2016-21553. Agency Information Collection Activities; Proposed Collection; Comment Request; Medical Device: Current Good Manufacturing Practice Quality System Regulations
-
Start Preamble
AGENCY:
Food and Drug Administration, HHS.
ACTION:
Notice.
SUMMARY:
The Food and Drug Administration (FDA) is announcing an opportunity for public comment on the proposed collection of certain information by the Agency. Under the Paperwork Reduction Act of 1995 (the PRA), Federal Agencies are required to publish notice in the Federal Register concerning each proposed collection of information including each proposed extension of an existing collection of information, and to allow 60 days for public comment in response to the notice. This notice solicits comments on recordkeeping requirements related to the medical devices current good manufacturing practice (CGMP) quality system (QS) regulation (CGMP/QS regulation).
DATES:
Submit either electronic or written comments on the collection of information by November 7, 2016.
ADDRESSES:
You may submit comments as follows:
Electronic Submissions
Submit electronic comments in the following way:Start Printed Page 62145
- Federal eRulemaking Portal: http://www.regulations.gov. Follow the instructions for submitting comments. Comments submitted electronically, including attachments, to http://www.regulations.gov will be posted to the docket unchanged. Because your comment will be made public, you are solely responsible for ensuring that your comment does not include any confidential information that you or a third party may not wish to be posted, such as medical information, your or anyone else's Social Security number, or confidential business information, such as a manufacturing process. Please note that if you include your name, contact information, or other information that identifies you in the body of your comments, that information will be posted on http://www.regulations.gov.
- If you want to submit a comment with confidential information that you do not wish to be made available to the public, submit the comment as a written/paper submission and in the manner detailed (see “Written/Paper Submissions” and “Instructions”).
Written/Paper Submissions
Submit written/paper submissions as follows:
- Mail/Hand delivery/Courier (for written/paper submissions): Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
- For written/paper comments submitted to the Division of Dockets Management, FDA will post your comment, as well as any attachments, except for information submitted, marked and identified, as confidential, if submitted as detailed in “Instructions.”
Instructions: All submissions received must include the Docket No. FDA-2016-N-2544 for “Agency Information Collection Activities; Proposed Collection; Comment Request; Medical Device: Current Good Manufacturing Practice Quality System Regulations.”
Received comments will be placed in the docket and, except for those submitted as “Confidential Submissions,” publicly viewable at http://www.regulations.gov or at the Division of Dockets Management between 9 a.m. and 4 p.m., Monday through Friday.
- Confidential Submissions—To submit a comment with confidential information that you do not wish to be made publicly available, submit your comments only as a written/paper submission. You should submit two copies total. One copy will include the information you claim to be confidential with a heading or cover note that states “THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION.” The Agency will review this copy, including the claimed confidential information, in its consideration of comments. The second copy, which will have the claimed confidential information redacted/blacked out, will be available for public viewing and posted on http://www.regulations.gov. Submit both copies to the Division of Dockets Management. If you do not wish your name and contact information to be made publicly available, you can provide this information on the cover sheet and not in the body of your comments and you must identify this information as “confidential.” Any information marked as “confidential” will not be disclosed except in accordance with 21 CFR 10.20 and other applicable disclosure law. For more information about FDA's posting of comments to public dockets, see 80 FR 56469, September 18, 2015, or access the information at: http://www.fda.gov/regulatoryinformation/dockets/default.htm.
Docket: For access to the docket to read background documents or the electronic and written/paper comments received, go to http://www.regulations.gov and insert the docket number, found in brackets in the heading of this document, into the “Search” box and follow the prompts and/or go to the Division of Dockets Management, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
Start Further InfoFOR FURTHER INFORMATION CONTACT:
FDA PRA Staff, Office of Operations, Food and Drug Administration, 8455 Colesville Rd., COLE-14526, Silver Spring, MD 20993-0002, PRAStaff@fda.hhs.gov.
End Further Info End Preamble Start Supplemental InformationSUPPLEMENTARY INFORMATION:
Under the PRA (44 U.S.C. 3501-3520), Federal Agencies must obtain approval from the Office of Management and Budget (OMB) for each collection of information they conduct or sponsor. “Collection of information” is defined in 44 U.S.C. 3502(3) and 5 CFR 1320.3(c) and includes Agency requests or requirements that members of the public submit reports, keep records, or provide information to a third party. Section 3506(c)(2)(A) of the PRA (44 U.S.C. 3506(c)(2)(A)) requires Federal Agencies to provide a 60-day notice in the Federal Register concerning each proposed collection of information, including each proposed extension of an existing collection of information, before submitting the collection to OMB for approval. To comply with this requirement, FDA is publishing notice of the proposed collection of information set forth in this document.
