98-20281. National Primary Drinking Water Regulations: Analytical Methods for Certain Pesticides and Microbial Contaminants  

  • [Federal Register Volume 63, Number 147 (Friday, July 31, 1998)]
    [Proposed Rules]
    [Pages 41134-41143]
    From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
    [FR Doc No: 98-20281]
    
    
    
    [[Page 41133]]
    
    _______________________________________________________________________
    
    Part III
    
    
    
    
    
    Environmental Protection Agency
    
    
    
    
    
    _______________________________________________________________________
    
    
    
    40 CFR Part 141
    
    
    
    National Primary Drinking Water Regulations: Analytical Methods for 
    Certain Pesticides and Microbial Contaminants; Proposed Rule
    
    Federal Register / Vol. 63, No. 147 / Friday, July 31, 1998 / 
    Proposed Rules
    
    [[Page 41134]]
    
    
    -----------------------------------------------------------------------
    
    ENVIRONMENTAL PROTECTION AGENCY
    
    40 CFR Part 141
    
    [WH-FRL-6132-3]
    RIN 2040-AD04
    
    
    National Primary Drinking Water Regulations: Analytical Methods 
    for Certain Pesticides and Microbial Contaminants
    
    AGENCY: Environmental Protection Agency (EPA).
    
    ACTION: Proposed rule.
    
    -----------------------------------------------------------------------
    
    SUMMARY: EPA is proposing to approve the use of a new membrane filter 
    medium for the simultaneous detection of total coliform and Escherichia 
    coli (E. coli) and three revised chemical analytical methods for 
    compliance determinations of acid herbicides and diquat in drinking 
    water. The Agency proposes to withdraw approval of the previous version 
    of the EPA Method for diquat. The Agency is also proposing to amend and 
    clarify laboratory certification requirements, and to amend the 
    regulations to make the current recommendation to ship microbiological 
    samples below 10  deg.C a requirement. The purpose of the proposal is 
    to approve new methods, withdraw an outdated EPA method, clarify 
    laboratory certification requirements and improve the reliability of 
    sample collection for total or fecal coliform. The rule is expected to 
    satisfy public requests for approval of new technologies in drinking 
    water analyses.
    
    DATES: Comments on this proposed rule must be received in writing by 
    September 29, 1998.
    
    ADDRESSES: Written or electronic comments on this proposed rule may be 
    submitted. Written comments may be sent to the W-97-05 Drinking Water 
    Analytical Methods Proposed Comment Clerk, U.S. Environmental 
    Protection Agency, Water Docket, MC 4101, 401 M Street, SW, Washington, 
    DC 20460. Please submit any references cited in your comments. EPA 
    would appreciate an original and 3 copies of your comments and 
    enclosures (including references).
        No facsimiles (faxes) will be accepted because EPA cannot ensure 
    that they will be submitted to the Water Docket. To ensure that EPA can 
    read, understand and therefore properly respond to comments, the Agency 
    would prefer that commenters cite, where possible, the paragraph(s) in 
    the proposed regulation (e.g., 141.24(e)) to which each comment refers. 
    Commenters should use a separate paragraph for each method or issue 
    discussed. Commenters who would like acknowledgment of receipt of their 
    comments should include a self-addressed, stamped envelope. For 
    information on submitting comments electronically see Supplementary 
    Information.
        The record for this rulemaking has been established under docket 
    number W-97-05. Copies of the supporting documents (including 
    references and methods cited in this notice) are available for review 
    at the U.S. Environmental Protection Agency, Water Docket, East Tower 
    Basement, 401 M Street, SW, Washington, DC 20460. For access to the 
    docket materials, call 202-260-3027 on Monday through Friday, excluding 
    Federal holidays, between 9:00 a.m. and 3:30 p.m. Eastern Time for an 
    appointment.
    
    FOR FURTHER INFORMATION CONTACT: Ms. Jeanne Campbell, Standards and 
    Risk Management Division, Office of Ground Water and Drinking Water 
    (MC-4607), U.S. Environmental Protection Agency, 401 M Street, SW, 
    Washington, DC 20460, telephone 202-260-7770. General information may 
    also be obtained from the EPA Safe Drinking Water Hotline. Callers 
    within the United States may reach the Hotline at 800-426-4791. The 
    Hotline is open Monday through Friday, excluding Federal holidays, from 
    9 a.m. to 5:30 p.m. Eastern Time.
        For technical information regarding microbiology methods, contact 
    Paul S. Berger, Ph.D., Office of Ground Water and Drinking Water (MC-
    4607), U.S. Environmental Protection Agency, Washington, DC 20460, 
    telephone 202-260-3039. For technical information regarding chemistry 
    methods, contact Richard Reding, Ph.D., Office of Ground Water and 
    Drinking Water, U.S. Environmental Protection Agency, Cincinnati, Ohio 
    45268, telephone 513-569-7961. For a list of Regional Contacts see 
    Supplementary Information.
    
    SUPPLEMENTARY INFORMATION:
    
    Potentially Regulated Entities
    
        EPA Regions, as well as States, Territories, and Tribes with 
    primacy to administer the regulatory program for public water systems 
    under the Safe Drinking Water Act, sometimes conduct analyses to 
    measure for contaminants in water samples, but often require the public 
    water systems themselves to conduct such analysis. If EPA has 
    established a maximum contaminant level (``MCL'') for a given drinking 
    water contaminant, the Agency also ``approves'' standardized testing 
    procedures (i.e., promulgated through rulemaking) for analysis of the 
    contaminant. Once EPA standardizes such test procedures, analysis using 
    those procedures (or approved alternate test procedures) is required. 
    Therefore, States, Territories, Tribes, and public water systems 
    required to test water samples are potentially regulated by the 
    standardization of testing procedures in this rulemaking. Categories 
    and entities that may ultimately be regulated include:
    
    ----------------------------------------------------------------------------------------------------------------
                                                                                  Examples of potentially regulated 
                                      Category                                                 entities             
    ----------------------------------------------------------------------------------------------------------------
    State and Territorial Governments and Indian Tribes........................  States, Territories, and Tribes    
                                                                                  that analyze water samples on     
                                                                                  behalf of public water systems    
                                                                                  required to conduct such analysis;
                                                                                  States, Territories, and Tribes   
                                                                                  that themselves operate public    
                                                                                  water systems required to conduct 
                                                                                  analytic monitoring.              
    Industry...................................................................  Industrial operators of public     
                                                                                  water systems.                    
    Municipalities.............................................................  Municipal operators of public water
                                                                                  systems.                          
    ----------------------------------------------------------------------------------------------------------------
    
        This table is not intended to be exhaustive, but rather provides a 
    guide for readers regarding entities likely to be regulated by this 
    action. This table lists the types of entities that EPA is now aware 
    could potentially be regulated by this action. Other types of entities 
    not listed in the table could also be regulated. To determine whether 
    your organization is or would be regulated by this action, you should 
    carefully examine the applicability language at 40 CFR 141.2 
    (definition of public water system). If you have questions regarding 
    the applicability of this action to a particular entity, consult the 
    person listed in the preceding FOR FURTHER INFORMATION CONTACT section.
    
    [[Page 41135]]
    
    Regional Contacts
    
    EPA Regional Offices
    
    I
    
    JFK Federal Bldg., One Congress Street, 11th Floor, Boston, MA 
    02203. Phone: 617-565-3602, Linda Murphy
    
    II
    
    290 Broadway, 24th Floor, New York, NY 10007. Phone: 212-637-3880, 
    Walter Andrews
    
    III
    
    841 Chestnut Building, Philadelphia, PA 19107. Phone: 215-597-6511, 
    Victoria Binetti
    
    IV
    
    345 Courtland Street, NE, Atlanta, GA 30365. Phone: 404-347-2207, 
    Stallings Howell
    
    V
    
    77 West Jackson Boulevard, Chicago, IL 60604. Phone: 312-886-6206, 
    Charlene Denys
    
    VI
    
    1445 Ross Avenue, Suite 1200, Dallas, TX 75202. Phone: 214-655-7150, 
    Larry Wright
    
    VII
    
    726 Minnesota Avenue, Kansas City, KS 66101. Phone: 913-551-7682, 
    Robert Morby
    
    VIII
    
    One Denver Place, 999 18th Street, Suite 500, Denver, CO 80202. 
    Phone: 303-293-1652, Jack Rychecky
    
    IX
    
    75 Hawthorne Street, San Francisco, CA 94105. Phone: 415-744-1817, 
    William Thurston
    
    X
    
    1200 Sixth Avenue, Seattle, WA 98101. Phone: 206-553-1893, Larry 
    Worley
    
    Electronic Comment Submission
    
        Electronic comments should be addressed to the Internet address: 
    ow-docket@epamail.epa.gov. Electronic comments must be submitted as an 
    ASCII file and avoid use of special characters and any form of 
    encryption. Electronic comments must be identified by the docket number 
    W-97-05. Comments and data will also be accepted on disks in 
    WordPerfect in 5.1 or 6.1 format or ASCII file format. Electronic 
    comments on this notice may be filed online at many Federal Depository 
    Libraries. Electronic comments will be transferred into a paper version 
    for the official record. EPA will attempt to clarify electronic 
    comments if there is an apparent error in transmission. Comments 
    provided electronically will be considered timely if they are submitted 
    electronically by 11:59 p.m. (Eastern time) on September 29, 1998.
    
