2015-29632. Medical Devices; Gastroenterology-Urology Devices; Classification of the Prostate Lesion Documentation System  

  • Start Preamble Start Printed Page 72899

    AGENCY:

    Food and Drug Administration, HHS.

    ACTION:

    Final order.

    SUMMARY:

    The Food and Drug Administration (FDA) is classifying the prostate lesion documentation system into class II (special controls). The special controls that will apply to the device are identified in this order and will be part of the codified language for the prostate lesion documentation system classification. The Agency is classifying the device into class II (special controls) in order to provide a reasonable assurance of safety and effectiveness of the device.

    DATES:

    This order is effective November 23, 2015. The classification was applicable on April 27, 2012.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    Robert J. De Luca, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. G214, Silver Spring, MD, 20993-0002, 301-796-6551, robert.deluca@fda.hhs.gov.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    I. Background

    In accordance with section 513(f)(1) of the Federal Food, Drug, and Cosmetic Act (the FD&C Act) (21 U.S.C. 360c(f)(1)), devices that were not in commercial distribution before May 28, 1976 (the date of enactment of the Medical Device Amendments of 1976), generally referred to as postamendments devices, are classified automatically by statute into class III without any FDA rulemaking process. These devices remain in class III and require premarket approval, unless and until the device is classified or reclassified into class I or II, or FDA issues an order finding the device to be substantially equivalent, in accordance with section 513(i) of the FD&C Act, to a predicate device that does not require premarket approval. The Agency determines whether new devices are substantially equivalent to predicate devices by means of premarket notification procedures in section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807) of the regulations.

    Section 513(f)(2) of the FD&C Act, as amended by section 607 of the Food and Drug Administration Safety and Innovation Act (Pub. L. 112-144), provides two procedures by which a person may request FDA to classify a device under the criteria set forth in section 513(a)(1). Under the first procedure, the person submits a premarket notification under section 510(k) of the FD&C Act for a device that has not previously been classified and, within 30 days of receiving an order classifying the device into class III under section 513(f)(1) of the FD&C Act, the person requests a classification under section 513(f)(2). Under the second procedure, rather than first submitting a premarket notification under section 510(k) of the FD&C Act and then a request for classification under the first procedure, the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence and requests a classification under section 513(f)(2) of the FD&C Act. If the person submits a request to classify the device under this second procedure, FDA may decline to undertake the classification request if FDA identifies a legally marketed device that could provide a reasonable basis for review of substantial equivalence with the device or if FDA determines that the device submitted is not of “low-moderate risk” or that general controls would be inadequate to control the risks and special controls to mitigate the risks cannot be developed.

    In response to a request to classify a device under either procedure provided by section 513(f)(2) of the FD&C Act, FDA will classify the device by written order within 120 days. This classification will be the initial classification of the device.

    In accordance with section 513(f)(1) of the FD&C Act, FDA issued an order on April 22, 2010, classifying the prostate mechanical imager into class III, because it was neither substantially equivalent to a device that was introduced or delivered for introduction into interstate commerce for commercial distribution before May 28, 1976, nor a device which was subsequently reclassified into class I or class II. On May 21, 2010, Artann Laboratories, Inc., submitted a request for classification of the prostate mechanical imager under section 513(f)(2) of the FD&C Act. The manufacturer recommended that the device be classified into class II (Ref. 1).

    In accordance with section 513(f)(2) of the FD&C Act, FDA reviewed the request for de novo classification in order to classify the device under the criteria for classification set forth in section 513(a)(1). FDA classifies devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls to provide reasonable assurance of the safety and effectiveness of the device for its intended use. After review of the information submitted in the request, FDA determined that the device can be classified into class II with the establishment of special controls. FDA believes these special controls, in addition to general controls, will provide reasonable assurance of the safety and effectiveness of the device.

    Therefore, on April 27, 2012, FDA issued an order to the requestor classifying the device into class II. FDA is codifying the classification of the device by adding 21 CFR 876.2050.

    Following the effective date of this final classification order, any firm submitting a premarket notification (510(k)) for a prostate lesion documentation system will need to comply with the special controls named in this final order. The device is assigned the generic name prostate lesion documentation system, and it is identified as a prescription device intended for use in producing an image of the prostate as an aid in documenting prostate abnormalities previously identified during a digital rectal Start Printed Page 72900examination. The device uses pressure sensors and image reconstruction software to produce a prostate image that highlights regional differences in intraprostatic tissue elasticity or stiffness. The device is limited to use as a documentation tool and is not intended for diagnostic purposes or for influencing any clinical decisions.

    FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks intable 1.

