98-14605. Medical Devices; Classification/Reclassification of Immunohistochemistry Reagents and Kits  

  • [Federal Register Volume 63, Number 106 (Wednesday, June 3, 1998)]
    [Rules and Regulations]
    [Pages 30132-30142]
    From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
    [FR Doc No: 98-14605]
    
    
    
    [[Page 30132]]
    
    =======================================================================
    -----------------------------------------------------------------------
    
    DEPARTMENT OF HEALTH AND HUMAN SERVICES
    
    Food and Drug Administration
    
    21 CFR Part 864
    
    [Docket No. 94P-0341]
    RIN 0910-ZA10
    
    
    Medical Devices; Classification/Reclassification of 
    Immunohistochemistry Reagents and Kits
    
    AGENCY: Food and Drug Administration, HHS.
    
    ACTION: Final rule.
    
    -----------------------------------------------------------------------
    
    SUMMARY: The Food and Drug Administration (FDA) is issuing a final rule 
    to classify/reclassify immunohistochemistry reagents and kits (IHC's) 
    into three classes depending on intended use. FDA is classifying/
    reclassifying into class I (general controls) and exempt from premarket 
    notification requirements IHC's used as adjunctive tests and presenting 
    a low risk to public health. FDA is classifying/reclassifying into 
    class II (special control) IHC's that detect or measure certain target 
    analytes and that provide prognostic or predictive data that is not 
    confirmed by routine histopathologic control specimens. The results of 
    the class II IHC's are reported independently to the clinician, and the 
    performance claims are widely accepted and supported by valid 
    scientific evidence. FDA is classifying/reclassifying into class III 
    (premarket approval) IHC's intended for any other use. The scope of 
    products covered by this final rule includes both pre-1976 devices that 
    have not been previously classified, as well as post-1976 devices that 
    are statutorily classified into class III. The intent of this final 
    rule is to regulate pre-1976 devices and post-1976 devices in a 
    consistent fashion. Therefore, FDA is classifying or reclassifying 
    these products as applicable.
    EFFECTIVE DATE:  This rule is effective August 17, 1998.
    
    FOR FURTHER INFORMATION CONTACT: Max Robinowitz, Center for Devices and 
    Radiological Health (HFZ-440), Food and Drug Administration, 2098 
    Gaither Rd., Rockville, MD 20850, 301-594-1293, ext. 136, or FAX 301-
    594-5941.
    
    SUPPLEMENTARY INFORMATION:
    
    I. Background
    
        Immunohistochemistry (IHC) is the diagnostic laboratory practice 
    that combines immunologic techniques, using specially prepared antibody 
    reagents, with the examination of intact cells and tissues under the 
    microscope by a pathologist or other trained laboratory scientist. An 
    IHC device is an in vitro diagnostic reagent or test kit that uses 
    immunological methods to identify antigens in tissues or intact cells. 
    An IHC reagent is the primary antibody of an IHC assay that is 
    developed to specifically target, react to, or combine with, a 
    particular cellular or tissue constituent, or antigen, using specific 
    immunological characteristics of the antibody. IHC's may be used 
    together with a secondary or reporter antibody, buffers, washing 
    solutions, and controls. If an IHC primary antibody reagent is sold 
    separately, there should be recommendations for what ancillary reagents 
    and equipment should be used with the IHC reagent to achieve the 
    performance characteristics claimed for the primary IHC reagent. If the 
    IHC is marketed as a test kit, there should be performance data with 
    the finished test kit.
    
    II. Highlights of the Final Rule
    
        In response to public comments, FDA has revised and clarified 
    certain provisions of the final regulation. The revisions maintain the 
    protection of the public health while reducing the regulatory burden on 
    manufacturers by lowering the classification of a number of IHC's. The 
    most significant changes from the proposed rule are as follows:
        1. Under the final rule, most IHC's are being classified as class I 
    devices, exempt from premarket notification. Class I includes all IHC's 
    being used as adjuncts to conventional histopathologic diagnostic 
    examination.
        2. The definition of class II IHC's has been changed to include 
    IHC's that are not directly confirmed by routine histopathologic 
    control specimens and with claims that are widely accepted and 
    supported by valid scientific evidence. Class II IHC's now include such 
    products as estrogen and progesterone receptors (ER/PR's).
        3. The definition of class III IHC's has been narrowed to include 
    only those IHC's that do not meet the criteria for class I or II.
        4. Accordingly, the rule lessens the regulatory burden for bringing 
    IHC's to market because most IHC's are now classified/reclassified as 
    class I or II. As post-1976 devices, most IHC's previously were class 
    III devices under section 513(f)(1) of the Federal Food, Drug, and 
    Cosmetic Act) (the act) (21 U.S.C. 360c(f)(1)).
        In addition, the agency clarifies and reinforces the following 
    points:
        1. This final rule regulates only IHC's being used for diagnostic 
    purposes. Neither the proposed rule nor the final rule would require 
    submissions for reagents or test kits used for research purposes only. 
    Nor does FDA require manufacturers of such research use only reagents 
    or test kits to comply with general controls; and
        2. IHC's used for diagnostic purposes have been and will continue 
    to be subject to the current good manufacturing practices (CGMP's) 
    under the act. The requirement to comply with CGMP's is a general 
    control that all devices must meet (unless expressly exempt under 
    section 513(d)(2)(A)) of the act without regard to their level of 
    classification or whether they have been previously classified. (See H. 
    Rept. 94-853 at 17 (1976).)
    
    III. The Final Rule
    
    A. General Approach
    
        FDA believes that the final rule establishes reasonable 
    requirements that can be implemented by the regulated industry without 
    unnecessary burden. To ensure safety and effectiveness, all classes of 
    IHC's will be subject to the following general controls: (1) Labeling 
    requirements for in vitro devices (Sec. 809.10 (21 CFR 809.10)), (2) 
    compliance with CGMP's, (3) registration and listing, (4) recordkeeping 
    and medical device reporting (MDR), and (5) labeling for prescription 
    use (Sec. 801.109 (21 CFR 801.109)). FDA has determined that these 
    controls are necessary for a reasonable assurance of safety and 
    effectiveness.
    
    B. Class I Exempt From Premarket Notification
    
        In the final rule, FDA has broadened the class I identification to 
    include all adjunctive IHC's. This change places the majority of IHC's 
    into class I. The final rule also modifies the language in the 
    regulation to clarify that class I IHC's are used to classify tumors.
        In response to comments submitted on the proposed rule, FDA 
    reconsidered the regulation of class I IHC's and decided to exempt them 
    from premarket notification (510(k)) requirements. In considering 
    whether to exempt class I devices from premarket notification, FDA 
    focuses on whether notification for the type of device is necessary for 
    the protection of the public health. For the devices exempted from 
    premarket notification by this rule, FDA has concluded that 
    notification is unnecessary primarily for the following reasons:
        1. The devices do not have a significant history of false or 
    misleading performance claims or of risks
    
    [[Page 30133]]
    
    associated with inherent characteristics of the device, such as device 
    design or materials. When making such determinations, FDA generally has 
    considered the frequency, persistence, cause, or seriousness of such 
    performance claims or risks, as well as other relevant factors.
        FDA is unaware of IHC failure being reported in the MDR data base. 
    IHC failures have been reported in the medical literature; the risks of 
    such failure, however, are mitigated by widely accepted practices that 
    are based on valid scientific evidence.
        2. In general, FDA will exempt a device from premarket notification 
    when the following factors apply: (a) Characteristics of the device 
    necessary for its safe and effective performance are well established; 
    (b) anticipated changes in the device that could affect safety and 
    effectiveness will either be readily detectable by users by visual 
    examination or other means, such as routine testing, before causing 
    harm (e.g. testing of a clinical laboratory reagent with positive and 
    negative controls), or not materially increase the risk of injury, 
    incorrect diagnosis, or ineffective treatment; and (c) any changes in 
    the device would not be likely to result in a change in the device's 
    classification. FDA makes these determinations based on its knowledge 
    of the device, including past experience and relevant reports or 
    studies on device performance.
        The characteristics of IHC's are well established. Although the 
    method is not generally quantitative, the results generated using this 
    technology are sufficiently accurate and precise to support 
    subclassification of tumors (neoplasms), and detection and measurement 
    of the presence or absence of clinically significant target analytes. 
    There are sufficient quality assurance techniques in the use of IHC's 
    to enhance the precision of the methodology and minimize the risks of 
    misdiagnosis.
        Because class I IHC's are identified as those that are used 
    adjunctively to support conventional histopathological diagnosis and 
    are controlled by readily available internal and external control 
    materials, minor changes to the IHC would not materially increase the 
    risk of injury, incorrect diagnosis, or ineffective treatment. 
    Adjunctive test results are evaluated and incorporated into the 
    diagnostic interpretation by the pathologist and are not usually 
    reported directly to the clinician. Because laboratories certified 
    under the Clinical Laboratory Improvement Amendments of 1988 (CLIA) are 
    required to run positive and negative quality control samples with all 
    special stains, reagent failures are likely to be easily identified by 
    pathologists. In addition, most slides will have normal along with 
    abnormal tissue included as part of the tissue sample on the slide and 
    this juxtaposition affords an additional opportunity to identify 
    inappropriate or uncommon staining patterns.
        Manufacturers are reminded that exemption from the requirement of 
    premarket notification is not an exemption from CGMP's and the other 
    applicable general controls.
        Because IHC's have been classified in accordance with the risk 
    associated with their intended use, a change in intended use or 
    indications for use of an IHC would likely result in a 
    reclassification. Such a change would not be considered minor and would 
    probably require a submission to the agency. For a discussion of 
    whether a change to a device would require a manufacturer to submit a 
    510(k), see the FDA's guidance entitled ``Deciding when to Submit a 
    510(k) for a Change to an Existing Device.''
    
