    Part VI
    
    
    
    
    
    Department of Health and Human Services
    
    
    
    
    
    _______________________________________________________________________
    
    
    
    Food and Drug Administration
    
    
    
    _______________________________________________________________________
    
    
    
    21 CFR Part 1020
    
    
    
    Federal Performance Standard for Diagnostic X-Ray Systems and Their 
    Major Components; Final Rule
     
    
    DEPARTMENT OF HEALTH AND HUMAN SERVICES
    
    Food and Drug Administration
    
    21 CFR Part 1020
    
    [Docket No. 92N-0108]
    
    Federal Performance Standard for Diagnostic X-Ray Systems and Their 
    Major Components
    
    AGENCY: Food and Drug Administration, HHS.
    
    ACTION: Final rule.
    
    -----------------------------------------------------------------------
    
    SUMMARY: The Food and Drug Administration (FDA) is issuing a final rule 
    to amend the Federal performance standard for diagnostic x-ray systems 
    and their major components (the performance standard). The amendments 
    revise the limits established for maximum patient entrance exposure 
    rates for fluoroscopic x-ray systems during high-level control and 
    other modes of operation. The revisions are being made due to concerns 
    regarding excessively large radiation exposure rates on some 
    fluoroscopic x-ray systems. In addition, an inadvertent error is being 
    corrected to improve the accuracy of the regulations.
    
    EFFECTIVE DATE: May 19, 1995.
    
    FOR FURTHER INFORMATION CONTACT: Patricia M. Dubill, Center for Devices 
    and Radiological Health (HFZ-84), Food and Drug Administration, 5600 
    Fishers Lane, Rockville, MD 20857, 301-594-4765.
    
    SUPPLEMENTARY INFORMATION:
    
    I. Background
    
        Under the Federal Food, Drug, and Cosmetic Act (21 U.S.C. 321 et 
    seq.) as amended, FDA is amending the performance standard for 
    diagnostic x-ray systems and their major components in Sec. 1020.32 (21 
    CFR 1020.32) by revising the limits established for maximum patient 
    entrance exposure rates for fluoroscopic x-ray systems during high-
    level control and other modes of operation. In the Federal Register of 
    August 15, 1972 (37 FR 16461), the performance standard was published 
    as a final rule which became effective on August 1, 1974. Since that 
    time, there have been several amendments to the performance standard. 
    These amendments take into account new technology, clarify 
    misinterpreted provisions, or incorporate additional requirements 
    determined to be necessary to provide for adequate radiation safety of 
    diagnostic x-ray systems.
        FDA determined that further changes to the standard were necessary, 
    in view of concerns regarding the potential for unnecessary radiation 
    exposures to patients from some current fluoroscopic x-ray system 
    designs. Therefore, FDA developed proposed draft amendments to the 
    performance standard, which were discussed during a public meeting of 
    the Technical Electronic Product Radiation Safety Standards Committee 
    (TEPRSSC) on November 14, 1990. The proposed amendments that were 
    discussed included the following:
        (1) Replacement of the current two-tier system of entrance exposure 
    rate limits applicable during normal fluoroscopy on systems with 
    automatic exposure rate control (AERC) with a single upper limit of 
    2.58x10-3 coulomb per kilogram (C/kg) per minute (10 roentgens per 
    minute (R/min)), regardless of mode of operation;
        (2) Establishment of an upper limit of 5.16x10-3 C/kg per 
    minute (20 R/min) on entrance exposure rate during high-level control 
    mode of fluoroscopy;
        (3) Establishment of a requirement for an AERC mode of operation 
    for any fluoroscopic x-ray system capable of operating at an entrance 
    exposure rate greater than 1.29x10-3 C/kg per minute (5 R/min); 
    and
        (4) Application of the entrance exposure rate limits discussed in 
    paragraphs (1) and (2) above during continuous recording of dynamic 
    fluoroscopic images with a videotape recorder or similar device. The 
    application of the entrance exposure rate limit during continuous 
    recording requires a distinction between the continuous mode of 
    operation and various modes of operation in which the x-ray exposure is 
    pulsed. Therefore, a definition of pulsed mode of operation is being 
    added. This is necessary because a consensus as to appropriate limits 
    for the various pulsed recording modes does not exist and further study 
    is needed before such a limit is imposed.
        TEPRSSC approved the content of the proposed amendments and 
    concurred with publication of them for public comment. FDA proposed the 
    amendments to the performance standard in the Federal Register of May 
    3, 1993 (58 FR 26407). Interested persons were given until August 2, 
    1993, to comment on the proposal. FDA received and considered one 
    comment each from a radiological physicist and a professional 
    organization representing radiologists and physicists. Both comments 
    supported the proposed changes and suggested several additional 
    amendments to the standard. These comments also addressed fluoroscopy-
    related issues on which FDA specifically solicited comments in the 
    preamble to the proposed rule. FDA appreciates having received these 
    comments and will consider them in the context of possible future 
    rulemaking activities related to fluoroscopic x-ray systems. A summary 
    of the comments pertaining to the proposal of May 3, 1993, and FDA's 
    responses are discussed below.
         In addition, in the final rule that published in the Federal 
    Register of May 3, 1993 (58 FR 26386 at 26398), in Sec. 1020.30(b) (21 
    CFR 1020.30(b), the acronym ``SID'' for source-image receptor distance 
    was inadvertently omitted. This document is also correcting that error.
    