    Information on Internet Access
    
        This Federal Register document has been placed on the Internet for 
    public review and downloading at the following location: http://
    www.epa.gov/fedrgstr.
    
    Availability and Sources for Methods
    
        Copies of final methods published by EPA are available for a 
    nominal cost through the National Technical Information Service (NTIS), 
    U.S. Department of Commerce, 5285 Port Royal Road, Springfield, VA 
    22161. NTIS also may be reached at 800-553-6847. Copies of draft EPA 
    methods may be obtained from USEPA, National Exposure Research 
    Laboratory (NERL)-Cincinnati, 26 West Martin Luther King Drive, 
    Cincinnati, OH 45268. FAX request for copies of draft EPA methods may 
    be sent to NERL-Cincinnati at 513-569-7757. All other methods must be 
    obtained from the publisher. Publishers (with addresses) for all 
    approved methods are cited at 40 CFR part 141 and in the References 
    section of today's rule.
    
    Table of Contents
    
    I. Statutory Authority
    II. Regulatory Background
    III. Explanation of Today's Action
        A. New Membrane Filter Medium for the Simultaneous Detection of 
    Total Coliform and E. coli
        B. Pesticide Methods
        C. New Information on Previously Approved Colisure Test
        D. Amendments and Clarifications
        E. Performance-based Measurement Systems
    IV. Regulation Assessment Requirements
    V. References
    
    I. Statutory Authority
    
        The Safe Drinking Water Act (SDWA), as amended in 1996, requires 
    EPA to promulgate national primary drinking water regulations (NPDWRs) 
    which specify maximum contaminant levels (MCLs) or treatment techniques 
    for drinking water contaminants SDWA section 1412, 42 U.S.C. 300g-1. 
    NPDWRs apply to public water systems pursuant to SDWA section 1401, 42 
    U.S.C. 300f(1)(A). According to SDWA section 1401(1)(D) of the Act, 
    NPDWRs include ``criteria and procedures to assure a supply of drinking 
    water which dependably complies with such maximum contaminant levels; 
    including quality control and testing procedure. * * *'' see 42 U.S.C. 
    300f(1)(D). In addition, SDWA section 1445(a) of the Act authorizes the 
    Administrator to establish regulations for monitoring to assist in 
    determining whether persons are acting in compliance with the 
    requirements of the SDWA see 42 U.S.C. 300j-4. EPA's promulgation of 
    analytical methods is authorized under these sections of the SDWA as 
    well as the general rulemaking authority in SDWA section 1450(a), 42 
    U.S.C. 300j-9(a).
    
    II. Regulatory Background
    
        EPA has promulgated analytical methods for all currently regulated 
    drinking water contaminants for which MCLs or monitoring requirements 
    have been promulgated. In most cases, the Agency has approved use of 
    more than one analytical method for measurement of a contaminant, and 
    laboratories may use any approved method for determining compliance 
    with an MCL or monitoring requirement. After any regulation is 
    published, EPA may amend the regulations to approve additional methods, 
    or modifications to approved methods, or withdraw methods that become 
    obsolete or amend other requirements (such as certification 
    requirements) for the use of approved methods. EPA takes these actions 
    as quickly as possible after new or revised methods are published.
    
    III. Explanation of Today's Action
    
        This proposal invites public comment on a new medium for 
    microbiological testing, on three revised chemical analytical methods 
    for acid herbicides and diquat, on withdrawal of an obsolete EPA method 
    for diquat and on methods-related changes to the regulations as 
    follows. EPA is proposing to approve the use of a new membrane filter 
    medium, 4-methylumbelliferyl-beta-D-galactopyranoside indoxyl-beta-D 
    gluconoride (MI) agar, for the determination of total coliforms and E. 
    coli, and to amend the regulations to codify the current recommendation 
    to ship microbiological samples below 10  deg.C. EPA is proposing a 
    revised method (EPA Method 515.3) for acid herbicides, but would not 
    withdraw approval of EPA Methods 515.1 or 515.2 for determination of 
    acid herbicides because these methods are not obsolete. EPA is also 
    proposing American Society for Testing and Materials (ASTM) method 
    D5317-93 for acid herbicides provided quality control limits specified 
    in EPA Method 515.1 are met. EPA is proposing to withdraw approval of 
    EPA Method 549.1 for measurement of diquat and to replace it with an 
    improved version, EPA Method 549.2. Withdrawal of EPA Method 549.1 
    would be effective 18 months after promulgation of a final rule that 
    would approve EPA Method 549.2 for diquat. EPA is proposing to amend 
    the certification regulations to clarify that analysis for certain 
    additional
    
    [[Page 41136]]
    
    contaminants does not require certification. Finally, the proposal 
    would also require a laboratory to successfully analyze a PE sample for 
    each method for which the laboratory seeks certification rather than 
    one PE sample per analyte.
    
    A. New Membrane Filter Medium for the Simultaneous Detection of Total 
    Coliform and E. coli
    
        EPA is proposing the approval of a new membrane filter medium known 
    as MI Agar for the simultaneous detection of total coliforms and E. 
    coli in drinking water samples and the enumeration of total coliforms 
    in raw water samples. In a procedure developed by EPA, a water sample 
    is passed through a membrane filter and the filter is placed on an MI 
    Agar plate (Brenner 1993). The medium is incubated at 35  deg.C for 16-
    24 hours and examined either for fluorescence of total coliforms under 
    longwave (366 nm) ultraviolet light or for the blue color under ambient 
    light that is characteristic of E. coli.
        The MI Agar procedure is based on the ability of total coliforms to 
    produce the enzyme beta-galactosidase which hydrolyzes the 4-
    methylumbelliferyl-beta-D-galactopyranoside in the medium to form a 
    product which fluoresces. The procedure is also based on the ability of 
    E. coli to produce the enzyme beta-glucuronidase which hydrolyzes 
    indoxyl-beta-D-glucuronide to form a blue color.
        In a comparison study between the membrane filter test using MI 
    Agar and already approved reference media (mEndo agar, Nutrient 
    Agar+MUG), the MI Agar performed equal to or better than the reference 
    media and non-coliform background counts were significantly lower for 
    MI Agar (Brenner 1993, 1996a,b). For total coliforms the false-positive 
    rate was 4.9% and the false-negative rate was less than 8.8%. The 
    false-positive and false-negative rates for E. coli were both 4.3% 
    (Brenner 1993). Based upon these results, EPA believes that the 
    membrane filter method using MI Agar is satisfactory as a compliance 
    method for total coliforms and E. coli.
    
