94-20042. Ambient Air Quality Surveillance Siting Criteria for Open Path Analyzers  

  • [Federal Register Volume 59, Number 159 (Thursday, August 18, 1994)]
    [Unknown Section]
    [Page 0]
    From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
    [FR Doc No: 94-20042]
    
    
    [[Page Unknown]]
    
    [Federal Register: August 18, 1994]
    
    
    -----------------------------------------------------------------------
    
    
    ENVIRONMENTAL PROTECTION AGENCY
    40 CFR Part 58
    
    [FRL-4842-4]
    
     
    
    Ambient Air Quality Surveillance Siting Criteria for Open Path 
    Analyzers
    
    AGENCY: Environmental Protection Agency (EPA).
    
    ACTION: Proposed rule.
    
    -----------------------------------------------------------------------
    
    SUMMARY: The EPA proposes to amend provisions of part 58 of chapter I 
    of title 40 of the Code of Federal Regulations to define the 
    appropriate ambient air monitoring criteria for open path (long-path) 
    analyzers. The proposed revisions to the Ambient Air Quality 
    Surveillance regulations would define the siting requirements for open 
    path analyzers used as State and Local Air Monitoring Stations (SLAMS), 
    which includes both National Air Monitoring Stations (NAMS) and 
    Photochemical Assessment Monitoring Stations (PAMS), as well as the 
    quality assurance procedures for this technology. These changes will 
    allow the ambient air monitoring community to effectively use open path 
    monitoring data for regulatory purposes.
    
    DATES: Comments must be received on or before September 19, 1994. 
    Requests for public hearing must be received by September 2, 1994. If a 
    hearing is held, comments must be received on or before 30 days from 
    the conclusion of the hearing.
    
    ADDRESSES: Comments should be submitted (in duplicate, if possible) to: 
    Air Docket (LE-131), Attention: Docket Number A-93-44, U.S. 
    Environmental Protection Agency, room M-1500, 401 M Street, SW., 
    Washington, DC 20460.
        Public hearing: A public hearing will be held, if requested, in 
    accordance with information provided in the DATES section of this 
    proposal, to provide interested parties an opportunity for oral 
    presentation of data, views, or arguments concerning the proposed 
    revisions. If anyone contacts EPA requesting a public hearing, it will 
    be held at the EPA's Environmental Research Center, Research Triangle 
    Park, North Carolina. Persons interested in attending the hearing or 
    wishing to present oral testimony should notify Ms. Lee Ann B. Byrd, 
    Monitoring and Reports Branch (MD-14), U.S. Environmental Protection 
    Agency, Research Triangle Park, North Carolina 27711, telephone number 
    (919) 541- 5367. Specific dates and other pertinent details of this 
    public hearing will be published in a separate Federal Register notice.
        Docket: Docket Number A-93-44, containing supporting information 
    used in developing these revised regulations, is available for public 
    inspection and copying between 8:30 a.m. and 12 noon, and between 1:30 
    p.m. and 3:30 p.m., Monday through Friday, at EPA's Air Docket Section 
    at the address noted above. As provided in 40 CFR Part 2, a reasonable 
    fee may be charged for copying.
    
    FOR FURTHER INFORMATION CONTACT: Ms. Lee Ann B. Byrd at telephone (919) 
    541-5367 concerning this action. The address is Monitoring and Reports 
    Branch (MD-14), U.S. Environmental Protection Agency, Research Triangle 
    Park, North Carolina 27711.
    
    SUPPLEMENTARY INFORMATION:
    
    Table of Contents
    
    I. Authority
    II. Background of Proposed Rule
    III. Discussion of Proposed Revisions to Regulation
        A. Section 58.1 Definitions
        B. Appendix A--Quality Assurance Requirements for State and 
    Local Air Monitoring Stations (SLAMS)
        C. Appendix B--Quality Assurance Requirements for Prevention of 
    Significant Deterioration (PSD) Air Monitoring
        D. Appendix E--Probe and Path Siting Criteria for Ambient Air 
    Quality Monitoring
    IV. Comments and the Public Docket
    V. Administrative Requirements
        A. Administrative Designation.
        B. Reporting and Recordkeeping Requirements
        C. Regulatory Flexibility Act
    
    I. Authority
    
        Authority: sections 110, 301(a), 313, and 319 of the Clean Air 
    Act as amended 42 U.S.C. 7410, 7601(a), 7613, 7619.
    
    II. Background of Proposed Rule
    
        The Clean Air Act, as amended in 1990, requires, in sections 
    181(b)(2), 185A, and 186(b)(2)(A), ambient air quality monitoring for 
    purposes of defining areas of nonattainment with the National Ambient 
    Air Quality Standards (NAAQS), evaluating progress toward achievement 
    of the NAAQS pursuant to State implementation plans (SIP's), and 
    reporting air quality data to the EPA to document the status and trends 
    of the Nation's air quality. These are nontrivial activities, and to 
    achieve the aforementioned objectives, the EPA must ensure that the 
    ambient air monitoring networks consist of high quality instruments 
    that produce accurate concentration measurements. As new monitoring 
    techniques are developed, the EPA evaluates the new methodology and, as 
    appropriate, determines how to effectively incorporate it into the 
    existing air quality monitoring program. To assess new ambient air 
    monitoring instruments for those pollutants with established NAAQS, the 
    EPA currently uses the Ambient Air Monitoring Reference and Equivalent 
    Method regulatory procedures detailed in title 40, chapter 1, part 53. 
    The EPA does not formally regulate the performance testing of ambient 
    air monitoring instruments, which measure pollutants without 
    established NAAQS. Methodology for collected ozone (O3) precursor 
    data (specified in the Photochemical Assessment Monitoring program as 
    volatile organic compounds and oxides of nitrogen) is reviewed in the 
    ``Technical Assistance Document for Sampling and Analysis of Ozone 
    Precursors,'' and any subsequent revisions, EPA/600-8-91/215, October 
    1991. Supplementing the part 53 performance requirements and the 
    aforementioned ``Technical Assistance Document,'' the part 58 Ambient 
    Air Quality Surveillance regulation specifies how to most appropriately 
    conduct routine ambient air monitoring through pollutant-specific 
    monitor siting criteria, operation schedules, monitoring network 
    design, and data reporting. Under the part 58 provisions, each SLAMS 
    must employ reference or equivalent methods, as determined according to 
    part 53, and meet all applicable siting requirements as contained in 
    part 58, before its data can be used for regulatory purposes. 
    Specifically, these regulatory actions include comparison with the 
    NAAQS and other SIP-related activities. It is important to note that 
    the NAMS and the PAMS are subsets of the SLAMS networks; therefore, 
    provisions for the SLAMS also apply to both the NAMS and PAMS, as 
    included in this proposal.
        A new technique for monitoring pollutants in ambient air has been 
    developed and introduced to the EPA. Instruments based on this new 
    technique, called open path (or long-path) analyzers, use ultraviolet, 
    visible, or infrared light to measure nitrogen dioxide (NO2), 
    O3, carbon monoxide (CO), sulfur dioxide (SO2), and other 
    gaseous pollutant concentrations over a path of several meters up to 
    several kilometers. The concentration measurements obtained by these 
    open path analyzers are path-integrated, or path-averaged, values. 
    Traditional fixed point analyzers measure pollutant concentrations at 
    one specific point by extracting an air sample from the atmosphere 
    through an inlet probe. A list of all EPA-approved reference and 
    equivalent ambient air monitoring methods is available through the 
    docket. Due to the fundamental difference in the measurement principles 
    of open path and point analyzers, there may be trade-offs in using each 
    type of instrument for certain applications. Because of the ability of 
    open path analyzers to measure pollutant concentrations over a path, 
    these new techniques are expected to provide better spatial coverage, 
    and thereby a better assessment of a general population's exposure to 
    air pollutants for certain applications. However, due to this same 
    path-averaging characteristic, open path analyzers could underestimate 
    high pollutant concentrations at specific points within the measurement 
    path for other ambient air monitoring situations. The applicability of 
    either technique to a particular monitoring scenario is dependent on a 
    number of factors including plume dispersion characteristics, 
    monitoring location, pollutant of interest, population density, site 
    topography, and monitoring objective. The EPA has considered these 
    factors in evaluating the advantages and disadvantages of using open 
    path analyzers for the various ambient air monitoring applications 
    detailed in 40 CFR part 58. Additionally, several studies of the 
    comparability of data collected with point and open path analyzers have 
    been conducted by the EPA and by other organizations. The most recent 
    EPA study of these two methodologies was completed during the summer of 
    1993 in Baytown, Texas. Results from this study and others are 
    available in the docket for public review. The EPA solicits comment on 
    these studies and on the comparability of using path-averaged and point 
    measurements in the Nation's ambient air monitoring programs.
        The EPA is currently assessing the performance of an open path 
    analyzer as a candidate method under part 53 to determine if it should 
    be designated as an equivalent method for one or more of those 
    pollutants. In parallel with this effort, the EPA has developed the 
    appropriate part 58 siting and quality assurance criteria for open path 
    analyzers, which are contained in this proposal.
        The existing part 58 monitoring network design criteria define the 
    monitoring objectives for a particular site in terms of measurement 
    scale. More specifically, each ambient air monitoring station is 
    located in such a way that it represents a particular air parcel or 
    volume. The regulation uses six measurement scales to describe the size 
    of these air parcels. These six scales are: microscale (dimensions of 
    several meters to approximately 100 meters), middle scale (100 to 500 
    meters), neighborhood scale (500 meters to 4 kilometers), urban scale 
    (4 to 50 kilometers), regional scale (tens to hundreds of kilometers), 
    and national or global scales. (National and global scales are 
    generally not applicable for a single air monitoring station. National 
    and global averages are more appropriately determined by networks of 
    various monitoring stations.) Within each of these measurement scales, 
    it is assumed that the pollutant concentrations are relatively 
    homogeneous; therefore, a monitor placed at any point in the area, 
    within the tolerances of this siting regulation, measures a 
    concentration representative of that area.
        Depending on the objective for a particular SLAMS, each pollutant 
    can be monitored on a particular measurement scale as defined in Table 
    5 of appendix D in part 58. The applicability of the first five scales 
    to monitoring the four pollutants referenced in this proposal, CO, 
    NO2, O3, and SO2, follows:
    
                  Summary of Spatial Scales for Selected SLAMS              
    ------------------------------------------------------------------------
                                                                   Ozone    
          Measurement Scale         CO     NO2    O3     SO2   precursors\1\
    ------------------------------------------------------------------------
    Micro-.......................  Yes..  No...  No...  No...  No.          
    scale........................                                           
    Middle scale.................  Yes..  Yes..  Yes..  Yes..  No.          
    Neighborhood scale...........  Yes..  Yes..  Yes..  Yes..  Yes.         
    Urban scale..................  No...  Yes..  Yes..  Yes..  Yes.         
    Regional.....................  No...  No...  Yes..  Yes..  No.          
    ------------------------------------------------------------------------
    \1\Ozone precursors, as defined in the PAMS program, include volatile   
      organic compounds, oxides of nitrogen, and selected carbonyls.        
    