With respect to the following collection of information, FDA invites comments on these topics: (1) Whether the proposed collection of information is necessary for the proper performance of FDA's functions, including whether the information will have practical utility; (2) the accuracy of FDA's estimate of the burden of the proposed collection of information, including the validity of the methodology and assumptions used; (3) ways to enhance the quality, utility, and clarity of the information to be collected; and (4) ways to minimize the burden of the collection of information on respondents, including through the use of automated collection techniques, when appropriate, and other forms of information technology.
Medical Devices: Current Good Manufacturing Practice Quality System Regulation—21 CFR Part 820—OMB Control Number 0910-0073—Extension
Under section 520(f) of the Federal Food, Drug, and Cosmetic Act (the FD&C Act) (21 U.S.C. 360j(f)), the Secretary of the Department of Health and Human Services has the authority to prescribe regulations requiring that the methods used in, and the facilities and controls used for, the manufacture, preproduction design validation (including a process to assess the performance of a device, but not including an evaluation of the safety and effectiveness of a device), packing, storage, and installation of a device conform to CGMP, as described in such regulations, to assure that the device will be safe and effective and otherwise in compliance with the FD&C Act.
The CGMP/QS regulation implementing authority provided by this statutory provision is found under part 820 (21 CFR part 820) and sets forth basic CGMP requirements governing the design, manufacture, packing, labeling, storage, installation, and servicing of all finished medical devices intended for human use. The authority for this regulation is covered under sections 501, 502, 510, 513, 514, 515, 518, 519, 520, 522, 701, 704, 801, and 803 of the FD&C Act (21 U.S.C. 351, 352, 360, 360c, 360d, 360e, 360h, 360i, 360j, 360l, 371, 374, 381, and 383). The CGMP/QS regulation includes requirements for purchasing and service controls, clarifies recordkeeping requirements for device failure and complaint investigations, clarifies requirements for verifying/validating production processes and process or product changes, and clarifies requirements for product acceptance activities quality Start Printed Page 62146data evaluations and corrections of nonconforming product/quality problems.
Requirements are compatible with specifications in the international standards “ISO 9001: Quality Systems Model for Quality Assurance in Design/Development, Production, Installation, and Servicing.” The CGMP/QS information collections will assist FDA inspections of manufacturers for compliance with QS requirements encompassing design, production, installation, and servicing processes.
Section 820.20(a) through (e) requires management with executive responsibility to establish, maintain, and/or review the following topics: (1) The quality policy, (2) the organizational structure, (3) the quality plan, and (4) the quality system procedures of the organization. Section 820.22 requires the conduct and documentation of QS audits and re-audits. Section 820.25(b) requires the establishment of procedures to identify training needs and documentation of such training.
Section 820.30(a)(1) and (b) through (j) requires, in respective order, the establishment, maintenance, and/or documentation of the following topics: (1) Procedures to control design of class III and class II devices and certain class I devices as listed therein; (2) plans for design and development activities and updates; (3) procedures identifying, documenting, and approving design input requirements; (4) procedures defining design output, including acceptance criteria, and documentation of approved records; (5) procedures for formal review of design results and documentation of results in the design history file (DHF); (6) procedures for verifying device design and documentation of results and approvals in the DHF; (7) procedures for validating device design, including documentation of results in the DHF; (8) procedures for translating device design into production specifications; (9) procedures for documenting, verifying, and validating approved design changes before implementation of changes; and (10) the records and references constituting the DHF for each type of device.
Section 820.40 requires manufacturers to establish and maintain procedures controlling approval and distribution of required documents and document changes. Section 820.40(a) and (b) requires the establishment and maintenance of procedures for the review, approval, issuance, and documentation of required records (documents) and changes to those records.
Section 820.50(a) and (b) requires the establishment and maintenance of procedures and requirements to ensure service and product quality, records of acceptable suppliers, and purchasing data describing specified requirements for products and services.
Sections 820.60 and 820.65 require, respectively, the establishment and maintenance of procedures for identifying all products from receipt to distribution and for using control numbers to track surgical implants and life-sustaining or supporting devices and their components.
