95-13831. Dental Devices; Effective Date of Requirement for Premarket Approval of Endodontic Dry Heat Sterilizer  

  • [Federal Register Volume 60, Number 109 (Wednesday, June 7, 1995)]
    [Proposed Rules]
    [Pages 30032-30037]
    From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
    [FR Doc No: 95-13831]
    
    
    
    =======================================================================
    -----------------------------------------------------------------------
    
    DEPARTMENT OF HEALTH AND HUMAN SERVICES
    
    Food and Drug Administration
    
    21 CFR Part 872
    
    [Docket No. 95N-0033]
    
    
    Dental Devices; Effective Date of Requirement for Premarket 
    Approval of Endodontic Dry Heat Sterilizer
    
    AGENCY: Food and Drug Administration, HHS.
    
    ACTION: Proposed rule; opportunity to request a change in 
    classification.
    
    -----------------------------------------------------------------------
    
    SUMMARY: The Food and Drug Administration (FDA) is proposing to require 
    the filing of a premarket approval application (PMA) or a notice of 
    completion of a product development protocol (PDP) for the endodontic 
    dry heat sterilizer, a medical device. The agency also is summarizing 
    its proposed findings regarding the degree of risk of illness or injury 
    designed to be eliminated or reduced by requiring the device to meet 
    the statute's approval requirements, and the benefits to the public 
    from use of the device. In addition, FDA is announcing the opportunity 
    for interested persons to request the agency to change the 
    classification of the device based on new information.
    
    DATES: Written comments by September 5, 1995; requests for a change in 
    classification by June 22, 1995. FDA intends that, if a final rule 
    based on this proposed rule is issued, PMA's will be required to be 
    submitted within 90 days of the effective date of the final rule.
    
    ADDRESSES: Submit written comments or requests for a change in 
    classification to the Dockets Management Branch (HFA-305), Food and 
    Drug Administration, rm. 1-23, 12420 Parklawn Dr., Rockville, MD 20857.
    
    FOR FURTHER INFORMATION CONTACT: Joseph M. Sheehan, Center for Devices 
    and Radiological Health (HFZ-84), Food and Drug Administration, 2094 
    Gaither Rd., Rockville, MD 20850, 301-594-4765.
    
    SUPPLEMENTARY INFORMATION:
    
