many comments were received on proposed Sec. 820.22(b) 
    regarding the reports exempt from FDA review. Most of the comments 
    objected to FDA reviewing evaluations of suppliers. FDA has decided not 
    to review such
    
    [[Page 52615]]
    
    evaluations at this time and will revisit this decision after the 
    agency gains sufficient experience with the new requirement to 
    determine its effectiveness. A thorough response to the comments is 
    found with the agency's response to other comments received on 
    Sec. 820.50 Purchasing controls. FDA has moved the section regarding 
    which reports the agency will refrain from reviewing from 
    Sec. 820.22(b) to new Sec. 820.180(c), ``Exemptions,'' under the 
    related records requirements. FDA believes this organization is easier 
    to follow.
    iii. Personnel (Sec. 820.25)
        59. A few comments stated that the requirement in Sec. 820.25 
    Personnel for the manufacturer to employ ``sufficient'' personnel 
    should be deleted, because whether there are ``sufficient'' personnel 
    is a subjective determination, and it is unnecessary to require it 
    since the manufacturer will know how best to staff the organization. A 
    few other comments stated that the provision should not base the 
    personnel requirements on ensuring that the requirements of the 
    regulation are ``correctly'' performed, because no manufacturer can 
    ensure that all activities are performed correctly. Another comment 
    stated that the term ``employ'' should be changed because personnel may 
    include qualified temporaries, contractors, and others who may not 
    typically be considered ``employees.''
        FDA disagrees with the suggestions that the terms ``sufficient'' 
    and ``correctly'' be deleted. Whether ``sufficient'' personnel are 
    employed will be determined by the requirements of the quality system, 
    which must be designed to ensure that the requirements of the 
    regulation are properly implemented. In making staffing decisions, a 
    manufacturer must ensure that persons assigned to particular functions 
    are properly equipped and possess the necessary education, background, 
    training, and experience to perform their functions correctly. However, 
    FDA changed ``ensure'' to ``assure'' to address the concerns that 
    people do make mistakes and management cannot guarantee that work is 
    correctly performed all of the time. Further, FDA agrees that the 
    manufacturer must determine for itself what constitutes ``sufficient'' 
    personnel with proper qualification in the first instance. However, if 
    the manufacturer does not employ sufficient personnel, or personnel 
    with the necessary qualifications to carry out their functions, the 
    manufacturer will be in violation of the regulation. FDA has often 
    found that the failure to comply with this requirement leads to other 
    significant regulatory violations. FDA agrees with the comment that the 
    term ``employ'' should be deleted so that the requirement covers all 
    personnel who work at a firm.
        60. In Sec. 820.25(b), ``Training,'' FDA deleted the requirement 
    that employees be trained ``by qualified individuals,'' because 
    Sec. 820.25(a) requires this. Several comments stated that FDA should 
    add the requirement that the training procedure include the 
    identification of training needs, to be consistent with the 
    requirements in ISO 9001:1994 and ISO/CD 13485. Other comments stated 
    that personnel need not be trained to the extent that they can quote 
    chapter and verse of the regulation as long as they can adequately 
    perform their assigned responsibilities. Several comments suggested 
    deleting the requirements in the last two sentences in favor of a 
    broad, general requirement that personnel be trained. A few comments 
    stated that the last two sentences should be retained because they are 
    crucial and sound requirements but that validation activities should be 
    included with verification activities.
        FDA amended the requirement so that the training procedure includes 
    the identification of training needs. FDA deleted the requirement on 
    understanding the CGMP requirements applicable to job functions to 
    avoid the perception that personnel would need to know ``chapter and 
    verse of the regulation.'' FDA notes, however, that a training program 
    to ensure personnel adequately perform their assigned responsibilities 
    should include information about the CGMP requirements and how 
    particular job functions relate to the overall quality system. FDA 
    further believes that it is imperative that training cover the 
    consequences of improper performance so that personnel will be apprised 
    of defects that they should look for, as well as be aware of the effect 
    their actions can have on the safety and effectiveness of the device. 
    In addition, FDA disagrees with comments that suggested that only 
    ``personnel affecting quality'' should be required to be adequately 
    trained. In order for the full quality system to function as intended, 
    all personnel should be properly trained. Each function in the 
    manufacture of a medical device must be viewed as integral to all other 
    functions. FDA has reorganized the last two sentences, however, to 
    place the requirements under Sec. 820.25(b), ``Training,'' and has 
    added validation activities as suggested by the comments.
        61. Many comments objected to the proposed requirements of 
    Sec. 820.25(c), ``Consultants,'' stating that requiring a manufacturer 
    to chose consultants that have sufficient qualifications and to keep 
    records subject to FDA review of all consultants used, along with 
    copies of their resumes and lists of previous jobs, would unreasonably 
    interfere with the manufacturer's business activities and restrict the 
    right of a manufacturer to hire consultants on any basis it chooses. 
    Other comments said that a manufacturer's employment of a consultant 
    has the same potential impact on the safety and effectiveness of 
    medical devices as employment of any other contractor for services, and 
    that consultants should, therefore, be covered by Sec. 820.50 
    Purchasing controls.
        FDA agrees in part with these comments. Although employing a 
    consultant is a business decision, when a manufacturer hires 
    consultants who do not have appropriate credentials, and manufacturing 
    decisions are made based on erroneous or ill-conceived advice, the 
    public suffers. Of course, the manufacturer is still ultimately 
    responsible for following the CGMP requirements and will bear the 
    consequences of a failure to comply. FDA notes that the use of 
    unqualified consultants has led to regulatory action for the failure to 
    comply with the CGMP regulation in the past. Thus, because of the 
    significant impact a consultant can have on the safety and 
    effectiveness of a device, FDA believes that some degree of control is 
    required in the regulation.
        The requirements are revised somewhat in response to comments, 
    however, to reflect that it is not FDA's goal to dictate whom a 
    manufacturer may use as a consultant, but instead to require that a 
    manufacturer determine what it needs to adequately carry out the 
    requirements of the regulation and to assess whether the consultant can 
    adequately meet those needs. The requirements related to consultants 
    have been added in Sec. 820.50 Purchasing controls because a consultant 
    is a supplier of a service.
    
    C. Design Controls (Subpart C)
    
        Since early 1984, FDA has identified lack of design controls as one 
    of the major causes of device recalls. The intrinsic quality of 
    devices, including their safety and effectiveness, is established 
    during the design phase. Thus, FDA believes that unless appropriate 
    design controls are observed during preproduction stages of 
    development, a finished device may be neither safe nor effective for 
    its intended use. The SMDA provided FDA with the
    
    [[Page 52616]]
    
    authority to add preproduction design controls to the device CGMP 
    regulation. Based on its experience with administering the original 
    CGMP regulation, which did not include preproduction design controls, 
    the agency was concerned that the original regulation provided less 
    than an adequate level of assurance that devices would be safe and 
    effective. Therefore, FDA has added general requirements for design 
    controls to the device CGMP regulation for all class III and II devices 
    and certain class I devices. FDA is not subjecting the majority of 
    class I devices to design controls because FDA does not believe that 
    such controls are necessary to ensure that such devices are safe and 
    effective and otherwise in compliance with the act. However, all 
    devices, including class I devices exempt from design controls, must be 
    properly transferred to production in order to comply with 
    Sec. 820.181, as well as other applicable requirements. For most class 
    I devices, FDA believes that the production and other controls in the 
    new quality system regulation and other general controls of the act 
    will be sufficient, as they have been in the past, to ensure safety and 
    effectiveness.
        62. Many comments were submitted in response to the addition of 
    design control requirements in general, many questioning how these new 
    requirements would be implemented and enforced. For instance, several 
    comments stated that the design control requirements do not reflect how 
    medical devices are actually developed, because the concept of a design 
    rarely originates with the manufacturer, who may not become involved 
    until relatively late in the design evolution. Others expressed concern 
    that FDA investigators will second-guess design issues in which they 
    are not educated or trained, and stated that investigators should not 
    debate whether medical device designs are ``safe and effective.''
        FDA agrees in part with the comments. The design control 
    requirements are not intended to apply to the development of concepts 
    and feasibility studies. However, once it is decided that a design will 
    be developed, a plan must be established to determine the adequacy of 
    the design requirements and to ensure that the design that will 
    eventually be released to production meets the approved requirements.
        Those who design medical devices must be aware of the design 
    control requirements in the regulation and comply with them. Unsafe and 
    ineffective devices are often the result of informal development that 
    does not ensure the proper establishment and assessment of design 
    requirements which are necessary to develop a medical device that is 
    safe and effective for the intended use of the device and that meets 
    the needs of the user.
        However, FDA investigators will not inspect a device under the 
    design control requirements to determine whether the design is 
    appropriate or ``safe and effective.'' Section 520(f)(1)(a) of the act 
    precludes FDA from evaluating the ``safety or effectiveness of a 
    device'' through preproduction design control procedures. FDA 
    investigators will evaluate the process, the methods, and the 
    procedures that a manufacturer has established to implement the 
    requirements for design controls. If, based on any information gained 
    during an inspection, an investigator believes that distributed devices 
    are unsafe or ineffective, the investigator has an obligation to report 
    the observations to the Center for Devices and Radiological Health 
    (CDRH).
        63. Several comments expressed concern that the application of 
    design controls would severely restrict the creativity and innovation 
    of the design process and suggested that design controls should not 
    apply too early in the design development process.
        FDA disagrees with the comments. It is not the intent of FDA to 
    interfere with creativity and innovation, and it is not the intent of 
    FDA to apply the design control requirements to the research phase. 
    Instead, the regulation requires the establishment of procedures to 
    ensure that whatever design is ultimately transferred to production is, 
    in fact, a design that will translate into a device that properly 
    performs according to its intended use and user needs.
        To assist FDA in applying the regulation, manufacturers should 
    document the flow of the design process so that it is clear to the FDA 
    investigator where research is ending and development of the design is 
    beginning.
        64. A few comments stated that design controls should not be 
    retroactive and that ongoing design development should be exempted.
        FDA agrees in part with the comments. FDA did not intend the design 
    requirements to be retroactive, and Sec. 820.30 Design controls will 
    not require the manufacturer to apply such requirements to already 
    distributed devices. When the regulation becomes effective on June 1, 
    1997, it will apply to designs that are in the design and development 
    phase, and manufacturers will be expected to have the design and 
    development plan established. The manufacturer should identify what 
    stage a design is in for each device and will be expected to comply 
    with the established design and development plan and the applicable 
    paragraphs of Sec. 820.30 from that point forward to completion. If a 
    manufacturer had a design in the development stage before June 1, 1997, 
    and cannot comply with any particular paragraph of Sec. 820.30, the 
    manufacturer must provide a detailed justification as to why such 
    compliance is not possible. However, designs will not have to be 
    recycled through previous phases that have been completed. 
    Manufacturers will be expected to comply in full by June 1, 1998. As 
    stated earlier, FDA wants to emphasize that it expects manufacturers to 
    be in a reasonable state of compliance with the design control 
    requirements from June 1, 1997, to June 1, 1998, because extra time was 
    given to the industry for implementing design controls before the final 
    regulation became effective.
        When changes are made to new or existing designs, the design 
    controls of Sec. 820.30 must be followed to ensure that the changes are 
    appropriate and that the device will continue to perform as intended. 
    FDA notes that the original CGMP regulation contained requirements for 
    specification controls and controls for specification or design changes 
    under Sec. 820.100(a).
        65. One comment asked how the proposed design controls would apply 
    to investigational device exemption (IDE) devices, since devices under 
    approved IDE's have been exempt from the CGMP regulation. Some comments 
    suggested that any changes to the IDE regulation should be done in a 
    separate rulemaking. Other comments stated that any change to the IDE 
    regulation should be worded so that all of Sec. 820.30 applies since 
    the IDE process is supplying information in support of the design 
    validation requirements but that all design requirements need not be 
    completed prior to the start of the IDE because the clinical evaluation 
    process often brings valuable information to the design project which 
    may need to be incorporated into the design before design transfer.
        The IDE regulation was published in 1976 and last updated in 1978, 
    and has been in effect since that time. Devices being evaluated under 
    IDE's were exempted from the original CGMP regulation because it was 
    believed that it was not reasonable to expect sponsors of clinical 
    investigations to ensure compliance with CGMP's for devices that may 
    never be approved for commercial distribution. However, sponsors of IDE 
    studies were required to ensure that investigational devices were 
    manufactured under a state of control.
    
    [[Page 52617]]
    
        With respect to the new regulation, FDA believes that it is 
    reasonable to expect manufacturers who design medical devices to 
    develop the designs in conformance with design control requirements and 
    that adhering to such requirements is necessary to adequately protect 
    the public from potentially harmful devices. The design control 
    requirements are basic controls needed to ensure that the device being 
    designed will perform as intended when produced for commercial 
    distribution. Clinical evaluation is an important aspect of the design 
    verification and validation process during the design and development 
    of the device. Because some of the device design occurs during the IDE 
    stage, it is logical that manufacturers who intend to commercially 
    produce the device follow design control procedures. Were a 
    manufacturer to wait until all the IDE studies were complete, it would 
    be too late to take advantage of the design control process, and the 
    manufacturer would not be able to fulfill the requirements of the 
    quality system regulation for that device.
        Therefore, FDA has concurrently amended the IDE regulation,
    
    812.1  Scope to state:
    
        (a) * * * An IDE approved under Sec. 812.30 or considered 
    approved under Sec. 812.2(b) exempts a device from the requirements 
    of the following sections of the Federal Food, Drug, and Cosmetic 
    Act (the act) and regulations issued thereunder: * * * good 
    manufacturing practice requirements under section 520(f) except for 
    the requirements found in Sec. 820.30, if applicable (unless the 
    sponsor states an intention to comply with these requirements under 
    Sec. 812.20(b)(3) or Sec. 812.140(b)(4)(v)) and color additive 
    requirements under section 721. (Emphasis added.)
    
        FDA does not expect any new information in IDE applications as a 
    result of this amendment, nor will FDA inspect design controls during 
    bioresearch monitoring inspections. FDA is simply making a conforming 
    amendment to the IDE regulation to make clear that design controls must 
    be followed when design functions are undertaken by manufacturers, 
    including design activity which occurs under an approved IDE. FDA will 
    evaluate the adequacy of manufacturers' compliance with design control 
    requirements in routine CGMP inspections, including preapproval 
    inspections for premarket approval applications (PMA's).
        66. Many written comments and oral comments at the August and 
    September 1995 meetings recommended that, because design controls are a 
    major addition to the regulation, the effective date for design 
    controls should be delayed until 18 months after publication of the 
    final rule.
        FDA has addressed these comments by extending the effective date of 
    the regulation until June 1, 1997, and by the inspectional strategy 
    described earlier.
        67. A couple of comments suggested that FDA lacked the authority to 
    establish the design control requirements.
        FDA disagrees with the comments. The act and its legislative 
    history make clear that FDA has the authority to impose those controls 
    necessary to ensure that devices are safe and effective. The SMDA gave 
    FDA explicit authority to promulgate design controls, including a 
    process to assess the performance of a device (see section 520(f)(1)(A) 
    of the act). The legislative history of the SMDA supports a 
    ``comprehensive device design validation regulation.'' H. Rept. 808, 
    101st Cong., 2d sess. 23 (emphasis added). Congress stated that the 
    amendment to the statute was necessary because almost half of all 
    device recalls over a 5-year period were ``related to a problem with 
    product design.'' Id. There is a thorough discussion on the evolution 
    of and need for the design controls in the preamble to the November 23, 
    1993 (58 FR 61952), proposal.
        68. A few comments objected to FDA requiring design controls for 
    any class I devices in Sec. 820.30(a).
        FDA believes that, for the class I devices listed, design controls 
    are necessary and has retained the requirements. Those relatively few 
    devices, while class I, require close control of the design process to 
    ensure that the devices perform as intended, given the serious 
    consequences that could occur if their designs were flawed and the 
    devices were to fail to meet their intended uses. In fact, some of the 
    devices included on the list have experienced failures due to design 
    related problems that have resulted in health hazards, injuries, or 
    death. Further, verification, or even validation, cannot provide the 
    assurance of proper design for some devices, especially those 
    containing extensive software. Thus, all automated devices must be 
    developed under the design control requirements.
        69. Several comments stated that FDA has underestimated the 
    complexity of a design project in requiring that the plans identify 
    ``persons responsible for each activity'' in proposed Sec. 820.30(b). 
    One comment stated that ``define responsibility for implementation'' 
    and ``activities shall be assigned'' were basically redundant 
    requirements. A few other comments stated that ISO 9001:1994 does not 
    call for the design plans to be ``approved'' and that this requirement 
    should be deleted because it would be burdensome.
        FDA agrees in part with the comments and has revised Sec. 820.30(b) 
    to require the plan to describe or reference design activities and 
    define responsibility for implementing the activities, rather than 
    requiring that the plan identify each person responsible for carrying 
    out each activity. In making this change, FDA notes that 
    Sec. 820.20(b)(1) requires manufacturers to establish the appropriate 
    responsibility for activities affecting quality, and emphasizes that 
    the assignment of specific responsibility is important to the success 
    of the design control program and to achieving compliance with the 
    regulation. Also, the design and development activities should be 
    assigned to qualified personnel equipped with adequate resources as 
    required under Sec. 820.20(b)(2). The requirements under Sec. 820.30(b) 
    were rewritten to be very similar to the requirements in ISO 9001:1994, 
    sections 4.4.2 and 4.4.3. FDA does not agree that the design plan 
    should not be ``approved.'' ISO 9001:1994, section 4.4.2 requires that 
    the plan be ``updated,'' and section 4.4.3 requires that the plan be 
    ``regularly reviewed.'' Therefore, the approval is consistent with ISO 
    9001:1994 and would not be unduly burdensome since the FDA does not 
    dictate how or by whom the plan must be approved. The regulation gives 
    the manufacturer the necessary flexibility to have the same person(s) 
    who is responsible for the review also be responsible for the approval 
    of the plan if appropriate.
        70. A few comments stated that the proposed requirement to describe 
    ``any interaction between or among different organizational and 
    technical groups'' in Sec. 820.30(b) for the design and development 
    plan should be deleted because it is overly broad, unnecessary, and 
    burdensome. One comment said that the communication expected between 
    these groups should be clarified.
        In response, FDA has amended the requirement as suggested by one 
    comment so that the plan shall identify and describe the interfaces 
    with different groups or activities that provide, or result in, input 
    to the design process. Many organization functions, both inside and 
    outside the design group, may contribute to the design process. For 
    example, interfaces with marketing, purchasing, regulatory affairs, 
    manufacturing, service groups, or information systems may be necessary 
    during the design
    
    [[Page 52618]]
    
    development phase. To function effectively, the design plan must 
    establish the roles of these groups in the design process and describe 
    the information that should be received and transmitted.
        71. One comment stated that the requirement in Sec. 820.30(b) that 
    manufacturers establish a design plan completely ignores the creative 
    and dynamic process of designing by requiring a plan to have complete 
    design and testing criteria established, with specifications, before 
    the design process is started.
        FDA disagrees with the comment. Section 820.30(b) does not require 
    manufacturers to complete design and testing criteria before the design 
    process begins. This section has been revised to state that ``plans 
    shall be reviewed, updated, and approved as design and development 
    evolves,'' indicating that changes to the design plan are expected. A 
    design plan typically includes at least proposed quality practices, 
    assessment methodology, recordkeeping and documentation requirements, 
    and resources, as well as a sequence of events related to a particular 
    design or design category. These may be modified and refined as the 
    design evolves. However, the design process can become a lengthy and 
    costly process if the design activity is not properly defined and 
    planned. The more specifically the activities are defined up front, the 
    less need there will be for changes as the design evolves.
        72. One comment stated that the language contained in proposed 
    Sec. 820.30(c) should more closely match that of ISO 9001. Many other 
    comments stated that the provision should not require the input 
    requirements to ``completely'' address the intended use of the device 
    because inputs could never ``completely'' address the intended use. 
    Several comments stated that the requirement of ISO 9001 that 
    ``incomplete, ambiguous or conflicting requirements shall be resolved 
    with those responsible for imposing these requirements'' should be 
    added to Sec. 820.30(c), ``Design input,'' because it is important that 
    the regulation identify the method of resolving conflicting 
    information.
        FDA agrees with the harmonization comment and has revised the 
    language to incorporate the requirement of section 4.4.4, ``Design 
    input,'' of ISO 9001:1994. FDA does not believe that it is necessary to 
    have identical language to harmonize quality system requirements. ISO 
    9001:1994, section 4.4.1, ``General,'' requires that the manufacturer 
    ``establish and maintain documented procedures to control and verify 
    the design of the product in order to ensure that the specified 
    requirements are met.'' FDA's regulation, under Sec. 820.30(a), imposes 
    the same requirements.
        Regarding the comments that input requirements cannot completely 
    address the intended use of the device, FDA recognizes that the 
    provision could be interpreted to impose a burden that may not always 
    be possible to meet and has deleted the word ``completely.'' FDA did 
    not intend the provision to suggest that a manufacturer must foresee 
    every possible event.
        FDA emphasizes, however, that the section requires the manufacturer 
    to ensure that the design input requirements are appropriate so the 
    device will perform to meet its intended use and the needs of the user. 
    In doing this, the manufacturer must define the performance 
    characteristics, safety and reliability requirements, environmental 
    requirements and limitations, physical characteristics, applicable 
    standards and regulatory requirements, and labeling and packaging 
    requirements, among other things, and refine the design requirements as 
    verification and validation results are established. For example, when 
    designing a device, the manufacturer should conduct appropriate human 
    factors studies, analyses, and tests from the early stages of the 
    design process until that point in development at which the interfaces 
    with the medical professional and the patient are fixed. The human 
    interface includes both the hardware and software characteristics that 
    affect device use, and good design is crucial to logical, 
    straightforward, and safe device operation. The human factors methods 
    used (for instance, task/function analyses, user studies, prototype 
    tests, mock-up reviews, etc.) should ensure that the characteristics of 
    the user population and operating environment are considered. In 
    addition, the compatibility of system components should be assessed. 
    Finally, labeling (e.g., instructions for use) should be tested for 
    usability.
        FDA agrees with the comments, in that it is important that 
    incomplete, ambiguous, or conflicting requirements be resolved with 
    those responsible for imposing these requirements. Therefore, FDA has 
    added the requirement that the procedures shall include a mechanism for 
    addressing incomplete, ambiguous, or conflicting requirements. FDA 
    notes that this must be done to ``ensure that the design requirements 
    are appropriate and address the intended use of the device,'' as 
    required under Sec. 820.30(c).
        73. A few other comments stated that ISO 9001:1994 does not call 
    for the design input to be ``approved'' and therefore, this requirement 
    should be deleted because it would be burdensome.
        FDA does not agree that the ``approval'' of design input 
    requirements should be deleted, nor that the requirement is 
    inconsistent with ISO. ISO 9001:1994, section 4.4.4, ``Design Input,'' 
    requires that the design input requirements be ``reviewed by the 
    supplier for adequacy.'' Therefore, the approval would not add any 
    additional burden because FDA does not dictate how or by whom the 
    design input requirements must be approved, thus giving the 
    manufacturer the necessary flexibility to have the same person(s) who 
    is responsible for the ``review for adequacy'' also be responsible for 
    the approval, if appropriate. Further, it is important that the design 
    input be assessed as early as possible in the development process, 
    making this an ideal time in the device's design development to have a 
    design review to ``approve'' the design input.
        74. A few comments stated that the proposed requirement under 
    Sec. 820.30(c) that ``design input shall be reviewed and approved by a 
    designated qualified individual'' should be deleted as it implies that 
    one person must be designated to review and approve a design, and that 
    there may not be one person who is qualified to assess all of the 
    design input requirements. Addressing the same point, several comments 
    suggested that the provision be revised to allow for more than one 
    person to review and approve the design. One comment said that the 
    FDA's requirement appears to be at odds with the team approach.
        FDA agrees with the concern expressed by the comments and has 
    modified the requirement to allow more than one individual to review 
    and approve the design input. FDA endorses the team approach and 
    believes that designs should be reviewed and evaluated by all 
    disciplines necessary to ensure the design input requirements are 
    appropriate.
        75. Two comments stated that proposed Sec. 820.30(c) should be 
    reworded to focus on systems for assuring adequate design input, not on 
    the input itself. One additional comment on this section said that the 
    design input requirements should include not only the device's intended 
    use and needs of the user, but the environmental limits of where it 
    will be used.
        FDA agrees that procedures for ensuring appropriate design controls 
    are of the utmost importance and has modified the section to clarify 
    that the
    
    [[Page 52619]]
    
    manufacturer must establish and maintain procedures to ensure that the 
    design requirements are properly addressed. FDA made this change to the 
    other paragraphs as well, but notes that Sec. 820.30(a), ``General,'' 
    requires the manufacturer to establish and maintain procedures to 
    control the design of the device in order to ensure that specified 
    design requirements are met. The sections that follow set forth some of 
    the requirements for which procedures must be established. It should be 
    emphasized that the input itself must also be appropriate; the 
    requirement is for the procedures to be defined, documented, and 
    implemented. Thus, if the input requirements related to a device fail 
    to address the intended use of the device, for example, the 
    manufacturer has failed to comply with the provision.
        FDA also agrees with the additional comment but believes that 
    identifying and establishing the environmental limits for safe and 
    effective device operation is inherent in the requirements for ensuring 
    that a device is appropriate for its intended use. Some factors that 
    must be considered when establishing inputs include, where applicable, 
    a determination of energy (e.g., electrical, heat, and electromagnetic 
    fields), biological effects (e.g., toxicity and biocompatibility) and 
    environmental effects (e.g., electromagnetic interference and 
    electrostatic discharge).
        76. Several comments stated that proposed Sec. 820.30(f), ``Design 
    output,'' should be rewritten or deleted because many of the 
    requirements were already stated in proposed Secs. 820.30(d), ``Design 
    verification,'' and 820.30(e), ``Design review,'' and, if retained, 
    should be reordered similar to ISO 9001.
        FDA agrees in part with the comments and has rewritten the 
    requirements of design output to be consistent with ISO 9001:1994, 
    section 4.4.5, ``Design output,'' and reordered the sections to be 
    consistent with ISO 9001:1994. FDA retained the provision, however, 
    because it does not agree that the section is redundant with the 
    sections on design verification, design validation, or design review. 
    Design output are the design specifications which should meet design 
    input requirements, as confirmed during design verification and 
    validation and ensured during design review. The output includes the 
    device, its labeling and packaging, associated specifications and 
    drawings, and production and quality assurance specifications and 
    procedures. These documents are the basis for the DMR. The total 
    finished design output consists of the device, its labeling and 
    packaging, and the DMR.
        77. One comment stated that the sentence ``Design output procedures 
    shall ensure that design output meets the design input requirements'' 
    is redundant with the requirement under design verification. Another 
    comment asked what is meant by ``release.''
        FDA agrees with the first comment and has deleted that sentence in 
    Sec. 820.30(d) but notes that the design output must be documented and 
    expressed in terms that can be verified against the design input 
    requirements.
        Design output can be ``released'' or transferred to the next design 
    phase at various stages in the design process, as defined in the design 
    and development plan. The design output is reviewed and approved before 
    release or transfer to the next design phase or production. The design 
    output requirements are intended to apply to all such stages of the 
    design process.
        78. One small manufacturer commented that the problems that 
    Sec. 820.30(e), ``Design review,'' is meant to reveal involve 
    coordination, cooperation, or communication difficulties among the 
    members of an organization and that these difficulties do not exist in 
    a small company. Therefore, the comment stated that the design review 
    requirements should not apply to small manufacturers.
        The purpose of conducting design reviews during the design phase is 
    to ensure that the design satisfies the design input requirements for 
    the intended use of the device and the needs of the user. Design review 
    includes the review of design verification data to determine whether 
    the design outputs meet functional and operational requirements, the 
    design is compatible with components and other accessories, the safety 
    requirements are achieved, the reliability and maintenance requirements 
    are met, the labeling and other regulatory requirements are met, and 
    the manufacturing, installation, and servicing requirements are 
    compatible with the design specifications. Design reviews should be 
    conducted at major decision points during the design phase.
        For a large manufacturer, design review provides an opportunity for 
    all those who may have an impact on the quality of the device to 
    provide input, including manufacturing, quality assurance, purchasing, 
    sales, and servicing divisions. While small manufacturers may not have 
    the broad range of disciplines found in a large company, and the need 
    to coordinate and control technical interfaces may be lessened, the 
    principles of design review still apply. The requirements under 
    Sec. 820.30(e) allow small manufacturers to tailor a design review that 
    is appropriate to their individual needs.
        79. One comment stated that the wording of proposed Sec. 820.30(e) 
    implies that only one design review is expected, and that design review 
    should be conducted at several stages of product development. Several 
    comments stated that to demand that every design review be conducted by 
    individuals who do not have direct responsibility for design 
    development is impractical, especially for small companies.
        FDA agrees with the first comment and has rewritten the requirement 
    to make clear that design reviews must be conducted at appropriate 
    stages of design development, which must be defined in the established 
    design and development plan. The number of design reviews will depend 
    on the plan and the complexity of the device. FDA also amended the 
    requirements so that the results of a design review include 
    identification of the design, the date, and the individual(s) 
    performing the review. Thus, multiple reviews can occur and the 
    manufacturer must document what is being reviewed, when, and by whom.
        FDA never intended to mandate that an individual without design 
    responsibility conduct the design reviews and, to clarify its position, 
    has rewritten the requirement. The requirement now states that the 
    procedures shall ensure that each design review includes an 
    individual(s) who does not have direct responsibility for the design 
    stage being reviewed. This requirement will provide an ``objective 
    view'' from someone not working directly on that particular part of the 
    design project, to ensure that the requirements are met. In making this 
    change, FDA also notes that it was not FDA's intention to prohibit 
    those directly responsible for the design from participating in the 
    design review.
        80. One comment stated that as part of the systematic review of the 
    adequacy of the device design, it is occasionally necessary to produce 
    a prototype device and have it evaluated by a physician who is an 
    expert in the area of the device's intended use. Thus, the comment 
    stated that the regulation should be revised to allow a means for a 
    manufacturer to ship a prototype device to a physician for evaluation. 
    One comment questioned whether design verification and validation can 
    be conducted using prototypes or machine shop models.
        FDA regulations do not prohibit the shipment of prototypes for 
    clinical or
    
    [[Page 52620]]
    
    other studies. Prototypes used in clinical studies involving humans may 
    be shipped in accordance with the IDE provisions in part 812 (21 CFR 
    part 812).
        FDA understands that it is not always practical to conduct clinical 
    studies on finished production units and, therefore, the use of 
    prototypes in clinical studies is acceptable. When prototype devices 
    are used on humans they must be verified as safe to the maximum extent 
    feasible. Final design validation, however, cannot be done on 
    prototypes because the actual devices produced and distributed are 
    seldom the same as the research and development prototypes. The final 
    verification and validation, therefore, must include the testing of 
    actual production devices under actual or simulated use conditions.
        81. A few comments stated that Sec. 820.30(d), ``Design 
    verification,'' should be rewritten and reordered similar to ISO 9001.
        FDA agrees with the comments and has rewritten and reordered this 
    section to be consistent with ISO 9001:1994. The language in revised 
    Sec. 820.30(f) and (g) incorporates the requirement of ISO 9001:1994, 
    sections 4.4.7, ``Design verification,'' and 4.4.8, ``Design 
    validation,'' respectively.
        Under the revised provisions, the design must be verified and 
    validated. It is important to note that design validation follows 
    successful design verification. Certain aspects of design validation 
    can be accomplished during the design verification, but design 
    verification is not a substitute for design validation. Design 
    validation should be performed under defined operating conditions and 
    on the initial production units, lots, or batches, or their equivalents 
    to ensure proper overall design control and proper design transfer. 
    When equivalent devices are used in the final design validation, the 
    manufacturer must document in detail how the device was manufactured 
    and how the manufacturing is similar to and possibly different from 
    initial production. Where there are differences, the manufacturer must 
    justify why design validation results are valid for the production 
    units, lots, or batches. Manufacturers should not use prototypes 
    developed in the laboratory or machine shop as test units to meet these 
    requirements. Prototypes may differ from the finished production 
    devices. During research and development, conditions for building 
    prototypes are typically better controlled and personnel more 
    knowledgeable about what needs to be done and how to do it than are 
    regular production personnel. When going from laboratory to scale-up 
    production, standards, methods, and procedures may not be properly 
    transferred, or additional manufacturing processes may be added. Often, 
    changes not reflected in the prototype are made in the device to 
    facilitate the manufacturing process, and these may adversely affect 
    device functioning and user interface characteristics. Proper testing 
    of devices that are produced using the same methods and procedures as 
    those to be used in routine production will prevent the distribution 
    and subsequent recall of many unacceptable medical devices.
        In addition, finished devices must be tested for performance under 
    actual conditions of use or simulated use conditions in the actual or 
    simulated environment in which the device is expected to be used. The 
    simulated use testing provision no longer requires that the testing be 
    performed on the first three production runs. However, samples must be 
    taken from units, lots, or batches that were produced using the same 
    specifications, production and quality system methods, procedures, and 
    equipment that will be used for routine production. FDA considers this 
    a critical element of the design validation. The requirement to conduct 
    simulated use testing of finished devices is found in the original CGMP 
    in Sec. 820.160, as part of finished device inspection. This 
    requirement has been moved to Sec. 820.30(g) because FDA believes that 
    simulated use testing at this point is more effective in ensuring that 
    only safe and effective devices are produced. Manufacturers must also 
    conduct such tests when they make changes in the device design or the 
    manufacturing process that could affect safety or effectiveness as 
    required in the original CGMP in Sec. 820.100(a)(2). The extent of 
    testing conducted should be governed by the risk(s) the device will 
    present if it fails. FDA considers these activities essential for 
    ensuring that the manufacturing process does not adversely affect the 
    device.
        Design validation may also be necessary in earlier stages, prior to 
    product completion, and multiple validations may need to be performed 
    if there are different intended uses. Proper design validation cannot 
    occur without following all the requirements set forth in the design 
    control section of the regulation.
        82. Several comments stated that adequate controls for verification 
    of design output are contained in proposed Sec. 820.30(d), ``Design 
    verification,'' and repeated in proposed Sec. 820.30(f), ``Design 
    output.'' One comment stated that this section will place undue burden 
    on designers and require additional documentation which will add little 
    value to a device's safety and effectiveness.
        FDA disagrees with the comments. Revised Sec. 820.30(f), ``Design 
    verification,'' and Sec. 820.30(g), ``Design validation,'' require 
    verification and validation of the design output. Section 820.30(d), 
    ``Design output,'' requires that the output be documented in a fashion 
    that will allow for verification and validation. These sections thus 
    contain different requirements that are basic to establishing that the 
    design output meets the approved design requirements or inputs, 
    including user needs and intended uses. All the requirements are 
    essential to assuring the safety and effectiveness of devices. FDA does 
    not believe that these requirements place undue burden on designers or 
    require additional documentation with no value added. These basic 
    requirements are necessary to assure the proper device performance, 
    and, therefore, the production of safe and effective devices, and are 
    acknowledged and accepted as such throughout the world.
        83. Several comments stated that the term ``hazard analysis'' 
    should be defined in reference to design verification. A couple of 
    comments stated that the proposed requirement for design verification, 
    to include software validation and hazard analysis, where applicable, 
    was ambiguous, and may lead an FDA investigator to require software 
    validation and hazard analysis for devices in cases where it is not 
    needed. One comment stated that FDA should provide additional guidance 
    regarding software validation and hazard analysis and what 
    investigators will expect to see. Another comment stated that by 
    explicitly mentioning only software validation and hazard analysis, FDA 
    was missing the opportunity to introduce manufacturers to some powerful 
    and beneficial tools for better device designs and problem avoidance.
        FDA has deleted the term ``hazard analysis'' and replaced it with 
    the term ``risk analysis.'' FDA's involvement with the ISO TC 210 made 
    it clear that ``risk analysis'' is the comprehensive and appropriate 
    term. When conducting a risk analysis, manufacturers are expected to 
    identify possible hazards associated with the design in both normal and 
    fault conditions. The risks associated with the hazards, including 
    those resulting from user error, should then be calculated in both 
    normal and fault conditions. If any risk is judged unacceptable, it 
    should be reduced to acceptable levels by the appropriate
    
