Code of Federal Regulations (Last Updated: July 5, 2024) |
![]() |
Title 10 - Energy |
![]() |
Chapter I - Nuclear Regulatory Commission |
![]() |
Part 35 - Medical Use of Byproduct Material |
![]() |
Subpart E - Unsealed Byproduct Material - Written Directive Required |
§ 35.220 - Possession of survey instruments.
Latest version.
-
A licensee authorized to use byproduct material for imaging and localization studies shall have in its possession a portable radiation detection survey instrument capable of detecting dose rates over the range of 0.1 millirem per hour to 100 millirem per hour, and a portable radiation measurement survey instrument capable of measuring dose rates over the range 1 millirem per hour to 1000 millirem per hour.