    I. Background
    
        Section 513 of the Federal Food, Drug, and Cosmetic Act (the act) 
    (21 U.S.C. 360c) requires the classification of medical devices into 
    one of three regulatory classes: Class I (general controls), class II 
    (special controls), and class III (premarket approval). Generally, 
    devices that were on the market before May 28, 1976, the date of 
    enactment of the Medical Device Amendments of 1976 (the amendments) 
    (Pub. L. 94-295), and devices marketed on or after that date that are 
    substantially equivalent to such devices, have been classified by FDA. 
    For the sake of convenience, this preamble refers to both the devices 
    that were on the market before May 28, 1976, and the substantially 
    equivalent devices that were marketed on or after that date as 
    ``preamendments devices.''
        Section 515(b)(1) of the act (21 U.S.C. 360e(b)(1)) establishes the 
    requirement that a preamendments device that FDA has classified into 
    class III is subject to premarket approval. A preamendments class III 
    device may be commercially distributed without an approved PMA or 
    notice of completion of a PDP until 90 days after FDA issues a final 
    rule requiring premarket approval for the device, or 30 months after 
    final classification of the device under section 513 of the act, 
    whichever is later. Also, a preamendments device, subject to the 
    rulemaking procedure under section 515(b) of the act, is not required 
    to have an approved investigational device exemption (IDE) (21 CFR part 
    812) contemporaneous with its interstate distribution until the date 
    identified by FDA in the final rule requiring the submission of a PMA 
    for the device.
        Section 515(b)(2)(A) of the act provides that a proceeding to issue 
    a final rule to require premarket approval shall be initiated by 
    publication of a notice of proposed rulemaking containing: (1) The 
    proposed rule; (2) proposed findings with respect to the degree of risk 
    of illness or injury designed to be eliminated or reduced by requiring 
    the device to have an approved PMA or a declared completed PDP and the 
    benefit to the public from the use of the device; (3) an opportunity 
    for the submission of comments on the proposed rule and the proposed 
    findings; and (4) an opportunity to request a change in the 
    classification of the device based on new information relevant to the 
    classification of the device.
        Section 515(b)(2)(B) of the act provides that if FDA receives a 
    request for a change in the classification of the device within 15 days 
    of the publication of the notice, FDA shall, within 60 days of the 
    publication of the notice, consult with the appropriate FDA advisory 
    committee and publish a notice denying the request for change of 
    classification or announcing its intent to initiate a proceeding to 
    reclassify the device under section 513(e) of the act. If FDA does not 
    initiate such a proceeding, section 515(b)(3) of the act provides that 
    FDA shall, after the close of the [[Page 30033]] comment period on the 
    proposed rule and consideration of any comments received, issue a final 
    rule to require premarket approval, or publish a notice terminating the 
    proceeding. If FDA terminates the proceeding, FDA is required to 
    initiate reclassification of the device under section 513(e) of the 
    act, unless the reason for termination is that the device is a banned 
    device under section 516 of the act (21 U.S.C. 360f).
        If a proposed rule to require premarket approval for a 
    preamendments device is made final, section 501(f)(2)(B) of the act (21 
    U.S.C. 351(f)(2)(B)) requires that a PMA or a notice of completion of a 
    PDP for any such device be filed within 90 days of the date of issuance 
    of the final rule or 30 months after final classification of the device 
    under section 513 of the act, whichever is later. If a PMA or a notice 
    of completion of a PDP is not filed by the later of the two dates, 
    commercial distribution of the device is required to cease. The device 
    may, however, be distributed for investigational use if the 
    manufacturer, importer, or other sponsor of the device complies with 
    the IDE regulations. If a PMA or a notice of completion of a PDP is not 
    filed by the later of the two dates, and no IDE is in effect, the 
    device is deemed to be adulterated within the meaning of section 
    501(f)(1)(A) of the act, and subject to seizure and condemnation under 
    section 304 of the act (21 U.S.C. 334) if its distribution continues. 
    Shipment of the device in interstate commerce will be subject to 
    injunction under section 302 of the act (21 U.S.C. 332), and the 
    individuals responsible for such shipment will be subject to 
    prosecution under section 303 of the act (21 U.S.C. 333). FDA has in 
    the past requested that manufacturers take action to prevent the 
    further use of devices for which no PMA has been filed and may 
    determine that such a request is appropriate for endodontic dry heat 
    sterilizers.
        The act does not permit an extension of the 90-day period after 
    issuance of a final rule within which an application or a notice is 
    required to be filed. The House Report on the amendments states that:
    
        the thirty month `grace period' afforded after classification of 
    a device into class III * * * is sufficient time for manufacturers 
    and importers to develop the data and conduct the investigations 
    necessary to support an application for premarket approval.
    
    (H. Rept. 94-853, 94th Cong., 2d sess. 42 (1976).)
    
    A. Classification of Endodontic Dry Heat Sterilizers
    
        In the Federal Register of August 12, 1987 (52 FR 30082), FDA 
    issued a final rule (Sec. 872.6730 (21 CFR 872.6730)) classifying the 
    endodontic dry heat sterilizer into class III. The preamble to the 
    proposal to classify the device published in the Federal Register of 
    December 30, 1980 (45 FR 86155), included the recommendation of the 
    Dental Device Classification Panel (the panel), of the Medical Devices 
    Advisory Committee, an FDA advisory committee, regarding the 
    classification of the device.
        The panel recommended that the device be in class III (premarket 
    approval) because the device presented an unreasonable risk of illness 
    or injury. According to the panel, the devices failed to sterilize 
    adequately various endodontic and dental instruments. The panel felt 
    that the failures could be the result of: (1) The device not reaching 
    and maintaining an adequate temperature because of a faulty thermostat 
    or (2) the result of unequal heat distribution by the glass beads 
    throughout the well despite sufficient heat. The panel believed that it 
    was not possible to establish an adequate performance standard for the 
    device because satisfactory performance had never been demonstrated. 
    The panel recommended the device to be subject to premarket approval to 
    assure that manufacturers of the device demonstrate satisfactory 
    performance and that further study was necessary to determine the 
    causes of the device's ineffectiveness.
        FDA agreed with the panel's recommendation that endodontic dry heat 
    sterilizers be classified into class III. FDA believed that there was 
    an unreasonable risk of illness or injury because of the potential 
    failure of the device to sterilize dental instruments adequately. FDA 
    believed that there was inadequate information to determine if general 
    controls or a performance standard would provide reasonable assurance 
    of safety and effectiveness.
    