Section 820.70(a) through (e), (g)(1) through (g)(3), (h), and (i) requires the establishment, maintenance, and/or documentation of the following topics: (1) Process control procedures; (2) procedures for verifying or validating changes to specification, method, process, or procedure; (3) procedures to control environmental conditions and inspection result records; (4) requirements for personnel hygiene; (5) procedures for preventing contamination of equipment and products; (6) equipment adjustment, cleaning, and maintenance schedules; (7) equipment inspection records; (8) equipment tolerance postings, procedures for utilizing manufacturing materials expected to have an adverse effect on product quality; and (9) validation protocols and validation records for computer software and software changes.
Sections 820.72(a), (b)(1), and (b)(2); and 820.75(a) through (c) require, respectively, the establishment, maintenance, and/or documentation of the following topics: (1) Equipment calibration and inspection procedures; (2) national, international, or in-house calibration standards; (3) records that identify calibrated equipment and next calibration dates; (4) validation procedures and validation results for processes not verifiable by inspections and tests; (5) procedures for keeping validated processes within specified limits; (6) records for monitoring and controlling validated processes; and (7) records of the results of revalidation where necessitated by process changes or deviations.
Sections 820.80(a) through (e) and 820.86, respectively, require the establishment, maintenance, and/or documentation of the following topics: (1) Procedures for incoming acceptance by inspection, test, or other verification; (2) procedures for ensuring that in process products meet specified requirements and the control of product until inspection and tests are completed; (3) procedures for, and records that show, incoming acceptance or rejection is conducted by inspections, tests or other verifications; (4) procedures for, and records that show, finished devices meet acceptance criteria and are not distributed until device master record (DMR) activities are completed; (5) records in the device history record (DHR) showing acceptance dates, results, and equipment used; and (6) the acceptance/rejection identification of products from receipt to installation and servicing.
Sections 820.90(a), (b)(1), and (b)(2) and 820.100 require, respectively, the establishment, maintenance and/or documentation of the following topics: (1) Procedures for identifying, recording, evaluating, and disposing of nonconforming product; (2) procedures for reviewing and recording concessions made for, and disposition of, nonconforming product; (3) procedures for reworking products, evaluating possible adverse rework effect and recording results in the DHR; (4) procedures and requirements for corrective and preventive actions, including analysis, investigation, identification and review of data, records, causes, and results; and (5) records for all corrective and preventive action activities.
Section 820.100(a)(1) through (a)(7) states that procedures and requirements shall be established and maintained for corrective/preventive actions, including the following: (1) Analysis of data from process, work, quality, servicing records, investigation of nonconformance causes; (2) identification of corrections and their effectiveness; (3) recording of changes made; and (4) appropriate distribution and managerial review of corrective and preventive action information. Section 820.120 states that manufacturers shall establish/maintain procedures to control labeling storage/application; and examination/release for storage and use, and document those procedures.
Sections 820.120(b) and (d); 820.130; 820.140; 820.150(a) and (b); 820.160(a) and (b); and 820.170(a) and (b), respectively, require the establishment, maintenance, and/or documentation of the following topics: (1) Procedures for controlling and recording the storage, examination, release, and use of labeling; (2) the filing of labels/labeling used in the DHR; (3) procedures for controlling product storage areas and receipt/dispatch authorizations; (4) procedures controlling the release of products for distribution; (5) distribution records that identify consignee, product, date, and control numbers; and (6) instructions, inspection and test procedures that are Start Printed Page 62147made available, and the recording of results for devices requiring installation.
Sections 820.180(b) and (c); 820.181(a) through (e); 820.184(a) through (f); and 820.186 require, respectively, the maintenance of records that are: (1) Retained at prescribed site(s), made readily available and accessible to FDA, and retained for the device's life expectancy or for 2 years; (2) contained or referenced in a DMR consisting of device, process, quality assurance, packaging and labeling, and installation, maintenance, and servicing specifications and procedures; (3) contained in a DHR and demonstrate the manufacture of each unit, lot, or batch of product in conformance with DMR and regulatory requirements include manufacturing and distribution dates, quantities, acceptance documents, labels and labeling, and control numbers; and (4) contained in a quality system record, consisting of references, documents, procedures, and activities not specific to particular devices.
