2023-01049. Medical Devices; Ophthalmic Devices; Classification of the Intense Pulsed Light Device for Managing Dry Eye  

  • Start Preamble

    AGENCY:

    Food and Drug Administration, Department of Health and Human Services (HHS).

    ACTION:

    Final amendment; final order.

    SUMMARY:

    The Food and Drug Administration (FDA, Agency, or we) is classifying the intense pulsed light device for managing dry eye into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the intense pulsed light device for managing dry eye's classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices.

    DATES:

    This order is effective January 20, 2023. The classification was applicable on February 23, 2021.

    Start Further Info

    FOR FURTHER INFORMATION CONTACT:

    Arkady Kaplan, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 1568, Silver Spring, MD 20993-0002, 301-796-6365, Morris.Kaplan@fda.hhs.gov.

    End Further Info End Preamble Start Supplemental Information

    SUPPLEMENTARY INFORMATION:

    I. Background

    Upon request, FDA has classified the intense pulsed light device for managing dry eye as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness. In addition, we believe this action will enhance patients' access to beneficial innovation, in part by placing the device into a lower device class than the automatic class III assignment.

    The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).

    FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate device by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).

    FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act. Section 207 of the Food and Drug Administration Modernization Act of 1997 established the first procedure for De Novo classification (Pub. L. 105-115). Section 607 of the Food and Drug Administration Safety and Innovation Act modified the De Novo application process by adding a second procedure (Pub. L. 112-144). A device sponsor may utilize either procedure for De Novo classification.

    Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).

    Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.

    Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under Start Printed Page 3637 section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device.

    When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C Act). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application to market a substantially equivalent device (see section 513(i) of the FD&C Act, defining “substantial equivalence”). Instead, sponsors can use the less-burdensome 510(k) process, when necessary, to market their device.

    II. De Novo Classification

    On April 20, 2020, FDA received Lumenis's request for De Novo classification of the Lumenis Stellar M22. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.

    We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.

    Therefore, on February 23, 2021, FDA issued an order to the requester classifying the device into class II. In this final order, FDA is codifying the classification of the device by adding 21 CFR 886.5201.[1] We have named the generic type of device intense pulsed light device for managing dry eye, and it is identified as a prescription device intended for use in the application of intense pulsed light therapy to the skin. The device is used in patients with dry eye disease due to meibomian gland dysfunction, also known as evaporative dry eye or lipid deficiency dry eye.

    FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.

    Table 1—Intense Pulsed Light Device for Managing Dry Eye Risks and Mitigation Measures

    Identified risksMitigation measures
    Tissue damage due to overheatingThermal safety assessment, Software verification, validation, and hazard analysis, and Labeling.
    Tissue damage or loss of vision due to light radiationClinical performance testing, and Labeling.
    Adverse tissue reactionBiocompatibility evaluation.
    Electrical shock or burnThermal safety assessment, Electrical safety testing, Software verification, validation, and hazard analysis, and Labeling.
    Interference with other devicesElectromagnetic compatibility testing; Software verification, validation, and hazard analysis; and Labeling.
    Pain or discomfortClinical performance testing, and Labeling.
    Failure to mitigate dry eye signs and/or symptomsClinical performance testing, and Labeling.

    FDA has determined that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of safety and effectiveness. For a device to fall within this classification, and thus avoid automatic classification in class III, it would have to comply with the special controls named in this final order. The necessary special controls appear in the regulation codified by this order. This device is subject to premarket notification requirements under section 510(k) of the FD&C Act.

    At the time of classification, intense pulsed light device for managing dry eye is/are for prescription use only. Prescription devices are exempt from the requirement for adequate directions for use for the layperson under section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR 801.5, as long as the conditions of 21 CFR 801.109 are met (referring to 21 U.S.C. 352(f)(1)).

