96-21846. Medical Devices; Reclassification of the Infant Radiant Warmer  

  • [Federal Register Volume 61, Number 167 (Tuesday, August 27, 1996)]
    [Proposed Rules]
    [Pages 44013-44019]
    From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
    [FR Doc No: 96-21846]
    
    
    -----------------------------------------------------------------------
    
    
    DEPARTMENT OF HEALTH AND HUMAN SERVICES
    21 CFR Part 880
    
    [Docket No. 85N-0285]
    
    
    Medical Devices; Reclassification of the Infant Radiant Warmer
    
    AGENCY: Food and Drug Administration, HHS.
    
    ACTION: Proposed rule.
    
    -----------------------------------------------------------------------
    
    SUMMARY: The Food and Drug Administration (FDA) is proposing to 
    reclassify the infant radiant warmer from class III (premarket 
    approval) into class II (special controls) based on new information 
    regarding the device. The infant radiant warmer is a device consisting 
    of an infrared heating element intended to maintain the infant's body 
    temperature by means of radiant heat. This document summarizes the 
    basis for the agency's findings that sufficient valid scientific 
    evidence is available to support reclassification of the infant radiant 
    warmer and to establish special controls to provide reasonable 
    assurance of the safety and effectiveness of the device. This action 
    implements the Medical Device Amendments of 1976 (the amendments) as 
    amended by the Safe Medical Devices Act of 1990 (the SMDA).
    
    DATES: Written comments by November 25, 1996. FDA proposes that any 
    final rule based on this proposal become final 30 days after 
    publication in the Federal Register.
    
    ADDRESSES: Submit written comments to the Dockets Management Branch 
    (HFA-305), Food and Drug Administration, 12420 Parklawn Dr., rm. 1-23, 
    Rockville, MD 20857.
    
    FOR FURTHER INFORMATION CONTACT: Janet L. Scudiero, Center for Devices 
    and Radiological Health (HFZ-410), Food and Drug Administration, 9200 
    Corporate Blvd., Rockville, MD 20850, 301-594-1287.
    
    SUPPLEMENTARY INFORMATION:
    
    Table of Contents
    
    I. Classification and Reclassification of Devices Under the Medical 
    Device Amendments of 1976
    II. Reclassification Under the Safe Medical Devices Act of 1990
    III. History of the Proceedings
    IV. Device Description
    V. Recommendation of the Panel
    VI. Summary of the Reasons for the Recommendation
    VII. Risks to Health
    VIII. Summary of Data Upon Which the Recommendation is Based
    IX. FDA's Tentative Findings
    X. Environmental Impact
    XI. Analysis of Impacts
    XII. Paperwork Reduction Act of 1995
    XIII. Request for Comments
    XIV. References
    
    I. Classification and Reclassification of Devices Under the Medical 
    Device Amendments of 1976
    
        Under section 513 of the Federal Food, Drug, and Cosmetic Act (the 
    act) (21 U.S.C. 360c), as established by the amendments (Pub. L. 94-
    295) and amended by the SMDA (Pub. L. 101-629), FDA must classify 
    devices into one of three regulatory classes: Class I, class II, or 
    class III. FDA's classification of a device is determined by the amount 
    of regulation necessary to provide reasonable assurance of safety and 
    effectiveness of a device. Except as provided in section 520(c) of the 
    act (21 U.S.C. 360j(c)), FDA may not use confidential information 
    concerning a device's safety and effectiveness as a basis for 
    reclassification of the device from class III into class II or class I.
        Under the original 1976 act, devices were to be classified into 
    class I (general controls) if there was information showing that the 
    general controls of the act were sufficient to assure safety and 
    effectiveness; into class II (performance
    
    [[Page 44014]]
    
    standards) if there was insufficient information showing that general 
    controls themselves would ensure safety and effectiveness, but there 
    was sufficient information to establish a performance standard that 
    would provide such assurance; and into class III (premarket approval) 
    if there was insufficient information to support classifying a device 
    into class I or class II and the device was a life-sustaining or life-
    supporting device or was for a use that is of substantial importance in 
    preventing impairment of human health.
        Most generic types of devices that were on the market before the 
    date of the original 1976 amendments (May 28, 1976) (generally referred 
    to as preamendments devices) have been classified by FDA under the 
    procedures set forth in section 513(c) and (d) of the act through the 
    issuance of classification regulations into one of these three 
    regulatory classes. Under sections 513(c) and (d) of the act, FDA 
    secures expert panel recommendations on the appropriate device 
    classifications for generic types of devices. FDA then considers the 
    panel's recommendations and, through notice and comment rulemaking, 
    issues classification regulations.
        For those devices introduced into interstate commerce for the first 
    time after May 28, 1976, the device is classified through the premarket 
    notification process under section 510(k) of the act (21 U.S.C. 
    360(k)). Those devices that FDA finds to be substantially equivalent to 
    a classified preamendments generic type of device are thereby 
    classified in the same class as the predicate preamendments device.
        Reclassification of classified preamendments devices is governed by 
    section 513(e) of the act. This section provides that FDA may, by 
    rulemaking, reclassify a device (in a proceeding that parallels the 
    initial classification proceeding) based on ``new information.'' The 
    reclassification can be initiated by FDA or by the petition of an 
    interested person.
        The term ``new information,'' as used in section 513(e) of the act, 
    includes information developed as a result of a reevaluation of the 
    data before the agency when a device was originally classified, as well 
    as information not presented, not available, or not developed at that 
    time. (See, e.g., Holland Rantos v. United States Department of Health, 
    Education, and Welfare, 587 F.2d 1173, 1174 n.1 (D.C. Cir. 1978); 
    Upjohn v. Finch, 422 F.2d 944 (6th Cir. 1970); Bell v. Goddard, 366 
    F.2d 177 (7th Cir. 1966).)
        Reevaluation of the data previously before the agency is an 
    appropriate basis for subsequent regulatory action where the 
    reevaluation is made in light of changes in ``medical science.'' (See 
    Upjohn v. Finch, supra, 422 F.2d at 951.) However, regardless of 
    whether data before the agency are past or new data, the ``new 
    information'' on which any reclassification is based is required to 
    consist of ``valid scientific evidence,'' as defined in section 
    513(a)(3) of the act and 21 CFR 860.7(c)(2). FDA relies upon ``valid 
    scientific evidence'' in the classification process to determine the 
    level of regulation for devices. For the purpose of reclassification, 
    the valid scientific evidence upon which the agency relies must be 
    publicly available. Publicly available information excludes trade 
    secret and/or confidential commercial information, e.g., the contents 
    of premarket approval applications (PMA's). (See section 520(c) of the 
    act, (21 U.S.C. 360j(c).)
    
