2022-28173. Medical Devices; Cardiovascular Devices; Classification of the Interventional Cardiovascular Implant Simulation Software Device  

  • Start Preamble

    AGENCY:

    Food and Drug Administration, HHS.

    ACTION:

    Final amendment; final order.

    SUMMARY:

    The Food and Drug Administration (FDA, Agency or we) is classifying the interventional cardiovascular implant simulation software device into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the interventional cardiovascular implant simulation software device's classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices.

    DATES:

    This order is effective December 28, 2022. The classification was applicable on September 8, 2021.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    Judy Ji, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 2543, Silver Spring, MD, 20993-0002, 301-796-6949, Judy.Ji@fda.hhs.gov.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    I. Background

    Upon request, FDA has classified the interventional cardiovascular implant simulation software device as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness. In addition, we believe this action will enhance patients' access to beneficial innovation, in part by placing the device into a lower device class than the automatic class III assignment.

    The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).

    FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate device by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).

    FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act. Section 207 of the Food and Drug Administration Modernization Act of 1997 (Pub. L. 105-115) established the first procedure for De Novo classification. Section 607 of the Food and Drug Administration Safety and Innovation Act (Pub. L. 112-144) modified the De Novo application process by adding a second procedure. A device sponsor may utilize either procedure for De Novo classification.

    Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).

    Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.

    Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device.

    When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for Start Printed Page 79802 future devices of that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C Act). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application to market a substantially equivalent device (see section 513(i) of the FD&C Act, defining “substantial equivalence”). Instead, sponsors can use the less-burdensome 510(k) process, when necessary, to market their device.

    II. De Novo Classification

    On May 7, 2020, FDA received FEops NV's request for De Novo classification of the FEops HEARTguide. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.

    We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.

    Therefore, on September 8, 2021, FDA issued an order to the requester classifying the device into class II. In this final order, FDA is codifying the classification of the device by adding 21 CFR 870.1405.[1] We have named the generic type of device interventional cardiovascular implant simulation software device, and it is identified as a prescription device that provides a computer simulation of an interventional cardiovascular implant device inside a patient's cardiovascular anatomy. It performs computational modeling to predict the interaction of the interventional cardiovascular implant device with the patient-specific anatomical environment.

    FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.

    Table 1—Interventional Cardiovascular Implant Simulation Software Device Risks and Mitigation Measures

    Identified risksMitigation measures
    Inaccurate simulation results leading to selection of suboptimal treatment plan, leading to prolonged procedure time and/or patient injurySoftware verification, validation, and hazard analysis; Computational modeling verification and validation; Performance validation with clinical data; Labeling; and Human factors testing.
    Delayed delivery of results due to software failure or use error, leading to delay of treatmentSoftware verification, validation, and hazard analysis; Human factors testing; and Labeling.
    Failure to properly interpret device results leading to selection of suboptimal treatment plan, leading to prolonged procedure time and/or patient injuryHuman factors testing, and Labeling.

    FDA has determined that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of safety and effectiveness. For a device to fall within this classification, and thus avoid automatic classification in class III, it would have to comply with the special controls named in this final order. The necessary special controls appear in the regulation codified by this order. This device is subject to premarket notification requirements under section 510(k) of the FD&C Act.

    At the time of classification, interventional cardiovascular implant simulation software device is for prescription use only. Prescription devices are exempt from the requirement for adequate directions for use for the layperson under section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR 801.5, as long as the conditions of 21 CFR 801.109 are met.

    III. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

    IV. Paperwork Reduction Act of 1995

    This final order establishes special controls that refer to previously approved collections of information found in other FDA regulations and guidance. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections of information in 21 CFR part 860, subpart D, regarding De Novo classification have been approved under OMB control number 0910-0844; the collections of information in 21 CFR part 814, subparts A through E, regarding premarket approval, have been approved under OMB control number 0910-0231; the collections of information in part 807, subpart E, regarding premarket notification submissions, have been approved under OMB control number 0910-0120; the collections of information in 21 CFR part 820, regarding quality system regulation, have been approved under OMB control number 0910-0073; and the collections of information in 21 CFR parts 801, regarding labeling, have been approved under OMB control number 0910-0485.

