2022-09749. Medical Devices; General and Plastic Surgery Devices; Classification of the Mountable Electromechanical Surgical System for Transluminal Approaches  

  • Start Preamble

    AGENCY:

    Food and Drug Administration, HHS.

    ACTION:

    Final amendment; final order.

    SUMMARY:

    The Food and Drug Administration (FDA, Agency, or we) is classifying the mountable electromechanical surgical system for transluminal approaches into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the mountable electromechanical surgical system for transluminal approaches' Start Printed Page 26994 classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices.

    DATES:

    This order is effective May 6, 2022. The classification was applicable on February 26, 2021.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    Virag Patel, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 4609, Silver Spring, MD 20993-0002, 301-796-0452, Virag.Patel@fda.hhs.gov.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    I. Background

    Upon request, FDA has classified the mountable electromechanical surgical system for transluminal approaches as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness. In addition, we believe this action will enhance patients' access to beneficial innovation, in part by placing the device into a lower device class than the automatic class III assignment.

    The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).

    FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate device by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).

    FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act. Section 207 of the Food and Drug Administration Modernization Act of 1997 (Pub. L. 105-115) established the first procedure for De Novo classification. Section 607 of the Food and Drug Administration Safety and Innovation Act (Pub. L. 112-144) modified the De Novo application process by adding a second procedure. A device sponsor may utilize either procedure for De Novo classification.

    Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).

    Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.

    Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device.

    When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C Act). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application to market a substantially equivalent device (see section 513(i) of the FD&C Act, defining “substantial equivalence”). Instead, sponsors can use the less-burdensome 510(k) process, when necessary, to market their device.

    II. De Novo Classification

    On April 17, 2019, FDA received Memic Innovative Surgery Ltd.'s request for De Novo classification of the Hominis Surgical System. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.

    We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.

    Therefore, on February 26, 2021, FDA issued an order to the requester classifying the device into class II. In this final order, FDA is codifying the classification of the device by adding 21 CFR 878.4961.[1] We have named the generic type of device mountable electromechanical surgical system for transluminal approaches, and it is identified as a software-controlled, patient bed- and/or operating table-mounted electromechanical surgical system with human/device interfaces that allows a qualified user to perform transluminal endoscopic or laparoscopic surgical procedures using surgical instruments attached to an electromechanical arm.

    FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1. Start Printed Page 26995

    Table 1—Mountable Electromechanical Surgical System for Transluminal Approaches Risks and Mitigation Measures

    Identified risksMitigation measures
    Thermal, electrical, or mechanical fault, or system malfunction resulting in tissue perforation or injury to patient or userNon-clinical performance testing; Electrical safety testing; Electromagnetic compatibility (EMC) testing; Software verification, validation, and hazard analysis; Human factors assessment; Clinical performance testing; Annual reporting; and Labeling.
    Use error resulting in patient injury: • Dehiscence or delayed healing at the device access site. • Hemorrhage. • Thromboembolism. • Transluminal risks.Non-clinical performance testing; Human factors assessment; Training; Clinical performance testing; Post-market surveillance; Annual reporting; Control on distribution; and Labeling.
    Adverse tissue reactionBiocompatibility evaluation, and Pyrogenicity testing.
    InfectionBiocompatibility evaluation; Pyrogenicity testing; Sterilization validation; Reprocessing validation; Shelf-life testing; Clinical performance testing; and Labeling.

    FDA has determined that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of safety and effectiveness. For a device to fall within this classification, and thus avoid automatic classification in class III, it would have to comply with the special controls named in this final order. The necessary special controls appear in the regulation codified by this order. This device is subject to premarket notification requirements under section 510(k) of the FD&C Act.

    III. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

    IV. Paperwork Reduction Act of 1995

    While this final order contains no collection of information, it does refer to previously approved FDA collections of information. Therefore, clearance by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (PRA) (44 U.S.C. 3501-3521) is not required for this final order. The previously approved collections of information are subject to review by OMB under the PRA. The collections of information in 21 CFR part 822, regarding postmarket surveillance of medical devices, have been approved under OMB control number 0910-0449; the collections of information in 21 CFR part 860, subpart D, regarding De Novo classification have been approved under OMB control number 0910-0844; the collections of information in 21 CFR part 814, subparts A through E, regarding premarket approval, have been approved under OMB control number 0910-0231; the collections of information in part 807, subpart E, regarding premarket notification submissions, have been approved under OMB control number 0910-0120; the collections of information in 21 CFR part 820, regarding quality system regulation, have been approved under OMB control number 0910-0073; and the collections of information in 21 CFR part 801, regarding labeling, have been approved under OMB control number 0910-0485.

