95-13205. Performance Requirements for Radiography Equipment  

  • [Federal Register Volume 60, Number 104 (Wednesday, May 31, 1995)]
    [Rules and Regulations]
    [Pages 28323-28325]
    From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
    [FR Doc No: 95-13205]
    
    
    
    =======================================================================
    -----------------------------------------------------------------------
    
    NUCLEAR REGULATORY COMMISSION
    
    10 CFR Part 34
    
    RIN 3150-AF28
    
    
    Performance Requirements for Radiography Equipment
    
    Agency: Nuclear Regulatory Commission.
    
    Action: Final rule.
    
    -----------------------------------------------------------------------
    
    Summary: The Nuclear Regulatory Commission (NRC) is amending the 
    regulations pertaining to performance requirements for radiography 
    equipment. The amended provision permits a licensee to use an alternate 
    value of torque for the performance testing criteria. The specified 
    torque test for the drive cable that is currently in the regulations is 
    not practical to meet, given the design of radiographic equipment. 
    Further, the amendment allows for the use of engineering analysis to 
    demonstrate that a modest change in an already approved design is 
    acceptable without the need to perform prototype tests. The amendment 
    is necessary to relieve licensees from compliance with an impractical 
    and unnecessary test criterion.
    
    EFFECTIVE DATE: June 30, 1995.
    
    FOR FURTHER INFORMATION CONTACT: Thomas Rich or J. Bruce Carrico, 
    Office of Nuclear Material Safety and Safeguards, U.S. Nuclear 
    Regulatory Commission, Washington, DC 20555, telephone (301) 415-7893 
    or (301) 415-7826.
    
    SUPPLEMENTARY INFORMATION:
    
    Background
    
        On January 10, 1990 (55 FR 843) the U.S. Nuclear Regulatory 
    Commission published a revision to 10 CFR 34.20. This regulation 
    required that significant safety improvements be made to radiography 
    equipment. Some of these are stated explicitly in the regulation and 
    some are required through an incorporation by reference of American 
    National Standards Institute N432-1980 (ANSI-N432). All newly 
    manufactured radiographic exposure devices and associated equipment 
    acquired by NRC licensees after January 10, 1992, must meet the 
    requirements specified in 10 CFR 34.20, including the provisions of 
    ANSI-N432. All equipment in use after January 10, 1996, must meet these 
    requirements. Vendors of source assemblies, associated equipment, and 
    radiography exposure devices have registered their designs with the NRC 
    or an Agreement State. This process allows both the user and regulatory 
    agency to determine if the equipment meets the applicable safety 
    requirements. Two vendors are located in Agreement States and three are 
    under NRC jurisdiction.
        It has come to the attention of the NRC staff that one of the test 
    criteria specified in section 8.9.2(c) of ANSI-N432 is not practical 
    and cannot be implemented. This test criterion is also not needed for 
    demonstration of safety, given the current design and use of 
    radiography equipment. The test in question is a prototype endurance 
    test of the entire radiography system and, in particular, is intended 
    to ensure the integrity of the source assembly for 20,000 operating 
    cycles. The tests have been performed for the specified number of 
    cycles and at the proper rotational speed, but not at the value of 
    [[Page 28324]] torque specified in ANSI-N432. The specific torque 
    translates to approximately 1345 newtons (300 pounds-force) tensile/
    compressive load on the teleflex cable. It is apparently not possible 
    to actually test at the torque values specified in the ANSI standard.
        The torque requirement specified in the ANSI test is not a 
    reasonable standard. First, it exceeds by a considerable amount the 
    torque that a human can exert on the radiography system while cranking 
    the source in and out by hand. Second, it would require that the drive 
    cable (Type 187 teleflex cable used in radiography cameras for the past 
    decade) be operated beyond the working load recommended by the supplier 
    of that component. The recommended working load for the standard cable 
    for 10,000 cycles is less than 583 newtons (130 pounds-force). For 
    20,000 cycles, which the ANSI endurance test specifies, the working 
    load would be lower.
        The design of the drive cable system in most radiography cameras 
    has been unchanged for more than a decade. The NRC staff is not aware 
    of any cable failures as a result of fatigue. Cable failures of this 
    type would be clearly visible to the radiographer and 10 CFR 34.30 
    requires reporting to NRC. Based on the good operating experience with 
    the standard cable and the fact that an individual is highly unlikely 
    to generate 1345 newtons of force continuously on the cable, the NRC 
    staff believes that testing equipment to the high torque requirement of 
    ANSI-N432 is not needed to ensure system safety.
        The NRC staff requested the American National Standards Institute 
    Committee N43, the organization responsible for development of the 
    standard, to clarify the basis for the test criterion. The committee's 
    response indicates that the requirement was adopted from an 
    International Standards Organization standard, that it was not aware of 
    the severity of the requirement, and that it was not aware of any 
    manufacturer that has tested equipment to this requirement. Based on 
    further discussion with the N43's working group subcommittee chairman, 
    the NRC staff understands that the working group intends to revise the 
    standard to incorporate a more realistic torque requirement for the 
    endurance test. However, considering the approval and publication 
    process, a revised standard would not be issued for at least 18 months. 
    At that time the NRC staff will evaluate the revised ANSI standard and 
    consider revising its regulations, if necessary and appropriate for 
    maintaining public safety, when radiography equipment is used.
    The Amendment
    