    II. Summary and Analysis of Comments
    
        1. Both of the comments received concurred with all of the proposed 
    amendments to the performance standard, except that one of the comments 
    suggested changing the proposed upper limit on entrance exposure rate 
    during the high-level control mode of operation. That comment advocated 
    a limit of 3.87x10-3 C/kg per minute (15 R/min), rather than the 
    proposed limit of 5.16x10-3 C/kg per minute (20 R/min), but noted 
    a lack of objective data to support the recommendation. The other 
    comment concurred with the proposed limit of 5.16x10-3 C/kg per 
    minute (20 R/min) and noted that a number of manufacturers provide 
    equipment which does not exceed 2.58x10-3 C/kg per minute (10 R/
    min) during fluoroscopy.
        The comment advocating a lower limit did not submit specific data 
    supporting that position. The second comment on the appropriate upper 
    limit to be imposed on entrance exposure rate during high-level control 
    mode of operation noted that a lower limit was consistent with certain 
    existing equipment, but nonetheless supported the limit proposed by 
    FDA. After reviewing these comments, FDA concludes that its choice of 
    an upper limit is a reasonable one. The objective in establishing such 
    a limit is to provide a mode of operation that will reasonably 
    accommodate unusual imaging requirements, but also will prevent 
    unnecessarily high exposure rates. The limit proposed was based on the 
    observation that the majority of fluoroscopic x-ray systems equipped 
    with high-level control, which were evaluated in conjunction with FDA 
    testing for compliance with the performance standard, are limited to 
    maximum entrance exposure rates of less than 5.16x10-3 C/kg per 
    minute (20 R/min). The proposed limit was also based on discussions 
    with medical physicists regarding their experience in limiting entrance 
    exposure rates, which indicated that a limit of 5.16x10-3 C/kg per 
    minute (20 R/min) eliminates unnecessarily high exposure rates without 
    compromising the use of the equipment. FDA interprets the lack of 
    significant adverse comment on this proposal as an indication of 
    general agreement with the proposed limit. Therefore, FDA rejects the 
    suggestion that a limit of 3.87x10-3 C/kg per minute (15 R/min) be 
    established for the maximum entrance exposure rate, and is establishing 
    the upper limit of 5.16x10-3 C/kg per minute (20 R/min) as 
    proposed.
        FDA notes that recommendations related to limits on entrance 
    exposure rate resulted from a workshop on fluoroscopy cosponsored by 
    FDA and the American College of Radiology (ACR) in October of 1992 
    (Ref. 2). These recommendations generally favored limiting exposure 
    rates from fluoroscopic x-ray systems in some manner and included 
    suggestions for limiting the maximum exposure rate to the input surface 
    of the image intensifier tube. FDA has chosen not to include these 
    suggestions in this final rule because they were beyond the scope of 
    the proposal. FDA will carefully consider the recommendations from the 
    workshop related to limiting exposure rates in conjunction with future 
    amendments.
        2. One of the comments suggested that the requirement for a 
    continuous audible signal during the high-level control mode of 
    operation be modified to specify that the signal be emitted at the 
    fluoroscopic imaging assembly (at an FDA-specified volume level), to 
    ensure that the signal is audible to the fluoroscopist.
        The current requirement for a continuous audible signal is that the 
    signal must be audible to the fluoroscopist, which would seem to meet 
    the intent of the comment. As with any medical device alarm or signal, 
    the volume level of the signal should be designed to be sufficiently 
    loud to alert the operator to the status of the device, taking into 
    account any background noise likely to be present during use. In 
    addition, the effect of an overly loud signal on the patient should be 
    considered in the signal design. FDA disagrees that the agency should 
    specify an emission location or a particular volume level for the 
    signal, but encourages manufacturers, in the design of their systems, 
    to consult available human factors guidelines for audio (and visual) 
    characteristics of medical device alarms and signals (e.g., Ref. 1), 
    and to test and evaluate their systems under representative operational 
    conditions.
        3. One of the comments referred to the workshop on fluoroscopy 
    discussed in comment 1 above, and noted a number of additional 
    suggestions for changes to fluoroscopic x-ray system performance that 
    were advocated by attendees at that workshop (Ref. 2).
        Those additional suggestions for changes are beyond the scope of 
    the current proposal and would require a separate proposal and 
    opportunity for comment. FDA agrees that the recommendations from the 
    workshop should be considered as the basis for possible future 
    amendments to the standard.
    