        Existing regulations in part 58 state that the pollutant 
    concentration within a particular measurement scale is nearly 
    homogeneous, and that a point measurement collected in this same scale 
    generally represents any other point within that scale. This basic 
    provision defines how ambient air monitoring data can be used to 
    represent the air quality in a neighborhood, city, or other geographic 
    region. Based on these current provisions, it is reasonable to accept 
    that a path-averaged measurement taken within the dimensions and other 
    siting specifications of that measurement scale would provide a value 
    descriptive of that same geographic region. In order to maintain data 
    comparability between open path and point analyzers, the revisions 
    contained in this proposal are based on the siting criteria currently 
    being used with conventional fixed point ambient air monitoring 
    networks. The most obvious difference between the proposed and existing 
    siting criteria is that the new requirements are defined in terms of a 
    ``probe'' (applicable to point analyzers), a ``monitoring path'' 
    (applicable to open path analyzers), or both. Some minor flexibility in 
    siting criteria was added for open path analyzers to compensate for the 
    additional difficulties in locating suitable sites for the various 
    equipment used with an open path analyzer, such as retroreflectors, 
    receivers, and transmitters. Nonetheless, these criteria should still 
    provide a concentration representative of the area to be monitored.
        It is important to note that criteria for open path measurement of 
    CO in a street canyon scenario, typically defined in terms of 
    microscale dimensions (up to 100 meters), is not included in this 
    proposal. The siting criteria currently used for microscale CO 
    monitoring is unique and narrow in scope in comparison to other 
    monitoring scenarios. Adapting the existing siting criteria to 
    accommodate path measurement techniques, as this proposal does for 
    other types of monitoring scales, would unduly restrict the usage of 
    open path analyzers for this particular application. In order to fully 
    address more appropriate siting criteria for microscale CO monitoring 
    using open path analyzers, the EPA must more fully evaluate the effects 
    of measuring path-averaged CO concentrations across roadways, 
    intersections, and at locations other than those currently defined in 
    the part 58 regulation. The EPA specifically solicits comments from the 
    public regarding the use of open path analyzers for measuring CO in 
    microscale applications.
    
    III. Discussion of Proposed Revisions to Regulations
    
    A. Section 58.1  Definitions
    
        Today's proposal would amend the definitions section of part 58 by 
    adding several new definitions that are necessary to clearly define the 
    proposed new requirements for open path analyzers. Definitions for 
    ``point analyzer'' and ``open path analyzer'' would be added to define 
    these two types of automated instruments and to clarify the distinction 
    between them, since the various new and existing requirements may apply 
    to one or the other or both types of analyzers. A new definition for 
    ``probe'' is proposed to specify the inlet where an air sample is 
    extracted from the atmosphere for delivery to a sampler or point 
    analyzer. This definition would clarify that location requirements 
    applicable to point analyzers apply to the analyzer's probe and not to 
    the analyzer (or sampler) itself, which could be located some distance 
    from the probe. Similarly, a new definition is proposed for 
    ``monitoring path'' to describe the path in the atmosphere over which 
    an open path analyzer measures and averages a pollutant concentration. 
    Closely associated with the term ``monitoring path'' are new 
    definitions for ``monitoring path length,'' to describe the scalar 
    length of the monitoring path, and ``optical measurement path length,'' 
    to describe the actual length of the optical beam of an open path 
    instrument. The length of the optical beam may be two or more times the 
    length of the monitoring path when one or more mirrors are used to 
    cause the optical beam to pass through the monitoring path more than 
    once.
        To help describe the new requirements for data quality assessment 
    procedures, the term ``effective concentration'' is proposed. It would 
    refer to the ambient concentration of a pollutant over the monitoring 
    path that would be equivalent to a much higher concentration of the 
    pollutant contained in a short calibration cell inserted into the 
    optical beam of an open path analyzer during a precision test or 
    accuracy audit. Specifically, effective concentration is proposed to be 
    defined as the actual concentration of the pollutant in the test cell 
    multiplied by the ratio of the optical measurement path length of the 
    test cell to the optical measurement path length of the atmospheric 
    monitoring path. Also, when a calibration cell is inserted into the 
    actual atmospheric measurement beam of an open path analyzer for a 
    precision or accuracy test, the resulting measurement reading would be 
    the sum of the pollutant concentration in the calibration cell and the 
    pollutant concentration in the atmosphere. The atmospheric pollutant 
    concentration must be measured separately and subtracted from the test 
    measurement to produce a ``corrected concentration,'' which would be 
    the true test result. Thus, the term ``corrected concentration'' is 
    proposed to define the result of such a precision or accuracy 
    assessment test after correction of the test measurement by subtracting 
    the atmospheric pollutant concentration.
        Finally, a formal definition of ``monitor'' is proposed to clarify 
    its use in the regulations as a generic term to refer to any type of 
    ambient air analyzer or sampler that is acceptable for use in a SLAMS 
    monitoring network under Appendix C of this part. A monitor could thus 
    be a point analyzer, an open path analyzer, or a sampler.
    
    B. Appendix A--Quality Assurance Requirements for State and Local Air 
    Monitoring Stations (SLAMS)
    
        Appendix A sets forth both general quality assurance requirements 
    applicable to SLAMS air monitoring as well as specific procedures for 
    assessing the quality of the monitoring data obtained in SLAMS 
    monitoring networks. While the general quality assurance requirements 
    (in section 2) would be directly applicable to open path analyzers 
    without change, the more specific data quality assessment procedures 
    (in section 3) must be modified somewhat to apply to open path 
    analyzers. Accordingly, changes to these procedures are proposed to 
    incorporate appropriate data quality assessment tests applicable to 
    open path monitoring instruments. To the extent possible, the new 
    requirements are similar or parallel to the existing requirements for 
    point analyzers.
        For both the precision test (Sec. 3.1) and the accuracy audit 
    (Sec. 3.2), the proposed new requirements specify that an optical 
    calibration or test cell containing a pollutant concentration standard 
    must be inserted into the optical measurement beam of the open path 
    analyzer. Both theory and testing indicate that the use of such a 
    calibration or test cell is equivalent in accuracy to measurement of 
    the equivalent pollutant concentration in air over the entire 
    monitoring path of an open path analyzer. Each concentration standard 
    must be selected such that it produces an ``effective concentration'' 
    equivalent to a specified ambient concentration over the monitoring 
    path. As noted previously, effective concentration is defined as the 
    actual concentration of the pollutant in the test cell multiplied by 
    the ratio of the optical measurement path length of the test cell to 
    the optical measurement path length of the atmospheric monitoring path. 
    The effective concentrations specified for the precision and accuracy 
    tests for open path analyzers would be the same as the test 
    concentrations currently specified in these procedures for point 
    analyzers.
        Ideally, precision and accuracy assessments should test a 
    monitoring instrument in its normal monitoring configuration. 
    Therefore, the proposed test procedures require that the test or 
    calibration cell containing the test pollutant concentration standard 
    be inserted into the actual atmospheric measurement beam of the open 
    path analyzer. The resulting test measurement of the pollutant 
    concentration would thus be the sum of the test concentration in the 
    cell and the pollutant concentration in the atmosphere, because the 
    measurement beam would pass through both the test cell and the 
    atmospheric monitoring path. Accordingly, a correction for the 
    atmospheric concentration is required to obtain the true test result. 
    In the proposed procedures, the atmospheric pollutant concentration 
    would be measured immediately before and again immediately after the 
    precision or accuracy test, and the average of these two measurements 
    would be subtracted from the test concentration measurement to produce 
    a ``corrected concentration,'' which would be reported as the test 
    result.
        The corrected concentration reported for a precision or accuracy 
    test may not be accurate if the atmospheric pollutant concentration 
    changes during the test. When the ambient concentration is variable, 
    the average of the pre- and post-test measurements may not be an 
    accurate representation of the ambient pollutant concentration during 
    the test. The proposed test procedures recommend that these tests 
    should be carried out, if possible, during periods when the atmospheric 
    pollutant concentration is low and steady. The lower the atmospheric 
    pollutant concentration, the steadier the concentration is likely to be 
    and the better the pre- and post-test measurements will represent the 
    actual atmospheric concentration during the test measurement. Further, 
    the procedures propose that if the pre- and post-test measurements of 
    the atmospheric concentration differ by more that 20 percent of the 
    effective concentration of the test standard, the test result would be 
    discarded and the test repeated.
        It is recognized that the proposed tests for precision and accuracy 
    for open path analyzers, as well as the existing tests for point 
    analyzers, are described in very general terms, and that additional, 
    more detailed information and guidance is usually necessary for an 
    analyzer operator to carry out these tests properly. Accordingly, 
    section 3 of appendix A is proposed to be amended by adding an explicit 
    indication that supplemental information and guidance to assist the 
    analyst in conducting these tests may be available in the publication, 
    ``Quality Assurance Handbook for Air Pollution Measurement Systems, 
    Volume II'' (EPA-600/4-77-027a, identified as Reference 3 at the end of 
    Appendix A), or in the operation or instruction manual associated with 
    the particular monitor being used.
        The proposed techniques for precision and accuracy assessment of 
    open path analyzers are based largely on consultations with the 
    manufacturer, along with EPA tests, of the differential optical 
    absorption spectrometer that is currently under consideration by EPA 
    for possible designation as equivalent methods under 40 CFR part 53. 
    However, it is desirable that the techniques be generic in nature, if 
    possible, so that they would be applicable to other types of open path 
    monitoring instruments as well. In addition, for some types of open 
    path instruments or for some installations or configurations, there may 
    be technical reasons why the proposed techniques for precision and 
    accuracy assessment may not be feasible, appropriate, or advisable. The 
    procedures, as currently proposed, allow for the use of an alternate 
    local light source or an alternate optical path that does not include 
    the normal atmospheric monitoring path, if such alternate configuration 
    is permitted by the operation or instruction manual associated with the 
    analyzer. Since the analyzer operation or instruction manual would be 
    subject to approval as part of the requirements for EPA designation of 
    an open path analyzer as an equivalent method, EPA would thereby have 
    control over the alternate configurations that would be allowable for 
    the precision and accuracy assessment tests.
        In view of these issues regarding the precision and accuracy 
    assessment techniques, EPA specifically solicits comments on: (1) The 
    suitability of the proposed techniques; (2) the advisability of a 
    technique that requires correction of the test result for the 
    atmospheric pollutant concentration versus a technique that does not 
    require that correction but does not test the normal atmospheric 
    measurement components and configuration; (3) the proposed technique 
    for correcting test measurements for the atmospheric pollutant 
    concentration, if required, and the 20 percent limit on the difference 
    between the pre- and post-test measurements of the atmospheric 
    concentration; and (4) whether the proposed techniques are sufficiently 
    generic in nature to apply to various other types of open path 
    analyzers that might be applicable to SLAMS monitoring, or how the 
    techniques could be made more generic.
    
