§ 803.17 - What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?  


Latest version.
  • § 803.17 What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?

    If you are a user facility, importer, or manufacturer, you must develop, maintain, and implement written MDR procedures for the following:

    (a) Internal systems that provide for:

    (1) Timely and effective identification, communication, and evaluation of events that may be subject to MDR requirements;

    (2) A standardized review process or procedure for determining when an event meets the criteria for reporting under this part; and

    (3) Timely transmission of complete medical device reports to manufacturers or to us, or to both if required.

    (b) Documentation and recordkeeping requirements for:

    (1) Information that was evaluated to determine if an event was reportable;

    (2) All medical device reports and information submitted to manufacturers and/or us;

    (3) Any information that was evaluated for the purpose of preparing the submission of annual reports; and

    (4) Systems that ensure access to information that facilitates timely followup and inspection by us.