    C. Class II
    
        In contrast to all adjunctive IHC's being placed into class I, the 
    final rule clarifies that class II IHC's are IHC's that generate 
    results that are not directly confirmed by routine histopathologic 
    internal and external control specimens. Class II IHC's are intended to 
    provide information that is ordinarily reported as independent 
    diagnostic information to the clinician. For an IHC to be classified 
    into class II, the claims associated with this information must be 
    widely accepted and supported by valid scientific evidence. FDA 
    believes that the manufacturer/sponsor can establish the acceptance of 
    the intended use of the IHC and valid scientific evidence through 
    sponsor-supported studies or scientific literature references, 
    materials from professional educational seminars, and/or the citation 
    of practice standards or guidelines, as described in the special 
    control noted in the paragraph below. These IHC's must be developed and 
    established by validation and correlation testing with well 
    characterized clinical specimens that support the intended use of the 
    IHC test system as an independent prognostic or predictive marker. FDA 
    believes that providing valid scientific evidence of performance claims 
    that are widely accepted and complying with the general controls should 
    be sufficient to ensure the safe and effective use of these IHC's.
        Class II IHC's are subject to a special control entitled ``FDA 
    Guidance for Submission of Immunohistochemistry Applications to the 
    FDA,'' FDA, Center for Devices and Radiologic Health, 1998. The updated 
    guidance will assist sponsors in collecting and presenting data to FDA 
    to establish that the claims associated with use of the device are 
    widely accepted and that there is valid scientific evidence to support 
    performance claims with clinical specimens. The special control is also 
    intended to provide guidance to manufacturers about labeling for use of 
    the device. This guidance has been issued as a Level 2 guidance 
    consistent with the ``Good Guidance Practices'' (GGP's) FDA adopted for 
    the development, issuance, and use of guidance documents (62 FR 8961, 
    February 27, 1997). Persons interested in obtaining this document 
    should refer to section VI of this document entitled ``Access to the 
    Special Control.''
        Several comments urged that IHC's for ER/PR's be classified as 
    class II devices rather than as class III, as proposed. FDA concurs 
    with this suggestion. By using well characterized clinical specimens 
    and validating their IHC's against appropriate FDA approved chemical 
    receptor assays, manufacturers can reliably characterize these products 
    and support their clearance as class II devices. FDA believes that 
    class II classification can now safely apply to IHC's for ER/PR's, 
    including hormone receptors in breast cancer, because clinical reliance 
    on such testing has been established in the medical literature and the 
    information derived from such test results are well understood.
    
    D. Class III
    
        In response to comments on the proposed rule and changes to the 
    class II classification, FDA has narrowed the scope of the class III 
    identification. Under the final rule, IHC's that do not meet the 
    criteria for class I or II will be classified into class III. 
    Manufacturers of these IHC's must submit valid scientific evidence to 
    support the new intended uses. An example of a class III IHC would be 
    markers used to identify new, clinically significant target analytes in 
    tissue specimens that cannot be confirmed by conventional 
    histopathologic examination.
        FDA has amended proposed Sec. 864.1860(c) to indicate that 
    postamendment class III IHC's cannot be commercially distributed unless 
    the manufacturer has an approval under section 515 of the act (21 
    U.S.C. 360e).
    
    IV. The Proposed Rule
    
        In the Federal Register of June 14, 1996 (61 FR 30197), FDA 
    published a
    
    [[Page 30134]]
    
    proposed rule to classify/reclassify IHC's. The proposed rule contained 
    the reasons for the proposed classification/reclassification, 
    summarized the Hematology and Pathology Device Panel's recommendation 
    regarding the classification of IHC devices, identified the risks to 
    health presented by the devices, included a summary of the data upon 
    which the proposed classification/reclassification was based, and 
    delineated the statutory authority under which FDA issues this rule. 
    Written comments were due August 30, 1996.
        The agency received 26 comments from individuals, manufacturers, 
    professional societies, and the U.S. Small Business Administration. A 
    summary of the written comments and FDA's response to them is provided 
    in section V of this document.
    
    V. Response to Comments
    
    A. Classification
    
        1. Two comments supported the classification of IHC reagents and 
    test kits into classes based on intended use as a balanced and 
    responsible level of regulation that would: (1) Not impinge on the 
    continued availability of these materials; (2) not negatively impact 
    the advance of new technology due to application of inappropriately 
    stringent regulatory controls; (3) not be overly burdensome to FDA or 
    industry; or (4) not be inconsistent with the needs and interests of 
    the medical professions, clinical laboratories, FDA, and industry.
        A third comment agreed with most of the proposed classification 
    designations. A fourth comment stated that IHC's intended for 
    adjunctive use were appropriately classified into class I. A fifth 
    comment stated that most immunohistochemical antibody reagents should 
    be regulated as class I because if they were ``over-regulated'' it 
    would be difficult to bring the antibodies to market and the reagents 
    were needed daily in the practice of surgical pathology.
        A sixth comment suggested that the proper classification for many 
    IHC reagents and test kits would be class I 510(k) exempt in vitro 
    diagnostic (IVD) devices. The comment argued that premarket 
    notification (510(k)) should not be necessary because: (1) 510(k) 
    clearance will not impact significantly on the expertise of the 
    pathologist nor on the quality or reproducibility of 
    immunocytochemistry/immunohistochemistry, which was the central factor 
    in the safe and effective use of immunocytochemistry/
    immunohistochemistry; (2) 510(k) clearance provided false reassurance 
    to the inexperienced end user in making diagnoses based on possibly 
    erroneous interpretations of data; and (3) of the negative implications 
    of the cost of 510(k) clearance.
        A seventh comment argued that the benefits do not outweigh the 
    costs for class I devices to be required to submit a 510(k). The 
    comment argued that manufacturers have no control over how accurately a 
    pathologist interprets results and that the correct focus should be on 
    CGMP standards and other key determinants of manufacturing consistency 
    and compliance.
        An eighth comment believed the majority of IHC's should be class I 
    and exempt from premarket notification requirements. The comment argued 
    that production of the antibody product was not the most critical and 
    subjective step in this diagnostic technique and that FDA's resources 
    were better spent in the area of ensuring reliable and consistent 
    production through the controls of CGMP's, medical device reporting, 
    registration, etc., to assure manufacturing consistency and compliance.
        FDA has considered these comments and has concluded that premarket 
    notification is unnecessary for the protection of the public health for 
    class I IHC's, which are those used to produce diagnostic information 
    that is confirmed readily by other tests or procedures. Section 
    513(d)(2)(A) of the act authorizes FDA to exempt class I IHC's from the 
    requirement of premarket notification in section 510(k) of the act (21 
    U.S.C. 360(k)). This exemption permits manufacturers to introduce into 
    commercial distribution those IHC's that fall within the class I 
    classification without obtaining premarket clearance from FDA. Ongoing 
    initiatives by professional organizations, manufacturers, and FDA are 
    directed at ensuring that pre- and postanalytic, as well as analytic 
    procedures, are properly performed. In the context of these 
    initiatives, FDA believes that classifying these devices as class I and 
    applying general controls will ensure that the majority of adjunctive 
    IHC's are used safely and effectively without the need to require 
    premarket notification. The Food and Drug Administration Modernization 
    Act of 1997, which became effective on February 19, 1998, does not 
    eliminate the need for this rule or require changes with respect to 
    FDA's determinations about classification of these products. The rule 
    establishes a classification scheme for all IHC's including many that 
    were not previously classified as well as class III IHC's. The class I 
    IHC's that are exempt from premarket notification under this rule do 
    not fall into the category of those class I devices that continue to 
    require premarket notification under the new legislation (section 
    510(l) of the act). Nor does the agency believe that the IHC's being 
    classified into class II by this rule are appropriate for exemption 
    from 510(k) submissions under new section 510(m) of the act.
        2. One comment requested clarification concerning the scope of the 
    proposed regulation as it pertains to ``ancillary reagents'' (including 
    detection systems). The comment recommended that ancillary reagents, 
    including secondary antibodies, buffers, and chromogens, should most 
    appropriately be regulated as general purpose reagents under 
    Sec. 864.4010 (21 CFR 864.4010), and subject to Sec. 864.1860 (21 CFR 
    864.1860) only when packaged with one or more primary antibodies as 
    components of a complete test system.
        FDA agrees in part with the comment. ``Ancillary reagents'' are 
    subject to Sec. 864.1860 when they are packaged with one or more 
    primary antibodies as a complete test system. In addition, ancillary 
    reagents are also subject to this regulation when they are sold with 
    performance claims for their use as a general detection system in 
    conjunction with primary antibodies that are sold separately. FDA 
    agrees that secondary antibodies, buffers, and chromogens may be 
    regulated as general purpose reagents under Sec. 864.4010 when these 
    reagents are sold without performance claims.
        3. Two comments requested clarification concerning whether devices 
    in commercial distribution prior to May 28, 1976, must comply with the 
    classification and requirements in the proposed rule, particularly the 
    proposed labeling recommended in the March 28, 1995, guidance listed as 
    Ref. 6 in the proposal (61 FR 30197 at 30199). The comment argued that 
    the regulation of ``pre 1976 devices which have not been previously 
    classified'' contradicts Sec. 807.85(b)(1) (21 CFR 807.85(b)(1)), which 
    exempts ``grandfathered'' products from 510(k) review.
        These comments misunderstand the meaning of Sec. 807.85(b)(1). 
    Section 807.85(b)(1) establishes exemptions from premarket notification 
    for private label distributors and repackagers who distribute devices 
    that already are being legally marketed without making any changes to 
    the device or its labeling beyond the addition of the private label 
    name. The exemptions in Sec. 807.85(b)(1) do not apply to device 
    manufacturers and distributors generally.
        It is true that the requirement to submit a premarket notification 
    before introducing a device into the market
    