    C. Appendix B--Quality Assurance Requirements for Prevention of 
    Significant Deterioration (PSD) Air Monitoring
    
        Appendix B sets forth both general quality assurance requirements 
    for PSD monitoring as well as specific procedures for assessing the 
    quality of the monitoring data obtained in PSD monitoring networks. The 
    amendments and procedures proposed for Appendix B to extend the 
    existing requirements to open path analyzers are essentially identical 
    to the changes proposed for Appendix A.
    
    D. Appendix E--Probe and Path Siting Criteria for Ambient Air Quality 
    Monitoring
    
        This proposal would amend Appendix E by adding new siting criteria 
    applicable to open path analyzers for monitoring of SO2, O3, 
    NO2, CO, and O3 precursors (defined in the PAMS program as 
    volatile organic compounds, oxides of nitrogen, and selected 
    carbonyls). Because of the substantial similarity in the siting 
    criteria for SO2, O3, and NO2 (both the existing 
    criteria for point monitors and proposed new criteria for open path 
    analyzers), the siting requirements for these three pollutants are 
    proposed to be combined, consolidated, and set forth in section 2 of 
    appendix E. The existing criteria for SO2, O3, and NO2 
    in sections 3, 5, and 6 would be deleted, and those sections would be 
    reserved. As noted below, the criteria for CO monitoring are somewhat 
    different, so they would be retained in a separate section 4. Siting 
    criteria for measuring O3 and its precursors as part of a PAMS 
    network are included in section 10. In all cases, the new open path 
    provisions would be incorporated into the existing provisions, as 
    appropriate.
        The proposed new open path siting requirements largely parallel the 
    existing requirements for point analyzers, with the revised provisions 
    applicable to either a ``probe'' (for point analyzers), a ``monitoring 
    path'' (for open path analyzers), or both, as appropriate. Accordingly, 
    criteria for the monitoring path of an open path analyzer are proposed 
    for horizontal and vertical placement, spacing from minor sources, 
    spacing from obstructions, spacing from trees, and spacing from 
    roadways. The open path requirements would apply to most of the 
    monitoring path--generally 80 or 90 percent--but not to the entire 
    monitoring path, to allow some needed flexibility in siting open path 
    analyzers. For example, using the proposed 80 percent requirement, a 
    monitoring path may be sited across uneven terrain, where up to 20 
    percent of the monitoring path may not fall within the proposed 3 to 15 
    meter specification for height above ground.
        In addition to the criteria common to both point and open path 
    analyzers mentioned above, two new provisions, applicable only to open 
    path analyzers, would limit the maximum length of the monitoring path 
    and the cumulative interferences on the path. The maximum monitoring 
    path length limit would help to ensure that open path monitoring data 
    represent the air volume that they are intended to measure according to 
    the monitoring objectives of the spatial scale identified for the site. 
    Similarly, the limit for the cumulative interferences on the monitoring 
    path would control the total amount of interferences from minor 
    sources, roadways, obstructions, and other factors that might unduly 
    influence the monitoring data collected by an open path analyzer. This 
    limit is necessary because a long monitoring path presents a much 
    greater opportunity to be affected by multiple interferences. It is 
    also recognized that State or local air monitoring agencies may 
    encounter difficulties in locating atmospheric monitoring equipment due 
    to vandalism, scarcity of available sites, and other considerations; 
    therefore, certain provisions are included in both the existing and the 
    proposed new provisions of the regulation to accommodate these 
    difficulties.
        In the consolidation of current sections 3, 5, and 6 to section 2, 
    Tables 2 and 3, which list the minimum separation distance between 
    O3 and NO2 stations and nearby roadways, would be combined 
    and redesignated as Table 1. As a result, Table 1 (in section 3), Table 
    4 (in section 7), Table 5 (in section 10), and Table 6 (in section 12) 
    would be renumbered as tables 2, 3, 4, and 5, respectively. Finally, 
    the summary of all the general siting requirements in renumbered Table 
    5 would be modified to include the new criteria for monitoring paths.
    
    IV. Comments and the Public Docket
    
        The EPA welcomes comments on all aspects of this proposed 
    rulemaking, specifically: (a) The appropriateness of using open path 
    (long-path) analyzers to measure CO, O3, SO2, NO2, and/
    or O3 precursors (defined in the PAMS program as volatile organic 
    compounds, oxides of nitrogen, and selected carbonyls); (b) the ability 
    of a monitoring agency to use an open path analyzer in a manner 
    consistent with these siting criteria; (c) using open path analyzers to 
    measure CO in microscale scenarios; (d) the precision and accuracy 
    assessment techniques as described in the proposed Appendix A and 
    Appendix B regulations; (e) using open path analyzers to measure 
    SO2 in source-oriented ambient air monitoring networks, 
    particularly in micro- and middle-scale applications; and (f) all 
    available and relevant study information on the comparability of open 
    path and point ambient air monitoring. All comments, with the exception 
    of proprietary information, should be directed to the EPA Air Docket 
    Section, Docket No. A-93-44.
        Those who wish to submit proprietary information for consideration 
    should clearly separate such information from other comments by:
         Labeling proprietary information ``Confidential Business 
    Information,'' and;
         Sending proprietary information directly to the contact 
    person listed (see FOR FURTHER INFORMATION CONTACT) and not to the 
    public docket.
        This will help ensure that proprietary information is not 
    inadvertently placed in the docket. If a commenter wants the EPA to use 
    a submission labeled as confidential business information as part of 
    the basis for the final rule, then a nonconfidential version of the 
    document, which summarizes the key data or information, should be sent 
    to the docket.
        Information covered by a claim of confidentiality will be disclosed 
    by the EPA only to the extent allowed and by the procedures set forth 
    in 40 CFR part 2. If no claim of confidentiality accompanies the 
    submission when it is received by the EPA, the submission may be made 
    available to the public without notifying the commenters.
    
    V. Administrative Requirements
    
    A. Administrative Designation
    
    Executive Order 12866
        Under Executive Order 12866 (58 FR 51735 (October 4, 1993)) the 
    Agency must determine whether the regulatory action is ``significant'' 
    and therefore subject to Office of Management and Budget (OMB) review 
    and the requirements of the Executive Order. The Order defines 
    ``significant regulatory action'' as one that is likely to result in a 
    rule that may:
        (1) Have an annual effect on the economy of $100 million or more or 
    adversely affect in a material way the economy, a sector of the 
    economy, productivity, competition, jobs, the environment, public 
    health or safety, or State, local, or tribal governments or 
    communities;
        (2) Create a serious inconsistency or otherwise interfere with an 
    action taken or planned by another Agency;
        (3) Materially alter the budgetary impact of entitlements, grants, 
    user fees, or loan programs or the rights and obligations or recipients 
    thereof; or
        (4) Raise novel legal or policy issues arising out of legal 
    mandates, the President's priorities, or the principles set forth in 
    the Executive Order.
        It has been determined that this rule is not a ``significant 
    regulatory action'' under the terms of Executive Order 12866 and is 
    therefore not subject to OMB review.
    
    B. Reporting and Recordkeeping Requirements
    
        All of the information collection requirements contained in part 58 
    have been approved by the OMB under the Paperwork Reduction Act of 
    1980, 44 U.S.C. 3501 et seq., and have been assigned OMB Control Number 
    2060-0084. This proposed amendment to Part 58 does not add any new 
    information collection requirements.
    