    [[Page 52621]]
    
    means, for example, by redesign or warnings. An important part of risk 
    analysis is ensuring that changes made to eliminate or minimize hazards 
    do not introduce new hazards. Tools for conducting such analyses 
    include Failure Mode Effect Analysis and Fault Tree Analysis, among 
    others.
        FDA disagrees with the comments that state the requirement is 
    ambiguous. Software must be validated when it is a part of the finished 
    device. FDA believes that this control is always needed, given the 
    unique nature of software, to assure that software will perform as 
    intended and will not impede safe operation by the user. Risk analysis 
    must be conducted for the majority of devices subject to design 
    controls and is considered to be an essential requirement for medical 
    devices under this regulation, as well as under ISO/CD 13485 and EN 
    46001. FDA has replaced the phrase ``where applicable'' with ``where 
    appropriate'' for consistency with the rest of the regulation.
        FDA believes that sufficient domestic and international guidelines 
    are available to provide assistance to manufacturers for the validation 
    of software and risk analysis. For example, ``Reviewer Guidance for 
    Computer Controlled Medical Devices Undergoing 510(k) Review,'' August 
    1991; ``A Technical Report, Software Development Activities,'' July 
    1987; and ISO-9000-3 contain computer validation guidance. Further, FDA 
    is preparing a new ``CDRH Guidance for the Scientific Review of Pre-
    Market Medical Device Software Submissions.'' Regarding guidance on 
    ``risk analysis,'' manufacturers can reference the draft EN (prEN) 
    1441, ``Medical Devices--Risk Analysis'' standard and the work 
    resulting from ISO TC 210 working group No. 4 to include ISO/CD 14971, 
    ``Medical Devices--Risk Management--Application of Risk Analysis to 
    Medical Devices.''
        FDA disagrees that it is missing the opportunity to introduce 
    manufacturers to some powerful and beneficial tools for better device 
    designs and problem avoidance because the manufacturer must apply 
    current methods and procedures that are appropriate for the device, to 
    verify and validate the device design under the regulation. Therefore, 
    FDA need not list all known methods for meeting the requirements. A 
    tool that may be required to adequately verify and validate one design 
    may be unnecessary to verify and validate another design.
        84. One comment stated that for some design elements it may be more 
    appropriate to reference data from another prior experimentation rather 
    than conduct new testing, and that the requirement to list verification 
    methods should be modified.
        FDA agrees in part with the comment. The revised language of 
    Sec. 820.30(f) will permit the use of data from prior experimentation 
    when applicable. When using data from previous experimentation, 
    manufacturers must ensure that it is adequate for the current 
    application.
        85. ``Design transfer,'' now Sec. 820.30(h), has been revised in 
    response to the many comments objecting to the requirements in the 
    proposed section on ``Design transfer.'' Specifically, the proposed 
    requirement for testing production units under actual or simulated use 
    conditions was rewritten and moved to current Sec. 820.30(g), ``Design 
    validation.''
        FDA again emphasizes that testing production units under actual or 
    simulated use conditions prior to distribution is crucial for ensuring 
    that only safe and effective devices are distributed and FDA has 
    therefore retained the requirement. ISO 9001:1994 discusses this 
    concept in notes 12 and 13. As noted above, it is not always possible 
    to determine the adequacy of the design by successfully building and 
    testing prototypes or models produced in a laboratory setting.
        The requirement for testing from the first three production lots or 
    batches has been deleted. While FDA believes that three production runs 
    during process validation (process validation may be initiated before 
    or during design transfer) is the accepted standard, FDA recognizes 
    that all processes may not be defined in terms of lots or batches. The 
    number three is, however, currently considered to be the acceptable 
    standard. Therefore, although the number requirement is deleted, FDA 
    expects validation to be carried out properly in accordance with 
    accepted standards, and will inspect for compliance accordingly.
        Revised Sec. 820.30(h) now contains a general requirement for the 
    establishment of procedures to ensure that the design basis for the 
    device is correctly translated into production methods and procedures. 
    This is the same requirement that is contained in Sec. 820.100(a) of 
    the original CGMP regulation.
        86. A few comments stated that the proposed requirements for 
    ``Design release'' would prohibit the release of components, partial 
    designs, and production methods before the design was final because the 
    requirements mandate a review of all drawings, analysis, and production 
    methods before allowing the product to go into production. Several 
    comments stated that the proposed section on ``Design release'' was a 
    duplication of requirements in other paragraphs of Sec. 820.30 and 
    should be deleted.
        FDA did not intend the requirements for ``Design release'' to 
    prohibit manufacturers from beginning the production process until all 
    design activities were completed. The intent of the requirement was to 
    ensure that all design specifications released to production have been 
    approved, verified, and validated before they are implemented as part 
    of the production process. This requirement is now explicitly contained 
    in Sec. 820.30(d).
        FDA agrees in part with the second set of comments and has moved 
    the requirement that design output be reviewed and approved to current 
    Sec. 820.30(d), ``Design output.'' The remainder of the requirements 
    have been deleted.
        87. Several comments on Sec. 820.30(i), ``Design changes,'' stated 
    that it is unnecessary to control all design changes and to do so would 
    inhibit change and innovation.
        FDA disagrees with the comments. Manufacturers are not expected to 
    maintain records of all changes proposed during the very early stages 
    of the design process. However, all design changes made after the 
    design review that approves the initial design inputs for incorporation 
    into the design, and those changes made to correct design deficiencies 
    once the design has been released to production, must be documented. 
    The records of these changes create a history of the evolution of the 
    design, which can be invaluable for failure investigation and for 
    facilitating the design of future similar products. Such records can 
    prevent the repetition of errors and the development of unsafe or 
    ineffective designs. The evaluation and documentation should be in 
    direct proportion to the significance of the change. Procedures must 
    ensure that after the design requirements are established and approved, 
    changes to the design, both pre- production and post-production are 
    also reviewed, validated (or verified where appropriate), and approved. 
    Otherwise, a device may be rendered unable to properly perform, and 
    unsafe and ineffective. ISO 9001:1994, section 4.4.9, similarly 
    provides that ``all design changes and modifications shall be 
    identified, documented, reviewed, and approved by authorized personnel 
    before their implementation.''
        Note that when a change is made to a specification, method, or 
    procedure, each manufacturer should evaluate the
    
    [[Page 52622]]
    
    change in accordance with an established procedure to determine if the 
    submission of a premarket notification (510(k)) under Sec. 807.81(a)(3) 
    (21 CFR 807.81(a)(3)), or the submission of a supplement to a PMA under 
    Sec. 814.39(a) (21 CFR 814.39) is required. Records of this evaluation 
    and its results should be maintained.
        88. Several comments recommended that only changes after design 
    validation and design transfer to full-scale production need to be 
    documented.
        FDA disagrees with the comments. The safety and effectiveness of 
    devices cannot be proven by final inspection or testing. Product 
    development is inherently an evolutionary process. While change is a 
    healthy and necessary part of product development, quality can be 
    ensured only if change is controlled and documented in the development 
    process, as well as the production process. Again, manufacturers are 
    not expected to maintain records of changes made during the very early 
    stages of product development; only those design changes made after the 
    approval of the design inputs need be documented. Each manufacturer 
    must establish criteria for evaluating changes to ensure that the 
    changes are appropriate for its designs.
        89. One comment on proposed Sec. 820.30(i), ``Design changes,'' 
    stated that validation of design changes is not always necessary and 
    the regulation should provide for other methods to be used. FDA agrees 
    with the comments and has amended the requirement to permit 
    verification where appropriate. For example, a change in the 
    sterilization process of a catheter will require validation of the new 
    process, but the addition of more chromium to a stainless steel 
    surgical instrument may only require verification through chemical 
    analysis. Where a design change cannot be verified by subsequent 
    inspection and test, it must be validated.
        90. Many comments noted that the acronym for proposed design 
    history record (DHR) was the same as that of ``device history record'' 
    (DHR), and suggested that the name of the ``design history record'' be 
    changed. Several comments stated that the requirements of the ``design 
    history record'' should be deleted because they were redundant with the 
    requirements of the ``device master record.''
        FDA agrees with the first set of comments and has changed the name 
    to ``design history file.''
        FDA disagrees with the second set of comments. The DMR contains the 
    documentation necessary to produce a device. The final design output 
    from the design phase, which is maintained or referenced in the DHF, 
    will form the basis or starting point for the DMR. Thus, those outputs 
    must be referred to or placed in the DMR. The total finished design 
    output includes the final device, its labeling and packaging, and the 
    DMR that includes device specifications and drawings, as well as all 
    instructions and procedures for production, installation, maintenance, 
    and servicing. The DHF, in contrast, contains or references all the 
    records necessary to establish compliance with the design plan and the 
    regulation, including the design control procedures. The DHF 
    illustrates the history of the design, and is necessary so that 
    manufacturers can exercise control over and be accountable for the 
    design process, thereby maximizing the probability that the finished 
    design conforms to the design specifications.
        91. A few comments stated that the proposed requirements in 
    Sec. 820.30(j) for the design history record should allow a single 
    design history record for each device family or group having common 
    design characteristics.
        FDA agrees with the comments. The intent of the DHF is to document, 
    or reference the documentation of, the activities carried out to meet 
    the design plan and requirements of Sec. 820.30. A DHF is, therefore, 
    necessary for each type of device developed. The DHF must provide 
    documentation showing the actions taken with regard to each type of 
    device designed, not generically link devices together with different 
    design characteristics and give a general overview of how the output 
    was reached.
        92. Some comments stated that the requirement that the DHF contain 
    ``all'' records necessary to demonstrate that the requirements are met 
    should be deleted because not ``all'' efforts need documentation.
        FDA received similar comments on almost every section of the 
    regulation that had the word ``all.'' The proposed requirement does not 
    state that all records must be contained in the DHF, but that all 
    records necessary to demonstrate that the requirements were met must be 
    contained in the file. FDA has deleted the word ``all'' but cautions 
    manufacturers that the complete history of the design process should be 
    documented in the DHF. Such records are necessary to ensure that the 
    final design conforms to the design specifications. Depending on the 
    design, that may be relatively few records. Manufacturers who do not 
    document all their efforts may lose the information and experience of 
    those efforts, thereby possibly requiring activities to be duplicated.
    
    D. Document Controls (Subpart D)
    
        93. One comment stated that subpart D of part 820 should be titled 
    ``Document Controls,'' instead of the proposed ``Document and Record 
    Controls'' because the ``record'' requirements are addressed in subpart 
    M. One comment stated that removal of obsolete or unneeded documents 
    should be performed to maintain the integrity of the product 
    configuration and the quality system. The comment suggested adding a 
    requirement for a verification step for document distribution and 
    removal to ensure this important element of a quality system is 
    performed correctly. A few comments stated that proposed Sec. 820.40 
    should be rewritten to be similar to ISO 9001 and to delete the 
    requirement that documents be ``accurate'' because the comments feared 
    that typographical errors would be considered violations.
        FDA agrees with the first comment and has changed the title 
    accordingly. FDA agrees in part with the second comment. The 
    verification of document distribution and removal is very important and 
    can directly affect the quality of a product. Section 820.40, which 
    requires that the manufacturer establish and maintain procedures to 
    control all documents, including those that are obsolete and/or to be 
    removed, requires that the removal (or prevention of use) of obsolete 
    documents be verified. FDA agrees in part with the last set of comments 
    and has rewritten the section, following ISO 9001:1994, to be a general 
    requirement for procedures to control documents that are required under 
    the regulation. The procedures established must, among other things, 
    ensure control of the accuracy and usage of current versions of the 
    documents and the removal or prevention from use of obsolete documents, 
    as well as ensure that the documentation developed is adequate to 
    fulfill its intended purpose or requirement. FDA retained the 
    requirement that the procedures ensure that documents meet the 
    requirements of the regulation because that is the purpose of 
    controlling the documents. FDA deleted the term ``accurate'' but notes 
    that a typographical error can change the meaning of a document and 
    have undesirable consequences.
        94. Several comments on proposed Sec. 820.40(a), ``Document 
    approval and issue,'' as well as other sections throughout the 
    regulation, suggested that the term ``signature'' be replaced by the 
    term ``identification.'' Such a change would allow for electronic or 
    computerized identification in lieu of formal written signatures. Other 
    comments stated that ``or stamps''
    
    [[Page 52623]]
    
    should be added after ``signature'' since they are legally recognized 
    in some foreign countries.
        FDA is aware that many documentation systems are now maintained 
    electronically, and is in the process of developing an agency-wide 
    policy that will be implemented through rulemaking on the use of 
    electronic signatures. The agency identified several important issues 
    related to the use of such signatures, including how to ensure that the 
    identification is in fact the user's ``signature.'' These issues are 
    discussed in FDA's ANPRM on the use of electronic signatures, published 
    in the Federal Register on July 21, 1992 (57 FR 32185), and the 
    proposed regulation published in the Federal Register on August 31, 
    1994 (59 FR 45160). Therefore, FDA has not revised the regulation to 
    use the term ``identification,'' but notes that the quality system 
    regulation's use of the term ``signature'' will permit the use of 
    whatever electronic means the agency determines is the equivalent of a 
    handwritten signature. FDA recommends that manufacturers use the two 
    Federal Register documents as guidance until the regulation is 
    finalized. FDA has not added the term ``or stamps'' to the regulation; 
    however, stamps could be acceptable if the manufacturer has a formal 
    procedure on how stamps are used in place of handwritten signatures. 
    The procedure would have to address many of the same issues addressed 
    in the electronic signature Federal Register documents, most 
    importantly how the stamps would be controlled and how the manufacturer 
    would ensure that the stamp was in fact the user's ``signature.''
        95. Several comments stated that proposed Sec. 820.40(b), 
    ``Document distribution,'' should be rewritten to be consistent with 
    ISO 9001.
        In response, FDA has deleted the section. The requirements for 
    making documents available at all appropriate locations (ISO 9001:1994, 
    section 4.5.2(a)) and the requirements for promptly removing obsolete 
    documents (ISO 9001:1994, section 4.5.2(b)) have been moved, in revised 
    form, to Sec. 820.40(a). In response to comments, FDA has added that 
    obsolete documents, in lieu of being promptly removed from points of 
    use, may be ``otherwise prevented from unintended use.''
        96. Several comments suggested major changes to proposed 
    Sec. 820.40(c), ``Documentation changes.'' Some stated that the 
    requirements should be revised to be consistent with ISO 9001. Others 
    stated that the requirements related to validation should be rewritten 
    and moved to another section under this part, because Sec. 820.40(c) 
    should only address document changes, not device changes. Several 
    comments stated that the reference to determining whether a 510(k) or 
    PMA supplement is required after making changes to a device should be 
    deleted because it is covered under different parts of the act and 
    regulations. One comment stated that the requirement in Sec. 820.40(c) 
    for changes to be ``approved by individuals in the same functions/
    organizations that performed the original review and approval, unless 
    specifically designated otherwise'' is unrealistic and does not reflect 
    the way things are done in real life.
        FDA agrees with many of the comments and has substantially 
    rewritten Sec. 820.40(c), now designated as Sec. 820.40(b), to relate 
    specifically to changes to a document. The requirements are now very 
    similar to the ISO 9001:1994 requirements in section 4.5.3. FDA has 
    retained the requirement that the approved changes must be communicated 
    in a timely manner to appropriate personnel. FDA has had many 
    experiences where manufacturers made corrections to documents, but the 
    changes were not communicated in a timely manner to the personnel 
    utilizing the documents. The result of these untimely communications 
    was the production of defective devices.
        In addition, FDA has moved the requirement for validating 
    production and process changes to Sec. 820.70(b), ``Production and 
    process changes,'' and notes that changes to the design specifications, 
    at any time during the lifetime of the design of the device, must 
    conform to the requirements in Sec. 820.30(i), ``Design changes.''
        FDA has also deleted the sentence referencing 510(k)'s and PMA 
    supplements because FDA believes this is covered elsewhere, but notes 
    that this sentence is in the preamble above for Sec. 820.30(i).
        FDA disagrees that the requirement for changes to be ``approved by 
    an individual(s) in the same function or organization that performed 
    the original review and approval, unless specifically designated 
    otherwise'' should be deleted and notes that this is a requirement of 
    ISO 9001:1994 as well. The intent of the requirement is to ensure that 
    those who originally approved the document have an opportunity to 
    review any changes because these individuals typically have the best 
    insight on the impact of the changes. The requirement is flexible, 
    however, because it permits the manufacturer to specifically designate 
    individuals who did not perform the original review and approval to 
    review and approve the changes. To designate such individuals, the 
    manufacturer will need to determine who would be best suited to perform 
    the function, thus ensuring adequate control over the changes. In this 
    way, review and approval will not be haphazard.
        97. One comment on proposed Sec. 820.40(d), ``Documentation change 
    record,'' stated that this section should be deleted because the other 
    paragraphs of Sec. 820.40 adequately cover the proposed requirements. 
    Two comments suggested replacing the section with the requirements of 
    section 4.5.2 of ISO 9001.
        FDA has deleted Sec. 820.40(d) and placed the revised requirements 
    in paragraphs (a) and (b) of this section. The general requirement of 
    Sec. 820.40 now requires the manufacturer to establish adequate 
    procedures to control all documents required by part 820. The 
    procedures must cover the requirements listed in Sec. 820.40 (a) and 
    (b). Thus, the manufacturer must establish a procedure for ensuring 
    that only the current and approved version of a document is used, 
    achieving the objective of the ``Master list or equivalent document 
    control procedure,'' required in ISO 9001:1994, section 4.5.2.
        The other requirement in Sec. 820.40(d), ``Document change 
    record,'' was to maintain a record of changes, to include a description 
    of the changes, among other things. FDA has retained this requirement 
    and has moved it into Sec. 820.40(b), ``Document changes,'' because the 
    agency believes this information to be important and useful when 
    investigating and performing corrective or preventive actions.
        FDA believes Sec. 820.40 on Document controls now adequately 
    harmonizes with ISO 9001:1994, sections 4.5.1, 4.5.2, and 4.5.3.
    
    E. Purchasing Controls (Subpart E)
    
        98. One comment stated that the proposed CGMP regulation omits any 
    discussion of contract reviews, such as that contained in ISO 9001, 
    section 4.3. Rather than leaving these procedures to the 
    interpretations of individual manufacturers and investigators, the 
    comment stated that FDA should explicitly state its general policy 
    regarding contract reviews in the regulation.
        FDA agrees with the concepts underlying the contract review 
    requirements of ISO 9001:1994, but believes these principles are 
    already reflected in requirements in the regulation, such as 
    Secs. 820.50 Purchasing controls and 820.160 Distribution.
    
    [[Page 52624]]
    
    Therefore, the agency has not added a separate section on contract 
    review.
        99. One comment stated that the requirements in Sec. 820.50 amount 
    to overregulation. The comment stated that components are purchased by 
    providing a specification sheet. They are then inspected upon receipt, 
    and defective components are returned. According to the comment, under 
    Sec. 820.50, the manufacturer would be required to spend more time on 
    paperwork, and product would still have to be inspected upon receipt. 
    Another comment stated that the cost of the quality assurance 
    documentation program is going to be significantly higher for a company 
    that runs a Just In Time (JIT) program than what FDA estimated.
        FDA disagrees with the comments. The failure to implement adequate 
    purchasing controls has resulted in a significant number of recalls due 
    to component failures. Most of these were due to unacceptable 
    components provided by suppliers. Since FDA is not regulating component 
    suppliers, FDA believes that the explicit addition to CGMP requirements 
    of the purchasing controls of ISO 9001:1994 is necessary to provide the 
    additional assurance that only acceptable components are used. To 
    ensure purchased or otherwise received product or services conform to 
    specifications, purchasing must be carried out under adequate controls, 
    including the assessment and selection of suppliers, contractors, and 
    consultants, the clear and unambiguous specification of requirements, 
    and the performance of suitable acceptance activities. Each 
    manufacturer must establish an appropriate mix of assessment and 
    receiving acceptance to ensure products and services are acceptable for 
    their intended uses. The specifications for the finished device cannot 
    be met unless the individual parts of the finished device meet 
    specifications. The most efficient and least costly approach is to 
    ensure that only acceptable products and services are received. This 
    means that only suppliers, contractors, and consultants that meet 
    specifications should be used.
        The regulation has been written to allow more flexibility in the 
    way manufacturers may ensure the acceptability of products and 
    services. Under the requirements, manufacturers must clearly define in 
    the procedures the type and extent of control they intend to apply to 
    products and services. Thus, a finished device manufacturer may choose 
    to provide greater in-house controls to ensure that products and 
    services meet requirements, or may require the supplier to adopt 
    measures necessary to ensure acceptability, as appropriate. FDA 
    generally believes that an appropriate mix of supplier and manufacturer 
    quality controls are necessary. However, finished device manufacturers 
    who conduct product quality control solely in-house must also assess 
    the capability of suppliers to provide acceptable product. Where audits 
    are not practical, this may be done through, among other means, 
    reviewing historical data, monitoring and trending, and inspection and 
    testing.
        After evaluation of all of the comments on Sec. 820.50, FDA has 
    decided to change the wording of Sec. 820.50(a) and adopt the wording 
    of ISO 9001:1994 to make clear that manufacturers have flexibility in 
    determining the degree of assessment and evaluation necessary for 
    suppliers, contractors, and consultants. Thus the degree of supplier 
    control necessary to establish compliance may vary with the type and 
    significance of the product or service purchased and the impact of that 
    product or service on the quality of the finished device. In addition, 
    the requirement for manufacturers to establish assessment criteria has 
    been deleted but the evaluation still must include a description how 
    the assessment was made (according to what criteria or objective 
    procedure) and the results must be documented. Each manufacturer must 
    now define the type and extent of control it will exercise over 
    suppliers, contractors, and consultants. This is consistent with the 
    1994 version of ISO 9001.
        Thus, FDA believes that the flexibility of the regulation will 
    allow manufacturers to implement JIT procedures without additional 
    cost. In fact, the new regulation is more conducive to JIT practices by 
    permitting the assessment or evaluation of product or services up 
    front, thereby lessening the degree of in-house control that may be 
    necessary.
        100. Several comments said that it was unclear what FDA meant by 
    the phrase ``or held by other persons under contract conform to 
    specifications'' and that this phrase should be deleted.
        FDA agrees with the comments and has deleted the phrase. The phrase 
    was intended to mean product and services which were purchased or 
    processed in some manner by other organizations. Section 820.50 now 
    applies to ``purchased or otherwise received product and services'' to 
    convey this meaning. FDA emphasizes that the requirements apply to all 
    product and service received from outside of the finished device 
    manufacturer, whether payment occurs or not. Thus, a manufacturer must 
    comply with these provisions when it receives product or services from 
    its ``sister facility'' or some other corporate or financial affiliate. 
    ``Otherwise received product'' would include ``customer supplied 
    product'' as in ISO 9001:1994, section 4.7, but would not apply to 
    ``returned product'' from the customer.
        101. One comment stated that ``manufacturing materials'' should be 
    deleted from the first sentence of the introductory text of the 
    proposed Sec. 820.50, as the assessment of the manufacturers of 
    manufacturing materials would be a monumental task.
        FDA disagrees with the comment. The first sentence of the 
    introductory text of Sec. 820.50 is rewritten to be a general 
    requirement that each manufacturer must establish procedures to ensure 
    that received product and services (purchased or otherwise received) 
    conform to specified requirements. All manufacturers are expected to 
    apply controls to manufacturing materials appropriate to the 
    manufacturing material, the intended use, and the effect of the 
    manufacturing materials on safety and effectiveness. For example, the 
    procedures necessary to ensure that a mold release agent conforms to 
    specified requirements may be less involved than the procedures for 
    controlling latex proteins. The provision allows the manufacturer the 
    flexibility of establishing the procedures to meet its needs and to 
    ensure that the product conforms to specified requirements.
        102. One comment said that FDA should delete the last sentence of 
    the introductory text of proposed Sec. 820.50 because it is unnecessary 
    for manufacturers to develop specifications for services that are 
    unrelated to product or process quality, and because the terms 
    ``service'' and ``other persons'' lack definition. Other comments 
    stated that ``all'' should be deleted in the general requirement.
        FDA disagrees with the comments. First, as used in the regulation, 
    ``service'' means parts of the manufacturing or quality system that are 
    contracted to others, for example, plating of metals, testing, and 
    sterilizing, among others. Second, FDA believes that all suppliers of 
    such services must be assessed and evaluated, just like a supplier of a 
    product. As always, the degree of control necessary is related to the 
    product or service purchased. FDA has, however, deleted the term 
    ``provided by other persons'' because it was unnecessary. FDA did not 
    delete the word ``all'' because, as discussed above, component 
    manufacturers are not subject to this regulation, so it is the
    
    [[Page 52625]]
    
    finished device manufacturer who is responsible for ``all'' product and 
    services.
        103. One comment stated that many suppliers of components to the 
    medical device industry have their quality systems certified to an ISO 
    9000 standard by an independent third party auditor, and that such 
    registration of component manufacturers should be considered in vendor 
    assessment plans.
        FDA agrees in part with the comment in that certification may play 
    a role in evaluating suppliers, but cautions manufacturers against 
    relying solely on certification by third parties as evidence that 
    suppliers have the capability to provide quality products or services. 
    FDA has found during inspections that some manufacturers who have been 
    certified to the ISO standards have not had acceptable problem 
    identification and corrective action programs. Therefore, the initial 
    assessment or evaluation, depending on the type and potential effect on 
    device quality of the product or service, should be a combination of 
    assessment methods, to possibly include third party or product 
    certification. However, third party certification should not be relied 
    on exclusively in initially evaluating a supplier. If a device 
    manufacturer has established confidence in the supplier's ability to 
    provide acceptable products or services, certification with test data 
    may be acceptable.
        104. Some comments stated that consultants should not be included 
    in the regulation at all. Others stated that it was not consistent with 
    ISO 9001.
        FDA added ``consultants'' to Sec. 820.50(a) in response to the 
    comments from Sec. 820.25(c). FDA disagrees that ``consultants'' should 
    be deleted because over the years FDA has observed that a surprising 
    number of firms hire consultants who have no particular expertise in 
    the area in which the firm is seeking assistance. Section 820.50 
    addresses this problem by ensuring that a consultant's capability for 
    the specific tasks for which he or she is retained be assessed and 
    documented. Further, FDA does not believe this requirement is 
    inconsistent with ISO 9001:1994 because ISO uses the term 
    ``subcontractor.'' The term ``subcontractor'' includes consultants.
        105. One comment said that requiring evaluation of potential 
    suppliers, contractors, and consultants ``on the basis of their ability 
    to meet requirements'' is vague and should be clearly defined.
        FDA disagrees that the phrase is vague. Suppliers, contractors, and 
    consultants selected by manufacturers of medical devices should have a 
    demonstrated capability of providing products and services that meet 
    the requirements established by the finished device manufacturer. The 
    capability of the product or service suppliers should be reviewed at 
    intervals consistent with the significance of the product or service 
    provided and the review should demonstrate conformance to specified 
    requirements.
        106. One comment questioned the usefulness of Sec. 820.50, given 
    that the requirements under Sec. 820.80 Receiving, in-process, and 
    finished device acceptance, require manufacturers to establish and 
    maintain procedures for acceptance of incoming components.
        The intent of Sec. 820.50 is to ensure that device manufacturers 
    select only those suppliers, contractors, and consultants who have the 
    capability to provide quality product and services. As with finished 
    devices, quality cannot be inspected or tested into products or 
    services. Rather, the quality of a product or service is established 
    during the design of that product or service, and achieved through 
    proper control of the manufacture of that product or the performance of 
    that service. Section 820.50 thus mandates that products be 
    manufactured and services be performed under appropriate quality 
    assurance procedures. Finished device manufacturers are required under 
    Sec. 820.50 to establish the requirements for, and document the 
    capability of, suppliers, contractors, and consultants to provide 
    quality products and services.
        Section 820.80 is specific to a device manufacturer's acceptance 
    program. While finished device manufacturers are required to assess the 
    capability of suppliers, contractors, and consultants to provide 
    quality products and services, inspections and tests, and other 
    verification tools, are also an important part of ensuring that 
    components and finished devices conform to approved specifications. The 
    extent of incoming acceptance activities can be based, in part, on the 
    degree to which the supplier has demonstrated a capability to provide 
    quality products or services. An appropriate product and services 
    quality assurance program includes a combination of assessment 
    techniques, including inspection and test.
        107. Several comments stated that it was not clear how a 
    manufacturer could evaluate an off-the-shelf component that is 
    purchased from a distributor rather than directly from its 
    manufacturer, and stated that it would not be helpful to audit the 
    distributor.
        FDA agrees that auditing a distributor would not meet the intent of 
    Sec. 820.50. Manufacturers should remember that the purpose of 
    assessing the capability of suppliers is to provide quality products 
    and to provide a greater degree of assurance, beyond that provided by 
    receiving inspection and test, that the products received meet the 
    finished device manufacturer's requirements. The agency recognizes that 
    finished device manufacturers may not always be able to audit the 
    supplier of a product. In such cases, the manufacturer must apply other 
    effective means to assure that products are acceptable for use.
        108. Many comments from both domestic and foreign firms in response 
    to proposed Sec. 820.22(b) said that making supplier audit reports 
    subject to FDA review would have a major adverse impact on the 
    relationships between the finished device manufacturers and their 
    suppliers and service providers. Some stated that the requirement would 
    cause suppliers to refuse to sell components to medical device 
    manufacturers, especially suppliers who provide only a small part of 
    their production to device manufacturers. Others said that this policy 
    is not consistent with FDA's policy for internal audits.
        FDA recognizes that quality audits of suppliers have a significant 
    and demonstrated value as a management tool for corrective action, 
    quality improvement, and overall assurance of component and service 
    quality, and does not seek to undermine their value. Therefore, based 
    on the concerns raised by the comments, FDA will not review supplier 
    audit reports during a routine FDA inspection for compliance with part 
    820, as noted in Sec. 820.180(c), ``Exceptions.'' The audit procedures, 
    the evaluation procedures, and documents other than the supplier audit 
    reports themselves that demonstrate conformance with Sec. 820.50 will 
    be subject to review by an FDA investigator.
        109. One comment stated that it was unclear what is meant by the 
    requirement to specify ``quality requirements'' that must be met by 
    suppliers, contractors, and consultants, as stated in Sec. 820.50(a).
        The term ``quality requirements'' means the quality control and 
    quality assurance procedures, standards, and other requirements 
    necessary to assure that the product or service is adequate for its 
    intended use. FDA does not believe the term is unclear.
        110. Several comments on proposed Sec. 820.50(b), ``Purchasing 
    forms,'' suggested that the term ``forms'' be replaced by ``data.'' 
    Other comments stated that use of the term would not allow electronic 
    data exchange. One comment stated that the use of an
    
    [[Page 52626]]
    
    exclusive form for purchasing is unnecessary and redundant, and that it 
    is unduly burdensome to require detailed documentation on those 
    commonly available items such as fasteners. The comment stated that it 
    is common practice to use prints or drawings to fulfill the purpose of 
    the form.
        FDA agrees in part with the comments, but does not believe that 
    Sec. 820.50(b) prohibits the use of drawings or prints, assuming that 
    the documents contain data clearly describing the product or service 
    ordered, and that the specified requirements are met. However, 
    Sec. 820.50(b) has been rewritten and now requires manufacturers to 
    establish purchasing ``data.'' This provides manufacturers with the 
    flexibility to use both written and electronic means to establish 
    purchasing information.
        111. One comment stated that the inclusion of an additional 
    provision mandating that suppliers notify manufacturers of any change 
    in their product or service places an undue burden on suppliers and 
    inhibits their ability to make minor adjustments within the parameters 
    of agreed upon specifications and quality requirements. Many other 
    comments stated that the requirement in Sec. 820.50(b) is feasible only 
    for components that are custom made for the manufacturer, and is 
    meaningless for off-the-shelf components purchased from distributors. 
    Other comments stated that the requirement is part of the original CGMP 
    regulation and experience has shown that suppliers are not willing to 
    supply device manufacturers with such information. A few other comments 
    stated that ``any'' should be deleted because the term is too broad and 
    could result in burdensome reporting of variables which are irrelevant 
    to the continued performance or specifications of the product or 
    service.
        FDA agrees in part with the comments and has amended the 
    requirement to state that such agreement should be obtained ``where 
    possible.'' FDA still believes that this change information is very 
    important to the manufacturer, and that the manufacturer should obtain 
    information on changes to the product or service. Where a supplier 
    refuses to agree to provide such notification, depending on the product 
    or service being purchased, it may render him an unacceptable supplier. 
    However, where the product is in short supply and must be purchased, 
    the manufacturer will need to heighten control in other ways.
        FDA has also deleted the term ``any'' to give manufacturers the 
    flexibility to define in the agreement the types of changes that would 
    require notification.
        112. One comment stated that Sec. 820.50(b) should incorporate a 
    provision that would allow manufacturers to cite published standards in 
    purchasing forms as one suitable method for specifying purchased item 
    quality requirements.
        FDA believes the addition is unnecessary, because the regulation 
    permits manufacturers to clearly describe or reference requirements. A 
    reference could be to a standard.
        113. One comment stated that it is unclear whether the requirement 
    for a signature to approve purchasing documents pertains to approval of 
    the form used for purchasing or approval of the individual purchasing 
    transaction. The comment also stated that a signature approval by 
    transaction is not practical for firms using electronic document 
    transmittals.
        FDA has rewritten the requirement to be more clear. The requirement 
    is for approval of purchasing data or information on the purchasing 
    document used to purchase a product or service. Thus, each manufacturer 
    must review and approve the purchasing data before release of the data. 
    Approval of each purchasing transaction is not required. FDA addressed 
    the use of electronic signatures in response to another comment, and 
    notes that FDA is in the process of developing an agency-wide policy on 
    the use of electronic signatures.
        114. One comment stated that purchasing is carried out verbally in 
    many small firms, without the use of component-specific purchasing 
    forms, and that the regulation should be revised to allow such verbal 
    purchasing to continue.
        FDA disagrees with the comment. About 15 percent of the recalls 
    each year are due to unacceptable purchased products. Many of these 
    products are unacceptable because the finished device manufacturer did 
    not properly describe the product. The requirements for purchased 
    products and services must be documented to ensure that the supplier, 
    contractor, and consultant provide a product or service which conforms 
    to specified requirements. This requirement, and the goal it seeks to 
    achieve, are applicable to both small and large companies.
        115. One comment stated that the requirement that purchasing forms 
    spell out the specifications for manufacturing materials in all cases 
    is excessive, and that the need for specifications should be based on 
    the criticality of and risk associated with the use of the specific 
    manufacturing material.
        FDA agrees that the specifications for many manufacturing materials 
    may be so well established that the trade name of the product may be 
    sufficient to describe the material needed. For other materials, 
    specific written specifications may be necessary to ensure that the 
    desired materials are received. The extent of the specification detail 
    necessary to ensure that the product or service purchased meets 
    requirements will be related to the nature of the product or service 
    purchased, taking into account the effect the product or service may 
    have on the safety or effectiveness of the finished device, among other 
    factors. The term ``specification'' has been replaced with the term 
    ``specified requirements'' to better reflect the intent of the 
    requirement.
        116. FDA has deleted the last two sentences of Sec. 820.50(b) in 
    the Working Draft and has replaced them with a reference to 
    Sec. 820.40, the general document control provision. This does not 
    change the requirement but simply eliminates any confusion about the 
    reviews and approvals being duplicative.
    