    B. Pesticide Methods
    
    Draft EPA Method 515.3, Acid Herbicides
        EPA is proposing a draft of EPA Method 515.3 (EPA 1996) as a 
    modification of, but not replacement for, EPA Method 515.1 (EPA 1995). 
    Both methods require hydrolysis and derivatization of the herbicides, 
    and rely on use of liquid-liquid extraction (LLE) and gas 
    chromatographic analysis with an electron capture detector. EPA Method 
    515.3 uses a different reagent to derivatize the acid herbicides, 
    smaller volumes of drinking water sample and less extraction solvent. 
    EPA Method 515.2 (EPA 1995) for acid herbicides, which uses solid phase 
    extraction (SPE) will also continue to be approved because many 
    laboratories prefer SPE as a solvent-saving alternative to the LLE step 
    in EPA Methods 515.1 and 515.3.
        EPA has been asked to offer alternatives to diazomethane as the 
    derivatizing reagent because it can form explosive products if it is 
    not handled carefully and in small amounts. EPA's current alternative 
    reagent, trimethylsilyldiazomethane (TMSD), produces acceptable results 
    and is being retained as an option in EPA Method 515.1. Use of TMSD 
    however, can produce unwanted interferences in the chromatography. EPA 
    is developing Method 515.3 to find an effective alternative reagent 
    that produces fewer impurities.
        EPA is proposing a draft of EPA Method 515.3 for comment while 
    development work is continuing. EPA realizes the draft method contains 
    method detection limits (MDLs) for some analytes that exceed detection 
    limits specified at 40 CFR 141.24(h)(18). These detection limits 
    specify a minimum sensitivity required to demonstrate that a 
    contaminant has not been detected in a compliance sample. Systems with 
    nondetects are often eligible for a reduced frequency for repeat 
    monitoring. Data in tables 2 and 3 of draft EPA Method 515.3 list MDLs 
    for two types of derivatization. Table 2 is for derivatization with 
    diazomethane, and MDLs for four (2,4-D, dinoseb, pentachlorophenol and 
    picloram) of the six regulated acid herbicides exceed detection limits 
    specified in the regulations. Table 3 is for derivatization with 
    tetrabutylammonium hydroxide and methyl iodide, and MDLs for three 
    (2,4-D, dinoseb and picloram) acid herbicides exceed the regulatory 
    limits. However, because EPA Method 515.3 has sufficient sensitivity 
    and reproducibility to determine that any regulated acid herbicide is 
    reliably and consistently below the MCL, this method would be used to 
    screen for acid herbicides in follow-up samples that are required when 
    a regulated herbicide has been detected in prior monitoring by a more 
    sensitive measurement. Also, in routine use, some laboratories may be 
    able to improve the sensitivity of EPA Method 515.3 and achieve the 
    required detection limits for 2,4-D, dinoseb, pentachlorophenol and 
    picloram.
        EPA proposes to approve draft EPA Method 515.3 for compliance 
    determinations of all six regulated acid herbicides (2,4-D, 2,4,5-TP, 
    dalapon, dinoseb, pentachlorophenol and picloram) and the unregulated 
    acid herbicide, dicamba. Final approval for compliance monitoring would 
    be contingent upon public comment and EPA's evaluation of the final 
    development work, which would include second laboratory validation of 
    the method.
    ASTM Method D5317-93, Acid Herbicides
        ASTM Method D5317-93 (ASTM 1996) uses procedures (sodium hydroxide 
    hydrolysis, diazomethane derivatization, liquid-liquid extraction and 
    gas chromatography with electron capture detection) that are equivalent 
    to those used in EPA Method 515.1. EPA is proposing to approve this 
    1993 version of ASTM D5317 for determination of five acid herbicides in 
    drinking water samples. The approval would be contingent upon 
    superseding the variable control limit for the laboratory fortified 
    blank (LFB), specified in section 6g of D5317-93, with the fixed upper 
    control limit of 30% specified in sections 9.3.2 and 9.7.2 
    of EPA Method 515.1 (Rev. 4.1). The variable control limit criterion 
    for the LFB specified in D5317-93 is equivalent to the variable limit 
    specified in section 10.7.2 of the 1989 version, Rev. 4.0, of EPA 
    Method 515.1. EPA published Rev. 4.1 of EPA Method 515.1 to specify a 
    fixed upper limit of  30% for the LFB in 1995, which was 
    after the last update of ASTM D5317-93.
        The detection limits listed for pentachlorophenol (PCP) and 2,4-D 
    in Table 1 of the ASTM method exceed the limits specified at 40 CFR 
    141.24(h)(18). However, the method has sufficient sensitivity and 
    reproducibility to determine that either herbicide is reliably and 
    consistently below the MCL. Thus, D5317-93 could be useful to screen 
    for PCP or 2,4-D in follow-up compliance samples that might be required 
    when detection of the target acid herbicide has been established by 
    prior monitoring. Also, in routine use, some laboratories may be able 
    to improve the sensitivity of D5317-93 so that it would be suitable for 
    lower level compliance monitoring of PCP or 2,4-D. Although D5317-93 
    should be applicable to compliance determinations of dalapon and 
    dinoseb, the method does not contain performance data for these two 
    analytes. Therefore, EPA proposes to approve D5317-93 for compliance 
    determinations of four regulated acid herbicides: PCP, picloram, 2,4-D 
    and 2,4,5-TP, and the unregulated acid
    
    [[Page 41137]]
    
    herbicide, dicamba, but not for determinations of dalapon or dinoseb. 
    EPA also proposes to add a footnote to the tables of approved methods 
    to require that the  30% control limits for the LFB 
    specified in EPA Method 515.1 (Rev. 4.1) supersede the variable limits 
    specified in D5317-93.
    Draft Method 549.2, Diquat
        EPA is proposing a draft version of EPA Method 549.2 (EPA 1997a) as 
    a revision of EPA Method 549.1. Both versions use ion exchange with a 
    specially conditioned C8 cartridge or disk to remove diquat 
    from the water sample. In the revised version, EPA Method 549.2, the 
    sample pH is not adjusted to 10.5 before extraction because EPA has 
    determined that this adjustment does not improve the extraction of 
    diquat. Also, at pH 10.5, a precipitate is formed in several hard water 
    samples. The precipitate causes very long delays in passing the sample 
    through the solid phase sorbent, which result in unacceptably low 
    recoveries of diquat. Changes resulting in EPA Method 549.2 are 
    supported by data contained in the method that demonstrate acceptable 
    extraction efficiency of method analytes at pH 7. No precipitation 
    occurred in the pH range of 7 to 9 in artificial matrices simulating 
    extremely hard water.
        Data included in the EPA Method 549.2 shows that recovery of diquat 
    is more variable and lower than with normal phase liquid-solid 
    extraction (LSE) of organic compounds (cf. Method 525.2). Matrix 
    effects are expected to be more of a problem with ion exchange because 
    it is affected by the ionic strength and ionic composition of the 
    drinking water sample, both of which can vary greatly with time and 
    with the type of source.
        In addition to matrix effects, EPA has observed significant 
    variability in diquat recovery between brands of media. The variability 
    appears to be greater with cartridge than with disk media. EPA seeks 
    comment on what additional QC steps could be added to the method to 
    detect and correct this problem. For example, EPA could increase the 
    frequency of required matrix spikes from the current rate of 10%. EPA 
    also seeks advice on ways to detect and correct for performance 
    differences between brands (or within lots) of extraction media.
        EPA believes EPA Method 549.2 is superior to EPA Method 549.1 
    because it solves the precipitation problem. The diquat recovery 
    problem described above occurs with either version of the method. 
    Therefore, the recovery rate is equivalent for both methods. Final 
    approval of EPA Method 549.2 and withdrawal of EPA Method 549.1 would 
    be contingent upon public comment and EPA's evaluation of the final 
    development work. This work would include second laboratory validation 
    with a variety of matrices spiked with diquat.
    Other Methods for Acid Herbicides and Diquat
        Under section 12(d) of the National Technology Transfer and 
    Advancement Act, the Agency is directed to use voluntary consensus 
    standards in its regulatory activities unless to do so would be 
    inconsistent with applicable law or otherwise impractical. Voluntary 
    consensus standards are technical standards which are developed or 
    adopted by voluntary consensus standard bodies. EPA searched for 
    acceptable voluntary consensus standards to propose with the methods 
    EPA is proposing in this rule for diquat (EPA Method 549.2) and for 
    acid herbicides (EPA Method 515.3). There are no methods for diquat 
    that are published by ASTM or the Standard Methods for Examination of 
    Water and Wastewater (Standard Methods or SM). These organizations 
    have, however, published a method for acid herbicides (ASTM D 5317-93 
    and SM 6640B, respectively) that EPA evaluated for today's proposal.
        As discussed above, EPA is proposing to approve ASTM D 5317-93 for 
    five acid herbicides: 2,4-D, 2,4,5-TP, dalapon, dicamba, 
    pentachlorophenol and picloram. EPA also evaluated SM 6640B (APHA 
    1995), which Standard Methods developed to be similar to EPA Method 
    515.1 (Rev. 4.0). SM 6640B measures all of the unregulated and 
    regulated acid herbicides with MDLs that meet the detection limits 
    specified at 40 CFR 141.24(h)(18). EPA does not propose SM 6640B in 
    today's rule because SM 6640B uses sample preparation and quality 
    control procedures that are not similar enough to be judged equivalent 
    to procedures specified in EPA Method 515.1, Rev. 4.1. In the next 
    paragraphs, EPA provides reasons why, if used as written, use of SM 
    6640B could lead to under reporting of acid herbicide contamination in 
    compliance samples.
        The sample preparation instructions in sections 2, 4 b and c of SM 
    6640B refer to the hydrolysis step as ``an optional procedure for 
    simplified alkaline solvent wash of difficult matrices''. Hydrolysis is 
    not optional nor is its primary function to clean-up ``difficult 
    matrices'' in the EPA method. As explained in section 2.1 of EPA Method 
    515.1, the regulated herbicides are ``applied as a variety of salts and 
    esters''. Thus ``it is vital to hydrolyze them to the parent acid prior 
    to extraction'' and derivatization. EPA has previously reinforced the 
    necessity of hydrolysis with a footnote in the tables of approved 
    methods at 40 CFR 141.24(e) and 141.40(n)(11) to state that hydrolysis 
    is mandatory, not optional. EPA believes, that as written, analysts 
    using SM 6640B might believe the ``alkaline solvent wash,'' which is 
    not a wash but a critical chemical conversion step, is not needed in 
    relatively clean matrices, such as drinking water.
        Two significant problems with the procedures specified in SM 6640B 
    for use of an internal standard are the possibility of interferences 
    and losses due to addition of the standard at the wrong time. To 
    calibrate the instrument, EPA Method 515.1 requires use of an internal 
    standard. EPA selected 4,4'-dibromooctofluorobiphenyl (DBOB) because 
    DBOB is not likely to be a sample contaminant. Instructions in 3c of 
    the quality control section of SM 6640 B specify use of 1,2,3-
    trichloropropane (TCP), which may be a sample contaminant because TCP 
    is both a solvent used as an industrial degreaser and a potential 
    chemical by-product of drinking water disinfection. Addition of TCP 
    rather than DBOB to sample extracts may lead to erroneous calibration 
    because of interference from TCP in waters contaminated with TCP runoff 
    or discharge, or with TCP synthesized during disinfection of the 
    drinking water supply. Sections 4b and 6f of SM 6640B allow the analyst 
    the discretion not to use an internal standard (6f) or to add it prior 
    to, rather than after, derivatization of the sample (4b). EPA Method 
    515.1 does not allow use of an internal standard to be optional and it 
    must be added after derivatization of the sample extract so as not to 
    incur losses from incomplete derivatization of the internal standard.
        To propose the current version of SM 6640B for compliance 
    monitoring of acid herbicides would require EPA to annotate the listing 
    in the CFR with several footnotes to supersede sample preparation and 
    quality control procedures that are specified in SM 6640B. Because 
    footnotes in the CFR have the potential to become separated from the 
    written method used by the analyst, EPA keeps these footnotes to a 
    minimum. Rather than propose SM 6640B in today's rule, EPA prefers to 
    work directly with the Standard Methods committee to revise SM 6640 B 
    for publication in the 21st edition of Standard Methods or in the 
    supplement to the 20th edition. EPA solicits compelling reasons why it 
    would be
    