    C. Regulatory Flexibility Act
    
        Pursuant to section 605(b) of the Regulatory Flexibility Act, 5 
    U.S.C. 605(b), the Administrator certifies that this rule will not have 
    a significant economic impact on a substantial number of small 
    entities. This rulemaking package does not impose any additional 
    requirements on small entities, rather, it is this proposal's intent to 
    provide all entities with the option to choose the most suitable 
    ambient air method for their particular application. This proposal 
    provides the appropriate siting and quality assurance criteria for a 
    new ambient air monitoring technology (open path analyzers) as they are 
    used in various applications. All of the criteria listed in this 
    rulemaking package parallel existing requirements and vary only as 
    necessary due to technological differences between measurement 
    techniques. It is possible that a beneficial impact may be encountered 
    by some small entities that use this new technology in certain 
    scenarios.
    
    List of Subjects in 40 CFR Part 58
    
        Air pollution control, Ambient air monitoring, Ambient air 
    monitoring networks and siting criteria, Intergovernmental relations, 
    National ambient air monitoring program, Quality assurance 
    requirements, Reporting and recordkeeping requirements, State and local 
    agency ambient air monitoring programs.
    
        Dated: August 4, 1994.
    Carol M. Browner,
    Administrator.
    
        For reasons set forth in the preamble, title 40, chapter I, part 58 
    of the Code of Federal Regulations is proposed to be amended as 
    follows:
    
    PART 58--[AMENDED]
    
        1. The authority citation for part 58 continues to read as follows:
    
        Authority: 42 U.S.C. 7410, 7601(a), 7613, and 7619.
    
        2. In Sec. 58.1, the following definitions are added:
    
    
    Sec. 58.1  Definitions.
    
    * * * * *
        (z) Point analyzer is an automated analytical method that measures 
    pollutant concentration in an ambient air sample extracted from the 
    atmosphere at a specific inlet probe point and that has been designated 
    as a reference or equivalent method in accordance with part 53 of this 
    chapter.
        (aa) Probe is the actual inlet where an air sample is extracted 
    from the atmosphere for delivery to a sampler or point analyzer for 
    pollutant analysis.
        (bb) Open path analyzer is an automated analytical method that 
    measures the average atmospheric pollutant concentration in situ along 
    one or more monitoring paths having a monitoring path length of 5 
    meters or more and that has been designated as a reference or 
    equivalent method under the provisions of part 53 of this chapter.
        (cc) Monitoring path for an open path analyzer is the actual path 
    in space over which the pollutant concentration is measured and 
    averaged.
        (dd) Monitoring path length of an open path analyzer is the length 
    of the monitoring path in the atmosphere over which the average 
    pollutant concentration measurement is determined. See also, ``optical 
    measurement path length.''
        (ee) Optical measurement path length is the actual length of the 
    optical beam over which measurement of the pollutant is determined. 
    Generally, the optical measurement path length is:
        (1) Equal to the monitoring path length for a (bistatic) system 
    having transmitter and receiver at opposite ends of the monitoring 
    path;
        (2) Equal to twice the monitoring path length for a (monostatic) 
    system having a transmitter and receiver at one end of the monitoring 
    path and a mirror or retroreflector at the other end; or
        (3) Equal to some multiple of the monitoring path length for more 
    complex systems having multiple passes of the measurement beam through 
    the monitoring path.
        (ff) Effective concentration pertains to testing an open path 
    analyzer with a high-concentration calibration or audit standard gas 
    contained in a short test cell inserted into the optical measurement 
    beam of the instrument. Effective concentration is the equivalent 
    ambient-level concentration that would produce the same spectral 
    absorbance over the actual atmospheric monitoring path length as 
    produced by the high-concentration gas in the short test cell. 
    Quantitatively, effective concentration is equal to the actual 
    concentration of the gas standard in the test cell multiplied by the 
    ratio of the path length of the test cell to the actual atmospheric 
    monitoring path length.
        (gg) Corrected concentration pertains to the result of an accuracy 
    or precision assessment test of an open path analyzer in which a high-
    concentration test or audit standard gas contained in a short test cell 
    is inserted into the optical measurement beam of the instrument. When 
    the pollutant concentration measured by the analyzer in such a test 
    includes both the pollutant concentration in the test cell and the 
    concentration in the atmosphere, the atmospheric pollutant 
    concentration must be subtracted from the test measurement to obtain 
    the corrected concentration test result. The corrected concentration is 
    equal to the measured concentration minus the average of the 
    atmospheric pollutant concentrations measured (without the test cell) 
    immediately before and immediately after the test.
        (hh) Monitor is a generic term for an instrument, sampler, 
    analyzer, or other device that measures or assists in the measurement 
    of atmospheric air pollutants and is acceptable for use in ambient air 
    surveillance under the provisions of appendix C to this part, including 
    both point and open path analyzers that have been designated as 
    reference or equivalent methods under part 53 of this chapter and air 
    samplers that are specified as part of a manual method that has been 
    designated as a reference or equivalent method under part 53 of this 
    chapter.
    
    Appendix A [Amended]
    
        3. Appendix A is amended as follows:
        a. The fourth paragraph of section 3 introductory text is revised.
        b. Section 3.1 is revised.
        c. The text preceding the table in the second paragraph, and the 
    seventh, and eighth paragraphs of Sec. 3.2 are revised; and a new 
    paragraph is added between the seventh and eighth paragraphs.
        d. Table A-1 is revised.
    
    Appendix A--Quality Assurance Requirements for State and Local Air 
    Monitoring Stations (SLAMS)
    
    * * * * *
    
    3. Data Quality Assessment Requirements
    
    * * * * *
        Assessment results shall be reported as specified in section 4. 
    Concentration and flow standards must be as specified in Secs. 2.3 
    or 3.4. In addition, working standards and equipment used for 
    accuracy audits must not be the same standards and equipment used 
    for routine calibration. Additional information and guidance in the 
    technical aspects of conducting these tests may be found in 
    Reference 3 or in the operation or instruction manual associated 
    with the analyzer or sampler. Concentration measurements reported 
    from analyzers or analytical systems (indicated concentrations) 
    should be based on stable readings and must be derived by means of 
    the same calibration curve and data processing system used to obtain 
    the routine air monitoring data (see Reference 1 and Reference 3, 
    section 2.0.9.1.3(d)). Table A-1 provides a summary of the minimum 
    data quality assessment requirements, which are described in more 
    detail in the following sections.
    
    3.1  Precision of Automated Methods
    
        A one-point precision check must be carried out at least once every 
    2 weeks on each automated analyzer used to measure SO2, NO2, 
    O3, and CO. The precision check is made by challenging the 
    analyzer with a precision check gas of known concentration (effective 
    concentration for open path analyzers) between 0.08 and 0.10 ppm for 
    SO2, NO2, and O3 analyzers, and between 8 and 10 ppm for 
    CO analyzers. To check the precision of SLAMS analyzers operating on 
    ranges higher than 0 to 1.0 ppm SO2, NO2, and O3, or 0 
    to 100 ppm for CO, use precision check gases of appropriately higher 
    concentration as approved by the appropriate Regional Administrator or 
    the Regional Administrator's designee. However, the results of 
    precision checks at concentration levels other than those specified 
    above need not be reported to the EPA. The standards from which 
    precision check test concentrations are obtained must meet the 
    specifications of Sec. 2.3.
        Except for certain CO analyzers described below, point analyzers 
    must operate in their normal sampling mode during the precision 
    check, and the test atmosphere must pass through all filters, 
    scrubbers, conditioners, and other components used during normal 
    ambient sampling and as much of the ambient air inlet system as is 
    practicable. If permitted by the associated operation or instruction 
    manual, a CO point analyzer may be temporarily modified during the 
    precision check to reduce vent or purge flows, or the test 
    atmosphere may enter the analyzer at a point other than the normal 
    sample inlet, provided that the analyzer's response is not likely to 
    be altered by these deviations from the normal operational mode.
        If a precision check is made in conjunction with a zero or span 
    adjustment, it must be made prior to such zero or span adjustments. 
    Randomization of the precision check with respect to time of day, 
    day of week, and routine service and adjustments is encouraged where 
    possible.
        Open path analyzers are tested by inserting a test cell 
    containing a precision check gas concentration into the optical 
    measurement beam of the instrument. If possible, the normally used 
    transmitter, receiver, and, as appropriate, reflecting devices 
    should be used during the test, and the normal monitoring 
    configuration of the instrument should be altered as little as 
    possible to accommodate the test cell for the test. However, if 
    permitted by the associated operation or instruction manual, an 
    alternate local light source or an alternate optical path that does 
    not include the normal atmospheric monitoring path may be used. The 
    actual concentration of the precision check gas in the test cell 
    must be selected to produce an ``effective concentration'' in the 
    range specified above. Generally, the precision test concentration 
    measurement will be the sum of the atmospheric pollutant 
    concentration and the precision test concentration. If so, the 
    result must be corrected to remove the atmospheric concentration 
    contribution. The ``corrected concentration'' is obtained by 
    subtracting the average of the atmospheric concentrations measured 
    by the instrument immediately before and immediately after the 
    precision check test from the precision test concentration 
    measurement. If the difference between these before and after 
    measurements is greater than 20 percent of the effective 
    concentration of the test gas, discard the test result and repeat 
    the test. If possible, open path analyzers should be tested during 
    periods when the atmospheric pollutant concentrations are relatively 
    low and steady.
        Report the actual concentration (effective concentration for 
    open path analyzers) of the precision check gas and the 
    corresponding concentration measurement (corrected concentration, if 
    applicable, for open path analyzers) indicated by the analyzer. The 
    percent differences between these concentrations are used to assess 
    the precision of the monitoring data as described in Sec. 5.1.
    