    Table 1—Prostate Lesion Documentation System Risks and Mitigation Measures

    Identified risksMitigation measures
    Failure to consistently produce an accurate imagePerformance Testing (non-clinical and clinical)
    Software Verification, Validation, and Hazard Analysis
    Labeling
    Misinterpretation of displayed imagesLabeling
    User errorLabeling
    Microbial contamination from reusable componentsLabeling
    Validation of Reprocessing Methods and Instructions
    Adverse tissue reactionBiocompatibility Testing
    Electromagnetic incompatibilityElectromagnetic Compatibility Testing
    Electrical injuryElectrical Safety Testing
    Thermal injuryThermal Safety Testing
    Mechanical injuryMechanical Safety Testing

    FDA believes that the measures set forth in the following special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of the safety and effectiveness:

    • Non-clinical and clinical performance testing must demonstrate the accuracy and reproducibility of the constructed image.
    • Appropriate analysis/testing must validate electromagnetic compatibility, electrical safety, thermal safety, and mechanical safety.
    • Appropriate software verification, validation, and hazard analysis must be performed.
    • All elements of the device that may contact the patient must be demonstrated to be biocompatible.
    • Methods and instructions for reprocessing of any reusable components must be properly validated.
    • The labeling must include specific information needed to ensure proper use of the device.

    Prostate lesion documentation systems are prescription devices restricted to patient use only upon the authorization of a practitioner licensed by law to administer or use the device; see 21 CFR 801.109 (Prescription devices).

    Section 510(m) of the FD&C Act provides that FDA may exempt a class II device from the premarket notification requirements under section 510(k) if FDA determines that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device. For this type of device, FDA has determined that premarket notification is necessary to provide reasonable assurance of the safety and effectiveness of the device. Therefore, this device is not exempt from premarket notification requirements. Persons who intend to market this type of device must submit to FDA a premarket notification, prior to marketing the device, which contains information about the prostate lesion documentation system they intend to market.

    II. Environmental Impact

    We have determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

    III. Paperwork Reduction Act of 1995

    This final administrative order establishes special controls that refer to previously approved collections of information found in other FDA regulations. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections of information in part 807, subpart E, regarding premarket notification submissions have been approved under OMB control number 0910-0120, and the collections of information in 21 CFR part 801, regarding labeling have been approved under OMB control number 0910-0485.

    IV. Reference

    The following reference has been placed on display in the Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852, and is available for viewing by interested persons between 9 a.m. and 4 p.m., Monday through Friday; it is also available electronically at http://www.regulations.gov.

    1. DEN100016: De novo request per section 513(f)(2) of the FD&C Act from Artann Laboratories, Inc., dated May 21, 2010.

    Start List of Subjects

    List of Subjects in 21 CFR Part 876

    • Medical devices
    End List of Subjects

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 876 is amended as follows:

    Start Part

    PART 876—GASTROENTEROLOGY-UROLOGY DEVICES

    End Part Start Amendment Part

    1. The authority citation for 21 CFR part 876 continues to read as follows:

    End Amendment Part Start Authority

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

    End Authority Start Amendment Part

    2. Add § 876.2050 to subpart C to read as follows:

    End Amendment Part
    Prostate lesion documentation system.

    (a) Identification. A prostate lesion documentation system is a prescription device intended for use in producing an image of the prostate as an aid in documenting prostate abnormalities previously identified during a digital rectal examination. The device uses pressure sensors and image reconstruction software to produce a prostate image that highlights regional differences in intraprostatic tissue elasticity or stiffness. The device is limited to use as a documentation tool and is not intended for diagnostic purposes or for influencing any clinical decisions.Start Printed Page 72901

    (b) Classification. Class II (special controls). The special controls for this device are:

    (1) Non-clinical and clinical performance testing must demonstrate the accuracy and reproducibility of the constructed image.

    (2) Appropriate analysis/testing must validate electromagnetic compatibility, electrical safety, thermal safety, and mechanical safety.

    (3) Appropriate software verification, validation, and hazard analysis must be performed.

    (4) All elements of the device that may contact the patient must be demonstrated to be biocompatible.

    (5) Methods and instructions for reprocessing of any reusable components must be properly validated.

    (6) The labeling must include specific information needed to ensure proper use of the device.

    Start Signature

    Dated: November 16, 2015.

    Leslie Kux,

    Associate Commissioner for Policy.

    End Signature End Supplemental Information

    [FR Doc. 2015-29632 Filed 11-20-15; 8:45 am]

    BILLING CODE 4164-01-P

Document Information

Effective Date:
11/23/2015
Published:
11/23/2015
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final order.
Document Number:
2015-29632
Dates:
This order is effective November 23, 2015. The classification was applicable on April 27, 2012.
Pages:
72899-72901 (3 pages)
Docket Numbers:
Docket No. FDA-2015-N-3720
Topics:
Medical devices
PDF File:
2015-29632.pdf
CFR: (1)
21 CFR 876.2050