    II. Reclassification Under the Safe Medical Devices Act of 1990
    
        The SMDA further amended the act to change the definition of a 
    class II device. Under the SMDA, class II devices are those devices for 
    which there is insufficient information to show that general controls 
    themselves will ensure safety and effectiveness, but there is 
    sufficient information to establish special controls to provide such 
    assurance, including the issuance of a performance standard, postmarket 
    surveillance, patient registries, development and dissemination of 
    guidelines, and other appropriate actions necessary to provide 
    reasonable assurance of the safety and effectiveness of the device. 
    Thus, the definition of a class II device was changed from 
    ``performance standards'' to ``special controls.''
    
    III. History of the Proceedings
    
        In the Federal Register of August 24, 1979 (44 FR 49873), FDA 
    published a proposed rule to classify the infant radiant warmer into 
    class III. The preamble included the classification recommendation of 
    the General Hospital and Personal Use Devices Panel (the panel). The 
    panel's recommendation included a summary of the reasons why the device 
    should be subject to premarket approval and identified certain risks to 
    health presented by the device, including electrical shock, possible 
    eye damage due to long-term exposure to infrared radiation, patient 
    injury, hospital staff burns, insensible water loss, and hyperthermia 
    or hypothermia. The panel also recommended that a high priority for the 
    application of section 515(b) of the act (21 U.S.C. 360e)(premarket 
    approval requirement) be assigned to the infant radiant warmer.
        In the Federal Register of October 21, 1980 (45 FR 69694), FDA 
    published a final rule classifying the infant radiant warmer into class 
    III (21 CFR 880.5130). Concern for possible long-term effects of 
    infrared radiation on the skin and eyes of infants was the sole reason 
    for classifying the device into class III. FDA believed that the other 
    risks to health identified in the proposed rule could be addressed by 
    labeling or by a standard.
        In the Federal Register of September 6, 1983 (48 FR 40272), FDA 
    published a notice of intent to initiate proceedings to require 
    premarket approval of 13 preamendments class III devices assigned a 
    high priority by FDA for the application of premarket approval 
    requirements. Among other things, the notice described the factors FDA 
    considered in establishing priorities for initiating proceedings under 
    section 515(b) of the act for issuing final rules requiring 
    preamendments class III devices to have approved PMA's or product 
    development protocols (PDP's) which have been declared completed. Using 
    these factors, FDA concurred with the panel's recommendation that the 
    infant radiant warmer should be subject to a high priority for 
    initiating a proceeding to require premarket approval.
        In the Federal Register of January 15, 1986 (51 FR 1910), FDA 
    published a proposed rule to require filing of a PMA or a notice of 
    completion of a PDP for the infant radiant warmer. In accordance with 
    section 515(b) of the act and 21 CFR 860.132, FDA also announced an 
    opportunity for interested persons to request a change in 
    classification of the device based on new information. FDA identified 
    the following potential risks to health associated with the use of 
    infant radiant warmers: Insensible water loss, special risk group 
    infants with very low birth weight, hypothermia and hyperthermia, 
    damage to the eyes and skin, increased oxygen consumption, operator 
    error, and other safety risks common to many devices (e.g., electric 
    shock, inadequate stability, and burns to the user).
        On January 30, 1986, the Health Industries Manufacturers 
    Association submitted a petition (Ref. 1) to reclassify the infant 
    radiant warmer from class III into class II. The petition was submitted 
    under section 513(e) of the act. Consistent with the act and the 
    regulations, FDA referred the petition to the panel for its 
    recommendation on the requested change in classification.
    
    [[Page 44015]]
    
        On May 21, 1986, during a meeting by teleconference, the panel 
    unanimously recommended that the infant radiant warmer be reclassified 
    from class III into class II and that any change in classification not 
    take effect until the effective date of a performance standard for the 
    generic type of device established under section 514 of the act (21 
    U.S.C. 360d) (Ref. 2 at p. 75).
        In the Federal Register of May 27, 1987 (52 FR 19735), FDA 
    published a notice of intent to initiate a proceeding to reclassify the 
    infant radiant warmer from class III into class II. Subsequent to that 
    notice, FDA determined that the deliberations of the 1986 panel were 
    incomplete and that another panel meeting was necessary to allow the 
    panel to address specific recommendations and issues concerning the 
    reclassification of the infant radiant warmer (Ref. 2 at pp. 54 and 
    65). This additional panel meeting was held on May 11, 1994. A summary 
    of the panel's recommendation is set forth below.
    
    IV. Device Description
    
        FDA is proposing the following device description based on the 
    panel's recommendation and the agency's review.
        The infant radiant warmer is a device consisting of an infrared 
    heating element intended to be placed over an infant to maintain the 
    infant's body temperature by means of radiant heat. The device may also 
    contain a temperature monitoring sensor, a heat output control 
    mechanism, and an alarm system (infant temperature, manual mode if 
    present, and failure alarms) to alert operators of a temperature 
    condition over or under the set temperature, manual mode time limits, 
    and device component failure, respectively. The device may be placed 
    over a pediatric hospital bed or it may be built into the bed as a 
    complete unit.
    