    [[Page 30135]]
    
    after May 28, 1976, does not apply to devices that were legally 
    marketed prior to that date. However, as explained in 21 CFR 
    807.81(a)(3), a manufacturer of a device that was marketed prior to the 
    1976 amendments is required to file a 510(k) if the devices was 
    significantly changed or modified in design, components, methods of 
    manufacture, or intended use. A first time manufacturer of a device 
    that the manufacturer believes to be the same or substantially 
    equivalent to a device that is already marketed also must submit a 
    510(k) to establish that substantial equivalence, unless the product 
    has been exempted from notification under 513(d)(2)(A).
        As discussed previously, preamendment devices have been and will 
    continue to be subject to general controls, such as CGMP's and the 
    existing labeling requirements (Sec. 809.10) for in vitro devices. 
    Although manufacturers of preamendment class II IHC's that are not 
    required to submit 510(k)'s will have no need to utilize FDA's guidance 
    being established as a special control, manufacturers of preamendment 
    devices that are modified in a way that will require submission of a 
    new 510(k) should consult the special control: ``FDA Guidance for 
    Submission of Immunohistochemical Applications to the FDA'' when 
    submitting premarket notifications for class II devices. Because this 
    special control is a guidance, no manufacturer is bound to follow the 
    details of the document (see response to comment 13 in section V of 
    this document).
        4. Three comments argued that additional regulation will do nothing 
    to lower the risk of misinterpretation of results. The comment stated 
    that IHC's have almost always been used as an adjunct to other 
    diagnostic techniques and that the proposed regulations would not 
    necessarily accomplish FDA's stated objectives of reducing risks to 
    patients.
        FDA's regulation of IHC assays is limited to oversight of the 
    manufacturers of IHC reagents or test kits; the rule does not regulate 
    the end users or their laboratories. FDA recognizes that safe and 
    effective IHC's do not by themselves guarantee that an IHC in the end 
    user's laboratory will be used accurately and reliably. FDA agrees that 
    IHC assays are multistep IVD test systems that require the expert 
    supervision of a qualified pathologist or laboratory scientist to 
    ensure that all the preanalytic, analytic, and postanalytic steps are 
    performed accurately and reliably.
        FDA believes, however, that the building blocks of those assays 
    should be safe, effective, and properly labeled for their intended use. 
    The risks associated with use of an IHC include the likelihood of 
    obtaining a false result, while the effectiveness of an IHC is 
    dependent upon the likelihood of the IHC performing as claimed by the 
    manufacturer. In accordance with Sec. 809.10, the label must include, 
    among other things, the intended use, indications for use, the 
    instructions for use, and limitations. The manufacturer is required to 
    support any performance claims for accuracy, precision, sensitivity, 
    and specificity included on the label of the IHC device with valid 
    scientific evidence. The labeling also should include statements that 
    remind the end user of the variable nature of the specimens to be 
    examined by the IHC, i.e., biologic variability of the tissues and 
    patients, the need for procedures relating to preanalytic fixation, 
    handling, processing, storage, and the variability and subjectivity in 
    the interpretation of the IHC slides.
        Contrary to the assertion of some comments, FDA believes that such 
    regulation does reduce risks associated with use of IHC's. The 
    requirements that labeled performance claims be supported by valid 
    scientific evidence and that labeling include instructions for use, 
    limitations, and information about variability significantly increases 
    the likelihood that the end user will have a product that will be used 
    safely and effectively in the laboratory.
        In response to comments that implied that industry experts did not 
    believe regulation of IHC's was necessary to reduce risks associated 
    with their use, FDA notes that its classification/reclassification 
    initiatives with respect to IHC's are based on input from public 
    workshops, advisory panels to the FDA, and industry petitions for 
    reclassification, as well as FDA experience with assessment of the 
    safety and effectiveness of IHC devices.
        FDA is aware that its regulation of IHC's is supported by other 
    assurances of safe and effective performance of the assays. For 
    example, there is widespread participation by end users in voluntary 
    and mandatory training in IHC assays and proficiency testing in IHC 
    assays by other government and professional organizations. End users 
    may also use voluntary guidelines to ensure reliable and accurate 
    performance of IHC assays within their laboratories, e.g., ``The 
    National Committee for Clinical Laboratory Standards (NCCLS) Quality 
    Assurance for Immunocytochemistry; Proposed Guideline, MM4-P,'' 
    February, 1997. (The approved NCCLS guideline is expected within 2 
    years.) However, FDA believes such voluntary standards and practices 
    cannot serve as a complete substitute for government regulation of 
    these devices. The existence of such guidelines and widespread 
    compliance with their recommendations has contributed to FDA's 
    determination that most of these devices can be regulated at the least 
    stringent level of control and be exempt from premarket notification.
        5. One comment did not support placing IHC's in which test results 
    were ``ordinarily reported as independent diagnostic information'' into 
    class II because the manner in which IHC test results were reported was 
    determined independent of the IHC supplier or FDA. The comment stated 
    that because there may be significant laboratory-to-laboratory and 
    within-laboratory variation in how results were reported, it would be 
    difficult to consistently determine device classification on the basis 
    of how results were reported.
        FDA agrees that the IHC manufacturer is not responsible for how 
    each end user laboratory scientist will report the results of an IHC 
    assay. However, the manufacturer is responsible for recommendations and 
    performance claims on the product's label (see Sec. 809.10). Such 
    indications and directions for use are important for the proper 
    performance of the assay and as a reference for compliance with the 
    CLIA requirements for the end user laboratory (42 CFR 493.1211). An 
    individual laboratory that chooses to use the device differently or 
    report results in a manner contrary to labeled recommendations is 
    responsible for that decision and validation of that use.
        FDA defines independent diagnostic information as information that: 
    (1) Is the sole or a major determinant of a diagnosis; (2) is used by 
    itself as the basis for a significant medical decision; or (3) may not 
    be readily confirmed by other diagnostic tests or clinical procedures. 
    FDA believes it is possible to identify IHC's for which test results 
    ordinarily are reported as independent diagnostic information to the 
    ordering clinician, and for which the claims associated with these data 
    are widely accepted and supported by valid scientific evidence. Those 
    IHC's that generate independent diagnostic information and where the 
    claims are not widely accepted will be reviewed as class III devices 
    and approved for marketing if there is valid scientific evidence to 
    support those claims.
        6. One comment stated that it was unclear why Ki-67 was class II, 
    while hematoxylin and eosin (H & E) staining, which was the more 
    critical assay, was class I. The comment added that class II reagents 
    had no characteristics clearly
    