    3.2  Accuracy of Automated Methods
    
    * * * * *
        The audit is made by challenging the analyzer with at least one 
    audit gas of known concentration (effective concentration for open 
    path analyzers) from each of the following ranges that fall within 
    the measurement range of the analyzer being audited:
    * * * * *
        For point analyzers, the audit shall be carried out by allowing 
    the analyzer to analyze the audit test atmosphere in its normal 
    sampling mode such that the test atmosphere passes through all 
    filters, scrubbers, conditioners, and other sample inlet components 
    used during normal ambient sampling and as much of the ambient air 
    inlet system as is practicable. The exception given in Sec. 3.1 for 
    certain CO analyzers does not apply for audits.
        Open path analyzers are audited by inserting a test cell 
    containing the various audit gas concentrations into the optical 
    measurement beam of the instrument. If possible, the normally used 
    transmitter, receiver, and, as appropriate, reflecting devices 
    should be used during the audit, and the normal monitoring 
    configuration of the instrument should be modified as little as 
    possible to accommodate the test cell for the audit. However, if 
    permitted by the associated operation or instruction manual, an 
    alternate local light source or an alternate optical path that does 
    not include the normal atmospheric monitoring path may be used. The 
    actual concentrations of the audit gas in the test cell must be 
    selected to produce ``effective concentrations'' in the ranges 
    specified in this Sec. 3.2. Generally, each audit concentration 
    measurement result will be the sum of the atmospheric pollutant 
    concentration and the audit test concentration. If so, the result 
    must be corrected to remove the atmospheric concentration 
    contribution. The ``corrected concentration'' is obtained by 
    subtracting the average of the atmospheric concentration measured by 
    the instrument immediately before and immediately after the audit 
    test (or preferably before and after each audit concentration level) 
    from the audit concentration measurement. If the difference between 
    the before and after measurements is greater than 20 percent of the 
    effective concentration of the test gas standard, discard the test 
    result for that concentration level and repeat the test for that 
    level. If possible, open path analyzers should be audited during 
    periods when the atmospheric pollutant concentrations are relatively 
    low and steady.
        Report both the audit test concentrations (effective 
    concentrations for open path analyzers) and the corresponding 
    concentration measurements (corrected concentrations, if applicable, 
    for open path analyzers) indicated or produced by the analyzer being 
    tested. The percent differences between these concentrations are 
    used to assess the accuracy of the monitoring data as described in 
    Sec. 5.2.
    * * * * *
    
                                    Table A-1.--Minimum Data Assessment Requirements                                
    ----------------------------------------------------------------------------------------------------------------
           Method            Assessment method          Coverage           Minimum frequency     Parameters reported
    ----------------------------------------------------------------------------------------------------------------
    Precision:                                                                                                      
        Automated Methods  Response check at      Each analyzer........  Once per 2 weeks.....  Actual              
         for SO2, NO2,      concentration                                                        concentration\2\   
         O3, and CO.        between .08 & .10                                                    and measured       
                            ppm (8 & 10 ppm for                                                  concentration.\3\  
                            CO)\2\.                                                                                 
        Manual methods     Collocated samplers..  1 site for 1-5 sites;  Once per week........  Two concentration   
         including lead.                           2 sites for 6-20                              measurements.      
                                                   sites; 3 sites > 20                                              
                                                   sites; (sites with                                               
                                                   highest conc.).                                                  
    Accuracy:                                                                                                       
        Automated Methods  Response check at:     1. Each analyzer. 2.   1. Once per year. 2.   Actual              
         for SO2, NO2,      .03-.08 ppm,12 .15-    25% of analyzers (at   Each calendar          concentration\2\   
         O3, and CO.        .20 ppm;12 .35-.45     least 1).              quarter.               and measured       
                            ppm;12 .80-.90                                                       (indicated)        
                            ppm;12 (if                                                           concentration\3\   
                            applicable).                                                         for each level.    
        Manual methods     Check of analytical    Analytical system....  Each day samples are   Actual concentration
         for SO2 and NO2.   procedures with                               analyzed, at least     and measured       
                            audit standard                                twice per quarter.     (indicated)        
                            solutions.                                                           concentration for  
                                                                                                 each audit         
                                                                                                 solution.          
        TSP, PM-10.......  Check of sampler flow  1. Each sampler. 2.    1. Once per year. 2.   Actual flow rate and
                            rate.                  25% of samplers (at    Each calendar          flow rate indicated
                                                   least 1).              quarter.               by the sampler.    
        Lead.............  1. Check sample flow   1. Each sampler. 2.    1. Include with TSP.   1. Same as for TSP. 
                            rate as for TSP. 2.    Analytical system.     2. Each quarter.       2. Actual          
                            Check analytical                                                     concentration &    
                            system with Pb audit                                                 measured           
                            strips.                                                              (indicated)        
                                                                                                 concentration of   
                                                                                                 audit samples      
                                                                                                 (g Pb/    
                                                                                                 strip).            
    ----------------------------------------------------------------------------------------------------------------
    \1\Concentration times 100 for CO.                                                                              
    \2\Effective concentration for open path analyzers.                                                             
    \3\Corrected concentration, if applicable, for open path analyzers.                                             
    
    * * * * *
    
    Appendix B [Amended]
    
        4. Appendix B is amended as follows:
        a. The first paragraph of section is revised.
        b. Section 3.1 is revised.
        c. The text preceding the table in the first paragraph, and the 
    third, and fourth paragraphs of section 3.2 are revised; and a new 
    paragraph is added between the third and fourth paragraphs.
        d. Table B-1 is revised.
    
    Appendix B--Quality Assurance Requirements for Prevention of 
    Significant Deterioration (PSD) Air Monitoring
    
    * * * * *
    
    3. Data Quality Assessment Requirements
    
        All ambient monitoring methods or analyzers used in PSD 
    monitoring shall be tested periodically, as described in this 
    section 3, to quantitatively assess the quality of the data being 
    routinely collected. The results of these tests shall be reported as 
    specified in section 6. Concentration standards used for the tests 
    must be as specified in section Sec. 2.3. Additional information and 
    guidance in the technical aspects of conducting these tests may be 
    found in Reference 3 or in the operation or instruction manual 
    associated with the analyzer or sampler. Concentration measurements 
    reported from analyzers or analytical systems must be derived by 
    means of the same calibration curve and data processing system used 
    to obtain the routine air monitoring data. Table B-1 provides a 
    summary of the minimum data quality assessment requirements, which 
    are described in more detail in the following sections.
    
    3.1  Precision of Automated Methods
    
        A one-point precision check must be carried out at least once 
    every 2 weeks on each automated analyzer used to measure SO2, 
    NO2, O3, and CO. The precision check is made by 
    challenging the analyzer with a precision check gas of known 
    concentration (effective concentration for open path analyzers) 
    between 0.08 and 0.10 ppm for SO2, NO2, and O3 
    analyzers, and between 8 and 10 ppm for CO analyzers. The standards 
    from which precision check test concentrations are obtained must 
    meet the specifications of section 2.3. Except for certain CO 
    analyzers described below, point analyzers must operate in their 
    normal sampling mode during the precision check, and the test 
    atmosphere must pass through all filters, scrubbers, conditioners 
    and other components used during normal ambient sampling and as much 
    of the ambient air inlet system as is practicable. If permitted by 
    the associated operation or instruction manual, a CO point analyzer 
    may be temporarily modified during the precision check to reduce 
    vent or purge flows, or the test atmosphere may enter the analyzer 
    at a point other than the normal sample inlet, provided that the 
    analyzer's response is not likely to be altered by these deviations 
    from the normal operational mode.
        Open path analyzers are tested by inserting a test cell 
    containing a precision check gas concentration into the optical 
    measurement beam of the instrument. If possible, the normally used 
    transmitter, receiver, and, as appropriate, reflecting devices 
    should be used during the test, and the normal monitoring 
    configuration of the instrument should be altered as little as 
    possible to accommodate the test cell for the test. However, if 
    permitted by the associated operation or instruction manual, an 
    alternate local light source or an alternate optical path that does 
    not include the normal atmospheric monitoring path may be used. The 
    actual concentration of the precision check gas in the test cell 
    must be selected to produce an ``effective concentration'' in the 
    range specified above. Generally, the precision test concentration 
    measurement will be the sum of the atmospheric pollutant 
    concentration and the precision test concentration. If so, the 
    result must be corrected to remove the atmospheric concentration 
    contribution. The ``corrected concentration'' is obtained by 
    subtracting the average of the atmospheric concentrations measured 
    by the instrument immediately before and immediately after the 
    precision check test from the precision test concentration 
    measurement. If the difference between these before and after 
    measurements is greater than 20 percent of the effective 
    concentration of the test gas, discard the test result and repeat 
    the test. If possible, open path analyzers should be tested during 
    periods when the atmospheric pollutant concentrations are relatively 
    low and steady.
        If a precision check is made in conjunction with a zero or span 
    adjustment, it must be made prior to such zero or span adjustment. 
    The difference between the actual concentration (effective 
    concentration for open path analyzers) of the precision check gas 
    and the corresponding concentration measurement (corrected 
    concentration, if applicable, for open path analyzers) indicated by 
    the analyzer is used to assess the precision of the monitoring data 
    as described in 4.1. Report data only from automated analyzers that 
    are approved for use in the PSD network.
    