    V. Recommendation of the Panel
    
        In the public meeting held on May 11, 1994, the panel unanimously 
    affirmed its previous recommendation that the infant radiant warmer 
    should be reclassified from class III into class II (Ref. 3), and that 
    the appropriate special control is a voluntary standard. The panel 
    identified the Association for the Advancement of Medical 
    Instrumentation (AAMI) voluntary standard for infant radiant warmers as 
    the special control for the infant radiant warmer (Ref. 4).
        The panel further recommended the following restrictions on the use 
    of the device: A prescription statement in the labeling of the device 
    that restricts the device to use only upon the order of a physician, 
    only in health care facilities, and only by persons with specific 
    training and experience in the use of the device.
    
    VI. Summary of the Reasons for the Recommendation
    
        The panel gave the following reasons in support of its 
    recommendation to reclassify the infant radiant warmer from class III 
    into class II:
        1. General controls by themselves are insufficient to provide 
    reasonable assurances of the safety and effectiveness of the device.
        2. There is sufficient publicly available information to establish 
    special controls to provide reasonable assurance of the safety and 
    effectiveness of the device for its intended use.
        3. An existing voluntary standard (Ref. 4) is the special control 
    recommended by the panel.
        4. There is sufficient publicly available information to 
    demonstrate that the device is not potentially hazardous to the life, 
    health, or well-being of the infant. The panel identified no new risks 
    to health associated with the use of the device and determined that 
    some of the previously identified potential risks to health are no 
    longer risks or are no longer serious risks (Ref. 3 at p. 225). Thus, 
    the probable benefits to health of the device outweigh any probable 
    risks to health.
        The panel believes that the current and any subsequent 
    manufacturers of the infant radiant warmer can comply with this 
    voluntary standard, that FDA can ensure the safety and effectiveness of 
    the device made by new manufacturers through the premarket notification 
    procedures under section 510(k) of the act, and that a regulatory level 
    of class III is unnecessary.
    
    VII. Risks to Health
    
        When the infant radiant warmer was proposed for classification into 
    class III in 1979, the panel identified certain risks to health that 
    they believed the device presented. The risks to health were identified 
    as electrical shock, possible eye damage, patient injury, hospital 
    staff burns, insensible water loss, and hyperthermia or hypothermia (44 
    FR 49873 at 49874). When the device was classified into class III in 
    1980, FDA identified concern for possible delayed long-term effects of 
    infrared radiation on the skin and eyes of infants as the only risk to 
    health presented by the device. FDA also determined that the other 
    risks to health identified in the proposed rule could be addressed by 
    labeling or by a standard (45 FR 69694). Subsequently, in 1986, the 
    agency identified increased oxygen consumption as another potential 
    risk to health associated with the use of the device (51 FR 1910).
        Based on the review of the new data and information contained in 
    the petition and the panel members' personal knowledge of and 
    experience with the device, the panel on May 11, 1994, agreed that all 
    the potential risks to health (insensible water loss; special risk 
    group, very low birth weight infants; hyperthermia and hypothermia; 
    possible eye and skin damage; and increased oxygen consumption) 
    associated with the use of the infant radiant warmer could be 
    controlled by special controls (Ref. 3). The panel also believed that 
    the general risks to health (operator error, electric shock, inadequate 
    device stability, and burns to operators) could also be addressed by 
    special controls.
        On the basis of its review and the panel's recommendation, FDA now 
    believes that the use of the infant radiant warmer for maintaining an 
    infant's body temperature does not present a potential unreasonable 
    risk of illness and injury, and that special controls would provide 
    reasonable assurance of the safety and effectiveness of the device. In 
    addition to the AAMI standard, FDA has also incorporated the panel's 
    labeling recommendation as special controls for this device.
    
    VIII. Summary of the Data Upon Which the Proposed Recommendation is 
    Based
    
    A. Insensible Water Loss
    
        An increased rate of insensible water loss is the principle, well-
    documented risk to health associated with the use of infant radiant 
    warmers (Refs. 5 and 6). Insensible water loss is the continuous and 
    usually imperceptible loss of water, mainly from the skin, that occurs 
    to some extent in all newborn infants. It is a well recognized 
    condition of prematurity, its severity being inversely related to birth 
    weight (Ref. 7). Other factors that contribute to insensible water loss 
    in neonates include: Illness; environmental temperature and humidity; 
    and other therapies, especially phototherapy and respiratory support 
    (Ref. 5). Insensible water loss is also associated with the use of 
    incubators (Refs. 5 through 7).
        Bell (Ref. 6) evaluated four studies (Refs. 8 through 11), which 
    reported increased rates of insensible water loss of 40 to 190 percent 
    during the use of radiant warmers compared to the use of incubators. He 
    determined that the variations in the increased rates of insensible 
    water loss are related to the experimental conditions of the 
    investigations (mainly the different
    
    [[Page 44016]]
    
    weighing methods used in the studies). Bell concluded that insensible 
    water loss in infants under infant radiant warmers without phototherapy 
    is 40 to 100 percent higher than in infants in incubators.
        Increased insensible water loss places an infant at a risk of 
    dehydration and electrolyte imbalance and potentially interferes with 
    the infant's thermoregulation. Because both underestimation and 
    overestimation of fluid and electrolyte requirements can have serious 
    consequences to infants, especially to low birth weight infants, 
    guidance for parenteral fluid and electrolyte administration was 
    needed. Since the infant radiant warmer was classified in 1980, several 
    guidances which include recommendations for parenteral fluid and 
    electrolyte administration have been developed for premature and term 
    infants (Refs. 6, 12, and 13).
        The use of plastic heat shielding with infant radiant warmers has 
    been reported to reduce insensible water loss (Refs. 14 through 17). 
    However, this practice is not without risks, including both 
    underheating and overheating of infants (Refs. 2 and 18). The panel 
    agreed that the use of heat shielding should be at the discretion of 
    the informed physician (Ref. 2).
        Although an increased rate of insensible water loss is a risk to 
    health in the use of the infant radiant warmer, it can be managed by 
    careful monitoring of the infant and administration of parenteral or 
    oral electrolyte therapy when necessary. The new parenteral fluid and 
    electrolyte therapy guidances minimize this risk to health and support 
    the use of infant radiant warmers in the management of critically ill 
    infants to whom continual access by health professionals is essential.
        The panel believed that this risk to health is a well-understood 
    risk associated with the use of the infant radiant warmer and that it 
    is related to both the prematurity of the infant and the open bed 
    design of the device (Ref. 3). The panel agreed that this risk to 
    health is clinically manageable and that it could be controlled by 
    special controls.
    