    Start List of Subjects

    List of Subjects in 21 CFR Part 878

    • Medical devices
    End List of Subjects

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 878 is amended as follows:

    Start Part

    PART 878—GENERAL AND PLASTIC SURGERY DEVICES

    End Part Start Amendment Part

    1. The authority citation for part 878 continues to read as follows:

    End Amendment Part Start Authority

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

    End Authority Start Amendment Part

    2. Add § 878.4961 to subpart E to read as follows:

    End Amendment Part
    Mountable electromechanical surgical system for transluminal approaches.

    (a) Identification. A mountable electromechanical surgical system for transluminal approaches is a software-controlled, patient bed- and/or operating table-mounted electromechanical surgical system with human/device interfaces that allows a qualified user to perform transluminal endoscopic or laparoscopic surgical procedures using surgical instruments attached to an electromechanical arm.

    (b) Classification. Class II (special controls). The special controls for this device are:

    (1) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.

    (2) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.

    (3) The device manufacturer must conduct and complete post-market surveillance, including an impact of the training program on user learning, behavior, and performance, in accordance with an FDA-agreed-upon protocol. The device manufacturer must submit post-market surveillance reports that contain current data and findings in accordance with the FDA-agreed-upon protocol.

    (4) The device manufacturer must submit a report to FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:

    (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and

    (ii) Identification and rationale for changes made to the device, labeling or device-specific use training program, which did not require submission of a premarket notification during the reporting period.

    (5) Labeling must include:

    (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; Start Printed Page 26996

    (ii) A statement in the labeling that the safety and effectiveness of the device has not been evaluated for outcomes related to the treatment or prevention of cancer, including but not limited to risk reduction, overall survival, disease-free survival and local recurrence, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;

    (iii) Identification of compatible devices;

    (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;

    (v) Reprocessing instructions for reusable components;

    (vi) A shelf life for any sterile components;

    (vii) A description of the device-specific use training program;

    (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and

    (ix) A detailed summary of the post-market surveillance data collected under paragraph (b)(3) of this section and any necessary modifications to the labeling to accurately reflect outcomes based upon the post-market surveillance data collected under paragraph (b)(3) of this section.

    (6) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use.

    (7) Human factors validation testing must be performed and must demonstrate that the user interfaces of the system support safe use in an operating room environment.

    (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:

    (i) Device motion accuracy and precision;

    (ii) System testing;

    (iii) Instrument reliability;

    (iv) Thermal effects on tissue;

    (v) Human-device interface;

    (vi) Mounting hardware testing;

    (vii) Workspace access testing; and

    (viii) Performance testing with compatible devices.

    (9) Software verification, validation, and hazard analysis must be performed. Software documentation must include an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.

    (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.

    (11) Performance data must demonstrate the sterility of all patient-contacting device components.

    (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.

    (13) Performance data must validate the reprocessing instructions for the reusable components of the device.

    (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.

    (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.

    Start Signature

    Dated: April 29, 2022.

    Lauren K. Roth,

    Associate Commissioner for Policy.

    End Signature End Supplemental Information

    Footnotes

    1.  FDA notes that the “ACTION” caption for this final order is styled as “Final amendment; final order,” rather than “Final order.” Beginning in December 2019, this editorial change was made to indicate that the document “amends” the Code of Federal Regulations. The change was made in accordance with the Office of Federal Register's (OFR) interpretations of the Federal Register Act (44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and parts 21 and 22), and the Document Drafting Handbook.

    Back to Citation

    [FR Doc. 2022-09749 Filed 5-5-22; 8:45 am]

    BILLING CODE 4164-01-P

Document Information

Effective Date:
5/6/2022
Published:
05/06/2022
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final amendment; final order.
Document Number:
2022-09749
Dates:
This order is effective May 6, 2022. The classification was applicable on February 26, 2021.
Pages:
26993-26996 (4 pages)
Docket Numbers:
Docket No. FDA-2022-N-0175
Topics:
Medical devices
PDF File:
2022-09749.pdf
Supporting Documents:
» Medical Devices; General and Plastic Surgery Devices; Classification of the Mountable Electromechanical Surgical System for Transluminal
CFR: (1)
21 CFR 878.4961