    [[Page 41138]]
    
    beneficial to approve the current version of SM 6640B for compliance 
    monitoring.
    
    C. New Information on Previously Approved Colisure Test
    
        The Millipore Corporation has submitted data to EPA indicating that 
    the Colisure Test, which EPA previously approved for the determination 
    of total coliform and E. coli under the Total Coliform Rule, can be 
    read after a 24-hour incubation period. Currently, the regulation at 40 
    CFR 141.21(f)(3) (footnote 7 to the Table) specifies a minimum 
    incubation time of 28 hours. EPA is now in the process of evaluating 
    the data. The Agency has placed a copy of Millipore's submission in the 
    Water Docket and is requesting public comment on the minimum incubation 
    time for the Colisure Test.
    
    D. Amendments and Clarifications
    
        The Agency is proposing to amend and clarify laboratory 
    certification requirements and to amend the regulations to codify as a 
    requirement the current recommendation to ship microbiological samples 
    below 10  deg.C.
    Require Chilling of Microbiological Samples During Storage and Transit
        EPA is proposing to require that all source water samples collected 
    for the analysis of total coliforms or fecal coliforms under the 
    Surface Water Treatment Rule (SWTR), or any distribution system sample 
    collected for the analysis of heterotrophic bacteria, be held at a 
    temperature below 10  deg.C during storage and transit. Currently, in a 
    footnote to the table at 40 CFR 141.74(a)(1), EPA encourages, but does 
    not require, systems to meet this provision. The Agency would make the 
    holding temperature limit mandatory because it believes that microbial 
    densities in a water sample may change dramatically at higher 
    temperatures over a short period of time (hours) and would thus no 
    longer mirror the microbial densities in the water from which they 
    originated. The proposed change in holding temperature would render the 
    EPA methods consistent with SM 9060B (APHA 1995) which requires that 
    water samples for microbiological examination be held at temperatures 
    below 10  deg.C. The proposal would not change the holding time and 
    temperature provisions for total coliform, fecal coliform, or E. coli 
    samples collected in the distribution system to meet the requirements 
    of the Total Coliform Rule (TCR) at 40 CFR 141.21(f). Analysis under 
    the TCR require determination of presence and absence of microorganisms 
    and not their densities, and are therefore less sensitive to storage 
    and transit temperature changes.
    Updating CFR Section Citations
        The regulations at 40 CFR 141.28(a) applicable compliance 
    determinations by certified laboratories currently read as follows:
    
        For the purpose of determining compliance with Secs. 141.21 
    through 141.27, 141.41 and 141.42, samples may be considered only if 
    they have been analyzed by a laboratory certified by the State 
    except that measurements for turbidity, free chlorine residual, 
    temperature and pH may be performed by any person acceptable to the 
    State.''
    
    EPA proposes to change the term ``free chlorine residual'' in this 
    section to ``disinfectant residual'' for accuracy. The term ``free 
    chlorine'' is confusing because it begs the question whether a 
    laboratory must be certified to conduct analysis for disinfectant 
    residuals other than free chlorine residuals. The proposed change would 
    specify the more correct and broader term, disinfectant residual. 
    Current federal regulations do not require certification for measuring 
    any of the disinfectant residuals. States already approve the 
    measurement of disinfectant residuals by any person acceptable to the 
    State.
        EPA also proposes to amend Sec. 141.28(a) to add trihalomethane, 
    unregulated contaminant, filtration and disinfectant, and lead 
    monitoring sections (Secs. 141.30, 141.40, 141.74, 141.89, 
    respectively) and to remove obsolete section references to sodium and 
    corrosivity monitoring (Secs. 141.41 and 141.42). This change would 
    impose no new burden because certification requirements for these 
    parameters are currently specified elsewhere in federal regulations. 
    The change merely clarifies that 40 CFR 141.28(a) also applies to these 
    parameters.
        EPA proposes to amend 40 CFR 141.28(a) and 141.89(a)(1) to address 
    confusion about whether a laboratory must be certified to conduct 
    monitoring for certain additional contaminants. Currently, 40 CFR 
    141.89(a)(1) requires certification to conduct analyses for lead, 
    copper, alkalinity, calcium, conductivity, orthophosphate and silica. 
    However, the regulations specify acceptance criteria only for the 
    analysis of samples for lead and copper and not for alkalinity, 
    calcium, conductivity, orthophosphate and silica. Acceptance criteria 
    for these latter five contaminants were omitted because EPA intended 
    for the certification exception provisions of 40 CFR 141.28(a) to apply 
    to these contaminants the same way the exceptions apply for pH, 
    temperature, turbidity and disinfectant residuals. The certification 
    exceptions at 40 CFR 141.28 (a) state that certain analyses may be 
    conducted by any person acceptable to the State. The regulations at 40 
    CFR 141.28 (a) and 141.89(a)(1) would be amended to clarify EPA's 
    intent that analyses for alkalinity, calcium, conductivity, 
    disinfectant residual, orthophosphate, pH, silica, temperature and 
    turbidity may be performed by any person acceptable to the State. 
    Section 40 CFR 141.89 would no longer require certification for the 
    measurement of these contaminants.
    Requirement To Report Method Used To Analyze PE Samples
        EPA proposes to amend the NPDWRs at Secs. 141.23(k), 141.24(f) and 
    (h) and 141.89(a) in order to provide greater assurance that public 
    water systems dependably comply with MCLs. Those regulations describe 
    the Performance Evaluation (PE) sample analysis requirements associated 
    with compliance monitoring for certain contaminants regulated under the 
    SDWA. The regulations provide that analysis for the drinking water 
    contaminants regulated under those sections must be conducted by a 
    laboratory that has received certification by EPA or the State. These 
    regulations further provide that, in order for a laboratory to receive 
    certification to conduct analyses for the regulated contaminants, the 
    laboratory must analyze PE samples for the regulated contaminants. PE 
    sample testing is intended to ensure that the laboratory accurately 
    measures the contaminant of concern, i.e., is proficient in the 
    analysis for such contaminants.
        To ensure that public water systems dependably comply with MCLs, 
    EPA proposes to amend the PE sample requirement to specify that a 
    laboratory needs to be certified not only for measurement of specific 
    contaminants, but also that such certification must specify the 
    method(s) used by that laboratory for evaluation of such 
    contaminant(s). The Agency's primary concern is that laboratories be 
    certified for the proficiency on the methods they actually use for 
    analysis for water samples used in compliance monitoring. In the 
    recently published fourth edition of the Manual for the Certification 
    of Drinking Water Laboratories (EPA 1997b), EPA's laboratory 
    certification guidance, the Agency recommended that EPA and State 
    certifying agencies require a laboratory to successfully analyze a PE 
    sample for each method for which the laboratory seeks certification 
    rather than one PE sample per analyte. Today's proposal would codify 
    that
    