    [[Page 30136]]
    
    distinguishable from those proposed to be in class I.
        The agency believes there are differences between H & E stains and 
    IHC's. Despite the critical nature of the assay, biologic stains such 
    as H & E have been placed in class I and exempted from 510(k) review 
    because FDA determined that the stains were well understood, with 
    commonly used controls that permit the user to readily detect 
    deviations in staining properties. For these reasons, FDA concluded 
    that general controls were sufficient and 510(k) submissions were not 
    necessary to establish reasonable assurance of safe and effective use 
    of H & E stains. IHC's, on the other hand, use monoclonal or polyclonal 
    antibodies that may require specific testing or reagents to verify that 
    the assay meets the manufacturer's specification for performance (see 
    also comment 14 of section V of this document). Under the final rule, 
    however, most IHC's will also be regulated as class I devices exempt 
    from premarket notification.
        FDA has made changes to the final rule that further distinguish 
    class I from class II IHC's. Class I IHC's are adjunctive IHC's. Class 
    II IHC's generate results that ordinarily are independently reported to 
    the clinician. However, the primary difference between a class I or II 
    IHC depends on the manufacturer's claims in the proposed product 
    labeling for an IHC reagent or test kit because it is these claims that 
    establish the intended use of the IHC. An identical device can be 
    subject to a range of regulatory controls--from the lowest to the 
    highest levels of regulation--depending on the claims being made and on 
    the issues of safety and effectiveness associated with those claims.
        Ki-67, the example referenced in the comment, is the name of a 
    monoclonal antibody clone that recognizes a nuclear antigen that is 
    expressed only in proliferating cells. A Ki-67 IHC will yield a 
    positive qualitative result in normal and abnormal proliferating cells. 
    This result correlates with the presence of mitotic activity. A Ki-67 
    IHC would be classified into class I, exempt from premarket 
    notification if: (a) The intended use of the assay result is to provide 
    adjunctive information that indicates the presence or absence of cell 
    proliferation in all or some of the cells within a tissue sample; (b) 
    the IHC can be controlled by the user with readily available positive 
    and negative tissues controls; (c) the result will be incorporated into 
    the pathologist's differential diagnosis; and (d) the result will not 
    be reported as independent information to the clinician.
        A Ki-67 IHC would be classified into class II if: (a) The sponsor 
    claims that the IHC results could be used as a stand-alone test to 
    determine prognosis independent of other findings; (b) the user must 
    use clinically well-characterized tissues to serve as positive and 
    negative controls; or (c) the analytic result will be reported as 
    independent information to the clinician. A Ki-67 IHC would be 
    classified into class III if the sponsor claims that the IHC will be 
    used in combination with a novel amplification method that would allow 
    this IHC to be used as a stand-alone detection system for 
    micrometastases in tissue, or some other new intended use.
        The previous comment was in response to the proposed rule. As 
    discussed previously, the final rule establishes that the majority of 
    IHC's are adjunctive and will fall within class I, exempt from 
    premarket notification. For those IHC's that are not adjunctive, the 
    majority will be class II because they have claims that are widely 
    accepted and supported by valid scientific evidence.
        7. FDA received conflicting comments about the necessity of 
    classifying certain types of IHC's in class III. Two comments stated 
    that IHC's intended for stand-alone use in making clinically 
    significant determinations, such as markers used for the detection of 
    medically important genetic mutations in tissues that were normal by 
    conventional histopathology, should be regulated as class III devices 
    until more information regarding the safety and effectiveness of these 
    tests became available. A third comment stated that, with the exception 
    of a limited number of class III devices that define a site-specific 
    therapeutic intervention and were used to provide circumstantial 
    information in support of H & E based histopathological diagnosis, 
    markers should be classified as class I exempt from premarket 
    notification. A fourth comment added that the higher classifications 
    were not relevant because currently there was no IHC for a prognostic/
    proliferation marker that was a reliable ``stand-alone'' indicator and 
    whose use was generally accepted.
        FDA agrees that the IHC's described in the first two comments 
    should be regulated as class III devices and not be commercially 
    marketed until a premarket approval application (PMA) establishes that 
    there is valid scientific evidence to support safe and effective use of 
    such products. The agency also agrees with the general point being made 
    by the third comment and has classified the majority of IHC's into 
    class I and exempted them from premarket notification requirements. 
    With respect to the fourth comment, the agency does not agree that the 
    identification of class III IHC's is not relevant. The fact that the 
    comment is unaware of products currently on the market that fit the 
    identification does not obviate the need for FDA to have regulations in 
    place for review of such products when they become available.
    
    B. Costs
    
        8. One comment stated the FDA may also have underestimated the cost 
    associated with the submission of 510(k) and PMA's and compliance 
    inspections for many firms engaged in manufacturing IHC's for research 
    purposes that will be required to register under the new rule.
        This comment was made under the mistaken assumption that this rule 
    applied to manufacturers of research products as well as to 
    manufacturers of IHC's marketed for diagnostic use. As noted 
    previously, this rule does not apply to manufacturers of research 
    products and, therefore, imposes no new burden on them. FDA also 
    believes that this comment was made under the mistaken assumption that 
    this rule would create a new requirement for firms to comply with 
    CGMP's. As discussed previously, the requirement to meet CGMP's is not 
    the result of this rulemaking; manufacturers of IHC devices marketed 
    for diagnostic use have always been required to comply with CGMP's 
    under section 520(f) of the act (21 U.S.C. 360j(f). Finally, FDA 
    reiterates that it has reconsidered its position since the proposed 
    rule and has established a classification scheme that does not require 
    510(k)'s for the majority of these IHC devices, and that places most 
    remaining IHC's in class II. Therefore, existing firms that are 
    currently in compliance with CGMP's should not experience any increased 
    costs because of this rule.
    
    C. Definition of IHC
    
        9. One comment supported the proposed definition for IHC reagents 
    and test kits because the definition distinguished between IHC reagents 
    and analyte specific reagents (ASR's) or flow cytometry reagents. The 
    comment also supported the use of performance claims and directions for 
    use with IHC's.
        FDA agrees with this comment. The definition of IHC's and labeling 
    requirements have been retained in the final rule.
        10. One comment was concerned that the IHC definition was 
    ``technology-specific'' and limited to only those devices that employ 
    monoclonal or polyclonal antibodies. The comment
    
    [[Page 30137]]
    
    argued that the operating technology used by the device should be of 
    secondary consideration, provided that the test was intended for 
    adjunctive use along with other conventional histopathology techniques.
        FDA disagrees with this comment. Although the operating technology 
    of the device is of primary importance in identifying an IHC, the 
    intended use of the device will establish its regulatory class. The 
    final rule provides a broad and inclusive regulatory path for 
    commercialization of new versions of currently available IHC devices or 
    IHC devices that are intended to detect a new analyte in tissues or 
    cells. This classification/reclassification is intended to decrease the 
    burden on FDA and industry by obviating the need to individually 
    classify IHC devices that detect previously identified or newly 
    identified analytes.
    