    B. Dates New Requirements Apply
    
        In accordance with section 515(b) of the act, FDA is proposing to 
    require that a PMA or a notice of completion of a PDP be filed with the 
    agency for the endodontic dry heat sterilizer within 90 days after 
    issuance of any final rule based on this proposal. An applicant whose 
    device was legally in commercial distribution before May 28, 1976, or 
    has been found by FDA to be substantially equivalent to such a device, 
    will be permitted to continue marketing the endodontic dry heat 
    sterilizer during FDA's review of the PMA or notice of completion of 
    the PDP. FDA intends to review any PMA for the device within 180 days, 
    and any notice of completion of a PDP for the device within 90 days of 
    the date of filing. FDA cautions that, under section 515(d)(1)(B)(i) of 
    the act, FDA may not enter into an agreement to extend the review 
    period for a PMA beyond 180 days unless the agency finds that `` * * * 
    the continued availability of the device is necessary for the public 
    health.''
        FDA intends that, under Sec. 812.2(d), the preamble to any final 
    rule based on this proposal will state that, as of the date on which a 
    PMA or a notice of completion of a PDP is required to be filed, the 
    exemptions in Sec. 812.2 (c)(1) and (c)(2) from the requirements of the 
    IDE regulations for preamendments class III devices will cease to apply 
    to any endodontic dry heat sterilizer which is: (1) Not legally on the 
    market on or before that date; (2) legally on the market on or before 
    that date but for which a PMA or notice of completion of a PDP is not 
    filed by that date; or (3) for which PMA approval has been denied or 
    withdrawn.
        If a PMA or a notice of completion of a PDP for the endodontic dry 
    heat sterilizer is not filed with FDA within 90 days after the date of 
    issuance of any final rule requiring premarket approval for the device, 
    commercial distribution of the device must cease. The device may be 
    distributed for investigational use only if the requirements of the IDE 
    regulations are met. FDA would not consider an investigation of an 
    endodontic glass bead sterilizer to pose a significant risk as defined 
    in the IDE regulation provided that instruments processed in the device 
    are terminally sterilized by a sterilization process which can be 
    biologically monitored, such as steam, ethylene oxide, or dry heat. If 
    the investigation cannot be so designed, the investigation would 
    constitute a significant risk. The requirements for significant risk 
    devices include submitting an IDE application to FDA for its review and 
    approval. An approved IDE is required to be in effect before an 
    investigation of the device may be initiated or continued. FDA, 
    therefore, cautions that IDE applications should be submitted to FDA at 
    least 30 days before the end of the 90-day period after the final rule 
    is published to avoid interrupting investigations.
    
    C. Description of Device
    
        Endodontic dry heat sterilizers are small electrically heated dry 
    heat sterilizers with a central well containing a heat transfer medium. 
    The types of [[Page 30034]] heat transfer media used in these units 
    have included glass beads, molten metal, metal beads, and salt. The 
    instruments which are to be sterilized are inserted directly into the 
    heat transfer medium. The units are defined in Sec. 872.6730 as devices 
    used to sterilize endodontic and other dental instruments by the 
    application of dry heat which is supplied by the glass beads which have 
    been heated by electricity.
        The proposed rule to require premarket approval of the endodontic 
    dry heat sterilizer applies to devices that were being commercially 
    distributed before May 28, 1976, and to devices that were introduced 
    into commercial distribution since that date which have been found to 
    be substantially equivalent to predicate endodontic dry heat 
    sterilizers.
    
    D. Proposed Findings With Respect to Risks and Benefits
    
        As required by section 515(b) of the act, FDA is publishing its 
    proposed findings regarding: (1) The degree of risk of illness or 
    injury designed to be eliminated or reduced by requiring endodontic dry 
    heat sterilizers to have an approved PMA or a declared completed PDP; 
    and (2) the benefits to the public from the use of the device.
    
