§ 86.1323-2007 - Oxides of nitrogen analyzer calibration.
-
Link to an amendment published at 79 FR 23704, Apr. 28, 2014.
This section describes the initial and periodic calibration of the chemiluminescent oxides of nitrogen analyzer.
(a) Prior to introduction into service and at least monthly thereafter, the chemiluminescent oxides of nitrogen analyzer must be checked for NO2 to NO converter efficiency. The Administrator may approve less frequent checks of the converter efficiency. Figure N84-9 is a reference for paragraphs (a) (1) through (11) of this section.
(1) Follow good engineering practices for instrument start-up and operation. Adjust the analyzer to optimize performance.
(2) Zero the oxides of nitrogen analyzer with zero-grade nitrogen.
(3) Connect the outlet of the NOX generator to the sample inlet of the oxides of nitrogen analyzer, which has been set to the most common operating range.
(4) Introduce into the NOX generator-analyzer system an NO-in-nitrogen (N2) mixture with an NO concentration equal to approximately 80 percent of the most common operating range. The NO2 content of the gas mixture shall be less than 5 percent of the NO concentration.
(5) With the oxides of nitrogen analyzer in the NO mode, record the concentration of NO indicated by the analyzer.
(6) Turn on the NOX generator O2 supply and adjust the O2 flow rate so that the NO indicated by the analyzer is about 10 percent less than indicated in paragraph (a)(5) of this section. Record the concentration of NO in this NO + O2 mixture.
(7) Switch the NOX generator to the generation mode and adjust the generation rate so that the NO measured by the analyzer is 20 percent of that measured in paragraph (a)(5) of this section. There must be at least 10 percent unreacted NO at this point. Record the concentration of residual NO.
(8) Switch the oxides of nitrogen analyzer to the NOX mode and measure total NOX. Record this value.
(9) Switch off the NOX generator but maintain gas flow through the system. The oxides of nitrogen analyzer will indicate the NOX in the NO + O2 mixture. Record this value.
(10) Turn off the NOX generator O2 supply. The analyzer will now indicate the NOX in the original NO-in-N2 mixture. This value should be no more than 5 percent above the value indicated in paragraph (a)(4) of this section.
(11) Calculate the efficiency of the NOX converter by substituting the concentrations obtained into the following equation:
a = concentration obtained in paragraph (a)(8) of this section,
b = concentration obtained in paragraph (a)(9) of this section,
c = concentration obtained in paragraph (a)(6) of this section,
d = concentration obtained in paragraph (a)(7) of this section.
(12) If converter efficiency is not greater than 90 percent, repair the analyzer. The repaired analyzer must achieve a converter efficiency greater than 90 percent before the analyzer may be used.
(b) Accuracy. The accuracy at the minimum limit of the NOX analyzer is defined in §86.1338-2007. In general the analyzer's minimum limit shall be the lowest concentration within a given range, in which it has an accuracy of ±2 percent of point.
(c) Initial and periodic calibration. Prior to its introduction into service and monthly thereafter, the chemiluminescent oxides of nitrogen analyzer shall be calibrated on all normally used instrument ranges. Use the same flow rate as when analyzing samples. Proceed as follows:
(1) Adjust analyzer to optimize performance.
(2) Zero the oxides of nitrogen analyzer with zero-grade nitrogen (N2).
(3) (i) Calibrate all operating ranges with a minimum of 9 NO-in-N2 calibration gases (e.g., 10, 20, 30, 40, 50, 60, 70, 80, and 90 percent of that range) and one zero-grade N2 gas. Sound engineering judgment shall dictate appropriate spacing and weighting of the calibration points.
(ii) For each range calibrated, if all deviations from a least-squares best-fit straight line are within ±2 percent of the value at each non-zero data point and within ±0.3 percent of full scale on the zero data point, then concentration values may be calculated using the linear calibration equation for that range. If the specified deviations are exceeded for ranges that have a minimum limit of 1 ppm or greater, then the best-fit non-linear equation that represents the data within these deviations may be used to determine concentration values. For ranges that have a minimum limit less than 1 ppm, only a linear or second order non-linear equation that represents the data within these deviations, may be used to determine concentration values.
(d) Chemiluminescent NOX analyzer interference check (i.e., quench check). Prior to its introduction into service and at least once per year thereafter, the quench check described in this section shall be performed on CLD NOX analyzers. CO2 and water vapor interfere with the response of a CLD by collisional quenching. The combined quench effect at their highest expected concentrations shall not exceed 2 percent.