    D. Estrogen and Progesterone Receptors
    
        11. Three comments recommended that ER/PR's be placed in class II 
    instead of class III, as had been proposed. One comment argued that 
    regulating hormone receptors as class III medical devices may limit the 
    availability of an important testing modality, forcing patients to rely 
    upon less accurate methodology for testing results. Two comments 
    maintained that ER/PR's should not be class III because they were not 
    used as stand-alone tests; the information they provided was 
    substantially dependent on other pathological or cytopathological 
    aspects of the specimen, and these tests did not have novel claims not 
    supported by current widely accepted scientific pathophysiologic 
    principles. A third comment recommended reclassifying ER/PR assays into 
    class II because it was likely that there was a sufficient accumulated 
    history of safe and effective use of the tests to support the 
    reclassification and because FDA had published a guideline for 
    premarket submissions of ER/PR assays that could be used as a special 
    control.
        FDA agrees with these comments and has modified the regulation 
    accordingly. The first IHC tests for ER/PR's were in vitro steroid-
    binding chemical assays that used dextran-coated charcoal to separate 
    bound from free fractions. These IHC tests were subject to class III 
    premarket approval because there was no substantially equivalent 
    legally marketed predicate device, a necessary requirement to qualify 
    for premarket notification (510(k)). There were additional safety and 
    effectiveness considerations raised by these devices, including the 
    likelihood that ER/PR results would be used as stand-alone test results 
    that would serve as the basis for choice of therapy and the inability 
    to confirm these results by other IVD tests or clinical procedures. 
    However, after evaluating the comments and reviewing the peer-reviewed 
    literature regarding use of these IHC's, FDA believes that IHC's for 
    estrogen, progesterone, or other hormone receptors now can be 
    classified/reclassified into class II under the final regulation when 
    their claims are widely accepted and there is valid scientific evidence 
    to support those claims.
        12. Two comments stated that there was confusion about which 
    products were covered under the proposed rule and used estrogen 
    receptor (ER) as an example. The comment suggested it was not 
    appropriate to place all ER's in a single class because that class 
    could not take into account differences between broad antigen 
    recognition and clones reacting with certain epitopes or populations of 
    ER, even though there was no clinical utility for some clones.
        FDA disagrees with these comments and believes they are based on a 
    misunderstanding of the proposed rule. FDA does not intend to require 
    premarket submissions for reagents or tests kits that are for 
    ``research use only.'' The regulation requires premarket submissions 
    only for ER/PR reagents or test kits that are intended to be marketed 
    ``for in vitro diagnostic use'' to obtain clinical information. If an 
    IHC reagent or test kit marketed for clinical use includes antibodies, 
    FDA requires the IHC manufacturer to identify the clones of those 
    monoclonal antibodies used in the IHC reagent or test kit that support 
    that intended use.
    
    E. Guidance Document
    
        13. One comment argued that a guidance document cannot be a special 
    control because using a draft guidance document as a special control is 
    an inappropriate use of guidance documents, and that it seemed to 
    contradict the interim policy announced by FDA concerning guidances to 
    use a guidance as if it were a rule.
        FDA disagrees that a guidance document cannot be a special control. 
    ``Guidelines (including guidelines for the submission of clinical data 
    in premarket notification submission * * *)'' are expressly listed in 
    section 513(a)(1)(B) of the act as an example of special controls. In 
    addition, FDA guidance documents are specifically listed as potential 
    special controls in the legislative history of the Safe Medical Devices 
    Act of 1990 (H. Committee Rept. 101-808, October 5, 1990, p. 28).
        Moreover, consistent with FDA's policy on GGP's, the agency 
    published for public comment a ``draft'' of this FDA guidance document 
    in advance of it being used as a special control (62 FR 8961, February 
    27, 1997). The guidance entitled ``FDA Guidance for Submission of 
    Immunohistochemical Applications to the FDA'' was developed by FDA in 
    conjunction with professional organizations, manufacturers of 
    immunohistochemical products, and the advisory committees of FDA. The 
    draft has been revised in response to public comments, and the guidance 
    is available to the public as delineated in section VI of this 
    document.
        FDA is using this guidance in conformance with its policy 
    concerning guidances. The guidance is intended to provide information 
    about acceptable ways to facilitate the gathering of data to ensure 
    reasonable safety and effectiveness of those IHC devices whose safety 
    and effectiveness cannot be ensured by general controls alone. Although 
    the guidance represents FDA's best thinking about ways to efficiently 
    and effectively gather and submit data to support the marketing of 
    these devices, neither the manufacturer nor the agency is bound by the 
    details of that guidance. As stated in the guidance document, 
    manufacturers are free to use alternative methods that achieve the same 
    underlying standard of safety and effectiveness.
    
    F. Impact of Proposed Rule
    
        14. One comment stated the author's belief that IHC's were utilized 
    under the guidance of board certified pathologists a significant 
    percentage of the time and had a performance record equal to or greater 
    than stains used since the turn of the century. This comment maintained 
    that undue restrictions on the use of the reagents would impact on the 
    availability of existing and future antibodies to the detriment of 
    patient care.
        FDA agrees in part with this comment. FDA is treating IHC's used to 
    provide adjunctive information the same as H & E stains by classifying 
    these products into class I and exempting them from the requirement to 
    submit 510(k)'s. Histopathologic and cytologic diagnostic tests that 
    use either conventional stains such as hematoxylin and eosin or IHC 
    methodologies are part of a multistep process that requires the direct 
    supervision of a qualified pathologist or other laboratory scientist to 
    ensure safe and effective results. However, FDA does not consider IHC's 
    to be equivalent to conventional biologic stains and has not exempted 
    IHC's from CGMP requirements although it did exempt the stains from 
    CGMP's. FDA considers IHC's to be
    
    [[Page 30138]]
    
    more complex to develop, manufacture, and standardize. FDA exempted 
    conventional biologic stains from premarket notification and compliance 
    with CGMP's because these stains have well-established chemical and 
    physical specifications and quality assurance. In addition, there are 
    voluntary organizations such as the Biologic Stain Commission that test 
    and certify the specifications of biologic stains. The final rule 
    ensures that all commercialized IHC reagents and test kits for in vitro 
    diagnosis are manufactured under general controls including CGMP, 
    thereby enhancing reliability and consistency for end users of these 
    products.
    
    G. Panel Meeting
    
        15. One comment stated that the October 21, 1994, meeting of the 
    Hematology and Pathology Devices Panel (the Panel) was procedurally 
    flawed. The comment referenced a complaint filed by a Washington, DC, 
    law firm that FDA inaccurately described the regulations to the Panel 
    members, and that this alleged misinformation was the basis for their 
    recommendations. The comment recommended that it be stated in the 
    administrative record that the advisory Panel meeting was procedurally 
    flawed, that the Panel recommendations should not be used to support 
    the decisions made by FDA about the classification of these products, 
    or that the Panel meeting should be invalidated and reconvened for 
    further consideration of the issue.
        This comment refers to a complaint about a FDA employee's public 
    comment that CGMP inspections for class I IVD device manufacturers are 
    so relatively low on the priority list for the agency actions that 
    there is a strong likelihood that these manufacturers will not get 
    timely CGMP inspections. The comment argues that this statement exerted 
    undue influence on the Panel. Because requiring compliance with CGMP's 
    was a high priority for pathologists and other laboratory scientists, 
    the comment asserts that the Panel recommended that IHC's should be 
    class II medical devices in large part to ensure timely CGMP 
    inspections.
        FDA does not agree with this comments' characterization of the 
    Panel meeting. While FDA agrees the Panel was concerned that IHC's be 
    subject to CGMP's, FDA believes the availability and need for a special 
    control is the basis for the Panel's recommendation that most IHC's be 
    classified as class II. However, as discussed previously, FDA has 
    reconsidered this recommendation of the Panel and amended the proposed 
    rule to place most IHC's in class I and exempt from premarket 
    notification. FDA does not believe that class II regulation is required 
    for all IHC reagents and test kits for the reasons discussed in section 
    III.B of this document. Unless specifically exempted, all manufacturers 
    of FDA regulated medical devices must comply with general controls, 
    which include CGMP's, regardless of whether or not the device is in 
    class I, II, or III, or exempt from premarket notification. While FDA 
    acknowledges that its limited resources do not allow inspections to be 
    as frequent as it might wish, the agency's experience shows that 
    competitors and dissatisfied customers will provide the agency with 
    information about circumstances that require more immediate followup.
    