    3.2  Accuracy of Automated Methods
    
        Each sampling quarter audit each analyzer that monitors for 
    SO2, NO2, O3, or CO at least once. The audit is made 
    by challenging the analyzer with at least one audit gas of known 
    concentration (effective concentration for open path analyzers) from 
    each of the following ranges that fall within the measurement range 
    of the analyzer being audited: * * *
        For point analyzers, the audit shall be carried out by allowing 
    the analyzer to analyze the audit test atmosphere in the same manner 
    as described for precision checks in Sec. 3.1. The exception given 
    in Sec. 3.1 for certain CO analyzers does not apply for audits.
        Open path analyzers are audited by inserting a test cell 
    containing an audit gas concentration into the optical measurement 
    beam of the instrument. If possible, the normally used transmitter, 
    receiver, and, as appropriate, reflecting devices should be used 
    during the audit, and the normal monitoring configuration of the 
    instrument should be modified as little as possible to accommodate 
    the test cell for the audit. However, if permitted by the associated 
    operation or instruction manual, an alternate local light source or 
    an alternate optical path that does not include the normal 
    atmospheric monitoring path may be used. The actual concentrations 
    of the audit gas in the test cell must be selected to produce 
    ``effective concentrations'' in the range specified in this section 
    3.2. Generally, each audit concentration measurement result will be 
    the sum of the atmospheric pollutant concentration and the audit 
    test concentration. If so, the result must be corrected to remove 
    the atmospheric concentration contribution. The ``corrected 
    concentration'' is obtained by subtracting the average of the 
    atmospheric concentrations measured by the instrument immediately 
    before and immediately after the audit test (or preferably before 
    and after each audit concentration level) from the audit 
    concentration measurement. If the difference between these before 
    and after measurements is greater than 20 percent of the effective 
    concentration of the test gas standards, discard the test result for 
    that concentration level and repeat the test for that level. If 
    possible, open path analyzers should be audited during periods when 
    the atmospheric pollutant concentrations are relatively low and 
    steady.
        The differences between the actual concentrations (effective 
    concentrations for open path analyzers) of the audit test gas and 
    the corresponding concentration measurements (corrected 
    concentrations, if applicable, for open path analyzers) indicated by 
    the analyzer are used to assess the accuracy of the monitoring data 
    as described in Sec. 4.2. Report data only from automated analyzers 
    that are approved for use in the PSD network.
    * * * * *
    
                                  Table B-1.--Minimum PSD Data Assessment Requirements                              
    ----------------------------------------------------------------------------------------------------------------
           Method            Assessment method          Coverage               Frequency         Parameters reported
    ----------------------------------------------------------------------------------------------------------------
    Precision:                                                                                                      
        Automated Methods  Response check at      Each analyzer........  Once per 2 weeks.....  Actual              
         for SO2, NO2,      concentration                                                        concentration\2\   
         O3, and CO.        between .08 and .10                                                  and measured       
                            ppm (8 and 10 ppm                                                    concentration.\3\  
                            for CO)\2\.                                                                             
        TSP, PM10, Lead..  Collocated samplers..  Highest concentration  Once per week or       Two concentration   
                                                   site in monitoring     every 3rd day for      measurements.      
                                                   network.               continuous sampling.                      
    Accuracy:                                                                                                       
        Automated Methods  Response check at:     Each analyzer........  Once per sampling      Actual              
         for SO2, NO2,      .03-.08 ppm;\1\\2\                            quarter.               concentration\2\   
         O3, and CO.        .15-.20 ppm;\1\\2\                                                   and measured       
                            .35-.45 ppm;\1\\2\                                                   (indicated)        
                            .80-.90 ppm;\1\\2\                                                   concentration\3\   
                            (if applicable).                                                     for each level.    
        TSP, PM10........  Sampler flow check...  Each sampler.........  Once per sampling      Actual flow rate and
                                                                          quarter.               flow rate indicated
                                                                                                 by the sampler.    
        Lead.............  1. Sample flow rate    1. Each sampler. 2.    1. Once/quarter. 2.    1. Same as for TSP. 
                            check. 2. Check        Analytical system.     Each quarter Pb        2. Actual          
                            analytical system                             samples are analyzed.  concentration and  
                            with Pb audit strips.                                                measured           
                                                                                                 concentration of   
                                                                                                 audit samples      
                                                                                                 (g Pb/    
                                                                                                 strip).            
    ----------------------------------------------------------------------------------------------------------------
    \1\Concentration shown times 100 for CO.                                                                        
    \2\Effective concentration for open path analyzers.                                                             
    \3\Corrected concentration, if applicable, for open path analyzers.                                             
    
    * * * * *
    
    Appendix E [Amended]
    
        5. Appendix E is amended as follows:
        a. The title of appendix E is revised.
        b. Section 1 is revised.
        c. Section 2 is added and sections 3, 5, and 6 are removed and 
    reserved.
        d. Section 4 is revised.
        e. In section 7 table 3 is removed and table 4 is redesignated as 
    table 3.
        f. The first paragraph of section 9 is revised.
        g. Section 10 is revised.
        h. Section 12 is revised.
    
    Appendix E--Probe and Monitoring Path Siting Criteria for Ambient Air 
    Quality Monitoring
    
    1. Introduction
    
        This appendix contains specific location criteria applicable to 
    ambient air quality monitoring probes and monitoring paths after the 
    general station siting has been selected based on the monitoring 
    objectives and spatial scale of representation discussed in Appendix 
    D of this part. Adherence to these siting criteria is necessary to 
    ensure the uniform collection of compatible and comparable air 
    quality data.
        The probe and monitoring path siting criteria discussed below 
    must be followed to the maximum extent possible. It is recognized 
    that there may be situations where some deviation from the siting 
    criteria may be necessary. In any such case, the reasons must be 
    thoroughly documented in a written request for a waiver that 
    describes how and why the proposed siting deviates from the 
    criteria. This documentation should help to avoid later questions 
    about the validity of the resulting monitoring data. Conditions 
    under which EPA would consider an application for waiver from these 
    siting criteria are discussed in section 11 of this appendix.
        The spatial scales of representation used in this appendix, 
    i.e., micro, middle, neighborhood, urban, and regional, are defined 
    and discussed in Appendix D of this part. The pollutant-specific 
    probe and monitoring path siting criteria generally apply to all 
    spatial scales except where noted otherwise. Specific siting 
    criteria that are phrased with a ``must'' are defined as 
    requirements and exceptions must be approved through the waiver 
    provisions. However, siting criteria that are phrased with a 
    ``should'' are defined as goals to meet for consistency but are not 
    requirements.
    * * * * *
    
    2. Sulfur Dioxide (SO2), Ozone (O3), and Nitrogen Dioxide 
    (NO2)
    
        Additional information on SO2, NO2, and O3 
    monitor siting criteria may be found in references 11 and 13.
        2.1  Horizontal and Vertical Placement. The probe or at least 80 
    percent of the monitoring path must be located between 3 and 15 
    meters above ground level. The probe or at least 90 percent of the 
    monitoring path must be at least 1 meter vertically or horizontally 
    away from any supporting structure, walls, parapets, penthouses, 
    etc., and away from dusty or dirty areas. If the probe or a 
    significant portion of the monitoring path is located near the side 
    of a building, then it should be located on the windward side of the 
    building relative to the prevailing wind direction during the season 
    of highest concentration potential for the pollutant being measured.
        2.2  Spacing from Minor Sources (applicable to SO2 and 
    O3 monitoring only). Local minor sources of SO2 can cause 
    inappropriately high concentrations of SO2 in the vicinity of 
    probes and monitoring paths for SO2. Similarly, local sources 
    of nitric oxide (NO) and ozone-reactive hydrocarbons can have a 
    scavenging effect causing unrepresentatively low concentrations of 
    O3 in the vicinity of probes and monitoring paths for O3. 
    To minimize these potential interferences, the probe or at least 90 
    percent of the monitoring path must be away from furnace or 
    incineration flues or other minor sources of SO2 or NO, 
    particularly for open path analyzers because of their potential for 
    greater exposure over the area covered by the monitoring path. The 
    separation distance should take into account the height of the 
    flues, type of waste or fuel burned, and the sulfur content of the 
    fuel. It is acceptable, however, to monitor for SO2 near a 
    point source of SO2 when the objective is to assess the effect 
    of this source on the represented population.
        2.3  Spacing from Obstructions. Buildings and other obstacles 
    may possibly scavenge SO2, O3, or NO2. To avoid this 
    interference, the probe or at least 90 percent of the monitoring 
    path must have unrestricted airflow and be located away from 
    obstacles so that the distance from the probe or monitoring path is 
    at least twice the height that the obstacle protrudes above the 
    probe or monitoring path. Generally, a probe or monitoring path 
    located near or along a vertical wall is undesirable because air 
    moving along the wall may be subject to possible removal mechanisms. 
    A probe must have unrestricted airflow in an arc of at least 270 
    degrees around the inlet probe, or 180 degrees if the probe is on 
    the side of a building. This arc must include the predominant wind 
    direction for the season of greatest pollutant concentration 
    potential. A sampling station having a probe located closer to an 
    obstacle than this criterion allows should be classified as middle 
    scale rather than neighborhood or urban scale, since the 
    measurements from such a station would more closely represent the 
    middle scale. A monitoring path must be clear of all trees, brush, 
    buildings, plumes, dust, or other optical obstructions, including 
    potential obstructions that may move due to wind, human activity, 
    growth of vegetation, etc. Temporary optical obstructions, such as 
    rain, particles, fog, or snow, should be considered when siting an 
    open path analyzer. Any of these temporary obstructions that are of 
    sufficient density to obscure the light beam will affect the ability 
    of the open path analyzer to continuously measure pollutant 
    concentrations.
        2.4  Spacing from Trees. Trees can provide surfaces for 
    SO2, O3, or NO2 adsorption or reactions and obstruct 
    wind flow. To reduce this possible interference, the probe or at 
    least 90 percent of the monitoring path should be 20 meters or more 
    from the drip line of trees. If a tree or trees could be considered 
    an obstacle, the probe or 90 percent of the monitoring path must 
    meet the distance requirements of 2.3 and be at least 10 meters from 
    the drip line of the tree or trees. Since the scavenging effect of 
    trees is greater for O3 than for other criteria pollutants, 
    strong consideration of this effect must be given to locating an 
    O3 probe or monitoring path to avoid this problem.
        2.5  Spacing from Roadways (applicable to O3 and NO2 
    only). In siting an O3 analyzer, it is important to minimize 
    destructive interferences from sources of NO, since NO readily 
    reacts with O3. In siting NO2 analyzers for neighborhood 
    and urban scale monitoring, it is important to minimize 
    interferences from automotive sources. Table 1 provides the required 
    minimum separation distances between a roadway and a probe and 
    between a roadway and at least 90 percent of a monitoring path for 
    various ranges of daily roadway traffic. A sampling station having a 
    point analyzer probe located closer to a roadway than allowed by the 
    Table 1 requirements should be classified as middle scale rather 
    than neighborhood or urban scale, since the measurements from such a 
    station would more closely represent the middle scale. The 
    monitoring path of an open path analyzer must not cross over a 
    roadway with an average daily traffic count of 10,000 vehicles per 
    day or more. In calculating the percentage of a monitoring path over 
    or near a roadway, one must consider the entire segment of the 
    monitoring path in the area of potential atmospheric interference 
    from automobile emissions. Therefore, this calculation must include 
    the length of the monitoring path over the roadway plus any segments 
    of the monitoring path that lie in the area between the roadway and 
    the minimum separation distance, as determined from Table 1. The sum 
    of these distances must not be greater than 10 percent of the total 
    monitoring path length.
    