        Section 34.20 is being amended, first, by inserting a new sentence 
    in paragraph (a) that will permit an applicant or licensee to submit an 
    engineering analysis to demonstrate the applicability of previously 
    performed testing on similar individual radiography equipment 
    components. This addition codifies a long-standing staff practice in 
    evaluating radiography equipment. For example, an engineering analysis 
    can demonstrate that a modest change in design is acceptable without 
    repeating a prototype test.
        Second, because of the flaw in the ANSI standard criteria, the 
    Commission is amending its regulation in 10 CFR 34.20 to eliminate the 
    impractical torque test. In its place, a radiography exposure device 
    and associated systems will be considered to be in compliance with the 
    performance requirements if the prototype equipment was tested using a 
    value of a torque representative of the torque that an individual using 
    the radiography equipment can realistically exert, provided the 
    exposure device and associated equipment are in compliance with all 
    other criteria in the referenced ANSI standard. To accomplish this 
    objective, a new paragraph (f) is being added to 10 CFR 34.20, to 
    specify that compliance with the ANSI-N432-1980 torque value for the 
    endurance test is not required, and that use of a realistic torque 
    value will satisfy the performance requirement. Furthermore, all 
    radiography equipment currently shown and sold by vendors meeting the 
    current part 34 requirements, will meet the revised Sec. 34.20. These 
    vendors have previously provided test results or engineering analysis 
    to either the Agreement State or NRC to demonstrate the products meet 
    Sec. 34.20. The revision imposes a practical performance requirement 
    that is consistent with industry practice while meeting NRC's objective 
    to provide radiographers with safe equipment. Therefore, the filing of 
    additional information with the Commission to demonstrate compliance 
    with the revised Sec. 34.20 requirement is not necessary.
        The Commission finds that public comment on this rule is 
    unnecessary because the purpose of the rule is to remove from the 
    regulations an impractical requirement and to stipulate in its stead a 
    practical standard that will permit continued use of a specific 
    component, the drive cable, of industrial radiography equipment long in 
    use without violating the Commission's regulation. The rule change 
    preserves the status quo for the particular component.
    
    Compatibility of Agreement State Regulations
    
        Section 34.20 is currently designated as a Division II Matter of 
    Compatibility for Agreement State regulations. The revisions addressed 
    in this rule correct a flaw in the regulations. The rule does not 
    affect the current compatibility designations and therefore, 10 CFR 
    34.20 continues to be designated as a Division II Matter of 
    Compatibility.
    
    Environmental Impact: Categorical Exclusion
    
        The NRC has determined that this final rule is the type of action 
    described in categorical exclusion 10 CFR 51.22(c)(2). Therefore, 
    neither an environmental impact statement nor an environmental 
    assessment has been prepared for this rule.
    
    Paperwork Reduction Act Statement
    
        This final rule does not contain a new or amended information 
    collection requirement subject to the Paperwork Reduction Act of 1980 
    (44 U.S.C. 3501 et seq.). Existing requirements were approved by the 
    Office of Management and Budget approval number 3150-0007.
    