    B. Special Risk Group--Very Low Birth Weight Infants
    
        To survive, very low birth weight infants, weighing 1,500 grams or 
    less, require aggressive diagnostic and therapeutic procedures, such as 
    emergency resuscitation, tracheal intubation, placement of catheters 
    and needles, and blood sampling (Ref. 1). The use of infant radiant 
    warmers has allowed essential access to the infants for the performance 
    of these necessary procedures while providing effective warming. This 
    is particularly important immediately after birth, during the first 
    days of life, and for the care of critically ill premature infants.
        Very low birth weight infants are especially susceptible to 
    increased rates of insensible water loss because of their larger 
    surface area to mass ratio, higher body water content, and the thinner 
    epidermal barrier of their skin (Refs. 2 (at pp. 56 and 57), 5, and 
    13). The advances in parenteral fluid and electrolyte therapy since 
    1980 provide specific guidance to minimize this risk for very low birth 
    weight infants (Refs. 6, 12, and 13).
        The panel believed that this potential risk to health is not a risk 
    related to the device, but that it is related to the prematurity of the 
    infants (Ref. 3). The panel stated that the use of the infant radiant 
    warmer has made the care of these infants more manageable, and the 
    panel commented that now even smaller premature infants than in 1986 
    are successfully treated in infant radiant warmers. The panel believed 
    that this risk can be controlled through special controls.
    
    C. Damage to the Eyes
    
        Infant radiant warmers operate by directing invisible infrared 
    radiation (IR) from an overhead heater to the infant's body. The 
    magnitude and spectral characteristics of the IR are controlled by the 
    design of the device and are important in assessing the potential risk 
    of exposure to IR.
        During its classification deliberations in 1979, the panel 
    considered infant radiant warmer performance data developed for FDA 
    under a contract (Ref. 19). However, that data did not sufficiently 
    address the panel's concern about the possibility of adverse effects on 
    the eyes of infants resulting from long-term exposure to IR. The 
    petition reported new performance data on five radiant warmers (Ref. 
    1). The new data provided measurements for individual wavelength 
    regions of the electromagnetic spectrum, including the ultraviolet (200 
    to 400 nanometers (nm)), visible (400 to 760 nm), and IR-A (760 to 
    1,400 nm) wavelength regions, and for the 1,400 to 4,500 nm wavelength 
    region which includes the IR-B (1,400 to 3,000 nm) wavelength region 
    and the 3,000 to 4,500 nm portion of the IR-C wavelength region (the 
    IR-C wavelength region extends from 3,000 to 100,000 nm). The petition 
    also reported total irradiance, including irradiance for wavelengths 
    extending beyond 4500 nm obtained by another measurement method. The 
    IR-A wavelength region is associated with the potential for damage to 
    the lens and retina of the eye. The IR-B and IR-C wavelength regions 
    are associated with the potential for thermal damage to the cornea of 
    the eye.
        All the infant radiant warmers emitted IR primarily in the IR-B and 
    IR-C wavelength regions (Ref. 1). No ultraviolet radiation and 
    negligible visible radiation (nondetectable to 0.026 milliwatt per 
    square centimeter (mW/cm2)) was detected. The range of maximum IR-
    A irradiance was 0.103 to 3.463 mW/cm2, and the range of maximum 
    total irradiance was 39.2 to 60.3 mW/cm2. These maximum 
    irradiances were obtained at full power and at high line voltage (130 
    volts). At lower heater power levels, proportionately more of the IR is 
    from the IR-C wavelength region.
        In clinical use, however, infant radiant warmers are rarely 
    operated at full power and at high line voltage (Ref. 1). The total 
    irradiances necessary to maintain the desired infant skin temperature 
    typically range from 12 to 25 mW/cm2, and typical IR-A irradiances 
    are less than 1.0 mW/cm2. Engel et al. reported mean total 
    irradiances of less than 10 mW/cm2 and 17.1 mW/cm2 for the 
    warming of two groups of critically ill premature infants (Refs. 20 and 
    21); in general, the smaller infants required higher irradiances. In 
    addition, the necessarily more frequent handling of critically ill 
    neonates, which may be as often as once every 10 minutes, may interrupt 
    delivery of a portion of the radiant heat to the infant and thus 
    increase the amount of radiant power required for heating (Ref. 2).
        The petition also summarized published information that was not 
    reviewed by the classification panel when the infant radiant warmer was 
    classified. Both Sliney and Freasier (Ref. 22) and Sliney and Wolbarsht 
    (Ref. 23) reported that a safe chronic ocular exposure level to IR-A 
    was 10 mW/cm2. The petition reported that the maximum amount of 
    IR-A of the tested infant radiant warmers ranged from 0.24 to 3.5 mW/
    cm2, and that in actual use, infant radiant warmers emit typically 
    less than 1 mW/cm2 of IR-A (Ref. 1). Thus, the potentials for 
    chronic injury to the lens and the retina are low because infant 
    radiant warmers emit significantly less IR-A radiation than the level 
    of IR-A radiation believed to be associated with injuries of the lens 
    and retina.
        The cornea and aqueous humor absorb almost all of the IR from 1,400 
    to 1,900 nm; the cornea absorbs all the IR above 1,900 nm (Ref. 23). 
    Thus, most IR emitted by infant radiant warmers is absorbed by the 
    anterior structures of
    