    III. Environmental Impact
    
        The agency has determined under 21 CFR 25.24(e)(3) that this action 
    is of a type that does not individually or cumulatively have a 
    significant effect on the human environment. Therefore, neither an 
    environmental assessment nor an environmental impact statement is 
    required.
    
    IV. Analysis of Impacts
    
         An analysis of the costs and benefits of this regulation, 
    conducted under Executive Order 12291, was discussed in the proposed 
    rule and the substance of that analysis has not changed. Executive 
    Order 12291 has been superseded by Executive Order 12866. FDA has 
    examined the impacts of this final rule under Executive Order 12866 and 
    the Regulatory Flexibility Act (Pub. L. 96-354). Executive Order 12866 
    directs agencies to assess all costs and benefits of available 
    regulatory alternatives and, when regulation is necessary, to select 
    regulatory approaches that maximize net benefits (including potential 
    economic, environmental, public health and safety, and other 
    advantages; distributive impacts; and equity). The agency believes that 
    this rule is consistent with the regulatory philosophy and principles 
    identified in the Executive Order. In addition, the rule is not a 
    significant regulatory action as defined by the Executive Order and so 
    is not subject to review under the Executive Order.
         The Regulatory Flexibility Act requires agencies to analyze 
    regulatory options that would minimize any significant impact of a rule 
    on small entities. This amendment to the performance standard will 
    affect the small fraction of fluoroscopic systems which are currently 
    provided with high-level controls or which do not provide AERC, and 
    have radiation outputs exceeding 1.29x10-3 C/kg per minute (5 R/
    min). Further, the additional testing performed by x-ray manufacturers 
    and government regulatory progams to ensure compliance with the 
    performance standard, and the quality assurance testing performed by 
    facilities can easily be included in current testing programs with 
    minimal additional costs. The agency certifies that the rule will not 
    have a significant economic impact on a substantial number of small 
    entities. Therefore, under the Regulatory Flexibility Act, no further 
    analysis is required.
         A copy of the threshold assessment supporting this determination 
    is on file with the Dockets Management Branch (address above) and may 
    be seen in that office between 9 a.m.and 4 p.m., Monday through Friday.
    
    V. References
    
        The following references have been placed on display in the Dockets 
    Management Branch (address above) and may be seen by interested persons 
    from 9 a.m. to 4 p.m., Monday through Friday.
        1. ``Human Factors Engineering Guidelines and Preferred 
    Practices for the Design of Medical Devices,'' AAMI HE48-R-93, 2d 
    ed., Association for the Advancement of Medical Instrumentation, 
    3330 Washington Blvd., suite 400, Arlington, VA 22201-4598, 1993.
        2. ``Proceedings of the ACR/FDA Workshop on Fluoroscopy,'' ACR, 
    October 1992.
    