    E. Risk Factors
    
        The panel identified the primary risk to health as infection by 
    stating that ``The inability of the device to sterilize adequately 
    endodontic and other dental instruments may lead to transmission of 
    microorganisms among patients and subsequent spread of infection.''
        A review of the literature on endodontic dry heat sterilizers has 
    identified the following problems associated with the use of these 
    devices which contribute to the inability of endodontic dry heat 
    sterilizers to sterilize instruments, including general medical 
    instruments.
    1. Temperature Variation Within the Well
        There are many reports in the literature describing the temperature 
    variation found within the wells of glass bead sterilizers (Refs. 2, 3, 
    4, 7, 10, and 11). Engelhardt et al. (Ref. 4) measured the temperature 
    distribution in four brands of glass bead sterilizers at two different 
    sites from the center and at six different depths in the well. He 
    reported that the temperature within the well varied significantly 
    depending upon location. The temperature was highest closest to the 
    wall and midway down from the surface (Ref. 4). Corner also reported 
    that near the periphery of the well the temperature varied by as much 
    as 10  deg.C over time (Ref. 5). According to Ingle, glass bead 
    sterilizers should not be used as a substitute for dry heat convection 
    or steam sterilizers because of the temperature variations (Ref. 7).
    2. Lack of Methods to Monitor the Recommended Exposure Times for 
    Sterilization of the Instruments
        The manufacturers' recommended exposure times for sterilization of 
    instruments vary from as short as 2 seconds to 45 seconds for 
    sterilizers whose purported operating temperatures are from 218  deg.C 
    to 260  deg.C. However, location of the instruments in the well, the 
    size and mass of the instruments, the number of instruments, and the 
    shape of the instruments must be factored into the amount of time 
    required for sterilization. Larger instruments composed of more metal 
    take more time to heat than smaller instruments. Koehler reported that 
    the time required to raise an instrument's temperature was dependent 
    upon its size. Small instruments such as root canal files heated 
    rapidly, while large instruments such as cotton pliers never reached 
    the specified operating temperature (Ref. 6). Corner reported that 
    instruments such as forceps, scalpels, spatulas, and scissors 
    sterilized in rapid succession caused the temperature in the well to 
    drop an average of 7  deg.C for each instrument and that it took 15 
    minutes for the temperature of the well to recover (Ref. 2). Smith 
    reported sterilization times of 15 seconds to kill orthodontic bands 
    contaminated with Staphylococcus albus and 45 seconds for bands 
    contaminated with Bacillus subtilis spores; but if five bands were 
    sterilized simultaneously, then the sterilization times doubled (Ref. 
    10). Fahid reported that a No. 60 file, which was the largest file 
    tested in the study, was the most difficult to sterilize. The 
    difficulty was attributed to two factors: the large mass of the file, 
    and the air trapped in the deep trough since air is a poor heat 
    conductor (Ref. 5). Engelhardt described sterilization times for 
    endodontic instruments ranging from 15 to more than 100 seconds in 
    glass bead sterilizers, and in some cases, the 100 seconds were not 
    sufficient to achieve sterilization (Ref. 4). Schutt et al. found that 
    it took 60 seconds to sterilize dental burs. He also emphasized that 
    the temperature at the depth of the immersion of the burs should be 
    measured and that the minimum temperature should be at least 175  deg.C 
    at 2 millimeters (mm) below the surface and 240  deg.C at 15 mm below 
    the surface (Ref. 9). It has been reported in the literature that glass 
    bead sterilizers have been shown to be effective only with small 
    instruments that can be imbedded into the heat transfer media and that 
    their effectiveness has not been demonstrated for instruments of larger 
    bulk. The insertion of large instruments would reduce the temperature 
    of the glass beads below the minimum temperature required for 
    sterilization (Ref. 1). Heat conduction in a large, partially imbedded 
    device would be variable.
        Precleaning of the instruments before insertion into the glass bead 
    sterilizer is critical to the effectiveness of the device. Engelhardt 
    demonstrated that if endodontic instruments were contaminated with a 
    protein load (blood), the time required for sterilization was more than 
    doubled. Such adverse conditions can easily be found in infected or 
    gangrenous pulp. Spores, which are more resistant to sterilization 
    processes than vegetative organisms, have been found in the oral cavity 
    and cultured from pulp material (Ref. 4).
    3. Lack of Methods to Monitor the Performance/Sterilization Efficacy of 
    the Device
        There are no identified methods for the routine monitoring of the 
    sterilization efficacy of the endodontic dry heat sterilizer such as 
    the ones which exist with the traditional sterilization methods, i.e., 
    steam autoclaves, hot air dry heat sterilizers, or ethylene oxide 
    sterilizers. Chemical and biological indicators are available for 
    routine monitoring of the efficacy of the cycle parameters and for the 
    validation of the process specifications for these traditional 
    sterilizers. The data in the literature, as noted above, suggest that 
    the user can not be assured that instruments inserted into an 
    endodontic dry heat sterilizer will be reliably exposed to the minimum 
    cycle parameters required for sterilization, i.e., exposure of the 
    device to a set temperature for a specified time.
    4. Variability of the Warm-up Times for Glass Bead Sterilizers
        Reported warm-up times for these devices range from 15 minutes to 
    50 minutes with the average of 15-20 minutes. However, Corner reported 
    that it took up to 30 minutes for the temperature of the glass beads to 
    stabilize even though the manufacturer claimed that the device reached 
    operating temperature within 10 minutes (Ref. 2). [[Page 30035]] 
    5. Maintenance of Sterility After Removal From the Device
        The instructions for use for most of the devices do not instruct 
    the user on the proper procedure to remove instruments from the device, 
    and on how to maintain sterility of the instruments or the processed 
    portion of the instrument during the cool down period. There also 
    exists the possibility that the heat transfer medium could serve as a 
    source of contamination between patients. Because of the reported 
    temperature gradients within the wells, there exists the possibility 
    that heat resistant microorganisms could survive in the cooler regions 
    near the top of the well and contaminate the instruments used upon the 
    next patient as they are removed from the well. Furthermore, because 
    endodontic dry heat sterilizers only process that portion of the 
    instrument which has been inserted into the glass beads, there is the 
    potential of contaminating a sterile field with a device which had not 
    been properly processed.
    6. Possibility of the Heat Transfer Medium Remaining Upon The Devices
        Occasionally the heat transfer media has been observed to adhere to 
    wet instruments. If the particles are not detected before the devices 
    are inserted into the site, then they could cause blockage of the wound 
    site or other adverse effects. This would cause significant problems if 
    the heat transfer media were glass beads or molten metal (Ref. 1).
    