    [[Page 44017]]
    
    the eye and is not transmitted to the lens and retina. Sliney and 
    Freasier (Ref. 22) and Sliney and Wolbarsht (Ref. 23) also reported 
    that the irradiance of 100 mW/cm2 was ``well below'' the threshold 
    irradiance level to prevent corneal injury. Thus, the potentials for 
    injury to the cornea and aqueous humor from exposure to IR emitted by 
    infant radiant warmers are low because the maximum irradiances of 
    infant radiant warmers range from 36.8 to 60.3 mW/cm2 and their 
    typical total use irradiances range from 12 to 25 mW/cm2 (Ref. 1). 
    For both the total irradiance and the IR-A irradiance, the margins for 
    safety are significant.
        To put this irradiance information in perspective, it should be 
    noted that premature infants' eyes are rarely opened and that blinking 
    of the eyes when opened keeps the corneal epithelium from drying out 
    (Ref. 24). Thus, there is a low probability that a significant amount 
    of IR actually enters the eyes of premature infants.
        There are two studies on the effects of IR on the eyes of neonates. 
    Johns et al. detected no adverse eye effects in infants warmed under 
    radiant warmers after followup times of up to 45 days (Ref. 25). This 
    study now has increased significance since Pitts and Cullen reported 
    that corneal damage heals rapidly (usually within 24 hours) and that 
    lens opacities formed within 24 hours after exposure heal earlier than 
    expected (usually within 1 month) (Ref. 26). Thus, any corneal or lens 
    effects, if present, would have been detected by Johns et al.
        In 1993, Baumgart et al. (Ref. 27) reported a retrospective study 
    of critically ill premature infants treated under radiant warmers and 
    incubators with longer followup times of 30 days to 6 years. The mean 
    followup time for the radiant warmer group was 29 months, and the mean 
    IR irradiance of the infant radiant warmer group was less than 30 mW/
    cm2. They found no long-term or short-term corneal or lens effects 
    in either group. The incidence of retinopathy of prematurity was higher 
    in the radiant warmer group, but this higher incidence was attributed 
    to prematurity and to the hospital's policy of placing the more 
    critically ill premature infants receiving oxygen in infant radiant 
    warmers rather than in incubators. It is noted that the incidence of 
    retinopathy of prematurity is associated with prolonged oxygen therapy 
    (Ref. 28).
        There are few recommended IR exposure levels specifically intended 
    for infants under infant radiant warmers. The Emergency Care Research 
    Institute proposed that 0.3 W/cm2 (300 mW/cm2) was a 
    reasonable total irradiance limit for an infant under an infant radiant 
    warmer in 1973 and 1984 (Refs. 24 and 18, respectively) and that the 
    near IR range between 700 to 1,200 nm should be limited to 40 mW/
    cm2. The 1994 International Electrotechnical Commission standard 
    for infant radiant warmers has irradiance limits of 100 mW/cm2 for 
    total IR irradiance and 10 mW/cm2 for IR-A (Ref. 29). The 1995 
    AAMI voluntary standard special control has irradiance limits of 60 mW/
    cm2 for total IR irradiance and 10 mW/cm2 for IR-A (Ref. 4). 
    The maximum irradiances of currently marketed infant radiant warmers 
    meet the AAMI voluntary standard special control irradiance limits 
    (Ref. 3).
        This new information concerning the IR irradiance characteristics 
    of infant radiant warmers and the irradiance levels associated with 
    acute and chronic injuries to the eyes have addressed the safety 
    concerns previously held about the unknown potential for IR-induced 
    long-term effects to the eyes of infants under infant radiant warmers. 
    The panel stated that in over 20 years of clinical use, there are no 
    reports in the literature of any adverse long-term effects to the eyes 
    of infants attributed to the IR radiation emitted by infant radiant 
    warmers (Ref. 3). They further commented that long-term developmental 
    health assessments of infants cared for in infant radiant warmers do 
    not mention any delayed eye conditions (Ref. 3, pp. 190 and 191). The 
    panel agreed that the potential risk to health of long-term damage from 
    overexposure of the eyes to total IR and IR-A could be controlled by 
    special controls.
    
    D. Damage to the Skin
    
        The IR emitted by infant radiant warmers is designed to be below 
    the threshold for thermal injury to the infant's skin (Ref. 24). The IR 
    is not of sufficient energy to cause photochemical reactions in the 
    skin. Most of the IR-A irradiance is reflected from the skin while IR-B 
    and IR-C irradiance are absorbed by the outer 1 millimeter of the skin 
    to accomplish the desired warming effect.
        The panel commented that there are no published reports of skin 
    damage in infants attributed to the use of radiant warmers and that 
    long-term developmental health assessments of infants cared for in 
    infant radiant warmers do not mention skin conditions (Ref. 3). The 
    panel believed that the potential risk of overexposure of the skin to 
    IR could be controlled by special controls.
    
