E7-18582. Agency Information Collection Activities; Submission for Office of Management and Budget Review; Comment Request; Medical Devices: Current Good Manufacturing Practice Quality System Regulations  

  • Start Preamble

    AGENCY:

    Food and Drug Administration, HHS.

    ACTION:

    Notice.

    SUMMARY:

    The Food and Drug Administration (FDA) is announcing that a proposed collection of information has been submitted to the Office of Management and Budget (OMB) for review and clearance under the Paperwork Reduction Act of 1995.

    DATES:

    Fax written comments on the collection of information by October 22, 2007.

    ADDRESSES:

    To ensure that comments on the information collection are received, OMB recommends that written comments be faxed to the Office of Information and Regulatory Affairs, OMB, Attn: FDA Desk Officer, FAX: 202-395-6974, or e-mailed to baguilar@omb.eop.gov. All comments should be identified with the OMB control number 0910-0073. Also include the FDA docket number found in brackets in the heading of this document.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    Denver Presley, Jr., Office of the Chief Information Officer (HFA-250), Food and Drug Administration, 5600 Fishers Lane, Rockville, MD 20857, 301-827-1472.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    In compliance with 44 U.S.C. 3507, FDA has submitted the following proposed collection of information to OMB for review and clearance:

    Medical Devices: Current Good Manufacturing Practice Quality System Regulations—21 CFR Part 820 (OMB Control Number 0910-0073)—Extension

    Under section 520(f) of the Federal Food, Drug, and Cosmetic Act (the act) (21 U.S.C. 360j(f)), the Secretary of the Department of Health and Human Services (the Secretary) has the authority to prescribe regulations requiring that the methods used in, and the facilities and controls used for, the manufacture, pre-production design validation (including a process to assess the performance of a device but not including an evaluation of the safety and effectiveness of a device), packing, storage, and installation of a device conform to current good manufacturing practices (CGMPs), as described in such regulations, to assure that the device will be safe and effective and otherwise in compliance with the act.

    The CGMP/Quality System (CGMP/QS) regulation implementing authority provided by this statutory provision is found under part 820 (21 CFR part 820) and sets forth basic CGMP requirements governing the design, manufacture, packing, labeling, storage, installation, and servicing of all finished medical Start Printed Page 53775devices intended for human use. The authority for this regulation is covered under the act , i.e. 21 U.S.C. 351, 352, 360, 360c, 360d, 360e, 360h, 360i, 360j, 360l, 371, 374, 381, and 383.

    The CGMP/QS regulation includes requirements for purchasing and service controls, clarifies recordkeeping requirements for device failure and complaint investigations, clarifies requirements for verifying/validating production processes and process or product changes, and clarifies requirements for product acceptance activities quality data evaluations and corrections of nonconforming product/quality problems.

    Requirements are compatible with specifications in International Standards, “ISO 9001: “Quality Systems Model for Quality Assurance in Design/Development, Production, Installation, and Servicing.” The CGMP/QS information collections will assist FDA inspections of manufacturers for compliance with QS requirements encompassing design, production, installation, and servicing processes.

    Section 820.20(a) through (e) requires management with executive responsibility to establish, maintain, and/or review the following topics: (1) The quality policy, (2) the organizational structure, (3) the quality plan, and (4) the QS procedures of the organization.

    Section 820.22 requires the conduct and documentation of QS audits and re-audits.

    Section 820.25(b) requires the establishment of procedures to identify training needs and documentation of such training.

    Section 820.30(a)(1) and (b) through (j), requires in respective order, the establishment, maintenance, and/or documentation of the following topics: (1) Procedures to control design of class III and class II devices, and certain class I devices as listed therein; (2) plans for design and development activities and updates; (3) procedures identifying, documenting, and approving design input requirements; (4) procedures defining design output, including acceptance criteria, and documentation of approved records; (5) procedures for formal review of design results and documentation of results in the design history file (DHF); (6) procedures for verifying device design and documentation of results and approvals in the DHF; (7) procedures for validating device design, including documentation of results in the DHF; (8) procedures for translating device design into production specifications; (9) procedures for documenting, verifying, validating approved design changes before implementation of changes; and (10) the records and references constituting the DHF for each type of device.

    Section 820.40 requires manufacturers to establish and maintain procedures controlling approval and distribution of required documents and document changes.

    Section 820.40(a) and (b) requires the establishment and maintenance of procedures for the review, approval, issuance and documentation of required records (documents) and changes to those records.

    Section 820.50(a)(1), (a)(2), (a)(3), and (b) requires the establishment and maintenance of procedures and requirements to ensure service and product quality, records of acceptable suppliers, and purchasing data describing specified requirements for products and services.