    F. Benefit of the Devices
    
        The endodontic dry heat sterilizer could be used to decontaminate 
    endodontic instruments during a procedure on a single patient provided 
    the instruments are properly cleaned to remove organic debris before 
    insertion into the unit. In theory the number of microorganisms that 
    would be introduced into the same site or into a new site on the same 
    patient during a single procedure would be reduced. Once the procedure 
    is over, the instruments should be processed using traditional methods 
    of decontamination and sterilization before use in the next patient.
    
    G. Need for Information for Risk/Benefit Assessment of the Device
    
        The data in the literature indicate the lack of uniform 
    sterilization parameters among the various glass bead sterilizers which 
    have been marketed. Because of the temperature variation found within 
    the wells of glass bead sterilizers, exposure of an instrument to an 
    adequate sterilizing temperature is difficult to determine and must be 
    confirmed independently for each instrument. Also determination of the 
    sterilization exposure time is dependent upon instrument size and mass. 
    As Koehler noted, some instruments never reach the appropriate 
    temperature because of their size and mass (Ref. 6); and, as noted in 
    the American Dental Association's ``Accepted Dental Therapeutics,'' 
    40th ed., endodontic dry heat sterilizers are not appropriate for large 
    bulk instruments (Ref. 1).
        Review of the claims being made for these devices suggests that 
    manufacturers are expanding the claims beyond those originally defined 
    in Sec. 872.6730. The claims have been expanded to include the 
    sterilization of general medical instruments and electrolysis and 
    acupuncture needles, and to devices not regulated by FDA such as 
    manicurist's instruments. The claims imply that these devices can be 
    used as a substitute for the traditional methods of sterilization. 
    Scarlett noted that endodontic dry heat sterilizers are not 
    sterilizers, but are decontaminating devices and that they should not 
    be used to sterilize instruments between patients (Ref. 8). No system 
    exists for (1) Monitoring the exposure of the instrument to 
    sterilization conditions, or (2) demonstrating that the sterilization 
    exposure parameters have been achieved within the well. Only the 
    portion of the instrument which is inserted into the heat transfer 
    medium has the potential of being sterilized; the portion which is not 
    inserted into the glass beads is not sterilized. The use of endodontic 
    dry heat sterilizers with general medical instruments and with the 
    implication as a substitute sterilization method raises serious safety 
    and effectiveness questions which the manufacturers of these devices 
    have not adequately addressed. There is the serious risk of infection 
    through the use of inadequately processed instruments.
        FDA believes that sufficient information may exist regarding the 
    risks and benefits associated with the device, but the information must 
    be assembled in such a way as to enable FDA to determine if the 
    information provides reasonable assurance of the safety and 
    effectiveness of the device for its intended use as defined in 21 CFR 
    860.7.
        FDA classified the endodontic dry heat sterilizer into class III 
    because it determined that insufficient information existed to 
    determine that general controls would provide reasonable assurance of 
    the safety and effectiveness of the device or to establish a 
    performance standard to provide such assurance. FDA has determined that 
    the special controls that may now be applied to class II devices under 
    the Safe Medical Devices Act of 1990 also would not provide such 
    assurance. FDA has weighed the probable risks and benefits to the 
    public health from the use of the device and believes that the 
    literature reports and other information discussed above present 
    evidence of significant risks associated with use of the device. These 
    risks must be addressed by the manufacturers of endodontic dry heat 
    sterilizers. FDA believes that the endodontic dry heat should undergo 
    premarket approval to establish effectiveness and to determine whether 
    the benefits to the patient are sufficient to outweigh any risk.
    