    F. Identification and Traceability (Subpart F)
    
    i. Identification (Sec. 820.60)
        117. A few comments on proposed Secs. 820.60 Identification and 
    traceability and 820.65 Critical device, traceability stated that the 
    two sections should be rewritten to delete the distinction between 
    critical and noncritical devices. Some stated they should be consistent 
    with ISO.
        FDA agrees in part with the comments and has rewritten Sec. 820.60 
    to be consistent with ISO 9001:1994 and broad enough to allow the 
    manufacturer the flexibility needed to identify product by whatever 
    means described by the required procedure. The term ``critical device'' 
    has also been deleted, and traceability is addressed solely in 
    Sec. 820.65.
        118. One comment stated that manufacturing materials should be 
    deleted from Sec. 820.60, as the requirements are excessive and not 
    economically justifiable with regard to such materials.
        FDA disagrees with the comment. The purpose of Sec. 820.60 is to 
    ensure that all products, including manufacturing materials used in the 
    manufacture of a finished device, are properly identified. This 
    requirement is intended to help prevent inadvertent use or release of 
    unacceptable product into manufacturing. It is as important that the 
    proper manufacturing materials be
    
    [[Page 52627]]
    
    used as it is that the proper component be used.
        119. A few comments thought that Sec. 820.60 Identification in the 
    Working Draft was redundant with Sec. 820.86 Acceptance status.
        FDA disagrees with the comments. Section 820.60 only requires that 
    product be identified but says nothing about the acceptance status of 
    that product. Section 820.86 requires that the acceptance status be 
    identified so that inadvertent use of product does not occur. The 
    manufacturer may choose to set up a system by which the identification 
    required by Sec. 820.60 can also show the acceptance status required by 
    Sec. 820.86, but this is up to the manufacturer.
    ii. Traceability (Sec. 820.65)
        120. A few comments stated that proposed Sec. 820.65 Critical 
    devices, traceability implies that traceability requirements exist for 
    all devices. Several other written comments and oral testimony at the 
    August and September 1995 meetings stated that the wording of the 
    Working Draft was too broad, vague, and ambiguous, and in effect would 
    require that all devices be traced.
        As noted above, FDA has deleted the critical device terminology. 
    Section 820.65 is now entitled Traceability and uses the definition 
    from the original CGMP of a critical device to provide the necessary 
    clarity and delineation for this requirement. Thus, traceability is 
    required for the critical devices listed in the Federal Register notice 
    of March 17, 1988 (53 FR 8854). However, FDA is using the definition of 
    critical device in the requirement of Sec. 820.65, rather than a 
    reference to the 1988 list of critical devices, because that list has 
    not been updated since 1988 and there are no plans to revise that list. 
    Therefore, it is imperative that manufacturers use the definition 
    within the requirement of Sec. 820.65 to determine if a particular 
    device needs to be traced; it may not be sufficient to rely solely on 
    the 1988 list. Manufacturers may find it advantageous to provide unit, 
    lot, or batch traceability for devices for which traceability is not a 
    requirement to facilitate control and limit the number of devices that 
    may need to be recalled due to defects or violations of the act.
        It is important that the traceability requirements in part 820 are 
    not confused with the Medical Device Tracking regulation in part 821 
    (21 CFR part 821). The tracking regulation is intended to ensure that 
    tracked devices can be traced from the device manufacturing facility to 
    the person for whom the device is indicated, that is, the patient. 
    Effective tracking of devices from the manufacturing facility, through 
    the distribution network (including distributors, retailers, rental 
    firms and other commercial enterprises, device user facilities, and 
    licensed practitioners) and, ultimately, to any person for whom the 
    device is intended is necessary for the effectiveness of remedies 
    prescribed by the act, such as patient notification (section 518(a) of 
    the act (21 U.S.C. 360h(a)) or device recall (section 518(e).) In 
    contrast, the traceability provision requires that a device that meets 
    the definition of a ``critical device'' can be traced from the 
    manufacturing facility only to the ``initial consignee'' as discussed 
    in Sec. 820.160 Distribution.
        121. Another comment on proposed Sec. 820.65 stated that critical 
    device component traceability could be interpreted to be required for 
    almost all electronic components and other components in a critical 
    device. The comment stated that the extent of component traceability 
    should be left to the manufacturer's discretion, since it is an 
    economic risk decision. Several comments stated that component 
    traceability should only be required ``where appropriate,'' that all 
    ``critical device'' components do not require traceability to comply 
    with the act.
        FDA disagrees that the traceability determination should be based 
    solely on economic risk. As noted in the preamble to the November 23, 
    1993, proposal (58 FR 61964), where traceability is important to 
    prevent the distribution of devices that could seriously injure the 
    user, traceability of components must be maintained so that potential 
    and actual problem components can be traced back to the supplier. The 
    revised requirement mandates traceability of components ``where 
    appropriate'' as recommended by the GMP Advisory Committee and limited 
    by the discussion in the scope, Sec. 820.1(a)(3). The critical 
    component definition in the original CGMP regulation may be used as 
    guidance. However, to carry out the requirement of the revised 
    provision, the manufacturer should perform risk analysis first on the 
    finished device, and subsequently on the components of such device, to 
    determine the need for traceability. FDA believes that the extent of 
    traceability for both active and inactive implantable devices should 
    include all components and materials used when such products could 
    cause the medical device not to satisfy its specified requirements. 
    ISO/CD 13485 also requires that the manufacturer's agents or 
    distributors maintain records of distribution of medical devices with 
    regard to traceability and that such records be available for 
    inspection. This requirement is found in Sec. 820.160 Distribution of 
    this regulation and is consistent with the requirements in Sec. 820.151 
    of the original CGMP.
        While FDA understands that traceability entails additional cost, 
    the agency notes that, if a product recall is necessary, more devices 
    would be subject to recall if units, lots, or batches of specific 
    devices are not traceable, with associated higher recall costs to the 
    manufacturer.
    
    G. Production and Process Controls (Subpart G)
    
    i. Production and Process Controls (Sec. 820.70)
        122. A few comments stated that the requirements in proposed 
    Sec. 820.70(a) General are similar to those in ISO 9001, but that ISO 
    9001 makes clear that the requirements apply only ``where applicable'' 
    and where deviations from device specifications would ``directly affect 
    quality.'' The comments suggested that FDA similarly employ such 
    language to avoid being too restrictive and overly burdensome.
        The requirements in Sec. 820.70(a) are intended to ensure that each 
    manufacturer produces devices that conform to their specifications. 
    Thus, where any deviations from specifications could occur during 
    manufacturing, the process control procedures must describe those 
    controls necessary to ensure conformance. Those controls listed in the 
    regulation may not always be relevant; similarly others may be 
    necessary. For example, where deviations from device specifications 
    could occur as a result of the absence of written production methods, 
    procedures, and workmanship criteria, such production controls are 
    required. Thus, FDA has retained the provision, but revised it slightly 
    to conform with the original CGMP requirements in Sec. 820.100(b)(1).
        As noted, the process control requirements apply when any deviation 
    from specifications could occur. FDA believes that such deviations must 
    be controlled, and that linking the requirements to deviations that 
    directly affect quality is inappropriate and subjective, and that it 
    could lead to the manufacture of potentially dangerous devices through 
    the lack of control of processes known to directly affect a device's 
    specifications. Therefore, the provision has not been restricted in 
    this manner. FDA has, however, revised the requirements to state 
    ``Where process controls are needed they shall include:'' to make it 
    clear that a manufacturer only has to comply with the requirements
    
    [[Page 52628]]
    
    stated in Sec. 820.70 (a)(1) through (a)(5) if the general criteria 
    described in Sec. 820.70(a) have been met.
        123. One comment stated that the second sentence of proposed 
    Sec. 820.70(a) was too restrictive, in that some processes can be 
    accomplished by adequately trained personnel without the use of 
    procedures.
        FDA disagrees with the comment because the establishment of 
    procedures is necessary to ensure consistency in manufacture. The 
    procedures may be tailored under the requirement to cover only those 
    controls necessary to ensure that a device meets its specifications. 
    FDA notes that the deletion of the word ``all'' does not alter the 
    requirements. The first sentence in the general requirement also serves 
    to tie the production and process controls to the design and 
    development phase where many of these controls are originally 
    established in order for the device to conform to its design 
    specifications.
        In addition to these changes, FDA has added the requirement that 
    production processes be ``monitored'' because a manufacturer must 
    monitor a controlled process to ensure that the process remains in 
    control.
        124. FDA deleted the requirement for process controls related to 
    ``installation and servicing'' from proposed Sec. 820.70 (a)(1) and 
    (a)(2) in response to comments. Such control is adequately assured by 
    the requirements in Secs. 820.170 Installation and 820.200 Servicing. 
    FDA amended Sec. 820.70(a)(3) in response to some comments that were 
    confused about compliance with ``applied reference standards.'' The 
    term ``applied'' was replaced with ``specified'' to make it clear that 
    the manufacturer must comply with reference standards or codes which he 
    or she has specified in the DMR. FDA has also deleted ``and process 
    control procedures'' because that requirement is inherent in 
    Sec. 820.70(a), ``General.'' FDA amended Sec. 820.70(a)(5) by adding 
    ``identified and approved'' in response to comments and to clarify that 
    the ``representative samples'' have to be identified and deemed 
    appropriate before they are used as reference standards.
        125. One comment believed that there is no longer a requirement 
    that process changes be validated. Other comments on the Working Draft 
    Sec. 820.70(b) stated the requirement was still confusing with respect 
    to ``unless inspection and test fully verifies,'' and when the 
    ``approval'' was to occur.
        Revised Sec. 820.70(b), ``Production and process changes,'' 
    addresses the requirement for production and process changes to be 
    ``verified or where appropriate validated according to Sec. 820.75.'' 
    This requirement for validation was moved from Sec. 820.40(c), in 
    revised form, to Sec. 820.70. Verification was added to give the 
    manufacturer the flexibility to verify changes that can be tested and 
    inspected because FDA believes that validation is not always necessary. 
    FDA has provided guidance on when changes should be validated in its 
    ``Guideline on General Principles of Process Validation.'' The agency 
    notes that wherever changes may influence a validated process, the 
    process must be revalidated as described in Sec. 820.75. A few examples 
    of processes that must be validated include sterilization, molding, and 
    welding.
        FDA has deleted the last part in Sec. 820.70(b) of the Working 
    Draft about approving changes and has replaced it with ``Changes shall 
    be approved in accordance with Sec. 820.40.'' This does not change the 
    requirement but simply refers back to Sec. 820.40 because this requires 
    the same review and approval. This was done to eliminate any confusion 
    about the reviews and approvals being duplicative.
        126. The EU Commission and others stated that environmental 
    conditions only affect the quality of certain devices and that the 
    requirements should, therefore, be limited in their application. Other 
    comments stated that the requirements in proposed Sec. 820.70(b), 
    ``Environmental control,'' were not consistent with the requirements in 
    the original CGMP, Sec. 820.46. Another comment requested that FDA 
    delete the reference to ``facilities'' inspection and limit the 
    requirement to review of the control system, as contained in the 
    original CGMP regulation.
        FDA has amended the requirements now in Sec. 820.70(c) to apply 
    only where environmental conditions could ``reasonably be expected to 
    have an adverse effect on product quality.'' The requirements for 
    procedures to ensure control of conditions, periodic inspection of 
    control systems, and documentation and review of results are similar to 
    the original CGMP requirements. However, the specific list of 
    conditions to be considered for control, which was carried over from 
    the original CGMP regulation to the proposal, was deleted in response 
    to a comment from the GHTF that the list would be better suited for a 
    guidance document. FDA agrees that it is not necessary to give examples 
    of conditions that may need controlling in a regulation, and notes that 
    lighting, ventilation, temperature, humidity, air pressure, filtration, 
    airborne contamination, and static electricity are among many 
    conditions that should be considered for control.
        FDA reworded the requirement to make it clear that the inspection 
    must be of the control system. FDA also added that the inspection of 
    the control system(s) shall include ``any necessary equipment,'' e.g., 
    pumps, filters, measurement equipment, etc. The sufficiency of 
    facilities is covered in a new Sec. 820.70(f), ``Buildings,'' that 
    requires that buildings be of suitable design and contain sufficient 
    space to allow for the proper manufacture of devices. Section 820.70(f) 
    is worded similarly to the original CGMP regulation Sec. 820.40, and is 
    intended to achieve the same objectives as that section.
        127. One comment stated that the last sentence of proposed 
    Sec. 820.70(b), ``Environmental control,'' should be deleted because it 
    is redundant with the audits required in Sec. 820.22(a). Another 
    comment said that environmental conditions are currently reviewed via 
    internal audit, which an FDA investigator cannot review.
        FDA disagrees with the comments. The inspection and review of 
    environmental control systems are routine quality assurance functions 
    that are part of the production quality assurance program. The audits 
    required by Sec. 820.22(a) are audits of the quality system, conducted 
    to ensure the adequacy of and conformance with the quality system 
    requirements. The requirement to conduct a quality audit is in addition 
    to other provisions in the regulation which require that a manufacturer 
    review its specific controls to ensure the requirements are met. FDA 
    may review the activities and results of environmental control system 
    inspections.
        128. The GHTF commented that the requirements of proposed 
    Sec. 820.70(c), ``Cleaning and sanitation,'' should be placed in 
    guidance.
        After careful consideration, FDA agrees that a separate section on 
    cleaning and sanitation is unnecessary. The objective of proposed 
    Sec. 820.70(c) is adequately met through the requirement of 
    Sec. 820.70(e), ``Contamination control,'' and Sec. 820.70(a), the 
    general process control procedure requirement. Contamination control 
    must include establishing and maintaining adequate cleaning procedures 
    and schedules, if such control is necessary to meet manufacturing 
    process specifications. In addition, Sec. 820.25 Personnel requires 
    that employees have a thorough understanding of their job functions, 
    which would include a requirement that the appropriate employees 
    comprehend
    
    [[Page 52629]]
    
    the cleanliness and sanitation procedures.
        129. The GHTF and others commented that the requirements of 
    proposed Sec. 820.70 (d)(1) through (d)(3) should be deleted and placed 
    in guidance because they are redundant with the first sentence in 
    proposed Sec. 820.70(d), ``Personnel health and cleanliness.''
        FDA agrees with the comments and has deleted Sec. 820.70 (d)(1) 
    through (d)(3). FDA has also rewritten the section, now entitled 
    ``Personnel,'' to require procedures to achieve the desired result, 
    rather than dictate the means to achieve the result. The section as 
    rewritten provides the manufacturer with more flexibility and is 
    consistent with ISO/CD 13485. Under this section, a manufacturer's 
    requirements must not permit unclean or inappropriately clothed 
    employees, or employees with medical conditions, to work with devices 
    where such conditions could reasonably be expected to have an adverse 
    effect on product quality. The procedures must also address acceptable 
    clothing, hygiene, and personal practices, if contact between personnel 
    and product or environment could reasonably be expected to have an 
    adverse effect on product quality.
        FDA also added the requirement, from ISO/CD 13485, that personnel 
    who are working temporarily (such as maintenance and cleaning 
    personnel) under special environmental conditions (such as a clean 
    room) be appropriately trained or supervised by someone trained to work 
    in such an environment.
        130. One comment stated that the requirements of Sec. 820.70(e), 
    ``Contamination control,'' should be deleted and placed in guidance. 
    Another comment stated that the reference to manufacturing materials 
    should be deleted because it is redundant with Sec. 820.70(g), 
    ``Equipment.''
        FDA has rewritten the section to delete the specific references to 
    contaminants that probably gave rise to the suggestion that the section 
    would be more appropriate as guidance. The section now contains a broad 
    requirement for the establishment of procedures to prevent 
    contamination of equipment or product by any substance that could 
    reasonably be expected to have an adverse effect on product quality. 
    Again, this revision adds flexibility.
        FDA disagrees with the comment that manufacturing materials should 
    be deleted from this section. Section 820.70(e) requires procedures to 
    ensure that manufacturing materials do not become contaminated. Section 
    820.70(g), in contrast, establishes requirements related solely to the 
    equipment used in the manufacturing process, and Sec. 820.70(h), 
    ``Manufacturing material,'' addresses requirements for the removal or 
    limitation of manufacturing materials. Thus, Sec. 820.70 (g) and (h) 
    are distinct and are intended to achieve different objectives.
        131. The only two comments received on proposed Sec. 820.70(f), 
    ``Sewage and refuse disposal,'' recommended that it be deleted because 
    it was unnecessary and/or covered by other Federal regulations.
        Section 820.70(f) has been deleted because the requirements are 
    adequately covered in the current requirements under Sec. 820.70(e), 
    ``Contamination control,'' and Sec. 820.70(c), ``Environmental 
    control.'' Under these sections, sewage, trash, byproducts, chemical 
    effluvium, and other refuse that could affect a device's safety, 
    effectiveness, or fitness-for-use must be adequately controlled.
        132. Two comments stated that the requirement related to equipment 
    in Sec. 820.70(g) should ensure that equipment meets ``specified 
    requirements,'' not be ``adequate for its intended use,'' because 
    intended use is determined during the design phase, and because it is 
    easier to assess whether equipment meets specified requirements.
        From these comments, FDA can see that the requirement should be 
    revised because it may have been misinterpreted. The requirement is 
    reworded as suggested. Under the requirement, the equipment must be 
    appropriately designed to facilitate maintenance, adjustment, cleaning, 
    and use. It must also meet the requirements that are necessary to 
    ensure its proper functioning for the manufacture of the device.
        133. A few comments stated that not all equipment requires 
    maintenance, and the requirement for a maintenance schedule in 
    Sec. 820.70(g)(1) should be revised to make that clear. The GHTF 
    recommended that the second sentence of proposed Sec. 820.70(g)(1), 
    which required that the maintenance schedule be posted or readily 
    available, be deleted and placed in guidance.
        FDA agrees that not all equipment may require maintenance and notes 
    that the general requirement of Sec. 820.70(a) requires process control 
    procedures that describe only those controls which are necessary. 
    Therefore, FDA did not revise the requirement.
        FDA has deleted the requirement that the maintenance schedule be 
    posted or readily available. Section 820.70(g), which directs a 
    manufacturer to ensure that equipment meets specified requirements, 
    requires that the manufacturer ensure that maintenance is carried out 
    on schedule to comply with the requirement. To satisfactorily meet this 
    requirement, FDA expects that the schedule will be posted on or near 
    the equipment to be maintained, or otherwise made readily available to 
    appropriate personnel. Deletion of the requirement, however, permits 
    the manufacturer added flexibility in complying with this section.
        134. Several comments stated that Sec. 820.70(g)(2), 
    ``Inspection,'' and (g)(3), ``Adjustment,'' should be deleted and 
    placed in guidance because the requirements are adequately covered in 
    Sec. 820.70(g)(1). Another comment stated that the requirement for 
    limitations or tolerances to be ``visibly posted on or near equipment'' 
    should be deleted.
        FDA believes that to adequately ensure that equipment continues to 
    meet its specifications, and to ensure that inherent limitations and 
    allowable tolerances are known, these requirements are imperative. FDA 
    notes inherent limitations and allowable tolerances must be visibly 
    posted on or near equipment or made readily available to personnel to 
    allow the manufacturer the flexibility to utilize any system to make 
    sure that the limitations or tolerances are readily available to the 
    personnel that need them. Both Sec. 820.70(g)(2) and (g)(3) are 
    requirements in the original CGMP regulation and the agency has found 
    them to be useful and necessary.
        135. One comment stated that requiring the removal of manufacturing 
    material to be documented in proposed Sec. 820.70(g)(4), 
    ``Manufacturing material,'' would result in impossible requirements, 
    such as the requirement to document how much cutting oil is lost during 
    a metal removing operation, such as drilling. Others commented that the 
    requirement needs to be amended to clarify that only manufacturing 
    materials that have an adverse effect or that are unwanted need to be 
    removed or limited.
        FDA disagrees with the first comment because Sec. 820.70(g)(4) (now 
    Sec. 820.70(h)) only requires that the fact that manufacturing material 
    was removed or reduced be documented, not how much was removed or how 
    much was lost due to processing. This requirement is carried over from 
    the original CGMP regulation, Sec. 820.60(d). FDA has amended the 
    section, however, to clarify that this requirement is necessary ``Where 
    a manufacturing material could reasonably be expected to have an 
    adverse effect on product quality.'' FDA
    
    [[Page 52630]]
    
    purposefully qualifies the general requirement by that which adversely 
    affects ``product quality'' (product as defined in Sec. 820.3(r)) and 
    limits the requirement for removal or reduction to ``an amount that 
    does not adversely affect the device's quality.''
        136. One comment on Sec. 820.70(h), ``Automated processes,'' (now 
    Sec. 820.70(i)), stated that the section should be revised to reflect 
    that software used in such systems must be validated for ``its intended 
    use,'' not simply validated. Another comment stated that most companies 
    buy software currently available on the market and do not make changes 
    to the software. It was recommended that Sec. 820.70(h) allow for use 
    of outside personnel for validation runs and not necessarily require 
    the development of a software validation procedure. One comment 
    suggested that the section should allow verification rather than 
    validation of off-the-shelf software. Several comments on ``automated 
    processes'' stated that the term ``data processing systems'' was 
    unclear and its inclusion rendered the requirement too broad. Others 
    asked for clarification of ``automated data processing systems.''
        FDA has modified the requirement to mandate validation for the 
    intended use of the software. In addition, the requirement that the 
    software be validated by individuals designated by the manufacturer has 
    also been deleted to make clear that validation may be performed by 
    those other than the manufacturer. However, whether the manufacturer 
    designates its own personnel or relies on outside assistance to 
    validate software, there must be an established procedure to ensure 
    validation is carried out properly.
        FDA has maintained the requirement for validation because the 
    agency believes that it is necessary that software be validated to the 
    extent possible to adequately ensure performance. Where source code and 
    design specifications cannot be obtained, ``black box testing'' must be 
    performed to confirm that the software meets the user's needs and its 
    intended uses.
        FDA emphasizes that manufacturers are responsible for the adequacy 
    of the software used in their devices, and activities used to produce 
    devices. When manufacturers purchase ``off-the- shelf'' software, they 
    must ensure that it will perform as intended in its chosen application.
        FDA has amended the requirement to state ``When computers or 
    automated data processing systems are used as part of production or the 
    quality system,'' for clarification. Software used in production or the 
    quality system, whether it be in the designing, manufacturing, 
    distributing, or tracing, must be validated.
    ii. Inspection, Measuring, and Test Equipment (Sec. 820.72)
        137. A few comments stated that it is unclear what is meant by the 
    requirement in proposed Sec. 820.84 Inspection, measuring, and test 
    equipment that equipment be capable of producing ``valid results.'' The 
    comments stated that such equipment may be ``suitable for its intended 
    purpose'' and still not always ``produce valid results.''
        FDA believes that the term ``valid results'' is commonly understood 
    and notes that it has been in the original CGMP regulation under 
    Sec. 820.61 for 18 years. The requirement is for the equipment to work 
    properly, thereby providing ``valid results.''
        FDA renumbered Sec. 820.84 as Sec. 820.72 in response to comments 
    that stated these requirements were more appropriate under subpart G 
    Production and Process Controls. FDA revised the requirement in new 
    Sec. 820.72(a), ``Control of inspection, measuring, and test 
    equipment,'' to make clear that the procedures must also ensure that 
    the equipment is maintained and moved the requirement that the 
    procedure include provisions for handling, preservation and storage of 
    equipment from Sec. 820.84(d) in the Working Draft to Sec. 820.72(a). 
    FDA deleted the term ``test software'' that was in Sec. 820.84(e) 
    because FDA believes that ``test software'' is now covered under 
    ``electronic inspection and test equipment'' in Sec. 820.72(a).
        138. A few comments stated that the last sentence in proposed 
    Sec. 820.84(a), ``Calibration,'' is unnecessary because the requirement 
    for trained personnel is redundant with Sec. 820.25(a) Personnel. A few 
    comments stated that FDA should identify what must be remedied in 
    proposed Sec. 820.84(a).
        FDA agrees that the requirement for trained personnel is redundant 
    and has deleted this sentence from Sec. 820.72(b), ``Calibration.'' FDA 
    has also added to this section the requirement that the calibration 
    procedure include provisions for remedial action to ``reestablish the 
    limits and to evaluate whether there was any adverse effect on the 
    device's quality'' to clarify this remedial action requirement and its 
    relationship to the requirements in Sec. 820.100 Corrective and 
    preventive action. 
        139. Several comments stated that Sec. 820.84(b), ``Calibration 
    standards,'' should allow for the use of international standards.
        FDA agrees and has rewritten the section, now Sec. 820.72(b)(1), 
    ``Calibration standards,'' to allow the use of international standards. 
    The standards used must be generally accepted by qualified experts as 
    the prevailing standards.
        140. FDA has deleted the requirement in proposed Sec. 820.84(c), 
    now Sec. 820.72(b)(2), ``Calibration records,'' that calibration 
    records be ``maintained by individuals designated by the manufacturer'' 
    because, on further reflection, the agency believes such a requirement 
    is unnecessary. As long as the required procedures and records are 
    maintained and displayed or readily available as required, the 
    objective of the section, ensuring that calibration is performed and 
    acceptable, will be met. FDA did add ``equipment identification'' to 
    the list of items that had to be documented in response to a comment 
    that requested clarification in this regard, so that equipment is 
    clearly identified in the calibration records even if the records are 
    not displayed on or near the particular piece of equipment.
        141. Two comments suggested deleting proposed Sec. 820.84(d) 
    because they believed it was unnecessary to establish procedures to 
    maintain equipment, because most manufacturers simply store equipment 
    in protective covers.
        As already noted, FDA has moved the requirement for establishing 
    maintenance procedures into the general requirement in Sec. 820.72. FDA 
    has retained the requirement because some equipment requires special 
    handling, preservation, and storage. For example, the temperature and 
    humidity of a room may affect the equipment and procedures would need 
    to be established taking those factors into account.
        142. Several comments stated that proposed Sec. 820.84(e), 
    ``Facilities,'' should be deleted because it is redundant with the 
    requirements under Sec. 820.70(g) and the general requirements of 
    proposed Sec. 820.84(a).
        FDA agrees that revised Sec. 820.84(a), which is now 
    Sec. 820.72(a), would require procedures to ensure that equipment is 
    protected from adjustments that could invalidate the calibration, in 
    that the section requires procedures to ensure that equipment is 
    properly maintained. The procedures that require equipment to be 
    routinely calibrated, inspected, and checked, will also ensure that 
    improperly calibrated equipment is not used. Therefore, FDA has deleted 
    proposed Sec. 820.84(e).
    
    [[Page 52631]]
    
    iii. Process Validation (Sec. 820.75)
        143. A few comments on proposed Sec. 820.75 Special processes 
    stated that the meaning of the term ``special processes'' was unclear. 
    Other comments stated that FDA should provide examples of processes 
    that would be considered ``special processes.'' Several comments stated 
    the term ``fully verified'' was unclear and should be deleted.
        In response to the comments, the term ``special processes'' has 
    been dropped from the regulation and the term ``process validation'' is 
    defined in Sec. 820.3(z)(1). The section now requires that when a 
    process ``cannot be fully verified by subsequent inspection and test, 
    the process shall be validated with a high degree of assurance. * * *'' 
    Examples of such processes include sterilization, aseptic processing, 
    injection molding, and welding, among others. The validation method 
    must ensure that predetermined specifications are consistently met. The 
    new Sec. 820.75, entitled ``Process validation,'' is consistent with 
    ISO 9001:1994, section 4.9, including the terminology ``fully 
    verified.'' FDA does not believe this terminology is unclear since it 
    has been used in ISO 9001:1987 and 1994 and explained in several 
    guidance documents.
        FDA amended this section by removing the requirement for the 
    signature of the individual(s) performing the process and placing the 
    signature requirement on the approval of the validation where FDA 
    believes it is more important and appropriate. FDA also added that 
    ``where appropriate, the major equipment validated'' must be 
    documented. Depending on the process that is validated, it may be 
    necessary to document the person performing the process or the 
    equipment or both in order to have adequate controls on the process.
        144. Several comments were received on proposed Sec. 820.75(a)(1) 
    through (a)(4) that stated that the requirements were redundant with 
    other parts of the regulation and should be modified or deleted.
        FDA disagrees with the comments and believes that, due to the 
    importance of process validation and correct performance of the 
    validated process, the requirements are necessary. The requirements 
    have been rearranged in the revised section.
        145. Comments on the first sentence of proposed Sec. 820.75(b) 
    stated that it was unclear and unrealistic. Other comments stated that 
    the requirement for continuous monitoring is not practical or 
    necessary.
        In response to the comments, FDA has revised the requirements. 
    Section 820.75(b) applies to the performance of a process after the 
    process has been validated. In contrast, Sec. 820.75(a) relates to the 
    initial validation of the process. FDA deleted the term ``continuous'' 
    because the agency concurs that monitoring can be accomplished at a 
    determined interval and frequency depending on the type of validated 
    process being monitored and controlled. FDA notes that the interval and 
    frequency should be periodically evaluated for adequacy, especially 
    during any evaluation or revalidation that occurs in accordance with 
    the requirements in new Sec. 820.75(c).
        New Sec. 820.75(b)(1), which was proposed Sec. 820.75(c) of the 
    Working Draft, requires that validated processes be performed by a 
    qualified individual(s). FDA notes that Sec. 820.75(b)(1) is similar to 
    the requirements under Sec. 820.25 Personnel but emphasizes that 
    validated processes must not only be performed by personnel with the 
    necessary education, background, training, and experience for their 
    general jobs but must be performed by personnel qualified for those 
    particular functions. Revised Sec. 820.75(b)(2), which was proposed 
    Sec. 820.75(d) of the Working Draft, contains the amended documentation 
    requirements for validated processes, to include the monitoring and 
    control methods and data. FDA notes that it is always ``appropriate'' 
    to document the equipment used in the process where the manufacturer 
    uses different equipment on different manufacturing lines. To 
    investigate a problem with the device, the manufacturer will need to 
    know which equipment was used, since the problem could be with the 
    equipment itself. The same holds true for the individual(s) performing 
    the process.
        Section 820.75(c) contains requirements on process revalidation in 
    response to several comments and concerns on when revalidation 
    activities were necessary. FDA believes that the new arrangement of 
    Sec. 820.75 should clarify the requirement.
    
    H. Acceptance Activities (Subpart H)
    
    i. Receiving, In-Process, and Finished Device Acceptance (Sec. 820.80)
        146. One comment stated that the emphasis on testing and inspection 
    in proposed Sec. 820.80 completely ignores the quality goals, the 
    benefit of requiring purchasing controls, and statements made in the 
    preamble of the proposal reflecting FDA's negative opinion about 
    manufacturers relying solely on testing and inspection. A few comments 
    on the Working Draft stated that ``acceptance activities'' should be 
    defined as inspections, tests, or other verification activities so that 
    the regulation does not require all of these activities but gives the 
    manufacturer the flexibility to choose the appropriate method.
        FDA agrees with the comments and has replaced the term ``inspection 
    and test'' with ``acceptance activities'' in Sec. 820.80. Further, FDA 
    now defines ``acceptance activities'' to include inspections, test, or 
    other verification activities, such as supplier audits.
        147. One comment stated that recordkeeping is a significant cost 
    factor in the operation of a total quality system, and that the revised 
    CGMP regulation should not add cost through duplication of 
    documentation. The comment said recording all quantitative data is 
    inappropriate and of little value.
        FDA agrees that unnecessary duplication of documentation should be 
    avoided. FDA believes that the quality system regulation requires the 
    minimum documentation necessary to ensure that safe and effective 
    devices are designed and produced. FDA similarly believes that 
    maintaining records of results of acceptance activities is imperative 
    to ensure that nonconforming product is not inadvertently used or 
    distributed. FDA has, however, deleted from Sec. 820.80(a) the 
    requirement for recording the results of inspections and testing 
    because Sec. 820.80(e) requires that the results of acceptance 
    activities be recorded. The requirement in Sec. 820.80(a) was therefore 
    unnecessary. Further, the regulation does not specify quantitative data 
    but simply requires that the results be recorded. FDA believes that it 
    is essential for the manufacturer to maintain records which provide 
    evidence that the product has gone through the defined acceptance 
    activities. These records must clearly show whether the product has 
    passed or failed the acceptance activities according to the defined 
    acceptance criteria. Where product fails to pass acceptance activities, 
    the procedures for control of nonconforming product must be 
    implemented, to include investigations where defined. If the acceptance 
    records are not clear about how the product failed, then the 
    manufacturer may end up duplicating the acceptance activities in order 
    to perform appropriate investigations.
        148. Several comments stated that proposed Sec. 820.80(b), 
    ``Receiving inspection and testing,'' did not allow for urgent use of 
    incoming items. The comments said that urgent use should be permitted 
    if forward traceability is maintained so that recall and
    
    [[Page 52632]]
    
    replacement is possible if the material is subsequently found to be 
    nonconforming. One comment stated that the requirements in proposed 
    Sec. 820.80(b) were too specific and did not allow flexibility.
        FDA agrees in part with the comments. FDA has permitted 
    manufacturers to use incoming items that had not yet been proven 
    acceptable for use, provided that the manufacturer maintained control 
    of the unapproved items and could retrieve the product that contained 
    the unapproved items before distribution. Therefore, the requirement 
    that product ``shall not be used or processed until * * * verified'' 
    has been deleted from Sec. 820.80(b), now entitled ``Receiving 
    acceptance activities.'' However, FDA emphasizes that while the product 
    can be used in production prior to verification, it cannot be 
    distributed prior to verification. FDA does not permit the distribution 
    of unapproved product through an urgent use provision, because all 
    finished devices must comply with Sec. 820.80(d), ``Final acceptance 
    activities,'' before they are released for distribution.
        In addition to the changes noted above, FDA has deleted the 
    requirement that ``individual(s) designated by the manufacturer shall 
    accept or reject incoming'' product. FDA does not believe this 
    requirement is necessary in Sec. 820.80(b) because Sec. 820.80(e) 
    requires that the identification of the individual(s) conducting the 
    acceptance activities be recorded.
        149. Several comments stated that an absolute requirement under 
    proposed Sec. 820.80(c), ``In-process inspection and testing,'' for in-
    process testing was inconsistent with the preamble, which stated that 
    an appropriate mix of controls should be established. Other comments 
    stated that in-process inspection and testing is unnecessary if the 
    process is validated and the devices are subject to final inspection. A 
    few comments on the Working Draft stated that the term ``held'' was too 
    restrictive and was not consistent with the requirements and the 
    preamble discussion for Sec. 820.80(b).
        FDA agrees with the comments in part, but believes that Sec. 820.80 
    as now written, with the inclusion of ``where appropriate,'' does not 
    mandate in-process inspection and testing. FDA acknowledges that in-
    process acceptance activities may not be necessary or possible for 
    every device, for example, medical socks. Further, the requirement 
    states that in-process product must be controlled until the required 
    inspection and test, or other verification activities, have been 
    performed. This will permit manufacturers to use, under defined 
    conditions and procedures, product that has not completed the 
    acceptance activities described in Sec. 820.80(b) and (c). This does 
    not means that manufacturers can ignore the requirements in 
    Sec. 820.80(b) and (c) because these requirements must be completed in 
    order to comply with Sec. 820.80(d), which must be satisfied before 
    devices are released for distribution.
        150. FDA received a similar comment on proposed Sec. 820.80(d), 
    ``Final inspection and test,'' which said that the provision requires 
    finished device inspection for all devices, without defining what 
    inspection is expected. The comment suggested that Sec. 820.80(d) could 
    be interpreted to require actual product inspection, which has been 
    shown to be ineffective as a means of controlling product quality. One 
    comment stated that signatures should not be the only approved method 
    for identification of the individual(s) responsible for release. The 
    comment stated that use of inspection stamps and initials should be 
    allowed.
        FDA has rewritten Sec. 820.80(d) to require that manufacturers 
    establish and maintain procedures for finished device acceptance to 
    ensure that each production run, lot, or batch of finished devices 
    meets specified requirements. Manufacturers have the flexibility to 
    choose a combination of methods, including finished device inspection 
    and test, provided such methods will accomplish the required result.
        FDA believes that it is important for the person responsible for 
    release to have personally documented and dated that release. This can 
    be accomplished through use of an inspection stamp, if the stamp is 
    controlled as discussed above under Sec. 820.40 Document controls. 
    Therefore, FDA has retained the requirement for a signature.
        151. Several comments on proposed Sec. 820.80(e), ``Inspection and 
    test records,'' stated that manufacturers should not be required to 
    record the use of general equipment in inspection and test records, 
    because this requirement would be burdensome to large manufacturers who 
    use many common pieces of equipment. A few comments stated that the 
    record requirements under Sec. 820.80(e) are overly prescriptive and go 
    well beyond ISO 9001's comparable requirements. The comments stated 
    that recordkeeping should be specified by the manufacturer in the 
    spirit of ISO 9001, and should include only the minimum records 
    necessary to show that finished device inspections are performed in 
    accordance with established procedures.
        FDA agrees that it may not be necessary to document every piece of 
    equipment used in acceptance activities. The requirement, renamed 
    ``Acceptance records,'' now provides that equipment used shall be 
    documented ``where appropriate.'' For some critical operations and 
    testing, identification of the equipment used will be imperative for 
    proper investigations into nonconforming product.
        The requirements, as revised, are similar to those in ISO 
    9001:1994. As discussed above, certain information must be captured on 
    acceptance records for the records to be useful in evaluating 
    nonconformance. Through many years of experience, FDA has determined 
    what it believes to be a minimum requirement for these records. Section 
    820.80(e) reflects that determination.
    ii. Acceptance Status (Sec. 820.86)
        152. Several comments on proposed Sec. 820.86, ``Inspection and 
    test status,'' stated that the section was not flexible enough to allow 
    identification of the inspection and test status of product by various 
    means, because the requirement was for the status to be ``visible.'' 
    One comment questioned why ``component acceptance'' was addressed 
    separately.
        FDA agrees that the inspection and test status may be identified by 
    any method that will achieve the result, which might include acceptable 
    computerized identification, markings, etc. The section has been 
    rewritten to reflect this intent, has been renamed ``Acceptance 
    status,'' and is now consistent with ISO 9001:1994. FDA also agrees 
    that ``component acceptance'' is covered by ``manufacturing'' and has 
    deleted the term.
        153. FDA has deleted proposed Sec. 820.86(b) which required that 
    records identify those responsible for release of the product, because 
    the agency believes that the records required by Sec. 820.80(e) will 
    identify those responsible for release of product.
    