    [[Page 41139]]
    
    recommendation. EPA believes this is a reasonable change because 
    laboratories always have had to report to the PE sample provider what 
    method was used to analyze the PE sample. Many, if not most, State 
    laboratory certification officers currently require this information.
        As currently written, the PE sample requirements at 
    Secs. 141.23(k), 141.24(f) and (h) and 141.89(a) allow certification 
    for any method approved for measurement of the target analyte provided 
    that the laboratory successfully analyzed the PE sample with any one of 
    the approved methods. Under today's proposal, EPA would clarify the 
    regulations to carry out the original intent of requiring a 
    certification certificate to describe both the analyte and approved 
    method covered by the certificate. The proposed change would, for 
    example, require successful analysis of a PE sample with both EPA 
    Method 524.2 and 502.2 to obtain certification for VOCs if the 
    laboratory elects to use both mass spectrometer and electrolytic 
    conductivity/photo ionization technologies, respectively, to measure 
    VOCs. Under today's proposal, no laboratory would be required to use 
    methods that it would not otherwise use. The requirement would merely 
    require proficiency in using the methods the laboratory has chosen to 
    use. This amendment would only affect laboratories that both (1) choose 
    to be certified for the same analyte with more than one method and (2) 
    are not already required by their certification authority to be 
    certified by method and by analyte. EPA is interested in any comments 
    that provide compelling reasons to show that this modification to the 
    PE sample requirements would be unnecessary or unreasonable. EPA is 
    also seeking comments on this requirement in regards to the national 
    standards for accreditation of environmental laboratories which 
    National Environmental Laboratory Accreditation Conference (NELAC) is 
    developing, and the Agency's implementation of a Performance-Based 
    Measurement System in the future. NELAC is sponsored by EPA as a 
    voluntary association of state and federal officials with the purpose 
    of developing national performance standards for environmental 
    laboratories. NELAC standards are available on the Internet at http://
    www.epa.gov/ttn/nelac/.
    
    E. Performance-Based Measurement System
    
        On October 6, 1997, EPA published a Notice of the Agency's intent 
    to implement a Performance-Based Measurement System (PBMS) in all of 
    its programs to the extent feasible (62 FR 52098). The Agency is 
    currently determining the specific steps necessary to implement PBMS in 
    its programs and preparing an implementation plan. Final decisions have 
    not yet been made concerning the implementation of PBMS in water 
    programs. However, EPA is currently evaluating what relevant 
    performance characteristics should be specified for monitoring methods 
    used in the water programs under a PBMS approach to ensure adequate 
    data quality. EPA would then specify performance requirements in its 
    regulations to ensure that any method used for determination of a 
    regulated analyte is at least equivalent to the performance achieved by 
    other currently approved methods. Our expectation is that EPA will 
    publish its PBMS implementation strategy for water programs in the 
    Federal Register by the end of calendar year 1998.
        Once EPA has made its final determinations regarding implementation 
    of PBMS in programs under the Safe Drinking Water Act, EPA would 
    incorporate specific provisions of PBMS into its regulations, which may 
    include specification of the performance characteristics for 
    measurement of regulated contaminants in the drinking water program 
    regulations.
    
    IV. Regulation Assessment Requirements
    
    A. Executive Order 12866
    
        Under Executive Order 12866 (58 FR 51735; October 4, 1993), the 
    Agency must determine whether the regulatory action is ``significant'' 
    and therefore subject to OMB review and the requirements of the 
    Executive Order. The Order defines ``significant regulatory action'' as 
    one that is likely to result in a rule that may:
        (1) Have an annual effect on the economy of $100 million or more, 
    or adversely affect in a material way the economy, a sector of the 
    economy, productivity, competition, jobs, the environment, public 
    health or safety, or State, local, or tribal governments or 
    communities;
        (2) Create a serious inconsistency or otherwise interfere with an 
    action taken or planned by another agency;
        (3) Materially alter the budgetary impact of entitlement, grants, 
    user fees, or loan programs or the rights and obligations of recipients 
    thereof; or
        (4) Raise novel legal or policy issues arising out of legal 
    mandates, the President's priorities, or the principles set forth in 
    the Executive Order.
        It has been determined that this rule is not a ``significant 
    regulatory action'' under the terms of Executive Order 12866 and is 
    therefore not subject to OMB review.
    
    B. Regulatory Flexibility Act
    
        Under the Regulatory Flexibility Act (RFA), EPA generally is 
    required to conduct a regulatory flexibility analysis describing the 
    impact of the regulatory action on small entities as part of 
    rulemaking. However, under section 605(b) of the RFA, if EPA certifies 
    that the rule will not have a significant economic impact on a 
    substantial number of small entities, EPA is not required to prepare a 
    regulatory flexibility analysis. Pursuant to section 605(b) of the 
    Regulatory Flexibility Act, 5 U.S.C. 605(b), the Administrator 
    certifies that this rule will not have a significant economic impact on 
    a substantial number of small entities (for purposes of EPA's analysis, 
    the Agency defines small entities as 10,000 or fewer customers or small 
    laboratories with sales revenue below $6 million annually).
        This rule proposes to approve revised methods for compliance 
    determination of selected chemical and microbial contaminants. Previous 
    versions of all methods for these contaminants except for diquat will 
    continue to be approved. EPA is proposing to withdraw approval of its 
    current method for diquat. EPA has determined that requiring pH 
    adjustment prior extraction is not necessary. Because the proposed 
    diquat method is similar in all other aspects to the currently approved 
    method, withdrawal of this method and adoption of the new method will 
    not have any economic burden on public water systems or the analytical 
    laboratories performing analyses on behalf of these systems whether 
    small entities or small businesses.
        The proposed amendment would also require that laboratories be 
    certified for proficiency in the analytical method they actually use. 
    Thus, in the case of laboratories that choose to be certified for an 
    analyte using more than one approved method, the regulation would 
    require such laboratories to analyze a PE sample for each method for 
    which certification is requested. Small laboratories are not expected 
    to consider this option. Currently most laboratories elect to be 
    certified for only one method and there is no reason to believe this 
    situation will change. Even if some small laboratories elected to seek 
    certification for more than one method for some analytes, EPA has 
    concluded that the number doing so would be few and the consequent 
    economic impact on
    
    [[Page 41140]]
    
    laboratories that are small businesses would not be significant.
        The proposal to hold source water samples for microbiological 
    analysis below 10  deg.C during transit/storage is also not expected to 
    cause any significant increase in monitoring cost for big or small 
    water systems. The Agency anticipates that only about 2,000 surface 
    water systems which do not currently filter water (approx. 1% of all 
    systems) will be affected by the cooling requirement for total coliform 
    and fecal coliform samples. All surface water systems (approx. 9% of 
    all systems) will be affected by the cooling requirement for 
    heterotrophic plate count (HPC). However, the HPC requirement is an 
    optional substitute for maintaining a detectable disinfection residual. 
    The requirement to hold samples below 10  deg.C can be easily met by 
    shipping samples in reusable ice packs. EPA estimates a one time cost 
    of less than $5 per sample for the ice packs; over a period of time 
    this represents only a slight increase in sample shipping cost under 
    current requirements. Based on the above stated considerations, the 
    Agency concluded that this proposal, if promulgated, will not have a 
    significant economic impact on a substantial number of small entities.
    