    Sections 820.60 and 820.65 require, respectively, the establishment and maintenance of procedures for identifying all products from receipt to distribution and for using control numbers to track surgical implants and life-sustaining or supporting devices and their components.

    Section 820.70(a)(1) through (a)(5), (b) through (e), (g)(1) through (g)(3), (h), and (i) requires the establishment, maintenance, and/or documentation of the following topics: (1) Process control procedures; (2) procedures for verifying or validating changes to specification, method, process, or procedure; (3) procedures to control environmental conditions and inspection result records; (4) requirements for personnel hygiene; (5) procedures for preventing contamination of equipment and products; (6) equipment adjustment, cleaning, and maintenance schedules; (7) equipment inspection records; (8) equipment tolerance postings; procedures for utilizing manufacturing materials expected to have an adverse effect on product quality; and (9) validation protocols and validation records for computer software and software changes.

    Sections 820.72(a), (b)(1), and (b)(2) and 820.75(a) through (c) require, respectively, the establishment, maintenance, and/or documentation of the following topics: (1) Equipment calibration and inspection procedures; (2) national, international or in-house calibration standards; (3) records that identify calibrated equipment and next calibration dates; (4) validation procedures and validation results for processes not verifiable by inspections and tests; (5) procedures for keeping validated processes within specified limits; (6) records for monitoring and controlling validated processes; and (7) records of the results of revalidation where necessitated by process changes or deviations.

    Sections 820.80(a) through (e) and 820.86 require, respectively, the establishment, maintenance, and/or documentation of the following topics: (1) Procedures for incoming acceptance by inspection, test or other verification; (2) procedures for ensuring that in-process products meet specified requirements and the control of product until inspection and tests are completed; (3) procedures for, and records that show, incoming acceptance or rejection is conducted by inspections, tests or other verifications; (4) procedures for, and records that show, finished devices meet acceptance criteria and are not distributed until device master record (DMR) activities are completed; (5) records in the device history record (DHR) showing acceptance dates, results and equipment used; and (6) the acceptance/rejection identification of products from receipt to installation and servicing.

    Sections 820.90(a), (b)(1), (b)(2), and 820.100 require, respectively, the establishment, maintenance and/or documentation of the following topics: (1) Procedures for identifying, recording, evaluating, and disposing of nonconforming product; (2) procedures for reviewing and recording concessions made for, and disposition of, nonconforming product; (3) procedures for reworking products, evaluating possible adverse rework effect and recording results in the DHR; (4) procedures and requirements for corrective and preventive actions, including analysis, investigation, identification and review of data, records, causes and results; and (5) records for all corrective and preventive action activities.

    Section 820.100(a)(1) through (a)(7) states that procedures and requirements shall be established and maintained for corrective/preventive actions, including the following: (1) Analysis of data from process, work, quality, servicing records; investigation of nonconformance causes; (2) identification of corrections and their effectiveness; (3) recording of changes made; and (4) appropriate distribution and managerial review of corrective and preventive action information.

    Section 820.120 states that manufacturers shall establish/maintain procedures to control labeling storage/application; and examination/release for storage and use, and document those procedures.

    Sections 820.120(b) and (d), 820.130, 820.140, 820.150(a) and (b), 820.160(a) Start Printed Page 53776and (b), and 820.170(a) and (b), respectively, require the establishment, maintenance, and/or documentation of the following topics: (1) Procedures for controlling and recording the storage, examination, release and use of labeling; (2) the filing of labels/labeling used in the DHR; (3) procedures for controlling product storage areas and receipt/dispatch authorizations; (4) procedures controlling the release of products for distribution; (5) distribution records that identify consignee, product, date, and control numbers; and (6) instructions, inspection and test procedures that are made available, and the recording of results for devices requiring installation.

    Sections 820.180(b) and (c), 820.181(a) through (e), 820.184(a) through (f), and 820.186 require, respectively, the maintenance of records that are: (1) Retained at prescribed site(s), made readily available and accessible to FDA and retained for the device's life expectancy or for 2 years; (2) contained or referenced in a DMR consisting of device, process, quality assurance, packaging and labeling, and installation, maintenance, and servicing specifications and procedures; (3) contained in a DHR and demonstrate the manufacture of each unit, lot or batch of product in conformance with DMR and regulatory requirements, include manufacturing and distribution dates, quantities, acceptance documents, labels and labeling, control numbers; and (4) contained in a quality system record (QSR), consisting of references, documents, procedures and activities not specific to particular devices.

