05-16914. Medical Devices; Immunology and Microbiology Devices; Classification of Ribonucleic Acid Preanalytical Systems  

  • Start Preamble

    AGENCY:

    Food and Drug Administration, HHS.

    ACTION:

    Final rule.

    SUMMARY:

    The Food and Drug Administration (FDA) is classifying ribonucleic acid (RNA) preanalytical systems into class II (special controls). The special control that will apply to the device is the guidance document entitled “Class II Special Controls Guidance Document: RNA Preanalytical Systems (RNA Collection, Stabilization, and Purification Systems for RT-PCR Used in Molecular Diagnostic Testing).” The agency is classifying the device into class II (special controls) in order to provide a reasonable assurance of safety and effectiveness of the device. Elsewhere in this issue of the Federal Register, FDA is announcing the availability of a guidance document that will serve as the special control for the device.

    DATES:

    This rule is effective September 26, 2005. The classification was effective April 18, 2005.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    Uwe Scherf, Center for Devices and Radiological Health (HFZ-440), Food and Drug Administration, 2098 Gaither Rd., Rockville, MD 20850, 240-276-0496.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    I. What is the Background of this Rulemaking?

    In accordance with section 513(f)(1) of the Federal Food, Drug, and Cosmetic Act (the act) (21 U.S.C. 360c(f)(1)), devices that were not in commercial distribution before May 28, 1976, the date of enactment of the Medical Device Amendments of 1976 (the amendments), generally referred to as postamendments devices, are classified automatically by statute into class III without any FDA rulemaking process. These devices remain in class III and require premarket approval, unless and until the device is classified or reclassified into class I or II, or FDA issues an order finding the device to be substantially equivalent, in accordance with section 513(i) of the act, to a predicate device that does not require premarket approval.

    The agency determines whether new devices are substantially equivalent to previous marketed devices by means of premarket notification procedures in section 510(k) of the act (21 U.S.C. 360(k)) and 21 CFR part 807 of FDA's regulations.

    Section 513(f)(2) of the act provides that any person who submits a premarket notification under section 510(k) of the act for a device that has not previously been classified may, within 30 days after receiving an order classifying the device in class III under section 513(f)(1) of the act, request FDA to classify the device under the criteria set forth in section 513(a)(1) of the act. FDA shall, within 60 days of receiving such a request, classify the device by written order. This classification shall be the initial classification of the device. Within 30 days after the issuance of an order classifying the device, FDA must publish a notice in the Federal Register announcing such classification (section 513(f)(2) of the act).

    In accordance with section 513(f)(1) of the act, FDA issued an order on February 18, 2005, classifying the PAXgeneTM Blood RNA System into class III, because it was not substantially equivalent to a device that was introduced or delivered for introduction into interstate commerce for commercial distribution before May 28, 1976, or a device which was subsequently reclassified into class I or class II. On February 28, 2005, PreAnalytiX GmbH, c/o Becton, Dickinson and Co., submitted a petition requesting classification of the PAXgeneTM Blood RNA System under section 513(f)(2) of the act. The manufacturer recommended that the device be classified into class II.

    In accordance with 513(f)(2) of the act, FDA reviewed the petition in order to classify the device under the criteria for classification set forth in 513(a)(1) of the act. Devices are to be classified into class II if general controls, by themselves, are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls to provide reasonable assurance of the safety and effectiveness of the device for its intended use. After review of the information submitted in the petition, FDA determined that the PAXgeneTM Blood RNA System can be classified into class II with the establishment of special controls. FDA believes these special controls will provide reasonable assurance of the safety and effectiveness of the device.

    The device is assigned the generic name RNA Preanalytical Systems and it is identified as a device intended to collect, store, and transport patient specimens, and stabilize intracellular RNA from the specimens, for subsequent isolation and purification of the intracellular RNA for reverse transcriptase polymerase chain reaction (RT-PCR) used in in vitro molecular diagnostic testing. The device may consist of sample collection devices, nucleic acid isolation and purification reagents, and processing reagents/equipment (tubes, columns, etc.). It also may contain instruments for automation of the nucleic acid isolation and purification steps.

    FDA has identified the following risks to health associated specifically with this type of device: (1) Inaccurate results and improper patient management, (2) delay in diagnosis, and (3) a need for patient specimen recollection.

    Failure of the system during specimen collection, or during RNA stabilization or purification could yield an RNA sample of low quality and quantity. Low quality RNA, when tested, could result in falsely low or falsely high RNA transcript signal levels leading to inaccurate diagnosis and/or improper patient management. Low quantity of RNA could render the samples unusable for downstream RT-PCR applications; specimens would need to be recollected, causing possible delay in diagnosis. In addition, depending on specimen type, recollection could pose additional patient risk (e.g., tissue biopsy). The degree of risk varies depending on the disease or condition/stage being diagnosed or managed. Results of RNA testing should always be considered in conjunction with other clinical factors.

    Start Printed Page 49863

    FDA believes that the class II special controls guidance document aids in mitigating the potential risks to health by providing recommendations on validation of performance characteristics, including RNA stability, purity, integrity, yield, repeatability, reproducibility, and suitability for use in RT-PCR assays. The guidance document also provides information on how to meet premarket (510(k)) submission requirements for the device. FDA believes that the special controls guidance document, in addition to general controls, addresses the risks to health identified previously and provides reasonable assurance of the safety and effectiveness of the device. Therefore, on April 18, 2005, FDA issued an order to the petitioner classifying the device into class II. FDA is codifying this device by adding § 866.4070.