    C. Unfunded Mandates Reform Act
    
        Title II of the Unfunded Mandates Reform Act of 1995 (UMRA), Pub.L. 
    104-4, establishes requirements for Federal agencies to assess the 
    effects of their regulatory actions on State, local, and tribal 
    governments and the private sector. Under section 202 of the UMRA, EPA 
    generally must prepare a written statement, including a cost-benefit 
    analysis, for proposed and final rules with ``Federal mandates'' that 
    may result in expenditures to State, local, and tribal governments, in 
    the aggregate, or to the private sector, of $100 million or more in any 
    one year. Before promulgating an EPA rule for which a written statement 
    is needed, section 205 of the UMRA generally requires EPA to identify 
    and consider a reasonable number of regulatory alternatives and adopt 
    the least costly, most cost-effective or least burdensome alternative 
    that achieves the objectives of the rule. The provisions of section 205 
    do not apply when they are inconsistent with applicable law. Moreover, 
    section 205 allows EPA to adopt an alternative other than the least 
    costly, most cost-effective or least burdensome alternative if the 
    Administrator publishes with the final rule an explanation why that 
    alternative was not adopted. Before EPA establishes any regulatory 
    requirements that may significantly or uniquely affect small 
    governments, including tribal governments, it must have developed under 
    section 203 of the UMRA a small government agency plan. The plan must 
    provide for notifying potentially affected small governments, enabling 
    officials of affected small governments to have meaningful and timely 
    input in the development of EPA regulatory proposals with significant 
    Federal intergovernmental mandates, and informing, educating, and 
    advising small governments on compliance with the regulatory 
    requirements.
        EPA has determined that this rule does not contain any Federal 
    mandate that may result in expenditures of $100 million or more for 
    State, local, and tribal governments, in the aggregate, or the private 
    sector in any one year. Therefore, today's proposed rule is not subject 
    to the requirements of section 202 and 205 of the UMRA.
        Today's rule proposes use of additional analytical methods by 
    public water systems and laboratories conducting analysis for 
    contaminants in drinking water. Therefore, the proposal would provide 
    regulatory relief in the form of increased operational flexibility to 
    laboratory analysts. Earlier versions of all methods approved in this 
    proposed rule (except for the diquat method) would continue to be 
    approved. The withdrawal of the earlier version of the diquat method 
    would not adversely affect analytical laboratories because the new 
    version is simpler and easier to use. The proposed amendment requiring 
    that laboratories be certified based on the proficiency on the method 
    they actually use, would impose a minor requirement for laboratories 
    that choose to be certified for an analyte by more than one method. 
    Currently most laboratories elect to be certified by one method and 
    there is no reason to believe laboratories would be compelled to incur 
    the cost of an additional PE sample in the future. The requirement to 
    hold source water samples below 10 deg.C during transit/storage would 
    affect only a small fraction (1-9%) of the water utilities and the 
    effect on monitoring cost would be very minor, and attributable to a 
    slight increase in sample shipping cost.
        Based on the information presented above, EPA has determined that 
    this rule contains no regulatory requirements that might significantly 
    or uniquely affect small governments. Therefore, today's proposed rule 
    is not subject to section 203 of UMRA.
    
    D. Paperwork Reduction Act
    
        In accordance with the Paperwork Reduction Act of 1980, 44 U.S.C. 
    3501 et seq., EPA must submit an information collection request 
    covering information collection requirements in a rule to the Office of 
    Management and Budget (OMB) for review and approval. This rule contains 
    no information collection requirements. Therefore, preparation of an 
    information collection request to accompany this rule is unnecessary.
    
    E. Science Advisory Board and National Drinking Water Advisory Council, 
    and Secretary of Health and Human Services
    
        In accordance with section 1412(d) and (e) of the SDWA, the Agency 
    submitted this proposal to the Science Advisory Board, the National 
    Drinking Water Advisory Council, and the Secretary of Health and Human 
    Services for their review. They had no comments.
    
    F. National Technology Transfer and Advancement Act
    
        Under section 12(d) of the National Technology Transfer and 
    Advancement Act, the Agency is directed to use voluntary consensus 
    standards in its regulatory activities unless to do so would be 
    inconsistent with applicable law or otherwise impractical. Voluntary 
    consensus standards are technical standards (e.g., material 
    specifications, test methods, sampling procedures, business practices, 
    etc.) that are developed or adopted by voluntary consensus standard 
    bodies. Where available and potentially applicable voluntary consensus 
    standards are not used by EPA, the Act requires the Agency to provide 
    Congress, through the Office of Management and Budget (OMB), an 
    explanation for the reasons for not using the standard.
        In preparing today's proposal, EPA searched for consensus methods 
    that would be acceptable for compliance determinations under the SDWA 
    for the measurement of diquat and six acid herbicides. EPA found no 
    methods for diquat but found two methods (ASTM D5317-93 and SM 6640B) 
    for the acid herbicides. As explained in the preamble to this proposed 
    rule, EPA is proposing D5317-93 provided the QC criterion is superseded 
    by a more stringent EPA criterion. EPA is not proposing SM 6640B 
    because of significant shortcomings in the sample preparation and 
    quality control sections of the method instructions. EPA will offer to 
    work with Standard Methods to revise SM 6640B for publication in future 
    editions of Standard Methods for the Examination of Water and 
    Wastewater. EPA invites comments on the potential use of voluntary 
    consensus standards in this rulemaking. EPA invites public comments on 
    the
    
    [[Page 41141]]
    
    Agency's proposal as well as identification and information about other 
    voluntary consensus standards, which the Agency could consider for 
    determination of diquat or acid herbicides under the SDWA.
    
    G. Executive Order 13045--Protection of Children From Environmental 
    Health Risks and Safety Risks
    
        Today's action is not subject to Executive Order 13045 (62 FR 19885 
    (April 23, 1997)), which requires agencies to identify and assess the 
    environmental health and safety risks of their rules on children. 
    Pursuant to the definitions in section 2-202, Executive Order 13045 
    only applies to rules that are economically significant as defined 
    under Executive Order 12886 and concern an environmental health or 
    safety risk that may disproportionately affect children. This rule is 
    not economically significant and does not concern a risk 
    disproportionately affecting children.
    
    V. References
    
        APHA 1995. Nineteenth edition of Standard Methods for the 
    Examination of Water and Wastewater, 1995, American Public Health 
    Association, 1015 Fifteenth Street NW, Washington, DC 20005.
        ASTM 1996. Annual Book of ASTM Standards, 1996, Vol. 11.02, 
    American Society for Testing and Materials, 101 Barr Harbor Drive, 
    West Conshohocken, PA 19428.
        Brenner 1993. Brenner, K.P, et al., ``New medium for the 
    simultaneous detection of total coliform and Escherichia coli in 
    water'', Appl. Environ. Microbiol. 59:3534-3544.
        Brenner 1996a. Brenner, K.P., et al., ``Comparison of the 
    recoveries of Escherichia coli and total coliform from drinking 
    water by the MI Agar method'', Appl. Environ. Microbiol. 62:203-208.
        Brenner 1996b. Brenner, K.P., et al., ``Interlaboratory 
    evaluation of MI Agar and the U.S. Environmental Protection Agency--
    approved membrane filter method for the recovery of total coliform 
    and Escherichia coli from drinking water'', J. Microbiol. Methods 
    27:111-119.
        EPA 1995. ``Methods for the Determination of Organic Compounds 
    in Drinking Water--Supplement III,'' EPA-600/R-95-131, August 1995, 
    NTIS PB95-261616.
        EPA 1996. Draft Method 515.3 is available from U.S. 
    Environmental Protection Agency, National Exposure Research 
    Laboratory (NERL)-Cincinnati, 26 West Martin Luther King Drive, 
    Cincinnati, OH 45268.
        EPA 1997a. Draft Method 549.2 is available from U.S. 
    Environmental Protection Agency, National Exposure Research 
    Laboratory (NERL)-Cincinnati, 26 West Martin Luther King Drive, 
    Cincinnati, OH 45268.
        EPA 1997b. Manual for the Certification of Laboratories 
    Analyzing Drinking Water, Fourth Edition, Office of Water Resource 
    Center (RC-4100), 401 M Street, SW, Washington, DC 20460, EPA 815-B-
    97-001, March 1997.
    