    E. Increased Oxygen Consumption
    
        Bell reviewed five studies (Ref. 6) that reported conflicting 
    results of statistically significant increased oxygen consumption rates 
    (Refs. 30 and 31) and unchanged oxygen or slightly increased 
    consumption rates (Refs. 11, 28, 32, and 33) in infants warmed under 
    radiant warmers compared to infants warmed in incubators. Because 
    increased oxygen consumption may be an indicator of a stress-related 
    increase in metabolism, these reports caused concern that the use of 
    infant radiant warmers stress the metabolism of infants.
        Bell evaluated these studies taking into account differences in the 
    various study parameters used, including differences in the 
    servocontrol skin temperatures and the humidity in the neonatal 
    nurseries (Ref. 6). He determined that only a small increase in oxygen 
    consumption (4 kilocalories per kilogram per 24 hours additional energy 
    expenditure) occurs in the infants under infant radiant warmers 
    compared to infants in incubators. Bell agreed with Wheldon and Rutter 
    (Ref. 31) that the net total heat loss of infants under radiant warmers 
    to the environment due to evaporation, convection, radiation, and 
    conduction does not exceed that of infants in incubators. He concluded 
    that the increased oxygen consumption of infants in infant radiant 
    warmers is of unknown clinical significance. Subsequently, Marks et al. 
    reported that premature infants under infant radiant warmers 
    experienced no short-term metabolic complications or adverse effects on 
    growth even though they had a 10 percent higher oxygen consumption 
    compared to infants in incubators (Ref. 34).
        The panel acknowledged that although oxygen consumption may be 
    greater in infants cared for in infant radiant warmers than in 
    incubators, the clinical significance of this, if any, is unknown (Ref. 
    3). They noted that other factors unrelated to the device can also 
    cause increased oxygen consumption. The panel agreed this potential 
    risk could be controlled by special controls.
    
    F. Hypothermia and Hyperthermia
    
        The risks to health of hypothermia and hyperthermia are low during 
    proper use of the device (Ref. 1). Infant radiant warmers are used to 
    treat and to prevent hypothermia. Both hypothermia and hyperthermia can 
    result from malfunctioning alarms and radiant heater components, and 
    hyperthermia can result from detachment of the skin temperature probe 
    from the infant. The device's temperature and failure alarm system is 
    designed to prevent
    
    [[Page 44018]]
    
    hypothermia and hyperthermia by alerting operators of unsafe 
    temperature conditions, skin temperature probe detachment from the 
    skin, probe failure and device failure. The petition (Ref. 1), current 
    device labeling (Ref. 3), the AAMI voluntary standard special control 
    (Ref. 4), and accepted medical practice (Refs. 1 and 3) all recommend 
    frequent monitoring of infants under infant radiant warmers. They also 
    recommend that infant radiant warmers should be operated in the skin 
    temperature servocontrol mode rather than the manual mode to further 
    reduce the risks of both hypothermia and hyperthermia (Refs. 1 and 4). 
    The panel agreed that this risk to health could be controlled by 
    special controls.
    
    G. Other Risks
    
        Four other potential risks associated with the use of infant 
    radiant warmers are electrical shock due to improper design or 
    construction of the device, injury due to instability of the device, 
    burns to the operator if the device is constructed of materials that 
    absorb radiant heat, and operator error. Operator error can be 
    minimized by appropriate training and comprehensive device labeling. 
    The panel agreed that these are well-known risks that are generic to 
    many neonatal devices and that they can be controlled by special 
    controls (Ref. 3).
    
    H. Benefits of the Device
    
        The infant radiant warmer has the unique benefit of providing 
    greater accessibility to the infant than do incubators during routine 
    nursing and intensive care procedures without interrupting the delivery 
    of heat. Infant radiant warmers can also heat an infant faster than an 
    incubator. Ahlgren reported that only 5 to 10 minutes are required to 
    warm the infant's skin to the preset skin temperature with the infant 
    radiant warmer as compared to 45 to 50 minutes for the incubator (Ref. 
    35). Infant radiant warmers are recommended for the care of newborn 
    infants who lose large amounts of heat through evaporation of amniotic 
    fluid from their skin in the delivery room (Ref. 27). It is estimated 
    that 80 percent of all infants are placed under infant radiant warmers 
    at some time during their hospital stay (Ref. 1). Many practitioners 
    consider infant radiant warmers to be the only way of warming some very 
    low birth weight and critically ill infants (Refs. 3 and 6).
        The panel believes, based on publicly available, valid scientific 
    evidence, that the infant radiant warmer can be regulated as a class II 
    device (general and special controls) to reasonably assure the device's 
    safety and effectiveness (Ref. 3).
    
    IX. FDA's Tentative Findings
    
        FDA tentatively concurs with the recommendation of the panel that 
    infant radiant warmers should be reclassified into class II. The agency 
    believes that ``new information'' in the form of publicly available, 
    valid scientific evidence exists to establish special controls to 
    provide reasonable assurance of safety and effectiveness of the infant 
    radiant warmer for its intended use. The agency further identifies the 
    AAMI voluntary standard and labeling as the special controls. Moreover, 
    existing devices, within the generic type, have established a 
    reasonable record of safe and effective use. Consistent with the 
    purpose of the act, class II controls as defined by section 
    513(a)(1)(B) of the SMDA would provide the least amount of regulation 
    necessary to reasonably assure that current and future infant warmers 
    are safe and effective.
    
    X. Environmental Impact
    
        The agency has determined under 21 CFR 25.24(e)(2) that this action 
    is of a type that does not individually or cumulatively have a 
    significant effect on the human environment. Therefore, neither an 
    environmental assessment nor an environmental impact statement is 
    required.
    
    XI. Analysis of Impacts
    
        FDA has examined the impacts of the proposed rule under Executive 
    Order 12866 and the Regulatory Flexibility Act (Pub. L. 96-354). 
    Executive Order 12866 directs agencies to assess all costs and benefits 
    of available regulatory alternatives and, when regulation is necessary, 
    to select regulatory approaches that maximize net benefits (including 
    potential economic, environmental, public health and safety, and other 
    advantages; distributive impacts; and equity). The agency believes that 
    this proposed rule is consistent with the regulatory philosophy and 
    principles identified in the Executive Order. In addition, the proposed 
    rule is not a significant regulatory action as defined by the Executive 
    Order and so is not subject to review under the Executive Order.
        The Regulatory Flexibility Act requires agencies to analyze 
    regulatory options that would minimize any significant impact of a rule 
    on small entities. Because of the potential costs to comply with the 
    provisions of premarket approval (class III) by each manufacturer, the 
    agency believes that the economic impact to comply with special 
    controls (class II) would likely be less. Therefore, the agency 
    certifies that the proposed rule will not have a significant economic 
    impact on a substantial number of small entities. Therefore, under the 
    Regulatory Flexibility Act, no further analysis is required.
    
