2014-11174. Medical Devices; General Hospital and Personal Use Devices; Classification of the Intravascular Administration Set, Automated Air Removal System
-
Start Preamble
AGENCY:
Food and Drug Administration, HHS.
ACTION:
Final order.
SUMMARY:
The Food and Drug Administration (FDA) is classifying the intravascular administration set, automated air removal system into class II (special controls). The special controls that will apply to the device are identified in this order and will be part of the codified language for the intravascular administration set, automated air removal system's classification. The Agency is classifying the device into class II (special controls) in order to provide a reasonable assurance of safety and effectiveness of the device.
DATES:
This order is effective June 16, 2014. The classification was effective on March 4, 2014.
Start Further InfoFOR FURTHER INFORMATION CONTACT:
Alan Stevens, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave. Bldg. 66, Rm. 2561, Silver Spring, MD 20993-0002, 301-796-6294.
End Further Info End Preamble Start Supplemental InformationSUPPLEMENTARY INFORMATION:
I. Background
In accordance with section 513(f)(1) of the Federal Food, Drug, and Cosmetic Act (the FD&C Act) (21 U.S.C. 360c(f)(1)), devices that were not in commercial distribution before May 28, 1976 (the date of enactment of the Medical Device Amendments of 1976), generally referred to as postamendments devices, are classified automatically by statute into class III without any FDA rulemaking process. These devices remain in class III and require premarket approval, unless and until the device is classified or reclassified into class I or II, or FDA issues an order finding the device to be substantially equivalent, in accordance with section 513(i), to a predicate device that does not require premarket approval. The Agency determines whether new devices are substantially equivalent to predicate devices by means of premarket notification procedures in section 510(k) of the FD&C Act (21 Start Printed Page 28405U.S.C. 360(k)) and part 807 (21 CFR part 807) of the regulations.
Section 513(f)(2) of the FD&C Act, as amended by section 607 of the Food and Drug Administration Safety and Innovation Act (Pub. L. 112-144, July 9, 2012), provides two procedures by which a person may request FDA to classify a device under the criteria set forth in section 513(a)(1). Under the first procedure, the person submits a premarket notification under section 510(k) for a device that has not previously been classified and, within 30 days of receiving an order classifying the device into class III under section 513(f)(1), the person requests a classification under section 513(f)(2). Under the second procedure, rather than first submitting a premarket notification under section 510(k) and then a request for classification under the first procedure, the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence and requests a classification under section 513(f)(2). If the person submits a request to classify the device under this second procedure, FDA may decline to undertake the classification request if FDA identifies a legally marketed device that could provide a reasonable basis for review of substantial equivalence with the device or if FDA determines that the device submitted is not of “low-moderate risk” or that general controls would be inadequate to control the risks and special controls to mitigate the risks cannot be developed.
In response to a request to classify a device under either procedure provided by section 513(f)(2) of the FD&C Act, FDA will classify the device by written order within 120 days. This classification will be the initial classification of the device.
In accordance with section 513(f)(1) of the FD&C Act, FDA issued an order on October 23, 2008, classifying the AirPurge System into class III, because it was not substantially equivalent to a device that was introduced or delivered for introduction into interstate commerce for commercial distribution before May 28, 1976, or a device which was subsequently reclassified into class I or class II. On October 29, 2008, Anesthesia Safety Products, LLC submitted a request requesting classification of the AirPurge System under section 513(f)(2) of the FD&C Act. The manufacturer recommended that the device be classified into class II (Ref. 1).
In accordance with section 513(f)(2) of the FD&C Act, FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act. FDA classifies devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls to provide reasonable assurance of the safety and effectiveness of the device for its intended use. After review of the information submitted in the request, FDA determined that the device can be classified into class II with the establishment of special controls. FDA believes these special controls, in addition to general controls, will provide reasonable assurance of the safety and effectiveness of the device.
Therefore, on March 4, 2014, FDA issued an order to the requestor classifying the device into class II. FDA is codifying the classification of the device by adding 21 CFR 880.5445.
Following the effective date of this final classification order, any firm submitting a premarket notification (510(k)) for an intravascular administration set, automated air removal system will need to comply with the special controls named in this final order. The device is assigned the generic name intravascular administration set, automated air removal system, and it is identified as a prescription device used to detect and automatically remove air from an intravascular administration set with minimal to no interruption in the flow of the intravascular fluid. The device may include an air identification mechanism, software, an air removal mechanism, tubing, apparatus to collect removed air, and safety control mechanisms to address hazardous situations.
FDA has identified the following risks to health associated specifically with this type of device, as well as the mitigation measures required to mitigate these risks.
Table 1—Identified Risks and Mitigation Measures
Identified risk Mitigation measures Embolus Hazard Argument. Software. Electromagnetic Compatibility. Human Factors. Labeling. Nonclinical Performance Testing. Infusion Delivery Error Hazard Argument. Software. Electromagnetic Compatibility. Human Factors. Labeling. Nonclinical Performance Testing. Electric Shock Hazard Argument. Electrical Safety. Electromagnetic Compatibility. Adverse Tissue Reaction Hazard Argument. Biocompatibility. Infection Sterilization. Shelf Life. FDA believes that the following special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of the safety and effectiveness:
1. Provide an argument demonstrating that all reasonably foreseeable hazards have been adequately addressed with respect to the persons for whose use the device is represented or intended and Start Printed Page 28406the conditions of use for the device, which includes the following:
- Description of the device indications for use, design, and technology, use environments, and users in sufficient detail to determine that the device complies with all special controls.
- Demonstrate that controls are implemented to address device system hazards and their causes.
- Include a justification supporting the acceptability criteria for each hazard control.
