2022-12364. Medical Devices; Cardiovascular Devices; Classification of the Intravascular Bleed Monitor
-
Start Preamble
AGENCY:
Food and Drug Administration, HHS.
ACTION:
Final amendment; final order.
SUMMARY:
The Food and Drug Administration (FDA or we) is classifying the intravascular bleed monitor into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the intravascular bleed monitor's classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices.
DATES:
This order is effective June 8, 2022. The classification was applicable on March 1, 2019.
Start Further InfoFOR FURTHER INFORMATION CONTACT:
LT Stephen Browning, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 2106, Silver Spring, MD 20993-0002, 240-402-5241, Stephen.Browning@fda.hhs.gov.
End Further Info End Preamble Start Supplemental InformationSUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the intravascular bleed monitor as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness. In addition, we believe this action will enhance patients' access to beneficial innovation, in part by placing the device into a lower device class than the automatic class III assignment.
The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate device by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act. Section 207 of the Food and Drug Administration Modernization Act of 1997 established the first procedure for De Novo classification (Pub. L. 105-115). Section 607 of the Food and Drug Administration Safety and Innovation Act modified the De Novo application process by adding a second procedure (Pub. L. 112-144). A device sponsor may utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).
Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device.
When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application to market a substantially equivalent device (see 21 U.S.C. 360c(i), defining “substantial equivalence”). Instead, sponsors can use the less-burdensome 510(k) process, when necessary, to market their device.
II. De Novo Classification
On April 23, 2018, Saranas, Inc. submitted a request for De Novo classification of the Early Bird Bleed Monitoring System. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for Start Printed Page 34778 its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.
Therefore, on March 1, 2019, FDA issued an order to the requester classifying the device into class II. In this final order, FDA is codifying the classification of the device by adding 21 CFR 870.1345.[1] We have named the generic type of device intravascular bleed monitor, and it is identified as a probe, catheter, or catheter introducer that measures changes in bioimpedance and uses an algorithm to detect or monitor progression of potential internal bleeding complications.
FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.
Table 1—Intravascular Bleed Monitor Risks and Mitigation Measures
Identified risks Mitigation measures Adverse tissue reaction Biocompatibility evaluation. Infection Sterilization validation, Pyrogenicity testing, Shelf-life testing, and Labeling. Blood loss, bleeding, hematoma Human factors testing, Labeling, Animal performance testing, and Non-clinical performance testing. Embolization (micro or macro) with transient or permanent ischemia Human factors testing, Labeling, Animal performance testing, and Non-clinical performance testing. Vascular trauma ( i.e., dissection, rupture, perforation, tear, etc.) Human factors testing, Labeling, Animal performance testing, and Non-clinical performance testing. Electrical shock Electrical safety testing. Device failure due to interference with other devices Electromagnetic compatibility (EMC) testing, and Electrical safety testing. Device failure due to software malfunction Software verification, validation, and hazard analysis. FDA has determined that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of safety and effectiveness. In order for a device to fall within this classification, and thus avoid automatic classification in class III, it would have to comply with the special controls named in this final order. The necessary special controls appear in the regulation codified by this order. We encourage sponsors to consult with us if they wish to use a non-animal testing method they believe is suitable, adequate, validated, and feasible. We will consider if such an alternative method could be assessed for equivalency to an animal test method. This device is subject to premarket notification requirements under section 510(k) of the FD&C Act.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to previously approved collections of information found in other FDA regulations and guidance. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections of information in the guidance document “De Novo Classification Process (Evaluation of Automatic Class III Designation)” have been approved under OMB control number 0910-0844; the collections of information in 21 CFR part 814, subparts A through E, regarding premarket approval, have been approved under OMB control number 0910-0231; the collections of information in part 807, subpart E, regarding premarket notification submissions, have been approved under OMB control number 0910-0120; the collections of information in 21 CFR part 820, regarding quality system regulation, have been approved under OMB control number 0910-0073; and the collections of information in 21 CFR part 801, regarding labeling, have been approved under OMB control number 0910-0485.
Start List of SubjectsList of Subjects in 21 CFR Part 870
- Medical devices
Therefore, under the Federal Food, Drug, and Cosmetic Act, 21 CFR part 870 is amended as follows:
Start PartPART 870—CARDIOVASCULAR DEVICES
End Part Start Amendment Part1. The authority citation for part 870 continues to read as follows:
End Amendment Part Start Amendment Part2. Add § 870.1345 to subpart B to read as follows:
End Amendment PartIntravascular bleed monitor.(a) Identification. An intravascular bleed monitor is a probe, catheter, or catheter introducer that measures changes in bioimpedance and uses an algorithm to detect or monitor progression of potential internal bleeding complications.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) In vivo animal performance testing must demonstrate that the device performs as intended under anticipated conditions of use and evaluate the following:
(i) Device performance characteristics;
(ii) Adverse effects, including gross necropsy and histopathology; and
(iii) Device usability, including device preparation, device handling, and user interface.
(2) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following Start Printed Page 34779 performance characteristics must be tested:
(i) Tensile testing of joints and materials;
(ii) Mechanical integrity testing;
(iii) Friction testing;
(iv) Flush testing;
(v) Air leakage and liquid leakage testing;
(vi) Latching and unlatching testing;
(vii) Kink and bend testing;
(viii) Insertion force testing;
(ix) Torque testing;
(x) Corrosion testing; and
(xi) Dimensional tolerance testing.
(3) Performance data must support the sterility and pyrogenicity of the device components intended to be provided sterile.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(5) The patient contacting components of the device must be demonstrated to be biocompatible.
(6) Software verification, validation, and hazard analysis must be performed.
(7) Performance data must demonstrate electromagnetic compatibility (EMC), electrical safety, thermal safety, and mechanical safety.
(8) Human factors performance evaluation must demonstrate that the user can correctly use the device, based solely on reading the directions for use.
(9) Labeling must include:
(i) Instructions for use;
(ii) A shelf life and storage conditions;
(iii) Compatible procedures;
(iv) A sizing table; and
(v) Quantification of blood detected.
Dated: June 2, 2022.
Lauren K. Roth,
Associate Commissioner for Policy.
Footnotes
1. FDA notes that the “ACTION” caption for this final order is styled as “Final amendment; final order,” rather than “Final order.” Beginning in December 2019, this editorial change was made to indicate that the document “amends” the Code of Federal Regulations. The change was made in accordance with the Office of Federal Register's (OFR) interpretations of the Federal Register Act (44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and parts 21 and 22), and the Document Drafting Handbook.
Back to Citation[FR Doc. 2022-12364 Filed 6-7-22; 8:45 am]
BILLING CODE 4164-01-P
Document Information
- Effective Date:
- 6/8/2022
- Published:
- 06/08/2022
- Department:
- Food and Drug Administration
- Entry Type:
- Rule
- Action:
- Final amendment; final order.
- Document Number:
- 2022-12364
- Dates:
- This order is effective June 8, 2022. The classification was applicable on March 1, 2019.
- Pages:
- 34777-34779 (3 pages)
- Docket Numbers:
- Docket No. FDA-2021-N-0600
- Topics:
- Medical devices
- PDF File:
- 2022-12364.pdf
- Supporting Documents:
- » Medical Devices; Cardiovascular Devices; Classification of the Intravascular Bleed Monitor
- CFR: (1)
- 21 CFR 870.1345