    I. Nonconforming Product (Subpart I)
    
        154. FDA has rewritten Sec. 820.90 Nonconforming product to utilize 
    the term ``product'' throughout, as defined in Sec. 820.3(r), for both 
    shorthand purposes and consistency with ISO 9001:1994.
        155. One comment suggested deleting the term ``inadvertently'' and 
    adding the word ``distributed'' before ``installed'' in Sec. 820.90(a). 
    Several written comments and persons who testified at the August and 
    September 1995 meetings stated that Sec. 820.90(a) should be written so
    
    [[Page 52633]]
    
    that it is not interpreted to require investigations for every 
    nonconformance. A few comments stated that the term ``provide for'' was 
    too broad and unclear. Other comments stated that the requirement to 
    ``ensure'' nonconforming product was ``not used or distributed'' was 
    inconsistent with the provisions in Sec. 820.90(b) which allowed for 
    concessions under certain circumstances. One comment stated that the 
    requirement that persons responsible for nonconforming product be 
    ``notified'' should be deleted because it is overly burdensome and not 
    needed in all cases.
        FDA has reworded the general requirement for procedures to control 
    nonconforming product and has deleted the term ``inadvertently.'' FDA 
    has also added the requirement that the procedures provide for the 
    ``evaluation'' of nonconforming product because evaluation is key to 
    protecting against recurring nonconformance. The addition is consistent 
    with ISO 9001:1994.
        FDA has further revised Sec. 820.90 in response to the comments on 
    the Working Draft. First, the manufacturer must establish procedures to 
    ``control'' nonconforming product. Second, the procedures shall 
    ``address the identification, documentation, evaluation, segregation, 
    and disposition of nonconforming product,'' which gives the 
    manufacturers the flexibility to define how they are going to 
    ``control'' products that are nonconforming. Third, the evaluation 
    process addressed in the procedure ``shall include a determination of 
    the need for an investigation.'' Therefore, the procedures will need to 
    set forth the manufacturer's SOP on when investigations will take place 
    and provisions for trending and/or monitoring the situation in the 
    future. Fourth, FDA added ``The evaluation and any investigation shall 
    be documented,'' which would include the explanations for not 
    performing investigations and how nonconformances will be trended and/
    or monitored. Further, the phrase ``is not used or distributed'' has 
    been deleted to be consistent with Sec. 820.90(b).
        FDA disagrees that the notification requirement should be deleted. 
    Where some person or organization is responsible for nonconformances, 
    they must be notified to ensure that future nonconformances are 
    prevented. This requirement is also in ISO 9001:1994, section 4.13.1.
        156. FDA has rewritten Sec. 820.90(b)(1), ``Nonconformity review 
    and disposition,'' to make clear that the section requires procedures 
    that define the responsibility for review and authority for disposition 
    of nonconforming product and that set forth the review and disposition 
    process. FDA believes that proper disposition of nonconforming product 
    is essential for ensuring the safety and effectiveness of devices. 
    Manufacturers have made determinations that nonconforming product may 
    be used which have resulted in defective devices being distributed. 
    Thus, although it may be appropriate at times to use nonconforming 
    products, the disposition process must be adequately controlled.
        The revision requires that disposition and justification for 
    concessions be documented. FDA believes that the justification should 
    be based on scientific evidence, which a manufacturer should be 
    prepared to provide upon request. Concessions should be closely 
    monitored and not become accepted practice. This section is consistent 
    with ISO 9001:1994, section 4.13.2.
        Several comments on the Working Draft stated that the term 
    ``concession'' should be deleted because it is confusing. FDA has 
    rewritten the sentence to ensure the meaning of this requirement is 
    clear. The sentence now reads, ``Documentation shall include the 
    justification for the use of nonconforming product and the signature of 
    the individual(s) authorizing the use.''
        157. Several comments were received on proposed Sec. 820.90(b)(2). 
    One comment stated that the requirement should allow for other types of 
    disposition besides reprocessing. One comment suggested replacing the 
    term ``reinspection'' with ``evaluation,'' to allow for greater 
    flexibility in verification methods. Many comments suggested that the 
    requirement for identification of reprocessed product should be deleted 
    because they believed it would cause the consumer to forego purchasing 
    the product. Several comments requested that the term ``rework'' be 
    used instead of ``reprocessing'' to harmonize terminology with ISO 
    standards.
        FDA agrees in part with the comments. FDA, as noted in the 
    definition section, has substituted the term ``rework'' and the ISO 
    8402:1994 definition for the term ``reprocessing'' in response to the 
    comments. FDA believes that the revised Sec. 820.90(b)(1) clearly 
    allows for other methods of disposition besides rework. Section 
    820.90(b)(2), which governs rework when it is chosen as a method of 
    disposition, has been revised as requested by replacing the term 
    ``reinspection'' with ``reevaluation.'' The change will allow 
    manufacturers the flexibility to inspect or use other verification 
    activities.
        FDA has also deleted the requirement for identification of reworked 
    product from this section because FDA believes that it is adequately 
    covered in Secs. 820.60 Identification and 820.86 Acceptance status.
        Other minor changes made to the section include requiring that a 
    determination of any adverse effect of the rework upon the product be 
    made, whether there is ``repeated'' rework or not. FDA's intent is that 
    such a determination be made with any rework, given the potential 
    harmful effect rework could have on the product. The change harmonizes 
    Sec. 820.90 with ISO/CD 13485. In addition, the sentence requiring a 
    ``complete reinspection'' for reworked product was deleted because the 
    section already requires retesting and reevaluation of reworked 
    product. FDA has also substituted ``current'' for ``original or 
    subsequently modified'' approved specifications for clarity. The 
    requirements as written are consistent with the original CGMP 
    requirements in Secs. 820.115 and 820.116.
    
    J. Corrective and Preventive Action (Subpart J)
    
        158. A few comments suggested revising proposed Sec. 820.100 
    Corrective and preventive action to require procedures for implementing 
    corrective and preventive action, consistent with ISO 9001. One comment 
    stated that the procedures should provide for an initial halt of 
    distribution of suspect products or tight control and action concerning 
    products already distributed before taking the long term action listed 
    in this section.
        FDA agrees that it is essential that the manufacturer establish 
    procedures for implementing corrective and preventive action and has 
    revised Sec. 820.100(a) accordingly. The procedures must include 
    provisions for the remaining requirements in the section. These 
    procedures must provide for control and action to be taken on devices 
    distributed, and those not yet distributed, that are suspected of 
    having potential nonconformities.
        159. Other comments stated that the degree of remedial action 
    should be commensurate with the risk associated with a product failure.
        FDA agrees that the degree of corrective and preventive action 
    taken to eliminate or minimize actual or potential nonconformities must 
    be appropriate to the magnitude of the problem and commensurate with 
    the risks encountered. FDA cannot dictate in a regulation the degree of 
    action that
    
    [[Page 52634]]
    
    should be taken because each circumstance will be different, but FDA 
    does expect the manufacturer to develop procedures for assessing the 
    risk, the actions that need to be taken for different levels of risk, 
    and how to correct or prevent the problem from recurring, depending on 
    that risk assessment.
        FDA emphasizes that any death, even if the manufacturer attributes 
    it to user error, will be considered relevant by FDA and will have a 
    high risk potentially associated with it. User error is still 
    considered to be a nonconformity because human factors and other 
    similar tools should have been considered during the design phase of 
    the device. FDA acknowledges that a manufacturer cannot possibly 
    foresee every single potential misuse during the design of a device, 
    but when the manufacturer becomes aware of misuse, the corrective and 
    preventive action requirements should be implemented to determine if 
    redesign of the device or labeling changes may be necessary.
        160. Several comments on proposed Sec. 820.100(a)(1) stated that 
    requiring a manufacturer to analyze ``all'' processes, work operations, 
    and other factors listed, is excessive and unrealistic. Some comments 
    stated that there should not be a requirement to conduct an analysis 
    for ``potential causes'' of nonconformances. A few comments stated that 
    including ``quality audits'' in the list was inconsistent with the FDA 
    policy of not reviewing internal audits. A few comments stated that the 
    requirement that the analysis include ``trend analysis'' should be 
    modified because it places unnecessary emphasis on only one statistical 
    method or tool. Other comments stated that statistical tools are not 
    always necessary and that the requirement should be modified.
        FDA agrees in part with the comments. It was not FDA's intent to 
    require that processes unrelated to an existing nonconformity be 
    analyzed. Instead, Sec. 820.100(a)(1) requires an analysis of those 
    items listed that could be related to the problem. To prevent 
    confusion, the word ``all'' has been deleted. The requirement is 
    similar to that of ISO 9001:1994, section 4.14.3(a).
        The inclusion of ``quality audits'' as a valuable feedback 
    mechanism for the manufacturer does not conflict with FDA's policy of 
    not reviewing internal quality audits. Internal audits are valuable and 
    necessary tools for the manufacturer to evaluate the quality system. 
    The audit reports should be used to analyze the entire quality system 
    and provide feedback into the system to close the feedback loop, so 
    that corrective or preventive actions can be taken where necessary. FDA 
    will review the corrective and preventive action procedures and 
    activities performed in conformance with those procedures without 
    reviewing the internal audit reports. FDA wants to make it clear that 
    corrective and preventive actions, to include the documentation of 
    these activities, which result from internal audits and management 
    reviews are not covered under Sec. 820.180(c).
        FDA has further revised the requirement to delete the reference to 
    trend analysis in response to the comments. The provision now requires 
    that ``appropriate statistical methodology'' be employed where 
    necessary to detect recurring quality problems. This revision is made 
    because there may be other statistical tools available beyond ``trend 
    analysis.'' FDA emphasizes that the appropriate statistical tools must 
    be employed when it is necessary to utilize statistical methodology. 
    FDA has seen far too often the misuse of statistics by manufacturers in 
    an effort to minimize instead of address the problem. Such misuse of 
    statistics would be a violation of this section.
        FDA has retained the requirement for analysis to identify 
    ``potential causes of nonconforming product,'' however, because FDA 
    believes this is an important aspect of preventive action. FDA notes 
    that ISO 9001:1994, section 4.14.1, specifically acknowledges that 
    corrective and preventive actions are associated with actual and 
    potential nonconformities.
        161. Several comments stated that proposed Sec. 820.100(a)(2) was 
    redundant with requirements in Sec. 820.198 Complaints.
        FDA agrees in part with the comments and has written the section to 
    require investigation of the cause of nonconformities relating to 
    process, product, and the quality system, consistent with ISO 
    9001:1994, section 4.14.2(b). The requirement in this section is 
    broader than the requirement for investigations under Sec. 820.198, 
    because it requires that nonconforming product discovered before or 
    after distribution be investigated to the degree commensurate with the 
    significance and risk of the nonconformity. At times a very indepth 
    investigation will be necessary, while at other times a simple 
    investigation, followed by trend analysis or other appropriate tools 
    will be acceptable. In addition, in contrast to Sec. 820.198, the 
    requirement in this section applies to process and quality system 
    nonconformities, as well as product nonconformities. For example, if a 
    molding process with its known capabilities has a normal 5 percent 
    rejection rate and that rate rises to 10 percent, an investigation into 
    the nonconformance of the process must be performed.
        162. One comment stated that proposed Sec. 820.100(a)(3) should not 
    require identification of action necessary to correct ``other quality 
    problems.'' Another stated that the section should be harmonized with 
    ISO. One comment thought that the requirement should be to identify 
    action to correct problems identified by ``trend analysis.''
        FDA agrees that harmonization is important and has harmonized the 
    terminology (and intent) of the section with ISO 9001:1994, sections 
    4.14.2(c) and 4.14.3(b). However, FDA disagrees that the section should 
    not require identification of action necessary to correct ``other 
    quality problems'' because the objective of Sec. 820.100 is to correct 
    and prevent poor practices, not simply bad product. Correction and 
    prevention of unacceptable quality system practices should result in 
    fewer nonconformities related to product. Therefore, this section 
    addresses problems within the quality system itself. For example, it 
    should identify and correct improper personnel training, the failure to 
    follow procedures, and inadequate procedures, among other things.
        FDA also disagrees with the suggestion to link the requirement in 
    Sec. 820.100(a)(3) to trend analysis and has deleted the reference to 
    trend analysis in Sec. 820.100(a)(1) to give the manufacturer the 
    flexibility to use whatever method of analysis is appropriate.
        163. FDA has revised Sec. 820.100(a)(4) to reflect that preventive, 
    as well as corrective, action must be verified or validated. The 
    section is now consistent with ISO 9001:1994, sections 4.14.2(d) and 
    4.14.3(c). Two comments stated that the definitions of validation and 
    verification cause confusion here, but FDA believes that these concerns 
    should be resolved with the amended definitions under Sec. 820.3 (z) 
    and (aa).
        164. FDA has also revised Sec. 820.100(a)(5) in the same manner, to 
    relate the requirements to preventive action. This section is 
    consistent with ISO 9001:1994, section 4.14.1, third paragraph.
        165. One comment suggested that proposed Sec. 820.100(a)(6) be 
    revised to reflect that minor quality problems may not need to be 
    disseminated to those directly responsible for ensuring quality and to 
    be reviewed by management.
    
    [[Page 52635]]
    
        FDA agrees in part with this comment. The revised Sec. 820.100 
    (a)(6) and (a)(7) require that procedures ensure that information is 
    disseminated to those directly responsible for assuring quality or the 
    prevention of such problems, and provide for submitting relevant 
    information on identified quality problems, as well as corrective and 
    preventive actions, for management review. This revision should address 
    the concern raised by the comment because only certain information need 
    be directed to management. The manufacturer's procedures should clearly 
    define the criteria to be followed to determine what information will 
    be considered ``relevant'' to the action taken and why. FDA emphasizes 
    that it is always management's responsibility to ensure that all 
    nonconformity issues are handled appropriately. This section is now 
    consistent with ISO 9001:1994, section 4.14.3(d).
        166. Two comments stated that the records required under 
    Sec. 820.100(b) should be treated as part of the internal audit.
        FDA disagrees with these comments because this information is 
    directly relevant to the safety and effectiveness of finished medical 
    devices. FDA has the authority to review such records and the 
    obligation to do so to protect the public health. Comparable 
    information and documentation is reviewed by the FDA under the 
    requirements of the original CGMP, Secs. 820.20 (a)(3) and (a)(4) and 
    820.162. Manufacturers will be required to make this information 
    readily available to an FDA investigator, so that the investigator may 
    properly assess the manufacturer's compliance with these quality system 
    requirements.
    
    K. Labeling and Packaging Control (Subpart K)
    
    i. Device Labeling (Sec. 820.120)
        167. Several comments on proposed Sec. 820.162 Device labeling 
    stated that the section should be deleted and placed in guidance 
    because it is unnecessary and redundant with requirements under 
    Secs. 820.80 and 820.86. A few comments stated that the section should 
    be changed to be the same as that in the original CGMP regulation, 
    under Secs. 820.120 and 820.121. Another comment stated that labeling 
    and packaging requirements should be in subpart K of part 820 and 
    handling, storage, distribution, and installation requirements should 
    be in subpart L of part 820 because labeling and packaging functionally 
    occur before distribution and installation.
        FDA believes that the section, as written, is consistent with the 
    requirements in the original CGMP. Section 820.120 relates specifically 
    to labeling and its requirements are in addition to those in both 
    Secs. 820.80 and 820.86. Further, FDA believes that the degree of 
    detail in this section is necessary because these same requirements 
    have been in place for 18 years, yet numerous recalls every year are 
    the result of labeling errors or mixups. FDA therefore believes that 
    more, not less, control is necessary.
        FDA has reordered the subparts but notes that the handling and 
    storage requirements apply throughout the production process.
        168. One comment stated that ``to maintain labeling integrity and 
    to prevent labeling mixups'' should be deleted from the general 
    requirement because the requirements are detailed in the following 
    sections. Other comments stated that all labels need not be affixed to 
    the device and others stated that ``legible and affixed'' may not be 
    appropriate for all implantable devices.
        FDA agrees with the comments and has revised the requirements 
    accordingly.
        169. A few comments stated that what is now Sec. 820.120(b), 
    ``Labeling inspections,'' should allow automated readers to be used in 
    place of a ``designated individual(s)'' to examine the labeling.
        FDA disagrees with the comments because several recalls on labeling 
    have been attributed to automated readers not catching errors. The 
    requirement does not preclude manufacturers from using automated 
    readers where that process is followed by human oversight. A 
    ``designated individual'' must examine, at a minimum, a representative 
    sampling of all labels that have been checked by the automated readers. 
    Further, automated readers are often programmed with only the base 
    label and do not check specifics, such as control numbers and 
    expiration dates, among other things, that are distinct for each label. 
    The regulation requires that labeling be inspected for these items 
    prior to release.
        170. FDA has amended Sec. 820.120(b) to add ``any'' to additional 
    processing instructions in response to a comment for clarity. FDA has 
    amended Sec. 820.120(d) to include ``The label and labeling used for 
    each production unit, lot, or batch shall be documented in the DHR'' in 
    response to comments questioning whether the labeling used should be 
    recorded in the DMR or the DHR. FDA also amended Sec. 820.120(e) by 
    adding ``or shall accompany the device through distribution'' and 
    deleting ``itself or its label'' for clarity.
        171. A few comments on proposed Sec. 820.165 Critical devices, 
    labeling stated that this section should be deleted to eliminate any 
    distinction between critical and noncritical devices.
        FDA agrees in part and has deleted Sec. 820.165, but has added the 
    requirement on control numbers to Sec. 820.120(e).
    ii. Device Packaging (Sec. 820.130)
        172. Two comments on proposed Sec. 820.160 Device packaging stated 
    that the section should be changed to allow manufacturers to use third 
    parties, if desired, for packaging. Another comment stated that it is 
    very difficult if not impossible to protect from intentional damage, 
    such as tampering.
        FDA agrees with the comments and has changed the requirement, now 
    in Sec. 820.130, accordingly. FDA believes, however, that any 
    intentional tampering would not be covered because the requirement 
    states ``during customary conditions.''
    
    L. Handling, Storage, Distribution, and Installation (Subpart L)
    
    i. Handling (Sec. 820.140)
        173. One comment on proposed Sec. 820.120 Handling suggested that 
    the procedures be ``designed to prevent,'' rather than be established 
    to ``ensure that,'' problems delineated in the section do not occur. 
    The comment stated that the word ``prevent'' would add clarity, without 
    compromising the meaning of the sentence. Another comment stated that 
    the handling procedures should apply ``prior to distribution,'' not 
    during ``any stage of handling.'' One comment stated that the 
    requirement does not cover the need for special precautions in handling 
    used devices which may be contaminated, and that this is an important 
    issue covered by ISO/CD 13485.
        FDA does not believe that Sec. 820.120, now Sec. 820.140, as 
    written is unclear. The procedures are expected to ensure that mixups, 
    damage, deterioration, contamination, or other adverse effects do not 
    occur. FDA amended the requirement, however, to remove ``any stage of'' 
    so it reads ``during handling.'' The requirement continues to apply to 
    all stages of handling in which a manufacturer is involved, which may 
    in some cases go beyond initial distribution.
        The comparable provision in ISO/CD 13485 states, ``If appropriate, 
    special provisions shall be established, documented and maintained for 
    the handling of used product in order to prevent contamination of other 
    product, the manufacturing environment and
    
    [[Page 52636]]
    
    personnel.'' FDA agrees with this requirement and has therefore added 
    the term ``contamination'' to Secs. 820.140 Handling and 820.150 
    Storage.
    ii. Storage (Sec. 820.150)
        174. Two comments stated that proposed Sec. 820.122 Storage should 
    be amended to be similar to ISO 9001, and that the rest of the 
    requirements should be deleted and included in a guidance document. One 
    comment stated that the term ``obsolete'' should be deleted because, 
    although a device may no longer be sold, thereby making it obsolete, 
    the components for that device may still be stored for customer support 
    of the existing devices.
        FDA agrees that Sec. 820.122, now Sec. 820.150, could be more 
    consistent with ISO 9001 and has revised the section to harmonize with 
    ISO 9001:1994. FDA has not deleted the term ``obsolete.'' FDA 
    understands that a device may no longer be sold, but that parts and 
    subassemblies may still be required for customer support; therefore, 
    those components or subassemblies are not ``obsolete.'' FDA's intent in 
    this requirement is to ensure that only the appropriate product be used 
    or distributed.
        FDA has deleted the requirement that control numbers or 
    identifications be legible and visible because it believes the 
    requirement is inherent in Sec. 820.150(a), which requires the 
    manufacturer to establish procedures to prevent mixups. To do this, a 
    manufacturer must ensure that product can be properly identified.
        175. A comment stated that restricting access to designated areas 
    through the use of keys, bar code readers, or other means, should be 
    sufficient to meet the intent of the requirement in proposed 
    Sec. 820.122(b), without the need for written procedures for 
    authorizing receipt.
        FDA has not deleted the requirement for procedures, now in 
    Sec. 820.150(b), to authorize receipt of product because the agency 
    believes that strict control over product in storage areas and stock 
    rooms results in decreased distribution of nonconforming product. Thus, 
    even where locked storage rooms are utilized, the procedures should 
    detail, among other things, who is permitted access and what steps 
    should be followed prior to removal.
    iii. Distribution (Sec. 820.160)
        176. A few comments on proposed Sec. 820.124 Distribution stated 
    that there are times when ``first in, first out'' inventory procedures 
    may not be in the best interest of the customer. The comments said that 
    especially when expiration dating is defined and labeled, a ``first in, 
    first out'' system should not be required. The GHTF and other EU 
    comments stated that if a new section ``Contract review,'' similar to 
    ISO 9001:1994, section 4.3 was not added to the regulation, the 
    requirement that ``purchase orders are reviewed to ensure that 
    ambiguities and errors are resolved before devices are released for 
    distribution'' should be added to this section.
        FDA agrees with the comments. FDA has amended the requirement in 
    Sec. 820.160 to state that the procedures must ensure that ``expired 
    devices or devices deteriorated beyond acceptable fitness for use'' are 
    not distributed. FDA has also added the sentence on reviewing purchase 
    orders.
        177. A few comments on proposed Sec. 820.124(b) stated that class I 
    devices should be exempt, or that the requirement should apply only to 
    critical devices, because all devices do not require control numbers. 
    Other comments stated that the term ``consignee'' should be defined, or 
    the word ``primary'' should be added before ``consignee'' for clarity.
        FDA agrees in part with the comments and in Sec. 820.160(b) has 
    added the term ``initial'' before ``consignee'' to make clear that the 
    requirement for maintaining distribution records extends to the first 
    consignee. FDA has retained the word ``consignee'' and notes that it is 
    a person to whom the goods are delivered. FDA has also clarified 
    Sec. 820.160(b)(4) by requiring ``Any control number(s) used.'' 
    Therefore, if the manufacturer is required by Sec. 820.65 to have 
    control numbers, these must be recorded along with any control numbers 
    voluntarily used. Logically, control numbers are used for traceability 
    so they should be recorded in the DHR distribution records. FDA 
    disagrees, however, that the requirement to maintain distribution 
    records should not apply to class I devices. The information required 
    by this section is basic information needed for any class of product in 
    order to conduct recalls or other corrective actions when necessary.
    iv. Installation (Sec. 820.170)
        178. Several comments received on proposed Sec. 820.126, 
    Installation stated that not all devices require installation. Several 
    comments on the Working Draft asked that, ``The results of the 
    installation inspection shall be made available to FDA upon request'' 
    be deleted because this was redundant with FDA's access to these 
    documents under Sec. 820.180.
        FDA agrees with the first set of comments. As discussed in 
    Sec. 820.1, the installation requirements only apply to devices that 
    are capable of being installed. However, to further clarify the 
    requirements in Sec. 820.170, FDA has made clear that the requirement 
    applies to ``devices requiring installation.'' FDA also agrees that the 
    sentence on document availability is redundant with Sec. 820.180 for 
    all records and has deleted the sentence.
        179. Several comments raised the issue of applying the regulation 
    requirements to third party installers.
        FDA has rewritten Sec. 820.170. Persons who install medical devices 
    have been regulated under the original CGMP under Sec. 820.3(k) which 
    describes a manufacturer as one who ``assembles or processes a finished 
    medical device,'' and continue to be regulated under this quality 
    system regulation under Sec. 820.3(o). Section 820.152 Installation of 
    the original CGMP discussed the manufacturer or its authorized 
    representative and persons other than the manufacturer's 
    representative. This regulation eliminates that terminology. Under the 
    revised requirement in Sec. 820.170(a), the manufacturer establishes 
    installation and inspection instructions, and where appropriate test 
    procedures. The manufacturer distributes the instructions and 
    procedures with the device or makes them available to person(s) 
    installing the device. Section 820.170(b) requires that the person(s) 
    installing the device follow the instructions and procedures described 
    in Sec. 820.170(a) and document the activities described in the 
    procedures and instructions to demonstrate proper installation.
        The revised provisions in Sec. 820.170(b) explicitly require that 
    the installation be performed according to the manufacturer's 
    instructions, regardless of whether the installer is employed by or 
    otherwise affiliated with the manufacturer. Section 820.170(b) requires 
    records to be kept by whomever performs the installation to establish 
    that the installation was performed according to the procedures. Such 
    records will be available for FDA inspection. FDA does not expect the 
    manufacturer of the finished device to maintain records of installation 
    performed by those installers not affiliated with the manufacturer, but 
    does expect the third party installer or the user of the device to 
    maintain such records.
        FDA believes that making these requirements explicit in the 
    regulation is necessary to ensure that devices are safe and effective, 
    and that they perform as intended after installation. FDA notes
    
    [[Page 52637]]
    
    again that installers are considered to be manufacturers under the 
    original CGMP regulation and that their records are, and will continue 
    to be, subject to FDA inspections when the agency deems it necessary to 
    review such records.
    
    M. Records (Subpart M)
    
    i. General Requirements (Sec. 820.180)
        180. Several comments under Sec. 820.180 General requirements 
    suggested that FDA delete the requirement that records be stored to 
    allow ``rapid retrieval'' because a reasonable time frame should be 
    allowed. One comment stated that the wording of the section needed to 
    be amended to allow records to be located in different places, 
    especially for foreign manufacturers and distributors. Two comments 
    stated that the requirement should be qualified by ``subject to 
    conflicting legal requirements in other countries'' because some 
    countries have ``blocking statutes'' that would prohibit the release of 
    some information. One comment stated that wherever the word ``all'' 
    appeared in the requirements, FDA should remove it.
        FDA has rearranged this section, and notes that records must be 
    kept in a location that is ``reasonably accessible'' to both the 
    manufacturer and FDA investigators, and that records must be made 
    ``readily available.'' FDA expects that such records will be made 
    available during the course of an inspection. If the foreign 
    manufacturer maintains records at remote locations, such records would 
    be expected to be produced by the next working day or 2, at the latest. 
    FDA has clarified that records can be kept at other than the inspected 
    establishment, provided that they are made ``readily available'' for 
    review and copying. This should provide foreign manufacturers and 
    initial distributors the necessary flexibility.
        FDA has not qualified Sec. 820.180 in response to the comments on 
    the ``blocking statues'' because if manufacturers want to import 
    medical devices into the United States, then they must comply with 
    applicable statutory and regulatory requirements, including part 820. 
    The records section of this regulation is essentially the same as that 
    of the original CGMP and FDA has not found these ``blocking statutes'' 
    to present a problem. Further, countries increasingly realize the 
    importance of a global market, thus FDA does not anticipate this issue 
    to be a problem in the future.
        In response to the comment on the term ``all'', FDA notes that 
    where a requirement exists for ensuring that records are maintained in 
    a certain fashion, a manufacturer must keep all records subject to the 
    regulation in that manner. The revised section makes clear that it is 
    ``all records required'' by the regulation to which the section's 
    requirements pertain.
        181. A few comments on Sec. 820.180(b), ``Record retention 
    period,'' stated that the section should be amended because all quality 
    records may not be tied to a specific device; therefore, such quality 
    records may not need to be maintained over the lifetime of a device. A 
    few comments stated that the retention period requirement is unclear 
    and burdensome, while others stated that the period should be left to 
    the manufacturer to define. One comment suggested the deletion of the 
    requirements related to photocopying records in proposed 
    Sec. 820.180(b) because it is technology that is not necessarily being 
    used.
        FDA believes that all records should be retained for a period 
    equivalent to the design and expected life of the device, but in no 
    case less than 2 years, whether the records specifically pertain to a 
    particular device or not. The requirement has been amended to make 
    clear that all records, including quality records, are subject to the 
    requirement. FDA believes this is necessary because manufacturers need 
    all such records when performing any type of investigation. For 
    example, it may be very important to access the wording of a complaint 
    handling procedure at the time a particular complaint came in when 
    investigating a trend or a problem that extends to several products or 
    over an extended period of time. Further, FDA does not believe that 
    allowing the manufacturer to define the retention period will serve the 
    public's best interest with regard to safety concerns and hazard 
    analysis.
        In response to the comment on photocopying, FDA has deleted the 
    last two sentences. The agency believes that this requirement is 
    outdated and does not necessarily reflect the technology being utilized 
    today. Section 820.180 requires that records be readily available for 
    inspection and copying by FDA, and FDA will interpret ``copying'' to 
    include the printing of computerized records, as well as photocopying.
        182. One comment on proposed Sec. 820.180(c) stated that all 
    quality audit reports should be subject to FDA review and public 
    disclosure. A few other comments stated that for a management 
    representative to certify that ``corrective action has been taken'' 
    would be difficult because some corrective actions are long term and 
    may not be completed at the time of certification.
        FDA disagrees with the comment that quality audit reports should be 
    subject to FDA review for the reasons given in the preamble of the 
    original CGMP regulation, published in the Federal Register on July 21, 
    1978 (43 FR 31508), and believes that the disclosure of the audit 
    reports themselves would be counterproductive to the intent of the 
    quality system. FDA has added Sec. 820.180(c), ``Exceptions,'' to 
    address which records FDA, as a matter of policy, will not request to 
    review or copy during a routine inspection; such records include 
    quality audit reports. FDA may request an employee in management with 
    executive responsibility to certify in writing that the management 
    reviews, quality audits, and supplier audits (where conducted) have 
    been performed, among other things. FDA may also seek production of 
    these reports in litigation under applicable procedural rules or by 
    inspection warrant where access to the records is authorized by 
    statute. Again, FDA emphasizes that its policy of refraining from 
    reviewing these reports extends only to the specific reports, not to 
    the procedures required by the sections or to any other quality 
    assurance records, which will be subject to review and copying.
        FDA agrees with the comments on the timing of corrective actions 
    and has amended the certification requirement to state ``corrective 
    action has been undertaken.''
    ii. Device Master Record (DMR) (Sec. 820.181)
        183. A few comments on proposed Sec. 820.181 Device master record 
    stated that the requirement for a ``qualified'' individual to prepare 
    the DMR should be deleted because it is unclear or redundant with the 
    requirements in Sec. 820.25.
        FDA has not deleted the requirement for the DMR to be prepared, 
    dated, and approved by a qualified individual because the agency 
    believes this is necessary to assure consistency and continuity within 
    the DMR. The section is consistent with the original CGMP, 
    Sec. 820.181. FDA has, however, substituted the phrase ``prepared and 
    approved in accordance with Sec. 820.40'' to be consistent with the 
    requirements already in Sec. 820.40 and to eliminate any redundancy.
        184. Two comments on Sec. 820.181(a) stated that ``software design 
    specifications'' should not be included in the DMR because these 
    documents will be located in the DHF. Another comment requested that 
    the requirement that the DMR contain ``software source code'' 
    information be amended because
    
