2013-18012. Specifications for Packagings

  • Start Preamble

    CFR Correction

    Start Amendment Part

    In Title 49 of the Code of Federal Regulations, Parts 178 to 199, revised as of October 1, 2012, in § 178.68, on page 80, paragraph (i)(2) is correctly reinstated and on page 81, paragraph (l)(2) is revised to read as follows:

    End Amendment Part
    Specification 4E welded aluminum cylinders.
    * * * * *

    (i) * * *

    (2) If the weld is at midlength of the cylinder, the test may be made as specified in paragraph (i)(1) of this section or must be made between wedge shaped knife edges (60° angle) rounded to a 1/2 inch radius. There must be no evidence of cracking in the sample when it is flattened to no more than 6 times the wall thickness.

    * * * * *

    (l) * * *

    (2) Guided bend test. A bend test specimen must be cut from the cylinder used for the physical test specified in paragraph (j) of this section. Specimen must be taken across the seam, must be a minimum of 11/2 inches wide, edges must be parallel and rounded with a file, and back-up strip, if used, must be removed by machining. The specimen shall be tested as follows:

    (i) The specimen must be bent to refusal in the guided bend test jig as illustrated in paragraph 6.10 of CGA C-3 (IBR, see § 171.7 of this subchapter). The root of the weld (inside surface of the cylinder) must be located away from the ram of the jig. The specimen must not show a crack or other open defect exceeding 1/8 inch in any direction upon completion of the test. Should this specimen fail to meet the requirements, specimens may be taken from each of 2 additional cylinders from the same lot and tested. If either of the latter specimens fails to meet requirements, the entire lot represented must be rejected.

    (ii) Alternatively, the specimen may be tested in a guided bend test jig as illustrated in Figure 12.1 of The Aluminum Association's 2002 publication, “Welding Aluminum: Theory and Practice.” The root of the weld (inside surface of the cylinder) must be located away from the mandrel of the jig. No specimen must show a crack or other open defect exceeding 1/8 inch in any direction upon completion of the test. Should this specimen fail to meet the requirements, specimens may be taken from each of 2 additional cylinders from the same lot and tested. If either of the latter specimens fails to meet requirements, the entire lot represented must be rejected.

    * * * * *
    End Preamble

    [FR Doc. 2013-18012 Filed 7-24-13; 8:45 am]

    BILLING CODE 1505-01-D