(1) CO2 quench check procedure: (i) For the procedure described in this paragraph, variations are acceptable provided that they produce equivalent %CO2quench results. Connect a pressure-regulated CO2 span gas to one of the inlets of a three-way valve. Its CO2 concentration should be approximately twice the maximum CO2 concentration expected during testing. The valve must be leak-free, and its wetted parts must be made of a stainless steel or other inert material. Connect a pressure-regulated zero-grade N2 gas to the other inlet of the three-way valve. Connect the single outlet of the valve to the balance-gas port of a properly operating gas divider. Connect a pressure-regulated NO span gas, which has approximately twice the typical NO concentration expected during testing, to the span-port of the gas divider. Configure the gas divider such that nearly equal amounts of the span gas and balance gas are blended with each other. Viscosity corrections shall be applied appropriately to ensure correct mass flow determinations.
(ii) With the CO2 flowing to the balance port and the NO flowing to the span port, measure a stable CO2 concentration from the gas divider's outlet with a properly calibrated NDIR analyzer. Record this concentration in percent (%); this is “%CO2”. This value will be used in the water vapor quench check calculations that are detailed in the following section. After the %CO2 measurement, measure the NO concentration at the gas divider outlet with the CLD analyzer in the NO mode. Record this concentration in ppm; this is “NOCO2”. Then switch the three-way valve such that 100 percent N2 flows to the balance port inlet. Monitor the CO2 concentration of the gas divider's outlet until its concentration stabilizes at zero. Then measure the stable NO concentration from the gas divider's outlet. Record this value in ppm; this is “NON2”. Calculate %CO2quench as follows:
%CO2quench = (1.00−(NOCO2/NON2)) × 100
(2) Water vapor quench check procedure:
(i) For all dry CLD analyzers it must be demonstrated that for the highest expected water vapor concentration (i.e., “%H2Oexp” as calculated later in this section), the water removal technique maintains CLD humidity at less than or equal to 5 gwater/kgdry air (or about 0.008 percent H2O), which is 100% RH at 3.9 °C and 101.3 kPa. This humidity specification is also equivalent to about 25% RH at 25 °C and 101.3 kPa. This may be demonstrated by measuring the temperature at the outlet of a thermal dehumidifier, or by measuring humidity at a point just upstream of the CLD. Humidity of the CLD exhaust might also be measured as long as the only flow into the CLD is the flow out of the dehumidifier.
(ii) For all “wet” CLD analyzers the following water vapor quench check procedure shall be followed. Measure an NO span gas, which has 90% to 100% of the typical NO expected during testing, using the CLD in the NO mode. Record this concentration in ppm; this is “NOdry”. Then bubble the same NO span gas through distilled water in a sealed vessel at 25 °C ±10 °C. This temperature specification imposed to ensure that the H2Ovol calculation (refer to (iii) of this section) returns an accurate result. To prevent subsequent condensation, this temperature must also be less than any temperature that the wetted sample will experience between the sealed vessel's outlet and the CLD. Record the vessel's water temperature in °C; this is “Tsat”. Record the vessel's absolute pressure in kPa; this is “Psat”. Measure the wetted span gas with the CLD, and record this value in ppm; this is “NOwet”.
(iii) Calculations for water quench must consider dilution of the NO span gas with water vapor and scaling of the water vapor concentration to that expected during testing.
(A) Calculate the volume fraction of water vapor in the wetted span gas, as H2Ovol = (exp(3.69−(81.28/Tsat)) + 1.61)/Psat. This calculation approximates some of the thermodynamic properties of water based on the “1995 Formulation for the Thermodynamic Properties of Ordinary Water Substance for General and Scientific Use”, issued by The International Association for the Properties of Water and Steam (IAPWS). However, this approximation should only be used as prescribed in this section because it is an exponential fit that is accurate for data at 25 °C ±10 °C. Then, assuming a diesel fuel atomic hydrogen to carbon ratio of 1.8, and an intake and dilution air humidity of 75 grains (10.71 gwater/kgdry air or 54.13 percent RH at 25 °C and 101.3 kPa),
(B) Calculate the maximum percent water vapor expected during testing; as %H2Oexp = (0.90 × %CO2) + 1.69. %CO2 is the value measured during the %CO2 quench check.
(C) Calculate the expected wet concentration of NO in ppm; as NOexp = NOdry × (1.00−H2Ovol)
(iv) Calculate the percent water vapor quench as:
%H2Oquench = ((NOexp−NOwet)/NOexp) × (%H2Oexp/H2Ovol)
(3) Add the %CO2quench and the %H2Oquench values. Their sum may not exceed the limit set in paragraph (d). If their sum is greater than this limit, then the CLD instrument may not be used to perform testing unless it is repaired. The analyzer must be shown to pass this quench check after the repair before it may be used for testing.