    Start List of Subjects

    List of Subjects in 21 CFR Part 870

    • Medical devices
    End List of Subjects

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 870 is amended as follows:

    Start Part

    PART 870—CARDIOVASCULAR DEVICES

    End Part Start Amendment Part

    1. The authority citation for part 870 continues to read as follows:

    End Amendment Part Start Authority

    Start Printed Page 79803 Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

    End Authority Start Amendment Part

    2. Add § 870.1405 to subpart B to read as follows:

    End Amendment Part
    Interventional cardiovascular implant simulation software device.

    (a) Identification. An interventional cardiovascular implant simulation software device is a prescription device that provides a computer simulation of an interventional cardiovascular implant device inside a patient's cardiovascular anatomy. It performs computational modeling to predict the interaction of the interventional cardiovascular implant device with the patient-specific anatomical environment.

    (b) Classification. Class II (special controls). The special controls for this device are:

    (1) Software verification, validation, and hazard analysis, with identification of appropriate mitigations, must be performed, including a full verification and validation of the software according to the predefined software specifications.

    (2) Computational modeling verification and validation activities must be performed to establish the predictive capability of the device for its indications for use.

    (3) Performance validation testing must be provided to demonstrate the accuracy and clinical relevance of the modeling methods for the intended implantation simulations, including the following:

    (i) Computational modeling results must be compared to clinical data supporting the indications for use to demonstrate accuracy and clinical meaningfulness of the simulations;

    (ii) Agreement between computational modeling results and clinical data must be assessed and demonstrated across the full intended operating range ( e.g., full range of patient population, implant device sizes and patient anatomic morphologies). Any selection criteria or limitations of the samples must be described and justified;

    (iii) Endpoints ( e.g., performance goals) and sample sizes established must be justified as to how they were determined and why they are clinically meaningful; and

    (iv) Validation must be performed and controls implemented to characterize and ensure consistency ( i.e., repeatability and reproducibility) of modeling outputs:

    (A) Testing must be performed using multiple qualified operators and using the procedure that will be implemented under anticipated conditions of use; and

    (B) The factors ( e.g., medical imaging dataset, operator) must be identified regarding which were held constant and which were varied during the evaluation, and a description must be provided for the computations and statistical analyses used to evaluate the data.

    (4) Human factors evaluation must be performed to evaluate the ability of the user interface and labeling to allow for intended users to correctly use the device and interpret the provided information.

    (5) Device labeling must be provided that describes the following:

    (i) Warnings that identify anatomy and image acquisition factors that may impact simulation results and provide cautionary guidance for interpretation of the provided simulation results;

    (ii) Device simulation inputs and outputs, and key assumptions made in the simulation and determination of simulated outputs; and

    (iii) The computational modeling performance of the device for presented simulation outputs, and the supporting evidence for this performance.

    Start Signature

    Dated: December 21, 2022.

    Lauren K. Roth,

    Associate Commissioner for Policy.

    End Signature End Supplemental Information

    Footnotes

    1.  FDA notes that the “ACTION” caption for this final order is styled as “Final amendment; final order,” rather than “Final order.” Beginning in December 2019, this editorial change was made to indicate that the document “amends” the Code of Federal Regulations. The change was made in accordance with the Office of Federal Register's (OFR) interpretations of the Federal Register Act (44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and parts 21 and 22), and the Document Drafting Handbook.

    Back to Citation

    [FR Doc. 2022-28173 Filed 12-27-22; 8:45 am]

    BILLING CODE 4164-01-P

Document Information

Effective Date:
12/28/2022
Published:
12/28/2022
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final amendment; final order.
Document Number:
2022-28173
Dates:
This order is effective December 28, 2022. The classification was applicable on September 8, 2021.
Pages:
79801-79803 (3 pages)
Docket Numbers:
Docket No. FDA-2022-N-3185
Topics:
Medical devices
PDF File:
2022-28173.pdf
Supporting Documents:
» Medical Devices; Cardiovascular Devices; Classification of the Interventional Cardiovascular Implant Simulation Software Device
CFR: (1)
21 CFR 870.1405