      Table 1.--Minimum Separation Distance Between Roadways and Probes or  
     Monitoring Paths for Monitoring Neighborhood--and Urban-Scale Ozone and
                                Nitrogen Dioxide                            
    ------------------------------------------------------------------------
                                                                   Minimum  
                                                                 separation 
          Roadway average daily traffic, vehicles per day       distance\1\,
                                                                   meters   
    ------------------------------------------------------------------------
    10,000.........................................           10 
    15,000....................................................           20 
    20,000....................................................           30 
    40,000....................................................           50 
    70,000....................................................          100 
    110,000........................................          250 
    ------------------------------------------------------------------------
    \1\Distance from the edge of the nearest traffic lane. The distance for 
      intermediate traffic counts should be interpolated from the table     
      values based on the actual traffic count.                             
    
        2.6  Cumulative Interferences on a Monitoring Path. The 
    cumulative length or portion of a monitoring path that is affected 
    by minor sources, obstructions, trees, or roadways must not exceed 
    10 percent of the total monitoring path length.
        2.7  Maximum Monitoring Path Length. The monitoring path length 
    must not exceed 1 kilometer for analyzers in neighborhood, urban, or 
    regional scale. For middle scale monitoring sites, the monitoring 
    path length must not exceed 300 meters. In areas subject to frequent 
    periods of dust, fog, rain, or snow, consideration should be given 
    to a shortened monitoring path length to minimize loss of monitoring 
    data due to these temporary optical obstructions. For certain 
    ambient air monitoring scenarios using open path analyzers, shorter 
    path lengths may be needed in order to ensure that the monitoring 
    station meets the objectives and spatial scales defined for SLAMS in 
    Appendix D. Therefore, the Regional Administrator or the Regional 
    Administrator's designee may require shorter path lengths, as needed 
    on an individual basis, to ensure that the SLAMS meet the Appendix D 
    requirements. Likewise, the Administrator or the Administrator's 
    designee may specify the maximum path length used at monitoring 
    stations designated as NAMS or PAMS as needed on an individual 
    basis.
    * * * * *
    
    4. Carbon Monoxide (CO)
    
        Open path analyzers may be used to measure CO for only middle or 
    neighborhood scale measurement applications if the open path 
    analyzer is designated as a SLAMS. Additional information on CO 
    monitor siting criteria may be found in reference 12.
        4.1  Horizontal and Vertical Placement. Because of the 
    importance of measuring population exposure to CO concentrations, 
    air should be sampled at average breathing heights. However, 
    practical factors require that the inlet probe be higher. The 
    required height of the inlet probe for CO monitoring is therefore 
    3\1/2\ meter for a microscale site, which is a 
    compromise between representative breathing height and prevention of 
    vandalism. The recommended 1 meter range of heights is also a 
    compromise to some extent. For consistency and comparability, it 
    would be desirable to have all inlets at exactly the same height, 
    but practical considerations often prevent this. Some reasonable 
    range must be specified and 1 meter provides adequate leeway to meet 
    most requirements.
        For the middle and neighborhood scale stations, the vertical 
    concentration gradients are not as great as for the microscale 
    station. This is because the diffusion from roads is greater and the 
    concentrations would represent larger areas than for the microscale. 
    Therefore, the probe or at least 80 percent of the monitoring path 
    must be located between 3 and 15 meters above ground level for 
    middle and neighborhood scale stations. The probe or at least 90 
    percent of the monitoring path must be at least 1 meter vertically 
    or horizontally away from any supporting structure, walls, parapets, 
    penthouses, etc., and away from dusty or dirty areas. If the probe 
    or a significant portion of the monitoring path is located near the 
    side of a building, then it should be located on the windward side 
    of the building relative to both the prevailing wind direction 
    during the season of highest concentration potential and the 
    location of sources of interest, i.e., roadways.
        4.2  Spacing from Obstructions. Buildings and other obstacles 
    may restrict airflow around a probe or monitoring path. To avoid 
    this interference, the probe or at least 90 percent of the 
    monitoring path must have unrestricted airflow and be located away 
    from obstacles so that the distance from the probe or monitoring 
    path is at least twice the height that the obstacle protrudes above 
    the probe or monitoring path. A probe or monitoring path located 
    near or along a vertical wall is undesirable because air moving 
    along the wall may be subject to possible removal mechanisms. A 
    probe must have unrestricted airflow in an arc of at least 270 
    degrees around the inlet probe, or 180 degrees if the probe is on 
    the side of a building. This arc must include the predominant wind 
    direction for the season of greatest pollutant concentration 
    potential. A monitoring path must be clear of all trees, brush, 
    buildings, plumes, dust, or other optical obstructions, including 
    potential obstructions that may move due to wind, human activity, 
    growth of vegetation, etc. Temporary optical obstructions, such as 
    rain, particles, fog, or snow, should be considered when siting an 
    open path analyzer. Any of these temporary obstructions that are of 
    sufficient density to obscure the light beam will affect the ability 
    of the open path analyzer to continuously measure pollutant 
    concentrations.
        4.3  Spacing from Roadways. Street canyon and traffic corridor 
    stations (microscale) are intended to provide a measurement of the 
    influence of the immediate source on the pollution exposure of the 
    population. In order to provide some reasonable consistency and 
    comparability in the air quality data from microscale stations, a 
    minimum distance of 2 meters and a maximum distance of 10 meters 
    from the edge of the nearest traffic lane must be maintained for 
    these CO monitoring inlet probes. This should give consistency to 
    the data, yet still allow flexibility of finding suitable locations.
        Street canyon/corridor (microscale) inlet probes must be located 
    at least 10 meters from an intersection and preferably at a midblock 
    location. Midblock locations are preferable to intersection 
    locations because intersections represent a much smaller portion of 
    downtown space than do the streets between them. Pedestrian exposure 
    is probably also greater in street canyon/corridors than at 
    intersections. Also, the practical difficulty of positioning 
    sampling inlets is less at midblock locations than at the 
    intersection. However, the final siting of the monitor must meet the 
    objectives and intent of appendix D, sections 2.4, 3, 3.3, and 
    appendix E, section 4.
        In determining the minimum separation between a neighborhood 
    scale monitoring station and a specific line source, the presumption 
    is made that measurements should not be unduly influenced by any one 
    roadway. Computations were made to determine the separation 
    distance, and table 2 provides the required minimum separation 
    distance between roadways and a probe or 90 percent of a monitoring 
    path. Probes or monitoring paths that are located closer to roads 
    than this criterion allows should not be classified as a 
    neighborhood scale, since the measurements from such a station would 
    closely represent the middle scale. Therefore, stations not meeting 
    this criterion should be classified as middle scale. 
    
      Table 2.--Minimum Separation Distance Between Roadways and Probes or  
       Monitoring Paths for Monitoring Neighborhood Scale Carbon Monoxide   
    ------------------------------------------------------------------------
                                                                   Minimum  
                                                                  separation
                                                                 distance\1\
                                                                  for probes
          Roadway average daily traffic, vehicles per day        or 90% of a
                                                                  monitoring
                                                                    path,   
                                                                   meters   
    ------------------------------------------------------------------------
    10,000..........................................          10 
    15,000.....................................................          25 
    20,000.....................................................          45 
    30,000.....................................................          80 
    40,000.....................................................         115 
    50,000.....................................................         135 
    60,000..........................................         150 
    ------------------------------------------------------------------------
    \1\Distance from the edge of the nearest traffic lane. The distance for 
      intermediate traffic counts should be interpolated from the table     
      values based on the actual traffic count.                             
    
        4.4 Spacing from Trees and Other Considerations. Since CO is 
    relatively nonreactive, the major factor concerning trees is as 
    obstructions to normal wind flow patterns. For middle and 
    neighborhood scale stations, trees should not be located between the 
    major sources of CO, usually vehicles on a heavily traveled road, 
    and the monitor. The probe or at least 90 percent of the monitoring 
    path must be 10 meters or more from the drip line of trees which are 
    between the probe or monitoring path and the road and which extend 
    at least 5 meters above the probe or monitoring path. For microscale 
    stations, no trees or shrubs should be located between the probe and 
    the roadway.
        4.5  Cumulative Interferences on a Monitoring Path. The 
    cumulative length or portion of a monitoring path that is affected 
    by obstructions, trees, or roadways must not exceed 10 percent of 
    the total monitoring path length.
        4.6  Maximum Monitoring Path Length. The monitoring path length 
    must not exceed 1 kilometer for analyzers used for neighborhood 
    scale monitoring applications, or 300 meters for middle scale 
    monitoring applications. In areas subject to frequent periods of 
    dust, fog, rain, or snow, consideration should be given to a 
    shortened monitoring path length to minimize loss of monitoring data 
    due to these temporary optical obstructions. For certain ambient air 
    monitoring scenarios using open path analyzers, shorter path lengths 
    may be needed in order to ensure that the monitoring station meets 
    the objectives and spatial scales defined for SLAMS in Appendix D. 
    Therefore, the Regional Administrator or the Regional 
    Administrator's designee may require shorter path lengths, as needed 
    on an individual basis, to ensure that the SLAMS meet the Appendix D 
    requirements. Likewise, the Administrator or the Administrator's 
    designee may specify the maximum path length used at monitoring 
    stations designated as NAMS or PAMS as needed on an individual 
    basis.
    * * * * *
    