- A traceability analysis demonstrating that all credible hazards have at least one corresponding control and that all controls have been verified and validated in the final device design.
2. Appropriate software verification, validation, and hazard analysis must be performed.
3. The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible.
4. Performance data must demonstrate the sterility of fluid path contacting components and the shelf life of these components.
5. The device must be designed and tested for electrical safety and electromagnetic compatibility (EMC).
6. Nonclinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
- Device system and component reliability testing must be conducted.
- Fluid ingress protection testing must be conducted.
- Testing of safety controls must be performed to demonstrate adequate mitigation of hazardous situations, including sensor failure, flow control failure, improper device position, device malfunction, infusion delivery error, and release of air to the patient.
7. A human factors validation study must demonstrate that use hazards are adequately addressed.
8. The labeling must include the following:
- The device's air identification and removal response time.
- The device's minimum air volume identification sensitivity.
- The minimum and maximum flow rates at which the device is capable of reliably detecting and removing air.
- Quantification of any fluid loss during device air removal operations as a function of flow rate.
Intravascular administration set, automated air removal systems are prescription devices restricted to patient use only upon the authorization of a practitioner licensed by law to administer or use the device (21 CFR 880.5445(a); see section 520(e) of the FD&C Act (21 U.S.C. 360j(e)) and 21 CFR 801.109 (Prescription devices.)). Prescription-use restrictions are a type of general controls as defined in section 513(a)(1)(A)(i) of the FD&C Act.
Section 510(m) of the FD&C Act provides that FDA may exempt a class II device from the premarket notification requirements under section 510(k) of the FD&C Act, if FDA determines that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device. For this type of device, FDA has determined that premarket notification is necessary to provide reasonable assurance of the safety and effectiveness of the device. Therefore, this device type is not exempt from premarket notification requirements. Persons who intend to market this type of device must submit to FDA a premarket notification, prior to marketing the device, which contains information about the intravascular administration set, automated air removal system they intend to market.
II. Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.
III. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to previously approved collections of information found in other FDA regulations. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections of information in part 807, subpart E, regarding premarket notification submissions have been approved under OMB control number 0910-0120, and the collections of information in 21 CFR part 801, regarding labeling have been approved under OMB control number 0910-0485.
IV. Reference
The following reference has been placed on display in the Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852, and may be seen by interested persons between 9 a.m. and 4 p.m., Monday through Friday, and is available electronically at http://www.regulations.gov.
1. K080644: De Novo Request per 513(f)(2) pursuant to the Agency's not substantially equivalent (NSE) determination, dated October 23, 2008, from Anesthesia Safety Products, LLC, dated October 29, 2008.
Start List of SubjectsList of Subjects in 21 CFR Part 880
- Medical devices
Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 880 is amended as follows:
Start PartPART 880—GENERAL HOSPITAL AND PERSONAL USE DEVICES
End Part Start Amendment Part1. The authority citation for 21 CFR part 880 continues to read as follows:
End Amendment Part Start Amendment Part2. Add § 880.5445 to subpart F to read as follows:
End Amendment PartIntravascular Administration Set, Automated Air Removal System.(a) Identification. An intravascular administration set, automated air removal system, is a prescription device used to detect and automatically remove air from an intravascular administration set with minimal to no interruption in the flow of the intravascular fluid. The device may include an air identification mechanism, software, an air removal mechanism, tubing, apparatus to collect removed air, and safety control mechanisms to address hazardous situations.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) Provide an argument demonstrating that all reasonably foreseeable hazards have been adequately addressed with respect to the persons for whose use the device is represented or intended and the conditions of use for the device, which includes the following:
(i) Description of the device indications for use, design, and technology, use environments, and users in sufficient detail to determine that the device complies with all special controls.
(ii) Demonstrate that controls are implemented to address device system hazards and their causes.
(iii) Include a justification supporting the acceptability criteria for each hazard control.
(iv) A traceability analysis demonstrating that all credible hazards have at least one corresponding control and that all controls have been verified and validated in the final device design.Start Printed Page 28407
(2) Appropriate software verification, validation, and hazard analysis must be performed.
(3) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible.
(4) Performance data must demonstrate the sterility of fluid path contacting components and the shelf life of these components.
(5) The device must be designed and tested for electrical safety and electromagnetic compatibility (EMC).
(6) Nonclinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Device system and component reliability testing must be conducted.
(ii) Fluid ingress protection testing must be conducted.
(iii) Testing of safety controls must be performed to demonstrate adequate mitigation of hazardous situations, including sensor failure, flow control failure, improper device position, device malfunction, infusion delivery error, and release of air to the patient.
(7) A human factors validation study must demonstrate that use hazards are adequately addressed.
(8) The labeling must include the following:
(i) The device's air identification and removal response time.
(ii) The device's minimum air volume identification sensitivity.
(iii) The minimum and maximum flow rates at which the device is capable of reliably detecting and removing air.
(iv) Quantification of any fluid loss during device air removal operations as a function of flow rate.
Dated: May 9, 2014.
Leslie Kux,
Assistant Commissioner for Policy.
[FR Doc. 2014-11174 Filed 5-15-14; 8:45 am]
BILLING CODE 4160-01-P
Document Information
- Effective Date:
- 6/16/2014
- Published:
- 05/16/2014
- Department:
- Food and Drug Administration
- Entry Type:
- Rule
- Action:
- Final order.
- Document Number:
- 2014-11174
- Dates:
- This order is effective June 16, 2014. The classification was effective on March 4, 2014.
- Pages:
- 28404-28407 (4 pages)
- Docket Numbers:
- Docket No. FDA-2014-N-0438
- Topics:
- Medical devices
- PDF File:
- 2014-11174.pdf
- CFR: (1)
- 21 CFR 880.5445