    [[Page 52638]]
    
    source codes for commercialized software will not be available to the 
    device manufacturers. Another comment stated that the source code 
    should not be in the DMR because it will already be in the DHF.
        FDA deleted the reference to ``software source code'' because this 
    is already covered with the requirement for ``software 
    specifications.'' The final software specifications should be 
    transferred into production. Therefore, the final software 
    specification for the particular device or type of device should be 
    located or referenced in the DMR, while any earlier version should be 
    located or referenced in the DHF. FDA believes that it is more 
    important for manufacturers to construct a document structure that is 
    workable and traceable, than to worry about whether something is 
    contained in one file as compared to another. The DMR is set up to 
    contain or reference the procedures and specifications that are current 
    on the manufacturing floor. The DHF is meant to be more of a historical 
    file for utilization during investigations and continued design 
    efforts.
        185. One comment on Sec. 820.181(c) stated that the DMR should not 
    contain quality system documents, but rather the quality control 
    documents related to the specific device. Three comments stated that 
    validation and verification information belongs in the DHF, not the 
    DMR.
        FDA agrees in part with the comments and has revised the section to 
    clarify that the quality records required in the DMR relate to the 
    specific current design, not the more general requirements of the 
    quality system, which are addressed under new Sec. 820.186. However, 
    the comments are incorrect that all validation and verification 
    information is related solely to design. There are requirements for 
    validation and verification pertaining to device processing that may be 
    better kept in the DMR instead of the DHF. The documentation of such 
    verification and validation activities relating to processes that are 
    performed for several different devices or types of devices can be 
    placed or referenced in the location that best suits the manufacturer. 
    Again, it is more important that the manufacturer store and retrieve 
    information in a workable manner, than keep such information in 
    particular files.
        186. FDA notes that the regulation contains a few requirements 
    which apply ``where appropriate'' or ``at appropriate stages.'' FDA 
    emphasizes that the procedures that the manufacturer places in the DMR 
    must clearly define the requirements the manufacturer is following and 
    when particular activities are appropriate. The manufacturer will have 
    failed to comply with the requirements of the section if the procedures 
    simply state that the review or activity occurs at ``appropriate 
    stages.''
        The same principle applies for every section of this regulation, 
    which is written to be flexible enough to cover the manufacture of all 
    types of devices. Manufacturers must adopt quality systems appropriate 
    for their specific products and processes. In establishing these 
    procedures, FDA will expect manufacturers to be able to provide 
    justifications for the decisions reached.
    iii. Device History Record (Sec. 820.184)
        187. One comment on Sec. 820.184 stated that labeling should not be 
    required in the DHR because it is already required in the DMR. Another 
    comment stated that some devices have 25 or more labels and that only 
    the primary identification labels are necessary in the DHR. One comment 
    stated the requirement should be amended because it explicitly requires 
    that dates and quantities for each batch be in the DHR, while only 
    implying through the general requirement that the DHR must also contain 
    the batch test data.
        FDA agrees that it may not be necessary to include all labeling 
    used in the DHR. However, FDA continues to believe, as it explained in 
    the preamble to proposed regulation published in the Federal Register 
    on November 23, 1993 (58 FR 61952 at 61968), that increased control 
    over labeling is necessary due to the many labeling errors resulting in 
    recalls. Therefore, FDA has retained a requirement related to labeling 
    in the DHR, but revised it to make it less burdensome. The requirement 
    was amended to ``the primary identification label and labeling'' which 
    is consistent with that contained in the original CGMP regulation, 
    Sec. 820.185. FDA believes that the requirement that the DHR contain 
    the primary label and labeling used for each production unit, coupled 
    with the labeling controls in Sec. 820.120, should help to ensure that 
    proper labeling is used and, hopefully, decrease the number of recalls 
    due to improper labeling.
        FDA agrees with the last comment and has added in Sec. 820.184 
    ``(d) The acceptance records which demonstrate the device is 
    manufactured in accordance with the DMR'' to explicitly state the 
    requirement to avoid any confusion.
        188. FDA has deleted the requirement for the DHR to be ``readily 
    accessible and maintained by a designated individual(s)'' because it 
    believes that the objective of that requirement is met through 
    Secs. 820.40 Document controls and 820.180 General requirements.
        FDA has also added ``device identification'' to the requirement 
    under Sec. 820.184(f) because it believes that any identification or 
    control number used should be documented in the DHR to facilitate 
    investigations, as well as corrective and preventive actions. FDA notes 
    that this provision does not add any requirement for identification or 
    traceability not already expressed in Secs. 820.60 and 820.65.
    iv. Quality System Record (Sec. 820.186)
        189. Several comments stated that the regulation should more 
    closely harmonize with ISO 9001:1994. A few comments stated that the 
    regulation should include the requirements for a quality manual. One 
    comment stated that general quality system procedures and instructions 
    should not be required in the DMR because the DMR is device specific, 
    and many quality system procedures are not tied to a particular device.
        FDA agrees in part with these comments and has developed new 
    Sec. 820.186 Quality system record. This section separates the 
    procedures and documentation of activities that are not specific to a 
    particular type of device from the device specific records.
    v. Complaint Files (Sec. 820.198)
        190. Two comments on proposed Sec. 820.198 Complaint files stated 
    that the requirements were very detailed and that much of the language 
    should be placed in a guidance document.
        FDA disagrees with the comments. These requirements are essentially 
    the same as the original CGMP requirements under Sec. 820.198, and 18 
    years of experience with these requirements shows that many 
    manufacturers still do not understand and properly handle complaints. 
    Therefore, FDA believes that the amount of detail in Sec. 820.198 is 
    appropriate and necessary. In an effort to make the requirements more 
    clear, however, the section has been reorganized to better illustrate 
    how complaint information should be handled.
        Section 820.198(a) sets forth the general requirement for 
    establishing and maintaining a complaint handling procedure and 
    includes a few items that the procedure needs to address. Section 
    820.198(b) discusses the initial review and evaluation of the 
    complaints in order to determine if complaints are ``valid.'' It is 
    important to note that this evaluation is not the same as a complaint 
    investigation. The evaluation
    
    [[Page 52639]]
    
    is performed to determine whether the information is truly a complaint 
    or not and to determine whether the complaint needs to be investigated 
    or not. If the evaluation decision is not to investigate, the 
    justification must be recorded. Section 820.198(c) then describes one 
    subset of complaints that must be investigated, but explains that 
    duplicative investigations are not necessary. In cases where an 
    investigation would be duplicative, a reference to the original 
    investigation is an acceptable justification for not conducting a 
    second investigation. Section 820.198(d) describes another subset of 
    complaints that must be investigated (those that meet the MDR criteria) 
    and the information that is necessary in the record of investigation of 
    those types of complaints. Section 820.198(e) sets out the type of 
    information that must be recorded whenever complaints are investigated. 
    The information described in Sec. 820.198 (e)(1) through (e)(5) would 
    most likely be attained earlier in order to perform the evaluation in 
    Sec. 820.198(b). This information need not be duplicated in the 
    investigation report as long as the complaint and investigation report 
    can be properly identified and tied together. Section 820.198 (e)(1) 
    through (e)(5) are considered to be basic information essential to any 
    complaint investigation. If there is some reason that the information 
    described in Sec. 820.198(e) cannot be obtained, then the manufacturer 
    should document the situation and explain the efforts made to ascertain 
    the information. This will be considered to be acceptable as long as a 
    reasonable and good faith effort was made. For example, a single phone 
    call to a hospital would not be considered by FDA to be a reasonable, 
    good faith effort to obtain information. Section 820.198(f) is the same 
    as Sec. 820.198(d) of the original CGMP, where the manufacturing 
    facility is separate or different from that of the formally designated 
    complaint handling unit. In such cases, it is important that the 
    facility involved in the manufacturing of the device receive or have 
    access to complaint and investigation information. In order to give 
    manufacturers the flexibility of using computer or automated data 
    processing systems, the term ``reasonably accessible,'' from 
    Sec. 820.180, is used. Section 820.198(g) is the complaint 
    recordkeeping requirement for distributors. In order to give 
    manufacturers the same flexibility as described in Sec. 820.198(f), FDA 
    has included ``reasonably accessible'' in Sec. 820.198(g).
        Throughout Sec. 820.198, when there is reference to the MDR 
    regulation or to the types of events that are reportable under the MDR 
    regulation, this section simply refers to events or complaints that 
    ``represent an event which is required to be reported to FDA under part 
    803 or 804 of this chapter.''
        191. A few comments on Sec. 820.198(a) stated that the section 
    should allow for more than one ``formally designated unit'' to handle 
    complaints, especially for large corporations where it would not be 
    feasible or beneficial for all divisions to have a single complaint 
    handling unit. A few other comments stated that Sec. 820.198(a)(2) on 
    oral complaints should be deleted because it is too subjective.
        FDA disagrees with these comments. Large corporations may have 
    different complaint handling units for different product types or 
    different manufacturing establishments. However, there should be only 
    one formally designated complaint handling unit for each product type 
    or establishment. If a corporation chooses to operate with different 
    complaint handling units for products and/or establishments, the 
    manufacturer must clearly describe and define its corporate complaint 
    handling procedure to ensure consistency throughout the different 
    complaint handling units. A system that would allow multiple 
    interpretations of handling, evaluating, categorizing, investigating, 
    and following up, would be unacceptable. Each manufacturer should 
    establish in its procedures which one group or unit is ultimately 
    responsible for coordinating all complaint handling functions.
        FDA also disagrees that the requirement that oral complaints be 
    documented upon receipt should be deleted. A December 1986 General 
    Accounting Office (GAO) report entitled ``Medical Devices; Early 
    Warning of Problems Is Hampered by Severe Underreporting,'' (Ref. 11) 
    showed that approximately 83 percent of the hospitals report complaints 
    orally. FDA believes that these oral complaints must be captured in the 
    complaint handling process.
        192. FDA, as noted above, has added to Sec. 820.198(c) the phrase 
    ``unless such investigation has already been performed for a similar 
    complaint and another investigation is not necessary'' to clarify that 
    duplicative investigations are not required if the manufacturer can 
    show that the same type of failure or nonconformity has already been 
    investigated.
        193. Several comments on proposed Sec. 820.198(b), now 
    Sec. 820.198(d), stated that the evaluation of complaints pertaining to 
    death, injury, or hazard to health should be removed from this section 
    because it is redundant with the MDR regulation. Several other comments 
    on Sec. 820.198(b) stated that complaints pertaining to death, injury, 
    or hazard to health need not be maintained separately, as long as they 
    are identified.
        FDA disagrees that the requirements are redundant, but believes 
    that they expressly state what is expected in the handling of this type 
    of complaint information. The requirements have been moved to a 
    separate section, Sec. 820.198(d).
        FDA agrees with the second set of comments and has revised the 
    section to permit such complaints to be ``clearly identified.'' This 
    will give a manufacturer flexibility in choosing a means of ensuring 
    that these types of complaints can be immediately recognized and 
    segregated for purposes of prioritizing and meeting other requirements.
        FDA has substituted the term ``promptly'' for the term 
    ``immediately'' to be more consistent with the new MDR regulation 
    timeframes. FDA has also clarified that Sec. 820.198 (d)(1) through 
    (d)(3) are in addition to the information that must be recorded in 
    Sec. 820.198(e).
        194. A few comments on proposed Sec. 820.198 (c) and (d) stated 
    that FDA should make clear that some of the requirements will not 
    always be applicable. For example, the comments stated that a record of 
    corrective action cannot be made if such action is not required, and is 
    not taken.
        Where corrective action is not necessary and is not taken, it 
    cannot be documented. The section was revised to make that clear. As 
    stated in the preamble to the proposal (58 FR 61952 at 61968), the 
    manufacturer's procedures should clearly identify when corrective 
    action will be taken.
        In addition, FDA combined provisions in Sec. 820.198 (c) through 
    (e) to eliminate redundancy and added the requirement that the records 
    include any device identification, as well as control number used, to 
    facilitate corrective and preventive actions. FDA has also deleted the 
    term ``written'' in Sec. 820.198(e) to be consistent with FDA's 
    statements on electronic and computer systems.
        195. FDA deleted the requirements in proposed Sec. 820.198(f) in 
    response to comments because it agrees that it is not necessary to 
    repeat the requirements of the MDR regulation in the quality system 
    regulation. Section 820.198(a) requires that all complaints be 
    evaluated to determine whether they are subject to
    
    [[Page 52640]]
    
    the requirements of the MDR regulation under part 803 or 804.
        196. A few comments on proposed Sec. 820.198(g), now 
    Sec. 820.198(f), stated that duplicate records are not needed in this 
    age of computer systems, and that the requirement as written would be 
    counterproductive.
        FDA agrees with the comments and has rewritten the section to allow 
    the complaints and records of investigation to be reasonably accessible 
    at the formally designated complaint unit and the manufacturing site, 
    where these locations are distinct. A manufacturer's procedures must 
    ensure that the manufacturing site is alerted to complaints concerning 
    devices produced at that site.
        197. Several comments on proposed Sec. 820.198(h), now 
    Sec. 820.198(g), stated that the requirement is unnecessary, given that 
    FDA can inspect a foreign manufacturer that imports devices, and is 
    burdensome.
        FDA has revised the section to permit the records to be reasonably 
    accessible, similar to Sec. 820.198(f), which should alleviate any 
    burden. However, the agency must have access to these records in the 
    United States.
        198. Several comments on proposed Sec. 820.198 (i) and (j) stated 
    that the requirements should be deleted because they are redundant with 
    the MDR requirements in part 803.
        FDA disagrees that all of the requirements in Sec. 820.198 (i) and 
    (j) are redundant. The requirement that procedures ensure that 
    complaints are processed uniformly and in a timely manner, and 
    evaluated to determine whether they are reportable under part 803 or 
    804, has been moved up to Sec. 820.198(a). These are basic requirements 
    for complaint handling. If the complaint is determined to be of the 
    type subject to part 803 or 804, those requirements apply. The 
    requirements of parts 803 and 804 are not repeated in this regulation. 
    FDA has deleted Sec. 820.198(j).
    
    N. Servicing (Subpart N)
    
        199. Numerous comments were received on the servicing requirements 
    that were proposed. Many of these comments dealt with competitive 
    issues between manufacturers that perform or contract out their own 
    servicing and third party service organizations. The comments received, 
    as well as the recommendations from the GMP Advisory Committee, were 
    split on many issues. Therefore in this regulation, FDA has chosen to 
    codify only longstanding requirements for servicing performed by 
    original manufacturers and remanufacturers. The requirements in 
    Sec. 820.200 are similar to those in ISO 9001:1994, with some 
    supplemental requirements for clarification on monitoring service 
    reports, on the relationship of service reports and complaints, and on 
    the type of information FDA believes is essential in any service 
    report. As described above in the definition section of this preamble, 
    a separate rulemaking will specify and clarify the requirements for 
    third party service organizations.
        200. Other comments on proposed Sec. 820.200(a) stated that it is 
    impractical to return a used device to its original specifications 
    because a certain amount of wear and tear should be expected, without 
    detriment to the safety and effectiveness of the device. Several 
    comments on Sec. 820.200(a) stated that the term ``records'' should be 
    replaced by ``reports,'' to be consistent with ISO 9001.
        FDA agrees and has revised the requirements in Sec. 820.200(a) to 
    be similar to the requirements in ISO 9001:1994 as recommended by 
    comments at the GMP Advisory Committee meeting to require that the 
    servicing instructions and procedures ensure that the device will meet 
    ``specified requirements'' for the device's intended use. FDA is aware 
    that with use and age, a device may be serviced to function as 
    intended, but may not meet original specifications.
        FDA agrees with the comments and has modified the language in 
    Sec. 820.200(b), (c), and (d) to use the term ``service reports.''
        201. A few comments on proposed Sec. 820.200(b), ``Service report 
    evaluation,'' questioned whether full corrective action was necessary 
    for every service report and whether service calls need to be handled 
    as complaints only when there is a death, injury, or hazard to safety.
        FDA has rewritten this section into Sec. 820.200(b) and (c) to 
    clarify the agency's intent and to use terms consistent with those used 
    in Sec. 820.198. Section 820.200(b) now states that ``Each manufacturer 
    shall analyze service reports with appropriate statistical methodology 
    in accordance with Sec. 820.100.'' Full corrective action may not be 
    required for every service report. However, if the analysis of a 
    service report indicates a high risk to health, or that the frequency 
    of servicing is higher than expected, the remainder of the corrective 
    and preventive action elements are applicable, in accordance with the 
    corrective and preventive action procedures established under 
    Sec. 820.100.
        Section 820.200(c) provides that when a service report ``represents 
    an event which must be reported to FDA under part 803 or 804 of this 
    chapter,'' it is automatically considered by FDA to be a complaint that 
    must be handled according to Sec. 820.198. FDA emphasizes that this 
    provision is not intended to limit ``complaints'' to MDR reportable 
    events.
        202. FDA has also added in Sec. 820.200(d) the requirements for 
    recording the name of the device, any device identification(s) and 
    control number(s) used, as well as test and inspection data, because 
    FDA believes such documentation in the service report will facilitate 
    investigations. This additional documentation provision does not add 
    any requirement for identification or traceability not already 
    expressed in Secs. 820.60 and 820.65. Therefore, Sec. 820.200(d) as 
    amended focuses on the type of information that should be captured on 
    the service report instead of where the information should be sent.
    
    O. Statistical Techniques (Subpart O)
    
        203. FDA amended Sec. 820.250(a) to be consistent with the 
    requirements in ISO 9001:1994, section 4.20.
        204. Several comments on Sec. 820.250(b) stated that the provision 
    as written seems to require the use of sampling plans, and that every 
    manufacturer does not necessarily use sampling plans. Another comment 
    stated that sampling plans are not often used during reviews of 
    nonconformities, quality audits, or complaints, and that these examples 
    should, therefore, be deleted. Two other comments questioned the 
    meaning of ``regularly reviewed.''
        FDA's intent was not to require the use of sampling plans, but to 
    require that where they are used, they should be written and valid. 
    Section 820.250 was revised to make that clear. Sampling plans are not 
    always required, but any time sampling plans are used, they must be 
    based on a valid statistical rationale. Further, FDA acknowledges that 
    the most common use of sampling plans is during receiving acceptance, 
    and has deleted the examples. FDA has also clarified the review 
    requirement by stating ``to ensure that when changes occur the sampling 
    plans are reviewed.''
    
    VI. Summary of Changes From the July 1995 Working Draft to the 
    Final Rule and Rationale
    
        Note: Minor changes to improve grammar, readability, and 
    clarity, as well as changes in terminology and organization for the 
    sake of consistency throughout the regulation, are not listed.
    
    [[Page 52641]]
    
    A. Section 820.1  Scope
    
        1. Inserted sentence, ``If a manufacturer engages in only some 
    operations subject to the requirements in this part, and not in others, 
    that manufacturer need only comply with those requirements applicable 
    to the operations in which it is engaged'' for further clarification of 
    the scope in response to many comments.
        2. Amended sentence on component manufacturers to read, ``This 
    regulation does not apply to manufacturers of components or parts of 
    finished devices, but such manufacturers are encouraged to use 
    appropriate provisions of this regulation as guidance'' as a result of 
    the many written comments and oral testimony at the August and 
    September 1995 meetings.
        3. Inserted sentence on how to interpret the phrase ``where 
    appropriate'' in the regulation, as recommended by the GMP Advisory 
    Committee. This sentence is consistent with International Organization 
    for Standards (ISO)/CD 13485-- ``Application of Quality Systems to 
    Medical Devices.''
    
    B. Section 820.3  Definitions
    
        4. Amended the definition of Complaint by inserting ``after it is 
    released for distribution'' in response to comments for clarification 
    and to harmonize with ISO/CD 13485.
        5. Amended the definition of Component by deleting ``packaging'' 
    for clarification that every piece of packaging is not necessarily a 
    component, only the materials that are part of the ``finished, 
    packaged, and labeled device.''
        6. Amended the definition of Design output to clarify its 
    relationship with the Device Master Record.
        7. Amended the definition of Design review to delete ``and propose 
    the development of solutions'' in order to allow the manufacturer the 
    flexibility to determine whom the appropriate person(s) is to propose 
    solutions.
        8. Deleted the definition of End of life in response to the many 
    written comments and oral testimony at the August and September 1995 
    meetings.
        9. Amended the definition of Manufacturer to delete component 
    manufacturers and to remove the terms ``servicer'' and ``refurbisher.'' 
    The obligations of servicers and refurbishers will be addressed in a 
    separate rulemaking later this year. The terms ``installation'' and 
    ``remanufacturing'' were added to codify longstanding FDA policy and 
    interpretations of the original CGMP regulation.
        10. Amended the definition of Manufacturing material in response to 
    comments requesting clarification and separation of examples.
        11. Deleted the definition of Record to avoid confusion. Record 
    will continue to be defined by the act and case law.
        12. Removed the definition of Refurbisher for reasons discussed in 
    paragraph 28, section V.A. of this document.
        13. Inserted the definition of Remanufacturer for reasons discussed 
    in paragraph 28, section V.A. of this document, and made the language 
    consistent with that of the 510(k) provision and the PMA amendment/
    supplement requirements.
        14. Changed the term Reprocessing to Rework and adopted a 
    definition consistent with ISO 8402 Quality Management and Assurance 
    Vocabulary Standard in response to comments for closer harmonization of 
    terminology.
        15. Removed the definition of Servicing, and Servicer which was 
    proposed to the GMP Advisory Committee, for reasons discussed above.
        16. Amended the definition of Validation as recommended by the GMP 
    Advisory Committee for further clarity by delineating the terms 
    validation, process validation, and design validation.
        17. Amended the definition of Verification for further clarity in 
    response to comments and to more closely harmonize with ISO 8402.
    
    C. Section 820.5  Quality System
    
        18. Deleted the requirements in Sec. 820.5(a) and (b) because these 
    requirements are now found in Sec. 820.20.
    
    D. Section 820.20  Management Responsibility
    
        19. Moved the requirements from Sec. 820.186 and rewrote into new 
    Sec. 820.20(d) and (e) for clarity, better organization, and closer 
    harmonization with ISO/CD 13485.
    
    E. Section 820.25  Personnel
    
        20. Inserted the phrase, ``establish procedures for identifying 
    training needs'' in Sec. 820.25(b) in response to comments to add this 
    requirement and to harmonize with the requirement in ISO/CD 13485.
        21. Deleted the sentence in Sec. 820.25 on understanding the ``CGMP 
    requirements applicable to their job function'' to provide 
    manufacturers with the flexibility to appropriately train personnel.
    
    F. Section 820.30  Design Controls
    
        22. Amended the requirements in Design and development planning for 
    clarity and to more closely harmonize with ISO/CD 13485.
        23. In Sec. 820.30(c), inserted the sentence, ``The procedures 
    shall include a mechanism for addressing incomplete, ambiguous, or 
    conflicting requirements'' in response to comments to add this 
    requirement and to harmonize with the requirement in ISO/CD 13485.
        24. In Sec. 820.30(d), deleted the sentence, ``Design output 
    procedures shall ensure that design output meets the design input 
    requirements'' because this was redundant with the requirement in 
    Sec. 820.30(f) Design verification.
        25. Amended Sec. 820.30(e) Design review to clarify that the 
    procedures shall ensure that an independent person is included in 
    design reviews.
        26. Section 820.30(f) Design verification and validation was split 
    into two paragraphs, (f) Design verification and (g) Design validation 
    and the requirements were separated between the two paragraphs, in 
    response to many written comments and oral testimony at the August and 
    September 1995 meetings and to improve clarity and consistency with 
    ISO/CD 13485.
        27. Amended the requirement for Sec. 820.30(i) Design changes to 
    add the phrase ``before their implementation'' due to an inadvertent 
    omission in the July 1995 Working Draft.
    
    G. Section 820.50  Purchasing Controls
    
        28. Deleted the last two sentences in Sec. 820.50(b) and inserted 
    ``Purchasing data shall be approved in accordance with Sec. 820.40'' 
    because the last two sentences were redundant with the requirements in 
    Sec. 820.40.
    
    H. Section 820.65  Traceability
    
        29. Substituted the definition of critical device from the original 
    CGMP for the phrase ``where necessary to ensure the protection of the 
    public health,'' in response to many comments requesting clarification 
    as to when traceability is necessary.
        30. Added ``where appropriate'' for the traceability of components 
    in response to the recommendation of the GMP Advisory Committee, the 
    written comments, and to harmonize with ISO/CD 13485.
    
    I. Section 820.70  Production and Process Controls
    
        31. Inserted ``identified and approved'' in Sec. 820.70(a)(5) 
    before ``representative samples'' to clarify that the samples have to 
    be established and deemed appropriate before they are used as a 
    standard.
        32. Substituted in Sec. 820.70(b) ``where appropriate validated 
    according to Sec. 820.75'' for ``unless inspection and test
    
    [[Page 52642]]
    
    fully verifies the results of the changes'' because it was redundant 
    with the requirements set forth in Sec. 820.75.
        33. Amended the requirement in Sec. 820.70(c) to apply only ``Where 
    environmental conditions could reasonably be expected to have an 
    adverse effect on product quality,'' in response to comments and to be 
    consistent with the original CGMP requirements.
        34. Amended the requirements in Sec. 820.70(d) and (e) to include 
    ``could reasonably be expected to have an adverse effect on product 
    quality'' to consistently qualify when these provisions are 
    appropriate.
        35. Amended the requirement in Sec. 820.70(h) to apply only ``Where 
    a manufacturing material could reasonably be expected to have an 
    adverse effect on product quality,'' in response to comments and to be 
    consistent with the original CGMP requirements.
        36. Rearranged the wording in Sec. 820.70(i) to clarify ``automated 
    data processing systems.''
    
    J. Section 820.72  Inspection, Measuring, and Test Equipment
    
        37. Renumbered Sec. 820.84 as Sec. 820.72 for better organization 
    because Inspection, measuring, and test equipment requirements are more 
    appropriate under Subpart G--Production and Process Controls than under 
    Subpart H--Acceptance Activities.
        38. Section 820.72(b) ``Calibration standards'' and (c) 
    ``Calibration records'' were reorganized as paragraphs(1) and (2), 
    respectively under paragraph (b) ``Calibration.''
        39. Amended Sec. 820.72(b) to include provisions for remedial 
    action to ``reestablish the limits and to evaluate whether there was 
    any adverse effect on the device's quality'' in response to comments 
    which questioned whether this was adequately covered under 
    Sec. 820.100.
        40. Section 820.84(d), ``Maintenance,'' is reorganized into 
    Sec. 820.72(a) ``Control of inspection, measuring, and test equipment'' 
    and ``test software'' is deleted because it is considered to be covered 
    under ``electronic inspection and test equipment'' in the general 
    requirement.
    
    K. Section 820.75  Process Validation
    
        41. Section 820.75(a) is amended for clarity. The phrase ``with a 
    high degree of assurance'' was deleted from the definition of 
    ``Validation'' and added as a requirement under process validation.
        42. Section 820.75(b)(2) was amended to state ``where appropriate, 
    the individual(s) performing the process or the major equipment used'' 
    in response to comments requesting that flexibility be given to the 
    manufacturer to determine when these items needed to be documented.
        43. Section 820.75 (c) and (d) were redesignated as paragraphs 
    (b)(1) and (b)(2) for better organization and flow.
        44. Section 820.75(c) was added to address comments and concerns on 
    when revalidation activities were necessary.
    
    L. Section 820.80  Receiving, In-process, and Finished Device 
    Acceptance
    
        45. Section 820.80(c) was amended to add ``where appropriate'' to 
    reinforce the discussion in the preamble that in- process testing is 
    not always necessary depending upon the type of device and the 
    manufacturing set-up.
    
    M. Section 820.90  Nonconforming Product
    
        46. Amended the requirement in Sec. 820.90(a) to include, ``The 
    evaluation of nonconformance shall include a determination of the need 
    for an investigation * * *. The evaluation and any investigation shall 
    be documented.'' in response to many written comments and oral 
    testimony at the August and September 1995 meetings on whether every 
    nonconformance had to be investigated.
        47. Amended the requirement in Sec. 820.90(b)(1) to read, 
    ``Documentation shall include the justification for use of 
    nonconforming product'' in response to several comments confused about 
    the meaning of the term ``concession.''
        48. In Sec. 820.90(b)(2), substituted the term ``rework'' for the 
    term ``reprocessing'' for reasons described in the definitions section.
        49. Deleted the sentence, ``Reprocessed product shall be clearly 
    identified during reprocessing, and shall be subjected to 
    reevaluation'' in Sec. 820.90(b)(2) because the requirement was 
    redundant with the requirements in Secs. 820.60 Identification and 
    820.86 Acceptance status.
    
    N. Section 820.100  Corrective and Preventive Action
    
        50. Amended Sec. 820.100(a)(7) to clarify what information is to be 
    submitted to management for review.
    
    O. Section 820.120  Device Labeling
    
        51. Inserted ``where appropriate'' before ``use'' in 
    Sec. 820.120(a) because every device may not have a label directly 
    affixed to the device itself (e.g. implantable devices).
        52. Inserted the sentence, ``The label and labeling used for each 
    production unit, lot, or batch shall be documented in the DHR'' into 
    Sec. 820.120(d) in response to comments questioning whether the 
    labeling used should be recorded in the device master record or the 
    device history record.
    
    P. Section 820.160  Distribution
    
        53. Inserted the requirement in Sec. 820.160 ``that purchase orders 
    are reviewed to ensure that ambiguities and errors are resolved before 
    devices are released for distribution'' in response to the GHTF 
    comments and other EU comments that the regulation did not address the 
    requirements in ISO 9001, section 4.3, ``Contract Review.''
    
    Q. Section 820.170  Installation
    
        54. Amended the installation requirements for clarity and deleted 
    the last sentence in Sec. 820.170(b), ``The results of the installation 
    inspection shall be made available to FDA upon request'' because this 
    sentence is redundant with the requirements in Sec. 820.180 for all 
    records.
    
    R. Section 820.181   Device Master Record (DMR)
    
        55. In Sec. 820.181 deleted the phrase ``dated, and signature of 
    the qualified individual(s) designated by the manufacturer'' and 
    inserted ``and approved in accordance with Sec. 820.40'' to be 
    consistent with the requirements already in Sec. 820.40.
        56. In Sec. 820.181 deleted the phrase ``and software source code 
    for customized software'' because comments stated that this was already 
    covered with the requirement for ``software specifications.''
    
    S. Section 820.186  Quality System Record (QSR)
    
        57. Amended the requirement in Sec. 820.186 by adding the sentence,
    
        The QSR shall include, or refer to the location of, procedures 
    and the documentation of activities required by this part that are 
    not specific to a particular type of device(s), including but not 
    limited to the records required by Sec. 820.20. Each manufacturer 
    shall ensure that the QSR is prepared and approved in accordance 
    with Sec. 820.40.
    
    Deleted the requirements in Sec. 820.186(a) through (d) because those 
    requirements are now found in Sec. 820.20. This change was in response 
    to comments and suggestions made by the GHTF for further harmonization 
    with ISO/CD 13485 and for clarity.
    
    T. Section 820.198  Complaint Files
    
        58. In Sec. 820.198 deleted the terminology ``pertaining to death,
    
    [[Page 52643]]
    
    injury, or any hazard to safety'' throughout this section and inserted 
    ``an event which must be reported to the FDA under part 803 or 804 of 
    this chapter'' to reference the MDR regulation.
        59. Added the phrase ``unless such investigation has already been 
    performed for a similar complaint and another investigation is not 
    necessary'' in Sec. 820.198(c) in response to comments which thought a 
    second investigation was always mandated by this requirement.
        60. Amended Sec. 820.198(d) by changing the word ``immediately'' to 
    ``promptly'' to be consistent with the new MDR regulation. Added, ``In 
    addition to the information required by Sec. 820.198(e),'' to clarify 
    that an investigation under Sec. 820.198(d) was to include requirements 
    under paragraphs (d)(1) through (d)(3) and under paragraphs (e)(1) 
    through (e)(8).
        61. Substituted the phrase ``reasonably accessible'' for 
    ``concurrently maintained'' in Sec. 820.198 (f) and (g) as recommended 
    by the GMP Advisory Committee to clarify FDA's intent of allowing these 
    records to be available in other media forms besides the hard copies 
    which were previously required.
    
    U. Section 820.200  Servicing
    
        62. Amended Sec. 820.200(a) to adopt language consistent with ISO/
    CD 13485, which was suggested at the GMP Advisory Committee meeting, in 
    order to clarify the requirement and further harmonize.
        63. Deleted the last two sentences in Sec. 820.200(a) on providing 
    information to third party servicers since this industry will be 
    addressed in a separate rulemaking, as discussed above.
        64. Section 820.200(d) was amended for clarity and to focus on the 
    service report and what type of information should be captured on the 
    report instead of where the information should be sent.
    
    V. Section 820.250  Statistical Techniques
    
        65. Amended Sec. 820.250(b) by inserting the phrase, ``to ensure 
    that when changes occur the sampling plans are reviewed'' in response 
    to comments for clarification on when the plans needed to be reviewed.
    
    VII. Environmental Impact
    
        The agency has determined under 21 CFR 25.24(a)(8) and (a)(10) that 
    this action is of a type that does not individually or cumulatively 
    have a significant effect on the human environment. Therefore, neither 
    an environmental assessment nor an environmental impact statement is 
    required.
    
    VIII. Intergovernmental Partnership
    
        The agency has analyzed this rulemaking in accordance with the 
    principles and criteria set forth in Executive Order 12875, ``Enhancing 
    the Intergovernmental Partnership'' and in the Unfunded Mandates Reform 
    Act of 1995 (Pub. L. 104-4). Executive Order 12875 states that no 
    agency or executive department shall issue any regulation that is not 
    required by statute and that creates a mandate upon a State, local, or 
    tribal government unless the Federal Government supplies funds 
    necessary to comply with the mandate, or the agency provides the Office 
    of Management and Budget (OMB) a description of the agency's 
    consultation with affected State, local, and tribal governments, the 
    nature of their concerns, any written communications submitted to the 
    agency by such units of government, and the agency's position 
    supporting the need to issue the regulation containing the mandate. 
    Executive Order 12875 does not apply to this final rule because the 
    regulatory requirements are not generally applicable to government 
    facilities but to finished device manufacturers. The agency notes, 
    however, that the membership of the advisory committee established to 
    review this regulation and make recommendations to the agency on the 
    feasibility and reasonableness of the regulation (GMP Advisory 
    Committee) must include three members who are officers or employees of 
    any State or local government or of the Federal Government, and that in 
    1995 this committee included two State government representatives and 
    one Federal Government representative.
        The agency has also examined the consistency of this final rule 
    with the Unfunded Mandates Reform Act of 1995. The Unfunded Mandates 
    Reform Act requires (in section 202) that agencies prepare an 
    assessment of anticipated costs and benefits before proposing any rule 
    that may result in an annual expenditure by State, local, and tribal 
    governments, in the aggregate, or by the private sector, of $100 
    million (adjusted annually for inflation). FDA believes that the 
    private sector expenditures for this rule fall below $100 million 
    annually but nonetheless, due to uncertainties of these estimates, the 
    agency has prepared for the private sector an assessment of anticipated 
    costs and benefits for the 1993 proposed rule and this final rule as 
    described in section IX. of this document.
    