    Regulatory Analysis
    
        The NRC has prepared this final rule to cure a defect in its 
    regulations that places an unnecessary and unwarranted burden on 
    certain of its licensees that use sources and devices for radiography. 
    There is no other procedure available to the NRC to efficiently and 
    effectively rectify the matter. There is no cost to the licensed and 
    regulated community in the promulgation of this rule. This discussion 
    constitutes the regulatory analysis for this rule.
    Backfit Analysis
    
        The NRC has determined that the backfit rule, 10 CFR 50.109, does 
    not apply to this final rule, and therefore, that a backfit analysis is 
    not required for this final rule because these amendments do not 
    involve any provisions that would impose backfits as defined in 10 CFR 
    50.109(a)(1).
    
    List of Subjects in 10 CFR Part 34
    
        Criminal penalties, Incorporation by reference, Packaging and 
    containers, Radiation protection, Radiography, Reporting and 
    recordkeeping requirements, Scientific equipment, Security measures.
    
        For the reasons set out in the preamble and under the authority of 
    the Atomic Energy Act of 1954, as amended, [[Page 28325]] the Energy 
    Reorganization Act of 1974, as amended, and 5 U.S.C. 552 and 553, the 
    Commission is adopting the following amendments to 10 CFR Part 34.
    
    PART 34--LICENSES FOR RADIOGRAPHY AND RADIATION SAFETY REQUIREMENTS 
    FOR RADIOGRAPHIC OPERATIONS
    
        1. The authority citation for Part 34 continues to read as follows:
    
        Authority: Secs. 81, 161, 182, 183, 68 Stat. 935, 948, 953, 954, 
    as amended (42 U.S.C. 2111, 2201, 2232, 2233); sec. 201, 88 Stat. 
    1242, as amended (42 U.S.C. 5841).
        Section 34.32 also issued under sec. 206, 88 Stat. 1246 (42 
    U.S.C. 5846).
    
        2. Section 34.20 is amended by revising paragraph (a) and adding a 
    new paragraph (f) to read as follows:
    
    
    Sec. 34.20  Performance requirements for radiography equipment.
    
    * * * * *
        (a) Each radiographic exposure device and all associated equipment 
    must meet the requirements specified in American National Standards 
    Institute N432-1980, ``Radiological Safety for the Design and 
    Construction of Apparatus for Gamma Radiography,'' (published as NBS 
    Handbook 136, issued January 1981). This publication has been approved 
    for incorporation by reference by the Director of the Federal Register 
    in accordance with 5 U.S.C. 552(a). This publication may be purchased 
    from the Superintendent of Documents, U.S. Government Printing Office, 
    Washington, DC 20402. Copies of the document are available for 
    inspection at the Nuclear Regulatory Commission library, 11545 
    Rockville Pike, Rockville, Maryland, 20852-2738. A copy of the document 
    is also on file at the Office of the Federal Register, 800 North 
    Capitol Street NW., Suite 700, Washington, DC 20408.
        Engineering analyses may be submitted by an applicant or licensee 
    to demonstrate the applicability of previously performed testing on 
    similar individual radiography equipment components. Upon review, the 
    Commission may find this an acceptable alternative to actual testing of 
    the component pursuant to the referenced standard.
    * * * * *
        (f) Notwithstanding paragraphs (a), (d), and (e) of this section, 
    equipment used in industrial radiographic operations need not comply 
    with section 8.9.2(c) of the Endurance Test in American National 
    Standards Institute N432-1980, if the prototype equipment has been 
    tested using a torque value representative of the torque that an 
    individual using the radiography equipment can realistically exert on 
    the lever or crankshaft of the drive mechanism.
    
        Dated at Rockville, MD, this 17th day of May, 1995.
    
        For the Nuclear Regulatory Commission.
    James M. Taylor,
    Executive Director for Operations.
    [FR Doc. 95-13205 Filed 5-30-95; 8:45 am]
    BILLING CODE 7590-01-P
    
    

Document Information

Effective Date:
6/30/1995
Published:
05/31/1995
Department:
Nuclear Regulatory Commission
Entry Type:
Rule
Action:
Final rule.
Document Number:
95-13205
Dates:
June 30, 1995.
Pages:
28323-28325 (3 pages)
RINs:
3150-AF28
PDF File:
95-13205.pdf
CFR: (1)
10 CFR 34.20