    List of Subjects in 40 CFR Part 141
    
        Environmental protection, Analytical methods, Chemicals, 
    Incorporation by reference, Intergovernmental relations, 
    Microorganisms, Monitoring, National Primary Drinking Water 
    Regulations, Water supply.
    
        Dated: July 23, 1998.
    Carol M. Browner,
    Administrator.
    
        For the reasons set out in the preamble, part 141 of chapter I, 
    title 40, Code of Federal Regulations, are proposed to be amended as 
    follows:
    
    PART 141--NATIONAL PRIMARY DRINKING WATER REGULATIONS
    
        1. The authority citation for part 141 continues to read as 
    follows:
    
        Authority: 42 U.S.C. 300f, 300g-1, 300g-2, 300g-3, 300g-4, 300g-
    5, 300g-6, 300j-4, 300j-9.
    
        2. Section 141.21 is amended by revising paragraph (f)(3) and 
    adding a new (f)(6)(v) with a table to read as follows:
    
    
    Sec. 141.21  Coliform sampling.
    
    * * * * *
        (f) * * *
        (3) Public water systems must conduct total coliform analyses in 
    accordance with one of the analytical methods in the following table.
    
    ----------------------------------------------------------------------------------------------------------------
                       Organism                            Methodology                      Citation \1\            
    ----------------------------------------------------------------------------------------------------------------
    Total Coliforms \2\..........................  Total Coliform Fermentation  9221A, B                            
                                                    Technique 3,4,5.                                                
                                                   Total Coliform Membrane      9222A, B, C                         
                                                    Filter Technique \6\.                                           
                                                   Presence-Absence (P-A)       9221D                               
                                                    Coliform Test 5,7.                                              
                                                   ONPG-MUG Test \8\..........  9223                                
                                                   Colisure Test \9\                                                
    ----------------------------------------------------------------------------------------------------------------
    The procedures shall be done in accordance with the documents listed below. The incorporation by reference of   
      the following documents was approved by the Director of the Federal Register in accordance with 5 U.S.C.      
      552(a) and 1 CFR part 51. Copies of the documents may be obtained from the sources listed below. Information  
      regarding obtaining these documents can be obtained from the Safe Drinking Water Hotline at 800-426-4791.     
      Documents may be inspected at EPA's Drinking Water Docket, 401 M Street, SW, Washington, D.C. 20460           
      (Telephone: 202-260-3027); or at the Office of Federal Register, 800 North Capitol Street, NW, Suite 700,     
      Washington, D.C. 20408.                                                                                       
    \1\ Methods 9221A, B, 9222A, B, C, 9221D and 9223 are contained in Standard Methods for the Examination of Water
      and Wastewater, 18th edition, 1992 and 19th edition, 1995, American Public Health Association, 1015 Fifteenth 
      Street NW, Washington, D.C. 20005; either edition may be used.                                                
    \2\ The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged   
      but not required to hold samples below 10 deg. C during transit.                                              
    \3\ Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system       
      conducts at least 25 parallel tests between this medium and lauryl tryptose broth using the water normally    
      tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total       
      coliform, using lactose broth, is less than 10 percent.                                                       
    \4\ If inverted tubes are used to detect gas production, the media should cover these tubes at least one-half to
      two-thirds after the sample is added.                                                                         
    \5\ No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed     
      tubes.                                                                                                        
    \6\ MI agar also may be used. Preparation and use of MI agar is set forth in the article, ``New medium for the  
      simultaneous detection of total coliform and Escherichia coli in water'' by Brenner, K.P., et al., 1993, Appl.
      Environ. Microbiol. 59:3534-3544.                                                                             
    \7\ Six-times formulation strength may be used if the medium is filter-sterilized rather than autoclaved.       
    \8\ The ONPG-MUG Test is also known as the Autoanalysis Colilert System. A source for this test is referenced at
      Sec.  141.21(f)(5)(iii).                                                                                      
    \9\ The Colisure Test must be incubated for 28 hours before examining the results. If an examination of the     
      results at 28 hours is not convenient, then results may be examined at any time between 28 hours and 48 hours.
      A description of the Colisure Test may be obtained from the Millipore Corporation, Technical Services         
      Department, 80 Ashby Road, Bedford, MA 01730.                                                                 
    *                    *                    *                    *                    *                    *      
                   *                                                                                                
    
        (6) * * *
        (v) The membrane filter method with MI agar, a description of which 
    is cited in a footnote to the table at paragraph (f)(3) of this 
    section.
    * * * * *
        3. Section 141.23(k)(3)(ii) is amended by revising the introductory 
    text before the table to read as follows:
    
    [[Page 41142]]
    
    Sec. 141.23  Inorganic chemical sampling and analytical requirements.
    
    * * * * *
        (k) * * *
        (3) * * *
        (ii) For each contaminant that has been included in the PE sample 
    and for each method for which the laboratory desires certification 
    achieve quantitative results on the analyses that are within the 
    following acceptance limits:
    * * * * *
        4. Section 141.24 is amended by:
        a. Revising the section heading;
        b. Adding a sentence to the end of paragraph (e);
        c. Revising, in the table in paragraphs (e), the entries for ``2,4-
    D'', ``2,4,5-TP (Silvex)'', ``Dinoseb'', ``Diquat'', 
    ``Pentachlorophenol'' and ``Picloram''; and adding footnotes 4 and 5;
        d. Revising paragraphs (f)(17)(i)(A), (f)(17)(ii) introductory 
    text; and paragraph (f)(17)(ii)(A); and
        e. Revising paragraph (h)(19)(i)(A) to read as follows:
    
    
    Sec. 141.24  Organic chemicals, sampling and analytical requirements.
    
    * * * * *
        (e) * * * EPA Draft Methods 515.3 and 549.2 are available from U.S. 
    Environmental Protection Agency, National Exposure Research Laboratory 
    (NERL)-Cincinnati, 26 West Martin Luther King Drive, Cincinnati, OH 
    45268. ASTM Method D 5317-93 is available in the Annual Book of ASTM 
    Standards, 1996, Vol. 11.02, American Society for Testing and 
    Materials, 101 Barr Harbor Drive, West Conshohocken, PA 19428, or in 
    any edition published after 1993.
    
                                                                            
               Contaminant                           Method \1\             
                                                                            
    *    *    *                              *    *    *                    
    2,4-D \4\ (as acid, salts and      515.2, 555, 515.1, 515.3, D5317-93   
     esters)                            \5\                                 
    2,4,5-TP \4\ (Silvex)              515.2, 555, 515.1, 515.3, D5317-93   
                                        \5\                                 
    *    *    *                              *    *    *                    
    Dinoseb \4\                        515.2, 555, 515.1, 515.3             
    Diquat                             549.2                                
    *    *    *                              *    *    *                    
    Pentachlorophenol                  515.2, 525.2, 555, 515.1, 515.3,     
                                        D5317-93 \5\                        
    Picloram \4\                       515.2, 555, 515.1, 515.3, D5317-93   
                                        \5\                                 
    *    *    *                              *    *    *                    
                                                                            
    \4\ Accurate determination of the chlorinated esters requires hydrolysis
      of the sample as described in EPA Methods 515.1, 515.2, 515.3 and 555,
      and ASTM Method D5317-93.                                             
    \5\ Use of ASTM D5317-93 requires that the variable control limit for   
      the laboratory fortified blank (LFB), which is specified in section 6g
      of D5317-93, be superseded by the fixed upper control limit of 30%, which is specified in sections 9.3.2 and 9.7.2 of EPA      
      Method 515.1 (Rev. 4.1).                                              
    
        (f) * * *
        (17) * * *
        (i) * * *
        (A) Analyze Performance Evaluation (PE) samples provided by EPA, 
    the State, or by a third party (with the approval of the State or EPA) 
    at least once a year by each method for which the laboratory desires 
    certification.
    * * * * *
        (ii) To receive certification to conduct analyses for vinyl 
    chloride, the laboratory must:
        (A) Analyze Performance Evaluation (PE) samples provided by EPA, 
    the State, or by a third party (with the approval of the State or EPA) 
    at least once a year by each method for which the laboratory desires 
    certification.
    * * * * *
        (h) * * *
        (19) * * *
        (i) * * *
        (A) Analyze Performance Evaluation (PE) samples provided by EPA, 
    the State, or by a third party (with the approval of the State or EPA) 
    at least once a year by each method for which the laboratory desires 
    certification.
    * * * * *
        5. Section 141.28 is amended by revising paragraph (a) to read as 
    follows:
    
    
    Sec. 141.28  Certified laboratories.
    