    III. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

    IV. Paperwork Reduction Act of 1995

    This final order establishes special controls that refer to previously approved collections of information found in other FDA regulations and guidance. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections of information in the guidance document “De Novo Classification Process (Evaluation of Automatic Class III Designation)” have been approved under OMB control number 0910-0844; the collections of information in 21 CFR part 814, subparts A through E, regarding premarket approval, have been approved under OMB control number 0910-0231; the collections of information in part 807, subpart E, regarding premarket notification submissions, have been approved under OMB control number 0910-0120; the collections of information in 21 CFR part 820, regarding quality system regulation, have been approved under OMB control number 0910-0073; and the collections of information in 21 CFR part 801, regarding labeling, have been approved under OMB control number 0910-0485.

    Start List of Subjects

    List of Subjects in 21 CFR Part 886

    • Medical devices
    • Ophthalmic goods and services
    End List of Subjects

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner Start Printed Page 3638 of Food and Drugs, 21 CFR part 886 is amended as follows:

    Start Part

    PART 886—OPHTHALMIC DEVICES

    End Part Start Amendment Part

    1. The authority citation for part 886 continues to read as follows:

    End Amendment Part Start Authority

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

    End Authority Start Amendment Part

    2. Add § 886.5201 to subpart F to read as follows:

    End Amendment Part
    Intense pulsed light device for managing dry eye.

    (a) Identification. An intense pulsed light device for managing dry eye is a prescription device intended for use in the application of intense pulsed light therapy to the skin. The device is used in patients with dry eye disease due to meibomian gland dysfunction, also known as evaporative dry eye or lipid deficiency dry eye.

    (b) Classification. Class II (special controls). The special controls for this device are:

    (1) Clinical performance testing must evaluate adverse events and improvement of dry eye signs and symptoms under anticipated conditions of use.

    (2) Thermal safety assessment in a worst-case scenario must be performed to validate temperature safeguards.

    (3) Performance testing must demonstrate electrical safety and electromagnetic compatibility (EMC) of the device in the intended use environment.

    (4) Software verification, validation, and hazard analysis must be performed.

    (5) The patient-contacting components of the device must be demonstrated to be biocompatible.

    (6) Physician and patient labeling must include:

    (i) Device technical parameters;

    (ii) A summary of the clinical performance testing conducted with the device;

    (iii) A description of the intended treatment area location;

    (iv) Warnings and instructions regarding the use of safety-protective eyewear for patient and device operator;

    (v) A description of intense pulse light (IPL) radiation hazards and protection for patient and operator;

    (vi) Instructions for use, including an explanation of all user interface components; and

    (vii) Instructions on how to clean and maintain the device and its components.

    Start Signature

    Dated: January 17, 2023.

    Lauren K. Roth,

    Associate Commissioner for Policy.

    End Signature End Supplemental Information

    Footnotes

    1.  FDA notes that the “ACTION” caption for this final order is styled as “Final amendment; final order,” rather than “Final order.” Beginning in December 2019, this editorial change was made to indicate that the document “amends” the Code of Federal Regulations. The change was made in accordance with the Office of Federal Register's interpretations of the Federal Register Act (44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and parts 21 and 22), and the Document Drafting Handbook.

    Back to Citation

    [FR Doc. 2023-01049 Filed 1-19-23; 8:45 am]

    BILLING CODE 4164-01-P

Document Information

Effective Date:
1/20/2023
Published:
01/20/2023
Department:
Food and Drug Administration
Entry Type:
Rule
Action:
Final amendment; final order.
Document Number:
2023-01049
Dates:
This order is effective January 20, 2023. The classification was applicable on February 23, 2021.
Pages:
3636-3638 (3 pages)
Docket Numbers:
Docket No. FDA-2022-N-3256
Topics:
Medical devices, Ophthalmic goods and services
PDF File:
2023-01049.pdf
Supporting Documents:
» Medical Devices; Ophthalmic Devices; Classification of the Intense Pulsed Light Device for Managing Dry Eye
CFR: (1)
21 CFR 886.5201