Sections 820.198(a) through (c); and 820.200(a) through (d), respectively, require the establishment, maintenance, and/or documentation of the following topics: (1) Complaint files and procedures for receiving, reviewing, and evaluating complaints; (2) complaint investigation records identifying the device, complainant, and relationship of the device to the incident; (3) complaint records that are reasonably accessible to the manufacturing site or at prescribed sites; (4) procedures for performing and verifying that device servicing requirements are met and that service reports involving complaints are processed as complaints; and (5) service reports that record the device, service activity, and test and inspection data.
Section 820.250 requires the establishment and maintenance of procedures to identify valid statistical techniques necessary to verify process and product acceptability; and sampling plans, when used, which are written and based on valid statistical rationale; and procedures for ensuring adequate sampling methods.
The CGMP/QS regulation added design and purchasing controls, modified previous critical device requirements, revised previous validation and other requirements, and harmonized device CGMP requirements with QS specifications in the international standard “ISO 9001: Quality Systems Model for Quality Assurance in Design/Development, Production, Installation, and Servicing.” The rule does not apply to manufacturers of components or parts of finished devices, or to manufacturers of human blood and blood components subject to 21 CFR part 606. With respect to devices classified in class I, design control requirements apply only to class I devices listed in § 820.30(a)(2) of the regulation. The rule imposes burden upon: (1) Finished device manufacturer firms, which are subject to all recordkeeping requirements; (2) finished device contract manufacturers, specification developers; and (3) re-packer, re-labelers, and contract sterilizer firms, which are subject only to requirements applicable to their activities. In addition, remanufacturers of hospital single-use devices are now considered to have the same requirements as manufacturers in regard to the regulation.
The establishment, maintenance, and/or documentation of procedures, records, and data required by the regulation assists FDA in determining whether firms are in compliance with CGMP requirements, which are intended to ensure that devices meet their design, production, labeling, installation, and servicing specifications and, thus are safe, effective, and suitable for their intended purpose. In particular, compliance with CGMP design control requirements should decrease the number of design-related device failures that have resulted in deaths and serious injuries.
The CGMP/QS regulation applies to approximately 24,738 respondents. A query of the Agency's registration and listing database shows that approximately 13,294 domestic and 11,444 foreign establishments are respondents to this information collection.[1] Respondents to this collection have no reporting activities, but must make required records available for review or copying during FDA inspection. Except for manufacturers, not every type of firm is subject to every CGMP/QS requirement. For example, all are subject to Quality Policy (§ 820.20(a)), Document Control (§ 820.40), and other requirements, whereas only manufacturers and specification developers are subject to subpart C, Design Controls. The PRA burden placed on the 24,738 establishments is an average burden.
Start SignatureTable 1—Estimated Annual Recordkeeping Burden 1
Activity/21 CFR section Number of recordkeepers Number of records per recordkeeper Total annual records Average burden per recordkeeping Total hours Quality policy—820.20(a) 24,738 1 24,738 7 173,166 Organization—820.20(b) 24,738 1 24,738 4 98,952 Management review—820.20(c) 24,738 1 24,738 6 148,428 Quality planning—820.20(d) 24,738 1 24,738 10 247,380 Quality system procedures—820.20(e) 24,738 1 24,738 10 247,380 Quality audit—820.22 24,738 1 24,738 33 816,354 Training—820.25(b) 24,738 1 24,738 13 321,594 Design procedures—820.30(a)(1) 24,738 1 24,738 2 49,476 Design and development planning—820.30(b) 24,738 1 24,738 6 148,428 Design input—820.30(c) 24,738 1 24,738 2 49,476 Design output—820.30(d) 24,738 1 24,738 2 49,476 Design review—820.30(e) 24,738 1 24,738 23 568,974 Design verification—820.30(f) 24,738 1 24,738 37 915,306 Design validation—820.30(g) 24,738 1 24,738 37 915,306 Design