    Following the effective date of this final classification rule, any firm submitting a 510(k) premarket notification for an RNA preanalytical system will need to address the issues covered in the special controls guidance. However, the firm need only show that its device meets the recommendations of the guidance, or in some other way provides equivalent assurance of safety and effectiveness.

    Section 510(m) of the act provides that FDA may exempt a class II device from the premarket notification requirements under 510(k) of the act, if FDA determines that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device. For this type of device, FDA has determined that premarket notification is necessary to provide reasonable assurance of the safety and effectiveness of the device and, therefore, the type of device is not exempt from premarket notification requirements. Persons who intend to market this type of device must submit to FDA a premarket notification, prior to marketing the device, which contains information about the RNA Preanalytical Systems they intend to market.

    II. What is the Environmental Impact of This Rule?

    The agency has determined under 21 CFR 25.34(b) that this action is of type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

    III. What is the Economic Impact of This Rule?

    FDA has examined the impacts of the final rule under Executive Order 12866 and the Regulatory Flexibility Act (5 U.S.C. 601-612), and the Unfunded Mandates Reform Act of 1995 (Public Law 104-4). Executive Order 12866 directs agencies to assess all costs and benefits of available regulatory alternatives and, when regulation is necessary, to select regulatory approaches that maximize net benefits (including potential economic, environmental, public health and safety, and other advantages; distributive impacts; and equity). The agency believes that this final rule is not a significant regulatory action under the Executive order.

    The Regulatory Flexibility Act requires agencies to analyze regulatory options that would minimize any significant impact of a rule on small entities. Because classification of this device into class II will relieve manufacturers of the device of the cost of complying with the premarket approval requirements of section 515 of the act (21 U.S.C. 360e), and may permit small potential competitors to enter the marketplace by lowering their costs, the agency certifies that the final rule will not have a significant economic impact on a substantial number of small entities.

    Section 202(a) of the Unfunded Mandates Reform Act of 1995 requires that agencies prepare a written statement, which includes an assessment of anticipated costs and benefits, before proposing “any rule that includes any Federal mandate that may result in the expenditure by State, local, and tribal governments, in the aggregate, or by the private sector, of $100,000,000 or more (adjusted annually for inflation) in any one year.” The current threshold after adjustment for inflation is $115 million, using the most current (2003) Implicit Price Deflator for the Gross Domestic Product. FDA does not expect this final rule to result in any 1-year expenditure that would meet or exceed this amount.

    IV. Federalism

    FDA has analyzed this final rule in accordance with the principles set forth in Executive Order 13132. FDA has determined that the rule does not contain policies that have substantial direct effects on the States, on the relationship between the National Government and the States, or on the distribution of power and responsibilities among the various levels of government. Accordingly, the agency has concluded that the rule does not contain policies that have federalism implications as defined in the Executive order and, consequently, a federalism summary impact statement is not required.

    V. How Does This Rule Comply With the Paperwork Reduction Act of 1995?

    This final rule contains no collections of information. Therefore, clearance by the Office of Management and Budget under the Paperwork Reduction Act of 1995 is not required.

    VI. What References Are on Display?

    The following reference has been placed on display in the Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852, and may be seen by interested persons between 9 a.m. and 4 p.m., Monday through Friday.

    1. Petition from PreAnalytiX GmbH, c/o Becton, Dickinson and Co., dated February 28, 2005.

    Start List of Subjects

    List of Subjects in 21 CFR Part 866

    • Medical devices
    End List of Subjects Start Amendment Part

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 866 is amended as follows:

    End Amendment Part Start Part

    PART 866—IMMUNOLOGY AND MICROBIOLOGY DEVICES

    End Part Start Amendment Part

    1. The authority citation for 21 CFR part 866 continues to read as follows:

    End Amendment Part Start Authority

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

    End Authority Start Amendment Part

    2. Section 866.4070 is added to subpart E to read as follows:

    End Amendment Part
    RNA Preanalytical Systems.

    (a) Identification. RNA Preanalytical Systems are devices intended to collect, store, and transport patient specimens, and stabilize intracellular RNA from the specimens, for subsequent isolation and purification of the intracellular RNA for RT-PCR used in in vitro molecular diagnostic testing.

    (b) Classification. Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: RNA Preanalytical Systems (RNA Collection, Stabilization and Purification System for RT-PCR Used in Molecular Diagnostic Testing).” See § 866.1(e) for the availability of this guidance document.

    Start Signature
    Start Printed Page 49864

    Dated: August 9, 2005.

    Linda S. Kahan,

    Deputy Director, Center for Devices and Radiological Health.

    End Signature End Supplemental Information

    [FR Doc. 05-16914 Filed 8-24-05; 8:45 am]

    BILLING CODE 4160-01-S

Document Information

Effective Date:
9/26/2005
Published:
08/25/2005
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final rule.
Document Number:
05-16914
Dates:
This rule is effective September 26, 2005. The classification was effective April 18, 2005.
Pages:
49862-49864 (3 pages)
Docket Numbers:
Docket No. 2005N-0263
PDF File:
05-16914.pdf
CFR: (1)
21 CFR 866.4070