    List of Subjects in 21 CFR Part 1020
    
         Electronic products, Medical devices, Radiation protection, 
    Reporting and recordkeeping requirements, Television, X-rays.
        Therefore, under the Federal Food, Drug, and Cosmetic Act and under 
    authority delegated to the Commissioner of Food and Drugs, 21 CFR part 
    1020 is amended as follows:
    
    PART 1020--PERFORMANCE STANDARDS FOR IONIZING RADIATION EMITTING 
    PRODUCTS
    
        1. The authority citation for 21 CFR part 1020 continues to read as 
    follows:
    
        Authority: Secs. 501, 502, 515-520, 530-542, 701, 801 of the 
    Federal Food, Drug, and Cosmetic Act (21 U.S.C. 351, 352, 360e-360j, 
    360gg-360ss, 371, 381).
    
        2. Section 1020.30 is amended in paragraph (b) by alphabetically 
    adding a new definition for ``Pulsed mode'' and by revising the 
    definition for ``Source-image receptor distance'' to read as follows:
    
    
    Sec. 1020.30  Diagnostic x-ray systems and their major components.
    
    * * * * *
        (b) Definitions. * * *
    * * * * *
        Pulsed mode means operation of the x-ray system such that the x-ray 
    tube current is pulsed by the x-ray control to produce one or more 
    exposure intervals of duration less than one-half second.
    * * * * *
        Source-image receptor distance (SID) means the distance from the 
    source to the center of the input surface of the image receptor.
    * * * * *
        3. Section 1020.32 is amended by revising paragraph (d), by 
    redesignating paragraphs (e) through (h) as paragraphs (f) through (i), 
    respectively, and by adding new paragraph (e) to read as follows:
    
    
    Sec. 1020.32  Fluoroscopic equipment.
    