    IX. Economic Impact
    
    A. Summary
    
        FDA has examined the impacts of the final rule under Executive 
    Order 12866 and the Regulatory Flexibility Act (Pub. L. 96-354). 
    Executive Order 12866 directs agencies to assess all costs and benefits 
    of available regulatory alternatives and, when regulation is necessary, 
    to select regulatory approaches that maximize net benefits (including 
    potential economic, environmental, public health and safety, and other 
    advantages; distributive impacts; and equity). The agency believes that 
    this final rule is consistent with the regulatory philosophy and 
    principles identified in the Executive Order. As explained in detail 
    below, FDA finds that this final rule has an estimated total annual 
    incremental cost of $81.9 million to the U.S. industry and an estimated 
    average annual benefit of from $180 million to $220 million in lives 
    saved and is economically significant under Executive Order 12866. 
    Consequently, the agency has completed this full regulatory flexibility 
    analysis which demonstrates that this rule is consistent with the 
    principles set forth in the Executive Order and the Regulatory 
    Flexibility Act, and also with the Unfunded Mandates Reform Act as 
    described in section VIII. of this document. This analysis, together 
    with the preamble published in the Federal Register and supporting 
    analysis and materials, constitutes a final regulatory flexibility 
    analysis. In addition, this document has been reviewed by OMB as an 
    economically significant regulatory action under Executive Order 12866.
        The detailed data for this analysis were developed by Eastern 
    Research Group, Inc. (ERG), under contract to FDA and their two 
    reports: ``Economic Analysis of the Proposed Revisions to the Good 
    Manufacturing Practices Regulation for Medical Devices,'' and 
    ``Addendum to the Final Report'' are on file at the Dockets Management 
    Branch (HFA-305), Food and Drug Administration, 12420 Parklawn Dr., rm. 
    1-23, Rockville, MD 20857.
        The objective of this rule is to reduce the number of fatalities 
    and injuries attributable to defective medical devices. FDA finds that 
    private market incentives do not adequately reduce the risk of design-
    related device failures because neither physicians nor consumers have 
    all of the information needed to make adequate judgments of product 
    quality and legal tort remedies are slow, inefficient, and extremely 
    costly.
        The changes to the CGMP regulation will require manufacturers to 
    extend
    
    [[Page 52644]]
    
    their quality systems to include several new areas, such as design and 
    purchasing, and to clarify or expand selected existing requirements. 
    Several of the changes to the regulation make it more consistent with 
    ISO 9001:1994 quality standards. The rule will affect all medical 
    device establishments engaged in the design, manufacture, contract 
    sterilization, and packaging of medical devices.
        This analysis presents the costs and benefits of the final CGMP 
    rule and reflects the differences between the proposed and final 
    regulation. The complete methodology and preliminary economic analysis 
    was presented in the November 1993 ERG report, ``Economic Analysis of 
    Proposed Revisions to the Good Manufacturing Practices Regulation for 
    Medical Devices''. While the proposed rule covered component 
    manufacturers, the cost of compliance for such manufacturers was 
    inadvertently omitted from the November 1993 ERG report. However, FDA 
    has decided not to cover component manufacturers, therefore most of the 
    preliminary analysis remains valid (e.g., estimates of labor and 
    resource requirements, level of compliance, and number of firms remain 
    the same for the final analysis, except where noted).
        Based on the ERG study, the total annual incremental costs to the 
    U.S. industry of the final CGMP regulation are estimated to be about 
    $81.9 million. These costs are more than offset, however, by benefits 
    to public health and by economic benefits to the medical device 
    industry. FDA estimates that the benefits to public health will include 
    36 to 44 fewer deaths and 484 to 677 fewer serious injuries per year, 
    which are attributed to design-related device failures. Studies on the 
    value of a statistical-life have reported estimates ranging from $1.6 
    million to $8.5 million.1 Assuming an economic value of $5 million 
    per fatality avoided, the monetary value of saving 36 to 44 lives each 
    year will be $180 to $220 million. Therefore, the value of the public 
    health benefits of preventing deaths alone easily exceeds the cost of 
    compliance even without estimating benefits from a reduced number of 
    serious injuries. Moreover, additional economic benefits to medical 
    device establishments will result from cost savings due to fewer 
    design-related product recalls, better product quality, and greater 
    productivity. In addition, medical device establishments exporting to 
    the EU will greatly benefit from the harmonization of the CGMP 
    regulation with the ISO 9001:1994 quality standards. Because the EU is 
    adopting ISO 9001:1994 as a basis for its medical device manufacturing 
    quality system, the harmonization of the two quality requirements will 
    eliminate the need for device manufacturers to maintain different 
    quality systems for each market.
    ---------------------------------------------------------------------------
    
         1 Fisher, A.; Chestnut, L.; and Violette, D. (1989). ``The 
    Value of Reducing Risks of Death: A Note on New Evidence.'' Journal 
    of Policy Analysis and Management, 8 (pp. 88-100).
    ---------------------------------------------------------------------------
    
        FDA supports the international harmonization of standards and 
    regulations governing medical devices and the eventual mutual 
    recognition of CGMP inspections between major device markets. While 
    full achievement of this goal is still in the future, the harmonization 
    of quality standards is an important first step.
        FDA believes in a step wise approach toward harmonization and 
    eventual mutual recognition. For CGMP inspections or Quality System 
    Conformity Assessments, these goals comprise four basic steps. First, 
    the harmonization of quality system requirements is a fundamental 
    building block of all future work in this area. FDA believes that by 
    working with the GHTF, specifically Study Group #3 of the GHTF, it has 
    developed a final rule that incorporates the harmonized quality system 
    requirements which are recognized around the world. Second is the 
    harmonization of regulatory auditing or compliance inspections. This 
    work is currently underway in the GHTF in Study Group #4, which has 
    developed one draft document entitled ``Guidelines For Regulatory 
    Auditing Quality Systems of Medical Device Manufacturers,'' expected to 
    be finalized in 1997. The third step is for harmonization of the 
    policy, interpretation, and regulatory consequences of noncompliance 
    with the quality system requirements in this rule and in counterpart 
    requirements of other countries. Underlying these activities is an 
    ongoing need for confidence building between the parties working 
    towards mutual recognition. FDA believes that this regulation will 
    provide a sound foundation for the goal of mutual recognition of 
    inspections, a goal that will benefit industry, as well as the agency. 
    The Health Industry Manufacturers Association has stated that 
    reciprocity for quality assurance inspections could save the medical 
    device industry millions of dollars as well as provide significant 
    savings to governments.2
    ---------------------------------------------------------------------------
    
         2 Gilmartin, R.V. (1992). ``The Benefits of Cooperation for 
    Industry and Regulators Alike: A Global Perspective.'' Presented at 
    the Third Annual Global Medical Device Conference, October 2.
    ---------------------------------------------------------------------------
    
        For individual establishments, the economic impact of the 
    regulation will depend on a number of factors, such as the level of 
    current compliance, the type of activities performed at the 
    establishment, and the nature of the product. On average, the smaller 
    establishments will bear a relatively greater economic burden.
    
    B. Industry Profile
    
        Firms in the medical device industry are heterogeneous. They vary 
    in size, product type, product and process technology, and rate of new 
    product introductions. There are over 7,000 medical device 
    establishments involved in the production of approximately 4,000 
    different types of devices (Table 1). Sixty-two percent of these 
    establishments are very small (fewer than 20 employees), while 27 
    percent are of medium-size (20 to 99 employees), 7 percent are large 
    (100 to 249 employees), and 4 percent are very large (250 or more 
    employees). These size categories were developed to reflect size 
    categories within the medical device industry and differ from the Small 
    Business Administration definition. Under the Small Business 
    Administration definition, over 98% of all establishments would be 
    small.
    
                          Table 1.--Distribution of Affected Establishments by Employment Size                      
    ----------------------------------------------------------------------------------------------------------------
                                                                               Employment size \2\                  
                                                            --------------------------------------------------------
               Type of establishment             Total \1\    Small  (1-  Medium (20-  Large  (100- Very large (250)  
    ----------------------------------------------------------------------------------------------------------------
    Design and Production Manufacturer........        5,415        3,323        1,414          415             265  
    Contract manufacturer.....................          419          257          109           32              20  
    Specification developer...................          541          352          162           27               0  
    Repacker/relabeler........................          828          538          248           41               0  
    
    [[Page 52645]]
    
                                                                                                                    
    Contract sterilizer.......................           34           22           10            2               0  
                                               ---------------------------------------------------------------------
          Total...............................        7,237        4,492        1,943          517             285  
    ----------------------------------------------------------------------------------------------------------------
    \1\ Based on data from FDA's Registration and Listing Branch, 1992, adjusted to reflect 13 percent not required 
      to register and 6 percent exempt from CGMP requirements.                                                      
    \2\ ERG (1993), Section 3.                                                                                      
    
    C. Comments to November, 1993 Proposed Changes to the CGMP Regulation
    
        A small percentage of the public comments on the November 1993 
    proposed regulation addressed the economic impact analysis. The 
    majority of these comments made very general, nonspecific observations 
    and therefore cannot be addressed directly. Many of these comments 
    stated that FDA underestimated the regulatory burden that the proposed 
    CGMP regulation would place on medical device manufacturers. Others 
    stated that their companies would expend more than the per 
    establishment estimated costs; some discussed the hiring of additional 
    personnel to address the compliance requirements.
        In developing the cost estimates for the 1993 proposal, ERG 
    attempted to describe the labor hours (and associated costs) needed to 
    achieve an acceptable minimum level of compliance with each 
    requirement. These estimates took into account the incremental labor 
    and capital resources that would be needed to progress from the 
    existing compliance level to the new level required by the proposal. 
    For individual establishments, the economic impact of the CGMP 
    regulation would depend on a number of factors, such as the level of 
    current compliance, the type of activities performed, and the nature of 
    the product. Not surprisingly, those establishments that currently 
    undertake relatively few of the activities to be required would incur 
    greater compliance costs than the averages presented.
        In the final rule, FDA has eliminated or modified several 
    requirements to give medical device establishments greater flexibility 
    in selecting compliance methods. In general, the words ``where 
    appropriate'' were added to many requirements to make them less 
    prescriptive and allow establishments to determine if or when they are 
    appropriate for their product. For example, in Sec. 820.65 
    Traceability, the final requirement allows the manufacturer to identify 
    which components require traceability. In addition, many procedures may 
    not need to be changed, only documented. To further minimize compliance 
    costs, FDA intends to provide additional guidance materials. The DSMA 
    currently offers guidance materials and regional seminars on CGMP 
    matters.
    1. Health Industry Manufacturers Association (HIMA)
        HIMA commented that FDA understated the costs for personnel 
    training, maintenance of new systems, documentation revisions, and 
    operational costs.
        ERG agrees that it did not fully address the initial training 
    requirements in the cost analysis for the proposed CGMP regulation. New 
    costs for initial training were included in the cost analysis for the 
    final CGMP regulation. However, the existing CGMP regulation requires 
    periodic training of personnel. Therefore no incremental costs for 
    periodic training were estimated.
        ERG did not change its cost estimate for quality system maintenance 
    and procedure revisions. Estimates were made for the incremental 
    compliance costs associated with an annual review of each new 
    procedure, but these procedures would be revised only sporadically and 
    probable estimates of their future costs would be small and could not 
    be reasonably quantified.
        ERG recognized that companies will incur incremental costs to use 
    new procedures. Although a separate estimate of these operational costs 
    was not made, they were incorporated into the estimates of the 
    individual requirements where applicable.
    2. Other General Comments
        Some manufacturers of low-risk devices and some that have never 
    experienced a product recall or MDR event questioned the merit and 
    benefits of applying design controls to all products. In the proposed 
    and final CGMP regulation, FDA exempted almost all class I devices 
    because the public health benefits gained did not exceed the costs of 
    implementation. However, FDA believes that all class II and III devices 
    should be covered because their failure could adversely affect public 
    health. Even firms with excellent past records put their consumers at 
    future risk if their design systems are inadequate. ERG estimates that 
    strict compliance to the final CGMP regulation will avert about 43 
    deaths and over 600 serious injuries per year. In addition, the 
    literature on quality systems consistently states that firms 
    implementing such systems, which begin with design controls, report 
    cost savings in the long-run.
        A number of comments argued that the proposed CGMP regulation would 
    slow product innovation and increase health care costs. FDA believes 
    that the gains from improvements in quality control and greater 
    efficiencies will lessen the impact on both innovation and health care 
    costs and will not lower the innovation rate for products with 
    significant medical benefit. Manufacturers will also avoid the costs of 
    most design-related medical device recalls. ERG estimated that design-
    related recalls cost industry approximately $40 million per year. 
    Health care spending overall will also decrease as deaths, injuries and 
    malfunctions from medical device failures decrease.
        Some comments suggested that the proposed CGMP regulation would 
    hurt the domestic medical device industry's competitiveness and 
    encourage companies to move their operations to foreign countries. FDA 
    has sought to harmonize the final CGMP regulation with ISO 9001:1994 
    and ISO/CD 13485. Some comments had stated they would like to see even 
    greater harmonization in the final regulation. The harmonization of 
    regulatory requirements will benefit medical device establishments 
    because they will be able to maintain a single regulatory compliance 
    program. The harmonization of CGMP requirements is also a first step in 
    developing mutual recognition agreements between U.S. and foreign 
    governments. An FDA sponsored survey of innovative medical
    
    [[Page 52646]]
    
    device companies found that nearly 65 percent of them sold their 
    products outside the United States, including 40 percent of the small 
    and 70 percent of the medium-sized companies.3 Thus, a majority of 
    firms should benefit from harmonization efforts. Since foreign firms 
    exporting their products to the United States must comply with the U.S. 
    CGMP regulation, they will incur essentially the same incremental costs 
    to comply with the final CGMP regulation as domestic establishments.
    ---------------------------------------------------------------------------
    
        \3\ ERG (1994). FDA Survey of Establishments Introducing New 
    Medical Devices. (Task Order 3, Contract No. 223-91-8100.)
    ---------------------------------------------------------------------------
    
    3. Small Business Concerns
        Some comments representing small businesses were concerned about 
    the increase in procedural and documentation requirements. The 
    procedures and paperwork requirements will be simpler for small medical 
    device establishments relative to larger firms. Further, small 
    businesses can reduce compliance costs by using FDA guidance and 
    training materials, industry-generated guidance, and other technical 
    assistance that is available. FDA is preparing an extensive range of 
    technical support regarding the final CGMP regulation, including 
    guidance documents, workshops, and other materials and presentations.
        Several small businesses argued that the regulatory costs fall 
    disproportionately on small business, hindering industry growth. The 
    regulatory requirements apply equally to whoever is designing and 
    developing new devices. However, the vast majority of firms are small 
    and medium in size and these firms are least likely to have such design 
    control procedures already in place. As a result, their incremental 
    costs may be higher. Nevertheless, because procedures reflect the 
    complexity of the processes they guide, small and medium-sized 
    establishments should incur proportionately lower gross compliance 
    costs for those activities than larger establishments.
    4. Section 820.22  Quality audit
        Some comments believed that requiring quality audits to be 
    performed by individuals without direct responsibility for the matters 
    being audited poses a severe burden for small business. This 
    requirement is already present in the original CGMP regulation and thus 
    was not addressed in the economic analysis of the final regulation.
    5. Section 820.25  Personnel
        Comments stated that the requirement to maintain files on 
    consultants was onerous and interfered with manufacturers' selection 
    processes. FDA modified this requirement and moved it to Sec. 820.50 
    Purchasing, in the final CGMP regulation. Companies will now be 
    required to verify that consultants meet specified requirements and 
    define the type and extent of control they will exercise over them. The 
    incremental compliance costs were judged to be negligible.
    6. Section 820.30  Design control
        Comments believed that the requirement stipulating that devices be 
    sampled from three production runs before a device is released for 
    routine distribution was too prescriptive and burdensome. FDA has 
    modified the requirement in the final rule to require design validation 
    of initial production units, lots, or batches, or their equivalent. 
    This modification should give manufacturers greater flexibility in 
    implementing this requirement.
        Some comments from small businesses were critical of the 
    requirement that independent personnel perform design reviews and 
    stated that they will have to hire outside engineers for this task. In 
    the final rule FDA allows greater flexibility and states that the 
    independent personnel can be individual(s) who do not have direct 
    responsibility for the design stage being reviewed. Thus, staff 
    personnel (including engineers working on other components of the 
    device and nonengineering personnel) can perform design reviews.
    7. Section 820.40  Document control
        Some comments believed that the cost of implementing documentation 
    systems and other paperwork was understated. However, ERG's estimates 
    included the incremental compliance costs for formalizing a written 
    document control procedure and ERG considered paperwork requirements in 
    its estimation. The final rule also extends document control 
    requirements to the design phase and cost estimates for these 
    requirements were added to the economic assessment.
        Most companies consider document control procedures to be essential 
    and have realized some benefits from such procedures, typically in the 
    form of efficiency gains and avoided documentation mixups. These 
    potential benefits were not quantified.
    8. Section 820.50  Purchasing control
        Comments questioned the need to establish the quality of materials 
    purchased from long-established suppliers or from new suppliers of 
    small quantities of components. Historical records, however, even for 
    suppliers of small quantities, can be used to assess a supplier's 
    quality. Supplier audits are not mandated in the CGMP regulation, but 
    may be a useful tool in assessing a supplier's capabilities. Cost 
    estimates for auditing from one- half to four new suppliers per year 
    for small to very large establishments were included in the economic 
    assessment.
    9. Section 820.80  Receiving, in-process, and finished device 
    acceptance
        One comment believed that requiring manufacturers to retain the 
    quantitative results of testing was excessive. The final rule 
    stipulates that ``the results'' of acceptance activities are to be 
    recorded, but does not specify that all quantitative results must be 
    recorded. Because this requirement is consistent with current industry 
    practices, incremental costs were not assigned to this section.
    10. Section 820.90  Nonconforming product
        Comments noted that identifying a product as ``reprocessed'' has a 
    negative impact on sales. (FDA now uses the term ``reworked''.) This 
    language was revised in the final rule to clarify that reworked devices 
    need to be identified as such at the manufacturing facility to avoid 
    mixups. No costs were estimated for this requirement.
    
    D. Industry Costs
    
        ERG estimated the total annual incremental cost of the final rule 
    at $81.9 million. This includes $9.5 million in one-time costs that 
    were annualized over 5 years at a 10 percent discount rate. Table 2 
    lists the most costly of the new requirements.
        Costs were based on the incremental tasks each manufacturer must 
    perform to achieve compliance. ERG retained most of the methodology and 
    data from the proposed rule to estimate the costs of the final rule. 
    Where applicable, costs were estimated for additional or changed final 
    requirements. Also, the distribution of costs across establishment size 
    was modified to reflect new information on the rate of product 
    innovation.4 The rates of innovation per year used for this 
    analysis are: 0.4 percent for small, 1.3 percent for medium-sized, 2.6 
    percent for large, and 6.5 percent for very large establishments.
    ---------------------------------------------------------------------------
    
         4 ERG (1994). FDA Survey of Establishments Introducing New 
    Medical Devices. (Task Order 3, Contract No. 223-91-8100).
    
    [[Page 52647]]
    
    
    
                           Table 2.--Total Compliance Costs, by Most Costly Incremental Tasks                       
                                                      [$ millions]                                                  
    ----------------------------------------------------------------------------------------------------------------
                                                                                        Annual                      
                         Incremental tasks                         One-time   --------------------------    Total   
                                                                  annualized      Labor       Nonlabor    annualized
    -----------------------------------------------------------------\1\--------------------------------------------
    Design Controls:                                                                                                
        Design Verification...................................            NA          18.2         27.4         45.6
        Design Review.........................................            NA           6.2           NA          6.2
        Design Changes........................................            NA           4.0           NA          4.0
        Design and Development Planning.......................            NA           1.2           NA          1.2
    Other:                                                                                                          
        Quality Audit.........................................           0.5           4.7           NA          5.2
        Evaluation of Suppliers and Contractors...............           0.6           1.9          0.9          3.4
        Management Review.....................................            NA           2.2           NA          2.2
        Purchasing Data.......................................            NA           1.1           NA          1.1
    All Remaining.............................................           8.4           4.6          0.0         13.0
                                                               -----------------------------------------------------
            Total for Final Regulation........................           9.5          44.1         28.3        81.9 
    ----------------------------------------------------------------------------------------------------------------
    \1\ One-time costs annualized over 5 years at discount rate of 10 percent.                                      
    NA=Not Applicable.                                                                                              
    Note: Totals may not add due to rounding.                                                                       
    Source: ERG (1996), Section 4.                                                                                  
    
        The great majority of costs for all size establishments will be 
    associated with the establishment of design controls for new products. 
    Therefore, the more innovative establishments will experience greater 
    compliance costs than the less innovative establishments. The estimated 
    annual design control costs total $57.5 million, which represents 70 
    percent of the total annual incremental costs of compliance. The most 
    costly task within the design control category is design verification 
    ($45.6 million), which includes design validation. Other costly tasks 
    are design review ($6.2 million), which encompasses conducting and 
    documenting design reviews; design changes ($4.0 million), which 
    includes documenting and maintaining design change procedures; and 
    design and development planning ($1.2 million), which includes 
    documenting and maintaining plans for device design and development. 
    The requirement for extending the quality system audit ($5.2 million) 
    and the evaluation of suppliers and contractors ($3.4 million) are also 
    relatively high cost items.
        The estimated total cost of compliance for the final rule ($81.9 
    million) is $2.6 million less than the estimated cost of the November 
    1993 proposed rule ($84.5 million). Some cost increases were due to 
    added requirements for increased documentation. However, these cost 
    increases were offset partly by a decrease of $0.5 million from the 
    modification of some requirements (e.g. Secs. 820.65 Traceability and 
    820.160 Distribution). The remaining changes resulted from changes in 
    assumptions or new information about cost and compliance rates in 
    design control and supplier audits and from new information regarding 
    product innovation rates across establishment size.
        The projected average cost per establishment (see Table 3) varies 
    substantially across industry sectors and establishment size 
    categories. As expected, the average incremental costs are largest for 
    establishments that design medical devices: design and production 
    manufacturers and specification developers. For these two sectors, the 
    average per establishment costs are $15,994 for design and production 
    manufacturers and $14,767 for specification developers. Actual per 
    establishment costs will vary substantially depending on the product 
    type, design complexity, innovation rate, and level of design control 
    currently in place. The average incremental costs for the other three 
    sectors are significantly lower: $3,554 for contract manufacturer, 
    $1,995 for repacker/relabeler, and $2,040 for contract sterilizer.
    
                     Table 3.--Average Total Annualized \1\ Costs Per Establishment by Type and Size                
                                                        [Dollars]                                                   
    ----------------------------------------------------------------------------------------------------------------
                                                          Medium  (20-  Large  (100-  Very large  (250)         All    
    ----------------------------------------------------------------------------------------------------------------
    Design and Production Manufacturer....        11,085        25,800        22,748          12,258          15,994
    Specification Developer...............         9,927        24,052        20,583              NA          14,767
    Contract Manufacturer.................         2,357         4,027         5,802          10,678           3,554
    Repacker/Relabeler....................         1,471         2,588         3,969              NA           1,995
    Contract Sterilizer...................         1,491         2,621         3,999              NA           2,400
    ----------------------------------------------------------------------------------------------------------------
    \1\ One-time costs annualized over 5 years at a discount rate of 10 percent.                                    
    NA=Not Applicable.                                                                                              
    Source: ERG (1996), Section 6.                                                                                  
    
        Because average current compliance rates tend to vary directly with 
    establishment size and there are relatively few large and very large 
    establishments (7 and 4 percent of all medical device establishments, 
    respectively), the largest share of the costs are incurred by small 
    establishments, $35.2 million (43
    
    [[Page 52648]]
    
    percent) and medium-size establishments, $34.5 million (42 percent), 
    while the smallest share is incurred by very large establishments, $3.4 
    million (4 percent) (see Table 4).
    
                                    Table 4.--Total Annualized Costs by Size Category                               
                                                      [$ millions]                                                  
    ----------------------------------------------------------------------------------------------------------------
                                                                                        Annual                      
                        Establishment size                         One-time   --------------------------    Total   
                                                                  annualized      Labor       Nonlabor    annualized
    -----------------------------------------------------------------\1\--------------------------------------------
    Small (1-19)..............................................           4.9          18.2         12.1         35.2
    Medium (20-99)............................................           3.0          18.2         13.3         34.5
    Large (100-249)...........................................           1.0           5.1          2.8          8.8
    Very large (250)...............................           0.7           2.6          0.1          3.4
    All establishments........................................           9.5          44.1         28.3         81.9
    ----------------------------------------------------------------------------------------------------------------
    \1\ One-time costs annualized over five years at discount rate of 10 percent.                                   
    Note: Totals may not add due to rounding.                                                                       
    Source: ERG (1996), Section 4.                                                                                  
    
    E. Benefits From Proposed Changes to the CGMP Regulation
    
        ERG used the methodology and data from the proposed rule to 
    estimate the benefits of the final CGMP regulation. Adjustments to the 
    number and distribution of MDR's were made based on updated numbers of 
    closed cases. Also, more reliable estimates of industry savings from 
    avoided design-related recalls were incorporated.
        The changes to the CGMP regulation will provide public health 
    benefits to medical device users and economic benefits to the medical 
    device industry. Based on its review of medical device recalls over the 
    4-year period 1988 to 1991, FDA has estimated that 30 percent of all 
    medical device product recalls are due to inadequate design controls. 
    It is extremely difficult to judge how many of these recalls could 
    reasonably have been avoided, but ERG judged that a majority would have 
    been prevented if manufacturers had fully implemented the CGMP design 
    control requirements.
    1. Public Health Benefits
        ERG used the MDR database to estimate the public health benefits of 
    the final CGMP regulation. There were over 41,600 MDR's submitted to 
    FDA in 1991; 97 percent of these MDR's are closed (i.e., a review of 
    the case is completed). Of these closed cases, FDA determined that 9.3 
    percent of the fatalities and 12.4 percent of the serious injuries were 
    due to device failures. The bulk of the remaining incidents were due to 
    user problems, but also include cases where cause could not be clearly 
    established. To estimate the total number of deaths and serious 
    injuries for 1991 by cause, the 1988-1991 averages of device recalls 
    were used. To estimate the number of deaths and serious injuries due to 
    design-related causes, ERG assumed that the percent of MDR's that were 
    design-related was the same as that for recalls (30 percent).5 
    Based on these assumptions, medical devices contributed to an estimated 
    49 fatalities and 663 serious injuries in 1991 due to design-related 
    problems in class II and III devices (see Table 5). To correct for the 
    substantial under reporting of MDR's, ERG made an upward adjustment in 
    the number of MDR's of 20 percent for fatalities and 40 percent for 
    serious injuries. The number of estimated fatalities adjusted for 
    underreporting of MDR's would be 59, with 929 serious injuries.
    ---------------------------------------------------------------------------
    
        \5\ There is no code in the MDR database to identify design-
    related events.
    
              Table 5.--Number of Design-Related Reports and Estimated Avoided Deaths and Serious Injuries          
    ----------------------------------------------------------------------------------------------------------------
                                                          Fatalities                       Serious Injuries         
                                             -----------------------------------------------------------------------
                                               Class II    Class III     Total     Class II    Class III     Total  
    ----------------------------------------------------------------------------------------------------------------
    Number in 1991..........................         555         475       1,030       4,391      11,794      16,185
    Device-related..........................         105          59         164         330       1,881       2,211
    Design-related \1\......................          32          18          49          99         564         663
    Number avoided..........................          23          13          36          72         412         484
    Adjusted number of design-related MDR's                                                                         
     \2\....................................          38          21          59         139         790         929
    Adjusted Number avoided.................          28          15          43         101         576         677
    ----------------------------------------------------------------------------------------------------------------
    \1\ Assumes 30 percent of device-related MDR's are design-related, based on FDA recall data.                    
    \2\ Total number of fatalities and injuries increased by 20 and 40 percent, respectively, to adjust for under-  
      reporting.                                                                                                    
    Source: ERG (1996), Section 5.                                                                                  
    
        To develop an approximate idea of the preventability of these 
    incidents, ERG convened a panel of industrial engineers and regulatory 
    specialists with extensive experience in the design of medical devices. 
    The panel examined a random sample of 100 design-related medical device 
    recalls and judged whether implementation of design controls could have 
    prevented the recall. ERG found that the expected value of their 
    judgments implied that proper design controls would have prevented 
    about 73 percent of these recalls. Assuming the same preventability 
    ratio for design-related MDR events, ERG calculated that the proposal 
    would prevent about 36 to 43 deaths and 484 to 677 serious injuries per 
    year, depending on the degree of MDR underreporting.
        To verify the reasonableness of the estimates, FDA examined an 
    alternative method of estimating the number of
    
    [[Page 52649]]
    
    fatalities caused by design-related failures. For this calculation, 3 
    years of design-related recalls were assumed linked to MDR fatalities 
    that occurred for these devices 1 year before or 3 months after the 
    date of the recall. This approach, which provides a conservative 
    estimate because not all relevant fatalities and subsequent MDR's would 
    occur during this limited time period, found that about 60 deaths per 
    year were due to design-related device failures. If 73 percent of such 
    incidents could be avoided through compliance with the proposed CGMP 
    regulation, 44 deaths per year would be prevented.
        These estimates of the public health benefits from fewer design-
    related deaths and serious injuries represent FDA's best projections, 
    given the limitations and uncertainties of the data and assumptions. 
    The above numbers, however, do not capture the quality of life losses 
    to patients who experience less severe injuries than those reported in 
    MDR's, who experience anxiety as a result of treatment with an 
    unreliable medical device, or who experience inconvenience and 
    additional medical costs because of device failure.
        Medical device malfunctions are substantially more numerous than 
    deaths or injuries from device failures and also represent a cost to 
    society. Malfunctions represent a loss of product and an inconvenience 
    to users and/or patients. Additionally, medical device malfunctions 
    burden medical personnel with additional tasks, such as repeating 
    treatments, replacing devices, returning and seeking reimbursement for 
    failed devices, and providing reports on the circumstances of medical 
    device failures. No attempt was made to quantify these additional 
    costs.
    2. Industry Benefits
        The medical device industry would gain substantial economic 
    benefits from the proposed changes to the CGMP regulation in three 
    ways: Cost savings from fewer recalls, productivity gains from improved 
    designs, and efficiency gains for export-oriented manufacturers who 
    would now need to comply with only one set of quality standards.
        An average of 359 medical device recall events per year were 
    reported to FDA over the period 1988 to 1991. As stated above, FDA 
    estimates that design-related deficiencies contributed to 30 percent of 
    those recall events annually. Applying the 73 percent recall 
    preventability factor, ERG projects that there would be 67 fewer 
    recalls of class II and III devices each year under the final CGMP 
    regulation (see Table 6). Based on data from a recent survey of recall 
    costs, 67 fewer recalls implies that the industry would avoid roughly 
    $29 million worth of recall expenses per year by complying with the 
    final CGMP regulation.6
    ---------------------------------------------------------------------------
    
         6 Design-related medical device recalls cost the industry 
    approximately $40 million annually. (Eastern Research Group, Inc. 
    (1994). FDA Survey of Medical Device Recall Costs. (Task Order 3, 
    Contract Number 223-91-8100).
    
      Table 6.--Number of Avoided Design-Related Recall Events by Class of  
                                     Device                                 
                                [FY 1988-FY 1991]                           
    ------------------------------------------------------------------------
                                                         Average   Number of
                                                        number of   avoided 
                                                         design-    design- 
                       Device class                      related    related 
                                                          recall     recall 
                                                        events\1\  events\2\
    ------------------------------------------------------------------------
    I.................................................         15         NA
    II................................................         80         58
    III...............................................         12          9
                                                       ---------------------
      All Devices.....................................        107        67 
    ------------------------------------------------------------------------
    \1\ Office of Compliance and Surveillance, CDRH.                        
    \2\ ERG estimates based on random sample of 100 design-related recalls. 
    Source: ERG (1996), Section 5.                                          
    
        ERG also found that the design control requirements in the final 
    CGMP regulation would require manufacturers to integrate their design 
    and production operations and that most industry experts believe that 
    this change would lead to better quality products, more efficient 
    engineering, lower manufacturing costs, and reduced product development 
    time. These savings, however, could not be quantified.
        Still another benefit of the revised regulation relates to the 
    harmonization of the final CGMP regulation with the ISO 9001:1994 
    international standard. This change would especially benefit export-
    oriented establishments, because they would need to meet only one set 
    of quality standards. ERG could not derive quantitative measures of 
    this benefit. However, 65 percent of innovative medical device 
    companies export their products, thus a majority should benefit from 
    harmonization of CGMP regulation between major trading partners.7
    ---------------------------------------------------------------------------
    
         7 ERG (1994). FDA Survey of Establishments Introducing New 
    Medical Devices. (Task Order 3, Contract No. 223-91-8100.)
    ---------------------------------------------------------------------------
    
    F. Economic and Small Business Impact
    
        The ability of medical device establishments to pass on the added 
    cost of the final regulation will determine the economic impact to the 
    industry. The diversity of medical devices precludes any easy 
    characterization of their product markets. Under the current medical 
    care system, however, the demand for many medical devices tends to be 
    price inelastic because they are often prescribed by physicians and 
    frequently paid for by third parties. Thus, small price increases have 
    not typically prompted significant declines in industry sales. 
    Nonetheless, competitive pressures have increased substantially under 
    new health care cost-containment measures. Therefore, to examine the 
    potential effect of the costs of compliance on the industry's 
    competitive structure, ERG calculated the maximum impact on industry 
    average prices and products, using extreme scenarios. Financial data 
    characterizing the scope of FDA-regulated medical device establishments 
    are not available. To make estimates of the regulatory impact on price 
    and profits, ERG used a combination of census and Dun and Bradstreet 
    data (see ERG (1993) for methodology). ERG assumed that the firms 
    characterized in these data sources had the same size and product 
    distribution, and introduced new products at the same rate as the 
    population of FDA-regulated establishments. While the validity of these 
    assumptions is uncertain, it was the only data available to measure 
    regulatory impact. ERG presents two extreme scenarios, the first 
    reflects the magnitude of the potential impact on product prices if all 
    costs were passed forward. The second demonstrates the maximum drop in 
    profits if no costs were passed forward. In reality, some combination 
    of these scenarios will occur.
        Based on the assumption that all costs of compliance are passed 
    through to the end user, with no loss in sales and no offset for 
    avoided recalls or other industry productivity gains, ERG found that 
    the average increase in the price of medical devices would be less than 
    0.13 percent. Estimated price increases ranged from 0.04 percent for X-
    Ray Apparatus and Tubes (SIC 3844) to 0.34 percent for Dental Equipment 
    and Supplies (SIC 3843) (see Table 7). The maximum price increase was 
    calculated using aggregate compliance costs as a percentage of the 
    value of shipments. The price increases calculated by size of 
    establishment suggest that small establishments will be under greater 
    pressure to increase prices. The cost of compliance represented an 
    average of 1.36 percent of the value of shipments for small 
    establishments and 0.01 percent for very large establishments. These 
    differences in impacts by size reflect the finding that small
    
    [[Page 52650]]
    
    establishments have lower current compliance than large establishments.
        To estimate the potential impact of compliance costs on medical 
    device industry profits, ERG calculated after-tax compliance costs as a 
    percentage of after-tax income for each medical device SIC (see Table 
    7). Again, no adjustments were made for avoided recalls or expected 
    productivity gains. If manufacturers have no ability to increase prices 
    to offset the increase in compliance costs, this estimate represents an 
    upper bound of the potential effect on entity income. Under these 
    circumstances, the medical device sectors could incur reductions in 
    income ranging from about 0.81 percent (SIC 3845, Electromedical 
    Equipment) to about 4.27 percent (SIC 3843, Dental Equipment and 
    Supplies). ERG concluded that such impacts may affect some 
    establishments' decisions to develop new products where expected 
    profits are marginal or highly uncertain, but judged that the level of 
    incremental costs imposed by this regulation would not substantially 
    lower the innovation rate especially for products with significant 
    medical benefits.
    