transfer—820.30(h) 24,738 1 24,738 3 74,214 Design changes—820.30(i) 24,738 1 24,738 17 420,546 Design history file—820.30(j) 24,738 1 24,738 3 74,214 Document controls—820.40 24,738 1 24,738 9 222,642 Documentation approval and distribution and document changes—820.40(a) and (b) 24,738 1 24,738 2 49,476 Start Printed Page 62148 Purchasing controls—820.50(a) 24,738 1 24,738 22 544,236 Purchasing data—820.50(b) 24,738 1 24,738 6 148,428 Identification—820.60 24,738 1 24,738 1 24,738 Traceability—820.65 24,738 1 24,738 1 24,738 Production and process controls—820.70(a) 24,738 1 24,738 2 49,476 Production and process changes and environmental control—820.70(b) and (c) 24,738 1 24,738 2 49,476 Personnel—820.70(d) 24,738 1 24,738 3 74,214 Contamination control—820.70(e) 24,738 1 24,738 2 49,476 Equipment maintenance schedule, inspection, and adjustment—820.70(g)(1)-(g)(3) 24,738 1 24,738 1 24,738 Manufacturing material—820.70(h) 24,738 1 24,738 2 49,476 Automated processes—820.70(i) 24,738 1 24,738 8 197,904 Control of inspection, measuring, and test equipment—820.72(a) 24,738 1 24,738 5 123,690 Calibration procedures, standards, and records—820.72(b)(1)-(b)(2) 24,738 1 24,738 1 24,738 Process validation—820.75(a) 24,738 1 24,738 3 74,214 Validated process parameters, monitoring, control methods, and data—820.75(b) 24,738 1 24,738 1 24,738 Revalidation—820.75(c) 24,738 1 24,738 1 24,738 Acceptance activities—820.80(a)-(e) 24,738 1 24,738 5 123,690 Acceptance status—820.86 24,738 1 24,738 1 24,738 Control of nonconforming product—820.90(a) 24,738 1 24,738 5 123,690 Nonconforming product review/disposition procedures and rework procedures—820.90(b)(1)-(b)(2) 24,738 1 24,738 5 123,690 Procedures for corrective/preventive actions—820.100(a)(1)-(a)(7) 24,738 1 24,738 12 296,856 Corrective/preventive activities—820.100(b) 24,738 1 24,738 1 24,738 Labeling procedures—820.120(b) 24,738 1 24,738 1 24,738 Labeling documentation—820.120(d) 24,738 1 24,738 1 24,738 Device packaging—820.130 24,738 1 24,738 1 24,738 Handling—820.140 24,738 1 24,738 6 148,428 Storage—820.150(a) and (b) 24,738 1 24,738 6 148,428 Distribution procedures and records—820.160(a) and (b) 24,738 1 24,738 1 24,738 Installation—820.170 24,738 1 24,738 2 49,476 Record retention period—820.180(b) and (c) 24,738 1 24,738 2 49,476 Device master record—820.181 24,738 1 24,738 1 24,738 Device history record—820.184 24,738 1 24,738 1 24,738 Quality system record—820.186 24,738 1 24,738 1 24,738 Complaint files—820.198(a), (c), and (g) 24,738 1 24,738 5 123,690 Servicing procedures and reports—820.200(a) and (d) 24,738 1 24,738 3 74,214 Statistical techniques procedures and sampling plans—820.250 24,738 1 24,738 1 24,738 Total 8,410,920 1 There are no capital costs or operating and maintenance costs associated with this collection of information. Dated: September 1, 2016.
Leslie Kux,
Associate Commissioner for Policy.
Footnotes
1. Based on fiscal year 2015 data.
Back to Citation[FR Doc. 2016-21553 Filed 9-7-16; 8:45 am]
BILLING CODE 4164-01-P
Document Information
- Published:
- 09/08/2016
- Department:
- Food and Drug Administration
- Entry Type:
- Notice
- Action:
- Notice.
- Document Number:
- 2016-21553
- Dates:
- Submit either electronic or written comments on the collection of information by November 7, 2016.
- Pages:
- 62144-62148 (5 pages)
- Docket Numbers:
- Docket No. FDA-2016-N-2544
- PDF File:
- 2016-21553.pdf
- Supporting Documents:
- » Agency Information Collection Activities; Proposed Collection; Comment Request; Current Good Manufacturing Practice; Quality System Regulation
- » Agency Information Collection Activities; Announcement of Office of Management and Budget Approvals
- » Agency Information Collection Activities; Submission for Office of Management and Budget Review; Comment Request; Medical Devices; Current Good Manufacturing Practice Quality System Regulation
- » Agency Information Collection Activities; Proposed Collection; Comment Request; Medical Devices; Current Good Manufacturing Practice Quality System Regulation
- » Agency Information Collection Activities; Announcement of Office of Management and Budget Approvals
- » Agency Information Collection Activities; Submission for Office of Management and Budget Review; Comment Request; Medical Device: Current Good Manufacturing Practice Quality System Regulations
- » Agency Information Collection Activities; Proposed Collection; Comment Request; Medical Device: Current Good Manufacturing Practice Quality System Regulations