    Sections 820.198(a) through (c) and 820.200(a) through (d), respectively, requires the establishment, maintenance and/or documentation of the following topics: (1) Complaint files and procedures for receiving, reviewing, and evaluating complaints; (2) complaint investigation records identifying the device, complainant, and relationship of the device to the incident; (3) complaint records that are reasonably accessible to the manufacturing site or at prescribed sites; (4) procedures for performing and verifying that device servicing requirements are met and that service reports involving complaints are processed as complaints; and (5) service reports that record the device, service activity, and test and inspection data.

    Section 820.250 requires the establishment and maintenance of procedures to identify valid statistical techniques necessary to verify process and product acceptability; and sampling plans, when used, that are written and based on valid statistical rationale, and procedures for ensuring adequate sampling methods.

    The CGMP/QS regulation amends and revises the CGMP requirements for medical devices set out under part 820. The regulation adds design and purchasing controls; modifies previous critical device requirements; revises previous validation and other requirements; and harmonizes device CGMP requirements with QS specifications in the international standard, “ISO 9001: Quality Systems Model for Quality Assurance in Design/Development Production, Installation and Servicing.” The rule does not apply to manufacturers of components or parts of finished devices, nor to manufacturers of human blood and blood components subject to 21 CFR part 606. With respect to devices classified in class I, design control requirements, apply only to class I devices listed in § 820.30(a)(2) of the regulation. The rule imposes burden upon: (1) Finished device manufacturer firms which are subject to all recordkeeping requirements; (2) finished device contract manufacturers; specification developers; and (3) repacker, relabelers and contract sterilizer firms, which are subject only to requirements applicable to their activities. In addition, re-manufacturers of hospital single-use devices (SUDs) will now be considered to have the same requirements as manufacturers in regard to this regulation. The establishment, maintenance and/or documentation of procedures, records and data required by this regulation will assist FDA in determining whether firms are in compliance with CGMP requirements, which are intended to ensure that devices meet their design, production, labeling, installation, and servicing specifications and, thus are safe, effective and suitable for their intended purpose. In particular, compliance with CGMP design control requirements should decrease the number of design-related device failures that have resulted in deaths and serious injuries.

    The CGMP/QS regulation applies to approximately 8,963 respondents. These recordkeepers consist of 8,945 original respondents and an estimated 18 hospitals which remanufacture or reuse SUDs. They include manufacturers, subject to all requirements and contract manufacturers, specification developers, repackers, relabelers and contract sterilizers, subject only to requirements applicable to their activities. Hospital remanufacturers of SUDs are now defined to be manufacturers under guidelines issued by FDA's Center for Devices and Radiological Health's (CDRH), Office of Surveillance and Biometrics. Respondents to this collection have no reporting activities, but must make required records available for review or copying during FDA inspection. The regulation contains additional recordkeeping requirements in such areas as design control, purchasing, installation, and information relating to the remanufacture of SUDs. The estimates for this burden are derived from those incremental tasks that were determined when the new CGMP/QS regulation became final as well as those carryover requirements. The carryover requirements are based on decisions made by the agency on July 16, 1992, under OMB clearance submission 0910-0073, which still provides valid baseline data.

    FDA estimates respondents will have a total annual recordkeeping burden of approximately 3,076,370 hours. This figure also consists of approximately 143,052 hours spent on a startup basis by 650 new firms.

    In the Federal Register of Monday, July 9, 2007, FDA published a 60-day notice soliciting public comments on the information collection requirements for the proposed extension of this collection of information. In response to that notice, no comments were received.

    FDA estimates the burden for this collection of information as follows:

    Table 1.—Estimated Annual Recordkeeping Burden1

    CFR SectionNo. of RecordkeepersAnnual Frequency of RecordkeepingTotal Annual HoursHours Per RecordkeeperTotal Hours
    820.20(a)8,96318,9636.5858,977
    820.20(b)8,96318,9634.4339,706
    820.20(c)8,96318,9636.1755,302
    Start Printed Page 53777
    820.20(d)8,96318,9639.8988,644
    820.20(e)8,96318,9639.8988,644
    820.228,96318,96332.72293,269
    820.25(b)8,96318,96312.68113,651
    820.30(a)(1)8,96318,9631.7515,685
    820.30(b)8,96318,9635.9553,330
    820.30(c) and (d)8,96318,9631.7515,685
    820.30(e)8,96318,96323.39209,645
    820.30(f) and (g)8,96318,96337.42335,395
    820.30(h)8,96318,9633.3429,936
    820.30(i)8,96318,96317.26154,701
    820.30(j)8,96318,9632.6423,662
    820.408,96318,9638.9179,860
    820.40(a)and (b)8,96318,9632.0418,285
    820.50(a)(1) through (a)(3)8,96318,96321.90196,290
    820.50(b)8,96318,9636.0253,957
    820.68,96318,9630.322,868
    820.658,96318,9630.676,005
    820.70(a)(1) through (a)(5), (b), and (c)8,96318,9631.8516,582
    820.70(d)8,96318,9632.8725,724
    820.70(e)8,96318,9631.8516,582
    820.70(g)(1) through (g)(3)8,96318,9631.4312,817
    820.70(h)8,96318,9631.8516,582
    820.70(i)8,96318,9637.5067,223
    820.72(a)8,96318,9634.9244,098
    820.72(b)(1) and (b)(2)8,96318,9631.4312,817
    820.75(a)8,96318,9632.6924,110
    820.75(b)8,96318,9631.029,142
    820.75(c)8,96318,9631.119,949
    820.80(a) through (e)8,96318,9634.8043,022
    820.868,96318,9630.797,081
    820.90(a), (b)(1), and (b)(2)8,96318,9634.9544,367
    820.100 (a)(1) through (a)(7)8,96318,96312.48111,858
    820.100(b)8,96318,9631.2811,473
    820.120(b) and (d)8,96318,9630.454,033
    820.1308,96318,9630.454,033
    Start Printed Page 53778
    820.1408,96318,9636.3456,825
    820.150(a) and (b)8,96318,9635.6750,820
    820.160(a) and (b)8,96318,9630.676,005
    820.170(a) and (b)8,96318,9631.5013,445
    820.180(b) and (c)8,96318,9631.5013,445
    820.181(a) through (e)8,96318,9631.2110,845
    820.184(a) through (f)8,96318,9631.4112,638
    820.1868,96318,9630.403,585
    820.198(a) through (c)8,96318,9634.9444,277
    820.200(a) and (d)8,96318,9632.6123,393
    820.258,96318,9630.676,005
    Totals3,072,337
    1 There are no capital costs or operating and maintenance costs associated with this collection of information.

    Burden (labor) hour and cost estimates were originally developed under FDA contract by Eastern Research Group, Inc. (ERG), in 1996 when the CGMP/QS regulation became final. These figures are still accurate. Additional factors considered in deriving estimates included:

    • Establishment Type: Query has been made of CDRH's registration/listing databank and has counted 8,963 domestic firms subject to CGMPs. In addition, hospitals which reuse or remanufacture devices are now considered manufacturers under new FDA guidance. After investigations of many hospitals and the changes in enforcements of FDA's requirements for hospitals, the number of reuse or remanufactures of SUDs have decreased from the estimated 66 to an estimated 18 hospitals. Because the total number of registered firms is not static, the number of respondents will fluctuate from year to year resulting in slight changes to the overall burden. Currently, there are 8,963 firms subject to the CGMPs; an increase from the last renewal of 8,254.
    • Potentially Affected Establishments: Except for manufacturers, not every type of firm is subject to every CGMP/QS requirement. For example, all are subject to FDA's quality policy regulations (§ 820.20(a)), document control regulations (§ 820.40), and other requirements, whereas only manufacturers and specification developers are subject to FDA's design controls regulations (§ 820.30). The type of firm subject to each requirement was identified by ERG.

    FDA estimates the burden hours (and costs) based on the last approved renewal for this information collection.

    FDA estimates that some 650 “new” establishments (marketing devices for the first time) will expend some 143,052 “development” hours on a one-time startup basis to develop records and procedures for the CGMP/QS regulation.

    FDA estimates that annual labor hours are apportioned as follows: 40 percent goes to requirements dealing with manufacturing specifications, process controls and the DHR; 20 percent goes to requirements dealing with components and acceptance activities; 25 percent goes to requirements dealing with equipment, records (the DMR and QSR), complaint investigations, labeling/packaging and reprocessing/investigating product nonconformance; and 15 percent goes to quality audit, traceability, handling, distribution, statistical, and other requirements.

    Start Signature

    Dated: September 14, 2007.

    Jeffrey Shuren,

    Assistant Commissioner for Policy.

    End Signature End Supplemental Information

    [FR Doc. E7-18582 Filed 9-19-07; 8:45 am]

    BILLING CODE 4160-01-S

Document Information

Published:
09/20/2007
Department:
Food and Drug Administration
Entry Type:
Notice
Action:
Notice.
Document Number:
E7-18582
Dates:
Fax written comments on the collection of information by October 22, 2007.
Pages:
53774-53778 (5 pages)
Docket Numbers:
Docket No. 2007N-0229
PDF File:
e7-18582.pdf