    H. Practice of Medicine
    
        16. One comment argued that whether a reagent was used to make 
    ``significant medical decisions'' was an inappropriate criterion for 
    classification. The comment argued that classifications did not rely on 
    the intent of the manufacturer, but on the physician's usage, which the 
    comment argued was the practice of medicine and beyond the 
    responsibility of a manufacturer. The comment stated that the basis for 
    device classification in this rule was medical practice (e.g. 
    ``significant medical decisions,'' ``markers of clinically significant 
    genetic mutations,'' and ``adjunctive diagnostic information that was 
    ordinarily reported as independent diagnostic information to the 
    ordering clinician'') that was inconsistent with the current 
    requirements of law for determining classification and an attempt to 
    regulate the practice of medicine.
        FDA does not regulate medical practice. FDA regulates the 
    manufacturers of IVD tests to ensure reasonable safety and 
    effectiveness of these products for the claimed intended use and 
    indications for use. The rule focuses on the use to which the 
    information being generated by the IHC will be put because it is the 
    IHC's intended use that determines the level of safety and 
    effectiveness that must be assured. An IHC manufacturer must document 
    the safety and effectiveness of these intended uses and indications for 
    use with valid scientific evidence. If a laboratorian or clinician uses 
    an IHC test for purposes not recommended by the IHC manufacturer, these 
    would be off-label uses that become the responsibility of the 
    laboratory scientist or clinician to establish and validate.
        The level of risk to a patient associated with use of an IVD must 
    account for the consequences of inaccurate results. The level of risk 
    rises with the seriousness of consequences from a false result, the 
    likelihood of the false result occurring, and the number of persons 
    likely to be exposed to the risk of a false result. All of these risks 
    are weighed against the benefit of the assay if it is performed 
    accurately for its intended use and the risk from not having the 
    results from the IHC assay. When evaluation of risks and benefits 
    requires FDA to seek information about the use to which test results 
    are to be put, such data collection is not an intrusion into the 
    practice of medicine but the necessary review of information that is 
    essential to establish whether the product can be marketed as labeled 
    by the manufacturer with reasonable assurance of safety and 
    effectiveness.
    
    I. Prescription
    
        17. One comment stated that it was inappropriate to include 
    Sec. 801.109, which provides that antibodies be provided only upon 
    authorization by a physician, as a general control applying to IHC's. 
    The comment argued that it would put severe restrictions on a 
    researcher wishing to purchase the reagents and was a complete change 
    from the way IHC's were currently ordered. The comment maintained that 
    many of the requirements of Sec. 801.109 were inappropriate for IHC's, 
    such as frequency or duration of administration and side effects, and 
    that generating and tracking this information would be burdensome to 
    the manufacturers and result in added cost to the customer. The comment 
    added that this requirement appeared to impose a ``drug model'' on 
    device manufacturers. The comment recommended that the proper general 
    control was 21 CFR 801.119. A related comment questioned whether a 
    physician prescription was necessary for the research use of FDA-
    approved and marketed IHC reagents and stated that such a requirement 
    had a high potential to hinder legitimate biomedical research efforts. 
    Five other comments stated that a key concern was that IHC reagents be 
    purchased only on the order of a physician, even if the reagents were 
    being used for research use.
        FDA disagrees with these comments. As stated previously, the rule 
    does not apply to IHC's used for research and FDA does not require any 
    premarket submissions from manufacturers of products labeled and 
    intended ``for research use only.'' FDA does not restrict the purchase 
    of reagents or test kits used for research, and FDA does not require a 
    physician's prescription if
    
    [[Page 30139]]
    
    these products are not to be used for diagnosis or management of 
    patients.
        Section 801.109 applies only to IVD devices intended for clinical 
    use in the diagnosis and management of patients. These devices are 
    required to be in the possession of practitioners licensed by law to 
    use or order such devices. Physicians are not the only practitioners 
    allowed to use or order IVD tests. Other practitioners include 
    dentists, veterinarians, nurses, or others licensed by applicable State 
    law to use or order the use of the device.
    
    J. Research Use
    
        18. Several comments were concerned that the proposed rule would 
    limit basic research by requiring IHC's used only in research to be 
    subject to the requirements of this regulation. Another comment 
    requested clarification about FDA's position with respect to antibodies 
    intended for use as immunohistochemical research reagents and whether 
    such antibodies could be marketed as ASR's. The comment also questioned 
    whether low or moderate complexity clinical laboratories would be able 
    to use these products if the products were marketed as ASR's.
        As discussed previously, FDA does not require premarket submissions 
    from manufacturers or users of in vitro reagents or test kits that are 
    labeled ``for research use only.'' FDA introduced the ASR regulations 
    to allow manufacturers to simplify the commercialization of new ASR's 
    for diagnostic use before these reagents have established performance 
    characteristics. IHC reagents may be marketed as ASR's as long as they 
    comply with the ASR regulations(Sec. 809.10, 21 CFR 809.30, and 
    864.4020). The product must be manufactured under general controls, 
    which include CGMP's. The product cannot be sold with any performance 
    claims, intended use, indications for use or instructions for use. It 
    is the responsibility of the end user to validate the intended use, 
    indications for use, and performance characteristics of the ASR. It is 
    because of the high level of proficiency required of the end user that 
    the ASR regulations restrict the use of ASR's to high complexity 
    laboratories.
    
    K. Reimbursement Status
    
        19. One comment asked FDA to discuss with the Health Care Financing 
    Administration (HCFA) and announce the Medicare reimbursement status 
    of: (1) IHC reagents in the interim period while manufacturers prepared 
    and FDA cleared 510(k) submissions, and (2) IHC reagents that have been 
    designated as ASR's.
        Manufacturers who have questions about HCFA reimbursement should 
    address their questions directly to HCFA. FDA's regulatory decisions 
    are based on providing assurance of safety and effectiveness of these 
    devices and are made independent of HCFA's reimbursement decisions. 
    HCFA does consider FDA's clearance and approval of IVD devices as part 
    of HCFA's decision to approve reimbursement. HCFA's decision to 
    reimburse for IVD devices is a cost-benefit decision about whether the 
    device is reasonable and necessary to establish a diagnosis or for 
    patient management.
    
    L. Small Entities
    
        20. One comment from a trade association requested that FDA re-
    examine its assertion that ``the proposed rule will not have 
    significant impact on a large number of small entities.'' The comment 
    stated it was aware that FDA made no formal study in arriving at its 
    conclusion and has not placed any data in the docket to support the 
    decision. The comment stated that most of the suppliers of antibodies 
    to the research community are small businesses that would be severely 
    affected by a requirement to manufacture small quantities of a large 
    number of products under CGMP regulations. The comment argued that its 
    membership estimates the cost of an antibody submission at between 
    $10,000 and $40,000 per antibody when the manufacturer follows the 
    draft guidance document and that the sales volume of most of these 
    products could not justify this expense.
        Another comment stated that FDA had offered no analysis or study to 
    support its conclusion that there would be no significant economic 
    impact on a substantial number of small entities. The comment stated 
    that, in order for an agency to certify that a rule would not have a 
    significant economic impact on a substantial number of small entities, 
    an agency must first demonstrate that it had made a reasonable 
    preliminary assessment of what constituted a small entity in the 
    affected industry, the number of small entities likely to be affected, 
    and the impact of the regulation on those businesses. The comment 
    argued that FDA had an affirmative obligation to explain why reasonable 
    alternatives were rejected and to demonstrate that there had been 
    outreach to the affected industry.
        These comments were made under the mistaken assumption that this 
    rule applied to manufacturers of research IHC products. Manufacturers 
    of preamendment IHC medical devices for diagnostic use already are 
    required to comply with general controls applicable to all 
    manufacturers of devices, and this rule does not add any new obligation 
    with respect to that requirement. All postamendment IHC devices require 
    premarket approval or an order finding substantial equivalence unless 
    exempted by statute or regulation. The effect of this rule is to 
    establish that the majority of these postamendment devices will now be 
    in class I and exempt from any premarket submissions. Although these 
    devices will continue to be subject to general controls, the rule will 
    impose no new burdens for most of these devices. In fact, the rule will 
    reduce the economic burden for many of these manufacturers because they 
    will no longer be required to submit PMA or 510(k) applications for 
    most of their products.
        FDA has prepared an analysis of impact for this rule in section VII 
    of this document and alternatives to the final rule are discussed 
    there. In response to the comment on agency outreach to the affected 
    industry, FDA notes that it convened a public meeting of the Hematology 
    and Pathology Devices Panel in October 1994 and received written 
    comments from interested parties before, during, and after the meeting.
        FDA believes this final regulation will not have a significant 
    adverse impact on small businesses that currently are in the business 
    of manufacturing IHC's. FDA believes the regulation ensures the public 
    that IHC reagents and test kits are reasonably safe and effective for 
    their intended use. At the same time, FDA does not intend or expect the 
    regulation to impede the timely development of safe and effective 
    medical devices. The level of regulation is designed to be in 
    proportion to the need for regulatory oversight based on claims and 
    promotion that a manufacturer makes for its products and the risks the 
    products pose. A product that is to be sold and used as a ``for 
    research use only'' reagent or test kit does not fall within the scope 
    of the rule and the manufacturer of these devices currently does not 
    have to comply with CGMP's. However, an IHC manufacturer that wants to 
    promote reagents for diagnosis or management of patients is required to 
    comply with the final rule and provide valid scientific evidence to 
    support its claims for the intended use of the device, indications for 
    use, and performance characteristics, unless exempted by the rule.
        The agency notes that the final rule exempts most IHC's from 
    premarket submission requirements because the majority of IHC's are 
    adjunctive and will be classified as class I, exempt from
    