    XII. Paperwork Reduction Act of 1995
    
        FDA tentatively concludes that the labeling requirements in this 
    proposed rule are not subject to review by the Office of Management and 
    Budget because they do not constitute a ``collection of information'' 
    under the Paperwork Reduction Act of 1995 (Pub. L. 104-13). Rather, the 
    proposed labeling statements are ``public disclosure of information 
    originally supplied by the Federal Government to the recipient for the 
    purpose of disclosure to the public'' (5 CFR 1320.3(c)(2)).
    
    XIII. Request for Comments
    
        Interested persons may, on or before November 25, 1996, submit to 
    the Dockets Management Branch (address above) written comments 
    regarding this proposal. Two copies of any comments are to be submitted 
    except that individuals may submit one copy. Comments are to be 
    identified with the name of the device and the docket number found in 
    brackets in the heading of this document. Received comments may be seen 
    in the office above between 9 a.m. and 4 p.m., Monday through Friday.
    
    XIV. References
    
        The following information has been placed on display in the Dockets 
    Management Branch (address above) and may be seen by interested persons 
    between 9 a.m. and 4 p.m., Monday through Friday.
        1. Health Industry Manufacturers Association Petition with 
    incorporated errata, volumes 1 to 5, Washington, DC, 1986 (submitted 
    January 30, 1986; revised April 15, 1986).
        2. Transcript, General Hospital and Personal Use Devices Panel 
    (telephone conference call), May 21, 1986.
        3. Transcript, General Hospital and Personal Use Devices Panel 
    with the attached general device classification questionnaire and 
    supplemental data sheet, May 11, 1994.
        4. Association for the Advancement of Medical Instrumentation, 
    Infant Radiant Warmers (draft standard), May 1995.
        5. Baumgart, S., ``Radiant Energy and Insensible Water Loss in 
    the Premature Newborn Infant Nursed under a Radiant Warmer,'' 
    Clinics in Perinatology, 9:483-503, 1982.
        6. Bell, E. F., ``Infant Incubators and Radiant Warmers, Early 
    Human Development,'' 8:351-375, 1983 (included in Ref. 1).
    
    [[Page 44019]]
    
        7. Baumgart, S., W. D. Engel, W. W. Fox, and R. A. Polin, 
    ``Radiant Warmer Power and Body Size as Determinants of Insensible 
    Water Loss in the Critically Ill Neonate,'' Pediatric Research, 
    15:1495-1499, 1981 (included in Ref. 1).
        8. Williams, P. R., and W. Oh, ``Effects of Radiant Warmers on 
    Insensible Water Loss in Newborn Infants,'' American Journal of 
    Diseases in Childhood, 128:511-514, 1974 (included in Ref. 1).
        9. Wu, P. Y. K., and J. E. Hodgemen, ``Insensible Water Loss in 
    Preterm Infants. Changes with Postnatal Development and Non-Ionizing 
    Radiant Energy,'' Pediatrics, 54:704-711, 1974 (included in Ref. 1).
        10. Bell, W. F., M. R. Weinstein, and W. Oh, ``Heat Balance in 
    Premature Infants: Comparative Effects of Convectively Heated 
    Incubator and Infant Radiant Warmer, with and without Heatshield,'' 
    Journal of Pediatrics, 96:460-465, 1980 (included in Ref. 1).
        11. Jones, R. W. A., M. J. Rochefort, and J. D. Baum, 
    ``Increased Insensible Water Loss in New Born Infants Nursed under 
    Radiant Heaters,'' British Medical Journal, 1:1347-1350, 1976 
    (included in Ref. 1).
        12. Baumgart, S., C. B. Langman, W. W. Fox, and R. A. Polin, 
    ``Fluid, Electrolyte, and Glucose Maintenance in the Very Low Birth 
    Weight Infant,'' Clinical Pediatrics, 32:199-206, 1982 (included in 
    Ref. 1).
        13. Costarino, A. T., and S. Baumgart, ``Controversies in Fluid 
    and Electrolyte Therapy for the Premature Infant,'' Clinical 
    Perinatology, 15:863-878, 1988.
        14. Marks, K. H., Z. Freidman, and M. B. Maisels, ``A Simple 
    Device for Reducing Insensible Water Loss in Low-Birth Weight 
    Infants,'' Pediatrics, 60:223-226, 1980 (included in Ref. 1).
        15. Baumgart, S., W. D. Engel, W. W. Fox, and R. A. Polin, 
    ``Effect of Heat Shielding on Convective and Evaporative Heat Losses 
    and on Radiant Heat Transfer in the Premature Infant,'' Journal of 
    Pediatrics, 99:948-956, 1981 (included in Ref. 1).
        16. Baumgart, S., W. W. Fox, and R. A. Polin, ``Physiologic 
    Implications of Two Heat Shields for Infants under Infant Radiant 
    Warmers,'' Journal of Pediatrics, 100:787-790, 1982 (included in 
    Ref. 1).
        17. Fitch, C. W., and S. B. Korones, ``Heat Shield Reduces Water 
    Loss,'' Archives Disease in Childhood, 59:886-888, 1984 (included in 
    Ref. 1).
        18. Emergency Care Research Institute, ``Evaluation: Infant 
    Warmers,'' Health Devices, 13:119-145, 1984.
        19. Emergency Care Research Institute, ``The Development of a 
    Standard for Infant Warmers and Incubators,'' final report, FDA 
    Contract No. 223-75-5012, Plymouth Meeting, PA, 1976.
        20. Engel, W. D., S. Baumgart, W. W. Fox, and R. A. Polin, 
    ``Effect of Increased Radiant Power Output on State of Hydration in 
    the Critically Ill Neonate,'' Critical Care Medicine, 10:673-676, 
    1982 (included in Ref. 1).
        21. Engel, W. D., S. Baumgart, J. G. Schwartz, W. W. Fox, and R. 
    A. Polin, ``Insensible Water Loss in the Critically Ill Neonate,'' 
    American Journal of Diseases in Children, 135:516-520, 1981 
    (included in Ref. 1).
        22. Sliney, D. H., and B. C. Freasier, ``Evaluation of Optical 
    Radiation Hazards,'' Applied Optics, 12:1-24, 1973 (included in Ref. 
    1).
        23. Sliney, D., and M. Wolbarsht, ``Safety with Lasers and Other 
    Optical Sources,'' Plenum Press, New York, 1980, pp. 144-149.
        24. Emergency Care Research Institute, ``Evaluation: Infant 
    Warmers,'' Health Devices, 3:4-25, 1973.
        25. Johns, R., D. Schaffer, and G. Peckham, ``Evaluation of the 
    Effects of Infrared Radiation on the Eyes of Infants under Radiant 
    Warmers,'' Unpublished, 1977 (included in Ref. 1).
        26. Pitts, D. G., and A. P. Cullen, ``Determination of Infrared 
    Radiation Levels for Acute Ocular Cataractogenesis,'' Albrecht von 
    Graefes Archives Klinische Ophthalmologie, 217:285-297, 1981 
    (included in Ref. 1).
        27. Baumgart, S., A. Knauth, F. X. Casey, and G. E. Quinn, 
    ``Infrared Eye Injury Not Due to Radiant Warmer Use in Infants,'' 
    American Journal of Diseases in Childhood, 147:565-569, 1993.
        28. American Academy of Pediatrics and American College of 
    Obstetricians and Gynecologists, Guidelines for Perinatal Care, 
    Evanston and Washington, DC, respectively, 1983 (included in Ref. 
    1).
        29. International Electrotechnical Commission, ``Medical 
    Electrical Equipment, Part 2: Particular requirements for the safety 
    of infant radiant warmers'', 1994.
        30. LeBlanc, M., ``Relative Efficacy of Radiant and Convective 
    Heat in Incubators in Producing Thermoneutrality for the 
    Premature,'' Pediatric Research, 18:426-428, 1984 (included in Ref. 
    1).
        31. Wheldon, A. E., and N. Rutter, ``The Heat Balance of Small 
    Babies Nursed in Incubators and under Radiant Warmers,'' Early Human 
    Development, 6:131-143, 1982 (included in Ref. 1).
        32. Marks, K. H., R. C. Gunther, J. A. Rossi, and M. J. Maisels, 
    ``Oxygen Consumption and Insensible Water Loss in Premature Infants 
    under Radiant Heaters,'' Pediatrics, 66:228-232, 1980 (included in 
    Ref. 1).
        33. Darnall Jr., R. A., and R. L. Ariagno, ``Minimal Oxygen 
    Consumption in Infants Cared for under Overhead Radiant Warmers 
    Compared with Conventional Incubators,'' Journal of Pediatrics, 
    93:283-287, 1978 (included in Ref. 1).
        34. Marks, K. H., E. E. Nardis, and M. N. Momin, ``Energy 
    Metabolism and Substrate Utilization in Low Birth Weight Neonates 
    under Radiant Warmers,'' Pediatrics, 78:465-472, 1986.
        35. Ahlgren, E. W., ``Environmental Control of the Neonate 
    Receiving Intensive Care,'' International Anesthesiology Clinic, 
    12:173-215, 1974 (included in Ref. 1).
    