    II. PMA Requirements
    
        A PMA for this device must include the information required by 
    section 515(c)(1) of the act. Such a PMA should also include a detailed 
    discussion of the risks identified above, as well as a discussion of 
    the effectiveness of the device for which premarket approval is sought.
        A PMA should include valid scientific evidence obtained from well-
    controlled studies, with detailed data, in order to provide reasonable 
    assurance of the safety and effectiveness of the endodontic dry heat 
    sterilizer for its intended use. The data must include the following 
    information:
        a. A general description of the sterilizer including its 
    specifications, process parameters and process monitors;
        b. An overview of the sterilization process with accompanying 
    charts, graphs, or other visuals explaining all parameters;
        c. A description of any test packs used in validating the 
    performance of the endodontic dry heat sterilizer and in routine 
    monitoring of the device;
        d. Physical tests which demonstrate that the sterilizer achieves 
    and maintains the physical process lethality conditions within 
    specifications. The testing should describe how the process parameters 
    and specifications were determined;
        e. The microbiological performance tests must demonstrate that the 
    device can sterilize to an acceptable sterilization assurance level all 
    medical products identified in the labeling when used in accordance 
    with the directions for use. The tests should be consistent with those 
    used to validate sterilization processes including simulated and actual 
    use tests;
        f. Material compatibility tests must show that the medical devices 
    identified in the labeling are compatible with the 
    [[Page 30036]] sterilization process of the endodontic dry heat 
    sterilizer; and
        g. Final qualification tests from at least three consecutive runs 
    under worst case loading conditions as indicated in the labeling.
        Additional information about the validation of sterilization 
    processes can be found in: ``Guidance on Premarket Notification 
    (510(k)) Submissions for Sterilizers Intended for Use in Health Care 
    Facilities'' (available upon request from the Center for Devices and 
    Radiological Health, Division of Small Manufacturers Assistance (HFZ-
    220), Food and Drug Administration, 1350 Piccard Dr., Rockville, MD 
    20850); the American Association of Medical Instrumentation's (AAMI) 
    voluntary standards describing the validation requirements for 
    sterilization processes; and the publication entitled ``Sterile Medical 
    Devices, A GMP Workshop Manual, 4th Ed., HHS Publication (FDA) 84-4147.
        The PMA should contain a detailed discussion with supporting 
    simulated- and in-use studies, as described in the above guidance, of: 
    (1) All risks that have been identified in this proposed rule; and (2) 
    the effectiveness of the specific endodontic dry heat sterilizer that 
    is the subject of the application. In addition, the submission should 
    contain all data and information on: (1) Risks known to the applicant 
    that have not been identified in this proposed rule; (2) summaries of 
    all existing simulated- and in-use data from investigations on the 
    safety and effectiveness of the device for which premarket approval is 
    sought; and (3) the results of simulated- and in-use studies conducted 
    by or for the applicant. Applicants should submit any PMA in accordance 
    with the FDA's ``Guideline for the Arrangement and Content of a PMA 
    Application.'' The guideline is available from the Center for Devices 
    and Radiological Health, Division of Small Manufacturers Assistance 
    (address above).
    