     Table 7.--Maximum Potential Impact on Price or Profits by Industry and 
                                 Employment Size                            
    ------------------------------------------------------------------------
                                                         Total              
                                                      annualized   After-tax
                                                      compliance  compliance
                                                      costs as a  costs as a
                                                      percentage  percentage
                                                          of       of after-
                                                       shipments  tax income
    ------------------------------------------------------------------------
    Industry:                                                               
      3841 Surgical and medical instruments.........        0.12        2.00
      3842 Surgical appliance and supplies..........        0.14        1.78
      3843 Dental equipment and supplies............        0.34        4.27
      3844 x-ray apparatus and tubes................        0.04        0.88
      3845 Electromedical equipment.................        0.05        0.81
      3851 Ophthalmic goods.........................        0.24        3.54
      All...........................................        0.13        1.87
    Establishment size:                                                     
      Small (1-19)..................................        1.36          NA
      Medium (20-99)................................        0.35          NA
      Large (100-249)...............................        0.09          NA
      Very large (250)...................        0.01          NA
      All...........................................        0.13          NA
    ------------------------------------------------------------------------
    NA = not available.                                                     
    Source: ERG (1996), Section 6.                                          
    
        The Regulatory Flexibility Act requires agencies to analyze 
    regulatory options that would minimize any significant impact of a rule 
    on small entities. This section together with other discussions in this 
    preamble and supporting analysis and materials constitute the agency's 
    regulatory flexibility analysis. A description of the projected 
    reporting, recordkeeping, and compliance requirements including the 
    type of professional skills required is included in the ERG economic 
    analysis reports that are referenced above and on file at the Dockets 
    Management Branch (address above). In accordance with the Regulatory 
    Flexibility Act, FDA has considered the effect of this action on small 
    businesses and has determined that there will be a significant impact 
    on a substantial number of small businesses. Almost all medical device 
    establishments are classified as small under the Small Business 
    Administrations definition of size.8 The incremental costs are 
    greatest for establishments that design medical devices and that 
    currently have lower levels of compliance with the new design control 
    requirements. These requirements account for 70 percent of the total 
    incremental costs of the final rule but affect only design and 
    production manufacturers and specification developers (82 percent of 
    the total affected establishments). Other sectors of the industry will 
    incur substantially lower costs (see Table 3).
    ---------------------------------------------------------------------------
    
         8 The Small Business Administrations definition is by the 
    employment size at the company level. Detailed demographic and 
    financial data is not available by company size, therefore FDA used 
    establishment data. FDA does not know the impact on companies.
    ---------------------------------------------------------------------------
    
        The actual added cost per establishment will vary by the 
    establishment's current level of compliance, complexity of product 
    design, product type, and rate of product innovation. As indicated in 
    Table 3, the average medium-size and large manufacturers of devices 
    will incur greater compliance costs ($25,800 and $22,748 per 
    establishment, respectively) relative to small and very large 
    establishments ($11,085 and $12,258, respectively). However, the 
    potential impact on product price (measured as a percent of the value 
    of shipments) is greatest for small (1.36 percent) and medium-size 
    (0.35 percent) establishments. Large and very large establishments will 
    incur only a 0.09 percent and 0.01 percent increase, respectively, due 
    to much larger values of shipments and higher rates of compliance with 
    the final rule. Smaller establishments producing differentiated 
    products or marketing to niche markets may not be at a disadvantage 
    because of their ability to pass on the added cost of compliance. 
    However, those smaller establishments that compete with larger 
    establishments based on price alone would suffer a drop in profits if 
    they currently operate at lower levels of compliance than their 
    competitors.
        FDA believes that actual per establishment compliance costs will be 
    lower than estimated for the following reasons: First, the final CGMP 
    regulation closely parallels the ISO 9001:1994 quality standards, which 
    have been adopted as the quality standard for the EU and are becoming 
    the international quality standards for medical devices. Close to 65 
    percent of domestic medical device manufacturers export their products 
    and generate approximately one-third of their sales from exports.9 
    Compliance with the quality control requirements is necessary for firms 
    to maintain international competitiveness and in fact many U.S. medical 
    device manufacturers have become ISO certified since the 1993 
    publication of the proposed CGMP regulation and the EU implementation 
    of unified regulatory requirements.
    ---------------------------------------------------------------------------
    
         9 ERG (1994). FDA Survey of Establishments Introducing New 
    Medical Devices. (Task Order 3, Contract No. 223-91-8100.)
    ---------------------------------------------------------------------------
    
        Second, the FDA has extended the effective date of the final rule 
    to June 1, 1997, and has chosen not to take regulatory action for an 
    additional year on the design control requirements. This revised 
    effective date will also reduce the cost of implementation estimated 
    for the 1993 proposal where the proposed effective date was only 180 
    days after date of publication. The extension will give manufacturers a 
    longer time to implement the new requirements, allowing the costs to be 
    spread over almost a 2-year period as compared to 180 days. June 1998 
    coincides with the implementation of the EU's Inactive Medical Device 
    Directive. Therefore, the economic impact of complying with the new 
    quality system regulation will be shared with the economic impact of 
    complying with the new EU Medical Device Directive for any manufacturer 
    who also produces devices for sale in the EU, lessening the direct 
    impact of the new quality system regulation.
        Third, ERG estimates of the number of labor hours needed for design 
    controls assume that many establishments have little or no formal 
    system in place. Once an establishment has developed a system, minor 
    modifications to an establishment's existing product (for which many 
    510(k) applications and PMA supplements are submitted) may be less 
    costly than ERG assumed.
    
    [[Page 52651]]
    
        Finally, cost estimates assume that establishments will use in-
    house expertise or hired consultants for all compliance activities. In 
    fact, FDA and trade publications have disseminated much of the 
    information that would be needed by the firms. FDA has taken many steps 
    specifically to assist small businesses in complying with this final 
    rule. The two stage implementation of the regulation was a concerted 
    effort to reduce the regulatory burden on small businesses. Stage 1 was 
    set up to be a 1 year training and cooperative phase for the entire 
    medical device community. FDA and industry would be participating in a 
    number of cooperative efforts as well as joint training exercises. Most 
    importantly, FDA would be evaluating design controls and providing 
    industry with feedback in the nature of a report. During this time, to 
    truly allow it to be a learning experience for both the device 
    manufacturers and the FDA investigators, there would be no regulatory 
    actions taken as a result of these evaluations and reports. The biggest 
    benefactor of the two stage implementation would clearly be small 
    businesses.
        Further, several guidances have been prepared by FDA for this 
    regulation as a whole, as well as on subject matters that are 
    significant in this final rule. FDA plans to release the following 
    three guidances within 60 days after the final rule is published: (1) 
    DSMA's ``Medical Device Quality Systems Manual: A Small Entity 
    Compliance Guide,'' which includes discussion on the entire regulation 
    plus multiple examples of procedures and forms that can be adopted and 
    modified by manufacturers; (2) ``Design Control Guidance For Medical 
    Device Manufacturers,'' which is intended to assist manufacturers in 
    understanding the intent of the design control requirements. Assistance 
    is provided by interpreting the language of the regulation and 
    explaining the underlying concepts in practical terms; and (3) ``Do It 
    By Design: An Introduction to Human Factors in Medical Devices,'' which 
    contains background information about human factors as a discipline, 
    descriptions and illustrations of device problems, and a discussion of 
    human factors principles and methods as a part of the design control 
    system. FDA also plans to release the following guidances after 
    publication of this final rule: (1) A guidance on ``Validation,'' which 
    will include discussions on design validation, computer validation, and 
    process validation; and (2) a draft of the ``Design Control 
    Inspectional Strategy,'' which will be the questions that FDA 
    investigators will be asking when assuring compliance with the design 
    control requirements.
        FDA is also prepared to release shortly after publication of this 
    final rule a 4 hour series of videotapes discussing the Quality System 
    Regulation. The videotapes will also be accompanied by a guidebook 
    entitled ``The FDA and World Wide Quality System Requirements Guidebook 
    For Medical Devices.'' This guidebook will contain the entire Quality 
    System Regulation from FDA, the entire text of ISO 9001:1994, FDA 
    guidance from the regulation's preamble, and guidance on quality 
    systems from the GHTF.
        FDA has also tentatively scheduled two teleconferences. The first 
    teleconference, which would be to discuss the Quality System Regulation 
    and answer questions that have come up from manufacturers beginning to 
    implement the regulation, is tentatively scheduled for December 1996. A 
    second teleconference is tentatively scheduled for April/May of 1997 
    and will specifically address design controls and the final Design 
    Control Inspectional Strategy. FDA is also exploring the possibility of 
    conducting regional workshops in May of 1997 to further discuss the 
    design control requirements and their implementation.
        In addition to these activities, FDA and DSMA will continue to 
    provide guidance and workshops that can help small business with their 
    compliance activities, and will continue to participate in industry 
    association workshops, conferences, and meetings. While all of the 
    above-mentioned activities will be available to all manufacturers, 
    small manufacturers will benefit the most from these FDA activities 
    without having to pay substantial costs, as most of the guidance and 
    written material will be available on the world wide web, and the 
    teleconferences and other workshops sponsored or cosponsored by FDA 
    will be of nominal cost.
        Finally, as described elsewhere in this preamble, FDA intends to 
    conduct a midcourse review of the new design control requirements 
    during the transition year (June 1997 to June 1998). Specifically, the 
    results of the first several months of design control inspections will 
    be reviewed by early 1998, and any midcourse adjustments to the 
    inspectional strategy will be instituted and made public by the Spring 
    of 1998. Also during this midcourse review FDA will evaluate the 
    information gathered at that point and determine if the design control 
    requirements as written in this final rule are appropriate to obtain 
    the goals expressed in this preamble. Any necessary adjustments or 
    proposed revisions will be published in the Federal Register and 
    comments will be solicited as necessary during the spring of 1998. This 
    implementation strategy is responsive to requests by industry for FDA 
    to harmonize the quality system regulation's implementation with the 
    mandatory date for implementation of the EU's Medical Device Directive, 
    which is June 1998. However, if during the midcourse review of stage 
    one it is determined that the industry and/or FDA needs more time to 
    fully implement the design control requirements, FDA will publish that 
    decision in the Spring of 1998 prior to the June 1, 1998, regulatory 
    implementation date.
        Small businesses will also benefit in that FDA considered but 
    rejected applying design requirements to all class I devices, because 
    the added benefits to public health were not great enough to offset the 
    increased burden on industry. Two requirements were eliminated or 
    modified in the final rule that decreased the burden on industry: The 
    applicability of the CGMP regulation to component suppliers was 
    removed, and Sec. 820.65 Traceability was limited to traceability of 
    components where necessary to assure the protection of public health. 
    These changes will particularly aid small businesses. In addition, 
    revisions were made to many requirements in the final rule to make it 
    less prescriptive and to allow establishments greater flexibility in 
    implementing the requirements. Cost savings from these changes were not 
    estimated.
        In addition, revisions were made to many requirements in the final 
    rule to make it less comprehensive in scope, less prescriptive and to 
    allow establishments greater flexibility in implementing the 
    requirements. Cost savings from these changes were not estimated. Based 
    on the above, the agency has determined that the current rule 
    represents the least burdensome alternative that meets the public 
    health goal of reducing deaths and serious injuries attributable to 
    defective medical devices.
        In summary, FDA concludes that the estimated $81.9 million annual 
    incremental cost to comply with the final CGMP regulation is likely an 
    upward bound figure and that it would be substantially offset by 
    significant savings from avoided recalls and more importantly, the 
    avoidance of deaths and serious injuries due to design-related device 
    failures. FDA's estimate of public health benefits includes the
    
    [[Page 52652]]
    
    prevention of 36 to 44 deaths and 484 to 677 serious injuries annually. 
    Establishing design controls will also result in better designed and 
    higher quality devices and fewer device failures. This quality 
    improvement will in turn reduce the inconvenience and expense of 
    repetitive treatments or diagnoses. The agency also believes the actual 
    cost to comply with the final rule will be lower than estimated because 
    the industry compliance baselines used to estimate costs are from 1993. 
    Since that time, market pressures have induced many firms that export 
    to the EU to become ISO 9001:1994 certified. These firms would now be 
    in compliance with most of FDA's final CGMP regulation. Further, FDA 
    has provided continued education efforts over the past 15 years, to 
    mitigate industry costs.
    
    IX. Paperwork Reduction Act of 1995
    
        This final rule contains information collections that are subject 
    to review by OMB under the Paperwork Reduction Act of 1995 (Pub. L. 
    104-13). The title, description and respondents of the information 
    collection are shown below with an estimate of the annual incremental 
    increase in the recordkeeping burden that respondents must undertake to 
    achieve compliance with the final regulation.
        Title: Medical Devices, Quality System Regulations, Current Good 
    Manufacturing Practice Requirements.
        Description: This final quality system regulation amends and 
    revises the current good manufacturing practice requirements for 
    medical devices, set out at 21 CFR part 820. This final regulation 
    replaces quality assurance program requirements with quality system 
    requirements; adds design and purchasing controls; modifies the 
    critical device requirements; revises certain existing requirements, 
    such as validation and management responsibility, to clarify the intent 
    of the requirements; and harmonizes the CGMP regulations for medical 
    devices with quality system specifications in ISO 9001:1994 ``Quality 
    Systems-Model for Quality Assurance in Design, Development, Production, 
    Installation and Servicing.''
        Description of Respondents: Business or other for-profit and small 
    businesses or organizations.
    
    --------------------------------------------------------------------------------------------------------------------------------------------------------
                                                                                                                                                   Total    
                                                                 Number of        Annual       Total annual      Hours per                     operating and
                           CFR section                        record-keepers   frequency of       records      record-keeper    Total hours     maintenance 
                                                                               recordkeeping                                                       costs    
    --------------------------------------------------------------------------------------------------------------------------------------------------------
    820.20(a)...............................................           7,237               1           7,237           10.96          79,386  ..............
    820.20(b)...............................................           7,237               1           7,237            4.88          35,285  ..............
    820.20(c)...............................................           7,237               1           7,237           10.28          74,364  ..............
    820.20(d)...............................................           7,237               1           7,237           16.49         119,305  ..............
    820.20(e)...............................................           7,237               1           7,237           16.49         119,305  ..............
    820.22(a)...............................................           7,237               1           7,237           52.03         376,507  ..............
    820.25(b)...............................................           7,237               1           7,237           21.13         152,896  ..............
    820.30(a)(1)............................................           7,237               1           7,237            2.92          21,162  ..............
    820.30(b)...............................................           7,237               1           7,237            9.91          71,718  ..............
    820.30(c)...............................................           7,237               1           7,237            2.92          21,162  ..............
    820.30(d)...............................................           7,237               1           7,237            2.92          21,162  ..............
    820.30(e)...............................................           7,237               1           7,237           38.98         282,115  ..............
    820.30(f)...............................................           7,237               1           7,237           62.37         451,342     $27,359,420
    820.30(g)...............................................           7,237               1           7,237           62.37         451,342  ..............
    820.30(h)...............................................           7,237               1           7,237            5.56          40,236  ..............
    820.30(i)...............................................           7,237               1           7,237           28.77         208,173  ..............
    820.30(j)...............................................           7,237               1           7,237            4.40          31,848  ..............
    820.40..................................................           7,237               1           7,237           11.76          85,081  ..............
    820.40(b)...............................................           7,237               1           7,237  ..............  ..............  ..............
    820.50 (a)(1) to (a)(3).................................           7,237               1           7,237           31.12         225,240         898,500
    820.50(b)...............................................           7,237               1           7,237           10.04          72,679                
    820.60..................................................           7,237               1           7,237            0.54           3,914  ..............
    820.65..................................................           7,237               1           7,237  ..............  ..............  ..............
    820.70 (a)(1) to (a)(5).................................           7,237               1           7,237  ..............  ..............  ..............
    820.70 (b)-(c)..........................................           7,237               1           7,237  ..............  ..............  ..............
    820.70(d)...............................................           7,237               1           7,237            3.09          22,335  ..............
    820.70(e)...............................................           7,237               1           7,237  ..............  ..............  ..............
    820.70 (g)(1) to (g)(3).................................           7,237               1           7,237  ..............                                
    820.70(h)...............................................           7,237               1           7,237  ..............  ..............  ..............
    820.70(i)...............................................           7,237               1           7,237            9.41          68,092  ..............
    829.72(a)...............................................           7,237               1           7,237            5.83          42,165  ..............
    820.72 (b)(1) to (b)(3).................................           7,237               1           7,237  ..............  ..............  ..............
    820.75(a)...............................................           7,237               1            7,23           72.79          20,172  ..............
    820.75(b)...............................................           7,237               1           7,237  ..............  ..............  ..............
    820.75(b)(2)............................................           7,237               1           7,237            0.15           1,096  ..............
    820.75(c)...............................................           7,237               1           7,237            0.15           1,096  ..............
    820.80 (a)-(e)..........................................           7,237               1           7,237  ..............  ..............  ..............
    820.86..................................................           7,237               1           7,237  ..............  ..............  ..............
    820.90(a)...............................................           7,237               1           7,237            6.11          44,217  ..............
    820.90 (b)(1) to (b)(2).................................           7,237               1           7,237            6.11          44,217  ..............
    820.100 (a)(1) to (a)(7)................................           7,237               1           7,237           20.06         145,144  ..............
    820.100(b)..............................................           7,237               1           7,237  ..............  ..............  ..............
    820.120.................................................           7,237               1           7,237  ..............  ..............  ..............
    820.120(b)..............................................           7,237               1           7,237  ..............  ..............  ..............
    820.120(d)..............................................           7,237               1           7,237  ..............  ..............  ..............
    820.130.................................................           7,237               1           7,237  ..............  ..............  ..............
    820.140.................................................           7,237               1           7,237            9.45          68,418  ..............
    820.150 (a)-(b).........................................           7,237               1           7,237            9.45          68,418  ..............
    820.160 (a)-(b).........................................           7,237               1           7,237  ..............  ..............  ..............
    
    [[Page 52653]]
    
                                                                                                                                                            
    820.170 (a)-(b).........................................           7,237               1           7,237  ..............  ..............  ..............
    820.180.................................................           7,237               1           7,237  ..............  ..............  ..............
    820.181 (a)-(e).........................................           7,237               1           7,237  ..............  ..............  ..............
    820.184 (a)-(f).........................................           7,237               1           7,237  ..............  ..............  ..............
    820.186.................................................           7,237               1           7,237  ..............  ..............  ..............
    820.198 (a)-(c).........................................           7,237               1           7,237            3.71          26,850  ..............
    820.200(a) and 820.200(d)...............................           7,237               1           7,237            4.35          31,459  ..............
    820.250.................................................           7,237               1           7,237  ..............  ..............  ..............
                                                             -----------------------------------------------------------------------------------------------
        Totals..............................................           7,237               1           7,237          487.50       3,527,901     28,257,920 
    --------------------------------------------------------------------------------------------------------------------------------------------------------
    \1\ Incremental increase in hours and costs to achieve compliance with additional requirements.                                                         
    Note: Totals may not add due to rounding                                                                                                                
    
        Under OMB information collection 0910-0073, which expired on June 
    30, 1995, there were 375,266 burden hours approved for recordkeeping 
    requirements currently contained in part 820 to include 114,882 burden 
    hours as a one time start up expenditure for 750 new firms. The 
    additional requirements contained in this final rule will add 3,527,901 
    burden hours to the burden, resulting in a total annual recordkeeping 
    burden of 3,903,167 hours. The 3,527,901 burden hours includes 
    1,433,579 burden hours for a one time start up expenditure for 7,237 
    manufacturers and 2,094,321 burden hours expended annually by 7,237 
    manufacturers.
        The final rule estimate of recordkeeping burden includes about 9.6 
    times as many manufacturers with a one time start up expenditure, due 
    to the addition of the design control requirements, than did FDA's 
    estimate of the manufacturers that would have had a one time start up 
    expenditure under the old regulation. Further the recordkeeping burden 
    hour calculations for the new regulation were done under contract using 
    a more complex methodology involving the estimated noncompliance ratio 
    for small, medium, large, and very large manufacturers (as defined 
    above) times the number of manufacturers in each category and factors 
    in a rate of product innovation for new products, including 510(k) 
    devices. This methodology is more precise than the methodology 
    previously utilized. Therefore, it is very difficult to directly 
    compare the total burden hours in this final rule as compared to the 
    estimated burden hours filed for the old regulation which expired June 
    1995.
        Approximately 85 percent of the additional burden hours for the 
    final rule are from the following four subparts of part 820: (1) 
    Subpart B--Quality System Requirements; (2) Subpart C--Design Controls; 
    (3) Subpart E--Purchasing Controls; and (4) Subpart J--Corrective and 
    Preventive Action. Over 45 percent of the 3,527,901 burden hours are 
    attributed directly to the addition of design control requirements. The 
    recordkeeping burden hours for design control are significant because 
    of the nature of the new requirements, as well as in response to 
    numerous comments on the 1993 and 1995 proposals. The comments 
    requested that the regulation focus on procedures required under design 
    control as compared to prescriptive requirements on the design 
    activities. The quality system requirements, as well as the corrective 
    and preventive action requirements combined are approximately 31 
    percent of the additional recordkeeping burden hours and were in 
    response to two major issues: (1) Most importantly, FDA had identified 
    these two areas as two of the top four deficiencies found during 
    inspections of the medical device industry, across all sizes of 
    manufacturers; and (2) numerous comments requested harmonization with 
    the ISO 9000 series standards. The involvement of management with 
    executive responsibility, the concept of a total quality system which 
    is a closed feedback loop system, and the practice of using that closed 
    loop system in taking appropriate corrective and preventive action is 
    paramount in ensuring that safe and effective medical devices are 
    available to the public. The purchasing control requirements and the 
    respective recordkeeping burden are approximately 8 percent of the 
    additional recordkeeping burden. Purchasing requirements were the 
    overwhelming choice of the medical device industry as compared to the 
    option of the final rule encompassing component manufacturers. See the 
    discussion in section V.7. of this document.
        It is important to note that small manufacturers may comply with 
    this final rule with less procedures and paperwork than larger 
    manufacturers of the same product because the structure and interfaces 
    for a small manufacturer often require less documentation and 
    paperwork.
        Although the November 23, 1993, proposed rule provided a 90 day 
    comment period under the Paperwork Reduction Act of 1980, and this 
    final rule incorporates the comments received, as required by 44 U.S.C. 
    section 3507(d), FDA is providing additional opportunities for public 
    comment under the Paperwork Reduction Act of 1995, which applies to 
    this final rule and was enacted after the expiration of the comment 
    period.
        Therefore, the agency solicits public comment on the information 
    collection requirements in order to: (1) Evaluate whether the proposed 
    collection of information is necessary for the proper performance of 
    the functions of the agency, including whether the information will 
    have practical utility; (2) evaluate the accuracy of the agency's 
    estimate of the burden of the proposed collection of information, 
    including the validity of the methodology and assumptions used; (3) 
    enhance the quality, utility, and clarity of the information to be 
    collected; and (4) minimize the burden of the collection of information 
    on those who are to respond, including through the use of appropriate 
    automated, electronic, mechanical, or other technological collection 
    techniques or other forms of information technology, e.g., permitting 
    electronic submission of responses.
        Individuals and organizations may submit comments on the 
    information collection requirements by December 6, 1996, and should 
    direct comments to FDA's Dockets Management Branch (address above).
        Prior to the effective date of this final rule, FDA will publish a 
    notice in the Federal Register when the information collection 
    requirements in this rule are submitted for OMB approval, and again 
    when OMB makes a decision to approve, modify, or disapprove the
    
    [[Page 52654]]
    
    information collection requirements. Persons are not required to 
    respond to a collection of information unless it displays a currently 
    valid OMB control number.
    
    X. Congressional Review
    
        This final rule has been determined to be a major rule for purposes 
    of 5 U.S.C. 801 et seq., Subtitle E of the Small Business Regulatory 
    Enforcement Fairness Act of 1996 (Pub. L. 104-121). FDA is submitting 
    the information and reports as required by that statute.
    
    XI. References
    
        The following references have been placed on display in the 
    Dockets Management Branch and may be seen by interested persons 
    between 9 a.m. and 4 p.m., Monday through Friday.
        1. ``Device Recalls: A Study of Quality Problems,'' FDA, Center 
    for Devices and Radiological Health, Rockville, MD 20857, HHS 
    Publication FDA 90-4235, January 1990.
        2. ``FDA Medical Device Regulation From Premarket Review to 
    Recall,'' Office of Inspector General, Washington, DC, HHS 
    Publication OEI 09-90-00040, February 1991.
        3. ``Software Related Recalls for Fiscal Years FY 83--FY 91,'' 
    FDA, Center for Devices and Radiological Health, Rockville, MD 
    20857, May 1992.
        4. ISO 9001:1994 ``Quality Systems-- Model for Quality Assurance 
    in Design, Development, Production, Installation, and Servicing.''
        5. ISO draft revision of ISO/CD 13485 ``Quality Systems--Medical 
    Devices-- Supplementary Requirements to ISO 9001.''
        6. Federal Register notice entitled ``Medical Devices; Current 
    Good Manufacturing Practices (CGMP) Regulations Document; Suggested 
    Changes; Availability,'' November 30, 1990 (55 FR 49644).
        7. Federal Register notice entitled ``Medical Devices; Current 
    Good Manufacturing Practice (CGMP) Regulations; Proposed Revisions; 
    Request for Comments,'' November 23, 1993 (55 FR 61952).
        8. Federal Register notice entitled ``Medical Devices; Working 
    Draft of the Current Good Manufacturing Practice (CGMP) Final Rule; 
    Notice of Availability; Request for Comments; Public Meeting,'' July 
    24, 1995 (60 FR 37856).
        9. European Standard (EN) 46001 ``Quality Systems--Medical 
    Devices--Particular Requirements for the Application of EN 29001.''
        10. ``Guidelines on General Principles of Process Validation,'' 
    Center for Drugs and Biologics, and Center for Devices and 
    Radiological Health, FDA, Rockville, MD 20857, May 11, 1987.
        11. ``Medical Devices; Early Warning of Problems Is Hampered by 
    Severe Underreporting,'' United States General Accounting Office, 
    Washington, DC, GAO/PEMD-87-1.
    
    List of Subjects
    
    21 CFR Part 808
    
        Intergovernmental relations, Medical devices.
    
    21 CFR Part 812
    
        Health records, Medical devices, Medical research, Reporting and 
    recordkeeping requirements.
    
    21 CFR Part 820
    
        Medical devices, Reporting and recordkeeping requirements.
    
        Therefore, under the Federal Food, Drug, and Cosmetic Act and under 
    authority delegated to the Commissioner of Food and Drugs, 21 CFR parts 
    808, 812, and 820 are amended as follows:
    
    PART 808--EXEMPTIONS FROM FEDERAL PREEMPTION OF STATE AND LOCAL 
    MEDICAL DEVICE REQUIREMENTS
    
        1. The authority citation for 21 CFR part 808 is revised to read as 
    follows:
    
        Authority: Secs. 520, 521, 701 of the Federal Food, Drug, and 
    Cosmetic Act (21 U.S.C. 360j, 360k, 371).
    
        2. Section 808.1 is amended by adding new paragraph (d)(10) to read 
    as follows:
    
    
    Sec. 808.1  Scope.
    
    * * * * *
        (d) * * *
        (10) Part 820 of this chapter (21 CFR part 820) (CGMP requirements) 
    does not preempt remedies created by States or Territories of the 
    United States, the District of Columbia, or the Commonwealth of Puerto 
    Rico.
    * * * * *
    
    PART 812--INVESTIGATIONAL DEVICE EXEMPTIONS
    
        3. The authority citation for 21 CFR part 812 is revised to read as 
    follows:
    
        Authority: Secs. 301, 501, 502, 503, 505, 506, 507, 510, 513-
    516, 518-520, 701, 702, 704, 721, 801, 803 of the Federal Food, 
    Drug, and Cosmetic Act (21 U.S.C. 331, 351, 352, 353, 355, 356, 357, 
    360, 360c-360f, 360h-360j, 371, 372, 374, 379e, 381, 383); secs. 
    215, 301, 351, 354-360F of the Public Health Service Act (42 U.S.C. 
    216, 241, 262, 263b-263n).
    
        4. Section 812.1 Scope is amended by revising the fourth sentence 
    of paragraph (a) to read as follows:
    
    
    Sec. 812.1  Scope.
    
        (a) * * * An IDE approved under Sec. 812.30 or considered approved 
    under Sec. 812.2(b) exempts a device from the requirements of the 
    following sections of the Federal Food, Drug, and Cosmetic Act (the 
    act) and regulations issued thereunder: Misbranding under section 502 
    of the act, registration, listing, and premarket notification under 
    section 510, performance standards under section 514, premarket 
    approval under section 515, a banned device regulation under section 
    516, records and reports under section 519, restricted device 
    requirements under section 520(e), good manufacturing practice 
    requirements under section 520(f) except for the requirements found in 
    Sec. 820.30, if applicable (unless the sponsor states an intention to 
    comply with these requirements under Sec. 812.20(b)(3) or 
    Sec. 812.140(b)(4)(v)) and color additive requirements under section 
    721.
    * * * * *
        5. Part 820 is revised to read as follows:
    
    PART 820--QUALITY SYSTEM REGULATION
    
    Subpart A--General Provisions
    
    Sec.
    820.1  Scope.
    820.3  Definitions.
    820.5  Quality system.
    
    Subpart B--Quality System Requirements
    
    820.20  Management responsibility.
    820.22  Quality audit.
    820.25  Personnel.
    
    Subpart C--Design Controls
    
    820.30  Design controls.
    
    Subpart D--Document Controls
    
    820.40  Document controls.
    
    Subpart E--Purchasing Controls
    
    820.50  Purchasing controls.
    
    Subpart F--Identification and Traceability
    
    820.60  Identification.
    820.65  Traceability.
    
    Subpart G--Production and Process Controls
    
    820.70  Production and process controls.
    820.72  Inspection, measuring, and test equipment.
    820.75  Process validation.
    
    Subpart H--Acceptance Activities
    
    820.80  Receiving, in-process, and finished device acceptance.
    820.86  Acceptance status.
    
    Subpart I--Nonconforming Product
    
    820.90  Nonconforming product.
    
    Subpart J--Corrective and Preventive Action
    
    820.100  Corrective and preventive action.
    
    Subpart K--Labeling and Packaging Control
    
    820.120  Device labeling.
    820.130  Device packaging.
    
    Subpart L--Handling, Storage, Distribution, and Installation
    
    820.140  Handling.
    820.150  Storage.
    
    [[Page 52655]]
    
    820.160  Distribution.
    820.170  Installation.
    
    Subpart M--Records
    
    820.180  General requirements.
    820.181  Device master record.
    820.184  Device history record.
    820.186  Quality system record.
    820.198  Complaint files.
    
    Subpart N--Servicing
    
    820.200  Servicing.
    
    Subpart O--Statistical Techniques
    
    820.250  Statistical techniques.
    
        Authority: Secs. 501, 502, 510, 513, 514, 515, 518, 519, 520, 
    522, 701, 704, 801, 803 of the Federal Food, Drug, and Cosmetic Act 
    (21 U.S.C. 351, 352, 360, 360c, 360d, 360e, 360h, 360i, 360j, 360l, 
    371, 374, 381, 383).
    
    Subpart A--General Provisions
    
    
    Sec. 820.1  Scope.
    
        (a) Applicability.
        (1) Current good manufacturing practice (CGMP) requirements are set 
    forth in this quality system regulation. The requirements in this part 
    govern the methods used in, and the facilities and controls used for, 
    the design, manufacture, packaging, labeling, storage, installation, 
    and servicing of all finished devices intended for human use. The 
    requirements in this part are intended to ensure that finished devices 
    will be safe and effective and otherwise in compliance with the Federal 
    Food, Drug, and Cosmetic Act (the act). This part establishes basic 
    requirements applicable to manufacturers of finished medical devices. 
    If a manufacturer engages in only some operations subject to the 
    requirements in this part, and not in others, that manufacturer need 
    only comply with those requirements applicable to the operations in 
    which it is engaged. With respect to class I devices, design controls 
    apply only to those devices listed in Sec. 820.30(a)(2). This 
    regulation does not apply to manufacturers of components or parts of 
    finished devices, but such manufacturers are encouraged to use 
    appropriate provisions of this regulation as guidance. Manufacturers of 
    human blood and blood components are not subject to this part, but are 
    subject to part 606 of this chapter.
        (2) The provisions of this part shall be applicable to any finished 
    device as defined in this part, intended for human use, that is 
    manufactured, imported, or offered for import in any State or Territory 
    of the United States, the District of Columbia, or the Commonwealth of 
    Puerto Rico.
        (3) In this regulation the term ``where appropriate'' is used 
    several times. When a requirement is qualified by ``where 
    appropriate,'' it is deemed to be ``appropriate'' unless the 
    manufacturer can document justification otherwise. A requirement is 
    ``appropriate'' if nonimplementation could reasonably be expected to 
    result in the product not meeting its specified requirements or the 
    manufacturer not being able to carry out any necessary corrective 
    action.
        (b) Limitations. The quality system regulation in this part 
    supplements regulations in other parts of this chapter except where 
    explicitly stated otherwise. In the event that it is impossible to 
    comply with all applicable regulations, both in this part and in other 
    parts of this chapter, the regulations specifically applicable to the 
    device in question shall supersede any other generally applicable 
    requirements.
        (c) Authority. Part 820 is established and issued under authority 
    of sections 501, 502, 510, 513, 514, 515, 518, 519, 520, 522, 701, 704, 
    801, 803 of the act (21 U.S.C. 351, 352, 360, 360c, 360d, 360e, 360h, 
    360i, 360j, 360l, 371, 374, 381, 383). The failure to comply with any 
    applicable provision in this part renders a device adulterated under 
    section 501(h) of the act. Such a device, as well as any person 
    responsible for the failure to comply, is subject to regulatory action.
        (d) Foreign manufacturers. If a manufacturer who offers devices for 
    import into the United States refuses to permit or allow the completion 
    of a Food and Drug Administration (FDA) inspection of the foreign 
    facility for the purpose of determining compliance with this part, it 
    shall appear for purposes of section 801(a) of the act, that the 
    methods used in, and the facilities and controls used for, the design, 
    manufacture, packaging, labeling, storage, installation, or servicing 
    of any devices produced at such facility that are offered for import 
    into the United States do not conform to the requirements of section 
    520(f) of the act and this part and that the devices manufactured at 
    that facility are adulterated under section 501(h) of the act.
        (e) Exemptions or variances. (1) Any person who wishes to petition 
    for an exemption or variance from any device quality system requirement 
    is subject to the requirements of section 520(f)(2) of the act. 
    Petitions for an exemption or variance shall be submitted according to 
    the procedures set forth in Sec. 10.30 of this chapter, the FDA's 
    administrative procedures. Guidance is available from the Center for 
    Devices and Radiological Health, Division of Small Manufacturers 
    Assistance, (HFZ-220), 1350 Piccard Dr., Rockville, MD 20850, U.S.A., 
    telephone 1-800-638-2041 or 1-301-443-6597, FAX 301-443-8818.
        (2) FDA may initiate and grant a variance from any device quality 
    system requirement when the agency determines that such variance is in 
    the best interest of the public health. Such variance will remain in 
    effect only so long as there remains a public health need for the 
    device and the device would not likely be made sufficiently available 
    without the variance.
    
    
    Sec. 820.3  Definitions.
    
        (a) Act means the Federal Food, Drug, and Cosmetic Act, as amended 
    (secs. 201-903, 52 Stat. 1040 et seq., as amended (21 U.S.C. 321-394)). 
    All definitions in section 201 of the act shall apply to the 
    regulations in this part.
        (b) Complaint means any written, electronic, or oral communication 
    that alleges deficiencies related to the identity, quality, durability, 
    reliability, safety, effectiveness, or performance of a device after it 
    is released for distribution.
        (c) Component means any raw material, substance, piece, part, 
    software, firmware, labeling, or assembly which is intended to be 
    included as part of the finished, packaged, and labeled device.
        (d) Control number means any distinctive symbols, such as a 
    distinctive combination of letters or numbers, or both, from which the 
    history of the manufacturing, packaging, labeling, and distribution of 
    a unit, lot, or batch of finished devices can be determined.
        (e) Design history file (DHF) means a compilation of records which 
    describes the design history of a finished device.
        (f) Design input means the physical and performance requirements of 
    a device that are used as a basis for device design.
        (g) Design output means the results of a design effort at each 
    design phase and at the end of the total design effort. The finished 
    design output is the basis for the device master record. The total 
    finished design output consists of the device, its packaging and 
    labeling, and the device master record.
        (h) Design review means a documented, comprehensive, systematic 
    examination of a design to evaluate the adequacy of the design 
    requirements, to evaluate the capability of the design to meet these 
    requirements, and to identify problems.
        (i) Device history record (DHR) means a compilation of records 
    containing the production history of a finished device.
        (j) Device master record (DMR) means a compilation of records 
    containing the procedures and specifications for a finished device.
    