    * * * * *
        (d) Entrance exposure rates. For fluoroscopic equipment 
    manufactured before May 19, 1995, the following requirements apply:
        (1) Equipment with automatic exposure rate control (AERC). 
    Fluoroscopic equipment that is provided with AERC shall not be operable 
    at any combination of tube potential and current that will result in an 
    exposure rate in excess of 2.58x10-3 coulomb per kilogram (C/kg) 
    per minute (10 roentgens per minute (10 R/min)) at the point where the 
    center of the useful beam enters the patient, except:
        (i) During recording of fluoroscopic images, or
        (ii) When an optional high-level control is provided. When so 
    provided, the equipment shall not be operable at any combination of 
    tube potential and current that will result in an exposure rate in 
    excess of 1.29x10-3 C/kg per minute (5 R/min) at the point where 
    the center of the useful beam enters the patient, unless the high-level 
    control is activated. Special means of activation of high-level 
    controls shall be required. The high-level control shall be operable 
    only when continuous manual activation is provided by the operator. A 
    continuous signal audible to the fluoroscopist shall indicate that the 
    high-level control is being employed.
        (2) Equipment without AERC (manual mode). Fluoroscopic equipment 
    that is not provided with AERC shall not be operable at any combination 
    of tube potential and current that will result in an exposure rate in 
    excess of 1.29x10-3 C/kg per minute (5 R/min) at the point where 
    the center of the useful beam enters the patient, except:
        (i) During recording of fluoroscopic images, or
        (ii) When an optional high-level control is activated. Special 
    means of activation of high-level controls shall be required. The high-
    level control shall be operable only when continuous manual activation 
    is provided by the operator. A continuous signal audible to the 
    fluoroscopist shall indicate that the high-level control is being 
    employed.
        (3) Equipment with both an AERC mode and a manual mode. 
    Fluoroscopic equipment that is provided with both an AERC mode and a 
    manual mode shall not be operable at any combination of tube potential 
    and current that will result in an exposure rate in excess of 
    2.58x10-3 C/kg per minute (10 R/min) in either mode at the point 
    where the center of the useful beam enters the patient except:
        (i) During recording of fluoroscopic images, or
        (ii) When the mode or modes have an optional high-level control, in 
    which case that mode or modes shall not be operable at any combination 
    of tube potential and current that will result in an exposure rate in 
    excess of 1.29x10-3 C/kg per minute (5 R/min) at the point where 
    the center of the useful beam enters the patient, unless the high-level 
    control is activated. Special means of activation of high-level 
    controls shall be required. The high-level control shall be operable 
    only when continuous manual activation is provided by the operator. A 
    continuous signal audible to the fluoroscopist shall indicate that the 
    high-level is being employed.
        (4) Measuring compliance. Compliance with paragraph (d) of this 
    section shall be determined as follows:
        (i) If the source is below the x-ray table, the exposure rate shall 
    be measured at 1 centimeter above the tabletop or cradle.
        (ii) If the source is above the x-ray table, the exposure rate 
    shall be measured at 30 centimeters above the tabletop with the end of 
    the beam-limiting device or spacer positioned as closely as possible to 
    the point of measurement.
        (iii) In a C-arm type of fluoroscope, the exposure rate shall be 
    measured at 30 centimeters from the input surface of the fluoroscopic 
    imaging assembly, with the source positioned at any available SID, 
    provided that the end of the beam-limiting device or spacer is no 
    closer than 30 centimeters from the input surface of the imaging 
    assembly.
        (iv) In a lateral type of fluoroscope, the exposure rate shall be 
    measured at a point 15 centimeters from the centerline of the x-ray 
    table and in the direction of the x-ray source with the end of the 
    beam-limiting device or spacer positioned as closely as possible to the 
    point of measurement. If the tabletop is movable, it shall be 
    positioned as closely as possible to the lateral x-ray source, with the 
    end of the beam-limiting device or spacer no closer than 15 centimeters 
    to the centerline of the x-ray table.
        (5) Exemptions. Fluoroscopic radiation therapy simulation systems 
    are exempt from the requirements set forth in paragraph (d) of this 
    section.
        (e) Entrance exposure rate limits. For fluoroscopic equipment 
    manufactured on and after May 19, 1995, the following requirements 
    apply:
        (1) Fluoroscopic equipment operable at any combination of tube 
    potential and current that results in an exposure rate greater than 
    1.29x10-3 C/kg per minute (5 R/min) at the point where the center 
    of the useful beam enters the patient shall be equipped with AERC. 
    Provision for manual selection of technique factors may be provided.
        (2) Fluoroscopic equipment shall not be operable at any combination 
    of tube potential and current that will result in an exposure rate in 
    excess of 2.58x10-3 C/kg per minute (10 R/min) at the point where 
    the center of the useful beam enters the patient except:
        (i) During the recording of images from an x-ray image-intensifier 
    tube using photographic film or a video camera when the x-ray source is 
    operated in a pulsed mode.
        (ii) When an optional high-level control is activated. When the 
    high-level control is activated, the equipment shall not be operable at 
    any combination of tube potential and current that will result in an 
    exposure rate in excess of 5.16x10-3 C/kg per minute (20 R/min) at 
    the point where the center of the useful beam enters the patient. 
    Special means of activation of high-level controls shall be required. 
    The high-level control shall only be operable when continuous manual 
    activation is provided by the operator. A continuous signal audible to 
    the fluoroscopist shall indicate that the high-level control is being 
    employed.
        (3) Measuring compliance. Compliance with paragraph (e) of this 
    section shall be determined as follows:
        (i) If the source is below the x-ray table, the exposure rate shall 
    be measured at 1 centimeter above the tabletop or cradle.
        (ii) If the source is above the x-ray table, the exposure rate 
    shall be measured at 30 centimeters above the tabletop with the end of 
    the beam-limiting device or spacer positioned as closely as possible to 
    the point of measurement.
        (iii) In a C-arm type of fluoroscope, the exposure rate shall be 
    measured at 30 centimeters from the input surface of the fluoroscopic 
    imaging assembly, with the source positioned at any available SID, 
    provided that the end of the beam-limiting device or spacer is no 
    closer than 30 centimeters from the input surface of the fluoroscopic 
    imaging assembly.
        (iv) In a lateral type of fluoroscope, the exposure rate shall be 
    measured at a point 15 centimeters from the centerline of the x-ray 
    table and in the direction of the x-ray source with the end of the 
    beam-limiting device or spacer positioned as closely as possible to the 
    point of measurement. If the tabletop is movable, it shall be 
    positioned as closely as possible to the lateral x-ray source, with the 
    end of the beam-limiting device or spacer no closer than 15 centimeters 
    to the centerline of the x-ray table.
        (4) Exemptions. Fluoroscopic radiation therapy simulation systems 
    are exempt from the requirements set forth in paragraph (e) of this 
    section.
    * * * * *
    
        Dated: May 12, 1994.
    Michael R. Taylor,
    Deputy Commissioner for Policy.
    [FR Doc. 94-12155 Filed 5-17-94; 8:45 am]
    BILLING CODE 4160-01-F
    
    
    