    III. Comments
    
        Interested persons may, on or before September 5, 1995, submit to 
    the Dockets Management Branch (address above) written comments 
    regarding this proposal. Two copies of any comments are to be 
    submitted, except that individuals may submit one copy. Comments or 
    requests are to be identified with the docket number found in brackets 
    in the heading of this document. Received comments and requests may be 
    seen in the office above between 9 a.m. and 4 p.m., Monday through 
    Friday.
    
    IV. Opportunity to Reguest a Change in Classification
    
        Before requiring the filing of a PMA or a notice of completion of a 
    PDP for a device, FDA is required by section 515 (b)(2)(A)(i) through 
    (b)(2)(A)(iv) of the act and 21 CFR 860.132 to provide an opportunity 
    for interested persons to request a change in the classification of the 
    device based on new information relevant to its classification. Any 
    proceeding to reclassify the device will be under the authority of 
    section 513(e) of the act.
        A request for a change in the classification of the endodontic dry 
    heat sterilizer is to be in the form of a reclassification petition 
    containing the information required by Sec. 860.123 (21 CFR 860.123), 
    including new information relevant to the classification of the device, 
    and shall, under section 515(b)(2)(B) of the act, be submitted by June 
    22, 1995.
        The agency advises that, to ensure timely filing of any such 
    petition, any request should be submitted to the Dockets Management 
    Branch (address above) and not to the address provided in 
    Sec. 860.123(b)(1). If a timely request for a change in the 
    classification of the endodontic dry heat sterilizer is submitted, the 
    agency will, by August 7, 1995, after consultation with the appropriate 
    FDA advisory committee and by an order published in the Federal 
    Register, either deny the request or give notice of its intent to 
    initiate a change in the classification of the device in accordance 
    with section 513(e) of the act and 21 CFR 860.130 of the regulations.
    
    V. References
    
        The following references have been placed on display in the Dockets 
    Management Branch (address above) and may be seen by interested persons 
    between 9 a.m. and 4 p.m., Monday through Friday.
    
    1. American Dental Association, ``Accepted Dental Therapeutics,'' 
    40th ed., pp. 138-139, Chicago, IL, 1984.
    2. Corner, G. A., ``An Assessment of the Performance of a Glass Bead 
    Sterilizer,'' Journal of Hospital Infection, 10:308-311, 1987.
    3. Dayoub, M. B., and M. J. Devine, ``Endodontic Dry-Heat Sterilizer 
    Effectiveness,'' Journal of Endodontics, 2:343-344, 1976.
    4. Engelhardt, M. P., L. Grun, and H. Dahl, ``Factors Affecting 
    Sterilization in Glass Bead Sterilizers,'' Journal of Endodontics, 
    10:454-470, 1984.
    5. Fahid, A., and J. F. Tainter, ``The Influence of File Size, 
    Cleaning, and Time on the Effectiveness of Bead Sterilizers,'' Oral 
    Surgery, 58:443-445, 1984.
    6. Koehler, H. M., and J. J. Hefferren, ``Time-Temperature Relations 
    of Dental Instruments Heated in Root-Canal Instrument Sterilizers,'' 
    Journal of Dental Research, 41:182-195, 1962.
    7. Ingle, J. I., Endodontics, 3d Ed., Philadelphia, Lea & Febiger, 
    pp. 615-616.
    8. Jakush, J., ``Infection Control Procedures and Products: Cautions 
    and Common Sense,'' Journal of The American Dental Association, 
    117:293-301, 1988.
    9. Schutt, R. W., and W. J. Starsiak, ``Glass Bead Sterilization of 
    Surgical Dental Burs,'' International Journal of Oral and 
    Maxillofacial Surgery, 19:250-251, 1990.
    10. Smith, G. E., ``Glass Bead Sterilization of Orthodontic Bands,'' 
    American Journal of Orthodontics Dentofacial Orthopedics, 90:243-
    249, 1986.
    11. Windeler, A. S., and R. G. Walter, ``The Sporicidal Activity of 
    Glass Beads Sterilizers,'' Journal of Endodontics, 1:273-275, 1975.
    