    [[Page 30140]]
    
    premarket notification. Even when premarket submissions are required, 
    for the most part premarket notification (510(k)) is required, rather 
    than premarket approval. Most of the remaining IHC's will be classified 
    as class II devices because they provide independent information and 
    have claims that are widely accepted and supported by valid scientific 
    evidence. Moreover, FDA is providing guidance for those IHC's requiring 
    510(k)'s. The guidance entitled ``FDA Guidance for Submission of 
    Immunohistochemical Applications to the FDA'' serves as a special 
    control to assist sponsors in collecting and presenting these data to 
    FDA for clearance of their class II devices. The guidance may also 
    serve as a resource for manufacturers of class I IHC's who do not have 
    to submit 510(k)'s but will nevertheless want to properly develop and 
    validate their products prior to marketing. PMA's are only needed for 
    those IHC's that do not meet the class I and II criteria.
        The regulation does require manufacturers of class II and class III 
    IHC's to submit valid scientific evidence to support the intended use 
    of these products. In many cases, much of the necessary data may be 
    available in the peer reviewed/refereed scientific literature. In those 
    cases where published data are available, the burden on the 
    manufacturer is minimal, and the guidance being established as a 
    special control can provide small and large firms with information to 
    help identify and submit such data. However, published data may not be 
    available for other IHC reagents or test kits that the manufacturer 
    wishes to modify or for new intended uses or indications for use of 
    these IHC devices. In those cases, manufacturers will have to gather 
    new testing data to support the claims.
        There also may be IHC reagents or test kits that do not have the 
    potential volume of sales to justify any manufacturer's business 
    decision to comply with FDA's requirements for data to support the 
    reasonable assurance of safety and effectiveness for particular labeled 
    claims and uses. In those cases, the manufacturer may commercialize the 
    IHC products with lesser performance claims or as an ASR and transfer 
    the responsibility for validation of the finished assay to the user. In 
    addition, manufacturers of low use/low revenue products may choose to 
    commercialize the IHC under the humanitarian device exemption 
    procedures (21 CFR part 800, subpart H). Each IHC manufacturer, whether 
    a large or small firm, will be able to control the impact of the final 
    rule on its business by carefully evaluating the claims and uses it 
    intends to promote for particular products.
        The minimal level of IHC IVD device regulation will be the ASR 
    regulation. Class I ASR's are exempt from premarket notification, but 
    must be manufactured in compliance with general controls to be legally 
    marketed as IVD reagents for diagnosis and management of patients. 
    Because ASR's do not require data to support an intended use, 
    indications for use, or performance characteristics; ASR product 
    labeling cannot include any claims for intended use, indications for 
    use, or performance characteristics. The sale of ASR's is restricted to 
    high complexity laboratories that are able to take the responsibility 
    for establishing and validating the reagent for an intended use, 
    indications for use, and performance characteristics of the finished 
    assay (62 FR 62243, November 21, 1997).
        21. One comment requested that new hearings be held and that 
    representatives of all small companies who will be affected by the 
    regulation be given an opportunity to speak and be heard not only by 
    FDA but also by congressional representatives.
        The Administrative Procedures Act gives agencies discretion over 
    whether to hold oral hearings in connection with informal rulemakings 
    (5 U.S.C. 553(c)). FDA believes that providing an opportunity for 
    written comment on the proposed rule has provided sufficient 
    opportunity for small entities to comment on this rulemaking. Moreover, 
    FDA has already held a public hearing soliciting comment on the 
    classification of immunohistochemical devices. That hearing, which was 
    convened on October 21, 1994, was open to all interested parties, 
    including small business entities and their representatives. Input from 
    regulated industry played an important part in shaping FDA's proposal 
    for regulating IHC's. Moreover, FDA has made extensive changes to the 
    final rule based on the agency's evaluation of the written comments. 
    FDA believes that it would be an unnecessary use of scarce agency 
    resources to hold a hearing for this rulemaking. Furthermore, FDA has 
    no authority to require congressional attendance or participation at 
    the agency's hearings.
    
    VI. Access to the Special Control
    
        To receive the special control entitled ``FDA Guidance for 
    Submission of Immunohistochemistry Applications to the FDA,'' FDA, 
    Center for Devices and Radiologic Health, 1998, via fax machine, call 
    the CDRH Facts-On-Demand system at 800-399-0381 or 301-827-0111 from a 
    touch-tone telephone. At the first voice prompt, press 1 to access the 
    Division of Small Manufacturers Assistance (DSMA) Facts. At the second 
    voice prompt, press 2, and then enter the document No. 364 followed by 
    the pound sign (#). Then follow the remaining voice prompts to complete 
    your request.
        CDRH maintains an entry on the World Wide Web (www) for easy access 
    to information, including text, graphics, and files that may be 
    downloaded to a PC with access to the www. The CDRH home page is 
    updated on a regular basis and includes the guidance cited previously, 
    as well as other guidance documents; device safety alerts; Federal 
    Register reprints; information on premarket submissions (including 
    lists of approved applications and manufacturers' addresses); small 
    manufacturers' assistance; and information on video conferencing and 
    electronic submissions, mammography matters, and other device-oriented 
    information. The CDRH home page may be accessed at http://www.fda.gov/
    cdrh.
        A text-only version of the CDRH Web site is also available from a 
    computer or VT-100 compatible terminal by dialing 800-222-0185 
    (terminal settings are 8/1/N). Once the modem answers, press ENTER 
    several times and then select menu choice 1: FDA BULLETIN BOARD 
    SERVICE. From there follow instructions for logging in, and at the BBS 
    TOPICS PAGE, arrow down to the FDA home page (do not select the first 
    CDRH entry). Then select MEDICAL DEVICES AND RADIOLOGICAL HEALTH. From 
    there select CENTER FOR DEVICES AND RADIOLOGICAL HEALTH for general 
    information, or arrow down for specific topics.
    
    VII. Analysis of Impacts
    
        FDA has examined the impacts of the final rule under Executive 
    Order 12866 and the Regulatory Flexibility Act (5 U.S.C. 601-612) (as 
    amended by subtitle D of the Small Business Regulatory Enforcement 
    Fairness Act of 1966 (Pub. L. 104-121), and the Unfunded Mandates 
    Reform Act of 1995 (Pub. L. 104-4)). Executive Order 12866 directs 
    agencies to assess all costs and benefits of available regulatory 
    alternatives and, when regulation is necessary, to select regulatory 
    approaches that maximize net benefits (including potential economic, 
    environmental, public health and safety, and other advantages; 
    distributive impacts; and equity). The agency believes that this final 
    rule is consistent with the regulatory philosophy and principles 
    identified in the Executive Order. In addition, the final rule has been 
    determined to be a
    
    [[Page 30141]]
    
    significant regulatory action as defined by the Executive Order and so 
    is subject to review under the Executive Order.
    
    A. Description of Impact
    
        The intended purpose of this final rule is to regulate pre- and 
    postamendment IHC devices in a consistent manner. Presently, 
    preamendment IHC's are unclassified, while most postamendment IHC's are 
    statutorily classified into class III. Both pre- and postamendment 
    devices are currently subject to general controls, and postamendment 
    devices require FDA approval before marketing. This rule will 
    categorize IHC devices based on their potential risk to public health 
    into one of the three device classes. The great majority of IHC's will 
    be categorized as class I devices and will be exempt from premarket 
    notification. The IHC's that fall into class II will require premarket 
    clearance and be subject to a special control, in addition to general 
    controls. Currently, there are no IHC devices on the market that will 
    fall into class III.
        The economic impact of this rule on manufacturers of IHC's will be 
    negligible. Currently, manufacturers of all IHC devices are required to 
    follow general controls. Under this rule, most preamendment IHC devices 
    marketed with their original (pre-1976) claims will be categorized as 
    class I devices and consequently exempt from premarket notification 
    requirements. Therefore, there will be no change in the regulatory 
    requirements that manufacturers of these devices must follow. The 
    manufacturers of postamendment devices may realize an economic savings 
    as a result of this rule. Manufacturers of the postamendment devices, 
    which are currently statutorily classified into class III, would have 
    been required to submit 510(k)'s or PMA's to be legally marketed. The 
    final rule classifies most IHC's in class I and exempts them from 
    premarket notification, eliminating the requirement for manufacturers 
    to make premarket submissions for these devices. Most postamendment 
    devices that will require submissions have been classified into class 
    II and will not require a PMA approval. One comment suggested that the 
    cost to submit a 510(k) ranged from $10,000 to $40,000 per antibody 
    (see comment 21 of this document). The cost of preparing a PMA would be 
    much higher. In addition, the special control established by this rule 
    for class II IHC's is a guidance document intended to help 
    manufacturers prepare 510(k)'s efficiently and effectively.
        FDA can not reliably estimate the total number of manufacturers of 
    IHC's affected by this rule. Currently, there are fewer than 25 firms 
    listed with the agency as manufacturers of 510(k) or PMA IHC devices. 
    Most, if not all, of these firms are small, based on the Small Business 
    Administration's definition of a small medical device entity (fewer 
    than 500 employees).
    