    [[Page 52656]]
    
        (k) Establish means define, document (in writing or 
    electronically), and implement.
        (l) Finished device means any device or accessory to any device 
    that is suitable for use or capable of functioning, whether or not it 
    is packaged, labeled, or sterilized.
        (m) Lot or batch means one or more components or finished devices 
    that consist of a single type, model, class, size, composition, or 
    software version that are manufactured under essentially the same 
    conditions and that are intended to have uniform characteristics and 
    quality within specified limits.
        (n) Management with executive responsibility means those senior 
    employees of a manufacturer who have the authority to establish or make 
    changes to the manufacturer's quality policy and quality system.
        (o) Manufacturer means any person who designs, manufactures, 
    fabricates, assembles, or processes a finished device. Manufacturer 
    includes but is not limited to those who perform the functions of 
    contract sterilization, installation, relabeling, remanufacturing, 
    repacking, or specification development, and initial distributors of 
    foreign entities performing these functions.
        (p) Manufacturing material means any material or substance used in 
    or used to facilitate the manufacturing process, a concomitant 
    constituent, or a byproduct constituent produced during the 
    manufacturing process, which is present in or on the finished device as 
    a residue or impurity not by design or intent of the manufacturer.
        (q) Nonconformity means the nonfulfillment of a specified 
    requirement.
        (r) Product means components, manufacturing materials, in- process 
    devices, finished devices, and returned devices.
        (s) Quality means the totality of features and characteristics that 
    bear on the ability of a device to satisfy fitness-for-use, including 
    safety and performance.
        (t) Quality audit means a systematic, independent examination of a 
    manufacturer's quality system that is performed at defined intervals 
    and at sufficient frequency to determine whether both quality system 
    activities and the results of such activities comply with quality 
    system procedures, that these procedures are implemented effectively, 
    and that these procedures are suitable to achieve quality system 
    objectives.
        (u) Quality policy means the overall intentions and direction of an 
    organization with respect to quality, as established by management with 
    executive responsibility.
        (v) Quality system means the organizational structure, 
    responsibilities, procedures, processes, and resources for implementing 
    quality management.
        (w) Remanufacturer means any person who processes, conditions, 
    renovates, repackages, restores, or does any other act to a finished 
    device that significantly changes the finished device's performance or 
    safety specifications, or intended use.
        (x) Rework means action taken on a nonconforming product so that it 
    will fulfill the specified DMR requirements before it is released for 
    distribution.
        (y) Specification means any requirement with which a product, 
    process, service, or other activity must conform.
        (z) Validation means confirmation by examination and provision of 
    objective evidence that the particular requirements for a specific 
    intended use can be consistently fulfilled.
        (1) Process validation means establishing by objective evidence 
    that a process consistently produces a result or product meeting its 
    predetermined specifications.
        (2) Design validation means establishing by objective evidence that 
    device specifications conform with user needs and intended use(s).
        (aa) Verification means confirmation by examination and provision 
    of objective evidence that specified requirements have been fulfilled.
    
    
    Sec. 820.5  Quality system.
    
        Each manufacturer shall establish and maintain a quality system 
    that is appropriate for the specific medical device(s) designed or 
    manufactured, and that meets the requirements of this part.
    
    Subpart B--Quality System Requirements
    
    
    Sec. 820.20  Management responsibility.
    
        (a) Quality policy. Management with executive responsibility shall 
    establish its policy and objectives for, and commitment to, quality. 
    Management with executive responsibility shall ensure that the quality 
    policy is understood, implemented, and maintained at all levels of the 
    organization.
        (b) Organization. Each manufacturer shall establish and maintain an 
    adequate organizational structure to ensure that devices are designed 
    and produced in accordance with the requirements of this part.
        (1) Responsibility and authority. Each manufacturer shall establish 
    the appropriate responsibility, authority, and interrelation of all 
    personnel who manage, perform, and assess work affecting quality, and 
    provide the independence and authority necessary to perform these 
    tasks.
        (2) Resources. Each manufacturer shall provide adequate resources, 
    including the assignment of trained personnel, for management, 
    performance of work, and assessment activities, including internal 
    quality audits, to meet the requirements of this part.
        (3) Management representative. Management with executive 
    responsibility shall appoint, and document such appointment of, a 
    member of management who, irrespective of other responsibilities, shall 
    have established authority over and responsibility for:
        (i) Ensuring that quality system requirements are effectively 
    established and effectively maintained in accordance with this part; 
    and
        (ii) Reporting on the performance of the quality system to 
    management with executive responsibility for review.
        (c) Management review. Management with executive responsibility 
    shall review the suitability and effectiveness of the quality system at 
    defined intervals and with sufficient frequency according to 
    established procedures to ensure that the quality system satisfies the 
    requirements of this part and the manufacturer's established quality 
    policy and objectives. The dates and results of quality system reviews 
    shall be documented.
        (d) Quality planning. Each manufacturer shall establish a quality 
    plan which defines the quality practices, resources, and activities 
    relevant to devices that are designed and manufactured. The 
    manufacturer shall establish how the requirements for quality will be 
    met.
        (e) Quality system procedures. Each manufacturer shall establish 
    quality system procedures and instructions. An outline of the structure 
    of the documentation used in the quality system shall be established 
    where appropriate.
    
    
    Sec. 820.22  Quality audit.
    
        Each manufacturer shall establish procedures for quality audits and 
    conduct such audits to assure that the quality system is in compliance 
    with the established quality system requirements and to determine the 
    effectiveness of the quality system. Quality audits shall be conducted 
    by individuals who do not have direct responsibility for the matters 
    being audited. Corrective action(s), including a reaudit of deficient 
    matters,
    
    [[Page 52657]]
    
    shall be taken when necessary. A report of the results of each quality 
    audit, and reaudit(s) where taken, shall be made and such reports shall 
    be reviewed by management having responsibility for the matters 
    audited. The dates and results of quality audits and reaudits shall be 
    documented.
    
    
    Sec. 820.25  Personnel.
    
        (a) General. Each manufacturer shall have sufficient personnel with 
    the necessary education, background, training, and experience to assure 
    that all activities required by this part are correctly performed.
        (b) Training. Each manufacturer shall establish procedures for 
    identifying training needs and ensure that all personnel are trained to 
    adequately perform their assigned responsibilities. Training shall be 
    documented.
        (1) As part of their training, personnel shall be made aware of 
    device defects which may occur from the improper performance of their 
    specific jobs.
        (2) Personnel who perform verification and validation activities 
    shall be made aware of defects and errors that may be encountered as 
    part of their job functions.
    
    Subpart C--Design Controls
    
    
    Sec. 820.30  Design controls.
    
        (a) General. (1) Each manufacturer of any class III or class II 
    device, and the class I devices listed in paragraph (a)(2) of this 
    section, shall establish and maintain procedures to control the design 
    of the device in order to ensure that specified design requirements are 
    met.
        (2) The following class I devices are subject to design controls:
        (i) Devices automated with computer software; and
        (ii) The devices listed in the following chart.
    
    ------------------------------------------------------------------------
                  Section                              Device               
    ------------------------------------------------------------------------
    868.6810..........................  Catheter, Tracheobronchial Suction. 
    878.4460..........................  Glove, Surgeon's.                   
    880.6760..........................  Restraint, Protective.              
    892.5650..........................  System, Applicator, Radionuclide,   
                                         Manual.                            
    892.5740..........................  Source, Radionuclide Teletherapy.   
    ------------------------------------------------------------------------
    
        (b) Design and development planning. Each manufacturer shall 
    establish and maintain plans that describe or reference the design and 
    development activities and define responsibility for implementation. 
    The plans shall identify and describe the interfaces with different 
    groups or activities that provide, or result in, input to the design 
    and development process. The plans shall be reviewed, updated, and 
    approved as design and development evolves.
        (c) Design input. Each manufacturer shall establish and maintain 
    procedures to ensure that the design requirements relating to a device 
    are appropriate and address the intended use of the device, including 
    the needs of the user and patient. The procedures shall include a 
    mechanism for addressing incomplete, ambiguous, or conflicting 
    requirements. The design input requirements shall be documented and 
    shall be reviewed and approved by a designated individual(s). The 
    approval, including the date and signature of the individual(s) 
    approving the requirements, shall be documented.
        (d) Design output. Each manufacturer shall establish and maintain 
    procedures for defining and documenting design output in terms that 
    allow an adequate evaluation of conformance to design input 
    requirements. Design output procedures shall contain or make reference 
    to acceptance criteria and shall ensure that those design outputs that 
    are essential for the proper functioning of the device are identified. 
    Design output shall be documented, reviewed, and approved before 
    release. The approval, including the date and signature of the 
    individual(s) approving the output, shall be documented.
        (e) Design review. Each manufacturer shall establish and maintain 
    procedures to ensure that formal documented reviews of the design 
    results are planned and conducted at appropriate stages of the device's 
    design development. The procedures shall ensure that participants at 
    each design review include representatives of all functions concerned 
    with the design stage being reviewed and an individual(s) who does not 
    have direct responsibility for the design stage being reviewed, as well 
    as any specialists needed. The results of a design review, including 
    identification of the design, the date, and the individual(s) 
    performing the review, shall be documented in the design history file 
    (the DHF).
        (f) Design verification. Each manufacturer shall establish and 
    maintain procedures for verifying the device design. Design 
    verification shall confirm that the design output meets the design 
    input requirements. The results of the design verification, including 
    identification of the design, method(s), the date, and the 
    individual(s) performing the verification, shall be documented in the 
    DHF.
        (g) Design validation. Each manufacturer shall establish and 
    maintain procedures for validating the device design. Design validation 
    shall be performed under defined operating conditions on initial 
    production units, lots, or batches, or their equivalents. Design 
    validation shall ensure that devices conform to defined user needs and 
    intended uses and shall include testing of production units under 
    actual or simulated use conditions. Design validation shall include 
    software validation and risk analysis, where appropriate. The results 
    of the design validation, including identification of the design, 
    method(s), the date, and the individual(s) performing the validation, 
    shall be documented in the DHF.
        (h) Design transfer. Each manufacturer shall establish and maintain 
    procedures to ensure that the device design is correctly translated 
    into production specifications.
        (i) Design changes. Each manufacturer shall establish and maintain 
    procedures for the identification, documentation, validation or where 
    appropriate verification, review, and approval of design changes before 
    their implementation.
        (j) Design history file. Each manufacturer shall establish and 
    maintain a DHF for each type of device. The DHF shall contain or 
    reference the records necessary to demonstrate that the design was 
    developed in accordance with the approved design plan and the 
    requirements of this part.
    
    Subpart D--Document Controls
    
    
    Sec. 820.40  Document controls.
    
        Each manufacturer shall establish and maintain procedures to 
    control all documents that are required by this part. The procedures 
    shall provide for the following:
        (a) Document approval and distribution. Each manufacturer shall 
    designate an individual(s) to review for adequacy and approve prior to 
    issuance all documents established to meet the requirements of this 
    part. The approval, including the date and signature of the 
    individual(s) approving the document, shall be documented. Documents 
    established to meet the requirements of this part shall be available at 
    all locations for which they are designated, used, or otherwise 
    necessary, and all obsolete documents shall be promptly removed from 
    all points of use or otherwise prevented from unintended use.
        (b) Document changes. Changes to documents shall be reviewed and 
    approved by an individual(s) in the same function or organization that 
    performed the original review and approval, unless specifically 
    designated otherwise. Approved changes shall be
    
    [[Page 52658]]
    
    communicated to the appropriate personnel in a timely manner. Each 
    manufacturer shall maintain records of changes to documents. Change 
    records shall include a description of the change, identification of 
    the affected documents, the signature of the approving individual(s), 
    the approval date, and when the change becomes effective.
    
    Subpart E--Purchasing Controls
    
    
    Sec. 820.50  Purchasing controls.
    
        Each manufacturer shall establish and maintain procedures to ensure 
    that all purchased or otherwise received product and services conform 
    to specified requirements.
        (a) Evaluation of suppliers, contractors, and consultants. Each 
    manufacturer shall establish and maintain the requirements, including 
    quality requirements, that must be met by suppliers, contractors, and 
    consultants. Each manufacturer shall:
        (1) Evaluate and select potential suppliers, contractors, and 
    consultants on the basis of their ability to meet specified 
    requirements, including quality requirements. The evaluation shall be 
    documented.
        (2) Define the type and extent of control to be exercised over the 
    product, services, suppliers, contractors, and consultants, based on 
    the evaluation results.
        (3) Establish and maintain records of acceptable suppliers, 
    contractors, and consultants.
        (b) Purchasing data. Each manufacturer shall establish and maintain 
    data that clearly describe or reference the specified requirements, 
    including quality requirements, for purchased or otherwise received 
    product and services. Purchasing documents shall include, where 
    possible, an agreement that the suppliers, contractors, and consultants 
    agree to notify the manufacturer of changes in the product or service 
    so that manufacturers may determine whether the changes may affect the 
    quality of a finished device. Purchasing data shall be approved in 
    accordance with Sec. 820.40.
    
    Subpart F--Identification and Traceability
    
    
    Sec. 820.60  Identification.
    
        Each manufacturer shall establish and maintain procedures for 
    identifying product during all stages of receipt, production, 
    distribution, and installation to prevent mixups.
    
    
    Sec. 820.65  Traceability.
    
        Each manufacturer of a device that is intended for surgical implant 
    into the body or to support or sustain life and whose failure to 
    perform when properly used in accordance with instructions for use 
    provided in the labeling can be reasonably expected to result in a 
    significant injury to the user shall establish and maintain procedures 
    for identifying with a control number each unit, lot, or batch of 
    finished devices and where appropriate components. The procedures shall 
    facilitate corrective action. Such identification shall be documented 
    in the DHR.
    
    Subpart G--Production and Process Controls
    
    
    Sec. 820.70  Production and process controls.
    
        (a) General. Each manufacturer shall develop, conduct, control, and 
    monitor production processes to ensure that a device conforms to its 
    specifications. Where deviations from device specifications could occur 
    as a result of the manufacturing process, the manufacturer shall 
    establish and maintain process control procedures that describe any 
    process controls necessary to ensure conformance to specifications. 
    Where process controls are needed they shall include:
        (1) Documented instructions, standard operating procedures (SOP's), 
    and methods that define and control the manner of production;
        (2) Monitoring and control of process parameters and component and 
    device characteristics during production;
        (3) Compliance with specified reference standards or codes;
        (4) The approval of processes and process equipment; and
        (5) Criteria for workmanship which shall be expressed in documented 
    standards or by means of identified and approved representative 
    samples.
        (b) Production and process changes. Each manufacturer shall 
    establish and maintain procedures for changes to a specification, 
    method, process, or procedure. Such changes shall be verified or where 
    appropriate validated according to Sec. 820.75, before implementation 
    and these activities shall be documented. Changes shall be approved in 
    accordance with Sec. 820.40.
        (c) Environmental control. Where environmental conditions could 
    reasonably be expected to have an adverse effect on product quality, 
    the manufacturer shall establish and maintain procedures to adequately 
    control these environmental conditions. Environmental control system(s) 
    shall be periodically inspected to verify that the system, including 
    necessary equipment, is adequate and functioning properly. These 
    activities shall be documented and reviewed.
        (d) Personnel. Each manufacturer shall establish and maintain 
    requirements for the health, cleanliness, personal practices, and 
    clothing of personnel if contact between such personnel and product or 
    environment could reasonably be expected to have an adverse effect on 
    product quality. The manufacturer shall ensure that maintenance and 
    other personnel who are required to work temporarily under special 
    environmental conditions are appropriately trained or supervised by a 
    trained individual.
        (e) Contamination control. Each manufacturer shall establish and 
    maintain procedures to prevent contamination of equipment or product by 
    substances that could reasonably be expected to have an adverse effect 
    on product quality.
        (f) Buildings. Buildings shall be of suitable design and contain 
    sufficient space to perform necessary operations, prevent mixups, and 
    assure orderly handling.
        (g) Equipment. Each manufacturer shall ensure that all equipment 
    used in the manufacturing process meets specified requirements and is 
    appropriately designed, constructed, placed, and installed to 
    facilitate maintenance, adjustment, cleaning, and use.
        (1) Maintenance schedule. Each manufacturer shall establish and 
    maintain schedules for the adjustment, cleaning, and other maintenance 
    of equipment to ensure that manufacturing specifications are met. 
    Maintenance activities, including the date and individual(s) performing 
    the maintenance activities, shall be documented.
        (2) Inspection. Each manufacturer shall conduct periodic 
    inspections in accordance with established procedures to ensure 
    adherence to applicable equipment maintenance schedules. The 
    inspections, including the date and individual(s) conducting the 
    inspections, shall be documented.
        (3) Adjustment. Each manufacturer shall ensure that any inherent 
    limitations or allowable tolerances are visibly posted on or near 
    equipment requiring periodic adjustments or are readily available to 
    personnel performing these adjustments.
        (h) Manufacturing material. Where a manufacturing material could 
    reasonably be expected to have an adverse effect on product quality, 
    the manufacturer shall establish and maintain procedures for the use 
    and removal of such manufacturing material
    
    [[Page 52659]]
    
    to ensure that it is removed or limited to an amount that does not 
    adversely affect the device's quality. The removal or reduction of such 
    manufacturing material shall be documented.
        (i) Automated processes. When computers or automated data 
    processing systems are used as part of production or the quality 
    system, the manufacturer shall validate computer software for its 
    intended use according to an established protocol. All software changes 
    shall be validated before approval and issuance. These validation 
    activities and results shall be documented.
    
    
    Sec. 820.72  Inspection, measuring, and test equipment.
    
        (a) Control of inspection, measuring, and test equipment. Each 
    manufacturer shall ensure that all inspection, measuring, and test 
    equipment, including mechanical, automated, or electronic inspection 
    and test equipment, is suitable for its intended purposes and is 
    capable of producing valid results. Each manufacturer shall establish 
    and maintain procedures to ensure that equipment is routinely 
    calibrated, inspected, checked, and maintained. The procedures shall 
    include provisions for handling, preservation, and storage of 
    equipment, so that its accuracy and fitness for use are maintained. 
    These activities shall be documented.
        (b) Calibration. Calibration procedures shall include specific 
    directions and limits for accuracy and precision. When accuracy and 
    precision limits are not met, there shall be provisions for remedial 
    action to reestablish the limits and to evaluate whether there was any 
    adverse effect on the device's quality. These activities shall be 
    documented.
        (1) Calibration standards. Calibration standards used for 
    inspection, measuring, and test equipment shall be traceable to 
    national or international standards. If national or international 
    standards are not practical or available, the manufacturer shall use an 
    independent reproducible standard. If no applicable standard exists, 
    the manufacturer shall establish and maintain an in-house standard.
        (2) Calibration records. The equipment identification, calibration 
    dates, the individual performing each calibration, and the next 
    calibration date shall be documented. These records shall be displayed 
    on or near each piece of equipment or shall be readily available to the 
    personnel using such equipment and to the individuals responsible for 
    calibrating the equipment.
    
    
    Sec. 820.75  Process validation.
    
        (a) Where the results of a process cannot be fully verified by 
    subsequent inspection and test, the process shall be validated with a 
    high degree of assurance and approved according to established 
    procedures. The validation activities and results, including the date 
    and signature of the individual(s) approving the validation and where 
    appropriate the major equipment validated, shall be documented.
        (b) Each manufacturer shall establish and maintain procedures for 
    monitoring and control of process parameters for validated processes to 
    ensure that the specified requirements continue to be met.
        (1) Each manufacturer shall ensure that validated processes are 
    performed by qualified individual(s).
        (2) For validated processes, the monitoring and control methods and 
    data, the date performed, and, where appropriate, the individual(s) 
    performing the process or the major equipment used shall be documented.
        (c) When changes or process deviations occur, the manufacturer 
    shall review and evaluate the process and perform revalidation where 
    appropriate. These activities shall be documented.
    
    Subpart H--Acceptance Activities
    
    
    Sec. 820.80  Receiving, in-process, and finished device acceptance.
    
        (a) General. Each manufacturer shall establish and maintain 
    procedures for acceptance activities. Acceptance activities include 
    inspections, tests, or other verification activities.
        (b) Receiving acceptance activities. Each manufacturer shall 
    establish and maintain procedures for acceptance of incoming product. 
    Incoming product shall be inspected, tested, or otherwise verified as 
    conforming to specified requirements. Acceptance or rejection shall be 
    documented.
        (c) In-process acceptance activities. Each manufacturer shall 
    establish and maintain acceptance procedures, where appropriate, to 
    ensure that specified requirements for in-process product are met. Such 
    procedures shall ensure that in-process product is controlled until the 
    required inspection and tests or other verification activities have 
    been completed, or necessary approvals are received, and are 
    documented.
        (d) Final acceptance activities. Each manufacturer shall establish 
    and maintain procedures for finished device acceptance to ensure that 
    each production run, lot, or batch of finished devices meets acceptance 
    criteria. Finished devices shall be held in quarantine or otherwise 
    adequately controlled until released. Finished devices shall not be 
    released for distribution until: (1) The activities required in the DMR 
    are completed; (2) the associated data and documentation is reviewed; 
    (3) the release is authorized by the signature of a designated 
    individual(s); and (4) the authorization is dated.
        (e) Acceptance records. Each manufacturer shall document acceptance 
    activities required by this part. These records shall include: (1) The 
    acceptance activities performed; (2) the dates acceptance activities 
    are performed; (3) the results; (4) the signature of the individual(s) 
    conducting the acceptance activities; and (5) where appropriate the 
    equipment used. These records shall be part of the DHR.
    
    
    Sec. 820.86  Acceptance status.
    
        Each manufacturer shall identify by suitable means the acceptance 
    status of product, to indicate the conformance or nonconformance of 
    product with acceptance criteria. The identification of acceptance 
    status shall be maintained throughout manufacturing, packaging, 
    labeling, installation, and servicing of the product to ensure that 
    only product which has passed the required acceptance activities is 
    distributed, used, or installed.
    
    Subpart I--Nonconforming Product
    
    
    Sec. 820.90  Nonconforming product.
    
        (a) Control of nonconforming product. Each manufacturer shall 
    establish and maintain procedures to control product that does not 
    conform to specified requirements. The procedures shall address the 
    identification, documentation, evaluation, segregation, and disposition 
    of nonconforming product. The evaluation of nonconformance shall 
    include a determination of the need for an investigation and 
    notification of the persons or organizations responsible for the 
    nonconformance. The evaluation and any investigation shall be 
    documented.
        (b) Nonconformity review and disposition. (1) Each manufacturer 
    shall establish and maintain procedures that define the responsibility 
    for review and the authority for the disposition of nonconforming 
    product. The procedures shall set forth the review and disposition 
    process. Disposition of nonconforming product shall be documented. 
    Documentation shall include the justification for use of nonconforming 
    product and the signature of the individual(s) authorizing the use.
    
    [[Page 52660]]
    
        (2) Each manufacturer shall establish and maintain procedures for 
    rework, to include retesting and reevaluation of the nonconforming 
    product after rework, to ensure that the product meets its current 
    approved specifications. Rework and reevaluation activities, including 
    a determination of any adverse effect from the rework upon the product, 
    shall be documented in the DHR.
    
    Subpart J--Corrective and Preventive Action
    
    
    Sec. 820.100  Corrective and preventive action.
    
        (a) Each manufacturer shall establish and maintain procedures for 
    implementing corrective and preventive action. The procedures shall 
    include requirements for:
        (1) Analyzing processes, work operations, concessions, quality 
    audit reports, quality records, service records, complaints, returned 
    product, and other sources of quality data to identify existing and 
    potential causes of nonconforming product, or other quality problems. 
    Appropriate statistical methodology shall be employed where necessary 
    to detect recurring quality problems;
        (2) Investigating the cause of nonconformities relating to product, 
    processes, and the quality system;
        (3) Identifying the action(s) needed to correct and prevent 
    recurrence of nonconforming product and other quality problems;
        (4) Verifying or validating the corrective and preventive action to 
    ensure that such action is effective and does not adversely affect the 
    finished device;
        (5) Implementing and recording changes in methods and procedures 
    needed to correct and prevent identified quality problems;
        (6) Ensuring that information related to quality problems or 
    nonconforming product is disseminated to those directly responsible for 
    assuring the quality of such product or the prevention of such 
    problems; and
        (7) Submitting relevant information on identified quality problems, 
    as well as corrective and preventive actions, for management review.
        (b) All activities required under this section, and their results, 
    shall be documented.
    
    Subpart K--Labeling and Packaging Control
    
    
    Sec. 820.120  Device labeling.
    
        Each manufacturer shall establish and maintain procedures to 
    control labeling activities.
        (a) Label integrity. Labels shall be printed and applied so as to 
    remain legible and affixed during the customary conditions of 
    processing, storage, handling, distribution, and where appropriate use.
        (b) Labeling inspection. Labeling shall not be released for storage 
    or use until a designated individual(s) has examined the labeling for 
    accuracy including, where applicable, the correct expiration date, 
    control number, storage instructions, handling instructions, and any 
    additional processing instructions. The release, including the date and 
    signature of the individual(s) performing the examination, shall be 
    documented in the DHR.
        (c) Labeling storage. Each manufacturer shall store labeling in a 
    manner that provides proper identification and is designed to prevent 
    mixups.
        (d) Labeling operations. Each manufacturer shall control labeling 
    and packaging operations to prevent labeling mixups. The label and 
    labeling used for each production unit, lot, or batch shall be 
    documented in the DHR.
        (e) Control number. Where a control number is required by 
    Sec. 820.65, that control number shall be on or shall accompany the 
    device through distribution.
    
    
    Sec. 820.130  Device packaging.
    
        Each manufacturer shall ensure that device packaging and shipping 
    containers are designed and constructed to protect the device from 
    alteration or damage during the customary conditions of processing, 
    storage, handling, and distribution.
    
    Subpart L--Handling, Storage, Distribution, and Installation
    
    
    Sec. 820.140  Handling.
    
        Each manufacturer shall establish and maintain procedures to ensure 
    that mixups, damage, deterioration, contamination, or other adverse 
    effects to product do not occur during handling.
    
    
    Sec. 820.150  Storage.
    
        (a) Each manufacturer shall establish and maintain procedures for 
    the control of storage areas and stock rooms for product to prevent 
    mixups, damage, deterioration, contamination, or other adverse effects 
    pending use or distribution and to ensure that no obsolete, rejected, 
    or deteriorated product is used or distributed. When the quality of 
    product deteriorates over time, it shall be stored in a manner to 
    facilitate proper stock rotation, and its condition shall be assessed 
    as appropriate.
        (b) Each manufacturer shall establish and maintain procedures that 
    describe the methods for authorizing receipt from and dispatch to 
    storage areas and stock rooms.
    
    
    Sec. 820.160  Distribution.
    
        (a) Each manufacturer shall establish and maintain procedures for 
    control and distribution of finished devices to ensure that only those 
    devices approved for release are distributed and that purchase orders 
    are reviewed to ensure that ambiguities and errors are resolved before 
    devices are released for distribution. Where a device's fitness for use 
    or quality deteriorates over time, the procedures shall ensure that 
    expired devices or devices deteriorated beyond acceptable fitness for 
    use are not distributed.
        (b) Each manufacturer shall maintain distribution records which 
    include or refer to the location of:
        (1) The name and address of the initial consignee;
        (2) The identification and quantity of devices shipped;
        (3) The date shipped; and
        (4) Any control number(s) used.
    
    
    Sec. 820.170  Installation.
    
        (a) Each manufacturer of a device requiring installation shall 
    establish and maintain adequate installation and inspection 
    instructions, and where appropriate test procedures. Instructions and 
    procedures shall include directions for ensuring proper installation so 
    that the device will perform as intended after installation. The 
    manufacturer shall distribute the instructions and procedures with the 
    device or otherwise make them available to the person(s) installing the 
    device.
        (b) The person installing the device shall ensure that the 
    installation, inspection, and any required testing are performed in 
    accordance with the manufacturer's instructions and procedures and 
    shall document the inspection and any test results to demonstrate 
    proper installation.
    
    Subpart M--Records
    
    
    Sec. 820.180  General requirements.
    
        All records required by this part shall be maintained at the 
    manufacturing establishment or other location that is reasonably 
    accessible to responsible officials of the manufacturer and to 
    employees of FDA designated to perform inspections. Such records, 
    including those not stored at the inspected establishment, shall be 
    made readily available for review and copying by FDA employee(s). Such 
    records shall be legible and shall be stored to
    
    [[Page 52661]]
    
    minimize deterioration and to prevent loss. Those records stored in 
    automated data processing systems shall be backed up.
        (a) Confidentiality. Records deemed confidential by the 
    manufacturer may be marked to aid FDA in determining whether 
    information may be disclosed under the public information regulation in 
    part 20 of this chapter.
        (b) Record retention period. All records required by this part 
    shall be retained for a period of time equivalent to the design and 
    expected life of the device, but in no case less than 2 years from the 
    date of release for commercial distribution by the manufacturer.
        (c) Exceptions. This section does not apply to the reports required 
    by Sec. 820.20(c) Management review, Sec. 820.22 Quality audits, and 
    supplier audit reports used to meet the requirements of Sec. 820.50(a) 
    Evaluation of suppliers, contractors, and consultants, but does apply 
    to procedures established under these provisions. Upon request of a 
    designated employee of FDA, an employee in management with executive 
    responsibility shall certify in writing that the management reviews and 
    quality audits required under this part, and supplier audits where 
    applicable, have been performed and documented, the dates on which they 
    were performed, and that any required corrective action has been 
    undertaken.
    
    
    Sec. 820.181  Device master record.
    
        Each manufacturer shall maintain device master records (DMR's). 
    Each manufacturer shall ensure that each DMR is prepared and approved 
    in accordance with Sec. 820.40. The DMR for each type of device shall 
    include, or refer to the location of, the following information:
        (a) Device specifications including appropriate drawings, 
    composition, formulation, component specifications, and software 
    specifications;
        (b) Production process specifications including the appropriate 
    equipment specifications, production methods, production procedures, 
    and production environment specifications;
        (c) Quality assurance procedures and specifications including 
    acceptance criteria and the quality assurance equipment to be used;
        (d) Packaging and labeling specifications, including methods and 
    processes used; and
        (e) Installation, maintenance, and servicing procedures and 
    methods.
    
    
    Sec. 820.184  Device history record.
    
        Each manufacturer shall maintain device history records (DHR's). 
    Each manufacturer shall establish and maintain procedures to ensure 
    that DHR's for each batch, lot, or unit are maintained to demonstrate 
    that the device is manufactured in accordance with the DMR and the 
    requirements of this part. The DHR shall include, or refer to the 
    location of, the following information:
        (a) The dates of manufacture;
        (b) The quantity manufactured;
        (c) The quantity released for distribution;
        (d) The acceptance records which demonstrate the device is 
    manufactured in accordance with the DMR;
        (e) The primary identification label and labeling used for each 
    production unit; and
        (f) Any device identification(s) and control number(s) used.
    
    
    Sec. 820.186  Quality system record.
    
        Each manufacturer shall maintain a quality system record (QSR). The 
    QSR shall include, or refer to the location of, procedures and the 
    documentation of activities required by this part that are not specific 
    to a particular type of device(s), including, but not limited to, the 
    records required by Sec. 820.20. Each manufacturer shall ensure that 
    the QSR is prepared and approved in accordance with Sec. 820.40.
    
    
    Sec. 820.198  Complaint files.
    
        (a) Each manufacturer shall maintain complaint files. Each 
    manufacturer shall establish and maintain procedures for receiving, 
    reviewing, and evaluating complaints by a formally designated unit. 
    Such procedures shall ensure that:
        (1) All complaints are processed in a uniform and timely manner;
        (2) Oral complaints are documented upon receipt; and
        (3) Complaints are evaluated to determine whether the complaint 
    represents an event which is required to be reported to FDA under part 
    803 or 804 of this chapter, Medical Device Reporting.
        (b) Each manufacturer shall review and evaluate all complaints to 
    determine whether an investigation is necessary. When no investigation 
    is made, the manufacturer shall maintain a record that includes the 
    reason no investigation was made and the name of the individual 
    responsible for the decision not to investigate.
        (c) Any complaint involving the possible failure of a device, 
    labeling, or packaging to meet any of its specifications shall be 
    reviewed, evaluated, and investigated, unless such investigation has 
    already been performed for a similar complaint and another 
    investigation is not necessary.
        (d) Any complaint that represents an event which must be reported 
    to FDA under part 803 or 804 of this chapter shall be promptly 
    reviewed, evaluated, and investigated by a designated individual(s) and 
    shall be maintained in a separate portion of the complaint files or 
    otherwise clearly identified. In addition to the information required 
    by Sec. 820.198(e), records of investigation under this paragraph shall 
    include a determination of:
        (1) Whether the device failed to meet specifications;
        (2) Whether the device was being used for treatment or diagnosis; 
    and
        (3) The relationship, if any, of the device to the reported 
    incident or adverse event.
        (e) When an investigation is made under this section, a record of 
    the investigation shall be maintained by the formally designated unit 
    identified in paragraph (a) of this section. The record of 
    investigation shall include:
        (1) The name of the device;
        (2) The date the complaint was received;
        (3) Any device identification(s) and control number(s) used;
        (4) The name, address, and phone number of the complainant;
        (5) The nature and details of the complaint;
        (6) The dates and results of the investigation;
        (7) Any corrective action taken; and
        (8) Any reply to the complainant.
        (f) When the manufacturer's formally designated complaint unit is 
    located at a site separate from the manufacturing establishment, the 
    investigated complaint(s) and the record(s) of investigation shall be 
    reasonably accessible to the manufacturing establishment.
        (g) If a manufacturer's formally designated complaint unit is 
    located outside of the United States, records required by this section 
    shall be reasonably accessible in the United States at either:
        (1) A location in the United States where the manufacturer's 
    records are regularly kept; or
        (2) The location of the initial distributor.
    
    Subpart N--Servicing
    
    
    Sec. 820.200  Servicing.
    
        (a) Where servicing is a specified requirement, each manufacturer 
    shall establish and maintain instructions and procedures for performing 
    and verifying that the servicing meets the specified requirements.
        (b) Each manufacturer shall analyze service reports with 
    appropriate
    
    [[Page 52662]]
    
    statistical methodology in accordance with Sec. 820.100.
        (c) Each manufacturer who receives a service report that represents 
    an event which must be reported to FDA under part 803 or 804 of this 
    chapter shall automatically consider the report a complaint and shall 
    process it in accordance with the requirements of Sec. 820.198.
        (d) Service reports shall be documented and shall include:
        (1) The name of the device serviced;
        (2) Any device identification(s) and control number(s) used;
        (3) The date of service;
        (4) The individual(s) servicing the device;
        (5) The service performed; and
        (6) The test and inspection data.
    
    Subpart O--Statistical Techniques
    
    
    Sec. 820.250  Statistical techniques.
    
        (a) Where appropriate, each manufacturer shall establish and 
    maintain procedures for identifying valid statistical techniques 
    required for establishing, controlling, and verifying the acceptability 
    of process capability and product characteristics.
        (b) Sampling plans, when used, shall be written and based on a 
    valid statistical rationale. Each manufacturer shall establish and 
    maintain procedures to ensure that sampling methods are adequate for 
    their intended use and to ensure that when changes occur the sampling 
    plans are reviewed. These activities shall be documented.
    
        Dated: October 1, 1996.
    David A. Kessler,
    Commissioner of Food and Drugs.
    Donna E. Shalala,
    Secretary of Health and Human Services.
    [FR Doc. 96-25720 Filed 10-3-96; 11:22 am]
    BILLING CODE 4160-01-P
    
    
    

Document Information

Effective Date:
6/1/1997
Published:
10/07/1996
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final rule.
Document Number:
96-25720
Dates:
The regulation is effective June 1, 1997. For more information on compliance with 21 CFR 820.30 see section IV. of this document.
Pages:
52602-52662 (61 pages)
Docket Numbers:
Docket No. 90N-0172
RINs:
0910-AA09: Implementation of the Safe Medical Devices Act of 1990
RIN Links:
https://www.federalregister.gov/regulations/0910-AA09/implementation-of-the-safe-medical-devices-act-of-1990
PDF File:
96-25720.pdf
CFR: (93)
21 CFR 814.39(a)
21 CFR 820.70(a)
21 CFR 820.72(a)
21 CFR 820.84(a)
21 CFR 820.100(a)(3)
More ...