    VI. Environmental Impact
    
        The agency has determined under 21 CFR 25.24(a)(8) that this action 
    is of a type that does not individually or cumulatively have a 
    significant effect on the human environment. Therefore, neither an 
    environmental assessment nor an environmental impact statement is 
    required.
    
    VII. Analysis of Impacts
    
        FDA has examined the impacts of the proposed rule under Executive 
    Order 12866 and the Regulatory Flexibility Act (Pub. L. 96-354). 
    Executive Order 12866 directs agencies to assess all costs and benefits 
    of available regulatory alternatives and, when regulation is necessary, 
    to select regulatory approaches that maximize net benefits (including 
    potential economic, environmental, public health and safety, and other 
    advantages; distributive impacts; and equity). The agency believes that 
    this proposed rule is consistent with the regulatory philosophy and 
    principles identified in the Executive Order. In addition, the proposed 
    rule is not a significant regulatory action as defined by the Executive 
    Order and so is not subject to review under the Executive Order.
        The Regulatory Flexibility Act requires agencies to analyze 
    regulatory options that would minimize any significant impact of a rule 
    on small entities. Because PMA's for this device could have been 
    required by FDA as early as February 12, 1990, and because firms that 
    distributed this device prior to May 28, 1976, or whose device has been 
    found by FDA to be substantially equivalent will be permitted to 
    continue marketing the endodontic dry heat sterilizer during FDA's 
    review of the PMA or notice of completion of the [[Page 30037]] PDP, 
    the agency certifies that the proposed rule will not have a significant 
    economic impact on a substantial number of small entities.
        Therefore, under the Regulatory Flexibility Act, no further 
    analysis is required.
    
    List of Subjects in 21 CFR Part 872
    
        Medical devices.
    
        Therefore, under the Federal Food, Drug, and Cosmetic Act and under 
    the authority delegated to the Commissioner of Food and Drugs, it is 
    proposed that 21 CFR part 872 be amended as follows:
    
    PART 872--DENTAL DEVICES
    
        1. The authority citation for 21 CFR part 872 is revised to read as 
    follows:
    
        Authority: Secs. 501, 510, 513, 515, 520, 522, 701 of the 
    Federal Food, Drug, and Cosmetic Act (21 U.S.C. 351, 360, 360c, 
    360e, 360j, 360l, 371).
    
        2. Section 872.6730 is amended by revising paragraph (c) to read as 
    follows:
    
    Sec. 872.6730
    
      Endodontic dry heat sterilizer.
    
    * * * * *
    
        (c) Date premarket approval application (PMA) or notice of 
    completion of product development protocol (PDP) is required. A PMA or 
    a notice of completion of a PDP is required to be filed with the Food 
    and Drug Administration on or before (90 days after the effective date 
    of a final rule based on this proposed rule), for any endodontic dry 
    heat sterilizer that was in commercial distribution before May 28, 
    1976, or that has on or before (90 days after the effective date of a 
    final rule based on this proposed rule), been found to be substantially 
    equivalent to the endodontic dry heat sterilizer that was in commercial 
    distribution before May 28, 1976. Any other endodontic dry heat 
    sterilizer shall have an approved PMA or declared completed PDP in 
    effect before being placed in commercial distribution.
    
        Dated: May 24, 1995.
    
    D. B. Burlington,
    
    Director, Center for Devices and Radiological Health.
    
    [FR Doc. 95-13831 Filed 6-6-95; 8:45 am]
    BILLING CODE 4160-01-P
    
    

Document Information

Published:
06/07/1995
Department:
Food and Drug Administration
Entry Type:
Proposed Rule
Action:
Proposed rule; opportunity to request a change in classification.
Document Number:
95-13831
Dates:
Written comments by September 5, 1995; requests for a change in classification by June 22, 1995. FDA intends that, if a final rule based on this proposed rule is issued, PMA's will be required to be submitted within 90 days of the effective date of the final rule.
Pages:
30032-30037 (6 pages)
Docket Numbers:
Docket No. 95N-0033
PDF File:
95-13831.pdf
CFR: (2)
21 CFR 860.123(b)(1)
21 CFR 872.6730