    B. Response to Comments by Small Business
    
        Some small businesses and the Small Business Administration 
    commented that the proposed rule would impose a severe economic burden 
    on IHC manufacturers, driving some companies out of business. These 
    comments misunderstood the scope of the proposed rule by assuming that 
    it would apply to IHC's used for research. In fact, there will be no 
    new regulatory costs for research firms. As discussed previously, FDA 
    has classified the majority of IHC devices as class I, exempt from 
    premarket notification. The final rule also narrowed the identification 
    of class III devices so that many devices that would have been class 
    III under the proposal will be class II under the final rule and not 
    require a PMA.
        There were also comments from small businesses that stated the 
    rule, as proposed, would have a negative effect on new product 
    introduction. With the changes made to the proposal, the agency 
    believes that the final rule will have no negative effect on new 
    product introduction and will introduce consistency in the regulation 
    of IHC's. Currently, postamendment IHC's require PMA's or 510(k)'s. 
    With this rule, most new products will be classified as class I exempt 
    from premarket notification.
    
    C. Summary
    
        In the proposed rule, FDA considered requiring 510(k)'s or PMA's 
    for all IHC's. In response to comments, the agency reconsidered its 
    position and determined that the necessary safeguards to public health 
    could be achieved with general controls alone for the majority of 
    currently marketed IHC's. Because this rule classifies these 
    postamendment devices into class I, exempt from premarket notification, 
    or into class II, the cost of the rule will be far below the $100 
    million threshold that determines an economically significant 
    regulation under Executive Order 12866 and the Unfunded Mandates Reform 
    Act. Because the rule will safeguard the public health and impose 
    almost no new burden on industry, the agency certifies that the rule 
    will not have a significant impact on a substantial number of small 
    entities.
    
    VIII. References
    
        The following references have been placed on display in the Dockets 
    Management Branch (address above) and may be seen by interested persons 
    between 9 a.m. and 4 p.m., Monday through Friday.
        1. Baddoura, F. K. , C. Cohen, E. R. Unger, P. B. DeRose, and M. 
    Chenggis, ``Image Analysis for Quantitation of Estrogen Receptor in 
    Formalin-fixed Paraffin-embedded Sections of Breast Carcinoma,'' 
    Modern Pathology, January 1991, 4(1):91-95.
        2. Esteban, J. M., P. L. Kandalaft, P. Mehta, T. L. Odom-Maryon, 
    S. Bacus, and H. Battifora, ``Improvement of the Quantification of 
    Estrogen and Progesterone Receptors in Paraffin-embedded Tumors by 
    Image Analysis,'' American Journal of Clinical Pathology, January 
    1993, 99(1):32-38.
        3. Esteban, J. M., C. Ahn, P. Mehta, and H. Battifora, 
    ``Biologic Significance of Quantitative Estrogen Receptor 
    Immunohistochemical Assay by Image Analysis in Breast Cancer,'' 
    American Journal of Clinical Pathology, August 1994, 102(2):158-162.
        4. Esteban, J. M., C. Ahn, H. Battifora, and B. Felder, 
    ``Predictive Value of Estrogen Receptors Evaluated by Quantitative 
    Immunohistochemical Analysis in Breast Cancer,'' American Journal of 
    Clinical Pathology, October 1994, 102 (4 Supplement 1), S9-S12.
        5. Molino, A., R. Micciolo, M. Turazza, F. Bonetti, Q. Piubello, 
    A. Corgnati, L. Sperotto, G. Martignoni, A. Bonetti, R. Nortilli, et 
    al., ``Estrogen Receptors in 699 Primary Breast Cancers: A 
    Compatison of Immunohistochemical and Biochemical Methods,'' Breast 
    Cancer Research Treatment, June 1995, 34(3):221-228.
    
    IX. Environmental Impact
    
        The agency has determined under 21 CFR 25.24(e)(2) that this action 
    is of a type that does not individually or cumulatively have a 
    significant effect on the human environment. Therefore, neither an 
    environmental assessment nor an environmental impact statement is 
    required.
    
    List of Subjects in 21 CFR Part 864
    
        Blood, Medical devices, Packaging and containers.
        Therefore, under the Federal Food, Drug, and Cosmetic Act and under 
    authority delegated to the Commissioner of Food and Drugs, 21 CFR part 
    864 is amended as follows:
    
    PART 864--HEMATOLOGY AND PATHOLOGY DEVICES
    
        1. The authority citation for 21 CFR part 864 continues to read as 
    follows:
    
        Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.
    
        2. Section 864.1860 is added to subpart B to read as follows:
    
    [[Page 30142]]
    
    Sec. 864.1860  Immunohistochemistry reagents and kits.
    
        (a) Identification. Immunohistochemistry test systems (IHC's) are 
    in vitro diagnostic devices consisting of polyclonal or monoclonal 
    antibodies labeled with directions for use and performance claims, 
    which may be packaged with ancillary reagents in kits. Their intended 
    use is to identify, by immunological techniques, antigens in tissues or 
    cytologic specimens. Similar devices intended for use with flow 
    cytometry devices are not considered IHC's.
        (b) Classification of immunohistochemistry devices--(1) Class I 
    (general controls). Except as described in paragraphs (b)(2) and (b)(3) 
    of this section, these devices are exempt from the premarket 
    notification requirements in part 807, subpart E of this chapter. This 
    exemption applies to IHC's that provide the pathologist with adjunctive 
    diagnostic information that may be incorporated into the pathologist's 
    report, but that is not ordinarily reported to the clinician as an 
    independent finding. These IHC's are used after the primary diagnosis 
    of tumor (neoplasm) has been made by conventional histopathology using 
    nonimmunologic histochemical stains, such as hematoxylin and eosin. 
    Examples of class I IHC's are differentiation markers that are used as 
    adjunctive tests to subclassify tumors, such as keratin.
        (2) Class II (special control, guidance document: ``FDA Guidance 
    for Submission of Immunohistochemistry Applications to the FDA,'' 
    Center for Devices and Radiologic Health, 1998). These IHC's are 
    intended for the detection and/or measurement of certain target 
    analytes in order to provide prognostic or predictive data that are not 
    directly confirmed by routine histopathologic internal and external 
    control specimens. These IHC's provide the pathologist with information 
    that is ordinarily reported as independent diagnostic information to 
    the ordering clinician, and the claims associated with these data are 
    widely accepted and supported by valid scientific evidence. Examples of 
    class II IHC's are those intended for semiquantitative measurement of 
    an analyte, such as hormone receptors in breast cancer.
        (3) Class III (premarket approval). IHC's intended for any use not 
    described in paragraphs (b)(1) or (b)(2) of this section.
        (c) Date of PMA or notice of completion of a PDP is required. As of 
    May 28, 1976, an approval under section 515 of the Federal Food, Drug, 
    and Cosmetic Act is required for any device described in paragraph 
    (b)(3) of this section before this device may be commercially 
    distributed. See Sec. 864.3.
    
        Dated: February 6, 1998.
    D.B. Burlington,
    Director, Center for Devices and Radiological Health.
    [FR Doc. 98-14605 Filed 6-2-98; 8:45 am]
    BILLING CODE 4160-01-F
    
    
    

Document Information

Effective Date:
8/17/1998
Published:
06/03/1998
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final rule.
Document Number:
98-14605
Dates:
This rule is effective August 17, 1998.
Pages:
30132-30142 (11 pages)
Docket Numbers:
Docket No. 94P-0341
RINs:
0910-ZA10
PDF File:
98-14605.pdf
CFR: (3)
21 CFR 801.109
21 CFR 864.1860
21 CFR 864.4010