2022-04154. Medical Devices; Orthopedic Devices; Classification of the Screw Sleeve Bone Fixation Device  

  • Start Preamble

    AGENCY:

    Food and Drug Administration, HHS.

    ACTION:

    Final amendment; final order.

    SUMMARY:

    The Food and Drug Administration (FDA, Agency, or we) is classifying the screw sleeve bone fixation device into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the screw sleeve bone fixation device's classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices.

    DATES:

    This order is effective March 1, 2022. The classification was applicable on May 1, 2020.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    Jesse Muir, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 4508, Silver Spring, MD 20993-0002, 240-402-6679, Jesse.Muir@fda.hhs.gov.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    I. Background

    Upon request, FDA has classified the screw sleeve bone fixation device as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness. In addition, we believe this action will enhance patients' access to beneficial innovation, by placing the device into a lower device class than the automatic class III assignment.

    The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).

    FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate device by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).

    FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act. Section 207 of the Food and Drug Administration Modernization Act of 1997 (Pub. L. 105-115) established the first procedure for De Novo classification. Section 607 of the Food and Drug Administration Safety and Innovation Act (Pub. L. 112-144) modified the De Novo application process by adding a second procedure. A device sponsor may utilize either procedure for De Novo classification.

    Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).

    Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.

    Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device.

    When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C Act). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application to market a substantially equivalent device (see section 513(i) of the FD&C Act, defining “substantial equivalence”). Instead, sponsors can use the less-burdensome 510(k) process, when necessary, to market their device.

    II. De Novo Classification

    On December 13, 2018, FDA received Woven Orthopedic Technologies, LLC's request for De Novo classification of the OGmend® Implant System. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.

    We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.

    Therefore, on May 1, 2020, FDA issued an order to the requester classifying the device into class II. In this final order, FDA is codifying the classification of the device by adding 21 CFR 888.3043.[1] We have named the generic type of device “screw sleeve bone fixation device,” and it is intended to be implanted in conjunction with a non-resorbable, metallic bone screw where the screw has lost purchase due to loosening, backout, or breakage. The device fits between the screw threads and surrounding bone and provides increased surface area to create an Start Printed Page 11294 interference fit to restore stability of the implant construct.

    FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.

    Table 1—Screw Sleeve Bone Fixation Device Risks and Mitigation Measures

    Identified risksMitigation measures
    Loss of function/mechanical integrity resulting from: Device malposition. Device breakage. Damage to screw during insertion. Deterioration due to aging. Insufficient restoration of screw fixation.In vivo performance testing; Non-clinical performance testing; Shelf life testing; and Labeling.
    RevisionIn vivo performance testing; Non-clinical performance testing; and Labeling.
    Adverse tissue reactionBiocompatibility evaluation; In vivo performance testing; Non-clinical performance testing; and Labeling.
    InfectionSterilization validation; and Shelf life testing.
    Febrile response due to endotoxinsPyrogenicity testing.

    FDA has determined that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of safety and effectiveness. For a device to fall within this classification, and thus avoid automatic classification in class III, it would have to comply with the special controls named in this final order. The necessary special controls appear in the regulation codified by this order. We encourage sponsors to consult with us if they wish to use a non-animal testing method that they believe is suitable, adequate, validated, and feasible. We will consider if such an alternative method could be assessed for equivalency to an animal test method. This device is subject to premarket notification requirements under section 510(k) of the FD&C Act.

    III. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

    IV. Paperwork Reduction Act of 1995

    This final order establishes special controls that refer to previously approved collections of information found in other FDA regulations and guidance. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections of information in 21 CFR part 860, subpart D, regarding De Novo classification have been approved under OMB control number 0910-0844; the collections of information in 21 CFR part 814, subparts A through E, regarding premarket approval, have been approved under OMB control number 0910-0231; the collections of information in part 807, subpart E, regarding premarket notification submissions, have been approved under OMB control number 0910-0120; the collections of information in 21 CFR part 820, regarding quality system regulation, have been approved under OMB control number 0910-0073; and the collections of information in 21 CFR part 801, regarding labeling, have been approved under OMB control number 0910-0485.

    Start List of Subjects

    List of Subjects in 21 CFR Part 888

    • Medical devices
    End List of Subjects

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 888 is amended as follows:

    Start Part

    PART 888—ORTHOPEDIC DEVICES

    End Part Start Amendment Part

    1. The authority citation for part 888 continues to read as follows:

    End Amendment Part Start Authority

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

    End Authority Start Amendment Part

    2. Add § 888.3043 to subpart D to read as follows:

    End Amendment Part
    Screw sleeve bone fixation device.

    (a) Identification. A screw sleeve bone fixation device is intended to be implanted in conjunction with a non-resorbable, metallic bone screw where the screw has lost purchase due to loosening, backout, or breakage. The device fits between the screw threads and surrounding bone and provides increased surface area to create an interference fit to restore stability of the implant construct.

    (b) Classification. Class II (special controls). The special controls for this device are:

    (1) In vivo performance testing under anticipated conditions of use must demonstrate:

    (i) The device provides sufficient stability to allow for fracture healing; and

    (ii) A lack of adverse biologic response to the implant through histopathological and histomorphometric assessment.

    (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must:

    (i) Assess the stability of the device in a rescue screw scenario;

    (ii) Demonstrate that the device can be inserted and removed without damage to the implant or associated hardware;

    (iii) Demonstrate the device can withstand dynamic loading without device failure; and

    (iv) Characterize wear particle generation.

    (3) The device must be demonstrated to be biocompatible.

    (4) The device must be demonstrated to be non-pyrogenic.

    (5) Performance data must demonstrate the sterility of the device.

    (6) Performance data must support the labeled shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the established shelf life.

    (7) Labeling must include:

    (i) A detailed summary of the device technical parameters;

    (ii) Information describing all materials of the device;

    (iii) Instructions for use, including device removal; and

    (iv) A shelf life.

    Start Signature
    Start Printed Page 11295

    Dated: February 22, 2022.

    Lauren K. Roth,

    Associate Commissioner for Policy.

    End Signature End Supplemental Information

    Footnotes

    1.  FDA notes that the “ACTION” caption for this final order is styled as “Final amendment; final order,” rather than “Final order.” Beginning in December 2019, this editorial change was made to indicate that the document “amends” the Code of Federal Regulations. The change was made in accordance with the Office of Federal Register's (OFR) interpretations of the Federal Register Act (44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and parts 21 and 22), and the Document Drafting Handbook.

    Back to Citation

    [FR Doc. 2022-04154 Filed 2-28-22; 8:45 am]

    BILLING CODE 4164-01-P

Document Information

Effective Date:
3/1/2022
Published:
03/01/2022
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final amendment; final order.
Document Number:
2022-04154
Dates:
This order is effective March 1, 2022. The classification was applicable on May 1, 2020.
Pages:
11293-11295 (3 pages)
Docket Numbers:
Docket No. FDA-2022-N-0114
Topics:
Medical devices
PDF File:
2022-04154.pdf
Supporting Documents:
» Medical Devices; Orthopedic Devices; Classification of the Screw Sleeve Bone Fixation Device
CFR: (1)
21 CFR 888.3043