2021-26738. Medical Devices; Gastroenterology-Urology Devices; Classification of the Ingested, Transient, Space Occupying Device for Weight Management and/or Weight Loss  

  • Start Preamble

    AGENCY:

    Food and Drug Administration, Department of Health and Human Services (HHS).

    ACTION:

    Final amendment; final order.

    SUMMARY:

    The Food and Drug Administration (FDA or we) is classifying the ingested, transient, space occupying device for weight management and/or weight loss into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the ingested, transient, space occupying device for weight management and/or weight loss's classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices.

    DATES:

    This order is effective December 10, 2021. The classification was applicable on April 12, 2019.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    April Marrone, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. G218, Silver Spring, MD 20993-0002, 240-402-6510, April.Marrone@fda.hhs.gov.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    I. Background

    Upon request, FDA has classified the ingested, transient, space occupying device for weight management and/or weight loss as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness. In addition, we believe this action will enhance patients' access to beneficial innovation by placing the device into a lower device class than the automatic class III assignment.

    The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).

    FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate device by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).

    FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act. Section 207 of the Food and Drug Administration Modernization Act of 1997 established the first procedure for De Novo classification (Pub. L. 105-115). Section 607 of the Food and Drug Administration Safety and Innovation Act modified the De Novo application process by adding a second procedure (Pub. L. 112-144). A device sponsor may utilize either procedure for De Novo classification.

    Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).

    Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.

    Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device.

    When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application to market a substantially equivalent device (see 21 U.S.C. 360c(i), defining “substantial equivalence”). Instead, sponsors can use the less-burdensome 510(k) process, when necessary, to market their device.

    II. De Novo Classification

    On November 15, 2018, Gelesis, Inc. submitted a request for De Novo classification of the ingested, transient, space occupying device for weight management and/or weight loss. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.

    We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the Start Printed Page 70372 establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.

    Therefore, on April 12, 2019, FDA issued an order to the requester classifying the device into class II. In this final order, FDA is codifying the classification of the device by adding 21 CFR 876.5982.[1] We have named the generic type of device ingested, transient, space occupying device for weight management and/or weight loss, and it is identified as an ingested material that transiently occupies space in the stomach. The device passes from the body via the natural gastrointestinal tract.

    FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.

    Table 1—Ingested, Transient, Space Occupying Device for Weight Management and/or Weight Loss Risks and Mitigation Measures

    Identified risksMitigation measures
    Device related gastrointestinal adverse events, including: • Obstruction • DilationClinical performance testing, Non-clinical performance testing, Labeling, and Shelf life testing.
    • Diarrhea
    • Constipation
    • Dehydration
    Weight gainClinical performance testing, and Labeling.
    Interaction with medicationClinical performance testing, Non-clinical performance testing, and Labeling.
    Adverse tissue reactionBiocompatibility evaluation.
    InfectionNon-clinical performance testing, and Shelf life testing.

    FDA has determined that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of safety and effectiveness. For a device to fall within this classification, and thus avoid automatic classification in class III, it would have to comply with the special controls named in this final order. The necessary special controls appear in the regulation codified by this order. This device is subject to premarket notification requirements under section 510(k).

    At the time of classification, ingested, transient, space occupying devices for weight management and/or weight loss are for prescription use only. Prescription devices are exempt from the requirement for adequate directions for use for the layperson under section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR 801.5, as long as the conditions of 21 CFR 801.109 are met.

    III. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

    IV. Paperwork Reduction Act of 1995

    This final order establishes special controls that refer to previously approved collections of information found in other FDA regulations and guidance. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections of information in the guidance document “De Novo Classification Process (Evaluation of Automatic Class III Designation)” have been approved under OMB control number 0910-0844; the collections of information in 21 CFR part 814, subparts A through E, regarding premarket approval, have been approved under OMB control number 0910-0231; the collections of information in part 807, subpart E, regarding premarket notification submissions, have been approved under OMB control number 0910-0120; the collections of information in 21 CFR part 820, regarding quality system regulation, have been approved under OMB control number 0910-0073; and the collections of information in 21 CFR part 801 regarding labeling, have been approved under OMB control number 0910-0485.

    Start List of Subjects

    List of Subjects in 21 CFR Part 876

    • Medical devices
    End List of Subjects

    Therefore, under the Federal Food, Drug, and Cosmetic Act, and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 876 is amended as follows:

    Start Part

    PART 876—GASTROENTEROLOGY-UROLOGY DEVICES

    End Part Start Amendment Part

    1. The authority citation for part 876 continues to read as follows:

    End Amendment Part Start Authority

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

    End Authority Start Amendment Part

    2. Add § 876.5982 to subpart F to read as follows:

    End Amendment Part
    Ingested, transient, space occupying device for weight management and/or weight loss.

    (a) Identification. This device is an ingested material that transiently occupies space in the stomach. The device passes from the body via the natural gastrointestinal tract.

    (b) Classification. Class II (special controls). The special controls for this device are:

    (1) The patient-contacting components of the device must be demonstrated to be biocompatible for its intended use.

    (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, as follows:

    (i) Performance bench testing in a simulated use model must evaluate device disintegration and device Start Printed Page 70373 hydration state throughout the gastrointestinal tract;

    (ii) Bioburden and moisture content assessments must evaluate device infection risk throughout the labeled shelf life; and

    (iii) Performance data must support the shelf life of the device by demonstrating continued package integrity and device functionality over the labeled shelf life.

    (3) Clinical performance testing must demonstrate the device performs as intended and evaluate the following:

    (i) Weight change;

    (ii) All adverse events, including obstruction, dilation, diarrhea, constipation, and dehydration; and

    (iii) Interaction with representative medications.

    (4) Physician and patient device labeling must state:

    (i) The clinical benefit of the device as assessed by using percent total body weight loss;

    (ii) Treatment must be offered in combination with diet and exercise;

    (iii) Instructions on how to use the device as intended including how to avoid interaction with medication; and

    (iv) The shelf life of the device.

    Start Signature

    Dated: December 6, 2021.

    Lauren K. Roth,

    Associate Commissioner for Policy.

    End Signature End Supplemental Information

    Footnotes

    1.  FDA notes that the “ACTION” caption for this final order is styled as “Final amendment; final order,” rather than “Final order.” Beginning in December 2019, this editorial change was made to indicate that the document “amends” the Code of Federal Regulations. The change was made in accordance with the Office of Federal Register's (OFR) interpretations of the Federal Register Act (44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and parts 21 and 22), and the Document Drafting Handbook.

    Back to Citation

    [FR Doc. 2021-26738 Filed 12-9-21; 8:45 am]

    BILLING CODE 4164-01-P

Document Information

Effective Date:
12/10/2021
Published:
12/10/2021
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final amendment; final order.
Document Number:
2021-26738
Dates:
This order is effective December 10, 2021. The classification was applicable on April 12, 2019.
Pages:
70371-70373 (3 pages)
Docket Numbers:
Docket No. FDA-2021-N-0585
Topics:
Medical devices
PDF File:
2021-26738.pdf
Supporting Documents:
» Medical Devices; Gastroenterology-Urology Devices; Classification of the Ingested, Transient, Space Occupying Device for Weight Management and/or Weight Loss
CFR: (1)
21 CFR 876.5982