    9. Probe Material and Pollutant Sample Residence Time
    
        For the reactive gases, SO2, NO2, and O3, special 
    probe material must be used for point analyzers. Studies20-24 
    have been conducted to determine the suitability of materials such 
    as polypropylene, polyethylene, polyvinylchloride, Tygon, aluminum, 
    brass, stainless steel, copper, pyrex glass and teflon for use as 
    intake sampling lines. Of the above materials, only pyrex glass and 
    teflon have been found to be acceptable for use as intake sampling 
    lines for all the reactive gaseous pollutants. Furthermore, EPA\25\ 
    has specified borosilicate glass or FEP teflon as the only 
    acceptable probe materials for delivering test atmospheres in the 
    determination of reference or equivalent methods. Therefore, 
    borosilicate glass, FEP teflon, or their equivalent must be used for 
    existing and new NAMS or SLAMS.
    * * * * *
    
    10. Photochemical Assessment Monitoring Stations (PAMS)
    
        10.1  Horizontal and Vertical Placement. The probe or at least 
    80 percent of the monitoring path must be located 3 to 15 meters 
    above ground level. This range provides a practical compromise for 
    finding suitable sites for the multi-pollutant PAMS. The probe or at 
    least 90 percent of the monitoring path must be at least 1 meter 
    vertically or horizontally away from any supporting structure, 
    walls, parapets, penthouses, etc., and away from dusty or dirty 
    areas.
        10.2  Spacing from Obstructions. The probe or at least 90 
    percent of the monitoring path must be located away from obstacles 
    and buildings such that the distance between the obstacles and the 
    probe or monitoring path is at least twice the height that the 
    obstacle protrudes above the probe or monitoring path. There must be 
    unrestricted airflow in an arc of at least 270 deg. around the probe 
    inlet. Additionally, the predominant wind direction for the period 
    of greatest pollutant concentration (as described for each site in 
    section 4.2 of Appendix D) must be included in the 270 deg. arc. If 
    the probe is located on the side of the building, 180 deg. clearance 
    is required. A monitoring path must be clear of all trees, brush, 
    buildings, plumes, dust, or other optical obstructions, including 
    potential obstructions that may move due to wind, human activity, 
    growth of vegetation, etc. Temporary optical obstructions, such as 
    rain, particles, fog, or snow, should be considered when siting an 
    open path analyzer. Any of these temporary obstructions that are of 
    sufficient density to obscure the light beam will affect the ability 
    of the open path analyzer to continuously measure pollutant 
    concentrations.
        10.3  Spacing from Roadways. It is important in the probe and 
    monitoring path siting process to minimize destructive interferences 
    from sources of NO since NO readily reacts with O3. Table 4 
    below provides the required minimum separation distances between 
    roadways and PAMS (excluding upper air measuring stations):
    
             Table 4.--Separation Distance Between PAMS and Roadways        
                         [Edge of Nearest Traffic Lane]                     
    ------------------------------------------------------------------------
                                                                    Minimum 
                                                                  separation
                                                                   distance 
                                                                    between 
          Roadway average daily traffic, vehicles per day          roadways 
                                                                      and   
                                                                   stations 
                                                                      in    
                                                                   meters\1\
    ------------------------------------------------------------------------
    <10,000.....................................................>10
    15,000......................................................          20
    20,000......................................................          30
    40,000......................................................          50
    70,000......................................................         100
    >110,000....................................................        250 
    ------------------------------------------------------------------------
    \1\Distance from the edge of the nearest traffic lane. The distance for 
      intermediate traffic counts should be interpolated from the table     
      based on the actual traffic flow.                                     
    
        10.4  Spacing from Trees. Trees can provide surfaces for 
    adsorption and/or reactions to occur and can obstruct normal wind 
    flow patterns. To minimize these effects at PAMS, the probe or at 
    least 90 percent of the monitoring path should be placed at least 20 
    meters from the drip line of trees. Since the scavenging effect of 
    trees is greater for O3 than for the other criteria pollutants, 
    strong consideration of this effect must be given in locating the 
    PAMS probe or monitoring path to avoid this problem. Therefore, the 
    probe or at least 90 percent of the monitoring path must be at least 
    10 meters from the drip line of trees that are located between the 
    urban city core area and the probe or monitoring path along the 
    appropriate wind direction.
    * * * * *
    
    12. Summary
    
        Table 5 presents a summary of the general requirements for probe 
    and monitoring path siting criteria with respect to distances and 
    heights. It is apparent from Table 5 that different elevation 
    distances above the ground are shown for the various pollutants. The 
    discussion in the text for each of the pollutants described reasons 
    for elevating the monitor, probe, or monitoring path. The 
    differences in the specified range of heights are based on the 
    vertical concentration gradients. For CO, the gradients in the 
    vertical direction are very large for the microscale, so a small 
    range of heights has been used. The upper limit of 15 meters was 
    specified for consistency between pollutants and to allow the use of 
    a single manifold or monitoring path for monitoring more than one 
    pollutant.
    
                                                 Table 5.--Summary of Probe and Monitoring Path Siting Criteria                                             
    --------------------------------------------------------------------------------------------------------------------------------------------------------
                                                                                      Horizontal and                                                        
                                                           Height from ground to  vertical distance from                                                    
                              Scale [maximum monitoring       probe or 80% of     supporting structuresB    Distance from trees to    Distance from roadways
           Pollutant            path length, meters]         monitoring pathA       to probe or 90% of    probe or 90% of monitoring  to probe or monitoring
                                                                 (meters)            monitoring pathA           pathA (meters)            pathA (meters)    
                                                                                         (meters)                                                           
    --------------------------------------------------------------------------------------------------------------------------------------------------------
    SO2 CDEF...............  Middle [300m] Neighborhood,  3-15..................  >1....................  >10.......................  N/A.                  
                              Urban, and Regional [1km].                                                                                                    
    CO DEG.................  Micro, Middle [300m],        30.5, 3-15  >1....................  >10.......................  2-10, See Table 1 for 
                              Neighborhood [1km].                                                                                      middle and           
                                                                                                                                       neighborhood scales. 
    O3CDE..................  Middle [300m],               3-15..................  >1....................  >10.......................  See Table 2 for all   
                              Neighborhood, Urban, and                                                                                 scales.              
                              Regional [1km].                                                                                                               
    Ozone precursors (for    Neighborhood and Urban [1    3-15..................  >1....................  >10.......................  See Table 4 for all   
     PAMS)CDE.                km].                                                                                                     scales.              
    NO2CDE.................  Middle [300m], Neighborhood  3-15..................  >1....................  >10.......................  See Table 2 for all   
                              and Urban [1km].                                                                                         scales.              
    PbCDEFH................  Micro; Middle,               2-7 (Micro), 2-15 (All  >2 (All scales,         >10 (All scales)..........  5-15 (Micro), See     
                              Neighborhood, Urban and      other scales).          horizontal distance                                 Table 3 for all other
                              Regional.                                            only).                                              scales.              
    PM-10CDEFH.............  Micro; Middle,               2-7 (Micro), 2-15 (All  >2 (All scales,         >10 (All scales)..........  2-10 (Micro), See     
                              Neighborhood, Urban and      other scales).          horizontal distance                                 Figure 2 for all     
                              Regional.                                            only).                                              other scales.        
    --------------------------------------------------------------------------------------------------------------------------------------------------------
    A Monitoring path for open path analyzers is applicable only to middle or neighborhood scale CO monitoring and all applicable scales for monitoring SO2,
      O3, O3 precursors, and NO2.                                                                                                                           
    B When probe is located on a rooftop, this separation distance is in reference to walls, parapets, or penthouses located on roof.                       
    N/A--Not applicable.                                                                                                                                    
    C Should be >20 meters from the dripline of tree(s) and must be 10 meters from the dripline when the tree(s) act as an obstruction.                     
    D Distance from sampler, probe, or 90% of monitoring path to obstacle, such as a building, must be at least twice the height the obstacle protrudes     
      above the sampler, probe, or monitoring path. Sites not meeting this criterion may be classified as middle scale (see text).                          
    E Must have unrestricted airflow 270 deg. around the probe or sampler; 180 deg. if the probe is on the side of a building.                              
    F The probe, sampler, or monitoring path should be away from minor sources, such as furnace or incineration flues. The separation distance is dependent 
      on the height of the minor source's emission point (such as a flue), the type of fuel or waste burned, and the quality of the fuel (sulfur, ash, or   
      lead content). This criterion is designed to avoid undue influences from minor sources.                                                               
    G For microscale CO monitoring sites, the probe must be >10 meters from a street intersection and preferably at a midblock location.                    
    H For collocated Pb and PM-10 samplers, a 2-4 meter separation distance between collocated samplers must be met.                                        
    
    * * * * *
    [FR Doc. 94-20042 Filed 8-17-94; 8:45 am]
    BILLING CODE 6560-50-P
    
    
    

Document Information

Published:
08/18/1994
Department:
Environmental Protection Agency
Entry Type:
Uncategorized Document
Action:
Proposed rule.
Document Number:
94-20042
Dates:
Comments must be received on or before September 19, 1994. Requests for public hearing must be received by September 2, 1994. If a hearing is held, comments must be received on or before 30 days from the conclusion of the hearing.
Pages:
0-0 (1 pages)
Docket Numbers:
Federal Register: August 18, 1994, FRL-4842-4
Supporting Documents:
» Legacy Index for Docket A-93-44
» Ambient Air Quality Surveillance Siting Criteria for Open Path Analyzers [A-93-44-V-A-1]
» Ambient Air Quality Surveillance Siting Criteria for Open Path Analyzers
CFR: (2)
40 CFR 5.2
40 CFR 58.1