    List of Subjects in 21 CFR Part 880
    
        Medical devices.
        Therefore, under the Federal Food, Drug, and Cosmetic Act and under 
    authority delegated to the Commissioner of Food and Drugs, it is 
    proposed that 21 CFR part 880 be amended as follows:
    
    PART 880--GENERAL HOSPITAL AND PERSONAL USE DEVICES
    
        1. The authority citation for 21 CFR part 880 continues to read as 
    follows:
    
        Authority: Secs. 501, 510, 513, 515, 520, 701 of the Federal 
    Food, Drug, and Cosmetic Act (21 U.S.C. 351, 360, 360c, 360e, 360j, 
    371).
    
        2. Section Sec. 880.5130 is revised to read as follows:
    
    
    Sec. 880.5130  Infant radiant warmer.
    
        (a) Identification. The infant radiant warmer is a device 
    consisting of an infrared heating element intended to be placed over an 
    infant to maintain the infant's body temperature by means of radiant 
    heat. The device may also contain a temperature monitoring sensor, a 
    heat output control mechanism, and an alarm system (infant temperature, 
    manual mode if present, and failure alarms) to alert operators of a 
    temperature condition over or under the set temperature, manual mode 
    time limits, and device component failure, respectively. The device may 
    be placed over a pediatric hospital bed or it may be built into the bed 
    as a complete unit.
        (b) Classification. Class II (Special Controls). (1) Association 
    for the Advancement of Medical Instrumentation (AAMI) Voluntary 
    Standard for Infant Radiant Warmers; (2) prescription statement in 
    accordance with 21 CFR 801.109 (restricted to use by or upon the order 
    of qualified practitioners as determined by the States); (3) labeling 
    for use only in health care facilities and only by persons with 
    specific training and experience in the use of the device.
    
        Dated: August 1, 1996.
    D. B. Burlington,
    Director, Center for Devices and Radiological Health.
    [FR Doc. 96-21846 Filed 8-26-96; 8:45 am]
    BILLING CODE 4160-01-F
    
    
    

Document Information

Published:
08/27/1996
Department:
Health and Human Services Department
Entry Type:
Proposed Rule
Action:
Proposed rule.
Document Number:
96-21846
Dates:
Written comments by November 25, 1996. FDA proposes that any final rule based on this proposal become final 30 days after publication in the Federal Register.
Pages:
44013-44019 (7 pages)
Docket Numbers:
Docket No. 85N-0285
PDF File:
96-21846.pdf
CFR: (1)
21 CFR 880.5130