        (a) For the purpose of determining compliance with Secs. 141.21 
    through 141.27, 141.30, 141.40, 141.74 and 141.89, samples may be 
    considered only if they have been analyzed by a laboratory certified by 
    the State except that measurements for alkalinity, calcium, 
    conductivity, disinfectant residual, orthophosphate, pH, silica, 
    temperature and turbidity may be performed by any person acceptable to 
    the State.''
    * * * * *
        6. Section 141.40 is amended by revising the last sentence in 
    paragraph (n)(11), entry for ``dicamba'' in paragraph (n)(11), and by 
    adding two footnotes to the table in paragraph (n)(11) to read as 
    follows:
    
    
    Sec. 141.40  Special monitoring for inorganic and organic chemicals.
    
    * * * * *
        (n) * * *
        (11) * * * A source for ASTM D5317-93, APHA method 6610 and for EPA 
    Methods 505, 507, 508, 508.1, 515.2, 515.3, 525.2 and 531.1 is 
    referenced at Sec. 141.24(e).
    
                                                                            
               Contaminants                            Method               
                                                                            
    *    *    *                         *    *    *                         
    dicamba \1\                         515.2, 555, 515.1, 515.3, D5317-93  
                                         \2\                                
    *    *    *                         *    *    *                         
                                                                            
    \1\ Accurate determination of the chlorinated esters requires hydrolysis
      of the sample as described in EPA Methods 515.1, 515.2, 515.3 and 555,
      and ASTM Method D5317-93.                                             
    \2\ Use of ASTM D5317-93 requires that the variable control limit for   
      the laboratory fortified blank (LFB), which is specified in section 6g
      of D5317-93, be superseded by the fixed upper control limit of  30%, which is specified in sections 9.3.2 and 9.7.2 of EPA     
      Method 515.1 (Rev. 4.1).                                              
    
    * * * * *
        7. Section 141.74 is amended by revising the table and footnotes in 
    paragraph (a)(1) to read as follows:
    
    
    Sec. 141.74  Analytical and monitoring requirements.
    
        (a) * * *
        (1) * * *
    
    ----------------------------------------------------------------------------------------------------------------
                       Organism                            Methodology                      Citation \1\            
    ----------------------------------------------------------------------------------------------------------------
    Total Coliform \2\...........................  Total Coliform Fermentation  9221A, B, C                         
                                                    Technique 3,4,5.                                                
                                                   Total Coliform Membrane      9222A, B, C                         
                                                    Filter Technique \6\.                                           
                                                   ONPG-MUG Test \7\..........  9223                                
    Fecal Coliforms \2\..........................  Fecal Coliform Procedure     9221E                               
                                                    \8\.                                                            
                                                   Fecal Coliform Filter        9222D                               
                                                    Procedure.                                                      
    Heterotrophic bacteria \2\...................  Pour Plate Method..........  9215B                               
    Turbidity....................................  Nephelometric Method.......  2130B                               
                                                   Nephelometric Method.......  180.1 \9\                           
                                                   Great Lakes Instruments....  Method 2 \10\                       
    ----------------------------------------------------------------------------------------------------------------
    The procedures shall be done in accordance with the documents listed below. The incorporation by reference of   
      the following documents was approved by the Director of the Federal Register in accordance with 5 U.S.C.      
      552(a) and 1 CFR part 51. Copies of the documents may be obtained from the sources listed below. Information  
      regarding obtaining these documents can be obtained from the Safe Drinking Water Hotline at 800-426-4791.     
      Documents may be inspected at EPA's Drinking Water Docket, 401 M Street, SW, Washington, D.C. 20460           
      (Telephone: 202-260-3027); or at the Office of Federal Register, 800 North Capitol Street, NW, Suite 700,     
      Washington, D.C. 20408.                                                                                       
    
    [[Page 41143]]
    
                                                                                                                    
    \1\ Except where noted, all methods refer to Standard Methods for the Examination of Water and Wastewater, 18th 
      edition, 1992 and 19th edition, 1995, American Public Health Association, 1015 Fifteenth Street NW,           
      Washington, D.C. 20005; either edition may be used.                                                           
    \2\ The time from sample collection to initiation of analysis may not exceed 8 hours. Systems must hold samples 
      below 10 deg.C during transit.                                                                                
    \3\ Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system       
      conducts at least 25 parallel tests between this medium and lauryl tryptose broth using the water normally    
      tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total       
      coliform, using lactose broth, is less than 10 percent.                                                       
    \4\ Media should cover inverted tubes at least one-half to two-thirds after the sample is added.                
    \5\ No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed     
      tubes.                                                                                                        
    \6\ MI agar also may be used. Preparation and use of MI agar is set forth in the article, ``New medium for the  
      simultaneous detection of total coliform and Escherichia coli in water'' by Brenner, K.P., et al., 1993, Appl.
      Environ. Microbiol. 59:3534-3544.                                                                             
    \7\ The ONPG-MUG Test is also known as the Autoanalysis Colilert System. A source for this test is referenced at
      Sec.  141.21(f)(5)(iii).                                                                                      
    \8\ A-1 Broth may be held up to three months in a tightly closed screw cap tube at 4C.                          
    \9\ ``Methods for the Determination of Inorganic Substances in Environmental Samples'', EPA-600/R-93-100, August
      1993. Available at NTIS, PB94-121811.                                                                         
    \10\ GLI Method 2, ``Turbidity'', November 2, 1992, Great Lakes Instruments, Inc., 8855 North 55th Street,      
      Milwaukee, Wisconsin 53223.                                                                                   
    *                    *                    *                    *                    *                    *      
                   *                                                                                                
    
        8. Section 141.89 is amended by revising paragraph (a)(1) 
    introductory text and (a)(1)(i) to read as follows:
    
    
    Sec. 141.89  Analytical methods.
    
    * * * * *
        (a) * * *
        (1) Analyses for alkalinity, calcium, conductivity, 
    orthophosophate, pH, silica, and temperature may be performed by any 
    person acceptable to the State. Analyses under this section for lead 
    and copper shall only be conducted by laboratories that have been 
    certified by EPA or the State. To obtain certification to conduct 
    analyses for lead and copper, laboratories must:
        (i) Analyze Performance Evaluation samples, which include lead and 
    copper, provided by or acceptable to EPA or the State at least once a 
    year by each method for which the laboratory desires certification; and
    * * * * *
    
    [FR Doc. 98-20281 Filed 7-30-98; 8:45 am]
    BILLING CODE 6560-50-P
    
    
    

Document Information

Published:
07/31/1998
Department:
Environmental Protection Agency
Entry Type:
Proposed Rule
Action:
Proposed rule.
Document Number:
98-20281
Dates:
Comments on this proposed rule must be received in writing by September 29, 1998.
Pages:
41134-41143 (10 pages)
Docket Numbers:
WH-FRL-6132-3
RINs:
2040-AD04: National Primary and Secondary Drinking Water Regulations: Analytical Methods for Chemical and Microbiological Contaminants and Revisions to Laboratory Certification Requirements
RIN Links:
https://www.federalregister.gov/regulations/2040-AD04/national-primary-and-secondary-drinking-water-regulations-analytical-methods-for-chemical-and-microb
PDF File:
98-20281.pdf
CFR: (8)
40 CFR 141.21(f)(5)(iii)
40 CFR 141.21
40 CFR 141.23
40 CFR 141.24
40 CFR 141.28
More ...